Touch TENS INSTRUCTIONS FOR USE

Touch TENS
INSTRUCTIONS FOR USE
(Reduced file size for download)
Touch TENS
Instructions for Use
Section
Page No
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
Introduction
How TENS works
Side Effects
Contents
Installation of Batteries
Belt clip
The Programmes
Programme Settings
Controls and Display
Controls
Connecting Lead wires
Positioning of Electrode Pads
General Pad Advice
Cleaning
Troubleshooting
Caution and Warnings
Consumables and servicing
WEEE
Guarantee
Technical Specifications
EMC Precautions
1
1
2
2
3
4
4
6
7
8
11
12
14
14
15
16
18
19
19
20
21
1.
INTRODUCTION
Device Description & Principles of Design
The Touch Tens is a dual Channel TENS unit with pre-set energy output levels for easy use. A choice of
eight pre-set programmes is available . The operation of the controls is described in detail in Sections 9 &
10.
The Tenscare Touch TENS is a dual channel TENS.
It has:
two independent strength controls
60 mA output over 500 Ohm load
8 pre-set programmes including Burst, Strength and Frequency Modulation, and Han
Programme Timer
Compliance Memory
2. HOW “TENS” WORKS
T.E.N.S. stands for Transcutaneous Electrical Nerve Stimulation.
T.E.N.S. stimulates your body’s own natural defences against pain, namely the release of endorphins.
T.E.N.S. is totally safe and has been used successfully by thousands of pain sufferers.
TENS sends a gentle stimulation through the skin which works in TWO ways:Pain Gate
Stimulating the sensory nerves, which carry touch and temperature signals. These nerves go to the same
connections in the spine as the nerves carrying pain. A strong sensory signal will block the pain signal
travelling up the spine to the brain, this is known as closing the “Pain Gate” and takes effect quite quickly
after the unit is switched on. You can use TENS several times a day, for as long as you like.
Endorphin Release
At low frequency settings, and slightly stronger output, TENS drives the motor nerves to produce a small
repetitive muscle contraction. This is seen by the brain as exercise, and this promotes release of
Endorphins – your body’s own natural pain killer. The relief builds up and normally takes about 40 minutes
to reach a maximum level which can last for hours after the machine is switched off.
By using TENS you can expect to achieve a significant reduction in pain if not complete pain relief.
3.
SIDE EFFECTS
There are no known side effects to TENS use and long-term TENS use is not harmful.
4.
CONTENTS
The pack contains:
Touch TENS Pain Relief unit
Belt Clip
2 x Leadwire
Pack of 4, 50x50mm, self-adhesive electrode pads
(item code: E-CM5050)
2 x AA 1.5V(Type LR6) alkaline batteries
Instruction Manual
Storage Pouch
5.
INSTALLATION OF BATTERIES
Remove battery tray from side of unit
i)
ii)
Slide down button in centre of battery cover to unlock.
Pull tray out in direction of raised arrows.
iii)
Insert batteries
Ensure that the batteries are inserted the right way as shown in battery compartment.
iv)
Lock battery cover
When the batteries are running low a low battery indicator will show on the screen (battery symbol) and it
is important to change the batteries as soon as possible.
Rechargeable batteries
The unit will not work with rechargeable batteries.
Low voltage cut-out
may be above the normal operating voltage of rechargeable batteries
Storage
Remove batteries from your TENS machine if the unit is unlikely to be used for a long period.
Some types of battery may leak corrosive fluid.
Disposal of batteries
Always dispose of batteries safely.
Do not throw batteries onto a fire.
Risk of explosion
Battery Life
Batteries should last about 48 hours at full power.
Unused batteries have a nominal shelf life of 3 years, but will usually last longer than this.
Warning
Keep batteries out of the reach of small children
If battery leakage occurs and comes in contact with the skin or eyes, wash thoroughly with lots of water.
Do not attempt to recharge alkaline batteries
6. BELT CLIP
A belt clip is provided to allow you to attach the touch TENS to your clothing. It
can be inserted either way up. Slide along the base of the unit until
it locks in position.
You can also hang the unit from your neck using the hook provided. Use thread or wool that will snap
easily if the unit snags on something.
