Rapid Analysis of Aminothiols by UHPLC with Boron-Doped Diamond Electrochemical Detection Bruce Bailey, Marc Plante, David Thomas, Qi Zhang and Ian Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Thermo Fisher Scientific, 22 Alpha Road, Chelmsford, M Overview Purpose: In order to obtain satisfactory information concerning aminothiols, disulfides, and thioethers from biological samples scientists require a sensitive approach that can measure key compounds, simultaneously. A simple, accurate and rapid UHPLC method was developed for the analysis of these compounds using isocratic liquid chromatography and an amperometric electrochemical cell with a boron-doped diamond (BDD) working electrode. This allowed the accurate quantification of analytes to low picogram (pg) sensitivity. Methods: Direct analysis of aminothiols, disulfides and thioethers using UHPLC chromatographic techniques with a robust electrochemical cell using a BDD electrode is described. Results: The method enables the rapid separation of various aminothiols, disulfides, and thioethers at low levels from whole blood without significant matrix interferences. Introduction A number of biochemically important sulfur-containing compounds occur in vivo including: aminothiols (e.g., cysteine, glutathione [GSH], homocysteine), disulfides (e.g., cystine, glutathione disulfide [GSSG], homocystine), and thioethers (e.g., methionine) as shown in Figure 1. These aminothiols plays numerous physiological roles. GSH is a major cellular antioxidant and a cofactor for glutathione peroxidase, an enzyme that detoxifies hydrogen peroxide and lipid hydroperoxides. The high ratio of GSH/GSSG keeps the cell in a reducing environment, essential for its survival. Decreases in this ratio are associated with cellular toxicity and numerous diseases including neurodegeneration (e.g., Parkinsonʼs disease). Cysteine is also a cellular antioxidant, serves as a precursor to glutathione and is often found in protein structures as a disulfide link. Methionine is an essential amino acid and serves as a methyl donor when incorporated into S-adenosylmethionine. Although a variety of HPLC techniques have been developed for the measurement of thiols, disulfides, and thioethers most of these exhibit technical issues. UV requires derivatization which can adversely affect the GSH/GSSG ratio, while some electrochemical approaches using glassy carbon electrodes suffer from electrode fouling and loss of response, and require routine maintenance. Although porous graphite working electrodes are more forgiving, they still need maintenance. Borondoped diamond (BDD) however, enable the direct measurement of these analytes without electrode issues.1,2 The rapid method presented herein is capable of accurately determining several aminothiols simultaneously using UHPLC chromatographic techniques with a BDD working electrode. Examples showing the analysis of GSH and GSSG from plasma samples using a simple “dilute and shoot” protocol are provided. A UHPLC approach was chosen as it provides several advantages over standard HPLC conditions including: shorter cycle times between samples; improved resolution, and sharper peaks. A sharper peak is important when measuring GSSG as it occurs at a low concentration and is one of the last peaks eluting in biological samples. The use of longer UHPLC columns provides better resolution between compounds, which is important when analyzing biological samples. FIGURE 1. Molecular structures of A) glutathione (GSH), B) glutathione disulfide (GSSG), C) methionine, and D) homocysteine A) GSH (reduced form) C) Methionine B) GSSG (oxidized form) D) Homocysteine 2 Rapid Analysis of Aminothiols by UHPLC with Boron-Doped Diamond Electrochemical Detection Methods Analytical Conditions for Am Column: Pump Flow Rate: Mobile Phase: Column Temperature: Post Column Temperature Injection Volume: Cell Potential: Filter Constant: Cell Clean: Cell Clean Potential: Cell Clean Duration: Sample Prep: FIGURE 2A. Thermo Scientifi FIGURE 2B. 6041RS ultra Am FIGURE 2C. Thermo Scientifi A The Thermo Scientific™ Di - SR-3000 Solvent Rack - DGP-3600RS Dual Gradi - WPS-3000TBPL Thermos - ECD-3000RS Electrochem with integrated temperatu - Chromeleon 6.80 SR12 C Results and Di An instrumental prerequisit inert (free from leachable tr using an electrochemical de biocompatible materials in t contribute to elevated back introduction of