Water Test Kit I N S T R U C T I O N S – to be used for determining the percentage of water in most lube and/or hydraulic oils. Water Test Kit Instructions The WATER TEST KIT can be used to determine the percentage of water in various fluids. Typically these fluids are hydraulic and lubrication oils; however, other fluids can also be tested. The WATER TEST KIT has five ranges from .005% to 12% water. Table of Contents: Safety Precautions & Warnings 2 Kit Contents 3 Principle of Operation 3 Nomenclature & Part Numbers 4 .15% Low Range Test Procedure: 50 to 1,500 PPM; .005% to .15% 5 1.5% High Range Test Procedure: 0 to 15,000 PPM; .05% to 1.5% 6 3.0% and 6.0% Range 7 12.0% Range 8 Cleaning Vessel 8 Conversion Charts: Gauge Reading to % Water 9 Beyond 12% Water Content 10 Temperature Compensation Chart 11 Water Test Kit – page 1 (of 11) Water Test Kit Instructions Caution: Using the water test kit can be dangerous! You must read and understand the following precautions before you begin to use the water test kit! WARNING: The packets of Reagent A in this kit contain Calcium Hydride which can cause eye and skin burns. Calcium hydride, in contact with water, will produce hydrogen gas which is highly flammable. Avoid getting the chemical on your skin or in your eyes. Keep the kit and Reagent A away from water and water vapor. DO NOT use the kit near an open flame or sparks. DO NOT permit smoking when the kit is being used. Never use more than one packet of Reagent A per test. Excessive hydrogen gas could be produced causing injury to you or damage to the test vessel. FIRST AID: In case of contact with Reagent A powder, immediately flush eyes or skin with water for at least 15 minutes. Remove contaminated clothing. Wash skin contact area with soap and water. Call a physician. Water Test Kit – page 2 (of 11) Water Test Kit Instructions Water Test Kit can be used to test for water content measured as Parts Per Million or as a % of volume. The test can be performed on any liquid which is not based on water. On the lower two ranges, the direct reading gauge displays from 50 to 15000 Parts Per Million (.005% to 1.5%), with a resolution of 20 PPM (.005%) water. Any fluid can be tested as long as the two following requisites are met: A, the water MUST be in the form of free molecules or droplets and not be chemically or physically bound as in emulsions; B, the fluid should not attack the water vessel components which are anodized aluminum, fluorocarbon gasket, brass, and the epoxy used to mount the Reagent Chamber. Principle of Operation Reagent A reacts with water and produces hydrogen gas. When a sample and a dry solvent (Reagent B) are placed in the vessel in the proper ratio, the pressure produced is proportional to the amount of water present. A sample with .1% (1000 PPM) water content will produce 10 pounds per square inch (PSI) @ 25° C. Since the pressure gauge has a full scale value of 15 PSI, a maximum reading of 0.15% (1500 PPM) water is possible on the low range. The smallest division of the gauge on the PSI scale is 0.2 PSI or .002% (20 PPM). The gauge’s pointer can easily point to the space between the smallest divisions, thus providing the ability to read a result down to the nearest 0.001% (0.1 PSI) or 10 PPM of water. There are a total of five ranges with the following full scale values: 0.15%, 1.5%, 3.0%, 6.0%, and 12.0%. CAUTION: When releasing pressure from the vessel via the “Pressure Relief Button,” DO NOT aim the discharge spray at your face or eyes! When using the test vessel, the operator