http://informahealthcare.com/jmt ISSN: 0309-1902 (print), 1464-522X (electronic) J Med Eng Technol, 2014; 38(1): 1–4 ! 2014 Informa UK Ltd. DOI: 10.3109/03091902.2013.844207 INNOVATION MR-based measurement of spinal cord motion during flexion of the spine: implications for intradural spinal cord stimulator systems S. Viljoen1, C. A. Smittkamp2, B. D. Dalm1, S. Wilson1, C. G. Reddy1, G. T. Gillies3, and M. A. Howard III*1 Department of Neurosurgery, 2Department of Radiology, University of Iowa Hospitals and Clinics, Iowa City, IO 52242, USA, and 3Department of Mechanical and Aerospace Engineering, University of Virginia, Charlottesville, VA 22904, USA Abstract Keywords This study develops a means of delivering electrical stimuli directly to the pial surface of the spinal cord for treatment of intractable pain. This intradural implant must remain in direct contact with the cord as it moves within the spinal canal. Therefore, magnetic resonance imaging was used to measure the movement of the spinal cord between neutral and flexedback positions in a series of volunteers (n ¼ 16). Following flexion of the back, the mean change in the pedicle-to-spinal cord dorsal root entry zone distance at the T10-11 level was (8.5 6.0) mm, i.e. a 71% variation in the range of rostral-caudal movement of the spinal cord across all patients. There will be a large spectrum of spinal cord strains associated with this observed range of rostral-caudal motions, thus calling for suitable axial compliance within the electrode bearing portion of the intradural implant. Dorsal root entry zone, intradural implants, neuromodulation, spinal cord, spinal cord stimulator 1. Introduction and background A limiting factor in the efficacy of epidural spinal cord stimulation for the treatment of intractable pain is the inability of the standard devices to selectively modulate the targeted neural fibres without inadvertent stimulation of neighbouring non-targeted structures, e.g. the dorsal nerve rootlets. This is a consequence of the shunting effects of the relatively high conductivity CSF located between the epidural electrodes and the spinal cord surface, migration of the leads following implantation, the formation of epidural scar tissue and several other such problems. The result is limited therapeutic efficacy in up to half of all patients having a standard epidural stimulator [1]. The Human Spinal Cord Modulation System (HSCMS) is being developed to improve this situation. It is an intradural device [2] placed via durotomy directly onto the pial surface of the spinal cord, thus avoiding the electrical shunting effects of the CSF and lead migration issues and enabling more focused delivery of the stimulus currents into the targeted regions [3]. The design features of the wired version of the implanted component of this device are shown in Figure 1. The electrode bearing surface at the bottom of the device makes contact with the dorsal surface of the spinal cord. It consists of a thin membrane (0.6 mm thick) of silicone into which are embedded six stimulating electrodes. The individual leads from the electrodes are configured in loops which join to form a single lead cable that exits the spinal canal. This cable is held in a stable position relative to the spinal canal by a *Corresponding author. Email: [email protected] History Received 24 April 2013 Revised 3 September 2013 Accepted 8 September 2013 titanium strap (termed the Oya Strap) that is secured to the lateral margins of the boney spinal canal. The dura is closed underneath the Oya Strap, forming a water-tight seal that prevents CSF leak. The restoring forces in the compressed lead loops inside the subdural space create a gentle pressure on the electrode membrane [4] (515 mm Hg, which is within the range of normal intrathecal pressures) to keep it in contact with the spinal cord as it moves within the spinal canal. It is important that this membrane remains stably in place and does not lift off the surface during rostral-caudal displacement of the cord or slide laterally from its midline position, the latter to prevent potentially harmful contact with the dorsal root entry zones (DREZ). Therefore, the lead loops of the device must be designed such that they accommodate all of the relative motion between the implant on the spinal cord surface and the lead exit point which is at a fixed position at the dural exit site under the Oya strap. Moreover, the membrane must conform to the shape of the dorsal surface of the spinal cord in order for the electrodes to make uniform direct contact with the pial surface. This means that the radius of curvature of the membrane