Selectra/Flexor ALBUMIN IN URINE/CSF FS TEST PARAMETERS (Microalbumin) Serum application Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Albumin Serum AlbS Two-point 578 nm g/l 2 0.000 g/l 100.0 g/l TruCal Protein 3 6 Multi* Selectra 2 – Non Linear Selectra XL – Standard Cubic Spline 14 days No 0.010% No * * * * * * * * * 1.000 0.000 g/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE AlbS No 25 ml 300 µl 301 µl 2.0 µl 1.0 µl 5 ml 60 µl 60 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 0242 .. .. … Notes 1. Please refer to the package insert for Albumin in Urine/CSF FS *Microalbumin for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 0242 90 32 51 September 2008/7 Selectra/Flexor ALBUMIN IN URINE/CSF FS TEST PARAMETERS (Microalbumin) Urine/CSF application Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes MIcroalbumin MAU Two-point 405 nm mg/l 2 0.000 mg/l 300.0 mg/l TruCal Alb U/CSF 3 6 Multi* Selectra 2 – Non Linear Selectra XL – Standard Cubic Spline 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE TRF No 25 ml 250 µl 254 µl 12.0 µl 8.0 µl 5 ml 50 µl 50 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 0242 .. .. … Notes 1. Please refer to the package insert for Albumin in Urine/CSF FS *Microalbumin for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 0242 90 52 51 September 2008/7 Selectra/Flexor ALBUMIN FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Albumin ALB Endpoint 578 nm g/l I 0.000 g/l 70.000 g/l TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.00 0.000 g/l DUAL MODE Order information Cat. No. 1 0220 .. .. … Notes 1. Please refer to the package insert for Albumin FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Albumin No 25 ml 300 µl 301 µl 3.0 µl 2.0 µl 5 ml 0 µl 0 µl No 4.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Albumin No 25 ml 300 µl 301 µl 3.0 µl 2.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 0220 90 02 51 September 2008/7 Selectra/Flexor ALAT (GPT) FS (IFCC Mod.) TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset ALAT ALT Kinetic 340 nm U/l I 0 500 None - Order information Cat. No. 1 2701.. .. … Notes No * * * * * * * * * 1.00 0.00 DUAL MODE 1. Please refer to the package insert for ALAT (GPT) FS (IFCC Mod.) for detailed information as per the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion Reagent blank ALAT No 25 ml 240 µl 259 µl 24 µl 5 µl 5 ml 60 µl 60 µl 50,186 sec No 15 % 0.850 3.00 -0.100 0.300 -0.750 Yes (#) Factor -2143 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank ALAT No 25 µl 300 µl 319 µl 24 µl 5 µl 51,175 sec 15 % 0.850 3.00 -0.100 0.300 0.200 Yes (#) Factor -2143 Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyzer is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyzer 845 2701 90 02 51 845 2701 90 82 51 November 2007/6 Selectra/Flexor ALAT (GPT) FS (IFCC Mod.) TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset with Pyridoxal-5-Phosphate ALAT ALT Kinetic 340 nm U/l I 0 500 None - Order information Cat. No. 1 2701.. .. … Notes No * * * * * * * * * 1.00 0.00 DUAL MODE 1. Please refer to the package insert for ALAT (GPT) FS (IFCC Mod.) for detailed information as per the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion Reagent blank ALAT No 25 ml 240 µl 259 µl 24 µl 5 µl 5 ml 60 µl 60 µl 50,186 sec No 15 % 0.850 3.00 -0.100 0.300 -0.750 Yes (#) Factor -2143 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyzer is at least one month provided that contamination and evaporation are avoided. 4. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany Factor This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyzer 845 2701 90 02 51 845 2701 90 82 51 November 2007/6 Selectra/Flexor α-AMYLASE CC FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Amylase AMY KINETIC 405 U/L 0 0 2000 Cat. No. 1 0501 .. .. … Notes NO * * * * * * 1. Please refer to the package insert for α-Amylase CC FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 1.000 0.000 DUAL MODE MONO MODE R1 bottle Normal Volume Rerun Volume Probe Sample Normal Vol. Sample Vol. rerun R2 bottle 4µl 2µl 5 ml Normal Volume 50µl Rerun Volume Predilution Slope blank Delay, min. time 50µl Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs Diviation Reagent blank Cal. Low Limit Cal.High Limit Factor Order information 25 ml 200µl 202µl 103,133 secs 10.0 -0.100 1.500 -0.100 0.800 2.300 Yes Sample blank R1 bottle Normal volume Rerun volume Sample Normal Volume rerun Volume Point one, two Incubation times Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank Cal Low Limit Cal. High Limit Factor NO 25µl 250µl 251µl 4µl 3µl 129,59 secs 10.0 -0.100 1.200 -0.100 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany 0.800 2.000 Yes 5670 **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. 