the ECD-300 attached in series after the ic, 22 Alpha Road, Chelmsford, MA, USA n of various aminothiols, blood without significant matrix ning compounds occur in vivo [GSH], homocysteine), disulfides cystine), and thioethers (e.g., iols plays numerous physiological ofactor for glutathione peroxidase, lipid hydroperoxides. The high nvironment, essential for its th cellular toxicity and numerous nsonʼs disease). Cysteine is also utathione and is often found in s an essential amino acid and S-adenosylmethionine. Although a for the measurement of thiols, chnical issues. UV requires H/GSSG ratio, while some electrodes suffer from electrode maintenance. Although porous ey still need maintenance. Boronmeasurement of these analytes sented herein is capable of aneously using UHPLC electrode. Examples showing the s using a simple “dilute and shoot” veral advantages over standard tween samples; improved mportant when measuring GSSG he last peaks eluting in biological ides better resolution between ological samples. (GSH), B) glutathione disulfide SG (oxidized form) mocysteine Analytical Conditions for Aminothiol, Disulfide and Thioether Analysis Column: Pump Flow Rate: Mobile Phase: Column Temperature: Post Column Temperature Injection Volume: Cell Potential: Filter Constant: Cell Clean: Cell Clean Potential: Cell Clean Duration: Sample Prep: Thermo Scientific™ Accucore™ RP-MS column 2.6 μm, 2.1 × 150 mm 0.500 mL/min 0.1% pentafluoropropionic acid, 0.02% ammonium hydroxide, 2.5% acetonitrile, water 50.0 °C 25.0 °C 2 µL standards; 4 µL samples Thermo Scientific™ Dionex™ model 6041RS ultra Amperometric Analytical Cell with BDD electrode at +1600 mV 1.0 s On 1900 mV 10.0 s 5–20 µL whole blood + 200 µL 0.4 N PCA, mix and spin for 10 minutes at 13,000 RPM. The clear supernatant was transferred into an autosampler vial and placed on the autosampler at 10 °C. FIGURE 2A. Thermo Scientific™ Dionex™ UltiMate™ 3000 Electrochemical detector FIGURE 2B. 6041RS ultra Amperometric Analytical Cell with BDD electrode FIGURE 2C. Thermo Scientific™ Dionex™ nanoViper™ fingertight capillaries FIGURE 3. Overlay of analytic 3.0 2.8 µA 2.6 2.4 2.2 2.0 GSH s and thioethers using UHPLC chemical cell using a BDD Use of the 6041RS amperome electrochemical capabilities of oxidation of organic compound working electrode materials. Th voltammetric resolution of com fittings were employed to cope particles. These fingertight, virt operate at pressures up to 14,5 tubing which can slip when usi tubing and are available in sma band spreading. Capillaries us connections made prior to the after the injector valve. CysGly on concerning aminothiols, scientists require a sensitive ultaneously. A simple, accurate analysis of these compounds perometric electrochemical cell rode. This allowed the accurate nsitivity. Methods 1.8 1.6 1.4 1.2 1.0 0.8 0.6 0.4 0.2 -0.0 A B 0.0 C The Thermo Scientific™ Dionex™ UltiMate™ 3000 UHPLC system consists of: - SR-3000 Solvent Rack - DGP-3600RS Dual Gradient Pump - WPS-3000TBPL Thermostatted Analytical Autosampler - ECD-3000RS Electrochemical Detector with integrated temperature controlled column compartment - Chromeleon 6.80 SR12 Chromatography Data System software Results and Discussion An instrumental prerequisite for trace analysis is that the HPLC system must be inert (free from leachable transition metals) in order to achieve optimal sensitivity using an electrochemical detector. The system shown above in Figure 2A uses biocompatible materials in the flow path to reduce the influence of metal that can contribute to elevated background currents at the electrochemical cell. The recent introduction of the ECD-3000RS detector enables multiple electrodes to be attached in series after the HPLC column. 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 The direct electrochemical dete doped diamond electrochemica potential used for this study (+1 disulfide analytes. Advantages and method simplicity since no analysis the electrode surface +1900 mV. After a 1.5 minute r again stable and could be used method, the whole blood samp and then centrifuged. The clea autosampler vial and placed on rapid sample processing thus m thiols. Rapid UHPLC analysis a samples within three minutes b place. For biological studies, the dete compounds such as GSH, GSS illustrates the overlay of calibra 1–20 µg/mL. Peak resolution a The column used for this meth material which provides