cue card Automated multi-step protein purification System overview ................... 2 UNICORN overview ........... 4 ÄKTAxpress™ work flow: 1 Selecting protocols ........ 6 2 System preparations .... 10 3 Column preparations ... 12 4 Run preparations .......... 16 5 Starting a run ................. 18 6 Monitoring a run ........... 20 7 Evaluating the results ... 22 Affinity tag removal ............. 24 System cleaning .................... 25 Column cleaning .................. 26 Advanced zone ..................... 28 Flow charts ............................ 30 cue 18-1178-03 AA 2004-03 • p1 cue card cue card Automated multi-step protein purification Automated multi-step protein purification System overview System overview, continued Control panel Column block Pause/Continue 1 – Affinity column 1 Tubing overview RUN 2 – Affinity column 2 3 – Affinity column 3 or if using protocol E – Desalting column 1 4 – Affinity column 4 or – Ion exchange column 5 – Gel filtration or – Desalting column or in a prepare run – Affinity column 5 4 2 3 PAUSE On/Standby /Reset Deep well 5 System no., status and error codes Next breakpoint Eject/Load microplate Sled for microplate Deep well microplate, 96 square shaped wells Manual injection For details, see page 12. Waste Column valve Column connection 3 Loop valve Superloop connection 2 Outlets F3–F11 1 BYPASS Column 4 BYPASS connection 5 1 example OUT 2 3 4 2 IN 4 5 5 1 4 Left holder • Gel filtration column • HiPrep Desalting column 3 1 3 2 Superloop IN 4 5 OUT Right holder • Superloop™ • Pump rinsing solution F3 F4 F5 F6 Flow through sample 1 sample 2 sample 3 sample 4 F7 Nonselected peaks, F8 F9 F10 sample 1 sample 2 sample 3 sample 4 Buffers A1, A2 B1, B2 S1 S2 S3 S4 Sample inlets Waste* Sample 1 Sample 2 Sample 3 Sample 4 Buffers A3–A8 Grouping buffer containers If the same buffer is used for several systems, one buffer container can feed several systems. * Waste From injection valve port 2 (manual sample injection) * Waste From injection valve port 3 (e.g. system wash) For more information about buffer inlets, see page 8. F11 Metal ion waste Recommended tubing positions Waste Detailed flow charts Frac F3 A H Outlet Valve F4 F11 F5 F10 See page 30–34. F6 F9 F7 F8 Fraction Collector OutV 0.2 MPa Loop L1 Restr L3 L2 L4 L5 C3in Cond C2in C4in C5in C3 C4 C2 L1 Bypass BP C1in C5 Cond C1 BP L1 L5 L2 UV L4 BP Bypass BP L4 L5 L3 L3 L2 L1 UV Loop Valve C5out C1out C2out C1 C4out C3out C2 Manual inject C4 C3 LoopV2 C5 Waste ColV Waste Column Valve Injection Valve Safety MixInj P A8 Inlet Valve S4 A7 Read the ÄKTAxpress Safety Handbook before using the system. Press A6 Mix1 S1 A3 NO NC A1 B2 IN SW1 SW2 SW1 NC Samples NO B1 A2 SW2 AirS Pump cue 18-1178-03 AA 2004-03 • p3 S2 A4 IN cue 18-1178-03 AA 2004-03 • p2 S3 A5 Mix2 Mixer Buffers Pump Air sensor cue card cue card Automated multi-step protein purification Automated multi-step protein purification UNICORN™ for ÄKTAxpress, overview Run preparations – Method Editor UNICORN for ÄKTAxpress, overview, continued See 4 Run preparations, page 16 Method Editor wizard System Control wizard The Method Editor wizard is used to create Method Plans. The System Control wizard is used to start runs based on Method Plans. Example input information: Systems to use, number of samples per system and sample information. Prepare Purify Maintain Method plans Starting a run – System Control See 5 Starting a run, page 18 Separation systems Summary check list Default parameters View or modify default parameters Enter Advanced Zone Save Method Plan Prints Summary check list. Sample information The run starts with Pause. Evaluating the results – Evaluation Run procedures See 7 Evaluating the results, page 22 Result files RUN Advanced zone PAUSE Click Continue in UNICORN, or Press Continue button on system. RUN Pooling protocols Save Method Plan Reports END cue 18-1178-03 AA 2004-03 • p4 cue 18-1178-03 AA 2004-03 • p5 cue card cue card Automated multi-step protein purification Automated multi-step protein purification 1 Selecting protocols, columns and buffers Selecting protocols, columns and buffers, continued Protocols supported Columns supported Multi step protocols Affinity Chromatography AC Desalting Ion Exchange Chromatography Single step protocols AC AC AC AC DS DS IEX DS IEX DS IEX DS GF DS AC column selection Volume [ml] 1 or 5 1 or 5 1 or 5 1 or 5 10 53 1 2 1 2 1 or 6 1 or 6 1 1 120 120 1 or 5 ml column • Depends on required yield. • If many purification steps are used, start with a 5 ml column. • For large sample volumes, use a 5 ml column to save time. • To increase protein purity, overload the column. DS column selection HiTrap or HiPrep • Depends on loading volumes: – Up to 3 ml: Use 2 × HiTrap Desalting. – Up to 10 ml: Use HiPrep Desalting. Gel Filtration or Desalting GF Protocol name A B C D E F G Protocol steps 2 2 3 4 4 1 1 Optional tag removal • • • • • Max no. of samples 4 4 3 3 2 4 4 • • • • • • • • • • • HiTrap Q HP and SP HP: • • RESOURCE Q and S: 15 µm • Mono Q and S 5/50 GL: 10 µm 0.2 0.2 ++ + ++ ++ +++ Run time2), per sample [h] – at room temperature – in cold room 2.7 3.3 1.2 1.4 2.2 2.6 3.1 3.6 4.3 5.1 1) 2) For ordering numbers see page 40. For guidance on column selection, see also the User Manual. IEX column selection Required resolution The smaller bead the higher resolution: Sample characteristics1) – Buffer exchange – Charge homogeneity – Size homogeneity – Purity increase Run time2), total at max. number of samples [h] – at room temperature – in cold room GF Column HisTrap™ HP HiTrap™ Chelating GSTrap™ FF GSTrap HP HiTrap Desalting × 2 HiPrep™ 26/10 Desalting HiTrap Q HP HiTrap Q HP × 2 HiTrap SP HP HiTrap SP HP × 2 RESOURCE™ Q RESOURCE S Mono Q™ 5/50 GL Mono S™ 5/50 GL HiLoad™ 16/60 Superdex 75 prep grade HiLoad 16/60 Superdex 200 prep grade • + The longer column the higher resolution: 1.6 2.0 • 2 columns in series give higher resolution. Anion exchange. e.g. RESOURCE Q 10.4 12.8 4.3 5.2 6.3 7.4 9.3 10.7 8.6 10.1 0.6 0.7 • The pH of the buffer should be 1 pH unit above the pI of the protein(s). 6.1 7.6 Cation exchange. e.g. RESOURCE S Achieved after the last purification step. Run times are approximate, sample loading time is not included. Tip • The pH of the buffer should be 1 pH unit below the pI of the protein(s). Tip Use high purity liquids Grouping samples GF column selection For best purification results, use high purity deionized water and chemicals. Filtering of liquids through a 0.45 µm filter and degassing the liquids is recommended. Considerations when running several separation systems in parallel: Separation size To maximize the success rate, run similar proteins on the same system, for example: • Separation range: Some examples of supported columns. • To maximize GF performance, run small proteins <70 kD on one system and large proteins >70 kD on another system. – 3–70 kDa: Use Superdex™ 75 prep grade. – 10–600 kDa: Use Superdex 200 prep grade. • To maximize IEX performance, run proteins with low pI on one system and proteins with high pI on another system. cue 18-1178-03 AA 2004-03 • p6 34 µm cue 18-1178-03 AA 2004-03 • p7 cue card cue card Automated multi-step protein purification Automated multi-step protein purification Selecting protocols, columns and buffers, continued Buffers and solutions