The Design and Analysis of a Double Transtibial Composite Prosthesis and the Effect of Lateral Movement Loads by Max A. Willer An Engineering Project Submitted to the Graduate Faculty of Rensselaer Polytechnic Institute in Partial Fulfillment of the Requirements for the degree of Master of Engineering in Mechanical Engineering Approved: Ernesto Gutierrez-Miravete, Project Adviser Rensselaer Polytechnic Institute Hartford, CT December, 2014 i CONTENTS The Design and Analysis of a Double Transtibial Composite Prosthesis and the Effect of Lateral Movement Loads i LIST OF TABLES iii LIST OF FIGURES iv TERMINOLOGY / LIST OF SYMBOLS / ACRONYMS v ACKNOWLEDGEMENTS vii ABSTRACT viii 1. Introduction 1 1.1. Background 1 1.2. Challenges of this Study 3 2. Theory and Methodology 7 2.1. Reverse Engineering the Flex Foot Design 7 2.2. Lateral Forces, Friction, and Angle of Attack 11 2.3. Analytic Method Summary 13 3. Results and Discussion 17 3.1. Model Setup 17 3.2. Lateral Movement 21 4. Remarks and Conclusion 25 5. References 30 6. Appendices 32 6.1. Graphical Data ii LIST OF TABLES Table 2-1: Properties of the Flex Foot 8 Table 2-2: Consistent Units for Analysis 8 Table 2-3: Coefficients of Friction Equations 10 Table 3-1: 0/90 Ply Orientation, Stress and Stiffness vs Load 18 Table 3-2: 0/-45/90/45 Stiffness at 1500 N 20 Table 3-3: Summary of Baseline Stresses and TSAIW 24 Table 4-1: Deflection and Modulus Response of Selected Materials 26 Table 4-2: Model Dimension Lengths 27 Table 4-3: Weight of Prosthetic Using Selected Materials 27 iii LIST OF FIGURES Figure 1-1: History of Leg Prostheses 1 Figure 1-2: Official Rendering of the Flex Foot Prosthetic 2 Figure 1-3: Ankle Power Response Curve 3 Figure 1-4: Diagram of Tennis Sole 4 Figure 1-5: Definitions of ankle eversion and inversion 5 Figure 1-6: Connection bolts of Flex Foot bolts 5 Figure 1-7: Ankle eversion in tennis lateral motion 6 Figure 2-1: Geometric Analysis of Flex Foot Prosthetic 7 Figure 2-2: Dimensions of Baseline Prosthetic 9 Figure 2-3: Comparison of Baseline Model to Original 9 Figure 2-4: Plot of Friction Equations 11 Figure 2-5: Maximum Force and Angle of Attack 12 Figure 2-6: Loads and Constraints 13 Figure 2-7: Assembly with Fixed Reference Point Shown 14 Figure 2-8: Assembly with Contact Interaction Method Shown 14 Figure 2-9: Tangential Behavior 15 Figure 2-10: Profile of Assembly 15 Figure 3-1: Mesh Experiment 17 Figure 3-2: Visualization of Linear Stiffness vs Load 19 Figure 3-3: Various Ply Stiffness at 1500 N 19 Figure 3-4: 0/-45/90/45 Stiffness at 1500 N, Various Plies 20 Figure 3-5: [0/90] Vertical Stress and TSAIW 21 Figure 3-6: [0/90] 45 Degree Stress and TSAIW 22 Figure 3-7: [0/45/90/-45] Vertical Stress and TSAIW 23 Figure 3-8: [0/45/90/-45] 45 Degree Stress and TSAIW 23 Figure 4-1: Modulus Response of Selected Materials 26 Figure 4-2: Dimensions of Baseline Prosthetic 27 Figure 4-3: Weight of Prosthetic Using Selected Materials 28 iv TERMINOLOGY / LIST OF SYMBOLS / ACRONYMS Transtibial – occurring across or involving the tibia Abduction/Adduction – Ankle rotation around the shin axis Plantar Flexion / Dorsiflexion – Ankle rotation about the ankle joint axis Inversion / Eversion = Ankle rotation about the foot axis FEA – Finite Element Analysis FBD – Free Body Diagram 2D – 2 Dimensions E – Modulus of Elasticity (Msi) G – Modulus of Rigidity (Msi) ν – Poisson’s Ratio ρ – Density (lbf/in3) tp – Ply Thickness (in) YS – Yield Strength (ksi) UTS – Ultimate Tensile Strength (ksi) σ1t – Tensile strength in the 1 (longitudinal) direction (ksi) σ1c – Compressive strength in the 1 (longitudinal) direction (ksi) σ2t – Tensile strength in the 2 (transverse) direction (ksi) σ2c – Tensile strength in the 2 (transverse) direction (ksi) τ12f – Shear Strength (ksi) [Orientation number of plies]S – Laminate Layup which is characterized by ply orientation, number of plies and symmetry about the mid-plane (S, if applicable). v Abaqus – Computer Software used to perform modeling and FEA Isotropic – Same properties in all directions Orthotropic – Different properties in different directions TSAIW – An abbreviation for Tsai-Wu Abaqus uses vi ACKNOWLEDGMENTS I would first like to thank my Project Adviser, Professor Ernesto, for all of the help he has given me and for his patience. I would like to thank Professor Hufner for his help with composites analysis. I would also like to thank all those who are working in the field of prosthetics. I would also like to thank the faculty of Rensselaer Polytechnic Institute at Groton for their support throughout the Masters of Engineering program. vii ABSTRACT The purpose of this project is to evaluate the performance of a prosthetic device similar to the Flex Foot Cheetah design used by Olympian sprinters. This transtibial prosthetic has been evaluated before to determine if it offers more mechanical advantage in a forward sprint to a runner than does an anatomical leg, ankle and foot. This paper determines the properties of such a prosthetic and then analyzes what changes, if any, would benefit the prosthetic wearer in sports that require lateral movement in addition to forward sprinting. First a model was created by gathering all non-proprietary information about the Flex Foot Cheetah model. Dimensions were ascertained by image analysis and from given information by previous papers. The dimensioned model was then analyzed with two ply orientation types to determine orientation, number, and thickness of composite plies to have a similar baseline mechanical performance as the Flex Foot. These