PRODUCTS Shimadzu News 3/2003 AA-6300 for multi-element analysis with system validation Fully automatic atomic absorption spectrometer – determination of cadmium, copper and arsenic T he quantitative determination of main, transition and trace elements is part of the daily routine analyses in every control laboratory. These types of analysis are carried out presently using modern atomic absorption spectrometers, for example Shimadzu’s new AA-6300. with optimized surface. The system is therefore resistant to all known acids, organic solvents and reagents that are commonly used in atomic absorption spectrometry. Background correction Two standard background correction methods are available for reliable determination of trace elements in complex matrices in the flame- and graphite furnace mode. • The deuterium technique for correction of spectral interferences caused by molecular absorption and light scattering caused by particles • The high current pulse technique (high speed self reversal method), which can correct interferences caused by absorption line overlapping and structured background. The determination of cadmium and copper AA-6300 with GFA-EX7i and autosampler An important criterion for effective sample throughput is fully automatic operation. Current system configuration enables the determination of a maximum of 20 elements in an analysis sequence for 60 individual samples in flame or electrothermal atomisation. Very long analyses sequences, for example in unattended overnight operation or during online operation, require regular system validation during quality control of measurement and test equipment. high performance operation and is designed for trace analysis. 900 nm and consists of a CzernyTurner monochromator with a holographic grating (1800 lines/ mm). The detector system, consisting of a photomultiplier for a wavelength range of 185 - 600 nm and a Si-detector for wavelengths between 600 and 900 nm, offers For flame atomisation, the new AA-6300 is equipped with a titanium burner, a ceramic impact bead, a platinum/iridium capillary and a polypropylene spray chamber PRODUCTS Shimadzu News 3/2003 The highly efficient nebulisation unit combined with the stable optical system leads to excellent performance data of approximately 0.350 Abs for a copper solution at a concentration of 2 mg/L and a detection limit of 0.0015 mg/L. Table 1 lists the instrumental parameters for the quantitative determination of cad- mium and copper. Both sets of parameters are loaded automatically from the system and are processed sequentially. The background correction method is selected according to the element analysed: • for copper the deuterium technique is used • for cadmium, due to the known spectral interference of iron, the deuterium technique typically leads to overcompensation. Therefore, the high-speed selfreversal technique is used. The calibration curve for cadmium in a concentration range of 0.04 mg/L up to 0.1 mg/L is shown in Figure 2. High Sensitivity required! The graphite furnace mode has been used successfully for many years for the trace analysis of heavy metals, for instance arsenic, cadmium, lead and manganese. Growing environmental awareness as well as increased quality requirements and control mechanisms on our daily needs (such as drinking water) lead to almost unimaginable, constantly more refined and increasingly more accurate analytical methods. The Environmental Commission of the 0.500 0.100 0.400 All experimental work described below has been carried out on a Shimadzu AA-6300 system, which is equipped for electrothermal atomisation using a GFA-EX7i graphite furnace with fully automatic sample preparation station (Figure 1). The spectrometer operates in a wavelength range of 185 - 0.200 324.8 Slit width (nm) 1.0 0.5 Atomisation Flame Flame Lamp current D 2 BGC*(mA) – 6 Lamp current SR BGC**(mA) 8/100 – Acetylene (L/min) 1.8 1.9 Table 1: Instrumental parameters **Deuterium background correction for the determination of the elements **Self reversal background correction copper and cadmium. Element Pb As Mn Wavelength (nm) 283.3 193.7 279.5 Slit width (nm) 0.7 0.7 0.2 Atomisation Furnance Furnance Furnance Lamp current D 2 BGC*(mA) – – 10 Lamp current SR BGC**(mA) 8/300 12/400 – Ashing (°C) 800 600 800 Atomisation (°C) 2400 2200 2200 Matrix Modifier Pd (NO 3) 2 Pd (NO 3) 2 – Table 2: Instrumental parameters **Deuterium background correction for the determination of the elements **Self reversal background correction arsenic, lead and manganese. European Union decided on 11 March 1999 to again lower drastically the threshold values for heavy metals. The actual value of 50 micrograms of lead per litre was decreased to 10 micrograms per litre as lead can cause damage to the nervous system in small children and pregnant women. In 2003, the new drinking water law was ratified and water supply companies as well as regulatory agencies are now required to analyse the chemical parameters according to the new threshold values in compliance with Appendix 2 Part II of the drinking water regulation. 