must closely approximate that of the oval-shaped spinal cord. It should also subtend as large an angle as possible of the dorsal arc, while leaving a small buffer zone on either side that separates the lateral edges of the membrane from the dorsal rootlets. These requirements call for knowledge of (1) the spinal cord’s thoracic morphology and (2) its axial motion within the spinal canal as a function of spinal flexion. The first point was addressed via direct image-based determination of the inter-DREZ dorsal arc lengths found 20 13 J Med Eng Technol Downloaded from informahealthcare.com by Miss Sandy Dalgleish on 12/16/13 For personal use only. 1 J Med Eng Technol Downloaded from informahealthcare.com by Miss Sandy Dalgleish on 12/16/13 For personal use only. 2 S. Viljoen et al. Figure 1. Artist’s rendering of the HSCMS device. The 1.5 cm long electrode-bearing silicone membrane at the bottom of the device is placed onto the pial surface of the spinal cord. The leads from the electrodes loop upwards, three per side, to meet and exit as a bundle through the titanium Oya Strap which serves as the fixation structure for the device, attaching it to the surrounding vertebra. A dural seal (not shown) encloses the lead loops and membrane within the dura and allows for leak-free passage of the lead bundle through the exit port on the Oya Strap. An implantable pulse generator (not shown) is located in a nearby subcutaneous pocket. The lead bundle connects to it and conveys the electrical stimuli to the electrodes. across a large number (n ¼ 50) of patients [5]. The present study addresses the second point through implementation of an experimental protocol aimed at measuring the flexiondriven rostral-caudal movement of the spinal cord. 2. Materials and methods 2.1. Patient position within the magnet In order to allow for spinal flexion during imaging, a 1.5 T Magnetom Espree (Siemens, Erlangen, Germany) magnet was utilized to take advantage of its larger bore size, which was designed to accommodate bariatric patients. Informed consent was obtained from a total of 16 healthy volunteers ranging in age from 23–58. Each volunteer was first imaged in a supine neutral position and then imaged in a maximal attainable flexed position. To obtain the maximal flexion of the spine, patients were given three basic positioning instructions. The first was to rotate their hips/pelvis backwards towards the gantry as far as possible to remove the lumbar lordosis and straighten the lumbar spine. The second was to curl their upper back, neck and head forward so that their shoulders were as close to their knees as possible. The third instruction was then to tuck their chin down as close to their chest as possible. While attaining this flexed position in the bore, a variety of foam wedges and pillows were utilized for added support so that the patient could remain as still as possible during image acquisition. Maximal flexion was limited by volunteer flexibility in 14 of the patients. In only two patients was flexion limited by MR bore size. Not surprisingly, these were also the two tallest patients with heights of 2.03 and 1.98 m (60 800 and 60 600 ). 2.2. Imaging protocol Each volunteer had a vitamin E capsule taped to their midline lower thoracic spine for help in level localization. A sagittal HASTE sequence was performed initially as a localizer both for vertebral level counting and identification of a more focal field of view centred over the region of lowest thoracic spinal nerves and the conus medullaris. In order to acquire anatomic images with enough resolution to accurately measure intervals J Med Eng Technol, 2014; 38(1): 1–4 between spinal nerve dorsal root entry zones, a CISS sequence was selected for its high spatial resolution. Although this is a highly T2-weighted sequence, acquisition time still required 2 min 5 s. This length of time initially caused too much motion degradation during flexed imaging to make accurate measurements. The use of the pillows and foam wedges provided just enough support for volunteers to remain still and in a state of quiet respiration for the required 2 min duration. This minimized breathing-related artifacts which, like the residual motions induced by the cardiac cycle, otherwise limit the imaging resolution that can be obtained. Hence, the measurements we have are technically an average during quiet respiration. All neutral and flexed sequences were obtained utilizing TR ¼ 4.35 ms, TE ¼ 2.18 ms, slice thickness ¼ 0.8 mm, matrix size ¼ 192 192, one acquisition per average, 192 phase encoding steps, field of view ¼ 200 mm and a 70 flip angle. 