VacStar 50 NEO and VacStar 80 NEO Pre

P R E - I N S TA L L AT I O N
G U I D E for MODELS
DENTAL VACUUM SYSTEMS
VS50NEO & VS80NEO
C he ck VacS t ar NEO Model Bein g I n st alled:
 VS50NEO
 VS80NEO
Doctor:__________________________________________________
Address:__________________________________________________
Phone#:__________________________________________________
Dealer:__________________________________________________
Dealer Address: __________________________________________________
PRODUCT SPECIFICATIONS
Electrical
Voltage Rating (VAC)
Voltage Min./Max. (VAC)
Frequency Single Phase (Hz)
Full Load (Amps)
Model VS50NEO
Model VS80NEO
220
220
198/242
198/242
50/60
50/60
20
28
20 - 100
20 - 100
0.13
0.13
40 - 75
40 - 75
10
10
Water
Inlet Water Pressure (PSI)
Flow Rate (gal/min) per Pump
Inlet Water Temperature (°F)
Vacuum Level
Preset at Factory (in Hg)
All installations must conform to local codes.
PHYSICAL CHARACTERISTICS
Model VS50NEO
Model VS80NEO
Height
23 inches (58 cm)
Height
23 inches (58 cm)
Width
28 inches (71 cm)
Width
28 inches (71 cm)
Depth
18 inches (46 cm)
Depth
18 inches (46 cm)
18 in.
18 in.
28 in.
28 in.
23 in.
23 in.
Shipping Weight
Shipping Weight
175 lbs. (79 kg)
175 lbs. (79 kg)
ELECTRICAL CONNECTIONS
220V Connections:
 Each pump has its own 6-foot BX cable pre-wired to the electrical connection box for direct wiring to dedicated
220VAC, single phase 50/60 Hz circuit via a disconnect box with approved ground.
 Disconnect box should be mounted no more than 3 feet of each other and 3 feet of installation center line.
 The supplied cable is pre-wired in the electrical box of each pump. Connect cable wires by matching color.
Important: When not connecting a 24VDC remote switch, the ORG and YEL wires must
be tied together and the BRN wire insulated for pump operation.
24VDC Connections:
 24VDC Connections are used when installing a 24VDC remote switch. They are made via a supplied interconnect cable that is also pre-wired to the electrical connection box of each pump.
 Make the 24VDC connections shown below for the remote switch. When not using the remote switch, insulate
the BRN wire and tie the ORG and YEL wires together.
Green
Yellow Stripe
Remote Switch
Terminal Board
BRN
Black
ORN
YEL
Supplied 6-Foot BX Cable
from the Electrical Connection
Box of Each Pump
ED
CL
BRN
Interconnect Cable
ORN
Interconnect Cable
YEL
White
GRN
2
Yel
3
Org
Remote Switch
4
Brn
Connection without
24VDC Switch
220VAC Connections
A S S IF I
Interconnect Cable
24VDC Connections
MEDICAL ELECTRICAL EQUIPMENT
WITH RESPECT TO ELECTRICAL SHOCK, FIRE, MECHANICAL
AND OTHER SPECIFIED HAZARDS ONLY
IN ACCORDANCE WITH UL-60601-1, CAN/CSA C22.2 NO.601.1 66CA
UTILITY ROOM CONFIGURATION
VacStar NEO without Wall-Mounted Air/Water Separator
Note:
All VacStar NEOs are hard-wired
directly to the VacStar’s electrical box
via a 4-inch handy box complying
with local electrical codes. The 4-inch
handy box is not supplied.
Intake from Main Line
terminate with 1” FNPT fitting
Water Suppy 1/2”
copper tube terminate with 1/2” FNPT
shut-off valve
Drain
Hose
Floor Drain
Note: See Optional Drain Connections
VacStar NEO with Wall-Mounted Air/Water Separator
Intake from Main Line
terminate with 1” FNPT fitting
Water Suppy 1/2”
copper tube terminate with 1/2” FNPT
shut-off valve
Vent
Vent to outside with 2” schedule 40 pipe
Air/Water
Separator
WARNING:
Condensation of water will occur in vent piping. Slope piping
toward separator to avoid accumulation of water in vent.
Note:
All VacStar NEOs are hard-wired
directly to the VacStar’s electrical box
via a 4-inch handyPower
box complying
Connection
with local electrical codes.
The 4-inch
handy box is not supplied.
32 Inches
Maximum.
Drain
Hose
Floor Drain
Note: See Optional Drain Connections
All VacStar NEO vacuums comply with NFPA 99C level 3 requirements
PLUMBING INSTALLATION
SUB FLOOR INSTALLATION - The sub-floor plumbing layout shown below is the recommended layout for
VacStar system installations and should be used whenever possible.
2-Inch
Exhaust/Vent
Ceiling
Interior Wall
Notes: 1. See Optional Drain Connections
shown below.
Black
Flex
Tubing
2. 8-Foot Maximum Height from Main
Line to pump.
See
Note 1
Junction
Box
1/2-Inch Diameter
Riser
Minumum Slope:
1/4 inch per 10 Feet
From
Dental Unit
1/2 Inch Riser
45°Elbow
5. When main line is 1-1/2" I.D. or
larger, use 45° Y's and elbows only.
6. Long radius 90° elbows can be used
as alternates to 45° elbows.
Riser - See Note 2
MAIN LINE
See Plumbing Requirements for Main
Line Diameters needed for specific
units. See Notes 3, 4 & 5.
45°Y
Main Line
3. Consult Dental Unit Manufacturer's
Guidelines for correct reduced size
and height of termination of vacuum
line inside junction box.
