Surgery FlowSound® Audiovisual Volume Flow Analysis Bypass Graft Patency Related to Vocal Pitches Normal Vocal Range The Transonic® Flowmeter measures absolute volume flow – and Transonic’s proprietary FlowSound® communicates these values audibly, so that a musically attuned surgeon comfortable with FlowSound® can save precious minutes in confirming functionality without taking the eyes from the surgical field. Even more time can be saved when examining grafts with technical error. G3 1.5; 2 mm <0.9 High C Hum FlowSound® is a powerful tool in the arsenal of a CABG surgeon experienced in “FAST” – Flow-Assisted Surgical Technique. Middle C Introduction C4 E4 G4 C5 E5 G5 C6 F6 C7 1.5 2.5 3.5 6 10 15 25 50 100 3 mm <1.8 3 5 7 12 20 30 50 100 200 4 mm <3.5 6 10 14 25 40 60 100 200 400 Probe Flow in mL/min Transonic Flowprobes’ Flow - to - Pitch (FlowSound) Conversion ® Probe Size Acceptable FlowSound ® Flow Ranges 1.5; 2 mm Volume Flow ≥ C6 (an octave above high C) ≥ 25 mL/min 3 mm ≥ E5 (E above high C) ≥ 20 mL/min 4 mm ≥ C5 (High C) ≥ 25 mL/min Unlike Doppler’s“swoosh-swoosh” flow Fig. 1: FlowSound pitch-to-flow conversion for CABG 1.5 mm - 4 mm. Flowprobes. velocity sounds, the calibrated pitches of Comparing FlowSound pitches between systole and diastole provides FlowSound® provide direct audio feedback of feedback on the systolic/diastolic flow ratio and graft patency. Sounds within bypass graft VOLUME FLOW! the normal vocal range are generally too low for a patent CABG graft. FlowSound CABG Flow-Assisted Surgical Technique (FAST) Measurement Preparation • Apply sterile gel inside the Flowprobe. Turn on FlowSound. A steady hum (Fig. 1: piano keyboard low G3) will be heard. The hum is the sound pitch for any flow near zero and below the Flowmeter’s minimal zero offset. • When a Flowprobe is applied to a graft, FlowSound “plays” the graft’s pulsatile flow (see Fig. 1 & sidebar below). When the Flowprobe is held steady, a repetitive beatby-beat FlowSound is generated. • The Flowprobe measures any motion within its flow window. Therefore, if the probe is moved, a “motion artifact” will generate a FlowSound spike that distorts the Pulsatility Index (PI). If FlowSound® is jittery, it could indicate turbulence caused by too large a vessel stuffed into the Flowprobe’s window. FlowSound Pitch An increase in FlowSound’s Do Re Mi Fa So La Ti Do pitch by one-half tone (E to F C D E F G A B C [Mi to Fa]) corresponds to a 12.5% increase in volume flow. One full tone increase in pitch (F to G [Fa to So]) corresponds to a 25% volume flow increase. An octave (C to C [Do to Do]) change in pitch equates to a four-fold volume flow increase. FlowSoundAdvantage(CV-23-tn)RevA2014USltr Audiovisual Graft Patency Assessment• Listen to FlowSound and observe the heart beating. 1.Temporarily Eliminate Competitive Flow Competitive flow from the native coronary artery reduces and sometimes reverses graft flow. One should, therefore, measure the graft at its maximum flow capacity. Temporarily apply finger pressure to the native vessel proximal to the anastomosis. If competitive flow had been present, the FlowSound pitch will increase, indicating its elimination. 2. Diastolic Dominant Left Heart FlowSounds Contracted muscle resists inflow. Therefore, on a good graft to the left heart, one would expect low flow (a pitch within one’s vocal range) during systole, and a far higher pitched FlowSound (above one’s vocal range) during diastole for a “Diastolic-dominant Flow Profile.” 