7. THE PROGRAMMES
There is no one programme for a particular condition, and the best choice varies from one person to another,
even if they have the same type of pain.
Each user needs to select both the programme and the positioning of the pads that is best for them.
You may need to try a few positions/programmes before finding the one that suits you.
If you are using TENS for the first time, it is recommended that you start with Programme A, which
gives the gentlest stimulation.
The Touch TENS has 8 pre-set pain relief programmes.
Research has shown that some programmes are particularly effective for certain pains. These have been
indicated under the “Specific uses” section.
The following grid summarises the programmes available and
how they vary.
Pr
Output
Sensation
A
Constant
Tingling/Pins and needles sensation. Inc Pain Gate
rease strength slowly until a strong but
comfortable sensation.
B
Constant
200
C
Burst
D
Constant
Pulsing or tapping sensation associated Not for injured
4
with
muscles
muscle contractions. Increase the stren Endorphin Releas
gth to a level that just produces muscle a e
ctivity.
Stimulus is delivered in bursts, i.e. a
Not for injured
100in
group of pulses rather than a
muscles
2Hzburst
single pulse. You will get
Pain Gate/Endorp
muscle activity at a lower strength
hin
and thus you may find it
more comfortable.
Is similar to B, but has a Freq specifically Menstrual pain, nau
10
recommended for nausea, travel
sea, travel
sickness and
sickness and
morning sickness (see body map for posi morning sickness.
tion of electrode pads).
Not for injured
muscles
Endorphin Releas
e
E
Constant
This is similar to A, but has deeper penet Pain Gate
ration and a stronger feel. It
is more likely to produce muscle contract
ions at a lower strength
110
200
F
Strength This is similar to E, but with the strength Pain Gate
Modulated modulating continuously, this gives a ple
asant slow relaxing massage sensation.
110
200
G
Pain Gate
Frequency This changes the frequency every 3
2-110 in 200
Modulated minutes,
7steps
stepping from 2 to 110 Hz. This constant
change helps to prevent
accommodation occurring.
2, 10, 50, 80, 90, 100,110Hz,
HAN Alter This changes the frequency from 2 to 10 Pain Gate/Endorp 100/2
100
hin
nating
0 Hz every 3 seconds. It
/200
is claimed to be the most
effective form of TENS, but
may be less comfortable
H
Specific Uses
Frequen Pulse
cy
Width
Hz
µS
110
50
200
200
USAGE
TENS can be used for as long as is necessary. Continuous treatment is fine, but electrode pads should be
repositioned regularly (at least every 12 hours) to allow the skin to be exposed to the air.
The Endorphin Release modes work better when the strength is high enough to cause small muscle
contractions. Best results are achieved at between 20 and 40 minutes. Longer use may cause muscle
ache.
8. INFORMATION ABOUT THE PROGRAMME SETTINGS
• Frequency (measured in Hz - pulses per second)
PAIN GATE: A high frequency of 110Hz is good at blocking pain signals.
ENDORPHINE RELEASE: A low frequency of 4 or 10Hz allows for the release of endorphins, the body’s
natural morphine-like substances.
• Pulse Width (measure in μS - millionths of a second)
The Touch TENS unit has pulse widths of 50 and 200 μS. Higher Pulse Widths are more likely to cause
muscle contractions.
• Constant and Burst Modes
Constant mode is when the sensation is continuous as against Burst mode when the sensation, is as its
name implies, is one of on and off.
• Modulation Modes
Modulation is when either the frequency or pulse width sweep across range of settings. This enables the
body to receive many different signals and can be very beneficial and lessen any effect of accommodation.
See programmes F and G.
• HAN Stimulation
Invented by a Professor Han, this alternates between high and low frequencies to give - in theory - the best
pain relief.
9. CONTROLS AND DISPLAY
Channel 1 Strength
Programme selected
Keypad Lock
Channel 2 Strength
Low battery
Time Selected
Channel 1 Strength
UP
Channel 2 Strength
UP
ON/OFF
Programme selector
Channel 1 Strength
DOWN
Channel 2 Strength
DOWN
Time Selector
10. CONTROLS
ON/OFF
To turn the unit ON, press the ON/OFF button on the side of the unit and hold for 2
secs until the display shows.