models were then analyzed for stresses in the lateral direction at the maximum amount of force seen by a tennis player changing direction. Modifications are proposed to the Flex Foot design to accommodate these forces, such as resizing of the equivalent ankle area and widening the contact foot for additional stability and traction. viii 1. Introduction Background: The use of prosthetic limbs as a medical solution for lost limb functionality has been in practice for most of human history, from wooden crutches to wooden legs all the way up to the articulating hands of the present and future. Those who use prosthetic limbs have long been perceived as at a disadvantage in physical fitness and sport to those with full use of their limbs. This perception however is on the precipice of change as advanced material technology and fabrication techniques are applied towards the design and manufacturing of prosthetic devices. In fact, some prosthetic limbs have become so advanced, that they have caused controversy in the sporting world for giving an unfair advantage over natural human limbs. Figure 1-1: History of Leg Prostheses, Reference 1 Oscar Pistorius is a South African sprint runner and a double transtibial amputee, which means the prosthetics he wears replace legs and feet that have been amputated below the knee. In 2007 he gained attention by participating in able-bodied international sprinting competitions which resulted in the International Association of Athletics Federations (IAAF) banning "any technical device that incorporates springs, wheels or any other element that provides a user with an advantage over another athlete not using such a device." (Reference 2). This led to several studies as to whether there was indeed an advantage in 1 using the Flex Foot prostheses that Oscar Pistorius has become famous for using. One of the most famous of theses is the Bruggemann study, a series of tests published in 2008 that reported that Pistorius uses twenty-five percent less energy expenditure than ablebodied runners with use of their lower legs, ankles, and feet. (Reference 3). A series of other studies followed showing that Pistorius used less metabolic energy and foot to ground force than able-bodied runners, further pointing toward the mechanical advantage of the Flex Foot prosthesis over conventional running (Reference 4). Regardless of the controversy in the sport, the fact is that prosthetic devices have become so advanced as to give their wearers equal or higher functionality than those with natural limbs. Given that prosthetics are becoming so advanced that they give their wearers an ad- Figure 1-2: Official Rendering of the Flex Foot Prosthetic, Reference 5 vantage over those with human limbs, it is natural to desire to expand the practical uses to other uses, including other sports. The purpose of the Flex Foot design, as well as many other leg prosthetic designs historically, is to emulate the spring action of the lower leg as a runner is moving forward. While the design works exceedingly well for that one function, it lacks the specialization to be useful for sports that include movements outside of the forward direction. The Flex Foot’s narrow design gives an effective 2 running spring stiffness and an aerodynamic profile for speed, but in sports where the athlete needs traction and ankle mobility for lateral movements, the Flex Foot design may come up short. Thus it is this paper’s purpose to analyze the Flex Foot design, and modify it to properly suit leg movements in sports that include lateral motion. Challenges of this Study: The Flex Foot Cheetah by description is a laid up composite laminate. Most challengingly it is a proprietary design, which means to analyze it will require reverse engineering it. This will mean estimating the dimensions, materials, and composite layup structure to arrive at appropriate baseline model from which to design a lateral force transtibial prosthesis. To accomplish this, this paper uses the data collected from the Bruggemann study, which illustrates the static and dynamic responses of the Flex Foot Cheetah. Getting within a reasonable degree of accuracy to this model will help ensure a functional lateral movement prosthesis. Figure 1-3: Ankle Power Response Curve, Reference 3 The next challenge of this study will be to accurately model the ground reaction forces seen in lateral movement of the human body. Side to side movement sees greater forces in the human leg than running forward, since the foot and leg has to support the weight of the body as well as stop its horizontal momentum in an appropriate amount of time in 3 the direction of gravity and in the opposite direction of the body’s momentum. Depending on the playing surface, different sports solve withstanding lateral moving force with various shoe surfaces. Field sports like soccer and American football give their athletes traction with cleats while ice sports give their athletes steel blades. However, court sports like basketball, racquetball, and tennis require their athletes to simply use the traction of a flat sole for movement. Luckily, studies have been done on shoe materials and patterns best suited for the lateral movements of tennis. These studies have detailed data on the force required of the athletes interface with the ground as well as materials used. Thus the next challenge will be to use these studies to determine the contact surface area of the new lateral movement prosthetic as well as the lateral force against the ground it can resist. After modelling the ground forces, the dynamic response of the baseline Flex Foot de- Figure 1-4: Diagram of Tennis Sole, Refer- ence 6 sign will be studied to examine the critical points to redesign to allow the maximum effectiveness for lateral movement. The design will be iterated keeping in mind the following main requirement: A large majority of the weight of the prosthetic design is in the connecting assembly to the remaining lower leg. Thus the weight of the mechanical spring part of the prosthetic should be kept at a minimum. 