0.025 0.100 0.000 0.000 0.025 0.050 0.075 0.100 0.000 0.000 2.500 5.000 7.500 10.000 12.500 15.000 Conc. (ppb) Conc. (mg/L) Figure 2: Calibration curve for cadmium 2 Cu 228.8 The determination of arsenic, lead and manganese Abs. Abs. 0.300 0.050 Cd Wavelength (nm) 1000 mg/L 1000 mg/L 0.075 System requirements Element 17.500 20.000 The digital temperature control of the new GFA-EX7i graphite furnace allows optimisation of the system characteristics with high signal intensities at excellent reproducibility. The system is also intelligent enough to differen- Figure 3: Calibration curve for lead 3 PRODUCTS Shimadzu News 3/2003 AA-6300 for multi-element analysis with system validation Fully automatic atomic absorption spectrometer – determination of cadmium, copper and arsenic T he quantitative determination of main, transition and trace elements is part of the daily routine analyses in every control laboratory. These types of analysis are carried out presently using modern atomic absorption spectrometers, for example Shimadzu’s new AA-6300. with optimized surface. The system is therefore resistant to all known acids, organic solvents and reagents that are commonly used in atomic absorption spectrometry. Background correction Two standard background correction methods are available for reliable determination of trace elements in complex matrices in the flame- and graphite furnace mode. • The deuterium technique for correction of spectral interferences caused by molecular absorption and light scattering caused by particles • The high current pulse technique (high speed self reversal method), which can correct interferences caused by absorption line overlapping and structured background. The determination of cadmium and copper AA-6300 with GFA-EX7i and autosampler An important criterion for effective sample throughput is fully automatic operation. Current system configuration enables the determination of a maximum of 20 elements in an analysis sequence for 60 individual samples in flame or electrothermal atomisation. Very long analyses sequences, for example in unattended overnight operation or during online operation, require regular system validation during quality control of measurement and test equipment. high performance operation and is designed for trace analysis. 900 nm and consists of a CzernyTurner monochromator with a holographic grating (1800 lines/ mm). The detector system, consisting of a photomultiplier for a wavelength range of 185 - 600 nm and a Si-detector for wavelengths between 600 and 900 nm, offers For flame atomisation, the new AA-6300 is equipped with a titanium burner, a ceramic impact bead, a platinum/iridium capillary and a polypropylene spray chamber PRODUCTS Shimadzu News 3/2003 The highly efficient nebulisation unit combined with the stable optical system leads to excellent performance data of approximately 0.350 Abs for a copper solution at a concentration of 2 mg/L and a detection limit of 0.0015 mg/L. Table 1 lists the instrumental parameters for the quantitative determination of cad- mium and copper. Both sets of parameters are loaded automatically from the system and are processed sequentially. The background correction method is selected according to the element analysed: • for copper the deuterium technique is used • for cadmium, due to the known spectral interference of iron, the deuterium technique typically leads to overcompensation. Therefore, the high-speed selfreversal technique is used. The calibration curve for cadmium in a concentration range of 0.04 mg/L up to 0.1 mg/L is shown in Figure 2. High Sensitivity required! The graphite furnace mode has been used successfully for many years for the trace analysis of heavy metals, for instance arsenic, cadmium, lead and manganese. Growing environmental awareness as well as increased quality requirements and control mechanisms on our daily needs (such as drinking water) lead to almost unimaginable, constantly more refined and increasingly more accurate analytical methods. The Environmental Commission of the 0.500 0.100 0.400 All experimental work described below has been carried out on a Shimadzu AA-6300 system, which is equipped for electrothermal atomisation using a GFA-EX7i graphite furnace with fully automatic sample preparation station (Figure 1). The spectrometer operates in a wavelength range of 185 - 0.200 324.8 Slit width (nm) 1.0 0.5 Atomisation Flame Flame Lamp current D 2 BGC*(mA) – 6 Lamp current SR BGC**(mA) 8/100 – Acetylene (L/min) 1.8 1.9 Table 1: Instrumental parameters **Deuterium background correction for the determination of the elements **Self reversal background correction copper and cadmium. Element Pb As Mn Wavelength (nm) 283.3 193.7 279.5 Slit width (nm) 0.7 0.7 0.2 Atomisation Furnance Furnance Furnance Lamp current D 2 BGC*(mA) – – 10 Lamp current SR BGC**(mA) 8/300 12/400 – Ashing (°C) 800 600 800 Atomisation (°C) 2400 2200 2200 Matrix Modifier Pd (NO 3) 2 Pd (NO 3) 2 – Table 2: Instrumental parameters **Deuterium background correction for the determination of the elements **Self reversal background correction arsenic, lead and manganese. European Union decided on 11 March 1999 to again lower drastically the threshold values for heavy metals. The actual value of 50 micrograms of lead per litre was decreased to 10 micrograms per litre as lead can cause damage to the nervous system in small children and pregnant women. In 2003, the new drinking water law was ratified and water supply companies as well as regulatory agencies are now required to analyse the chemical parameters according to the new threshold values in compliance with Appendix 2 Part II of the drinking water regulation. 