To turn the unit OFF, press the ON/OFF button on the side of the unit and hold for 2 seconds
until the display stops.
The unit will turn OFF automatically :
When the Timer reaches zero,
If it is left at zero strength for more than 5 minutes.
At first use, or after changing batteries, the display shows that the unit is automatically set in Programme A
at zero strength.
On the top of the XL-Y1 unit there are four coloured buttons:
STRENGTH CONTROLS
The BLUE BUTTONS, marked ,  are strength controls.
The left hand buttons change strength in the left hand lead, Ch1, and the right hand
buttons change the right hand lead, Ch2.
To increase strength in steps of 1, press and release the top part of the two blue buttons marked .
The strength levels are shown on the LCD.
The strength control buttons will not operate until the unit is properly connected to you. Touch
TENS detects a disconnection and automatically returns the strength to zero.
The unit has 60 levels of strength.
If you hold down the button for 2 seconds, the strength will start scrolling.
You may feel nothing over the first few presses. Continue pressing until the sensation is strong but
comfortable. Further increases during use may be necessary if your body becomes used to the sensation.
To lower the strength, press and release the lower part of the blue buttons marked .
PROGRAMME CONTROL
The YELLOW BUTTON, marked #, is the Programme control.
The Touch TENS has eight different, specially developed programmes lettered A-H.
When the unit is first switched on it automatically enters
PROGRAMME A.
Each time you press and release the YELLOW BUTTON, the programme changes and the letter is shown
on the LCD.
Each time you change the programme, the strength level reverts back to zero.
This is a
safety feature to alleviate any sudden feeling of a surge, as each Programme gives a different sensation.
TIMER
The GREEN BUTTON, marked with a clock symbol, can be used to set the session
duration.
When you switch the unit on, it is automatically set at continuous, (displayed as CO) which
means that the unit will continue functioning until you switch it off.
You can set session times of 20,40, or 60 minutes by pressing the green button repeatedly.
The LCD shows the session duration next to the clock symbol. The unit
automatically counts down the minutes set and switches off when this reaches zero.
A LOW BATTERY display, with a symbol of empty batteries, will show when
you need to change the batteries. The unit will shut down about 2 minutes after this.
KEYPAD LOCK
If you do not press any keys for 30 seconds, the keypad will lock so that only the keys marked ▼
will work. This is to avoid accidental changes in setting.
To unlock, press either ▼ key.
OPEN CIRCUIT CUTOUT
If the Touch TENS is not correctly connected to your body, the strength will
automatically reset to zero. This is to prevent sudden changes if a broken connection is re-made.
MEMORY
The XL-Y1 has a Memory with three functions:
1) Programme Retention. When you switch the unit On, it will automatically start in the Programme which
was being used when it was switched Off.
2) Frequency of Use. Press Programme and Left blue ▲ together and hold down for 3
seconds. The display will show T (for no. of Times) and a number from 1 to 999, indicating the number
of times the unit has been used.
Press the same buttons again to return to normal controls.
3) Elapsed Time. Press Programme and Right blue ▲ together and hold down for 3
seconds. The display will show U (for Usage) and a number from 1 to 999 hours,
indicating the total number of hours the unit has been above zero strength.
Press the Programme button again to return to normal controls.
Memory Reset.
To reset Memory to zero, hold down the Programme and Timer buttons together for 3 seconds.
11. CONNECTING LEAD WIRES
Insert the leadwire plug into the base of the unit.
Attaching the electrode pads to Lead wire
Push the pin ends firmly into the pigtail ends of the electrode pads.
The leadwires may be damaged by rough handling, and should be treated with care.
Lead wire colour coding
The ends of the lead wire are coloured white or red. This coding is provided for some professional
uses, For most purposes, the orientation makes no difference, and you can ignore this colour coding.
12.
POSITIONING OF ELECTRODE PADS
The following body maps in this instruction booklet will show
you where to place the electrode pads for a range of common complaints, dependent on your symptoms.
For symptoms not illustrated, apply the pads around/near to the source of the pain or
seek advice from your doctor or physiotherapist.