4 It is for this requirement that the Flex Foot original design utilizes a composite layup with strengthening carbon fiber laminates, and this study will use this line of reasoning during material selection. With the main requirement in mind, the first iterations will likely expand the under-foot contact area in order to meet the amount of friction force required by the lateral movement of tennis or other court sports. Just like the standard tennis shoe is wider at the base than the standard running shoe to accommodate lateral forces, so will the new prosthetic device need to be wider to grant it the traction to resist lateral movement. Figure 1-6: Definitions of an- Figure 1-5: Connection kle eversion and inversion, bolts of Flex Foot pros- Reference 7 thetic, Reference 8 The next iterations will lead to fine tuning the ankle support axis that contributes towards inversion and eversion of the foot as related to the shin axis. This twisting action is important to get tuned correctly. Too loose and the under-foot contact area will overturn on its leading edge causing the leading edge to dig in and possibly lead to injury or send the player off balance. If the twisting action support is too stiff, the design will not allow the full under-foot contact area to make contact with the ground and provide full traction. Additional iterations beyond the scope of this project could include fine tuning the shin axis of the prosthetic device. The current Flex Foot design only utilizes two bolts to 5 sandwich the mechanical spring to the leg connecting section of the prosthetic. These iterations will likely beef up the lateral moment of inertia of the mechanical spring section and perhaps iterate on the bolt connecting pattern. There are many factors that go into a prosthetic limb design, just as there are many factors as to way our limbs evolved the way they did. This project attempts to be an open study into the factors that would allow this type of prosthetic to be used in sports that require more movement than just in a straight line. Figure 1-7: Ankle eversion in tennis lateral motion 6 2. Theory and Methodology 2.1 Reverse Engineering the Flex Foot Design The first step in redesigning the Flex Foot design is understanding the original design. Unfortunately, the original design of the Flex Foot is proprietary to Ossur, the Icelandic prosthetic design company that created the Flex Foot prosthetic. However, by using the resources of the Bruggeman study, Reference 3, we can work backwards using the dimensional and material properties reported to come to an educated guess at the original design. Figure 5 of Reference 3 displays the power through time stance phase via percentage of the time elapsed as its argument the Flex Foot design allows a runner to be more energy efficient than a runner with natural limbs. Table 2-1 of this paper uses the same basic mechanical properties Reference 3 uses to make its analysis. 2.1.1 Dimensions Using Figure 2-1, we can determine the neutral position geometric dimensions of the Flex Foot prosthetics. We take the figure and determine the pixel coordinates of significant points in the design. Then using Table 3 of Reference 3, we see the neutral height of the Flex Foot is 46 cm, we can determine the relationship between the pixels of the image and length dimensions. Figure 2-1: Geometric Analysis of Flex Foot Prosthetic 7 Table 2-1: Properties of the Flex Foot, Reference 3 𝐻𝑒𝑖𝑔ℎ𝑡, 𝑝𝑖𝑥𝑒𝑙 ℎ𝑒𝑖𝑔ℎ𝑡, 𝑐𝑜𝑛𝑣𝑒𝑟𝑠𝑖𝑜𝑛 𝑟𝑎𝑡𝑖𝑜, ℎ = 0.46𝑚 ℎ𝑝 = 584𝑝𝑥 − 316𝑝𝑥 = 268𝑝𝑥 𝑐= 0.46𝑚 = 0.0017164 𝑚⁄𝑝𝑥 268𝑝𝑥 (2.1) (2.2) (2.3) Knowing the conversion rate, we can discover the rest of the significant dimensions of the Flex Foot, and input them into the model. This study uses Abaqus analysis software, for which there are no internal units analysis. Thus we need to pick a consistent units system, indicated below. Table 2-2: Consistent Units for Analysis 8 Figure 2-2: Dimensions of Baseline Prosthetic In this way we guess at the dimensions of the Flex Foot. The below figure demonstrates the baseline model’s approximation to the original design. While this appears satisfactory for the geometry, we must next iterate on the possible composite material properties to accurately represent the spring response of the prosthetic. Figure 2-3: Comparison of Baseline Model to Original 𝐸𝑥𝑡𝑟𝑢𝑠𝑖𝑜𝑛 𝑤𝑖𝑑𝑡ℎ, 9 𝑤 = 0.07𝑚 2.1.2 Material Having gotten reasonably close to the Flex Foot geometric design, we now attempt to get reasonably close to the linear stiffness measured by Table 2-1. 