should make sure the pressure does not exceed 15 PSI, since over-pressurizing the gauge will damage it. Should you encounter a sample with a water content which will produce a pressure equal to or greater than 14 PSI, release the pressure immediately. The preferred procedure is to watch the gauge as the reaction takes place, observing the pressure. When the pressure reaches 14 PSI, intervene and stop the test by pressing the relief valve or loosening the lid. Water Test Kit – page 3 (of 11) Kit Contents: Each Water Test Kit contains the following items: Water Test Vessel consisting of cup, cap, gauge and gasket 1 ea Box of Reagent A containing 50 packets 1 ea Syringe 10 ml for sample 1 ea Syringe 30 ml for Reagent B 1 ea Instructions 1 ea Physical Data: P567855 Size: 4.25 H x 2.5 Dia. Weight: 9.8 oz. Material: anodized aluminum, chrome plated brass, fluorocarbon gasket, teflon tape, epoxy Water Test Kit Instructions Nomenclature & Part Numbers: The illustrations on this page identify the parts of the WATER TEST KIT Water Test Kit Contains: ITEM PART# Reagent A, Box 50 packets P567851 Test Vessel (Complete) P567854 Vessel (Cap & Cups) P567855 15 PSI Gauge P567856 Fluorocarbon Gasket P567857 Syringe 10 ml P567858 Syringe 30 ml P567859 Reagent B 500 ml GAUGE GASKET PRESSURE RELIEF BUTTON OIL-REAGENT A CHAMBER REAGENT B CHAMBER Cap Water Test Kit – page 4 (of 11) Cup Water Test Kit Instructions Low Range Test Procedure .15% Full Scale 1 Using the 30 ml syringe, measure 30 ml of sample oil and inject it into the OilReagent B Chamber. Be careful not to spill any fluids into the Reagent A Chamber. 2 Using the 10 ml syringe, measure 10 ml of Reagent B and inject it into the Oil-Reagent B Chamber. 3 Prepare a packet of Reagent A (according to the Reagent A Instruction sheet). Empty the rolled-up packet of Reagent A into the Reagent A Chamber, being careful not to spill any powder into the Oil Solvent Chamber. Place the rolledup packet with the open end facing down, into the Reagent A Chamber and flick the bottom of the packet a few times with your finger. 4 Keeping the cup vertical so as not to spill or mix the contents, tightly screw the cap onto the cup making sure there is a very tight seal on the gasket. 5 Shake the vessel vigorously for 20 seconds, then observe the pressure gauge to insure that the pressure has not exceeded 14 PSI. If the pressure has reached 14 PSI, STOP the test by pressing the relief valve or loosening the cap. Go to the next section titled “1.5% Full Scale” for instructions on testing samples with water content greater than .14%. 6 Shake the vessel vigorously for another 20 seconds, and then observe the pressure gauge to insure that the pressure has not exceeded 14 PSI. 7 Shake the vessel vigorously AGAIN for another twenty seconds, observing the pressure gauge to insure that the pressure has not exceeded 14 PSI. 8 Set the vessel down and wait one minute. Shake the vessel for 10 seconds once every minute and take the final reading 5 minutes after you have started the test. 9 Measure the temperature of the oil-Reagent B mixture and refer to the temperature compensation chart to correct the reading due to the effects of temperature. Reading the Gauge Correlation of Data: Low Range – Multiply the Gauge Reading by 0.01 10.0 PSI = 0.100% water content or 1,000 PPM 1.0 PSI = 0.010% water content or 100 PPM 0.2 PSI = 0.002% water content or 20 PPM 0.1 PSI = 0.001% water content or 10 PPM The scale is from 0 to 15 PSI (pounds per square inch) with the smallest scale division being 0.2 PSI. See Page 9 for Pressure Conversion Charts for the .15 % range Water Test Kit – page 5 (of 11) (See Page 8 - Cleaning Vessel) Water Test Kit Instructions High Range Test Procedure 1.5% Full Scale 1 Using the 10 ml syringe, measure 4 ml of sample oil and inject it into the OilReagent B Chamber. Be careful not to spill any fluids into the Reagent A Chamber. 