fast, h pressures. When operated at 0 than 300 bar with the last comp in Figure 3. The calibration curves for amin linearity of response to differen coefficients ranging from R2 = 0 (Table 1) over the range of 1–2 (%RSD) for the calibration curv in Table 1. The RSD values ran electrode provided good stabili Thermo Scientific Poster Note • PN70532_E 03/13S 3 A de and Thioether Analysis ific™ Accucore™ RP-MS column 150 mm oropropionic acid, 0.02% ammonium % acetonitrile, water µA 2.0 1 0.8 0.6 0.4 0.2 1.8 1.6 0 5 CysGly HCYS 2.2 GSH CysGly 2.6 2.4 1.2 0 Table 1. Regression data GSSG 2.8 tiMate™ 3000 Electrochemical detector ytical Cell with BDD electrode noViper™ fingertight capillaries 1.4 FIGURE 3. Overlay of analytical standards ranging from 1–20 µg/mL 3.0 blood + 200 µL 0.4 N PCA, mix and utes at 13,000 RPM. The clear as transferred into an autosampler vial the autosampler at 10 °C. 1.6 Methionine s; 4 µL samples ific™ Dionex™ model 6041RS ultra Analytical Cell with BDD electrode at FIGURE 4. Calibration cu Response (µA) Use of the 6041RS amperometric cell (Figure 2B) provides the unique electrochemical capabilities of the boron-doped diamond which enables the oxidation of organic compounds using higher electrode potentials than other working electrode materials. This platform provides both chromatographic and voltammetric resolution of compounds. The nanoViper (Figure 2C) fingertight fittings were employed to cope with the higher pressures due to smaller column particles. These fingertight, virtually zero-dead-volume (ZDV) capillaries can operate at pressures up to 14,500 psi and are much safer to use than PEEK™ tubing which can slip when using elevated pressures. They are made of PeekSil tubing and are available in small internal dimensions to minimize chromatographic band spreading. Capillaries used on this system were 150 micron ID for all connections made prior to the autosampler valve and 100 micron ID for those made after the injector valve. 1.4 Peak 1.2 20 ug/mL 1.0 CysGly 10 ug/mL 0.8 0.6 5 ug/mL 0.4 2 ug/mL 0.2 15 GSH 15 Meth 15 HCYS 1 ug/mL -0.0 # Poin 15 GSSG 15 min 0.0 C ™ 3000 UHPLC system consists of: Autosampler umn compartment Data System software sis is that the HPLC system must be in order to achieve optimal sensitivity em shown above in Figure 2A uses educe the influence of metal that can at the electrochemical cell. The recent nables multiple electrodes to be 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 The direct electrochemical detection of aminothiol compounds only using a borondoped diamond electrochemical cell has been previously described.1 The applied potential used for this study (+1600 mV) was sufficient to oxidize both thiol and disulfide analytes. Advantages of this approach include a stable electrode surface and method simplicity since no sample derivatization is required. After each analysis the electrode surface was regenerated by a 10 second clean cell pulse at +1900 mV. After a 1.5 minute re-equilibration at +1600 mV the electrode was once again stable and could be used for the analysis of the next sample. In the current method, the whole blood sample was added to the perchloric acid media, mixed and then centrifuged. The clear supernatant was then transferred into an autosampler vial and placed on the autosampler at 10 C. This approach enabled rapid sample processing thus minimizing issues related to the instability of the thiols. Rapid UHPLC analysis as shown in Figure 3 enables processing of these samples within three minutes before any major chemical transformations take place. For biological studies, the determination of aminothiol content should include compounds such as GSH, GSSG, methionine, and homocysteine. Figure 3 illustrates the overlay of calibration standards for these compounds ranging from 1–20 µg/mL. Peak resolution and retention time uniformity were both excellent. The column used for this method was the Accucore RP-MS 2.6 micron solid-core material which provides fast, high resolution separations but with lower system pressures. When operated at 0.5 mL/min at 50.0 C the backpressure was less than 300 bar with the last compound (GSSG) eluting under 2.8 minutes as shown in Figure 3. The calibration curves for aminothiol standards are shown in Figure 4. Good linearity of response to different concentrations was obtained with correlation coefficients ranging from R2 = 0.989–1.00 for the five compounds evaluated (Table 1) over the range of 1–20 µg/mL. The percent relative standard deviation (%RSD) for the calibration curves (five concentrations in triplicate) is also shown in Table 1. The RSD values ranged from 1.2% to 8.9%, indicating that the BDD electrode provided good stability during this study. 