original studies Radiological Incident in the Radiotherapy Department. Presentation and Actions of Intervention Mihaela Dumitru1, Laura Rebegea1,2, Dorel Firescu2,3 1) 1 „Sf.Ap.Andrei” Emergency Clinical Hospital; 2) „Dunarea de Jos” University of Galati, Faculty of Medicine, Clinic Department; 3) „Sf. Ap. Andrei” Emergency Clinical Hospital, Galati, Surgery Clinic II, Galati Romania This paper presents a radiological incident which occurred with the radiotherapy equipment, Theratron Elite 100 from the Radiotherapy Department - Emergency Clinical Hospital „Sf. Ap. Andrei”, Galati at the beginning of February 2009 and describes the actions of subsequent intervention. The radiological incident concerned the framework which lies in a 180° position, under the treatment table, which prevented the possibility of manually returning to the source using the T-bar, a problem which is not mentioned in the operator’s manual of the radiotherapy unit. The radiological incident consisted of a radioactive source blocking in the TRANSIT position between BEAM OFF and BEAM ON positions at the start of treatment. We want to state that all personnel - physicians, physicists, operators and service staff acted with professionalism and promptitude, conforming to emergency procedures and and regulations. After this incident the Radiotherapy Department’s Emergency Plan was modified and updated including the emergency procedures described in this paper; the night shift was removed from the Radiotherapy Department’s work program. Also, we want to mention that misinformation and confusion regarding this incident must be avoided in order not to raise false alarms with other hospital staff and the public. Keywords: radiological incident, radioactive source, emergency procedure Introduction The radiological incident description that occurred in the Radiotherapy Department of the Emergency Clinical Hospital „Sf. Ap. Andrei”, Galati, February, 2009 was a unique situation concerning the framework which lies in a 180° position, under the treatment table, which prevented the possibility of manually returning to the source using the T-bar. The present paper aims to demonstrate the exposure danger for occupational personnel who did act conforming to emergency procedures from the Radioprotection Program but we also would like to state that neither the public or the environment was affected by this incident. Also under discussion is that this situation is not mentioned in the operator’s manual and public opinion and the press created very bad media regarding this incident, stating incorrectly that that there had been a radioactive leakage (we work with sealed radioactive Co-60 source) in the Radiotherapy Department and that an explosion had happened which had hurt other staff from other hospital departments. Methods and materials A description of the incident and the actions taken. This paper presents a radiological incident occurring with Journal of Radiotheraphy & Medical Oncology August 2012 Vol. 18 No 1: 18-21 Address for correspondence: Dr. Laura Rebegea „Sf.Ap.Andrei” Emergency Clinical Hospital, Radiotherapy Department Brailei St. 177, 800830 Galati, Romania Email: [email protected] the radiotherapy equipment, Theratron Elite 100 from the Radiotherapy Department - Emergency Clinical Hospital „Sf. Ap. Andrei”, Galati at the beginning of February 2009. In February 2009, 90 patients per day ( both day and also at the night shift, because of the great number of patients) were in treatment at one radiotherapy equipment, Theratron Elite 100,; the other radiotherapy equipment, Rokus M 40 was not functional at that date. The radiotherapy equipment specifications: - Name: Theratron Elite 100, SAD=100 cm, SDD= 45cm; - Radiation source: Co-60 sealed source, activity of 14000Ci (at November 2002) and dose rate of 102,23cGy/min at the moment of incident. The radiological incident consisted of the radioactive source blocking in the TRANSIT position between BEAM OFF and BEAM ON positions at the start of treatment. After positioning the patient (an elderly woman) and fixing the treatment parameters, the operators left the treatment room, and from the command room started the treatment. However, the source did not reach the BEAM ON position but remained blocked in the IN TRANSIT position. At that moment the unit went into the „inhibited state”, the automatic collimator closure feature was activated but the unit power shut off. Under normal circumstances, the source would have returned to the fully shielded position, but in our case the source remained exposed. Emergency procedures (1) for the automatic retreat of the source (push the emergency button from the command room) did not have any result. This situation forced the 2 operators Radiological incident in the radiotherapy department to open the treatment room door and enter the treatment room (source situated in the IN TRANSIT position) and evacuate the patient; this action took no more that 20 seconds. After the patient’s evacuation, the operators closed the treatment room door turned off the key switch at the control console and notified the Radiation Safety Officers. When they arrived, the Radiation Safety Officers entered the corridor of the treatment room with a portable survey meter. The survey meter indicated that the radiation level had increased and they decided to leave the room. Afterwards the radiation safety officers measured the dose