MICRON FORCE CALIPER MICROTECH_150 -

PRECISION
FORCE MICRON CALIPER
MICROTECH
Patent #100613
Serial. №
Quality management system
ISO 9001:2008 (cert. № U227417)
www.microtech.ua
1 MAIN SPECIFICATIONS
Item №
Jaws length, mm
Range, for external
for internal
mm measurement measurements
s
141099192 0-150
38
16
Recommended
force
N
Max error,
mm
8±2
±0,005
Operating conditions from +10 ºС to +35 ºС
2 BUTTON FUNCTIONS
-MODE
- SET
2s for mm/inch
2 s for ZERO 4s for SLEEP
3.OPERATION INSTRUCTIONS
3.1 Wipe with a clean cloth, soaked in gasoline, measuring surface of the frame
and gauge calipers to remove anti-corrosion oil. Then wipe them with a clean dry
cloth.
3.2 If necessary, open the battery cover; insert the battery (type CR2032)
according to the polarity of the electrodes. Blinking display information or absence
suggests replacing battery.
3.3 Check the zero setting of the caliper. Sum the measuring jaws to contact with
each other. Force is generated with a measuring force indication. To create the
recommended force (8N) by the block adjustment and press SET button.
PROCESS OF ZERO SETTING AND MEASURING
You can see on MICROTECH youtube channel.
https://www.youtube.com/user/MicrotechTool
3.4
During the measurement, measuring jaws should to sum to the measured
object without knocking.
3.5
During the measurement avoid warps of measuring surfaces of the
instrument. Measuring surface must be fully in contact with the measurement
object.
3.6 Optimal measuring force for this size caliper 8±2N. During a
measuring process control optimal measuring force on Measuring force indication
window.
3.7 After finishing work wipe the measuring surfaces of the caliper with a
cloth soaked in gasoline and apply anticorrosion oil.
4 DATA OUPUT
MICROTECH caliper has data output 4Pin socket with RS-232 data output
signal. USB cable for connection and SPC MICROTECH Software available for order.
5 WARRIANTY
MICROTECH guarantees within the warranty period of 12 months.
Detailed information is available at: www.microtech-ua.com/warranty
Director MICROTECH, Ph.d. _______________/Boris Kramarenko/
СALIBRATION PROTOCOL
Item No______________
Measuring range ______mm
Reference temperature __C
Serial No__________
Resolution 0,001 mm
Calibration date ______
Nominal
mm
0.000
13.000
21.200
41.300
51.400
60.000
71.500
90.000
External
101.600
measurement
126.800
131.400
150.000
175.000
200.000
243.500
250.000
281.200
300.000
D 16.020
Internal
measurement
D 100.000
20,000
Depth measuring
Parallelism of measuring surfaces
Flatness of measuring surfaces
Results
Measured
mm
Roughness of measuring surfaces Ra<0.32
Head of the MICROTECH calibration lab.
Max. error
mm
μm
_____ /Tatyana Guz/