2.3. Measurement algorithm Imaging was obtained in the coronal plane. Three-dimensional multiplanar reconstruction software was utilized on a Carestream PACS station to aid in measurement. The T10 and T11 nerve roots were identified. A cranial caudal measurement was made in a plane parallel to the spinal canal between the DREZ of T10 and T11. (The exact position of the entry zones was confirmed by assessing sequential axial images to identify the most cranial aspect of the nerve originating from the spinal cord.) As shown in Figure 2, the difference between this measurement on the neutral and flexed images is a measure of spinal cord contraction/expansion along the rostral-caudal axis. Next, a cranial caudal measurement was made from the DREZ of the T10 nerve root along the same plane as the prior measurement, to the level of a plane orthogonal to the spinal canal at the level of the inferior T10 pedicles. The latter were selected as a reference point of the bony canal inside of which the spinal cord moves. The difference between these measurements represents cord movement within the bony canal. An example of a coronal image on which the relevant anatomical features are identified is shown in Figure 3. Lastly, a cranial caudal measurement of the change in conus tip position was made. In order to accurately accomplish this, the position of the conus tip was first identified on the neutral images with reference to a landmark within the bony spinal canal at the same cranial-caudal level. This landmark was then identified on flexed imaging and a cranial caudal measurement was made from that level to the level of the new conus position. This measurement also represents movement of the spinal cord within the canal. 3. Results The spinal cord should move rostrally during flexion and should lie in its most caudal location when the patient is in the neutral position. The measured change in the pedicle-to-spinal cord DREZ distance across all patients between the neutral and flexion positions ranged from 1.9–18.0 mm, with a mean and standard deviation of 8.5 6.0 mm. The inter-DREZ distance across all patients between the neutral and flexion positions ranged from 2.0–6.7 mm, with a mean and Measurement of spinal cord motion DOI: 10.3109/03091902.2013.844207 3 mean) are very large, 71%, 74% and 64%, respectively. This reflects the wide variability in the capacity of individual subjects to maximally flex the spine, as well as possible intersubject variability in spinal cord mechanical characteristics. J Med Eng Technol Downloaded from informahealthcare.com by Miss Sandy Dalgleish on 12/16/13 For personal use only. 4. Discussion Figure 2. The white line running transversely is at the base of the pedicle. The yellow vertical line (on the right) represents the distance from there to the DREZ. The change in that length represents how much the cord slides during flexed and neutral positions. The green vertical line (on the left) indicates the inter-DREZ distance and represents how much the cord stretches. Figure 3. Coronal image on which the nerve-root sheath exit zones are identified by orange arrows (dark arrows, three on each side), the pedicle locations are marked by blue arrows (light arrows, three on the left side), and two of the pedicle cross-sections are outlined in blue (the two light traces on the right). standard deviation of 3.5 2.6 mm. The mean and standard deviation for the rostral-caudal conus movement was found to be 6.4 4.1 mm within an overall range of 1.1–11.4 mm. The fractional variations in these findings (standard deviation/ Several previous studies have reported both gross anatomical and image-based measurements of spinal cord motions and related mechanical parameters, some of which are relevant to the HSCMS design issues discussed above. For instance, Figley and Stroman [6] found that the amplitude of the cardiac-driven pulsation of the anterior-posterior (A/P) component of bulk cord movement in 10 immobile patients was 0.60 0.34 mm in the cervical and upper thoracic region. However, as found in their subsequent study on eight immobile patients [7], this value was 50.10 mm in the lower thoracic region (T4/T5 and below), with excursions of the same size along the left/right axis, thus setting the lower limits on the size of bulk cord motions in those directions. Much larger changes (millimetre-scale) in the A/P position of the spinal cord relative to the dura were found in patients who purposely changed posture or carried out flexion/extension manoeuvres during imaging [8,9]. See also the reviews of Harrison et al. [10] and Cox [11]. Furthermore, it has long been known that flexion and extension of the spine