DS buffer suggestions If preparing for... Buffer types needed for AC: • binding buffer • elution buffer • cleavage buffer (optional) • extra wash buffer (optional) for DS: • one buffer type per run (two for protocol D) for IEX: • binding buffer suggested buffer binding using HisTrap binding using HiTrap Chelating extra wash2) 50 mM Tris-HCl pH 7.5, 0.5 M NaCl, 20–40 mM imidazole1) 50 mM Tris-HCl pH 7.5, 0.5 M NaCl, 5–40 mM imidazole1) 50 mM Tris-HCl pH 7.5, 0.5 M NaCl, 100 mM imidazole 50 mM Tris-HCl pH 7.5, 0.5 M NaCl, 50 mM imidazole 50 mM Tris-HCl pH 7.5, 0.5 M NaCl, 500 mM imidazole AC buffer suggestions for GST-tagged proteins If performing... suggested buffer include e.g. 10% glycerol in a suitable buffer (e.g. 50 mM Tris-HCl, pH 7.5, 150 mM NaCl) binding to AIEX suggested buffer (depends on the pI of the protein) 50 mM Tris-HCl pH 8.0 binding to CIEX 20 mM MES pH 6.0 elution from AIEX 50 mM Tris-HCl pH 8.0, 1 M NaCl elution from CIEX 20 mM MES pH 6.0, 1 M NaCl GF buffer suggestions If preparing for... suggested buffer further studies a suitable buffer, e.g. 50 mM Tris-HCl pH 7.5, 150 mM NaCl include e.g. 10% glycerol in a suitable buffer (e.g. 50 mM Tris-HCl pH 7.5, 150 mM NaCl) protein storage The imidazole concentration is protein dependent. x% of extra wash buffer is added to the binding buffer. The x% should be specified in the Method Wizard. See User Manual for details. 3) In the affinity tag removal protocols, the affinity columns can be equilibrated with either AC-binding/wash buffer or an alternative cleavage buffer prior to protease injection. 1) 2) binding using 50 mM Tris-HCl pH 7.5, GSTrap HP or FF 150 mM NaCl, 1 mM DTT cleavage using 50 mM Tris-HCl pH 7.5, PreScission™ protease 150 mM NaCl, 1 mM EDTA, on GSTrap HP or FF3) 1 mM DTT step elution protein storage If for example... If performing... step elution 20 mM MES pH 6.0 IEX buffer suggestions AC buffer suggestions for His-tagged proteins cleavage using TEV on HisTrap HP or HiTrap Chelating HP3) 50 mM Tris-HCl pH 8.0 CIEX • elution buffer for GF: • one buffer type per run suggested buffer AIEX 50 mM Tris-HCl, 10 mM reduced gluthathione, pH 8 Inlet buffers and solutions Inlet Purification run and column preparations Stripping, charging (AC) and blank run A1 Affinity binding buffer (A) Affinity binding buffer (A) GST A buffer B1 Affinity wash buffer (B) (additional wash) – 1 M NaCl A2 Ion exchange binding buffer (A) / Desalting buffer – B2 Ion exchange elution buffer (B) – 2 M NaCl A3 Affinity elution buffer (B) Affinity elution buffer (B) 30% isopropanol A4 Gel filtration buffer or Desalting buffer (final step) – 0.2 M NaOH A5 – Deionized water Deionized water (cleaning) A6 0.5 M NaOH (if NaOH wash between samples is chosen) – 0.5 M NaOH A7 Cleavage buffer (tag removal only) Metal ion charging solution (e.g. 0.1 M NiSO4) 1.0 M NaOH A8 – Metal ion stripping solution (e.g. His-Affinity A buffer with 50 mM EDTA) 6 M Guanidine HCl or 1% TritonX100 cue 18-1178-03 AA 2004-03 • p8 CIP – Cleaning in place Ion exchange binding buffer (A) cue 18-1178-03 AA 2004-03 • p9 cue card cue card Automated multi-step protein purification Automated multi-step protein purification 2 System preparations System preparations, continued Automatic filling of the inlet tubing (Prepare) Manual filling of the inlet tubing Other purging procedures Purging the pump with methanol General Before starting a run, all inlet tubing must be filled to remove air bubbles. Purge the pump and system A small pump flow of 0.1 ml/min may be used. Set the flow in System Control : Manual : Pump. Air remaining in the system may be removed by purging the pump and system by manually