2 Using the 30 ml syringe, measure 16 ml of Reagent B and inject it into the Oil-Reagent B Chamber. 3 Prepare a packet of Reagent A according to the Reagent A Instruction sheet. Empty the rolled-up packet of Reagent A into the Reagent A Chamber, being careful not to spill any powder into the Oil-Reagent B Chamber. Place the rolled-up packet with the open end facing down, into the Reagent A Chamber and flick the bottom of the packet and flick the bottom of the packet a few times with your finger. 4 Keeping the cup vertical so as not to spill or mix the contents, tightly screw the cap onto the cup making sure there is a very tight seal on the gasket. 5 Shake the vessel vigorously for twenty seconds, then observe the pressure gauge to insure that the pressure has not exceeded 14 PSI. If the pressure has reached 14 PSI, STOP the test by pressing the relief valve or loosening the cap. Go to the next section titled “3.0% Range” for instructions on testing samples with water content greater than 1.4%. 6 Shake the vessel vigorously for another twenty seconds, and then observe the pressure gauge to insure that the pressure has not exceeded 14 PSI. 7 Shake the vessel vigorously AGAIN for another twenty seconds, observing the pressure gauge to insure that the pressure has not exceeded 14 PSI. 8 Set the vessel down and wait one minute. Shake the vessel for 10 seconds once every minute and take the final reading 5 minutes after you have started the test. 9 Measure the temperature of the oil-Reagent B mixture and refer to the temperature compensation chart to correct the reading due to the effects of temperature. Correlation of Data: High Range – Multiply the Gauge Reading by 0.1 10.0 PSI = 1.00% water content or 10,000 PPM Reading the Gauge 1.0 PSI = 0.10% water content or 1,000 PPM 0.2 PSI = 0.02% water content or 200 PPM 0.1 PSI = 0.01% water content or 100 PPM The scale is from 0 to 15 PSI (pounds per square inch) with the smallest scale division being 0.2 PSI. See Page 9 for Pressure Conversion Charts for the 1.5 % range Water Test Kit – page 6 (of 11) (See Page 8 - Cleaning Vessel) Water Test Kit Instructions Should the fluid sample have greater than 1.4 % water content, the concentration can still be determined by using the following procedures: Test Procedure (Normal): 1.5 % Range Sample Reagent B 4 16 ml ml Total Volume 20 ml Test Procedure: 3.0 % Range For example, let’s say that after using 2 ml of sample and 18 ml of Reagent B, the test produces a reading of 12.0 PSI on the gauge. To convert this reading to the actual % of water, use the following calculation: Water % = Gauge Reading x .2 Use Sample Volume 2 Reagent B Volume 18 ml ml Total Volume ml 20 Water % = 12.0 x .2 = 2.4% See page 9 for pressure Conversion Charts for the 3.0% range Test Procedure: 6.0 % Range Should the reading on the gauge exceed 14 PSI again, stop the test and prepare a new test using 1 ml of sample and 19 ml of Reagent B. Using the following equation, determine the percentage of water: Water % = Gauge Reading x .4 Example: gauge reads 11.5 PSI Water % = 11.5 x .4 = 4.6% See page 9 for pressure Conversion Charts for the 6.0% range Water Test Kit – page 7 (of 11) Water Test