5670 * can be selected freely # calculated by the analyser 845 0501 90 02 51 November 2007/5 Selectra/Flexor ALKALINE PHOSPHATASE FS DGKC TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset AP DGKC APDG Kinetic 405 nm U/l I 0.0 U/l 2300.0 U/l None - Factor Cat. No. 1 0401 .. .. … Notes No * * * * * * * * * 1.00 0.00 U/l DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R. Abs. Deviation Reagent blank Order information 1. Please refer to the package insert for Alkaline phosphatase FS DGKC for detailed information about the test on the following: MONO MODE AP DGKC No 25 ml 200 µl 202 µl 4.0 µl 2.0 µl 5 ml 50 µl 50 µl 50, 186 sec No 10.0 % -0.100 Abs 3.00 Abs -0.100 Abs 0.500 Abs Yes (#) 3433.0 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank AP DGKC No 25 ml 250 µl 251 µl Factor 2757.0 5.0 µl 4.0 µl 70, 78 sec 10.0 % -0.100 3.00 -0.100 3.000 3.000 Yes (#) Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 0401 90 02 51 November 2007/6 Selectra/Flexor ALKALINE PHOSPHATASE FS IFCC 37° C TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset AP IFCC APIF Kinetic 405 nm U/l I 0.0 U/l 2800.0 U/l None - Order information Cat. No. 1 0441 .. .. … Notes No * * * * * * * * * 1.00 0.00 DUAL MODE 1. Please refer to the package insert for Alkaline phosphatase FS IFCC for detailed information about the test on the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion Reagent blank AP IFCC No 25 ml 200 µl 203 µl 5.0 µl 2.0 µl 5 ml 50 µl 50 µl 77, 79 sec No 10.0 % -0.100 3.00 -0.100 3.000 3.000 Yes (#) Factor 2757.0 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank AP IFCC No 25 ml 250 µl 251 µl Factor 2757.0 5.0 µl 4.0 µl 70, 78 sec 10.0 % -0.100 3.00 -0.100 3.000 3.000 Yes (#) Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 0441 90 02 51 November 2007/6 Selectra/Flexor APOLIPOPROTEIN A1 FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Apolipoprotein A ApoA Two-point 578 nm mg/dl 2 16.00 mg/dl 230.0 mg/dl TruCal Apo A1/B 3 4 Multi* Selectra 2 – Non Linear Selectra XL – 4PLL 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/dl Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE ApoA No 25 ml 250 µl 251 µl 2.0 µl 1.0 µl 5 ml 50 µl 50 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7102 .. .. … Notes 1. Please refer to the package insert for Apoliporotein A1 FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7102 90 02 51 September 2008/7 Selectra/Flexor APOLIPOPROTEIN B FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Apolipoprotein B ApoB Two-point 340 nm mg/dl 2 0.000 mg/dl 230.0 mg/dl TruCal ApoA1/ B 3 4 Multi* Selectra 2 – Non Linear Selectra XL – 4PLL 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/dl Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE ApoB No 25 ml 250 µl 252 µl 3.0 µl 1.0 µl 5 ml 50 µl 50 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7112 .. .. … Notes 1. Please refer to the package insert for Apoliporotein B FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7112 90 02 51 September 2008/7 Selectra/Flexor ANTISTREPTOLYSIN O FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibrator Modes Antistrept. O ASO Two-point 505 nm IU/ml 1 0.00 IU/ml 800.0 IU/ml TruCal ASO 3 6 Multi* Selectra 2 – Non Linear Selectra XL – 4PLL 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE ASO No 25 ml 250 µl 252 µl 3.0 µl 1.0 µl 5 ml 50 µl 50 µl No No 6 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7012.. .. … Notes 1. Please refer to the package insert for Antistreptolysin O FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7012 90 02 51 September 2008/7 Selectra/Flexor ASAT(GOT) FS (IFCC Mod) TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset ASAT AST Kinetic 340 nm U/l I 0 U/l 500 U/l None - Cat. No. 1 2601 .. .. … Notes No * * * * * * * * * 1.00 0.00 DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion Reagent Blank Factor Order information 1. Please refer to the package insert for ASAT (GOT) FS (IFCC Mod) for detailed information about the test on the following: MONO MODE ASAT No 25 ml 240 µl 259 µl 24 µl 5 µl 5 ml 60 µl 64 µl 50,186 sec No 15 % 0.600 Abs 3.00 Abs -0.100 Abs 0.300 Abs -0.500 Abs Yes(#) -2143 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank ASAT No 25 ml 300 µl 319 µl 24 µl 5 µl 51,175 sec 15 % 0.600 3.00 -0.100 0.300 0.200 Yes (#) Factor -2143 Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyzer is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyzer 845 2601 90 02 51 845 2601 90 82 51 November 2007/6 Selectra/Flexor ASAT(GOT) FS (IFCC Mod) with Pyridoxal-5-Phosphate TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset ASAT AST Kinetic 340 nm U/l I 0 U/l 500 U/l None - Cat. No. 1 2601 .. .. … Notes 1. Please refer to the package insert for ASAT (GOT) FS (IFCC Mod) for detailed information about the test on the following: No * * * * * * * * * 1.00 0.00 DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion Reagent Blank Factor Order information MONO MODE ASAT No 25 ml 240 µl 259 µl 24 µl 5 µl 5 ml 60 µl 64 µl 50,186 sec No 15 % 0.600 Abs 3.00 Abs -0.100 Abs 0.300 Abs -0.500 Abs Yes(#) -2143 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank Factor Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyzer is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyzer 845 2601 90 02 51 845 2601 90 82 51 November 2007/6 BILIRUBIN AUTO DIRECT FS Selectra/Flexor TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Bilirubin Direct DBili Two-point 546 nm mg/dl 1 0.000 mg/dl 10.00 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr. Depletion Reagent blank Cal Low limit Cal High limit Factor MONO MODE DBili No 25 ml 240 µl 250 µl 25.0 µl 15.0 µl 5 ml 60 µl 60 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 0.500 Abs 3.000 Abs Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs. Deviation Reagent blank Cal Low limit Cal High limit Order information Cat. No. 1 0821 .. .. … Notes 1. Please refer to the package insert for Bilirubin Auto Direct FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 0821 90 02 51 August 2010/8 BILIRUBIN AUTO TOTAL FS Selectra/Flexor TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Bilirubin Total TBili Two-point 546 nm mg/dl 1 0.000 mg/dl 30.00 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit TBili No 25 ml 240 µl 244 µl 6.0 µl 2.0 µl 5 ml 60 µl 60 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 0.500 Abs Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs. Deviation