rate at different points conforming to Fig. 1. The recorded values are indicated in Table I. Conforming to the inverse square distance low calculation, the dose rate in the exposure room had high values and staff exposure was not justified (conforming to specifications from the Operator’s Manual), and the patient had to be evacuated and the treatment room closed (1). The distance between the measurement point number 4 and the position of source was 7.5 meters (Fig. 1). Figure 1. The location of measuring points for the dose rate measurement at radiological incident data 19 on the front part of the treatment machine head. The T-bar has a 50cm length and the maneuver space has a 15 cm length, Figs 2 and 3. Fig. 2. Treatment unit’s position at the moment of the source blocking in the TRANSIT position. Fig. 1. The numbers in the figure specify the measure points: 1 – treatment unit command console, 2 – access door in treatment room, 3 – the middle of thecorridor , 4 – the end of corridor , 5 – dosimetry room. Table I. Dose rate values recorded at the radiological incident date Measure point* Dose rate values (µSv/h) 1 0,1 2 0,1 3 2 4 5 5 0,1 * the measure points are specified in fig. 1 Given the head of treatment machine position – under the treatment table, without sufficient space, the insertion of the T-bar was IMPOSSIBLE. The T-bar hole is situated Fig. 3. Irradiation head’s position at the moment of the source blocking in the TRANSIT position. In these conditions the radiation safety officers decided to notify the service unit – accredited Representative of MDS Nordion whose advice was to not intervene in this danger situation and to wait the arrival of service personal, to isolate the controlled zone and limit the staff’s and patient’s access even in the supervised zone beside the treatment room. The Radiation Safety Officers notified the Licensee who decided to wait for the arrival of the service personal. After that the competent authorities and the medical physics expert were also notified. 20 Dumitru et al Since the source remained in the TRANSIT position for an extended time, the Radiation Safety Officers made sure that all access to high radiation areas were prevented by locking doors and appointing knowledgeable staff to prevent unauthorized entry to these areas. The service team arrived the next day. They accessed the service menu of radiotherapy unit (for that they used a password which is not accessible to the radiotherapy department’s staff). For 3-4 hours they tried to rotate the gantry through repeated operations ON-OFF of the radiotherapy unit and the source returned itself in the fully shielded position. Immediately the decision was made to verify carefully with the radiotherapy installation and to command and replace the “5-years Kit”. First off all we have to reiterate that this decision – the waiting of the service team, after the patient was evacuated, did not put neither the occupational personnel, nor others from other departments of hospital, nor the public in danger because of the following reasons: - The shielding of treatment room, conforming with IAEA Reports Standards and Regulations (2), and conforming with NCRP 49 (3); - The treatment room walls are 150 cm of concrete with 2.35gm/cc; - The presence of the corridor with a wall of 100 cm, concrete with 2.35gm/cc, which attenuated the leakage and scattered radiation; - The installation’s particularity – the existing of the beam-stopper attached on the unit, a massive bloc of lead which protects almost 99.44% of the primary beam; behind this beam stopper the level of the dose rate is very low even during of exposure (Table II), so the public and environment was not affected during the incident. ■ The treatment installation passed into the INHIBIT state i.e.: ■ The Collimator closed at the minimum field, ■ Any maneuver of the installation was impossible, so the unnecessary motions of the treatment table or of the treatment machine, which could cause collisions, were impossible. For the staff who evacuated the patient from treatment room all measures to estimate the dose were taken; the electronic dosimeters indicated the following dose rates: Table II. Comparative dose rate values measured at different points and at different moments of time Measure point Dose rate (mSv/h) [February 2009*] Dose rate (mSv/h) [March 2010**] Behind beam-stopper at 1.5 m 0.58 0.5 Behind irradiation head on the collimator axe 2.71 2.35 T – bar insertion’s hole 2.61 2.26 * In February 2009 the reference dose rate was 102.23cGy/min; ** In March 2010 the reference dose rate was 88.66cGy/min. 30µSv/h and 25µSv/h, respectively, for each operator. The dosimetric films were immediately sent to the accredited dosimetric organisation for the staff affected. The values of the dose equivalent Hp (10) were 300±40µSv and 260±40µSv, respectively for each operator, were found to be in the range of the limits of a monthly equivalent dose [170-1700µSv]. The differences between the two determinations can be explained by the characteristics and the conditions of measurements: the positions of the two dosimeters respectively of the radioactive source. The dosimetric film accumulates through time the received dose, ( Some days had passed since the dosimetric films had been changed). The maximal admitted errors for Hp (10) ≤400µSv are +100% / -50%, conforming to AECB S-106 (E), IAEA – Safety Guide No. RS-G-1.3 % standards. The