produces centimetre-scale axial stretching motions of the spinal cord and dura along the rostral-caudal (R/C) direction relative to the other structures forming the spinal canal [12]. However, that work was carried out on fixed-tissue samples from cadavers and there are no reported equivalent series of in vivo measurements that use a similar protocol. While all of these studies have provided very useful background information, an effort focusing specifically on determination of the axial component of spinal cord motion within the region targeted for HSCMS implantation was needed to complete the picture. Moreover, from a biophysical perspective, the findings presented here can also be used to yield the ratio of the spinal cord’s mean stretch-to-mean axial movement over a full flexion cycle: 3.5 mm/8.5 mm &40%. That is, on average across all patients, it required 1 mm of net axial displacement of the cord to stretch it 0.4 mm in length. However, as discussed above, there were large variations between patients and this makes it difficult to draw generalized conclusions about the mechanics of the spinal cord in relation to certain aspects of its interaction with the HSCMS. For instance, estimates of the net strain within its parenchyma will depend strongly on the assumptions made regarding how the spinal cord’s stretch is distributed over its length, as that will be governed at least in part by the anisotropies in its structure in a given patient. In the worst case, for some patients, high levels of parenchymal strain could mean that the HSCMS electrodes might slide across the spinal cord’s surface unless the membrane is compliant enough to accommodate this component of the cord’s axial motion. On the other hand, in those cases where the stretching component is more localized within a given region of the typically 45 cm long cord, the strain would likely be smaller, foreseeably 1% or less, thus resulting in little relative motion between the HSCMS electrodes and the spinal cord. A large animal (ovine) J Med Eng Technol Downloaded from informahealthcare.com by Miss Sandy Dalgleish on 12/16/13 For personal use only. 4 S. Viljoen et al. in vivo trial is now underway to investigate chronic interactions between the HSCMS and the spinal cord surface and histologic examination of explanted samples for evidence of tissue damage will be carried out. (The preliminary data from an earlier acute (non-survival) ovine in vivo study [3] showed no evidence of tissue disruption.) Lastly, from a HSCMS design perspective, it is now clear that the lead loops of the devices implanted in some patients must be sufficiently large to permit a total rostral-caudal range of motion of up to 2 cm for the cord/membrane relative to the Oya Strap fixation point, i.e. 1 cm rostral and 1 cm caudal from the neutral position. HSCMS prototypes with loop area &160 mm2, when placed on a custom-designed silicone surrogate spinal cord arrangement [13,14] can accommodate axial movement of the surrogate cord over most of that range of displacements without lift-off of the membrane at either the rostral and caudal extremes of displacement. Careful bench top studies now underway will evaluate prototype devices with both larger and smaller loop areas and membrane lengths, to test for potential lift-off or other movement of the HSCMS as a function of spinal cord displacement and rotation. However, beyond these biomechanical design considerations, the natural scar formation processes will also play a role in stabilizing the position of the electrode-bearing membrane on the surface of the spinal cord. A separate series of in vivo studies, now also underway, are investigating that point. Lastly, other future studies will include extending the number and age range of volunteers in the imaging protocol discussed above, to broaden the scope of our findings. 5. Conclusions MR-based measurements of the rostral-caudal motions of the spinal cord during flexion of the spine were made in a series of 16 volunteers. The results indicated that there were large variations, &70%, in the magnitude of that motion from person to person, with implications for the design of the intradural components of the HSCMS. In particular, there will be a spectrum of spinal cord strains associated with flexion-driven motion of the cord. Having suitable axial compliance within the electrode bearing portion of the HSCMS will reduce the risk of potential irritation of the pial surface by the HSMCS in those patients where the intraparenchymal strains are large enough to induce some sliding motion of the electrodes. In patients with small levels of strain, there would be