Clinical Science (1982) 63,107-1 13 107 Cardiac output measured by transthoracic impedance cardiography at rest, during exercise and at various lung volumes A . T . E D M U N D S , S. G O D F R E Y * A N D M A R I O N T O O L E Y Department of Paediatrics and Neonatal Medicine, Hammersmith Hospital. London (Received 27 July114 December 1981; accepted 5 March 1982) Summary 1. Cardiac output measured by transthoracic impedance cardiography has been compared with simultaneous measurements made by the indirect Fick CO, rebreathing method in nine adults and 14 children. All were healthy normal volunteers. Sixty-six comparisons were made at rest and during steady exercise at work loads up to 100 W. 2. Impedance measurements of cardiac output were consistently higher than indirect Fick measurements of cardiac output, but after application of a correction factor related to packed cell volume there was close correlation between the results obtained by the two methods (r = 0-94). 3. The mean coefficient of variation of impedance measurements of cardiac output was 13% at rest and 5% during steady-state exercise. 4. Changes of lung volume due to breath holding or resulting from addition of an expiratory resistance did not affect the measurement of cardiac output by impedance. 5. Transthoracic impedance cardiography is a rapid, non-invasive technique for measurement of cardiac output. It requires very little active co-operation from the subject. The method would probably give reliable results for patients with respiratory illnesses such as acute asthma or bronchiolitis, during which changes of lung volume may be expected to occur. * Present address: Department of Paediatrics, Hadassah University Hospital, Mount Scopus, Jerusalem, Israel. Correspondence: Dr A. T. Edmunds, Royal Hospital for Sick Children, Sciennes Road, Edinburgh EH9 lLF, Scotland, U.K. 0143-5221/82/080107-07$01.50 Key words: cardiac output, exercise, impedance cardiography, indirect Fick, lung volumes. Introduction Impedance cardiography is a non-invasive technique which has numerous potential applications in clinical practice. Cardiac output is measured by detection of the changes in transthoracic electrical impedance associated with the cardiac cycle [l]. The method and its potential applications have recently been reviewed by Mohapatra [21. It relies on the over-simplified assumption that the thorax is a homogeneous column of blood, the volume of which increases by the left ventricular stroke volume during each systole [31. Because of this the results are largely empirical and measurements made under various physiological conditions must be verified by comparison with established standard methods. Comparison with several other methods has been made in adults and children with various results 13-11]. As far as we are aware cardiac output measured by the impedance technique has not previously been compared with simultaneous measurements made by the indirect Fick CO, rebreathing method. We report such a comparison made in normal subjects at rest and during exercise. We have also studied the effects of alteration of lung volume caused by breath holding and by breathing through an expiratory resistance on impedance measurements of cardiac output in normal subjects. This has been done in an attempt to determine the potential value of the impedance method for measuring cardiac output in patients with lung disease. 0 1982 The Biochemical Society and the Medical Research Society 108 A . T. Edmunds, S. Godfrey and M . Tooley TABLE1. Anthropometric data of 23 subjects Mean values SE are shown. Subject Males Females Age (years) Height (cm) Weight (kg) ~ Adults Children 4 5 9 5 29 ? 