𝐿𝑖𝑛𝑒𝑎𝑟 𝑆𝑡𝑖𝑓𝑓𝑛𝑒𝑠𝑠, 𝐸 = 38.7 𝑘𝑁⁄𝑚 As mentioned in numerous sources, as well as the product manufacturer’s description, the prosthetic is made out of a carbon-fiber composite lay-up. Within the industry this could refer to any number for material properties, laminate orientations, and intermediary materials. In this study, we use Reference 9 as an initial guide to the material properties and common orientations of performance composites. Composites can be used in a variety of applications with various requirements for stiffness and strength. Some applications require the high stiffness and strength that a carbon-fiber Nomex honeycomb sandwich provides, while other applications require the lower stiffness spring-type profile of many thin laminate plies. In this case the closest mechanical analogue is that of a leaf-spring, which in most light applications such as this one, uses the thin laminate ply style of composite design. Since the thicknesses of the plies can be of various thickness and. The above attempt at ascertaining the geometry of the Flex Foot worked well for the large geometries like the length and width but not as well for the thickness which is significantly smaller and more difficult to ascertain. In addition, the thickness of the plies used must be guessed at. Thus the initial material section will iterate on the ply thickness, orientations, and number of plies. A load will be placed at the top of the Flex Foot model, along with constraints to ensure the load and displacement only occurs in the vertical direction as seen in Figure 2-4. The displacement of the top of the model with then be calculated, resulting in a linear stiffness to compare to the Bruggeman study. 10 2.2 Lateral Forces, Friction, and Angle of Attack Table 2-3: Coefficients of Friction Equations, Reference 6 Figure 2-4: Plot of Friction Equations, Reference 6 It is important at this point to research the loads that the average tennis player impacts on the ground during side-to-side lateral movement. Luckily the Clarke paper that concentrated on shoe patterns and traction effects summarized it nicely. According to the Clarke Paper: “In order to mechanically test under conditions that best represent real-life play, ground reaction forces from a study conducted by Damm et al. were examined to understand the forces exerted by a tennis player during shoesurface interactions. Damm et al. measured three-dimensional ground reaction forces of tennis players performing a side jump followed by a pushoff movement on an acrylic hard court surface. The mean peak normal force found during the initial impact phase of the movement was found to be approximately 1,150 N, and during the phase of forefoot push-off the normal force reduced to relatively constant value of approximately 650 N.” 11 The Clarke paper summarizes its material results into a trend line graph of the shoe traction force to the normal force of the foot on the ground of a relatively rough tennis court surface: 𝐹𝑡 = 0.84𝐹𝑁 + 206.62 (2.4) 𝑅 2 = 0.98 (2.5) 𝐹𝑡 = 0.84(1150 𝑁) + 206.62 (2.6) 𝐹𝑡 = 1172.62 𝑁 (2.7) Figure 2-5: Maximum Force and Angle of Attack Maximum Angle of Attack: 1150 𝑁 𝜃 = arctan ( ) = 44.44 𝑑𝑒𝑔𝑟𝑒𝑒𝑠 1172.62 (2.8) Maximum Axial Force 𝐹 = √11502 + 1172.622 = 1642.4 𝑁 (2.9) This equation is put into analysis software as the governing friction equation in the dynamic analysis of the prosthetic coming into contact with the ground. The peak normal force of 1642.4 N applied downward axially along the shin at a 44.44 degree angle to the ground as described in the Clarke paper will be used as the analysis goal for what the prosthetic design can handle laterally. 12 2.3 Analytic Method Summary 2.3.1 Mesh We first take the baseline design and attempt at optimizing the mesh of the part. In this first part we don’t worry about the accuracy of the linear stiffness result, we just adjust the mesh fineness up until an acceptable point. 2.3.2 Boundary Conditions We simulate the loading on the prosthetic by applying a vertical load on the top set of nodes, this can be seen in Figure 2-6. These nodes are constrained in all directions but the vertical. This simulates the bolting pattern on the prosthetic that constrains it to the axis of the shin. Figure 2-6: Loads and Constraints The ground is constrained via tying it to a Reference Point in the Abaqus Interaction constructor. This ensures the ground is fixed in all degrees of freedom as seen in Figure 2-7. 13 Figure 2-7: Assembly with Fixed Reference Point Shown The interaction between the prosthetic model and the ground model is governed by a Kinematic Contact method between the lower face of the horizontal portion of the prosthetic and the upper face of the ground model. This sets up the contact rules for the modelling and can be seen in Figure 2-8. Figure 2-8: Assembly with Contact Interaction Method Shown The Interaction is also governed by an interaction property that defines the tangential behavior, in this case the sliding friction. We input the friction coefficient from Equation 2.4 as the governing interaction property as can be seen in Figure 2-9. Finally, we place the prosthetic model slightly above the ground model. This ensures that in the initial steps of the analysis, nodes on the prosthetic model cannot be construed 14 as being below the ground model, which would skew the results. An image of this is shown in Figure 2-10. Figure 2-9: Tangential Behavior Figure 2-10: Profile of Assembly 15 2.3.3 Number of Plies We then attempt to come to an accurate representation of the linear stiffness result of Reference 3. We do this simply be increasing the number of plies of the composite. We begin with a standard [0/90] orientation. This lets half of the fiber laminate assist with the linear stiffness in the longitudinal tension mode where composite laminate is strongest. 2.3.4 Ply Orientation After we arrive at a reasonable approximation of the linear stiffness, we determine the effect of the ply orientation. Various ply orientations will have different effects when we analyze lateral forces on the prosthetic. Before we test these effects, we need to ensure that the ply orientation types we have also match up with the measured linear stiffness of the baseline design. 