little relative motion between cord and HSCMS, meaning that there would be small risk of any skidding of the electrodes on the pial surface. Also, the net axial travel of the spinal cord relative to the Oya Strap fixation device of the HSCMS is within the range that can be accommodated without lift-off of the electrode bearing portion of the HSCMS for some of the devices that were preliminarily tested. More definitive and systematic evaluations are underway. Acknowledgements The first two authors have made equal contributions to this work and are therefore designated as co-first authors. We thank Hiroyuki Oya, MD, Hiroto Kawasaki, MD, W. R. Smoker, MD, and H. Chen for several useful J Med Eng Technol, 2014; 38(1): 1–4 discussions and technical assistance. We also thank R. Shurig and colleagues of Evergreen Medical Technologies LLC for skillful fabrication of the prototype HSCMS devices and for the illustration used in Figure 1. AMS subject classification: 92C10; 92C. Declaration of interest Authors Viljoen, Dalm, Reddy, Gillies and Howard may receive patent royalties from any commercial licensing of the HSCMS intellectual properties that might be negotiated by their respective institutions. References 1. Eldabe, S., Kumar, K., Buchser, E., and Taylor, R.S., 2010, An analysis of the components of pain, function, and health-related quality of life in patients with failed back surgery syndrome treated with spinal cord stimulation or conventional medical management. Neuromodulation, 13, 201–209. 2. Howard III, M.A., Utz, M., Brennan, T.J., Dalm, B.D., Viljoen, S., Jeffery, N.D., and Gillies, G.T., 2011, Intradural approach to selective stimulation in the spinal cord for treatment of intractable pain. Journal of Applied Physics, 110, 044702. 3. Flouty, O., Oya, H., Kawasaki, H., Kawasaki, H., Wilson, S., Reddy, C.G., Jeffery, N.D., Brennan, T.J., Gibson-Corley, K.N., Utz, M., Gillies, G.T., and Howard III, M.A., 2012, A new device concept for directly modulating spinal cord pathways: Initial in vivo experimental results. Physiological Measurement, 33, 2003–2015. 4. Oliynyk, M.S., Gillies, G.T., Oya, H., Wilson, S., Reddy, C.G., and Howard III, M.A., 2013, Dynamic loading characteristics of an intradural spinal cord stimulator. Journal of Applied Physics, 113, 026103. 5. Viljoen, S., Dalm, B.D., Reddy, C.G., Wilson, S., Smittkamp, C.A., Gillies, G.T., and Howard, III, M.A., Optimization of intradural spinal cord stimulator designs via analysis of thoracic spine imaging data. Journal of Medical and Biological Engineering, 33, 193–198. 6. Figley, C.R., and Stroman, P.W., 2007, Investigation of human cervical and upper thoracic spinal cord motion: Implications for imaging spinal cord structure and function. Magnetic Resonance in Medicine, 58, 185–189. 7. Figley, C.R., Yau, D., and Stroman, P.W., 2008, Attenuation of lower-thoracic, lumbar, and sacral spinal cord motion: Implications for imaging human spinal cord structure and function. AJNR American Journal of Neuroradiology, 29, 1450–1454. 8. Takiguchi, T., Yamaguchi, S., Tezuka, M., and Kitajima, T., 2009, Measurement of shift of the cauda equina in the subarachnoid space by changing position. Regional Anesthesia and Pain Medicine, 34, 326–329. 9. Lee, R.A., van Zundert, A.A.J., Botha, C.P., Lataster, L.M.A., van Zundert, T.C.R., van der Ham, W.C.J.M., and Wieringa, P.A., 2010, The anatomy of the thoracic spinal canal in different postures. Regional Anesthesia and Pain Medicine, 35, 364–369. 10. Harrison, D.E., Cailliet, R., Harrison, D.D., Troyanovic, S.J., and Harrison, S.O., 1999, A review of biomechanics of the central nervous system – part I: Spinal canal deformations resulting from changes in posture. Journal of Manipulative and Physiological Therapeutics, 22, 227–234. 11. Cox, J.M., 2000, A review of biomechanics of the central nervous system – part I: Spinal canal deformations caused by changes in posture (comment). Journal of Manipulative and Physiological Therapeutics, 23, 211–217. 12. Reid, J.D., 1960, Effects of flexion-extension movements of the head and spine upon the spinal cord and nerve roots. Journal of Neurology, Neurosurgery and Psychiatry, 23, 214–221. 13. Howard III, M.A., Utz, M., Brennan, T.J., Dalm, D.D., Viljoen, S., Kanwal, J.K., and Gillies, G.T., 2011, Biophysical attributes of an in vitro spinal cord surrogate for use in developing an intradural neuromodulation system. Journal of Applied Physics, 110, 074701. 14. Oya, H., Howard III, M.A., Shurig, R., and Gillies, G.T., 2012, Spinal canal surrogate for testing intradural implants. Journal of Medical Engineering & Technology, 36, 407–410.
© Copyright 2026 Paperzz