LUS-2-Brochure - Olympus Australia

Ultrasonic Lithotriptor
LUS-2
Two Functions In One Unit Improve Both Procedural And Cost Efficiency
The Two-In-One
LUS-2 Ultrasonic Lithotriptor —
Powerfully Disintegrates Calculi,
Then Simultaneously Removes The Fragments
Ultrasonic lithotripsy is already highly-regarded for being minimally
invasive, while disintegrating calculi and removing the fragments
simultaneously. Yet by carefully re-examining every element of its
conventional lithotripsy products, Olympus has developed new ways to
improve efficiency. The result is the LUS-2 Ultrasonic Lithotriptor.
Improvements include an ergonomic and lightweight transducer to lessen
hand fatigue, a more compact body to save space, and a wider array of
probes to be prepared for any situation. The LUS-2 Ultrasonic Lithotriptor
Several Easy-To-Adjust
Ultrasonic Output And
Suction Settings
from Olympus—Efficient ultrasonic lithotripsy for efficient medical
professionals.
Lightweight, Ergonomic,
And Autoclavable
Transducers
The LUS-2's transducer is not only
lightweight and ergonomically designed
to minimize hand fatigue, but also allows
fast and low-cost autoclave sterilization
that uses safe high-pressure steam,
instead of ethylene oxide gas or
formaldehyde.
Both Disintegration
And Removal In A
Single Compact Unit
The compact LUS-2 is capable of both
disintegration of calculi and removal of
the fragments, which simplifies your
procedures, saves space, and makes the
equipment easy to carry.
Wide Array of
Easy-To-Replace
Probes
Six types of optional probes are available,
and changing them is simple and easy.
The 2.0 mm O.D. probe has been newly
added, widening the selection and
versatility even further.
3.7
mm
LUS-2
1.5
mm
1.5
O.
D.
mm
Sh
O.
ort
D.
Lo
2.0
ng
mm
O.
2.3
D.
mm
NEW
O.
D.
3.4
mm
O.
D.
The 3 ultrasonic lithotripsy output settings
and 4 suction power settings are easy to
select.
Warning Indicator
Upon detecting a disconnection/damage
of the probe or a transducer malfunction,
a warning indicator on the front panel
lights up and an alarm sounds.
Recommended Combinations
pe
sco
tero
e
r
U
Pro
ES
O
e
Th
With
O.
D.
Conventional Model
h
Wit
ous
ane
t
u
erc
ro P
SP
E
O
The
ope
osc
r
h
Nep
WA33024A
A02958A
A02959A
Adapters for LUS-2 simplify the handling of
LUS-2 and prevent the probe from getting
kinked or damaged during the procedure.
Specifications
LUS-2 Ultrasonic Lithotriptor
Power Supply
Voltage
Power Fluctuation
Frequency
Input Current
Size
Dimensions
(including pump head
and bottle holder)
Weight
Protection Against
Electric Shock
Protection Against
Electric Shock With
Applied Part
Protection Against
Explosion
Hand Pieces
Classification
(electro-medical
equipment)
Ultrasonic Output
120/230-240V AC
Within ±10%
50/60 Hz
6A (at 120V AC),
3A (at 230-240V AC)
360 (W) x 160 (H) x 450 (D) mm
13 kg
Class I
Type BF (low frequency leakage
current below 100 µA)
Never use the LUS-2 in
a flammable situation.
Only the Olympus Transducer
for the LUS-2 can be used.
Foot switch
The output level can be set
by pressing the ultrasonic output
setting switches (level 1, 2, 3).
23.5 kHz
Foot switch
The suction level can be set
by pressing the suction setting
switches (level 0,1, 2, 3).
When probe damage/looseness
or transducer malfunction occur,
ultrasonic output is stopped, a
warning alarm sounds,
and the ultrasonic output error
indicator lights up.
On/Off Operation
Output Setting
Suction
Safety Features
Frequency
On/Off Operation
Suction Setting
Problems With
Ultrasonic Output
LUS-2 Standard Set
LUS-2
Ultrasonic Lithotriptor
MAJ-1001 Foot Switch For LUS-2
MB-721
Suction Tube
MAJ-1053 Drain Tube For LUS-2 (2 pcs.)
MAJ-1078 Debris Collection Bottle For LUS/LUS-2
Power Cord
MAJ-1215 Transducer For LUS-2
MAJ-1114 3mm/6mm Torque Wrench
Probe For LUS-2
MAJ-1215 Transducer
Maximum Outer Diameter
Of Hand Piece (when fully assembled)
Weight
Length Of Hand Piece Cord
34mm
125g (excluding hand piece cord and plug)
3m
MAJ-1001 Foot Switch
Size
Classification
(electro-medical
equipment)
Dimensions
Weight
Length Of Cord
Water Resistance
Degree Of Protection
Against Explosion
270 (W) x 86 (H) x 220 (D) mm
3.2kg
4m
Type IPX8 IEC60529 watertightness
(except plug section)
Category AP
MAJ-981
MAJ-982
MAJ-1216
MAJ-983
MAJ-984
MAJ-985
Outer Diameter
1.5mm
1.5mm
2.0mm
2.3mm
3.4mm
3.7mm
Working Length
384mm
593mm
400mm
484mm
399mm
407mm
Packed With Probe Unclogging Rod
MB-728
MB-727
Chart Of Adapter Connection
WA33036A
OES Pro Percutaneous
Nephroscope
WA33024A
Adapter
A02958A
Adapter
WA29042A
OES Pro Ureteroscope
WA00395A
WA00396A
Attachment
MAJ-1215
Transducer for LUS-2
A02959A
Adapter
A0395
A0396
Attachment
Printed in Japan R294UB-0706