selecting Pump Wash and System Wash. Both pump heads must be purged with methanol if: – The system has been left unused for a week, or longer, or – The pump has been run dry. System preparations can be made: – As individual Prepare methods, as described here, or – As part of the purification run, see page 16. If the inlet tubing is without air bubbles an automatic Prepare method can be used to fill the buffer and sample inlet tubing with buffer/solution. If there is air in the tubing, perform a Manual filling of the inlet tubing. Filling inlet tubing A1–A8 and S1–S4 • Immerse tubing in the liquid containers. Put the sample inlet tubing A1 into buffer A. Note: When performing a Prepare, Purification or Maintain run, an initial system wash will be included in the method. • Connect a syringe to the purge valve. • Immerse A1 tubing in the buffer to be used. Creating the method plan • Select System Control : Manual : PumpWash or System Control : Manual : SystemWash. • Click the Method Wizard icon in the Method Editor module. • Set the injection valve to position Waste: – Select System Control : Manual : Flow path : InjectionValve : Waste and click Execute. • Turn the purge valve to open it. • On the last page, click Finish (or to enter the Advanced zone, click Next). • Run the pump at 20 ml/min for 2 minutes: – In System Control : Manual : Pump, set a flow rate of 20 ml/min. – Click Execute to start the flow and run the pump for 2 minutes. – Stop the pump by clicking Pause. • In the Save As page, type the name of the method plan and click OK. Performing the run • Carefully move the inlet tubing A1 to the flask containing about 200 ml of methanol. • Run the pump at 50 ml/min for 2 minutes: – Set the flow rate to 50 ml/min and click Execute. – Click Continue to start the flow and run the pump for 2 minutes. – Stop the pump by clicking Pause. • Slowly draw solution into the syringe. When fluid starts to enter the syringe, continue to draw a few milliliters before closing the purge valve. Check that there is no visible air left in the tubing. • Run the pump at 20 ml/min for 1 minute: – Set the flow rate to 20 ml/min and click Execute. – Click Continue to start the flow and run the pump for 1 minute. – Stop the pump by clicking Pause. • Carefully move the inlet tubing A1 back to the flask containing water. • Repeat the procedure for A2–A8 and S1–S4. Set the inlet valve to the appropriate positions in System Control : Manual : Flowpath. • Repeat the procedure for B1 and B2. – To fill B1 set valve positions: In System Control : Manual : Pump : Gradient select Target 100%B and Mode A1/B1. When the inlet is filled, reset Target %B to 0%. – To fill B2 set valve positions: In System Control : Manual : Pump : Gradient select Target 100%B and Mode A2/B2. When the inlet is filled, reset Target %B to 0%. cue 18-1178-03 AA 2004-03 • p10 • Prepare two flasks: – 200 ml of methanol – 300 ml of deionized water • Put both Waste tubings from the injection valve into a waste flask. • Select Prepare : System (for column preparation see page 12). • Run the Prepare method plan as any other run using Instant Run. See page 18. Note: All inlet tubing that will be used in the next purification run should be filled manually before purging the pump. • Carefully immerse the inlet tubing A1 in the flask containing at least 300 ml of deionized water. • Select New or open an existing method plan to edit. • Select Fill buffer inlets and/or Fill Sample Inlets and specify inlets. The purging will maintain the pumping capacity and protect the pump piston seals. Tip To run the pump for a specific length of time, the instructions PauseTimer