Kit Instructions 12 % Range Should the reading on the gauge exceed 14 PSI again, stop the test and prepare a new test using 1/2 ml of sample and 19-1/2 ml of Reagent B. Using the following equation, determine the percent water: Water % = Gauge Reading x .8 Example: gauge reads 10.0 PSI Water % = 10 x .8 = 8.0% See page 9 for pressure Conversion Charts for the 12% range Cleaning the Vessel After each use, be sure to rinse out the cap and vessel thoroughly before storing the Water Test Kit. You can use any petroleum solvent to clean the vessel. Remove all traces of Reagent A (undissolved particles). Water Test Kit – page 8 (of 11) Water Test Kit Instructions Conversion Charts: Gauge Reading to % Water Gauge Reading = GR Range 0.15% GR x .01 GR % Water 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Range 1.5% GR x .1 GR % Water 0.01% 0.02% 0.03% 0.04% 0.05% 0.06% 0.07% 0.08% 0.09% 0.10% 0.11% 0.12% 0.13% 0.14% 0.15% 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Range 6% GR x .4 GR 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 % Water 0.4% 0.8% 1.2% 1.6% 2.0% 2.4% 2.8% 3.2% 3.6% 4.0% 4.4% 4.8% 5.2% 5.6% 6.0% Water Test Kit – page 9 (of 11) Range 3% GR x .2 GR % Water 0.1% 0.2% 0.3% 0.4% 0.5% 0.6% 0.7% 0.8% 0.9% 1.0% 1.1% 1.2% 1.3% 1.4% 1.5% 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Range 12% GR x .8 GR % Water 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 0.8% 1.6% 2.4% 3.2% 4.0% 4.8% 5.6% 6.4% 7.2% 8.0% 8.8% 9.6% 10.4% 11.2% 12.0% 0.2% 0.4% 0.6% 0.8% 1.0% 1.2% 1.4% 1.6% 1.8% 2.0% 2.2% 2.4% 2.6% 2.8% 3.0% Water Test Kit Instructions Beyond 12% Water Content It is possible to measure water content above 12%. The accuracy will depend on your laboratory technique and should be within 5 to 10% of reading. The technique used to prepare the sample is known as serial dilution. The process involves taking a sample with high water content and diluting it with a dry solvent. The resulting mixture is a sample with water content in the range of the Water Test Kit (WTK). This diluted sample mixture is tested using the WTK and the results are multiplied by a known factor to arrive at the original sample’s water content. The chart below shows how to take a high water content sample and make an approximate 2% sample for testing on the WTK’s 3 % range. The chart also shows the multiplication factor needed to convert the results of the 3% range test back to the original sample’s water content. The following is an example of the process. Process Example For this example, we use an oil sample with an assumed water content of 35%. In the chart below, go down the first column labeled “% Water in Sample” to 35%. In the second column labeled “Sample”, you will see that it directs you to take 2.9 ml of the high water content sample and mix it with 47.1 ml of dry solvent. Mix the diluted sample (DS) well and then perform the 3 % range test which is found on page 7 (i.e. Take 2 ml of DS and 18 ml of Reagent B). In this example the gauge reading will be 10 PSI. Water % = gauge reading x .2 (10 PSI x .2 = 2.0%). To convert the results of the 3% range test back to the original sample’s water content, multiply 2% x 17.5 (the factor listed below at 35 % level). 