The electrode pads must always be used in pairs,
so that the signal can flow in a circuit.
NOTE:
Always check unit is OFF before attaching or removing pads.
NOTE: Touch TENS resets strength to zero if the pad or lead is disconnected from your body.
This is to prevent sudden changes in sensation when the pad is re-connected.
For areas that are difficult to reach, why not ask a friend to help you attach the pads?
See Section 16 for information on where you should not place pads.
ELECTRODE PAD POSITIONS
Low Back Pain
Shoulder Pain
Ankle Pain
Neck and Shoulder
Tension
Knee Pain
Headache
Where only two pads are shown on the arm, shoulder, and leg,
use the other two pads either on the opposite limb
Sciatica
Elbow Pain
Leg Pain
13.
GENERAL PAD ADVICE









The electrode pads supplied are reusable but are for single patient use.
In order to obtain the best conductivity through the pads always ensure that they are in good
condition and tacky.
Before use make sure your skin is clean and dry.
Peel the electrode pads from their protective plastic shield by holding and lifting one corner
of the pad and pulling.
Do not pull on the pigtail wire of the pad.
After use always replace the pads on the plastic liner and replace in the re-sealable plastic bag.
If the pads dry out then it is best to buy a replacement pack of electrodes. In an emergency it
may be possible to restore some of the tackiness of the pad by adding a tiny drop of water on
each pad and spreading around. If too much water is added the pads will become too soft then it
is suggested in order to try re-establish some adhesiveness to place them sticky side up in
a refrigerator for a few hours.
In very hot weather the gel on the pads may become soft. In such cases place the pads, still
on their plastic liners and in their bag into a fridge until they return to their normal condition.
The pads provided are latex –free.
Replace pads when they lose their stickiness.
Poor connection may cause discomfort and skin irritation

Storage life of an unopened pack of electrodes is 2
years. This may be affected by very high temperatures or very low humidity.
14. CLEANING
Clean the case and lead wires at least once a week by wiping with a damp cloth and a solution of mild soap
and water. Wipe dry.
Do not immerse your TENS machine in water.
Do not use any other cleaning solution than soap and water.
15. TROUBLESHOOTING
If your TENS machine is not working properly please check the following:
Problem
Possible causes
No display Flat batteries
Batteries inserted
incorrectly
Solution
Replace batteries
Check
Damaged battery tray Contact supplier
contacts
Low
battery
display
Low batteries
Replace batteries
No
sensation
Incorrect connection
Have you applied both
electrode pads (per lead wire)
to ensure a complete circuit?
Not strong enough
Increase strength.
Most users will feel
something at a setting below
20
Output will Open circuit cut-out Ensure machine is attached
not
operating.
to your body correctly.
increase
Lead not connected to Try using the second lead.
above zero body or
Purchase replacement if
faulty/damaged
necessary.
(Lead wires can break at the bend where they leave
the machine, giving no, or intermittent, output)
Sudden
change in
sensation
If you disconnect and Always return strength to
re-connect
a
few zero after disconnecting the
minutes
later,
the lead or pads.
signal will feel quite a
lot stronger.
If the above review has failed to resolve your problem, or to report unexpected operation or events,
call TensCare or your local dealer (address on back cover) for advice.
16. CAUTIONS & WARNINGS
If you are in any doubt about any of these warnings please consult your medical adviser.
Do not use TENS :
 If you have a heart pacemaker or have a heart rhythm problem
TENS in the direct vicinity of a pacemaker may affect some models. If
TENS is applied on the front of the neck this can
affect your heart rate.
Very strong TENS across the chest may cause an extra heartbeat.
 If you have epilepsy
TENS may affect seizure threshold.
 During the first three months of pregnancy
It is not known whether TENS may affect foetal development
 On the abdomen in the later stages of pregnancy.
Stop use immediately if you experience unexpected contractions.
 When driving, operating machinery, or similar actions needing fine control
Loose pads, damaged leads, or sudden changes in contact may cause
brief involuntary muscle movements
 To mask or relieve undiagnosed pain
may delay diagnosis of progressive condition.
