Instructions for Cleaning Platinum Cones Prior to Return to Agilent

Instructions for Cleaning Platinum
Cones Prior to Return to Agilent
Platinum cones that are to be returned to Agilent for trade-in must be cleaned to
remove any hazardous materials that may have accumulated on the cone during
use.
Recommended cleaning procedure
1. Clean with pure water
a. Dip a cotton swab in pure water and clean both sides of the cone.
b. Rinse with pure water.
c. Ultrasonic clean the cones in pure water for more than 5 minutes.
2. Clean with a 2% Citranox Acid Cleaner and Detergent solution
(Alconox Inc.)
a. Ultrasonic clean the cones in a 2% Citranox solution for 2 to 3 minutes.
b. Rinse with pure water.
c. Ultrasonic clean in pure water for an additional 2 to 3 minutes.
3. Clean with a 2% nitric acid solution
a. Dip a cotton swab in a 2% nitric acid solution and clean both sides of the cone.
b. Rinse with pure water.
c. Ultrasonic clean in pure water for 2 to 3 minutes.
Be sure to note this method on the EHS form that must accompany the returned
cone.