2.3.5 Lateral Forces We then apply the lateral forces on the prosthetic and observe the stresses and deformations of the laminate. This will be achieved by applying a force downward on the prosthetic design and tilting the ground at increasing angle. This will reveal the stresses and deformations on the prosthetic design that will form the basis of the design iterations. 2.3.6 Design Iterations It is the hypothesis of this paper that the prosthetic design will not be adequately resistant to the maximum lateral forces applied at maximum tennis shoes traction. For a robust and accurate ankle simulation, this paper hypothesizes that the design will have to be thickened in the abduction/adduction axis to prevent excess twist in addition to being tweaked in the inversion/eversion axis for proper ground contact and stability. 16 3. Results and Discussion 3.1 Model Setup 3.1.1 Mesh Before attempting to discern the material properties of the Flex Foot, we first make sure that the baseline design’s mesh is fine enough to make accurate calculations. We begin with a coarse mesh, a 200 N load downward axially on the shin, a 20 ply [0 / 90] composite and a level ground. Maximum von Mises Stress (Pa e8) Maximum von Mises Stress 3 2.5 2 1.5 Maximum von Mises Stress 1 0.5 0 0 0.02 0.04 0.06 0.08 Mesh Size (m) Figure 3-1: Mesh Experiment As shown in Figure 3-1, adjusting the mesh size leads to slightly erratic maximum stresses. The most stable of the meshes was around 0.03 to 0.02. Thus we will be using a mesh of 0.03 to enable quicker compiling of the stress data when the plies of the laminate increase. 3.1.2 Number of Plies As stated in the methodology section, the thickness of the composite prosthetic design is difficult to ascertain via visual inspection. In addition the thickness directly correlates with the mechanical performance of the prosthetic, so estimating the number of plies is of the utmost importance. 17 The first step of the material methodology was to take the assumed geometry definition of the baseline model and see what material properties let us arrive at the Bruggemann study linear elasticity measurement of the Flex Foot device stated in Figure 2-1. 𝐿𝑖𝑛𝑒𝑎𝑟 𝑆𝑡𝑖𝑓𝑓𝑛𝑒𝑠𝑠, 𝐸 = 38.7 𝑘𝑁⁄𝑚 We begin with our baseline model as stated in section 3.1.1. We then add successive increasing loads in the shin axial direction and measure the total deflection. As we can see, the original 20 ply model has nowhere near the linear stiffness required by the Bruggeman study. We also see with this 20 ply model that the linear stiffness is not a constant as the load increases. 0/90 Vertical Load (N) 20 Ply Displacement (m) 20 Ply Linear Stiffness (Pa) 24 Ply Displacement (m) 24 Ply Linear Stiffness (Pa) 30 Ply Displacement (m) 30 Ply Linear Stiffness (Pa) 40 Ply Displacement (m) 40 Ply Linear Stiffness (Pa) 50 Ply Displacement (m) 50 Ply Linear Stiffness (Pa) 60 Ply Displacement (m) 60 Ply Linear Stiffness (Pa) 100 0.0280 3571 0.0240 4167 0.0175 5714 0.0125 8000 0.0095 10526 0.0080 12500 200 0.0400 5000 0.0310 6452 0.0210 9524 0.0190 10526 0.0110 18182 0.0081 24691 300 0.0540 5556 0.0375 8000 0.0250 12000 0.0200 15000 0.0120 25000 0.0093 32258 400 0.0600 6667 0.0440 9091 0.0400 10000 0.0240 16667 0.0130 30769 0.0100 40000 500 0.0720 6944 0.0525 9524 0.0450 11111 0.0275 18182 0.0150 33333 0.0108 46296 600 0.0770 7792 0.0560 10714 0.0550 10909 0.0325 18462 0.0170 35294 0.0115 52174 700 0.0800 8750 0.0600 11667 0.0400 17500 0.0240 29167 0.0175 40000 0.0120 58333 800 0.0910 8791 0.0660 12121 0.0440 18182 0.0260 30769 0.0180 44444 0.0125 64000 Bruggeman Linear Stiffness (Pa) 38700 38700 38700 38700 38700 38700 38700 38700 Table 3-1: 0/90 Ply Orientation, Stress and Stiffness vs Load We can visualize this data in a chart and compare it to the ideal stiffness value from the Bruggeman study. The plots of Figure 3-2 show how the variation in linear stiffness at different load increases as the number of plies increases. While in a simple beam this behavior would seem odd, due to the complex shape of the prosthetic, it can be expected that the prosthetic would not have a simple linear stiffness as the load is supported by the different sections. We now repeat the Bruggeman test in paragraph 2.2 of reference 3 which tests the prosthetic device on a “material testing machine T1-FR020TN.A50 (Zwick GmbH & Co, Ulm, Germany)” [3] at a controlled deflection up to a load of 1500 N. In this way they come to the estimation of linear stiffness. The results of this test simulation are in Figure 3-3. 18 Figure 3-2: Visualization of Linear Stiffness vs Load As we can see from the data in Figure 3-3, at 1500 N, a prosthetic design incorporating 40 plies comes closest to the Bruggeman test results of 38.7 kN/m linear stiffness at a loading of 1500 N. At a laminate thickness of 0.00025 as estimated by reference 9, this leaves us with a prosthetic thickness of 1 cm. From the limited specifications and images given, this thickness and linear stiffness best estimates the physical and mechanical properties of the Flex Foot baseline design. Figure 3-3: Various Ply Stiffness at 1500 N 19 3.1.3 Orientation of Plies The 0/90 orientation of plies is certainly not the only orientation option available for carbon fiber composites, though it is one of the more common. The 0/90 orientation is simple to manufacture, and well suited for an environment where the laminates are under a