2 x 17.5 = 35% % Water in Sample 100 90 80 70 60 50 40 35 30 25 20 15 10 Sample 1.0 ml 1.1 ml 1.3 ml 1.4 ml 1.7 ml 2.0 ml 2.5 ml 2.9 ml 3.3 ml 4.0 ml 5.0 ml 6.7 ml 10.0 ml Water Test Kit – page 10 (of 11) Solvent 49.0 ml 48.9 ml 48.8 ml 48.6 ml 48.3 ml 48.0 ml 47.5 ml 47.1 ml 46.7 ml 46.0 ml 45.0 ml 43.3 ml 40.0 ml Factor 50.0 45.0 40.0 35.0 30.0 25.0 20.0 17.5 15.0 12.5 10.0 7.5 5.0 Water Test Kit Instructions Temperature Compensation Factors °F Factors °F Factors °F Factors °F Factors 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 1.0916 1.0894 1.0872 1.0850 1.0828 1.0806 1.0784 1.0762 1.0741 1.0719 1.0698 1.0677 1.0656 1.0634 1.0613 1.0592 1.0572 1.0551 1.0530 1.0509 1.0489 1.0468 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 1.0448 1.0428 1.0407 1.0387 1.0367 1.0347 1.0327 1.0307 1.0288 1.0268 1.0248 1.0229 1.0209 1.0190 1.0171 1.0151 1.0132 1.0113 1.0094 1.0075 1.0056 1.0037 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 1.0019 1.0000 0.9981 0.9963 0.9944 0.9926 0.9908 0.9889 0.9871 0.9853 0.9835 0.9817 0.9799 0.9781 0.9763 0.9746 0.9728 0.9710 0.9693 0.9675 0.9658 0.9641 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 0.9623 0.9606 0.9589 0.9572 0.9555 0.9538 0.9521 0.9504 0.9487 0.9470 0.9454 0.9437 0.9420 0.4040 0.9387 0.9371 0.9355 0.9338 0.9322 0.9306 0.9290 0.9274 0.9258 °C Factors °C Factors °C Factors °C Factors °C Factors 0 1 2 3 4 5 6 7 8 9 1.0916 1.0876 1.0836 1.0797 1.0758 1.0719 1.0681 1.0643 1.0605 1.0567 10 11 12 13 14 15 16 17 18 19 1.0530 1.0493 1.0456 1.0420 1.0383 1.0347 1.0311 1.0276 1.0241 1.0205 20 21 22 23 24 25 26 27 28 29 1.0171 1.0136 1.0102 1.0068 1.0034 1.0000 0.9967 0.9933 0.9900 0.9868 30 31 32 33 34 35 36 37 38 39 0.9835 0.9803 0.9770 0.9739 0.9707 0.9675 0.9644 0.9613 0.9582 0.9551 40 41 42 43 44 45 46 47 48 49 0.9521 0.9490 0.9460 0.9430 0.9401 0.9371 0.9342 0.9313 0.9283 0.9255 Instructions: For temperature compensation, multiply the above factor times the gauge reading for the temperature of the sample in the test vessel. The product is the corrected percent water. Water Test Kit – page 11 (of 11) Portable Fluid Analysis Kit reorder component information Part Number P567860 Description Solvent dispensing bottle filters P567861 Sample bottles (50/carton) P567862 1000 ml solvent dispensing bottle P567863 Membrane holder & funnel assembly P567864 100X Microscope P176433 Easy-vac Hose P567865 3x5 analysis cards (50/pack) P567866 Beaker P567858 Syringe 10 ml P567859 Syringe 30 ml P567854 Test Vessel (complete) cap & cup, 15 PSI gauge, fluorocarbon gasket. P567869 0.8 Micron membrane filters (100/pack) P567868 5.0 Micron membrane filters (100/pack) P567851 Reagent A Powder (50 packets per kit) P176431 Hand-actuated vacuum pump P567863 Filter funnel P567855 Vessel (cap & cups) P567856 15 PSI Gauge P567857 Fluorocarbon gasket P567912 Laminate sheets (100/pack) Filtration Solutions Donaldson Company, Inc. PO Box 1299 Minneapolis, MN 55440-1299 U.S.A. Tel 800.846.1846 Fax 952.703.4652 [email protected] www.donaldson.com/en/ih Information in this document is subject to change without notice. © 2007 Donaldson Co., Inc. I n n o v a t i v e S o l u t i o n s f o r a l l Yo u r F i l t r a t i o n N e e d s Systems Systems for Sale and/or Rent Contact us for information regarding our Rental Fleet & Pricing High Flow Flushing Skids Low Flow Filter Carts Vacuum Dehydrators INDUSTRIAL FLUID PURIFICATION Coalescers LTC Transformer Filtration Rental Fleet Turnkey Flushing Services at your jobsite, performed by qualified factory trained technicians. Filter Elements: For Oil, Air, Fuel and Water applications. Filtration Solutions Donaldson Company, Inc. PO Box 1299 Minneapolis, MN 55440-1299 U.S.A. Tel 800.846.1846 Fax 952.703.4652 [email protected] www.donaldson.com/en/ih Information in this document is subject to change without notice. © 2007 Donaldson Co., Inc.
© Copyright 2026 Paperzz