electronic dosimeters were calibrated to signal optically and acoustically the high dose rate detection and the intervention was very rapid. Also, specific blood tests and cytogenetic tests were taken at the Public Health Institute in Bucharest. These results were also within normal limits. Discussions and Conclusions In the literature there are many incidents and accidents due, in many cases, to human mistakes but also due to technical deficiencies. Some human mistakes we can mention here are: wrong commissioning or calibrating machines or the treatment planning system that can lead to a systematic under-dose (e.g. at the Department of Radiation Oncology, Royal Adelaide Hospital the radiotherapy to cancer patients, incorrectly delivered 5% less than what had been prescribed for cancer patients treated on that treatment machine between July 2004 and July 2006.) or systematic over-exposure, maintenance problems, errors associated with human machine interfaces, errors related to reading or entering from the patients chart, mistakes involving the wrong patient, the wrong field or the wrong site, the wrong beam or source, errors in communication, mistakes associated with special situations, personnel changes, special treatments (4-6). A notorious incident happened at the National Oncology Institute (Instituto Oncológico Nacional, ION) of Panama when 28 radiation therapy patients were overexposed, treated with the cobalt-60 therapy unit at the ION, in late 2000 and early 2001 (7-9). They had received a dose greater than the normal dose with errors ranging from +10 to +105% (7, 10, 11). Twenty-three of the 28 overexposed patients died by September 2005, with at least 18 of the deaths being from radiation effects (8). In our case the incident occurred due to a technical deficiency and in Romania this is the first time when such an incident has occurred and been reported. This was the reason why we considered it useful to publish all the facts related to this incident. Radiological incident in the radiotherapy department However, after these events we must point out some important aspects. First of all, this situation: the source blocked in the IN TRANSIT position, the framework under the treatment table in rotation at 180° without sufficient space to use the T-bar IS NOT specified in the Operator’s Manual of therapy unit. Even under these circumstances, the operator personnel demonstrated their professionalism, courage, evacuating the patient rapidly from the exposure room and taking all necessary actions, conforming to emergency procedures. Also, the radiation safety officers had the capacity to keep the situation under control and to make good decisions conforming to the complexity and the seriousness of the situation. On the other hand, precautions are required to be taken when this radiological incident become public information, because of the false alarming of other hospital’s staff and the public. Misinformation and confusion must be avoided at all costs. Therefore, the information to be given out to the media and the public must be taken from qualified staff of the therapy department or from a responsible source of the hospital. Radiotherapy is very much dependent on human performance and the professionalism of a very strong team is the key to its success. After this incident the Radiotherapy Department’s Emergency Plan was modified and updated including the emergency procedures described in this paper; the night shift was removed from the Radiotherapy Department’s work program. Also, these procedures were included in the personneltraining program, and simulations of emergency interventions with all staff, are now a regular occurrence. Disclosures: The authors indicated no potential conflict of interest 21 References 1. Operator’s Manual of Theratron Elite 100. 2. Safety Reports Series, No 47. Radiation Protection in the Design of Radiotherapy Facilities. INTERNATIONAL ATOMIC ENERGY AGENCY, VIENNA, 2006. 3. National council on radiation protection and measurements, Structural Shielding Design and Evaluation for Medical Use of Xrays and Gamma-rays of Energies up to 10MeV, Rep. 49, NCRP, Washington, DC, (1976). 4. IAEA Year book 1996, “Accidents in Radiotherapy: Lessons Learned”, IAEA, Vienna, 1996 p.p. D40-D44. 5. Ash D., Bates T. Report on the clinical effects of inadvertent radiation under-dosage in 1045 patients. Clin. Onc. 6:214-225, 1994. 6. Task Group on Accident Prevention and Safety in Radiation Therapy. Prevention of accidental exposures to patients undergoing radiation therapy. A report of the International Commission on Radiological Protection. Annals of the ICRP. 30(3):770, 2000. 7. Borrás C. Overexposure of radiation therapy patients in Panama: problem recognition and follow-up measures. Rev Panam Salud Publica. 2006;20(2/3);173–87. 8. International Atomic Energy Agency. Accidental overexposure of radiotherapy patients in San Jose, Costa Rica. Vienna: IAEA; 1998. 9. International Atomic Energy Agency. Investigation of an accidental overexposure of radiotherapy patients in Panama. Vienna: IAEA; 2001. 10. British Journal of Radiology. Central axis depth dose data for use in radiotherapy. London: The British Institute of Radiology; 1996. (The British Journal of Radiology Supplement 25). 11. Young MEJ, Batho HF. Dose tables for linear radium sources calculated by an electronic computer. Brit J Radiol. 1964; 37:38– 44.
© Copyright 2026 Paperzz