0.025 0.100 0.000 0.000 0.025 0.050 0.075 0.100 0.000 0.000 2.500 5.000 7.500 10.000 12.500 15.000 Conc. (ppb) Conc. (mg/L) Figure 2: Calibration curve for cadmium 2 Cu 228.8 The determination of arsenic, lead and manganese Abs. Abs. 0.300 0.050 Cd Wavelength (nm) 1000 mg/L 1000 mg/L 0.075 System requirements Element 17.500 20.000 The digital temperature control of the new GFA-EX7i graphite furnace allows optimisation of the system characteristics with high signal intensities at excellent reproducibility. The system is also intelligent enough to differen- Figure 3: Calibration curve for lead 3 PRODUCTS Shimadzu News 3/2003 tiate the heating rates for each set temperature, for instance for lead (slow heating rates up to approximately 800 °C and fast heating rate at 2400 °C so that the preset atomisation temperature can be reached within a fraction of a second). This leads to results as shown in Figure 3 for the calibration of lead in the measuring range of 5 mg/L up to 20 mg/L where the concentration of the new threshold value (10 mg/L) is straight within the ideal calibration range. water sample with a certified arsenic concentration of 0.72 mg/L ± 0.05 mg/L was used as control standard. The concentration measured with the GFA-EX7i was 0.73 mg/L (average value from 3 determinations) and is in excellent agreement with the certified reference value. 0.200 0.150 0.100 System validation 0.050 0.000 The temperature step between the thermic pretreatment (ashing phase) and atomisation must be as small as possible in order to reach the final temperature selected as quickly as possible and to avoid a strong expansion of the inert gas. The digital gas control ensures that the flow through the graphite tube during atomisation and also the last 3 to 5 seconds of the ashing phase can be stopped reproducibly in order to avoid premature release of the atomic cloud and to increase the residence time of the atoms in the light path. The instrumental parameters are listed in Table 2. Arsenic is another element which is significant in environmental analysis because of its toxicity. Arsenic can be determined quantitatively using AAS analysis in the hydride and graphite furnace mode in samples such as drinking water, mineral water, wastewater and sludge, each with its associated matrix problems. Modern 0.0 2.5 Figure 4: Typical absorption signal for an arsenic standard sample graphite furnace systems such as the GFA-EX7i enable reliable routine determination of an arsenic concentration of 1 mg/L, using a suitable background correction technique such as the high-speed self-reversal method. This method for the determination of arsenic using graphite furnace atomic absorption spectrometry is described in DIN 38405-35 in the German Standard Methods (Deutsche Einheitsverfahren) for routine analysis. Figure 4 shows a typical absorption signal for an arsenic standard sample at a concentration of 6 mg/L, obtained using the digital graphite furnace. The element signal (red) and the background signal (blue) of 3 repeated measurements are overlayed as peak profiles. An international reference Atomic absorption spectrometry is a relative technique for the quantification of element concentrations, so the calibration curve as the important basic requirement for accurate measurement must be obtained prior to analysis of unknown samples. The quality of the calibration can be tested during routine operation using reference materials as control standards. Regular control of the analysis sequence using control standards is a standard function of the system software. The system validation during quality control however requires the use of a software package which is independent of the system software. Parameters tested during system validation are wavelength accuracy, baseline noise and drift, as well as performance data of selected elements in the flame and graphite furnace atomisation mode and calculation of the detection limits. A basic requirement for accurate measurement is the precise setting of the element-specific wavelengths. Using a slit width of 1 nm, the profile of the selected line is registered and the energy maximum is set. Setting of the wavelength may not exceed the predetermined tolerance limits, otherwise a wavelength calibration procedure must be carried out. Testing of the wavelength in the range of 253.7 up to 640.2 nm is shown in Figure 5. The set wavelength is at 640.20 nm with a standard deviation of 0.02 nm within the tolerance limits. The first semi-micro balance with UniBloc technology Always optimally calibrated Testing of additional critical parameters Similarly, additional critical parameters can be