Caution:
 Observe caution when using the Touch TENS at the same time as being connected to monitoring
equipment with body worn electrode pads
it may interfere with the signals being monitored.
 Strong electromagnetic fields (electrosurgery/microwave cookers/mobile phones) may affect the correct
operation of this unit—See section 21. If it appears to behave unusually, move it away from these devices.
 Use caution following recent surgical procedures
stimulation may disrupt the healing process.
Do not place electrode pads :
On broken skin
The pads could encourage infection
On skin which does not have normal sensation.
If the skin is numb, too great a strength may be used, which could result in skin inflammation.
On the front of the neck
Could cause the airway to close, giving breathing problems. May cause sudden drop in blood pressure (vasovagal response)
Over the eyesMay affect eyesight or cause headaches
Across the front of the head
Effect on patients who have had strokes or seizures is not known.
May affect sense of balance
The effects of stimulation of the brain are unknown,
Near malignant tumours
In vitro experiments have shown that electricity can promote cell growth
Do not:
Ignore any allergic reaction to the electrode pads:
If a skin irritation develops stop using TENS, and try a different type of electrode.
Alternative pads specially made for sensitive skin are available, see
section 16
Also do not :
Immerse your TENS unit in water or place it close to excessive heat.
It may cease to operate correctly
Attempt to open up the TENS unit.
This will invalidate the guarantee.
Mix old, new or different types of batteries.
Use this device leads, electrodes, and accessories other than those recommended by the manufacturer.
Performance may vary from specification
17. ACESSORIES AND SPARES
Expected Service Life
The machine will often last for more than 5 years, but is guaranteed for 2 years. Accessories (leads, pads,
and batteries) are not covered by the guarantee.
Lead life depends greatly on use. Always handle the leads with care
Pads should last 12-20 applications, depending on skin condition and humidity.
AA Alkaline Batteries should last about 48 hours continuous use.
Replacement electrode pads, new batteries and lead wires are available from your supplier or distributor
(see back cover for contact details), by mail order from TensCare, by telephone using a credit or debit card,
or through our website.
The following replacement parts may be ordered from Tenscare at www.tenscare.co.uk or +44(0) 1372
723434
E-CM5050
E-696-SS-3
L-BPT
L-CPT
B-AA
X- BC-TT
X- BH-TT
Pack of 4 50 x 50 electrode pads:
Pack of 4 50 x 50 hypoallergenic electrode pads
Lead wire for units with S/N lower than 120020001
Lead wire for units with S/N higher than 120020001
Batteries 1.5V AA batteries
Replacement belt clip
Replacement Battery Holder
18. DISPOSAL OF WASTE ELECTRICAL AND ELECTRONIC PRODUCTS (WEEE)
One of the provisions of the European Directive 2002/96/CE is that anything electrical or electronic
should not be treated as domestic waste and simply thrown away. To remind you of this Directive all
affected products are now being marked with a crossed-out wheelie bin symbol, as depicted below.
To comply with the Directive you can return your old electro-therapy unit to us for disposal. Simply print
a postage-paid PACKETPOST RETURNS label from our website www.tenscare.co.uk, attach this to an
envelope or padded bag with the unit enclosed, and post it back to us. Upon receipt we willyour old
device for components recovery and recyclinghelp to conserve the world's resources and minimise
adverse effects on the environment.
19. GUARANTEE
Your TensCare device is guaranteed for two years from the date of purchase. If a fault
develops return the unit to TensCare at the address below, together with a copy of your invoice and
details of the problem. The guarantee does not cover accessories (the batteries, electrode pads or lead
wires).
Please note that the Guarantee is invalidated if incorrect batteries have been fitted, or the unit has been
immersed in water, maltreated or tampered with.
20. TECHNICAL SPECIFICATION
Waveform
Asymmetrical Bi-Phasic
Amplitude
(over 500 ohm load)
60mA zero to peak +ve in 15 steps.
+/- 10%
Output plug
Fully shielded: touch proof
Channels
Dual Channel
Batteries
2 x AA Alkaline (Two AA batteries)
Weight
80 gms without batteries
Dimensions
115 x 53 x 27 mm
Safety Classification
Internal power
source. Type BFDesigned for
continuous use.