Reagent blank Cal Low limit Cal High limit Order information Cat. No. 1 0811 .. .. … Notes 1. Please refer to the package insert for Bilirubin Auto Total FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 0811 90 02 51 August 2010/8 Selectra/Flexor COMPLEMENT C3c FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Complement C3c C3c Two-point 340 nm mg/dl 2 0.000 mg/dl 360.00 mg/dl TruCal Protein 3 6 Multi* Selectra 2 – Non Linear Selectra XL - Standard Cubic Spline 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/dl Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit Order information Cat. No. 1 1802 .. .. … Notes 1. Please refer to the package insert for Complement C3c FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE C3c No 25 ml 250 µl 252 µl 4.0 µl 2.0 µl 5 ml 50 µl 50 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1802 90 02 51 September 2008/7 Selectra/Flexor COMPLEMENT C4 FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Complement C4 C4 Two-point 340 nm mg/dl 2 0.000 mg/dl 60.00 mg/dl TruCal Protein 3 6 Multi* Selectra 2 – Non Linear Selectra XL - Standard Cubic Spline 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/dl Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit Order information Cat. No. 1 1812 .. .. … Notes 1. Please refer to the package insert for Complement C4 FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE C4 No 25 ml 250 µl 254 µl 7.0 µl 3.0 µl 5 ml 50 µl 50 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1812 90 02 51 September 2008/7 Selectra/Flexor CALCIUM CPC FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Calcium CPC Ca Endpoint 578 nm mg/dl 2 0.000 mg/dl 20.000 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 1121 .. .. … Notes 1. Please refer to the package insert for Calcium CPC FS for detailed information about the test on the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Ca No 25 ml 240 µl 242 µl 4.0 µl 2.0 µl 5 ml 60 µl 60 µl No 6.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Ca No 25 ml 300 µl 301 µl 4.0 µl 3.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1121 90 02 51 September 2008/7 Selectra/Flexor CHOLINESTERASE FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Cholinesterase CHE Kinetic 405 nm U/l I 0.000 U/l 20000 U/l None No * * * * * * * * * 1.00 0.00 U/l DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R. Abs. Deviation Reagent Blank Factor Order information Cat. No. 1 1401 .. .. … Notes 1. Please refer to the package insert for Cholinesterase FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE CHE No 25 ml 200 µl 202 µl 4.0 µl 2.0 µl 5 ml 50 µl 50 µl 103, 133 sec No 10.0 % 0.000 Abs 3.000 Abs -0.100 Abs 0.300 Abs Yes (#) -68500 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany Factor **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1401 90 02 51 November 2007/7 Selectra/Flexor CHOLESTEROL FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Cholesterol Chol Endpoint 546 - 620 nm mg/dl I 0.000 mg/dl 800.0 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 1300 .. .. … Notes 1. Please refer to the package insert for Cholesterol FS for detailed information about the test on the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Cholesterol No 25 ml 300 µl 302 µl Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # 3.0 µl 1.0 µl 5 ml 0 µl 0 µl No 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Cholesterol No 25 ml 300 µl 301 µl Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # 3.0 µl 2.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1300 90 02 51 September 2008/7 CK-NAC FS Selectra/Flexor TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Creatine kinase CK Kinetic 340 nm U/l I 0.0 1000.0 None - Order information Cat. No. 1 1601 .. .. … Notes 1. Please refer to the package insert for CK-NAC FS for detailed information about the test on the following: No * * * * * * * * * 1.00 0.00 DUAL MODE MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion CK No 25 ml 200 µl 202 µl No 10.0 % -0.100 Abs 3.00 Abs -0.100 Abs 3.00 Abs 3.00 Abs Factor 4127 10.0 µl 8.0 µl 5 ml 50 µl 50 µl 103, 133 sec Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit Substr. Depletion Reagent blank CK No 25 ml 250 µl 252 µl Factor 4127 10.0 µl 8.0 µl 168, 254 sec 10.0 % -0.100 Abs 3.00 Abs -0.100 Abs 3.00 Abs 3.00 Abs Yes (#) Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1601 90 02 51 November 2007/6 Selectra/Flexor CHLORIDE FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Chloride Cl Endpoint 436 - 620 nm mmol/l 1 80.000 mmol/l 140.00 mmol/l Standard 3 1 * Linear 14 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 1200 .. .. … Notes 1. Please refer to the package insert for Chloride FS for detailed information about the test on the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Chloride No 25 ml 300 µl 301 µl 3.0 µl 2.0 µl 5 ml 0 µl 0 µl No 11.5 min -0.100 Abs 3.000 Abs -0.100 3.000 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Chloride No 25 ml 300 µl 301 µl Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit 3.0 µl 2.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. 