Application Note # CA-274195 Low level speciation of chromium in drinking waters using LC-ICP-MS Introduction Instrumentation Measuring only the total concentration of a particular element in environmental or clinical samples can often lead to errors in assessing toxicity, mobility, and bioavailability, as these effects can differ greatly between the various chemical species in which an element may occur. Mercury, arsenic, chromium, lead, selenium and tin can all be found in numerous organic and inorganic compounds, and will have varying effects on an ecosystem depending on their form. The LC system used was comprised of an inert Tertiary Solvent Delivery Module, a biocompatible autosampler fitted with a 200 μL sample loop, and a Anion exchange column (Hamilton PRP-X100, 4.1 mm ID x 250 mm, 10 μm) and PEEK™ guard column. When measuring Cr, it is important to make sure the LC systems are metal free. When using LC components, be sure to select the inert or biocompatible versions. Chromium (Cr) is a good example of this, although in this case the key distinction is the ionic state of the chromium ion. The inorganic trivalent form of chromium (Cr III) is relatively non-toxic and is an essential micronutrient. Hexavalent chromium (Cr VI) however, is a known carcinogen although it is widely used in numerous industries including plastics, metalworking and dyes/paints/inks. Many government environmental and occupational safety regulatory agencies have established regulations, guidelines and exposure limits regarding the use, storage and disposal of hexavalent chromium-containing materials.[1],[2],[3],[4] The aurora M90 ICP-MS was used for all chromium measurements, operated in high sensitivity mode using the standard sample introduction system. The ICP-MS setup allows automatic optimization of both ion optics and gas flows (plasma and CRI gases). Before connecting the ICP-MS to LC, it was tuned automatically by using the auto optimization routine included in the software. The CRI system was also optimized to maximize Cr52 sensitivity in the presence of a high Ar40C12 background signal, resulting in an approximate 30 times improvement in signal-to-noise when blank eluent solution was being measured. The ICP-MS nebulizer was then connected to the LC column using a 20 cm length of PEEK tubing (0 25 mm ID). The mobile phase used for this experiment, ammonium nitrate, can also be used for the simultaneous determination of selenium [8], [9]. Galaxie software was used to control the operation of the ICPMS software and hardware during data acquisition and to process all of the data. See Tables 1 to 3 for the ICP-MS and LC conditions used. Inductively Coupled Plasma Mass Spectrometry (ICPMS) is a very powerful analytical tool for measuring trace and ultra-trace concentrations of metals and metalloids. When coupled to a Liquid Chromatography (LC) system, species elute one by one from the LC column directly to the ICP-MS so that low level detection and compound-specific identification can be achieved simultaneously. [5], [6] Interfacing the LC system to ICP-MS is very straightforward; the LC column output is directly connected to the nebulizer of the ICP-MS by a piece of PEEK tubing, and no other hardware changes are required to either the LC or the ICP-MS. Using an aurora M90 ICP-MS with Collision/Reaction Interface (CRI) [7] technology allows for the elimination of major polyatomic interferences formed in the plasma, allowing for even lower detection limits of the analytes of interest. All instrument control during analysis and post-run data processing used Bruker’s Galaxie Workstation Data System, which offers full control of the ICP-MS, LC pump and autosampler for automated, unattended operation. Reagents and samples Deionized water (18MΩ/cm, Millipore MilliQ, Billerica, MA, USA) was for mobile phase preparation. Mobile phase (LC) Suprapur ammonia (Merck, Kilysth, Victoria, Australia) and AR Select Plus nitric acid (Mallinckrodt Baker, Phillipsburg, NJ, USA) were used to produce the 60mM ammonium nitrate mobile phase. The pH of the mobile phase was 9.3. The mobile phase was prepared daily. Calibration standards Chromium (+3) and chromium (+6) aqueous standards were obtained from Inorganic Ventures, Inc., (Lakewood, NJ, USA). curves shown calibrate successfully from 50 ng/L to 100 µg/L, with calibration correlations of 0.999 or better, thus demonstrating the linearity and reproducibility of the technique. All quantitation was performed on peak area. Figure 3 illustrates the speciation of Cr(III) and Cr(VI). Chromium(III) EDTA complex To stabilize the Cr(III) species and allow the determination of both species in