Instructions for Cleaning Platinum Cones Prior to Return to Agilent Platinum cones that are to be returned to Agilent for trade-in must be cleaned to remove any hazardous materials that may have accumulated on the cone during use. Recommended cleaning procedure 1. Clean with pure water a. Dip a cotton swab in pure water and clean both sides of the cone. b. Rinse with pure water. c. Ultrasonic clean the cones in pure water for more than 5 minutes. 2. Clean with a 2% Citranox Acid Cleaner and Detergent solution (Alconox Inc.) a. Ultrasonic clean the cones in a 2% Citranox solution for 2 to 3 minutes. b. Rinse with pure water. c. Ultrasonic clean in pure water for an additional 2 to 3 minutes. 3. Clean with a 2% nitric acid solution a. Dip a cotton swab in a 2% nitric acid solution and clean both sides of the cone. b. Rinse with pure water. c. Ultrasonic clean in pure water for 2 to 3 minutes. Be sure to note this method on the EHS form that must accompany the returned cone. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. © Agilent Technologies, Inc., 2012 Printed in the USA May 3, 2012 5991-0450EN
© Copyright 2026 Paperzz