1.1 11,5?0.8 Methods Subjects Comparison of cardiac output measured by impedance with simultaneous indirect Fick CO, measurements was made in studies of 23 normal healthy volunteers. Nine were adults (five women and four men), mean age 29 f SD 1 years. Fourteen were children (nine boys and five girls), mean age 11.8 rt 0.8 years. Their anthropometric data are shown in Table 1. All consented to participate in the study after full explanation, and were free to withdraw at any stage. Impedance cardiac output Impedance cardiac output was measured with the Minnesota Impedance Cardiograph model 304A. This utilizes four disposable aluminized strip electrodes. Two were placed around the neck, one with its lower border at the level of the supra-sternal notch anteriorly and the spinous process of the seventh cervical vertebra posteriorly and one parallel to the latter, but at least 3 cm above it. The lower pair of electrodes was applied horizontally, one with its upper border at the tip of the xiphisternum and the second parallel and at least 3 cm below it. A constant sinusoidal current of 4 mA r.m.s. with a frequency of 100 KHz was applied between the outer two electrodes. The potential difference, Z,, between the two inner electrodes was recorded via a high input impedance linear amplifier. The potential varies during the cardiac cycle. For the calculation of stroke volume the thorax is assumed to be a homogeneous cylinder of blood, the diameter of which changes uniformly during the cardiac cycle. The measured change of potential reflects the increase in volume of the cylinder during systole. This increase in volume, AV, can be calculated by using the formula of Kubicek et al. [31. 173 ? 4 141f4 Surface area (m') ~~ 64 ? 3 31?3 1.75 f 0.06 1.11f0.06 where AV represents stroke volume (ml), P is the resistivity of blood (ohmlcm), L is the mean distance between the recording electrodes, 2, is the basal thoracic impedance (ohm), (dZldt)max. is the maximum rate of fall of impedance in ohmsls and T is the ventricular ejection time (s). In this study the outputs taken from the impedance cardiograph were Z,, dZldt, an ECG trace and a phonocardiogram trace. A typical record is illustrated in Fig. 1. P was calculated according to packed cell volume from the data of Mohapatra et al. [121. Packed cell volume was measured on finger-prick blood specimens spun for 3 min on a Hawksley haematocrit centrifuge. L was calculated as the mean of the distances between the lower edge of the upper and upper edge of the lower recording electrodes in the mid-line anteriorly and posteriorly, 2, was taken as the mean of the maximum and minimum values on the digital display during the period of the record. The subjects breathed freely during measurements. Therefore the baseline for the measurement of (d2ldt)max. was taken as the start of the steep upstroke irrespective of its position relative to the calibration zero [2, 101. T was calculated as the time between the start of the steep upstroke and the start of the second heart sound, which coincides with the lowest point on the dZldt trace. Beats were analysed only if there was no obvious movement artifact. Traces from subjects at rest, during treadmill walking and during moderate exercise on the bicycle ergometer were rarely distorted, and most could be analysed without difficulty. At the highest workloads movement of the baseline in association with respiration became more prominent and a proportion of the traces were not suitable for analysis. When this happened the measurement was immediately repeated, while the steady-state exercise was continued unaltered. Cardiac output was calculated from the mean stroke volume measured over 8-12 consecutive beats during one or more complete respiratory cycles: Cardiac output = Mean AV x HR where HR is heart rate (beatslmin). Impedance cardiography Indirect Fick CO, rebreathing cardiac output Measurements of cardiac output by the indirect Fick CO, rebreathing method were made as described by Godfrey [131. As the subjects were all normal, end-tidal CO, concentration was measured and assumed to have a partial pressure equal to that of arterial blood (Paco,). Subjects breathed through a low resistance, low dead space (50 ml) valve. Inspired ventilation was measured with a dry gas meter and the expired gas flushed through a 4 litres mixing chamber. Gas was continuously sampled either from the mouthpiece or from the outlet of the mixing chamber. It was passed through an infrared analyser to measure the concentration of CO, and a paramagnetic oxygen analyser to measure the concentration of 0,. The analysers were calibrated frequently during