2. The stability of the reagent on board the analyser is 7 days provided that contamination and evaporation are avoided. * can be selected freely # calculated by the analyser 845 1200 90 02 51 September 2009/5 Selectra/Flexor CREATININE FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Creatinine Jaffe CreJ Two-point 505 nm mg/dl 2 0.000 mg/dl 15.00 mg/dl TruCal U 3 1 * Linear 1 day No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs. Deviation Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 1711 .. .. … Notes 1. Please refer to the package insert for Creatinine FS for detailed information about the test on the following: MONO MODE CreJ No 25 ml 200 µl 208 µl 10.0 µl 2.0 µl 5 ml 50 µl 50 µl No 50 130 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs. Deviation Reagent blank Cal Low limit Cal High limit Factor CreJ No 25 ml 250 µl 250 µl 10.0 µl 10.0 µl 51, 129 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No * * # Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is 9 days provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1711 90 02 51 February 2009/9 Selectra/Flexor CREATININE FS TEST PARAMETERS Compensated Method Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Creatinine Jaffe CreJ Two-point 505 nm mg/dl 2 0.000 mg/dl 15.00 mg/dl TruCal U 3 1 * Linear 1 day No No * * * * * * * * * 1.000 - 0.300 mg/dl DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs. Deviation Reagent blank Cal Low limit Cal High limit Factor To use the Creatinine FS Compensated Method on the Selectra/ Flexor the Correlation Offset must be changed as follows; When the units are mg/dL Correlation Offset - 0.3 When the units are µmol/L Correlation Offset - 27 Note: Ensure that the relevant assigned calibrator value is selected when using the Creatinine Compensated Method. MONO MODE CreJ No 25 ml 200 µl 208 µl 10.0 µl 2.0 µl 5 ml 50 µl 50 µl No 50, 130 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs. Deviation Reagent blank Cal Low limit Cal High limit Factor CreJ No 25 ml 250 µl 250 µl 10.0 µl 10.0 µl 51, 129 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No * * # * can be selected freely # calculated by the analyser 845 1711 90 02 51 February 2009/9 Selectra/Flexor CREATININE PAP FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Creatinine PAP CreP Enppoint 546 nm mg/dl 2 0.000 mg/dl 160.00 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 1759 .. .. … Notes 1. Please refer to the package insert for Creatinine PAP FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Incubation Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit CreP No 25 ml 200 µl 206 µl Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # 8.0 µl 2.0 µl 25 ml 100 µl 100 µl No No 6.5 min -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs. Deviation Reagent blank Cal Low limit Cal High limit Factor 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1759 90 02 51 April 2011/6 Selectra/Flexor CRP FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes CRP CRP Two-point 340 nm mg/l 2 2.00 mg/l 250.0 mg/l TruCal CRP 3 6 Multi* Selectra 2 – Non Linear Selectra XL – 4PLL 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE CRP No 25 ml 250 µl 260 µl 15.0 µl 5.0 µl 5 ml 50 µl 50 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7002 .. .. … Notes 1. Please refer to the package insert for CRP FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7002 90 02 51 September 2008/7 Selectra/Flexor CRP U-hs High sensitive (hs) application TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes CRP U-hs CRP Two-point 505 nm mg/l 2 0.05 mg/l 20.0 mg/l TruCal CRP hs 2 6 Multi* Selectra 2 – Non Linear Selectra XL - Standard Cubic Spline 7 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE CRP No 25 ml 150 µl 160 µl 15.0 µl 5.0 µl 5 ml 150 µl 150 µl No No 6 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7045 .. .. … Notes 1. Please refer to the package insert for CRP U-hs for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7045 90 42 51 September 2008/5 Selectra/Flexor CRP U-hs Universal (U) application TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes CRP U-hs CRP Two-point 505 nm mg/l 2 3.00 mg/l 500.0 mg/l TruCal CRP U 2 6 Multi* Selectra 2 – Non Linear Selectra XL - Standard Cubic Spline 10 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE CRP No 25 ml 198 µl 200 µl 4.0 µl 2.0 µl 5 ml 198 µl 198 µl No No 24 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 0.100 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7045 .. .. … Notes 1. Please refer to the package insert for CRP U-hs for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7045 90 32 51 September 2008/5 Selectra/Flexor CRP U-hs High sensitive (hs) application TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes CRP U-hs CRP Two-point 505 nm mg/l 2 0.05 mg/l 20.0 mg/l TruCal CRP hs 2 6 Multi* Selectra 2 – Non Linear Selectra XL - Standard Cubic Spline 7 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE CRP No 25 ml 150 µl 160 µl 15.0 µl 5.0 µl 5 ml 150 µl 150 µl No No 6, 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7045 .. .. … Notes 1. Please refer to the package insert for CRP U-hs for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7045 90 42 51 September 2008/5 Selectra/Flexor CRP U-hs Universal (U) application TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes CRP U-hs CRP Two-point 505 nm mg/l 2 3.00 mg/l 500.0 mg/l TruCal CRP U 2 6 Multi* Selectra 2 – Non Linear Selectra XL - Standard Cubic Spline 10 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE CRP No 25 ml 198 µl 200 µl 4.0 µl 2.0 µl 5 ml 198 µl 198 µl No No 24, 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 0.100 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7045 .. .. … Notes 1. Please refer to the package insert for CRP U-hs for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7045 90 32 51 September 2008/5 Selectra/Flexor IRON FS Ferene TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Iron Ferene Fe Two point 578 nm µg/dl Cat. No. 1 1911 .. .. … 0.000 µg/dl 1000.0 µg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 µg/dl DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Point one, two Slope blank Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Order information MONO MODE Iron Ferene No 25 ml 240 µl 250 µl Reagent blank Name Sample blank R1 bottle normal volume rerun volume Sample 15.0 µl normal Volume 5.0 µl rerun Volume 5 ml Incubation time 60 µl Low Absorbance 60 µl High Absorbance No R.Abs.L.Limit -3 236 sec R.Abs. H.Limit No -0.100 Abs 3.000 Abs Reagent blank -0.100 Cal Low limit 0.500 Cal High limit Factor Yes (#) Factor # Notes 1. Please refer to the package insert for Iron FS Ferene for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1911 90 02 51 September 2008/7 Selectra/Flexor FERRITIN FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Ferritin Ferr Two-point 578 nm µg/l I 0.00 µg/l 1000.00 µg/l TruCal Ferritin 3 5 Multi* Selectra 2 – Non Linear Selectra XL - Standard Cubic Spline 14 days No 1.000 % No * * * * * * * * * 1.000 0.000 µg/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit Order information Cat. No. 1 7059 .. .. … Notes 1. Please refer to the package insert for Ferritin FS for detailed information about the test on the following Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Ferr No 25 ml 200 µl 205 µl 10.0 µl 5.0 µl 5 ml 100 µl 100 µl No No 6 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7059 90 02 51 September 2008/7 Selectra/Flexor TEST PARAMETERS WITH FACTOR ACCORDING TO SZASZ Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Gamma-GT SZASZ GGT Kinetic 405 nm U/l I 0.000 U/l 750.0 U/l None - Order information Cat. No. 1 2801 .. .. … Notes No * * * * * * * * * 1.00 0.00 DUAL MODE 1. Please refer to the package insert for Gamma-GT FS (Szasz mod.) for detailed information about the test on the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion Reagent blank GGT No 25 ml 200 µl 210 µl No 10.0 % -0.100 3.00 -0.100 3.000 3.000 Yes (#) Factor 1421.0 20.0 µl 10.0 µl 5 ml 50 µl 50 µl 103, 133 sec GAMMA GT FS (Szasz mod.) Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank GGT No 25 ml 250 µl 260 µl Factor 1158.0 25.0 µl 15.0 µl 90, 58 sec 10.0 % -0.100 3.00 -0.100 3.000 Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany Yes (#) **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 2801 90 02 51 November 2007/6 Selectra/Flexor TEST PARAMETERS WITH FACTOR ACCORDING TO IFCC Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Gamma-GT SZASZ GGT Kinetic 405 nm U/l I 0.000 U/l 850.0 U/l None - Order information Cat. No. 1 2801 .. .. … Notes No * * * * * * * * * 1.00 0.00 DUAL MODE 1. Please refer to the package insert for Gamma-GT FS (Szasz mod.) for detailed information about the test on the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion Reagent blank GGT No 25 ml 200 µl 210 µl No 10.0 % -0.100 3.00 -0.100 3.000 3.000 Yes (#) Factor 1606.0 20.0 µl 10.0 µl 5 ml 50 µl 50 µl 103, 133 sec GAMMA GT FS (Szasz mod.) Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank GGT No 25 ml 250 µl 260 µl Factor 1309.0 25.0 µl 15.0 µl 90, 58 sec 10.0 % -0.100 3.00 -0.100 0.500 Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 4. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany Yes (#) **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 2801 90 02 51 November 2007/6 Selectra/Flexor GLUCOSE HEXOKINASE FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Glucose Hexokinase Gluc Endpoint 340 nm mg/dl 2 0.000 mg/dl 650.00 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 2511 .. .. … Notes 1. Please refer to the package insert for Glucose Hexokinase FS for detailed information about the test on the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Gluc No 25 ml 240 µl 242 µl 3.0 µl 1.0 µl 5 ml 60 µl 60 µl No 6.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Gluc No 25 ml 300 µl 301 µl 3.0 µl 2.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 2511 90 02 51 September 2008/7 Selectra/Flexor HbA1c FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration modes Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset HbA1c HbA1c one-point 620 nm ## 2 0 20 TruCal HbA1c liquid 4 level 2 5 (NaCl + 4 levels) Multi* Selectra 2 - Non linear Selectra XL – 4PLL 10 days No No * * * * * * * * * 1.000 0.000 DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume R3 bottle normal Volume rerun Volume predilution Slope blank Point one Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Order information Cat. No. 1 3329 .. .. … Notes 1. Please refer to the package insert for oneHbA1c FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE HbA1c No 25 ml 220 µl 223 µl 5.0 µl 2.0 µl 5 ml 75 µl 75 µl 5 ml 35 µl 35 µl No No 6.5 min -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Cal Low limit Multi dAbs* Cal High limit Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit 2. The stability of the reagent on board the analyser is at least 3 weeks provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany Reagent blank Cal Low limit Cal High limit Factor NOTE: For optimal performance the HbA1c should be run as a batch and a cuvette wash with NaOH as cleaner 14 performed after the assay. After the cuvette wash has finished NaOH should be replaced with the HCl. **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser ## Data entry by the user 845 3329 90 02 51 August 2010/7 Selectra/Flexor IMMUNOGLOBULIN A FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Immunoglobulin A Ig A Two-point 578 nm mg/dl I 0.000 mg/dl 900.0 mg/dl TruCal Protein 3 6 Multi* Selectra 2 – Non Linear