a single chromatographic run, each of the standards and samples were incubated at 70 ºC with 0.2 mM of EDTA (disodium salt, from Sigma Aldrich, Castle Hill, NSW Australia) for 1.5 hours. When reacting with EDTA, it is then the Cr(III)-EDTA complex that is retained on the column. Results and discussion ICP-MS is well known for its excellent sensitivity. The addition of interference reduction technology, such as CRI, on the aurora M90 ICP-MS allows Cr to be measured with even better sensitivity and signal to background ratios. Using CRI it is possible to use the main chromium isotope at m/z 52 with the removal of interferences including ArC and ClOH. Figures 1 and 2 show that it is possible to quantitate to ng/L (ppt) levels for both Cr(III) and Cr(VI). The calibration Figure 1: Calibration curve for Cr(III), using standards ranging from 0.05 to 100 ppb Table 1: aurora M90 ICP-MS operating conditions ICP Conditions CRI Settings Sample Introduction Parameters Settings Plasma flow (L/min) 18 Auxiliary flow (L/min) 1.65 Nebulizer flow (L/min) 0.24 Sheath flow (L/min) 0.98 Plasma RF power (kW) 1.40 Monitored ion Cr52 Dwell time (s) 0.5 Skimmer gas 60mL/min H2 Sampler gas OFF Pump rate (RPM) 25 Spraychamber temp (ºC) 3 Table 2: LC operating conditions Parameters Settings Mobile Phase 60 mM ammonium nitrate, pH 9.3 Flow rate 1 mL/min Run time 10 min Column Anion exchange, Hamilton PRPX100, 4.1 mm ID x 250 mm, 10 µm PEEK Guard column 1 Column temperature Ambient -30 Sample injection volume 200 µL -3.9 Detection aurora M90 ICP-MS Ion Optics (Volts) 1st Extraction lens -500 2nd Extraction lens -879 3rd Extraction lens -643 Corner lens -763 Mirror left 40 Mirror right 37 Mirror bottom 40 Entrance lens Entrance plate Fringe bias Figure 2: Calibration curve for Cr(VI), using standards ranging from 0.05 to 100ppb Analysis of samples To apply the chromium speciation method to real world sample types, six different brands of both Australian and International mineral waters were analysed. These were analysed both unspiked and spiked at levels of 0.1 ppb for each of the chromium species. Each of the mineral waters were designated by a letter, A through to F. As shown in Table 3, each of the mineral waters gave very good percentage recoveries. Figure 4 shows mineral water E, spiked with 0.1 µg/L Cr(III). From this chromatogram it is also possible to observe the stability of the Cr(VI) signal, which was not spiked. Conclusion This work has demonstrated that the aurora M90 ICP-MS with CRI technology is an excellent detector when coupled with liquid chromatography for chromium speciation. Low ng/L detection limits can be routinely obtained and very good recoveries were obtained on a variety of spiked mineral waters. Figure 3: Chromatogram showing mineral water (blue) and mineral water spiked with 1 ppb Cr(III) and Cr(VI) (red). Samples were stabilized with EDTA prior to analysis Table 3: Spike recovery data for 0.1 µg/L spikes of Cr(III) and Cr(VI) in commercial mineral waters Brand Cr(III) % recovery Cr(VI) % recovery Mineral water A 105 99 Mineral water B 98 102 Mineral water C 106 108 Mineral water D 103 95 Mineral water E 102 106 Mineral water F 104 104 Figure 4: Chromatogram showing Australian mineral water E before and after a 0.1 µg/L Cr(III) spike. References [1] United States Department of Labor, Occupational Health and Safety Administration [2] Environment Canada – Canadian Environmental Protection Act 1999 [3] UN World Health Organization – Guidelines for Drinking Water Quality [4] Australian National Health and Medical Research Council [5] H. Gurleyuk and D. Wallschager, JAAS 2001,16, 926 [6] F. Byrdy, L. Olson, N. Vela and J. Caruso, J. Chrom A 1995, 712, 311 [7]I. Kalinitchenko, Mass spectrometry apparatus and method, US Patent 7,329,863,132 B2, 12 February 2008 [8] M. Leist, The Speciation of Se(IV) and Se(VI) using the aurora M90 ICP-MS, Bruker application note [9] Y. Martinez-Bravo, A.F. Roig-Navarro, F.J. Lopez, F. Hernandez, J. Chrom A 2001, 926, 265 Keywords Instrumentation & Software Drinking water analysis aurora M90 ICP-MS LC-ICP-MS Galaxie software Speciation of chromium Authors Michael Leist, Ray Leiser and Andrew Toms Bruker Daltonik GmbH Bruker Daltonics Inc. Bremen · Germany Phone +49 (0)421-2205-0 Fax +49 (0)421-2205-103 [email protected] Billerica, MA · USA Phone +1 (978) 663-3660 Fax +1 (978) 667-5993 [email protected] www.bruker.com/chemicalanalysis to change specifications without notice. © Bruker Daltonics 02-2011, #CA-274195 Bruker Daltonics is continually improving its products and reserves the right For research use only. Not for use in diagnostic procedures.
© Copyright 2026 Paperzz