tested, for instance the stability of the system. The drift of light sources used for element- and background absorption may not exceed predetermined threshold values in order to guarantee the high quality of the analytical data over extended periods of operation. Additionally it is necessary to carry out regular control of the performance data in absorption, i.e. testing the sensitivity of the system and to assure excellent reproducibility of repeated measurements. It is of course important to use suitable control standards. Testing of the performance data in the flame mode using a copper standard is shown in Figure 6. We will gladly send you further information. Please note the appropriate number on your reader reply card. Info 273 IMPRINT PRODUCTS Shimadzu News 3/2003 High-performance measuring cell for the fully automatic semi-micro AUW-D analytical balance series S himadzu’s AUW-D series represents a special highlight in the newest developments of analytical and semi-micro balances. The AUW-D series consists of fully automatic semi-micro/ analytical balances in the dualrange operational mode. This means that the balances can be operated up to a maximum weighing capacity of 82 g at a display accuracy of 0.01 mg and up to 220 g at an accuracy of 0.1 mg. PSC is a fully automatic calibration function of the balance that responds to changes in the ambient temperature. An internal sensor determines the temperature variations and initiates a fully automatic calibration as soon as a certain tolerance level is reached. This way, the balance is optimally calibrated at any time. In addition, the so-called ‘Clock-Cal’ function can be used. The user can predefine up to three individual times per day at which the balance is automatically calibrated, for example at the start of the working day, at noon and in the evening. This guarantees that the balance is operational at any time, even without the presence operating personal. Of course, the balance can be calibrated at any other time the user wishes – just by pushing the button. Calibration with external weights is also extremely simple. In this case a weight is placed on the balance and the user simply follows the instructions on the display. Special applications? such as a set for density determination, an external printer, footoperated switch and external keyboard for simple data entry, is optional. All balances are equipped with a serial interface, which can transfer data via a zero-modem cable directly into a Windows application. Additional software is not necessary. Shimadzu’s patentedpending ‘WindowsDirect’ function makes the use of its balances very comfortable. With respect to FDA 21 CFR Part 11 requirements, the data can be transferred directly to the CLASS Agent software, which complies with the FDA requirements. Specific software offers various options to adjust the instrument depending on the ambient conditions at the site of operation. Via options such as ‘Anti-Wind, ‘Anti-Vibration’ et cetera, the balances can be operated under non-optimal conditions. racy of 0.1 mg. They offer a weighing capacity of 120 g – max. 320 g and feature a variety of accessories. Balances in the AUW series also feature a backgroundilluminated display, which significantly enhances readability in dark areas. Depending on the model, PSC, Clock-Cal and an internal automatic calibration weight are implemented. In general, Shimadzu’s wide selection of balances offers individual solutions at an excellent price/ quality ratio. AUW-series balances are accurate up to five decimal points Shimadzu NEWS, Customer Magazin of Shimadzu Deutschland GmbH, Duisburg Publisher: Design and Production: Shimadzu Deutschland GmbH ME Werbeagentur GWA · Düsseldorf Albert-Hahn-Str. 6 -10 · 47269 Duisburg Telefon: +49 (0) 203 76 87-0 Telefax: +49 (0) 203 76 66 25 Email: [email protected] Internet: www.shimadzu.de Editorial Team: All models in the AUW-D series are equipped with the high-performance UniBloc measuring cell. This guarantees very fast equilibration times and accurate, stable results. The balances feature userfriendly functions, such as ‘Perfect Self-Calibration’ (PSC), Clock-Cal and an integrated motor-driven calibration weight. Circulation: 7.300 Copies © Copyright: Shimadzu Deutschland GmbH, Duisburg, September 2003. Uta Steeger Windows is a Trademark of Microsoft Telefon: +49 (0) 203 76 87- 410 Corporation. Ralf Weber, Adlene Berg Figure 5: Testing of the wavelength In order to handle the numerous functions that a modern balance must perform, various applications (for instance piece counting, measurement of specific densities and various units such as ct, g, mg...) are integrated into the software. Switching between applications is possible at the touch of a button. The extensive range of accessories, Analytical balances of the AU series In addition to the semi-micro balances of the dual-range series, Shimadzu also offers a wide range of classical analytical balances. The AUW/AUX and AUY series are all equipped with UniBloc technology and feature a display accu- Figure 6: Testing of the performance data in the flame mode 4 5
© Copyright 2026 Paperzz