simple tensile load from one direction. However, since this study aims to test a composite prosthetic in more than one direction and brings twist of the composite into the equation, other laminate orientations that can accommodate twist were considered, the most common of which being a simple 0/-45/90/-45 degree pattern. This pattern, according to reference 9, has less linear stiffness in the two main longitudinal and transverse directions, but has a higher stiffness for in-plane shear. Thus we also test a material with the 0/-45/90/-45 degree pattern and see how many plies it must have to meet the Bruggeman study linear stiffness. As can be seen in Table 3-2 and Figure 3-4, the lower longitudinal and transverse moduli correlate to a higher number of plies to maintain the same linear stiffness at 1500 N. Load 1500 N Ply Test 20 28 36 40 48 1500 N Vertical Displacement (m) 0.1250 0.0830 0.0620 0.0450 0.0335 Linear Stiffness (Pa) 12000 18072 24194 33333 44776 Bruggeman Stiffness (Pa) 38700 38700 38700 38700 38700 Table 3-2: 0/-45/90/45 Stiffness at 1500 N Figure 3-4: 0/-45/90/45 Stiffness at 1500 N, Various Plies 20 As you can see, the [0/-45/90/45] ply pattern would need more plies to fully meet the Bruggeman Linear Stiffness value. This would lead to a thicker prosthetic, which would likely be heavier and worse off in the forward sprinting mode. In the next section we’ll see how it performs in a lateral movement mode. 3.2 Lateral Movement 3.2.1 Stresses and Tsai Wu Now that a general ply structure has been analyzed, we now look to see what the effects of lateral movement are on the stresses and failure criteria. In the vertical position, the baseline prosthetic has a maximum stress of 2.815e8 and a maximum Tsai Wu failure criterion of 0.47 as seen in Figure 3-5. The stress concentration is mainly in the equivalent of the ankle region. Figure 3-5: [0/90]Vertical Stress and TSAIW 21 In the 45 degree model, stresses are still concentrated around the ankle area, but asymmetric stresses on the sole of the prosthetic become apparent. The maximum von Mises stress decreases while the Tsai Wu failure maximum increases. This is likely due to the increase amount of torsion in the inversion/eversion axis. In this model, stresses and Tsai Wu criteria still don’t approach any failure point as designed. See Figure 3-6. This is important to note as we iterate the design. 3.2.2 Orientation of Plies Next we look at how the ply orientation affects the maximum stresses and Tsai Wu criterion. The additional number of plies will reduce the maximum stresses in the prosthetic and the additional diagonally oriented plies in the [0/45/90/-45] ply structure will likely reduce the maximum Tsai Wu failure criterion as well. stresses seen in the prosthetic. As Figure 3-6: [0/90] 45 Degree Stress and TSAIW can be seen in the 90 degree orientation in Figure 3-7, these intuitions hold true. Note that boundary conditions have slightly skewed the Tsai Wu criterion scale. The largest Tsai Wu values are seen in the ankle area with a value of 0.3016. Figure 3-8 shows the results 22 of the [0/45/90/-45] ply structure in the 45 degree orientation. Again, this orientation generates higher Tsai Wu failure numbers since a large amount of torsional shear is present in the ankle area in this location. Figure 3-7: [0/45/90/-45] Vertical Stress and TSAIW Figure 3-8: [0/45/90/-45] 45 Degree Stress and TSAIW 23 In addition to being located in the ankle region, the Tsai Wu values are particularly high for this laminate configuration. This high failure value means the laminate is quite close to delaminating due to the high shear stresses present in the ankle area. So while the [0/45/90/-45] laminate type was significantly better at stress distribution in the vertical direction, the [0/90] laminate behaves better in the 45 degree orientation. This is the behavior we want to see from a lateral movement prosthetic. Laminate [0/90] [0/45/90/-45] Vertical 45 Degree Stress (Pa) TSAIW Stress (Pa) TSAIW 2.82E+08 0.471 2.71E+08 0.64 1.09E+08 0.3016 1.71E+08 0.925 Table 3-3: Summary of Baseline Stresses and TSAIW 24 4. Remarks and Conclusion 4.1 Prosthetic Design The scope of this project began as taking a simple prosthetic limb design and redesigning it to fully model the mechanics of an ankle for sports with lateral movement. As the project progressed, it became apparent that even emulating the original design of the Flex Foot was a challenge. Composite design is many faceted, with millions of permutations of structures, materials, fibers, epoxies, and laminate patterns. A sports composite such as a golf club, racing fairing, bicycle helmet, or prosthetic already have years of design work behind them, which is why composites are even now just coming out to revolutionize the sporting industry. Thus taking a design that has already been tested and created and attempting to make even a close replica is a difficult task. Fortunately studies existed for this particular composite, particularly the Bruggemann study which gave some dimensions, spring constants, and energy curves with which to get a baseline design. 