No special moisture protection.
Operating: Temperature range:5 to 40C Humidity:
15 to 93% RH non-condensingTemperatu
Environmental
re range: -25-+70C RH 93% non
Specifications:
condensing
Storage:
TYPE BF
EQUIPMENT
Equipment providing a
degree of protection against electric
shock, with isolated applied part.
This symbol on the unit
means “ Refer to Instructions for use”
The unit is not water resistant, and
should be protected from liquids.
NB
The electrical specifications are nominal and
subject to variation from the listed values due to normal production tolerances of at least 5%
21.
EMC PRECAUTIONS
Use special precautions regarding EMC according to the information provided below.
Other portable and mobile RF communications equipment can affect performance. Do not use when
adjacent to or stacked with other electrical equipment. Use of leads or probes other than those listed in
section 18 may affect EMC performance.

Guidance and manufacturer’s declaration - electromagnetic
emissions
The Touch TENS is intended for use in the electromagnetic environment
specified below.
The customer or the user of the Touch TENS should assure that it is used in such an
environment.
Emissions test
Complia Electromagnatic environment - guidance
nce
RF emissions
CISPR 11
Group 1
The model itouch uses RF energy only for
its internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The model itouch is suitable for used in domestic establishment and in
establishment directly connected to a low voltage power
supply network which supplies buildings used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Not
applicable
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Not
applicable
The Touch TENS is intended for use in the electromagnetic environment
specified below. The customer or the user of the Touch TENS should assure
that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance
Level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the model
itouch, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter
Conducted
RF
IEC
61000-4-6
3 Vrms
3 Vrms
150 kHz to 80 MHz
3 V/m
3 V/m
80 MHz to 2,5 GHz
Radiated RF
IEC
61000-4-3
Recommended separation distance 3V
d = 1.2 P 80M Hz to 800MHz
d = 2.3 P 800MHz to 2.5GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, should
be less than the compliance level in each frequency
range.
Interference may occur in the vicinity of equipment
marked with the symbol on the left of this box
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply
in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
model itouch is used exceeds the applicable RF compliance level above, the model itouch should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the model itouch.
b.Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The Touch TENS is intended for use in the electromagnetic environment specified
below. The customer or the user of the Touch TENS should assure that it is used in
such an environment.
Immunity test
IEC 60601 test level
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Compliance
level
±6 kV contact
±8 kV air
Electromagnetic
environment
guidance
Floors should be
wood, concrete or
ceramic tile
. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for Input/output lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
Not applicable
Not applicable
Voltage dips, short
interruptions and
voltage variations on
power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT) for 0,5 cycle
40 % UT(60 % dip in UT ) for 5 cycles
70 % UT (30 % dip in UT )for 25
cycles
<5 % UT (>95 % dip in UT ) For 5s
Not applicable
Not applicable
Power frequency
(50/60 Hz) magnetic
field. IEC 61000-4-8
3 A/m
Not applicable
Not applicable
NOTE
UT is the a.c. mains voltage prior to application of the test level
Recommended separation distances between portable and mobile RF communications
equipment and the Touch TENS
The model itouch is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the model itouch can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the model itouch as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
Separation distance according to frequency of transmitter M
output power of
transmitter W
150kHz to 80MHz
80 MHz to 800 MHz
800MHz to 2.5GHz
d=1.2 P
d=1.2 P
d=2.3 P
0.01
0.12
0.12
0.23
0.1
0.38
.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
EC Declaration of Conformity
TensCare Ltd hereby declare that an examination of the production quality assurance system has been
carried out following the requirements of the UK national legislation according to Annex V of the Directive
93/42/EEC on medical devices. We certify that the production quality system conforms with the relevant
provisions of the aforementioned legislation, and the result entitles the organization to use the CE 0473
marking on this product.
Manufactured by:
TensCare Ltd, 9 Blenheim Road,
Epsom, Surrey KT19 9BE, UK
Tel: +44(0) 1372 723434
www.tenscare.co.uk
Pub No.: I-XLY1T-UK
Version 1.1 06/13