Selectra XL – 4PLL 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/dl Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE Ig M No 25 ml 250 µl 251 µl 2.0 µl 1.0 µl 5 ml 50 µl 50 µl No No -3 209 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7202 .. .. … Notes 1. Please refer to the package insert for Immunoglobulin A FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7202 90 02 51 September 2008/7 Selectra/Flexor IMMUNOGLOBULIN G FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Immunoglobulin G Ig G Two-point 578 nm mg/dl I 0.000 mg/dl 3100.0 mg/dl TruCal Protein 3 6 Multi* Selectra 2 – Non Linear Selectra XL – 4PLL 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/dl Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE Ig G No 25 ml 300 µl 301 µl 2.0 µl 1.0 µl 5 ml 60 µl 60 µl No No -3 209 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7212 .. .. … Notes 1. Please refer to the package insert for Immunoglobulin G FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7212 90 02 51 September 2008/7 Selectra/Flexor IMMUNOGLOBULIN M FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Immunoglobulin M Ig M Two-point 405 nm mg/dl I 0.000 mg/dl 800.0 mg/dl TruCal Protein 3 6 Multi* Selectra 2 – Non Linear Selectra XL – 4PLL 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/dl Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE Ig M No 25 ml 300 µl 301 µl 2.0 µl 1.0 µl 5 ml 60 µl 60 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7222 .. .. … Notes 1. Please refer to the package insert for Immunoglobulin M FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7222 90 02 51 September 2008/7 Selectra/Flexor LACTATE FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Lactate Lac Endpoint 340 nm mg/dl 1 0.000 U/l 120.0 U/l DiaSys 3 1 * Linear 0 No No * * * * * * * * * 1.00 0.00 U/l DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Incubation Time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 4001 .. .. … Notes 1. Please refer to the package insert for Lactate FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Lac No 25 ml 240 µl 242 µl 4.0 µl 2.0 µl 5 ml 60 µl 60 µl 6.5 min 0.000 Abs 3.000 Abs 0.000 Abs 3.000 Abs Yes (#) * * 1.00 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 4001 90 02 51 September 2008/5 Selectra/Flexor LDH FS DGKC TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset LDH DGKC LDHD Kinetic 340 nm U/l 0 0.000 U/l 2000.0 U/l None No * * * * * * * * * 1.00 0.00 U/l DUAL MODE MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R. Abs. Deviation Substr Depletion LDH DGKC No 25 ml 250 µl 252 µl 5.0 µl 2.0 µl 5 ml 63 µl 64 µl 50, 186 sec No 10.0 % -0.400 Abs 3.000 Abs -0.100 3.000 Abs 0.200 Abs -0.400 Abs Factor -10095.0 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank Factor Order information Cat. No. 1 4201 .. .. … Notes 1. Please refer to the package insert for LDH.FS DGKC for the detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 4201 90 02 51 November 2007/6 Selectra/Flexor LDH FS IFCC TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset LDH IFCC LDHI Kinetic 340 nm U/l I 0.000 U/l 1500.0 U/l None No * * * * * * * * * 1.00 0.00 U/l DUAL MODE MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion Reagent Blank LDH IFCC No 25 ml 200 µl 203 µl 4.0 µl 1.0 µl 5 ml 50 µl 50 µl 77, 159 sec No 10.0 % -0.100 Abs 3.000 Abs -0.100 3.000 Abs 3.000 Abs Yes(#) Factor 10080.0 Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank Factor Order information Cat. No. 1 4211 .. .. … Notes 1. Please refer to the package insert for LDH FS IFCC for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 4211 90 02 51 November 2007/7 Selectra/Flexor LIPASE DC FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Lipase DC LIPA Kinetic 578 nm U/l I 0.000 U/l 250.0 U/l TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.00 0.00 U/l DUAL MODE MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs. Deviation Reagent Blank LIPA No 25 ml 200 µl 203 µl No 10.0 % -0.100 Abs 3.00 Abs -0.100 Abs 3.00 Abs 3.00 Abs Yes(#) Factor # 5.0 µl 2.0 µl 5 ml 50 µl 50 µl 130, 106 sec Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit Substr. Depletion Reagent blank Factor Order information Cat. No. 1 4321 .. .. … Notes 1. Please refer to the package insert for Lipase DC FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 4321 90 02 51 September 2008/7 Selectra/Flexor MAGNESIUM XL FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Magnesium Mg Endpoint 546 nm mg/dl 2 0.000 mg/dl 5.000 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 4610 .. .. … Notes 1. Please refer to the package insert for Magnesium XL FS for detailed information about the test on the following: MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Magnesium No 25 ml 300 µl 301 µl 3.0 µl 2.0 µl 5 ml 0 µl 0 µl No 4.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Magnesium No 25 ml 300 µl 301 µl 3.0 µl 2.