4. Limit branches. Orient main line
under junction box or cabinet.
Floor
Sink
To
Pump
RISER to MAIN LINE DETAIL
OVERHEAD INSTALLATION - The overhead plumbing layout shown below is the alternate layout for VacStar
system installations and should be used only when unable to use the sub-floor plumbing layout.
45°Elbow
45°Elbow
MAIN LINE
See Plumbing Requirements for Main Line
Diameters needed for specific units.
1" Min
1/2 Inch Riser
Main Line
Interior Wall
1/2-Inch
Riser
Ceiling
2-Inch
Exhaust/Vent
1/2-Inch Diameter Riser
Dental
Unit
10-Foot Maximum Height from
Riser Trap to Main Line
Black
Flex
Tubing
Floor
Sink
Junction
Box
1/2-Inch
RISER TRAP DETAIL
(Using 45° Elbows)
45°Y
To
STS Pump
RISER to MAIN LINE DETAIL
Minumum Slope:
1/4 inch per 10 Feet
To Main
Line
1/2 Inch
Diameter
Note: Consult Dental Unit Manufacturer's Guidelines for
correct reduced size and height of termination of
vacuum line inside junction box.
Note: See Optional Drain
Connections
CONNECTION DETAILS ALL INSTALLATIONS 45° Elbow
MAKING
TURNS
MAIN LINE
O DO NOT use 90° elbows except where main
line terminates at vacuum pump.
CLEARING AN
OBSTRUCTION
DRAIN OPTIONS
Direct connection to vented
drain. No traps before vent.
OR
O Use only 45° elbows to make turns in main line.
45° Elbow
O If piping is diverted to clear an obstruction,
DO NOT MAKE A TRAP.
CLOSED VENTED DRAIN
All installations must conform to local codes.
Indirect connection
(Air gap) with a p-trap.
6”
1-1/2 Inch
P-Trap
OPEN DRAIN
PIPE
SITE REQUIREMENTS
Electrical
Minimum Circuit Breaker Rating
Minimum Wire Gauge Size
220-Volt Buck/Boost Transformer
VS50NEO
VS80NEO
2 X 15 or 30A
2 X #12 AWG or
1 X #10 AWG
#67002 (3.4 KVA)
2 X 20 or 40A
2 X #12 AWG or
1 X #8 AWG
#67002 (3.4 KVA)
Plumbing
Note:
Suction piping must slope at least a ¼” for each 10 feet of run towards the pump.
Use PVC Schedule 40 or Copper Type M.
Minimum CFM @ 0" Hg
Air Exhaust with Air/Water Separator
30
2" schedule 40 pipe
40
2" schedule 40 pipe
Overhead Plumbing
Main Line Diameter (Min./Max. ID in inches)
End Fitting
Overhead Main Line
1 to 1½"
/ ” FNPT
½" ID
1¼ to 2"
/ ” FNPT
½" ID
3 4
3 4
1 to 1½"
/ ” FNPT
3/4 ” to 1½”
3 4
Floor Plumbing
Main Line Diameter (Min./Max. ID in inches)
End Fitting
Branch Line Diameter (Min./Max. ID in inches)
3 4
1¼ to 2"
/ ” FNPT
1 to 1½”
Environmental
40 to 104°F
(5 to 40°C)
Ambient Temperature
(See ventilation requirements below.)
40 to 104°F
(5 to 40°C)
VENTILATION REQUIREMENTS
Equipment Room Temperature
The VacStar NEO equipment must be used in a controlled temperature environment. Maintain equipment room temperature between 40 and 105 degrees Fahrenheit. Adequate forced ventilation must be provided across the unit by
placing an appropriate exhaust fan opposite an equivalent air intake vent. The fan should be higher than the associated intake vent.
Exhaust Vent Protection.
If the exhaust piping is venting to the outside of the building, precautions must be taken to protect the equipment room
from weather elements and animal intrusion. This can be accomplished by using one of the three methods shown
below.
Shroud &
Screen
Screen
Roof-Mounted Outside Vent Protection
Shroud &
Screen
Wall-Mounted Outside Vent Protection
Make sure to use the required pipe type for associated system
NOTES
NOTES
For over 50 years, Air Techniques has been a leading innovator and manufacturer of dental
products. Our priority is ensuring complete satisfaction by manufacturing reliable products
and providing excellent customer and technical support. Whether the need is digital imaging,
utility room equipment or merchandise, Air Techniques can provide the solution via our network
of authorized professional dealers. Proudly designed, tested and manufactured in the U.S., our
products are helping dental professionals take their practices to the next level.
Air Techniques’ family of quality products for the dental professional include:

Digital Imaging
•
•
•
•
•
•

Utility Room
•
•
•
•
•
•

Digital Radiography
Intraoral Camera
Caries Detection Aid
Intraoral X-ray
Panoramic X-ray
Film Processors
Dry Vacuums
Wet Vacuums
Air Compressors
Amalgam Separator
Utility Accessories
Utility Packages
Merchandise
•
•
•
•
•
•
•
•
Surface Disinfectant
Enzymatic Cleaner
Hand Sanitizer and Lotion
Waterline Cleaner
Evacuation System Cleaner
Imaging Accessories
Chemistry
Processor Accessories
Corporate Headquarters
1295 Walt Whitman Road | Melville, New York 11747- 3062 | Phone: 800-247-8324 | Fax: 888-247-8481
Western Facility
291 Bonnie Lane, Suite 101 | Corona, CA 92880 - 2804 | Phone: 800-247-8324 | Fax: 951-898-7646
www. air tec hniques.c om
VacStar is a registered trademark of Air Techniques, Inc.
© Air Techniques, Inc. • P/N K1106 Rev. A • February 2016