4.2 Composite Choices Only two basic composite choices were considered for this study. Given more scope, more composite patterns and materials could be considered, not to mention use of metals and other plastics. The [0/90] composite configuration was a bit of an all-purpose composite pattern. While this paper had the hypothesis early on that the [0/45/90/-45] composite pattern would have a better response to the lateral force considerations, it turned out that the [0/90] was further from failure in the weakest portion of the prosthetic, the ankle. The [0/90] configuration also required less layers to reach the adequate spring constant proposed by the Bruggemann paper. This spring constant was a target to hit to get close to the Flex Foot’s original design, but with more scope changing this number would have led to an interesting dynamic analysis of a runner’s stride as well as that of a lateral sports player. 4.2.1 Metallic Side Study As a quick side study, we briefly take a look at other cutting edge materials that could be used in place of laminate composite for a prosthetic design similar to the Flex 25 Foot. Keeping in mind our model from Figure 2-2 and Figure 2-3, we are seeking a design that matches the response of the Flex Foot design while maintaining as little weight as possible. We first select a few high strength metals, Aluminum 7078 T6, TiAl4v, and SS 174, and determine the modulus of elasticity of the prosthetic part with various thicknesses: Prosthetic Response at: 1500 N Thickness: 0.004 Al 7075 T6 TiAl4v SS 17-4 Max Deflection (m) 0.110 0.060 0.080 0.047 0.070 0.035 0.035 0.028 0.023 Al 7075 T6 TiAl4v SS 17-4 Modulus of Elasticity (Pa) 13636 25000 18750 31915 21429 42857 42857 54545 66667 0.006 0.008 Table 4-1: Deflection and Modulus Response of Selected Materials Prosthetic Modulus (Pa) 70000 60000 50000 40000 Al 7075 T6 30000 TiAl4v 20000 SS 17-4 Bruggeman 10000 0 0 0.002 0.004 0.006 0.008 0.01 Thickness (m) Figure 4-1: Modulus Response of Selected Materials We can see that the different materials require different thicknesses to emulate at least the same spring constant of the prosthetic design. We can then compare the densities, thicknesses and the volumes to see which design is advantageous in terms of weight. 26 First we analyze our model to determine its total length and width: Section Shin Ankle Arc 1 Ankle Arc 2 Heel Arc Toe Arc Radius (m) NA 0.15 0.05 0.25 0.11 Arc Angle (degrees) NA 38 90 58 39 Length (m) 0.190 0.099 0.079 0.253 0.075 Total Length Width of Prosthetic 0.696 m 0.070 m Table 4-2: Model Dimension Lengths Figure 4-2: Dimensions of Baseline Prosthetic Then we look at how much each material would weigh at the thickness it has to be to support the prosthetic’s mechanical response. Material Composite Al 7075 T6 TiAl4v SS 17-4 Density (kg/m^3) 1600 2810 4430 7800 Thickness (m) 0.008 0.007 0.0065 0.0055 Volume (m^3) 0.000389744 0.000341026 0.000316667 0.000267949 Table 4-3: Weight of Prosthetic Using Selected Materials 27 Weight (kg) 0.624 0.958 1.403 2.090 Weight (kg) 2.500 Weight 2.000 1.500 1.000 0.500 0.000 Composite Al 7075 T6 TiAl4v SS 17-4 Material Figure 4-3: Weight of Prosthetic Using Selected Materials As you can see, even the strongest of the materials, the SS 17-4, could not be thin enough to overcome its much higher density than the composite. Even the lightest of the metals was almost a pound heavier. While this might not be a deal breaker for many, in a competitive sport, every pound counts. 4.3 Design Iterations The immediate design iterations to consider after the results found in this study would be adjusting the ankle section dimensions of the prosthetic and adjusting the width of the prosthetic footprint. Both of these ideas have parallels in sporting goods design. A running shoe is vastly different than a tennis shoe in these two areas. While a running shoe favors lightness and minimalism, a tennis shoe requires a large amount of ankle support due to the lateral movements. This support, in the form of a thicker and higher shoe wall helps the tennis player resist the higher forces of moving from side to side. On the same vein, the footprint of a tennis shoe is wider than that of a running shoe. Again in this case the running shoe prefers thinness and lightness while the tennis shoe requires a wider sole for ankle support and for lateral traction. These are the two obvious iterative areas to explore, tough other design changes exist that may help the Flex Foot design in a lateral sport. These include the connection point 28 between the leg connection of the prosthetic and the foot itself. In the original design, the foot is simply bolted on. But when a large amount of torsion is introduced by the lateral forces, a more integral connection could be considered to prevent high local stresses at the bolts. 4.4 Conclusion Based on the initial results of this analysis, the current Flex Foot design is structurally adequate enough to survive the larger forces of lateral movement in sports. Simple composite patterns can be used to build this prosthetics baseline design. However like in other sporting goods design, tweaks can be made to the foot to allow for better lateral dynamics. 29 5. References 1. “Perfecting the Prosthetic Leg: How incremental innovation works for patients.” Eucomed Medical Technology www.eucomed.org, http://www.eucomed.org/uploads/_key_themes/mdd/eucomed_incremental_innovati on.jpg 2. Knight, Tom. “IAAF call time on Oscar Pistorius’ dream”. The Telegraph. 