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 4610 90 02 51 September 2008/4 Selectra/Flexor MYOGLOBIN FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Myoglobin Myo Two-point 578 nm µg/l I 0.00 µg/l 520.00 µg/l TruCal Myoglobin 3 5 Multi* Selectra 2 – Non Linear Selectra XL - Standard Cubic Spline 1 days No 0.100 % No * * * * * * * * * 1.000 0.000 µg/l Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit Order information Cat. No. 1 7098 .. .. … Notes 1. Please refer to the package insert for Myoglobin FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Myo No 25 ml 210 µl 213 µl 7.0 µl 4.0 µl 5 ml 70 µl 70 µl No No 24 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7098 90 02 51 September 2008/5 Selectra/Flexor PREALBUMIN FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Interval Curve Fit Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Prealbumin PreAlb Two-point 405 nm g/L 3 0.04 g/L 1.50 g/L TruCal Protein 3 6 (NaCl + 5 levels) Multi* * 4PLL No No * * * * * * * * * 1.000 0.000 g/L DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr. Depletion Reagent blank Cal Low limit Cal High limit Factor MONO MODE PALB No 25 mL 250 µL 252 µL 3.0 µL 1.0 µL 5 mL 50 µL 50 µL No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs Multi dAbs Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit R.Abs. Deviation Reagent blank Cal Low limit Cal High limit Order information Cat. No. 1 0292 .. .. … Notes 1. Please refer to the package insert for Prealbumin FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 0292 90 02 51 February 2008/5 Selectra/Flexor PHOSPHATE FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Phosphate Phos Endpoint 340 nm mg/dl 2 -1.000 mg/dl 40.000 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.00 0.000 g/dl DUAL MODE Order information Cat. No. 1 5211 .. .. … Notes 1. Please refer to the package insert for Phosphate FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Phosphate No 25 ml 240 µl 242 µl 3.0 µl 1.0 µl 5 ml 60 µl 60 µl No 6.5 min -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Phosphate No 25 ml 300 µl 300 µl Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # 3.0 µl 3.0 µl 11.5 min 0.000 Abs 3.000 Abs 0.000 Abs 3.000 Abs 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 5211 90 02 51 September 2008/5 Selectra/Flexor RHEUMATOID FACTOR FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Rheumatoid factor RF Two-point 340 nm IU/ml I 0.000 IU/ml 500.0 IU/ml TruCal RF 3 6 Multi* Selectra 2 – Non Linear Selectra XL - Standard Cubic Spline 14 days No 1.000% No * * * * * * * * * 1.000 0.000 IU/ml Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit MONO MODE Rf No 25 ml 250 µl 260 µl 15.0 µl 5.0 µl 5 ml 50 µl 50 µl No No -3 236 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 7022 .. .. … Notes 1. Please refer to the package insert for Rheumatoid Factor FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7022 90 02 51 September 2008/8 Selectra/Flexor TOTAL PROTEIN FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Total Protein TP Endpoint 546 nm g/dl I 0.000 g/dl 15.000 gd/l TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.00 0.000 g/dl DUAL MODE Order information Cat. No. 1 2311 .. .. … Notes 1. Please refer to the package insert for Total Protein FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Total Protein Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit 5.0 µl 4.0 µl 5 ml 60 µl 60 µl No 4.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # No 25 ml 240 µl 241 µl Total Protein Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit 5.0 µl 4.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # No 25 ml 300 µl 301 µl 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 2311 90 02 51 September 2008/6 Selectra/Flexor TRANSFERRIN FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Modes Transferrin TRF Two-point 578 nm mg/dl I 0.000 mg/dl 800.0 mg/dl TruCal Protein 3 6 Multi* Selectra 2 – Non Linear Selectra XL – 4PLL 14 days No 1.000% No * * * * * * * * * 1.000 0.000 mg/dl Interval Cut off Max. inaccuracy Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr depletion Reagent blank Cal Low limit Cal High limit Order information Cat. No. 1 7252 .. .. … Notes 1. Please refer to the package insert for Transferrin FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE TRF No 25 ml 250 µl 252 µl 3.0 µl 1.0 µl 5 ml 50 µl 50 µl No No -3 130 sec -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 3.000 Abs No Multi dAbs* Multi dAbs* Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 7252 90 02 51 September 2008/7 Selectra/Flexor TRIGLYCERIDES FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Triglyceride TRIG Endpoint 546 nm mg/dl I 0.000 mg/dl 1400.0 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 5710 .. .. … Notes 1. Please refer to the package insert for Triglycerides FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Triglyceride No 25 ml 300 µl 302 µl 3.0 µl 1.0 µl 5 ml 0 µl 0 µl No 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Triglyceride No 25 ml 300 µl 301 µl 3.0 µl 2.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 5710 90 02 51 September 2008/7 Selectra/Flexor TRIGLYCERIDES FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Triglyceride TRIG Endpoint 546 nm mg/dl I 0.000 mg/dl 1400.0 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 5760 .. .. … Notes 1. Please refer to the package insert for Triglycerides FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Triglyceride No 25 ml 300 µl 302 µl 3.0 µl 1.0 µl 5 ml 0 µl 0 µl No 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Triglyceride No 25 ml 300 µl 301 µl 3.0 µl 2.