4 Rapid Analysis of Aminothiols by UHPLC with Boron-Doped Diamond Electrochemical Detection Enhanced peak shape (na of the 6041RS cell with a B volume of only 50 nL contr the noise of the electroche illustrates that amounts les determined by this method shown in Table 2 and rang ratio of 5 are also shown in 5 is also shown in this tabl LOD for GSSG is 175 pg o FIGURE 5. Sensitivity of 0.15 µA 0.12 0.10 0.08 0.06 0.04 CysGly GSH B 0.02 0.00 -0.02 0.0 0.5 1.0 1.5 Table 2. Signal-to-noise r Compound S/N Ratio LOD (pg, S/N of 5) CysG 37 54 FIGURE 4. Calibration curves for standards (1–20 µg/mL, n=3) 1.6 1.2 R² = 0.9997 R² = 0.9892 1 0.8 0 5 10 CysGly 15 0.80 20 25 GSH Methionine HCys GSSG Slope CysGly 15 8.9059 0.989 0.0481 Meth 15 3.4526 0.999 1 ug/mL 1.2458 1.00 15 GSSG 1.00 1.9282 15 0.0509 0.0736 1.00 0.0268 0.0170 4.2 4.4 4.6 4.8 5.0 ompounds only using a boronously described.1 The applied ent to oxidize both thiol and ude a stable electrode surface n is required. After each a 10 second clean cell pulse at 00 mV the electrode was once he next sample. In the current perchloric acid media, mixed en transferred into an 10 C. This approach enabled ated to the instability of the enables processing of these mical transformations take ol content should include homocysteine. Figure 3 ese compounds ranging from ormity were both excellent. RP-MS 2.6 micron solid-core ions but with lower system the backpressure was less g under 2.8 minutes as shown shown in Figure 4. Good obtained with correlation e compounds evaluated t relative standard deviation ns in triplicate) is also shown 9%, indicating that the BDD Enhanced peak shape (narrow peak width) in addition to improved design features of the 6041RS cell with a BDD electrode provides good sensitivity. The small cell volume of only 50 nL contributes to low background currents which helps minimize the noise of the electrochemical cell. The chromatogram shown in Figure 5 illustrates that amounts less than 400 pg on column of each compound can be determined by this method. The signal-to-noise ratios (S/N) for this standard are shown in Table 2 and range from 7.2–37. The limit of detection (LOD) with a S/N ratio of 5 are also shown in this table. The limit of detection (LOD) with S/N ratio of 5 is also shown in this table. For GSH the LOD is approximately 67 pg, while the LOD for GSSG is 175 pg on column. FIGURE 5. Sensitivity of low level analytical standard (400 pg on column) 0.15 µA 0.12 0.50 0.75 1.00 1.25 1.50 1.75 2.0 Figure 6 shows an overlay of chrom and a sample of deproteinized who easily measured in small samples method described. The levels dete reported levels using other techniq below the assay LOD using these detect this compound by simply inc processed. Conclusions • The method for the analysis for th thioethers proved to be simple an measurement in deproteinized wh autoxidation were minimized by r • Although the analysis of aminothi the method could be easily adapt compounds in plasma and tissue • The levels of GSH and GSSG de using other techniques. 0.10 0.08 0.06 0.04 0.02 References GSSG 4.0 Methionine HCYS 3.8 CysGly GSH 3.6 0.25 Table 3. Aminothiol levels (µM) o electrode CysGly GSH Level µM 55.9±13.0 1017 min 3.4 0.20 0.00 Correlation Coefficient. HCYS 0.40 -0.20 RSD. % 1.7446 0.60 -0.00 # Points 15 GSH R² = 0.9998 Peak GSH 2 ug/mL 1.40 1.00 Table 1. Regression data for standard calibration curves 5 ug/mL 1.60 R² = 0.9996 Concentration (µg/mL) 10 ug/mL µA CysGly 0.4 20 ug/mL 2.00 1.20 0.6 0 FIGURE 6. Overlay of chromatog whole blood (blue trace) 1.80 0.2 g from 1–20 µg/mL 3.2 R² = 0.9986 1.4 Response (µA) rovides the unique mond which enables the ode potentials than other both chromatographic and er (Figure 2C) fingertight ures due to smaller column me (ZDV) capillaries can safer to use than PEEK™ s. They are made of PeekSil s to minimize chromatographic re 150 micron ID for all d 100 micron ID for those made 0.2 ug/mL 0.00 -0.02 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 min 5.0 Table 2. Signal-to-noise ratio values calculated for 400 pg on column Compound S/N Ratio LOD (pg, S/N of 5) CysGly 37.0 54 GSH 29.4 67 Methionine 20.4 98 Hcys 7.2 278 GSSG 11.4 175 1. Bailey, B.A. et al. Direct determin levels using HPLC-ECD with a no electrode. Advanced Protocols in 2011, 594, 327-339. 