3. Systolic/Diastolic Balanced Right Heart FlowSounds The right side of the heart contracts less forcefully than the left heart. Therefore, bypass graft flow to a right heart coronary is less impeded during systole. Both systolic and diastolic FlowSounds to a good right heart graft will be above one’s vocal range. Surgery FlowSound® Audiovisual Volume Flow Analysis cont. Technical Error Troubleshooting Using FlowSound If an audiovisual inspection of graft patency indicates the need to examine the graft for technical error, FlowSound® provides instantaneous feedback on whether manipulation of the graft will affect flow. Again, be sure to rule out competitive flow first. Fig. 2.: Competitive flow is evident from the negative flow spike on the right side of the waveform caused by reversal of flow through the graft at the start of systole. Competitive flow spikes increase the PI reading, and should be eliminated for a valid graft patency test. Flow Reversal = Negative Waveform Spikes Since a native coronary artery connects to the base of the aorta, the systolic pressure wave arrives at the anastomotic site first via the native coronary circuit, before it travels the longer bypass graft route. In the presence of competitive flow from the native coronary, this reverses flow through the bypass graft and creates a short peak of retrograde (negative) flow on the flow waveform at the beginning of systole. Flowsound Graft Patency Assessment 1.If mean graft flow is above 25 mL/min (above 20 mL/min for a smaller patient), graft flow is good. FlowSound’s pitch will be well above the normal vocal range. Mean flow is displayed on the AureFlo® display and front panel LED of the Flowmeter. 2.When mean flow is suspect (<20 mL/min), technical error such as a misapplied stitch, thrombus formation and distal obstruction in the native coronary must be ruled out before one can conclude that the poor flow results from acceptable causes such as low blood pressure, diseased cardiac muscle or a limited flow delivery capacity of an arterial graft. • If the graft is partially constricted from a misapplied stitch at its distal anastomosis, the constriction often limits graft flow. Central pressure governs the graft’s flow profile more than systole/diastole. Because central pressure is highest during systole, you will hear some blood flow squeezing past the constriction during a portion of systole, producing a “systolic-dominant flow profile” where systolic pitch is higher than diastolic pitch. • With the Flowprobe applied to the graft, manipulate the distal anastomotic site (tugging on anastomosis may open up a constriction). A significant change in FlowSound as the result of minor repositioning of the graft indicates that there is a trouble spot on this site. • A fully thrombosed graft will exhibit no flow pulsatility at all. FlowSound will hum its G3 pitch. • In the case of sequential grafts, the distribution between its branches is determined by the flow resistance beyond their myocardial target sites. Test each target graft at its full flow capacity individually by temporarily constricting the other branches. Transonic Systems Inc. is a global manufacturer of innovative biomedical measurement equipment. Founded in 1983, Transonic sells “gold standard” transit-time ultrasound flowmeters and monitors for surgical, hemodialysis, pediatric critical care, perfusion, interventional radiology and research applications. In addition, Transonic provides pressure and pressure volume systems, laser Doppler flowmeters and telemetry systems. www.transonic.com AMERICAS EUROPE ASIA/PACIFIC JAPAN Transonic Systems Inc. 34 Dutch Mill Rd Ithaca, NY 14850 U.S.A. Tel: +1 607-257-5300 Fax: +1 607-257-7256 [email protected] Transonic Europe B.V. Business Park Stein 205 6181 MB Elsloo The Netherlands Tel: +31 43-407-7200 Fax: +31 43-407-7201 [email protected] Transonic Asia Inc. 6F-3 No 5 Hangsiang Rd Dayuan, Taoyuan County 33747 Taiwan, R.O.C. Tel: +886 3399-5806 Fax: +886 3399-5805 [email protected] Transonic Japan Inc. KS Bldg 201, 735-4 Kita-Akitsu Tokorozawa Saitama 359-0038 Japan Tel: +81 04-2946-8541 Fax: +81 04-2946-8542 [email protected]
© Copyright 2026 Paperzz