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 5760 90 02 51 September 2008/2 Selectra/Flexor URIC ACID FS TBHBA TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Uric Acid TBHBA UATB Endpoint 546 nm mg/dl 2 0.000 mg/dl 25.00 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs. L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Order information Cat. No. 1 3021 .. .. … Notes 1. Please refer to the package insert for Uric Acid FS TBHBA for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE UATB No 25 ml 240 µl 243 µl 5.0 µl 2.0 µl 5 ml 60 µl 60 µl No No 6,5 min -0.100 Abs 3.000 Abs -0.100 Abs 1.000 Abs Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution. * can be selected freely # calculated by the analyser October 2011/1 Selectra/Flexor URIC ACID FS TBHBA TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Uric Acid TBHBA UATB Endpoint 546 nm mg/dl 2 0.000 mg/dl 25.00 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl Order information Cat. No. 1 3021 .. .. … Notes 1. Please refer to the package insert for Uric Acid FS TBHBA for detailed information about the test on the following: DUAL MODE MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit 5 µl 3 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 Abs 0.500 Abs Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Reagent blank Cal Low limit Cal High limit Factor UATB No 25 ml 300 µl 302 µl Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 3021 90 02 51 September 2008/7 Selectra/Flexor URIC ACID (TOOS) FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Uric Acid TOOS UATO Endpoint 546 nm mg/dl 2 0.000 mg/dl 25.00 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 3001 .. .. … Notes 1. Please refer to the package insert for Uric Acid FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit UATO No 25 ml 240 µl 242 µl 5.0 µl 3.0 µl 5 ml 60 µl 60 µl No 6.5 min -0.100 Abs 3.000 Abs -0.100 0.500 Reagent blank Cal Low limit Cal High limit Factor Yes (#) * * # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 3001 90 02 51 September 2006/7 Selectra/Flexor UIBC FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset UIBC UIBC Two point 578 nm µg/dl 6.000 µg/dl 1000.0 µg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 µg/dl DUAL MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Point one, two Slope blank Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit MONO MODE UIBC No 25 ml 240 µl 249 µl Reagent blank Name Sample blank R1 bottle normal volume rerun volume Sample 18.0 µl normal Volume 9.0 µl rerun Volume 5 ml Incubation time 60 µl Low Absorbance 60 µl High Absorbance No R.Abs.L.Limit -3 236 sec R.Abs. H.Limit No -0.100 Abs 3.000 Abs Reagent blank -0.100 Cal Low limit 3.000 Cal High limit Factor Yes (#) Factor # Order information Cat. No. 1 1921 .. .. … Notes 1. Please refer to the package insert for UIBC FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 1921 90 02 51 Septemebr 2008/5 Selectra/Flexor UREA FS TEST PARAMETERS Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Urea UREA Kinetic 340 nm mg/dl I 0.000 mg/dl 300.0 mg/dl TruCal U 3 1 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl DUAL MODE Order information Cat. No. 1 3101 .. .. … Notes 1. Please refer to the package insert for Urea FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume Delay, min. time Slope blank Linearity Limit Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit Substr Depletion Reagent Blank Urea No 25 ml 240 µl 242 µl 3.0 µl 1.0 µl 5 ml 60 µl 60 µl 24, 79 sec No 10.0 % -0.100 Abs 3.000 Abs -0.100 Abs 0.3000 Abs 3.000 Abs Yes (#) Factor # Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Delay, min. time Linearity Limit Low Absorbance High Absorbance R.Abs L.Limit R.Abs H. Limit R.Abs Deviation Reagent blank Urea No 25 ml 300 µl 301 µl Factor # 3.0 µl 2.0 µl 12, 58 sec 10.0 % -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs 0.100 Yes (#) 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons. * can be selected freely # calculated by the analyser 845 3101 90 02 51 September 2008/7 Aplicação de Proteína Urinária para Selectra Proteína Total em Urina FS Referência Apresentação 102109910021 R1 5 x 25 ml + ST 1 x 3 ml SELECTRA Name Abbr. Name Mode Wavelengths Units Decimals Low Conc. High Conc. Calibrator Name Repeat Number Concentration Calibration Mode Interval Cut off Prozone Check Ref. Male Low Ref Male High Ref Female Low Ref Female High Ref Ped. Low Ref Ped. High Control 1 Control 2 Control 3 Correlat. Factor Correlat. Offset Este é um protocolo teórico e foi validado em versão de equipamento disponível na época. Proteína Urinária PTU Endpoint 600 nm mg/dl 0 0.000 mg/dl 3000 mg/dl Notas: 1. Por favor, recorra a bula do produto para informações detalhadas sobre os seguintes testes: Relevância clínica Método e Princípio Composição e Estabilidade dos Reagentes Amostras Calibradores e Controles Desempenho e Características considerando: - Faixa de medição - Especificidade/Interferentes - Sensibilidade/Limite de Detecção - Precisão (Reprodutibilidade, Repetibilidade) - Comparação de método - Valores de referência - Literatura 1 2 * Linear 7 days No No * * * * * * * * * 1.000 0.000 mg/dl 2. A estabilidade do reagente a bordo do analisador é pelo menos e semanas contanto que sejam evitadas a contaminação e a evaporação. DUAL MODE MONO MODE Name Sample blank R1 bottle normal Volume rerun Volume Sample normal Volume rerun volume R2 bottle normal Volume rerun Volume predilution Slope blank Point one, two Low Absorbance High Absorbance R.Abs. L.Limit R.Abs. H.Limit Reagent blank Cal Low limit Cal High limit Factor Name Sample blank R1 bottle normal volume rerun volume Sample normal Volume rerun Volume Incubation time Low Absorbance High Absorbance R.Abs.L.Limit R.Abs. H.Limit PTU No 50 mL 300 µl 301 µl Reagent blank Cal Low limit Cal High limit Factor Yes * * # 6.0 µl 3.0 µl 11.5 min -0.100 Abs 3.000 Abs -0.100 Abs 3.000 Abs Produzido por: DiaSys Diagnostic Systems GmbH Alte Strasse 9 IVD 65558 Holzheim Alemanha Distribuído por: Biosys Ltda. Rua Coronel Gomes Machado, 358 Niterói – Rio de Janeiro (21) 3907-2534 # Entrada pelo usuário * Calibrador + salina como ponto zero. Rev: 0 - 01/2014
© Copyright 2026 Paperzz