Magnetic Force Microscopy TN00031, R1 Preliminary steps This technical note is an guide to MFM imaging with the EasyScan 2 and Mobile S AFM systems. Requirements Equipment • EasyScan 2 with extension and dynamic modules, or Mobile S AFM • Software version 1.5 or higher • Magnetic Probe: Nanoworld MFMR or Applied Nanotech MAGT • Strong permanent magnets Knowledge • Dynamic mode AFM Theoretical Background The aim of Magnetic force microscopy (MFM) is to image a spatial distribution of a magnetic field. Usual samples for MFM can be magnetic tapes, hard disks and magneto-optical disks. FT = n ( n FM ) FT n Cantilever FM Sample Forces in MFM: Force gradient of a point dipole. When performing MFM measurements the tip shouldn't come into mechanical contact with the surface as the influence of the mechanical contact is usually stronger. Therefore during MFM the tip is scanned at a defined distance above the sample. The magnetic force which is measured in MFM can be approximated by the force gradient of a point dipole seen by the magnetic tip (see figure Forces in MFM (p.1)), where n is the unit vector normal to the cantilever plane and FM the magnetic force When the AFM is operated in dynamic mode, e. g. when the tip is vibrated in z direction (vertically to the sample) during the scanning, the influence of the magnetic force in this direction is dominant and thus determines the contrast the MFM image (see figure MFM in Dynamic Mode (p.2)). magnetic field lines FM Sample Contrast MFM in Dynamic Mode: MFM imaging in dynamic mode. The magnetic filed influences the spring constant of the cantilever probe in the following way: k o = k – F T' where k is the original spring constant. Therefore a force gradient pointing away from the tip decreases the spring constant and a force gradient pointing towards the tip increases the spring constant. Like in dynamic mode, this can be related to a decrease in topography for the lower spring constant and an increase in height for a higher spring constant.he effect of the magnetic force on the amplitude and phase signal are also comparable to the dynamic mode topography imaging. For the amplitude the change of the spring constant induces a shift of the resonance peak and the difference in PAGE 2 OF 11 TN00031 © 2006 NANOSURF AG the amplitude is thereby related to the magnitude of the magnetic field (see figure Amplitude and Phase Shift (p.3)). The same shift is also observed in the phase signal (see figure Amplitude and Phase Shift (p.3)). The advantage of the phase signal over the amplitude signal is its local linearity and the thus simpler and linear relation between the force and the phase shift: Q dϕ = ---- F T' k where Q is the Q-factor of the cantilever probe. f0' f0' f0 shift phase shift attenution frequency Phase shift Amplitude f0 frequency bla Amplitude and Phase Shift: Left: The magnetic tip sample interaction influences the spring constant of the cantilever probe ant provokes a shift in the resonance frequency. The resonance shift can be detected in the amplitude signal. Right: The magnetic tip sample interaction influences the spring constant of the cantilever probe and provokes a shift in the resonance frequency. This shift is related to the phase shift. Technical Implementation The most important change compared to the standard dynamic mode is that the probe needs to be scanned at a defined distance above the surface. This has been implemented in the following way: As shown in figure Implementation (p.4) at every line start (forward and backward scan) a spectroscopy is performed. Based on that the distance at which the probe will be scanned is calibrated. The probe is then lifted at the chosen height and linearly scanned over the sample surface. The feedback is switched off. TN00031 © 2006 NANOSURF AG PAGE 3 OF 11 y x linear line scan displacement from sample z sample Implementation: Linear scan at a given height over the sample surface. Caution Like in spectroscopy, the distance is always seen from the tip, i.e. a negative value means a displacement upwards as the tip is then moving away (backwards) from he surface. Entering a positive value could crash the tip into the sample. Preparation for the Measurement • Mount the MFM Probe on the AFM. Usually new MFM cantilevers are not magnetised. Thus we need to magnetise the tip with a strong permanent magnet as show in figure Magnetisation of the Tip (p.5): • Determine the north-pole of the permanent magnet (for example using a compass). • Mark the north-pole, in order to be able to magnetise the tip in a reproducible direction. • Bring the pole of the magnet to within 1-2 mm distance of the tip. • Move the magnet away from the tip in a direction perpendicular to the cantilever surface. • Put the scan head on the small sample stage. • Create a new cantilever type according to the specifications of the manufacturer (see Software reference). • Select cantilever type in the software. PAGE 4 OF 11 TN00031 © 2006 NANOSURF AG N Magnetisation of the Tip: • Select the phase contrast imaging mode. • Select the advanced user interface mode. • Create the chart arrangement in the image window as described in figure MFM Chart Arrangement (p.5). Phase (Filter:line fit) Topography (Filter: raw data) Amplitude (Filter: raw data) MFM Chart Arrangement: TN00031 © 2006 NANOSURF AG PAGE 5 OF 11 Measurement In this part you will find a step by step recipe. Approaching the Surface • Set the free vibration amplitude to 200mV. • Clear the check box ‘Auto. start imaging’ in the positioning panel as shown in figure Auto Start (p.6). A direct start could harm the tip. Auto Start: Stopping the auto. start scanning in the positioning window. • Clear the check box ‘Enable Constant Height mode’ as shown in figure Constant Height Mode (p.7), i.e. that the constant height mode is switched off. Otherwise the feedback is disabled and the tip could be harmed. • Approach the surface. PAGE 6 OF 11 TN00031 © 2006 NANOSURF AG Constant Height Mode: Switching off the constant height mode in the imaging window. Calibrating the Vibration Amplitude The calibration value of the vibration amplitude depends on the cantilever type, because of changes in the length and resonance frequency of the cantilever. As it is important to know the vibration amplitude it needs to be calibrated. This needs to be done only when a cantilever was changed. Use the following procedure to calibrate the vibration amplitude: • Measure a vibration amplitude vs. distance curve in spectroscopy mode as shown in figure Amplitude Calibration (p.8). • Determine the slope with the length tool and calculate the m / V calibration value. TN00031 © 2006 NANOSURF AG PAGE 7 OF 11 Amplitude Calibration: • Retract the tip from the surface again. • Set the free vibration amplitude to a value in order of the magnetic feature size you intend to measure. An example is given in the chapter Example Measurement and Hints (p.10). Caution Only change the amplitude calibration value when the tip is far away from the sample. Otherwise you could break the tip. Starting the Measurement • Coarse approach close to the sample again. • Ensure that the check box ‘Auto. start scanning’ is still cleared. • Approach the surface. • Start imaging the surface with a small range (1-2µm).This prevents you from harming the tip. As the scan at a defined distance above the surface is linear, we need to correct the tilt of the sample in order to avoid a tip crash. • Correct the tilt angle for the 0° and 90° scan angle. Caution Only scan a few topography lines to correct the tilt. Scanning to long in contact might harm the magnetic layer and demagnetize the tip. PAGE 8 OF 11 TN00031 © 2006 NANOSURF AG Hint If you have the scripting interface option installed you can also correct the tilt with the ‘Correct XY-Slope’ function in the script menu. • Set the desired scan distance in the ‘Relative Tip-Position’ as shown in figure Distance from Sample (p.9) at a safe distance (>=500nm) immediately after the tilt correction. Distance from Sample: Attention A negative value means a displacement away from the surface. A positive value will crash the tip into the sample. • Activate the constant height mode by checking the box ‘Enable Constant Height Mode’ shown in figure Constant Height Mode (p.7). • Decrease the distance manually until you reach the optimum contrast, an example is given in the chapter Example Measurement and Hints (p.10). TN00031 © 2006 NANOSURF AG PAGE 9 OF 11 Example Measurement and Hints The following examples were measured on a hard disk sample taken from a defect hard disk. This was a Western Digital WD100BB, 10GB, with only a single read-write head. Selecting the Vibration Amplitude The magnitude of the MFM (phase) signal, the phase noise and the resolution of the MFM signal depend on the vibration amplitude. When the amplitude is too small, the noise is too high, when the amplitude is too large, the resolution becomes low, and ultimately the signal will also become too small. It can be shown that the vibration amplitude should be of a similar magnitude as the smallest details you wish to see. The accurate vibration amplitude can be calculated by multiplying the previously measured V/m value from the spectroscopy with the mean feature size. In our the measurement shown in figure Vibration Amplitude (p.10), the mean feature size is 80nm, the amplitude calibration gave 5 mV/nm, therefore the corresponding vibration amplitude is mV amplitude = 80nm ⋅ 5 -------- = 400mV . nm Vibration Amplitude: Left: This image was recorded with 30nm (150mV) vibration amplitude and shows a higher resolution and a lower contrast compared to the image on the right. Right: This image was recorded with 80nm (400mV) vibration amplitude and shows a lower resolution but a higher contrast compared to the image on the left. PAGE 10 OF 11 TN00031 © 2006 NANOSURF AG Selecting the Relative Tip-Position In Order to achieve the best contrast, the distance to the sample must be as small as possible. At a certain point, however the tip will touch the surface and will be harmed when the contact is to strong. A safe procedure as illustrated in figure Scanning Distance (p.11) is to start from a safe tip to sample distance (e.g. -500nm) and reduce it step wise until the tip begins to feel the surface. Immediately increase the distance until the topography features disappear from the phase signal. Scanning Distance: Top Left: Magnetic contrast increased when the distance to the sample was decreased. Top Right: When the surface topography gets visible in the phase scan, you are probably scanning to near. Bottom Left: Here the topography and some dirt particles are visible in the phase scan. Also here the distance was to small. Bottom Right: Topography of the HD sample. ‘NANOSURF’ AND THE NANOSURF LOGO ARE TRADEMARKS OF NANOSURF AG, REGISTERED AND/OR OTHERWISE PROTECTED IN VARIOUS COUNTRIES. © 2006 NANOSURF AG, SWITZERLAND, TN00031, R1 PAGE 11 OF 11
© Copyright 2026 Paperzz