Br. J. Anaesth. (1981), 53, 777 CORRESPONDENCE CALIBRATION OF PEAK FLOW METERS Sir,—Fisher and Shaw (1980) criticize the standard and MiniWright Peak Flow Meters (PFM) for not reading correctly when tested with their flow and volume calibrator (Shaw, Fisher and Hughes, 1979). Had they read the paper by the late Dr C. B. McKerrow and myself (Wright and McKerrow, 1959) more carefully, they would have seen that their criticism is unjustified. We did not claim that the PFM measured air flow correctly in absolute units, but that it estimated, with sufficient precision for clinical purposes, the Peak Expiratory Flow (PEF) that the subject would produce when tested with an ideal instrument of negligible resistance and inertia. We obtained a calibration curve for this purpose by asking 20 subjects with widely varying PEF to blow alternately into a pneumotachograph and the PFM. This laborious and rather imprecise procedure was adopted because it was not technically possible to connect the two instruments in series. It was not clearly realized at the time that this is the only valid and satisfactory procedure. All portable, and most non-portable PFM have appreciable resistance or inertia, or both, and therefore modify the PEF to a varying extent. If they were calibrated to read the actual flow through them, the PEF of an individual would depend on the instrument used, which would be most unsatisfactory. As it is, any kind of PFM can be used to estimate the "ideal" PEF provided that it has been calibrated, either directly against an "ideal" instrument such as we used, or against an already calibrated instrument such as the Wright PFM. The latter procedure was adopted when calibrating the Minimeter (Wright, 1978; Oldham, Bcvan and McDermott, 1979; Perks ct al., 1979) and makes it possible to substitute the Mini for the Standard PFM without any confusion or appreciable loss of accuracy. I do not, however, claim that our original calibration of the PFM was absolutely correct and it i&open to anyone to repeat it. There are two sources of doubt. First, 20 subjects is a niggardly number on which to base the calibration of an instrument in world-wide use and a recalibration using a much larger number of subjects and a wider range of PEF, especially at the lower end of the scale, is long overdue. Second, the pneumotachograph we used had a resistance of only 15 mm H 2 O at 700 litre min "', which was considered to be negligible compared with 120mmH 2 O for the PFM, but its steady flow calibration was carried out with a 1000-litre min "' rotameter which had been calibrated on water and not on air. Fisher and Shaw's method could be used to obtain absolute calibration on air for flows up to 1000 litre min1 and I hope this will be done. If, as a result, our original calibration is found to have been seriously in error the Wright PFM will be adjusted accordingly. It must be remembered, however, that the estimate of PEF with secondary instruments is an inherently imprecise procedure. The resistance of the PFM modifies the PEF in different and unpredictable ways with different people. For example, the much higher resistance of the Minimeter does not usually reduce the PEF as much as would be expected, presumably because of the flexibility of the airways, whereas it has a marked effect on a rigid mechanical flow generator of the type used by Fisher and Shaw. In addition, low PEF cannot often be repeated because they exhaust the subject, so that calibration in this range will always be uncertain. For the time being, therefore, and perhaps for ever, the Wright PFM scales will continue to be calculated as they are now. B. M. WRIGHT Harrow, Middx REFERENCES Fisher, J., and Shaw, A. (1980). Calibration of some Wright peak flow meters. Br. J. Anaesth., 52, 461. Oldham, H. G., Bevan, M. M., and McDermott, M. (1979). Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter. Thorax, 34, 807. Perks, W. H., Tarns, I. P., Thompson, D. A., and Prowse, K. (1979). An evaluation of the mini-Wright peak flow meter. Thorax, 34, 79. Shaw, A., Fisher, J., and Hughes, D. W. (1979). A flow and volume calibrator for respiratory measuring equipment. J. Mtd. Eng. Technol., 3, 248. Wright, B. M. (1978). A miniature Wright peak-flow meter. Br. Mtd. J., 2, 1627. McKerrow, C. B. (1959). Maximum forced expiratory flow rate as a measure of ventilatory capacity. Br. Mtd. J., 2, 1041. Sir,—We can assure Dr Wright that we did read his paper most carefully, but our paper was devoted to a constructive appraisal of meters in present use. However, Dr Wright has raised some points which should be clarified. Meters are calibrated when their readings are compared with a known standard under the same measuring conditions. The tests described by Dr Wright only compare an individual meter with another instrument the dynamic calibration of which is also unknown. As each subject may not produce identical flows when blowing into the meter and reference instrument, standardization of measuring conditions is not achieved. Even if a large number of subjects are used, all that can be said is that the two instruments appear to give similar overall performances for that particular group. If healthy subjects are studied, it cannot be assumed that the correlation still holds good for those with respiratory deficiencies. Also, substantial discrepancies in performance may go undetected, for example at low flows. There is no substitution for proper calibration, for this enables an instrument's performance to be defined in terms of accuracy and repeatability. We cannot comment on the performance of the particular meters referred to in Wright's original paper. However, we did calibrate three of the original styled peak flow meters. For these meters the errors were fairly modest, being ± 10% for flows greater than 1501itremin"'. These errors were less than the errors for the 10 restyled and six mini-meters reported by us. This suggests a deterioration in the calibration over the years. What matters is the