study periods with gases of known concentration, which had been analysed chemically by means of the Haldane technique. Data were continuously recorded on an ink jet recorder. CO, rebreathing was performed with a sliding valve assembly. By means of this the subject could be switched instantaneously from breathing air through the valve, to rebreathe from a previously prepared bag of CO, in 0,. The bag CO, concentration was made slightly higher than the predicted mixed venous CO, concentration (Hco,) and the volume was adjusted according to the subject’s tidal volume to ensure good equilibration with alveolar gas in less than one circulation time. fico, was 109 measured as the plateau concentration of CO, achieved. If no satisfactory plateau was recorded it was estimated by using the extrapolation method of Denison [13, 141. In the calculation of cardiac output, measured values of HCO, were used for all subjects with no correction for the down-stream effect described by Jones et al. 1151. Measurements of steady-state oxygen consumption (Yo,) and cardiac output were made with the subject seated at rest on a bicycle ergometer and performing exercise at one or more workloads ranging from 16.5 to 100 W. Impedance cardiograph recordings were made during the measurement of steady-state Po, immediately before the performance of the CO, rebreathing manoeuvre, so that comparison of paired measurements could be made. Effects of change of lung volume and addition of expiratory resistance The effect of change of lung volume on the measurement of cardiac output by impedance was determined in seven resting subjects. Measurements were made during breath holding with glottis open at functional residual capacity (FRC), total lung capacity (TLC) and at residual volume (RV). For each subject vital capacity was first measured by spirometry. During the experiment the subject breathed through a mouthpiece and three-way tap with outlets to room air and to a water-sealed spirometer. Cardiac output was measured first at FRC, then at TLC and finally 110 A . T. Edmunds, S. Godfrey and M . Tooley at RV after a maximum expiratory manoeuvre into the spirometer. The effect of breathing against an expiratory resistance on the measurement of cardiac output by impedance was determined in six normal adults. With the subject seated in a constantpressure whole-body plethysmograph and breathing in time with a metronome, a variable expiratory resistance was adjusted until FRC increased by 50% of the resting inspiratory capacity. Subsequently outside the plethysmograph, again with the subject breathing in time with the metronome, a steady-state resting cardiac output was measured by impedance and indirect Fick methods. The measurements were then repeated with the previously determined resistance in the expiratory line between the mouthpiece and the mixing chamber. The reproducibility of the measurement of cardiac output by impedance was determined at rest in 11 subjects and during steady-state treadmill walking in three subjects. Six to ten measurements were made over a period of half an hour and the coefficient of variation was calculated. Results The linear relationship between the data obtained by simultaneous measurements of cardiac output by indirect Fick CO, rebreathing and by impedance is shown in Fig. 2. The data for nine adults and 14 children are plotted separately. Impedance measurements of cardiac output were consistently higher than indirect Fick measurements of cardiac output. This difference was more marked for adults than for children. The ratio of impedance cardiac output values to indirect Fick values was significantly related to packed cell volume (Fig. 3). This relationship could be used as a correction factor to reduce the calculated values of impedance cardiac output. From the regression line in Fig. 3 the impedance cardiac output/Fick cardiac output ratio was determined according to the measured packed cell volume. Impedance cardiac output was then divided by this ratio, which was used as the correction factor. The relation between indirect Fick cardiac output and impedance cardiac output for all subjects after correction for packed cell volume in this way is shown in Fig..4. There was a linear relationship between Vo, and cardiac output measured by both methods, as shown in Fig. 5. The reproducibility of the measurement of cardiac output by impedance cardiography in 11 subjects sitting at rest and in three subjects during . . .'