2. Park, H.J. et al. Validation of high diamond detection for assessing 2010, 407, 151-159. 3. Raffa, M. et al. Decreased glutath activities in drug-naïve first-episo 11, 124-131. © 2013 Thermo Fisher Scientific Inc. All rights trademark of SGE International Pty Ltd. All ot Inc. and its subsidiaries. This information is n manners that might infringe the intellectual pr Thermo Scientific Poster Note • PN70532_E 03/13S 5 /mL, n=3) R² = 0.9986 R² = 0.9997 R² = 0.9892 µA 1.60 1.40 1.00 R² = 0.9998 0.80 0.40 Methionine 25 CysGly 20 0.60 0.20 GSSG R² = 0.9996 GSH 1.20 Whole blood -0.00 GSSG Slope .989 0.0509 1.00 0.0268 0.0736 1.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00 4.25 4.50 min 4.75 5.00 Table 3. Aminothiol levels (µM) observed in whole blood (n=3) using the BDD electrode CysGly GSH Methionine GSSG Level µM 55.9±13.0 1017.2±26.2 107.4±52.7 71.1±25.9 0.0481 1.00 .999 5 ug/mL mix -0.20 0.00 relation fficient. 0.0170 improved design features sensitivity. The small cell ents which helps minimize shown in Figure 5 ach compound can be /N) for this standard are ection (LOD) with a S/N on (LOD) with S/N ratio of ximately 67 pg, while the d (400 pg on column) Figure 6 shows an overlay of chromatograms for a standard mixture of aminothiols and a sample of deproteinized whole blood. The levels of GSH and GSSG was easily measured in small samples of whole blood (<20 µL) using the UHPLC method described. The levels detected as shown in Table 3 are within the range of reported levels using other techniques.3 Although the level of homocysteine was below the assay LOD using these small sample volumes, it would be possible to detect this compound by simply increasing the volume of sample collected and processed. Conclusions • The method for the analysis for the measurement of thiols, disulfides and thioethers proved to be simple and reliable with sufficient sensitivity for their measurement in deproteinized whole blood. Technical issues related to GSH autoxidation were minimized by rapid sample preparation techniques. • Although the analysis of aminothiols in this work was related to whole blood, the method could be easily adapted to determine the levels of these compounds in plasma and tissue samples as well. • The levels of GSH and GSSG detected are within the range of reported levels using other techniques. References 0.2 ug/mL 4.0 4.5 min 5.0 00 pg on column e 2.00 1.80 ves 3.5 FIGURE 6. Overlay of chromatograms of standard (black trace) vs. whole blood (blue trace) Hcys 7.2 278 GSSG 11.4 175 1. Bailey, B.A. et al. Direct determination of tissue aminothiol, disulfide, and thioether levels using HPLC-ECD with a novel stable boron-doped diamond working electrode. Advanced Protocols in Oxidative Stress II, Methods in Molecular Biology, 2011, 594, 327-339. 2. Park, H.J. et al. Validation of high-performance liquid chromatography–boron-doped diamond detection for assessing hepatic glutathione redox status. Anal. Biochem. 2010, 407, 151-159. 3. Raffa, M. et al. Decreased glutathione levels and impaired antioxidant enzyme activities in drug-naïve first-episode schizophrenic patients BMC Psychiatry, 2011, 11, 124-131. © 2013 Thermo Fisher Scientific Inc. All rights reserved. PEEK is a trademark of Victrex PLC. Peeksil is a trademark of SGE International Pty Ltd. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. PO70532_E 03/13S 6 Rapid Analysis of Aminothiols by UHPLC with Boron-Doped Diamond Electrochemical Detection www.thermofisher.com/dionex ©2016 Thermo Fisher Scientific Inc. All rights reserved. PEEK is a trademark of Victrex PLC. Peeksil is a trademark of SGE International Pty Ltd. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Australia +61 3 9757 4486 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 China +852 2428 3282 PN70532_E 07/16S Denmark +45 70 23 62 60 France +33 1 60 92 48 00 Germany +49 6126 991 0 India +91 22 2764 2735 Italy +39 02 51 62 1267 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Netherlands +31 76 579 55 55 Singapore +65 6289 1190 Sweden +46 8 473 3380 Switzerland +41 62 205 9966 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA and Canada +847 295 7500
© Copyright 2026 Paperzz