performance of currently produced meters. Our calibration allows a proper evaluation of this to be achieved. Dr Wright suggests that it is all right to disguise the modifying effect of high resistance meters by alteration of the scale. This only leads to confusion and will prevent progress in our understanding of the important relationship between BRITISH JOURNAL OF ANAESTHESIA 778 pressure and flow during the forced expiratory manoeuvre. It also breaks a fundamental law of the science of measurement: that units of measurement such as litre per minute should have an absolute definition and meaning. Finally, Dr Wright states that the resistance of the meter has a marked effect on arigidmechanicalflowgenerator of the type used by us. This is untrue, for our calibration is independent of the resistance of the meter under test. T _ J. FISHER A. SKA1.1/ Glagow DOES KETAMINE METABOLITE II EXIST IN METABOLITE HI NH-CH 3 R NH, r°KETAMINE /* NH, r I^ • " O H R NH, CONJUGATED R N Duration of anaesthesia (min) Man VlVOi Sir,—In their review of the pharmacological properties of ketamine, Chang and Glazko (1974) postulated a tentative scheme for the biotransformation of this interesting anaesthetic agent (fig. 1). An N-demethylation, yielding nor-ketamine (MI) followed by a hydroxylation of the cyclohexane ring at two optional positions yields hydroxy-nor-ketamine (Mill and MIV, respectively). These two metabolites arc then conjugated and excreted or dehydrated to dehydro-nor-ketamine (Mil). R^ TABLE I. Concentrations (\wiollitre ') of ketamine, MI and Mil in simultaneously collected samples ofplasma and c.s.f. in five rats and one man during steady-state ketamine anaesthesia. n.d. = not detectable m NH, METABOLITE I Ketamine Ccrf 20 60 180 2.8 5.6 6.5 15 20 303 MI Mil c. c« CP c« cr 4.5 9.4 0.2 0.9 1.8 0.6 n.d. n.d. n.d. 0.2 0.5 1.2 1.5 1.6 8.4 4.4 3.8 n.d. n.d. n.d. n.d. n.d. n.d. 11.0 1.9 3.4 Rat 1 2 3 4 5 70 90 120 21.3 32.1 33.8 27.5 44.4 41.3 48.6 52.5 43.8 13.1 12.7 15.6 25.0 22.5 0.5 7.5 12.0 14.2 diffuse slowly or not at all into the c.s.f., we believe there is a strong argument for considering Mil as an artefact. If Mil does not exist in vivo, there is no obvious reason to postulate the presence of MVII. If ketamine administration does not result in Mil or MVII in vivo, man has never been exposed to these compounds unless given them directly. Consequently, during future research in this field, administration of Mil and MVII to man should be carried out with extreme care. P. STENBERG METABOUTE I T OH J . IDVALL CONJUGATED IV METABOLITE IV FIG. 1. Postulated biotransformation of ketamine according to Chang and Glazko (1974). Following the work of Chang and Glazko, no further basic research on this topic has appeared in the literature. Their postulated scheme of biotransformation has been adopted and reappears, although enlarged, in a recent update by Zsigmond and Domino (1980). These authors postulated the presence of another dehydrated metabolite, dehydroketamine (MVII). In our opinion, there are several reasons to question the in vivo existence of Mil. Using the bioassay of Chang and Glazko (1972), we were not able to detect the presence of Mil in cerebrospinal fluid, while simultaneously collected plasma samples (man or rat) contained easily determined concentrations of this compound (table I). These experiments included collections following prolonged anaesthesia times. In contrast, ketamine and MI were easily determined in c.s.f., with concentrations amounting to approximately 30-60% of those seen in plasma even within 15—20min after induction of anaesthesia. Since there is no obvious reason to believe that Mil should be less lipophilic or more strongly bound to plasma proteins than ketamine and MI, other explanations for this apparent "inhibited diffusibility" of Mil across the blood-brain barrier have to be considered. The bioassay utilizes gas-liquid chromatography and includes benzene extraction of a strongly alkalized sample and subsequent derivatization with heptafluorobutyric anhydride in the presence of pyridine. Since alpha and betahydroxyketoncssuch as Mill and MIV respectively (and possibly also their conjugates) would be expected to be converted easily to Mil under the conditions described, and since there are reasons to believe that these less lipophilic compounds would Malmo, Sweden REFERENCES Chang, T., and Glazko, A. J. (1972). A gas chromatographic assay for ketamine in human plasma. Anesthesiology, 36,401. (1974). Biotransformation and disposition of ketaminc. Int. Anesthesiol. Clin., 12, 157. Zsigmond, E. K., and Domino, E. F. (1980). Clinical pharmacology and current clinical uses of ketamine. 7th World Congress of Anaesthcsiologists, Hamburg. INHALATION OF GASTRIC CONTENTS Sir,—We report two cases, one fatal, of aspiration pneumonias following regurgitanon and inhalation of gastric content during general anaesthesia for emergency evacuation of an incomplete abortion. A 24-yr-old African female of average size was admitted at 09.00 h with a septic incomplete abortion. She received papaveretum 15mg and atropine 0.6 mg i.m. at 14.00 h. The haemoglobin concentration was 10.5gdl~'. She had starved for 15h. At 14.30h the patient inhaled pure oxygen for 3min through a Maplcson A system. Anaesthesia was induced at 14.40 h with Althesin 3 ml and diazepam lOmg i.v. The patient developed severe laryngospasm. Suxamethonium lOOmg was injected i.v. to facilitate trachea! intubation. On direct larynoscopy, approximately 15 ml of clear fluid was seen in the pharynx. After suction, a size 8 mm cuffed trachea] tube was inserted and the cuff inflated. Endotracheal suction before ventilation of the lungs revealed no obvious evidence of airway contamination. Anaesthesia was maintained with 50% nitrous oxide in oxygen with manual ventilation of the lungs; supplementary suxamethonium was given as required. Steroids were not administered and neither bronchoscopy nor bronchial toilet was performed.
© Copyright 2026 Paperzz