./ 22 23 21 20 - - Indirect Fick Q (Vmin) FIG. 2. Linear regression of simultaneous paired measurements of cardiac output (Q) by impedance cardiography vs indirect Fick CO, rebreathing. Measurements were made in resting subjects and Line of during exercise. a, Adults; , children. -, best fit for adults: y = 1 . 3 4 ~+ 1.79 (k2.1 SE), r = 0.92, n = 34. - - -, Line of best fit for children: y = 1.22X + 0.61 (f1.2 SE), r = 0.88, n 32. --, Line of identity. 34 35 36 37 3 8 39 40 41 42 43 44 45 46 47 PCV (%) FIG. 3. Linear regression for the ratio of cardiac output (Q) measured by impedance and measured by indirect Fick CO, rebreathing vs packed cell volume (PCV). y = 0 . 0 3 ~+ 0.28 (k0.26 SE), r = 0.39, n = 66. steady-state treadmill walking is shown in Table 2. The mean coefficient of variation at rest was 13% and during exercise was 5%. Analysis of variance demonstrated that after removal of the difference between subjects the standard Impedance cardiography // *= i .i .. g 12;1 1 g 10x 9.s 8 E i -2 1 2 3 4 4 1 v 7- 'E 0 111 5 6 7 8 9 10111213141516 6 54 3 - 2 - 0.2 0 0.4 0.6 Indirect Fick Q (I/min) FIG. 4. Linear regression of simultaneous paired measurements of cardiac output (Q) by impedance cardiography corrected for packed cell volume (PCV) and by indirect Fick CO, rebreathing. Measurements were made in resting subjects and during exercise. M, Adults; 0 , children. -, Line of best fit: y = 0.93~+ 0.31 (&1.17 SE), r = 0.94, n = 66. __- ,Line of identity. 15 14 - ;. .z2 Discussion In early studies of impedance cardiography a constant value of 150 ohm/cm was assumed for P 13-6, 91. More recently various values of P, related to the subjects' packed cell volume, have been used [7, 8, 10, 111. A number of different relationships between P and packed cell volume have been reported 1111. We have used the relationship reported by Mohapatra ef al. 1121 because it was obtained with fresh specimens of capillary blood from normal newborn babies, children and adults. Costeloe et al. 1101 found 1.2 1.4 1.6 13 1.2 1.4 1.6 - 1211 10- U * 98 7- e .- 6 - .0 deviation of the residual differences for the resting measurements was 10%. For the measurements during steady-state exercise it was 4.9%. Thus the 95% confidence limits for measurement of cardiac output by impedance cardiography were f10% during moderate steady-state exercise. Comparison of 19 paired estimates of cardiac output made in seven subjects breath holding at TLC and RV is shown in Fig. 6. A paired t-test showed no significant differences between the measurements made at the two lung volumes (P > 0.05).Comparison of paired measurements in six subjects made at rest and while breathing against an expiratory resistance is shown in Fig. 7. Again a paired f-test showed no significant difference between measurements made at rest and those made while breathing through the resistance (P> 0.05). 0.8 1.0 Po, (I/min) - 0 E 54 3 - 21 - 0 0.2 0.4 0.6 0.8 1.0 Po, (I/min) FIG. 5 . Linear regression for cardiac output vs fro, at rest and during exercise. (a) Impedance cardiac output (Q) corrected for packed cell volume. M, Adults [ y = 6 . 2 ~+ 3.7 ( + l a 3 SE), r = 0.93, n = 341; 0 , children [ y = 5 . 4 ~+ 2.7 (k1.1 SE), r =.0.79, n = 321. (b) Indirect Fick cardiac output (Q). H, Adults [ y = 7x + 3.1 (t1.0 SE), r = 0.97,n = 341; 0 , children [ y = 6x + 2.7 (L-0.8 SE), r = 0.9, n = 321. this gave an overestimation of cardiac output in infants, and derived a correction factor related to packed cell volume to bring their impedance measurements into line with simultaneous measurements of pulmonary capillary flow obtained by the nitrous oxide rebreathing method. Many factors including chest shape and the resistivity of tissue must affect measurement of cardiac output by impedance. Hence it is unlikely that the same correction factor would apply to newborn babies and to