10 Jan 2008. http://www.telegraph.co.uk/sport/othersports/athletics/2288489/IAAF-call- time-on-Oscar-Pistorius-dream.html 3. Bruggemann, Arampatzis, Emrich, and Potthast. “Biomechanics of double transtibial amputee sprinting using dedicated sprinting prostheses.” Institute of Biomechanics and Orthopaedics, German Sport University, Cologne, Germany. Sports Technology 2008, 1, No. 4-5, 220-227. DOI: 10.1002/jst.63 4. Weyand, Peter G.; Bundle, Matthew W.; McGowan, Craig P.; Grabowski, Alena; Brown, Mary Beth; Kram, Rodger; Herr, Hugh (17 February 2009), The fastest runner on artificial legs: different limbs, similar function?, Journal of Applied Physiology, retrieved 19 September 2012 5. “Flex Foot Cheetah Product Description.” Ossur Americas. Web. 2014. http://www.ossur.com/prosthetic-solutions/products/feet/feet/cheetah 6. Clarke, Dixon, Damm, and Carré. The effect of normal lad force and roughness on the dynamic traction developed at the shoe-surface interface in tennis.” Department of Mechanical Engineering, The University of Sheffield, Mappin Street, Sheffield S1 3JD, UK. Sports Eng (2013) 16:165-171. DOI 10.1007/s12283-013-0121-3. Tennis Friction Paper 7. “Unit VIII – The Foot and Ankle.” Virginia Commonwealth University Course Lectures. Web. 2014. http://www.courses.vcu.edu/DANC291-003/unit_8.htm 8. http://upload.wikimedia.org/wikipedia/commons/thumb/7/71/Oscar_Pistoris_at_Inter national_Paralympic_Day%2C_Trafalgar_Square%2C_London__20110908.jpg/640px-Os-car_Pistorius_at_International_Paralympic_Day%2C_Trafalgar_Square%2C_London_-_20110908.jpg 30 9. Mechanical Properties of Carbon Fibre Composite Materials, Fibre / Epoxy resin (120 degree Cure). Performance Composites Ltd. Web. 2009. http://www.performance-composites.com/carbonfibre/mechanicalproperties_2.asp 10. Metals Handbook, Vol.2 - Properties and Selection: Nonferrous Alloys and SpecialPurpose Materials, ASM International 10th Ed. 1990 11. SS 17-4 Data Bulletin. AK Steel Corporation. 9227 Centre Pointe Drive, West Chester, OH 45069. 2007. 12. Materials Properties Handbook: Titanium Alloys, R. Boyer, G. Welsch, and E. W. Collings, eds. ASM International, Materials Park, OH, 1994. 13. Metals Handbook, Vol. 3, Properties and Selection: Stainless Steels, Tool Materials and Special-Purpose Metals, Ninth Edition, ASM Handbook Committee., American Society for Metals, Materials Park, OH, 1980. 14. Structural Alloys Handbook, 1996 edition, John M. (Tim) Holt, Technical Ed; C. Y. Ho, Ed., CINDAS/Purdue University, West Lafayette, IN, 1996. 31 6. Appendix A 6.1 Mesh Experiment Mesh Size Results Mesh Size (m) Maximum von Mises Stress (108 Pa) 1.991 1.991 1.354 1.189 1.409 1.484 2.392 0.07 0.06 0.05 0.04 0.03 0.02 0.01 Maximum von Mises Stress (10e8 Pa) Maximum von Mises Stress (Pa e8) 6.1.1 3 2.5 2 1.5 Maximum von Mises Stress (108 Pa) 1 0.5 0 0 0.02 0.04 0.06 Mesh Size (m) 32 0.08 6.1.2 Mesh Raw Data 6.1.2.1 Mesh Size 0.07 6.1.2.2 Mesh Size 0.06 33 6.1.2.3 Mesh Size 0.05 6.1.2.4 Mesh Size 0.04 34 6.1.2.5 Mesh Size 0.03 6.1.2.6 Mesh Size 0.02 35 6.1.2.7 Mesh Size 0.01 6.1.2.8 Mesh Size 0.005 Mesh Size 0.005 turned out to be too fine for efficient processing. 36 6.2 Number of Plies 6.2.1 Summary Table 0/90 Vertical Load (N) 20 Ply Displacement (m) 20 Ply Linear Stiffness (Pa) 24 Ply Displacement (m) 24 Ply Linear Stiffness (Pa) 30 Ply Displacement (m) 30 Ply Linear Stiffness (Pa) 40 Ply Displacement (m) 40 Ply Linear Stiffness (Pa) 50 Ply Displacement (m) 50 Ply Linear Stiffness (Pa) 60 Ply Displacement (m) 60 Ply Linear Stiffness (Pa) 100 0.0280 3571 0.0240 4167 0.0175 5714 0.0125 8000 0.0095 10526 0.0080 12500 200 0.0400 5000 0.0310 6452 0.0210 9524 0.0190 10526 0.0110 18182 0.0081 24691 300 0.0540 5556 0.0375 8000 0.0250 12000 0.0200 15000 0.0120 25000 0.0093 32258 400 0.0600 6667 0.0440 9091 0.0400 10000 0.0240 16667 0.0130 30769 0.0100 40000 500 0.0720 6944 0.0525 9524 0.0450 11111 0.0275 18182 0.0150 33333 0.0108 46296 600 700 800 0.0770 NA NA 7792 #VALUE! #VALUE! 0.0560 NA NA 10714 #VALUE! #VALUE! 0.0550 NA NA 10909 #VALUE! #VALUE! 0.0325 0.0240 0.0260 18462 29167 30769 0.0170 0.0175 0.0180 35294 40000 44444 0.0115 0.0120 0.0125 52174 58333 64000 Bruggeman Linear Stiffness (Pa) 38700 38700 38700 38700 38700 38700 6.2.2 38700 38700 Summary Figure 70000 20 Ply Linear Stiffness (Pa) 60000 24 Ply Linear Stiffness (Pa) 50000 30 Ply Linear Stiffness (Pa) 40000 40 Ply Linear Stiffness (Pa) 30000 50 Ply Linear Stiffness (Pa) 20000 60 Ply Linear Stiffness (Pa) 10000 Bruggeman Linear Stiffness (Pa) 0 100 200 300 400 500 600 700 37 800 6.2.3 Experiment Setup 6.2.4 20 Ply, Displacement Response to Specific Load Values 6.2.4.1 20 Ply, 100 N 38 6.2.4.2 20 Ply, 200 N 6.2.4.3 20 Ply, 300 N 6.2.4.4 20 Ply, 400 N 39 6.2.4.5 20 Ply, 500 N 6.2.4.6 20 Ply, 600 N 6.2.4.7 20 Ply, 700 N 40 6.2.4.8 20 Ply, 800 N 6.2.5 24 Ply, Displacement Response to Specific Load Values 6.2.5.1 24 Ply, 100 N 6.2.5.2 24 Ply, 200 N 41 6.2.5.3 24 Ply, 300 N 6.2.5.4 24 Ply, 400 N 6.2.5.5 24 Ply, 500 N 42 6.2.5.6 24 Ply, 600 N 6.2.5.7 24 Ply, 700 N 6.2.5.8 24 Ply, 800 N 43 6.2.6 30 Ply, Displacement Response to Specific Load Values 6.2.6.1 30 Ply, 100 N 6.2.6.2 30 Ply, 200 N 6.2.6.3 30 Ply, 300 N 44 6.2.6.4 30 Ply, 400 N 6.2.6.5 30 Ply, 500 N 6.2.6.6 30 Ply, 600 N 45 6.2.6.7 30 Ply, 700 N 6.2.6.8 30 Ply, 800 N 6.2.7 40 Ply, Displacement Response to Specific Load Values 6.2.7.1 40 Ply, 100 N 46 6.2.7.2 40 Ply, 200 N 6.2.7.3 40 Ply, 300 N 6.2.7.4 40 Ply, 400 N 47 6.2.7.5 40 Ply, 500 N 6.2.7.6 40 Ply, 600 N 6.2.7.7 40 Ply, 700 N 48 6.2.7.8 40 Ply, 800 N 6.2.8 50 Ply, Displacement Response to Specific Load Values 6.2.8.1 50 Ply, 100 N 6.2.8.2 50 Ply, 200 N 49 6.2.8.3 50 Ply, 300 N 6.2.8.4 50 Ply, 400 N 6.2.8.5 50 Ply, 500 N 50 6.2.8.6 50 Ply, 600 N 6.2.8.7 50 Ply, 700 N 6.2.8.8 50 Ply, 800 N 51 6.2.9 60 Ply, Displacement Response to Specific Load Values 6.2.9.1 60 Ply, 100 N 6.2.9.2 60 Ply, 200 N 6.2.9.3 60 Ply, 300 N 52 6.2.9.4 60 Ply, 400 N 6.2.9.5 60 Ply, 500 N 6.2.9.6 60 Ply, 600 N 53 6.2.9.7 60 Ply, 700 N 6.2.9.8 60 Ply, 800 N 54
© Copyright 2026 Paperzz