older children or adults. We therefore derived a new correction factor for our subjects. It is again related to packed cell A . T. Edmunds, S . Godfrey and M . Tooley 112 TABLE2. Reproducibility of measurements of cardiac output by impedance cardiography in 11 subjects at rest and three subjects during steady-state exercise Mean cardiac outputs f SD are shown. Subject no. No. of observations Mean cardiac output (I/min) ~ Coefficient of variation Mean coefficient of variation (%) (%) ~~~ At rest 1 2 3 4 5 6 3.9 f 0.7 4.2 f 0.4 3.0 f 0.3 6.0 0.4 6.2 f 0.7 8 6 6 9 9 + 2.7 f 0.7 5.9 f 0.6 8 9 7 8 8 5.4 f 0.2 8.0f 0.4 3.9 f 0.4 5.4 f 0.6 10 10 10 7.8 f 0.4 13.5 It 0.6 13.2 f 0.6 8 8 8 9 10 11 13 10 4 5 10 II Treadmill walking 12 13 14 :} 5 5 7- 7 - z.- 6h E 2 > e: z C 3 2 5- 65- 4 - Q g z 3- E 2.,a 1 ~ 0 1 2 3 4 5 6 7 8 9 Q measured at TLC (I/min) FIG. 6. Relation between 19 paiied measurements of impedance cardiac output (Q) made at residual volume and total lung capacity in seven subjects at rest. - - -, Line of identity; 20% limits are shown by unbroken lines. volume and, probably because the range of packed cell volume in our subjects was relatively small, it could be expressed in linear form. After applying the above correction there was very close correlation (r = 0.94) between the simultaneous measurements of cardiac output made by the impedance and the indirect Fick methods (Fig. 4). This demonstrates that impedance cardiography can provide measurements of cardiac output in normal subjects at rest and during moderate exercise which are as accurate as those obtained with the indirect Fick CO, rebreathing 0 1 2 3 4 5 6 7 Q at rest (I/min) FIG. 7. Relation in six subjects between paired measurements of cardiac output (Q) while breathing against expiratory resistance and at rest. Measurements were made simultaneously by impedance (.) and by indirect Fick (0)in each subject. - - -, Line of identity; 20% limits are shown by unbroken lines. method. This accuracy was further confirmed by the finding of the expected linear relationship between cardiac output and Vo, (Fig. 5). The equations for the regression lines of cardiac output related to VO,for both adults and children obtained by impedance cardiography and by indirect Fick C O , rebreathing were similar to previously published equations derived from direct Fick measurements [ 161, indirect Fick Impedance cardiography measurements 131 and dye-dilution measurements [16, 171. The scatter of the indirect Fick results about the regression line was very similar to that in previous studies [ 13, 181 and that of the impedance data was only slightly greater. Analysis of variance demonstrated that between individual subjects there were significant differences in the slope of the relationship between impedance cardiac output and indirect Fick CO, rebreathing measurements of cardiac output. These differences were accounted for by the increasing scatter of measurements at the highest workloads on the bicycle ergometer. This scatter was not greater than expected when the coefficient of variation of the impedance measurements was taken into account. Therefore all data have been used in the calculation of the regression equations in Figs. 4 and 5. The mean coefficient of variation of impedance measurements at rest was 13% with 95% confidence limits of +_20%.This is similar to previously published values [3, 61. During steady-state treadmill walking the mean coefficient of variation was reduced to 5% with 95% confidence limits for the measurement of f 10%. This compares favourably with published values for reproducibility for 0, direct Fick [191, dyedilution [19, 201 and CO, rebreathing indirect Fick measurements [2 11. The reproducibility of measurements of impedance cardiac output was not determined at the highest workloads on the bicycle ergometer, but movement of the baseline was prominent in this situation and a proportion of the traces could not be analysed because of movement artifact. Thus the 95% confidence limits would almost certainly have been greater than f10%. Change of lung volume as a result of breath holding or induced by breathing against an expiratory resistance did not materially affect the measurement of cardiac output by impedance (Figs. 6 and 7). Thus impedance cardiography would probably be reliable for both clinical and experimental purposes in patients with respiratory illnesses such as acute asthma or bronchiolitis, during which changes of lung volume may be expected to occur. It has the added advantage that it is rapid, non-invasive and it can easily be carried out at the bedside. Acknowledgments We thank Mr A. 0. Hughes for assistance with statistical interpretation. A.T.E. was supported 113 throughout the study by a project grant from the Medical Research Council. References I l l BONJER, F.H., VAN DER BERG, J.W. & DIRKEN,M.N.J. (1952) The origin of the variations of body impedance occurring during the cardiac cycle. Circulation, 6,415-420. S.N. (1981) Non-invasive Cardiovascular Moni[21 MOHAPATRA, toring by Electrical Impedence Technique, pp. 1-185, 82. Pitman Medical, London. 131 KUBICEK, W.G., KARNEGIS, J.N., PATTERSON, R.P., WITSOE, D.A. & MATTSON,R.H. (1966) Development and evaluation of an impedance cardiac output system. Aerospace Medicine, 37,1208-1212. J.C. (1968) Determination of 141 HARLEY,A. & GREENFIELD, cardiac output in man by means of impedance plethysmography.Aerospace Medicine, 39,248-252. 151 KINNEN,E. (1970) Cardiac output from transthoracic impedance variations. Annals of the New York Academy of Sciences, 170,747-765. F.M. & 161 BAKER,L.E., JUDY,W.V.,GEDDES,L.E., LANGLEY, HILL,D.W. (1971) Measurement of cardiac output by means of electrical impedance. Cardiovascular Research Centre Bulletin, 9,135-145. Z., EHMKE,D.A., DURMISS, R.E., LEAVERTON, P.E. [71 LABABIDI, & LAVER,R.M. (1971) Evaluation of impedance cardiac output in children. Paediatrics, 47,870-879. J.C., MAHER,J.T., REEVES,J.T., CRUZ,J.C., 181 DENNISTON, CYMERMAN, A. & GROVER,R.F. (1976) Measurement of cardiac output by electrical impedance at rest and during exercise. Journal of Applied Physiology, 40,9 1-95. J.H., ORO, L. & EKELUND, L.G. 191 GABRIEL,S., ATTERHOG, (1976) Measurement of cardiac output in patients with myocardial infarction. Comparative evaluation of impedance and dye dilution methods. Scandinauian Journal of Clinical and Laboratory Investigation, 36,29-34. 1101 COSTELOE, K., STOCKS,J. & GODFREY,S. (1977) Cardiac output in the neonatal period using impedance cardiography. Paediatric Research, 11, 1171-1 177. [ I 11 ARENSON, H.M. (1978) The use of impedance techniques for the assessment of bloodflow during anaesthesia. Ph.D. Thesis, London University. 1121 MOHAPATRA,S.N., COSTEWE,K. & HILL, D.W. (1977) Blood resistivity and its implications for the calculation of cardiac output by the thoracic electrical impedance technique. Intensive Care Medicine, 3,63-67. S. (1974)Exercise Testing in Children, pp. 42-50, 1131 GODFREY, 47,95-101. W.B. Saunders Co. Ltd, London. (141 DENISON,D. (1968) Mixed venous blood gas tensions and respiratory stress in man. Ph.D. Thesis, London University. I151 JONES,N.L., CAMPBELL, E.J.M., MCHARDY, G.J.R., HIGGS, B.E. & CLODE,M. (1967) The estimation of carbon dioxide pressure of mixed venous blood during exercise. Clinical Science, 32,311-327. 1161 WADE, O.L. & BISHOP,J.M. (1962) Cardiac Output and Regional Blood Flow, pp. 5 1-72. Blackwell Scientific Publications, Oxford. 1171 ASTRAND,P.O., CUDDY,T.E., SALTIN,B. & STENBERG, J. f 1964) Cardiac outnut durine submaximal and maximal work. Journal of Applied i'hysiolo&, 19,268-274. I181 JONES, N.L., CAMPBELL,E.J.M., EDWARDS,R.H.T. & ROBERTSON, D.G. (1975) Clinical Exercise Testing, pp. 33-35. W.B. Saunders Co. Ltd, London. 1191 COURNAND, A,, RILEY,R.L., BREED,E.S., BALDWM, E.F. & RICHARDS,D.W. (1945) Measurement of cardiac output in man using a technique of catheterisation of the right auricle or ventricle. Journal of Clinical Investigation, 24,106-1 16. 1201 G F ~ B YG,., NILSSON,N.J. & SANNE, H. (1966) Serial determination of cardiac output at rest. British Heart Journal, 28,118-121. 1211 ZEIDIPARD,E., SILVERMAN, M. & GODPREY,S. (1972) Reproducibility of indirect (CO,) Fick method for calculation of cardiac output. Journal of Applied Physiology, 33,14 1-143.
© Copyright 2026 Paperzz