and EndTimer can be used. The instructions are found in: System Control : Manual : Other. cue 18-1178-03 AA 2004-03 • p11 • Run the pump at 10 ml/min for 10 minutes: – Set the flow rate to 10 ml/min and click Execute. – Click Continue to start the flow and run the pump for 10 minutes. – Stop the pump by clicking End. • Carefully move the inlet tubing A1 to the flask containing the buffer that will be used during the next purification run. Note: Remember to fill inlet tubing A1 with the buffer to be used, as described on page 10. cue card cue card Automated multi-step protein purification Automated multi-step protein purification 3 Column preparations Tip Check for leakage during column equilibration. • Run the Prepare method plan as any other run using Instant Run. See page 18. Note: Check that appropriate buffers/solutions have been prepared and placed at the correct inlets. See tables page 8. 1 Creating the method plan Connecting columns • Click the Method Wizard icon in the Method Editor module. Drop-to-drop • Select New or open an existing method plan to edit. • Select Prepare : Columns and select column preparation options as required. GF AC IEX 1/16" Female/ M6 Male To avoid air bubbles, use the drop-to-drop procedure when connecting columns: HiTrap 5 ml Column preparations can be made: – As individual Prepare/Maintain methods, as described here, or – As part of the purification run, see page 16. Connecting tubing to columns 2 Performing the run HiTrap 1 ml Column preparations Column preparations, continued 1/16" Female/ M6 Female 1/16" Male/ M6 Female and GSTrap, HisTrap 2 Set the column valve to the required position. Union M6 male to 1/16" female 18-1112-57 Union M6 female to 1/16" male 18-1112-58 Union M6 female to 1/16" female (PEEK) 18-1123-94 Union 1/16" male to 1/16" male 18-1120-93 On-line filter 18-1112-44 DS On-line filter 1/16" Male/ M6 Female 1 Start the pump with a low flow rate. Ordering numbers: 1/16" Female/ M6 Male Column block Column positioning overview AC(1) AC(2) AC(3) AC(4) GF B 4 AC(1) AC(2) AC(3) AC(4) DS C 3 AC(1) AC(2) AC(3) IEX DS D 3 AC(1) AC(2) AC(3) IEX DS E 2 AC(1) AC(2) DS IEX GF F 4 – – – – DS G 4 – – – – GF RESOURCE 6 ml 4 RESOURCE 1 ml A Mono Q 5/50 GL Protocol No. of Column type (sample) in column position samples 1 2 3 4 5 HiPrep 26/10 Desalting IEX HiLoad 16/60 Superdex 3 Fix the tubing to the column drop-to-drop. 1/16" Female/ M6 Male 1/16" Male/ 1/16" Male 1/16" Female/ M6 Female Column block Tip • On the last page, click Finish (or to enter the Advanced zone, click Next). Replace the union’s internal capillary with a 1 mm peek tubing (brown) to reduce backpressure, if required. • Finally, in the Save As page, type the name of the method plan and click OK. View film in On-line Cue Card cue 18-1178-03 AA 2004-03 • p12 Using ÄKTAxpress cue 18-1178-03 AA 2004-03 • p13 350 mm tubing1) 1) The standard system tubing should be replaced by the 350 mm capillary tubing available in the accessory kit. cue card cue card Automated multi-step protein purification Automated multi-step protein purification Column preparations, continued For specification of buffers and inlets, see page 8. Column equilibration AC DS IEX Customized equilibration AC GF Prepare/Column equilibration will equilibrate columns with buffer before starting a purification run. All four column types can be selected in the same run. Column Inlet Volume Flow [ml/min] RT CR HisTrap HP, HiTrap Chelating (1 ml / 5 ml) GSTrap HP/FF (1 ml / 5 ml) HiTrap Desalting HiPrep 26/10 Desalting HiTrap Q/SP HP RESOURCE Q/S (1 ml / 6 ml) Mono Q/S HiLoad Superdex 75/200 pg A1 A1 A2/A4 A2/A4 A2 A2 A2 A4 5 CV 5 CV 5 CV 5 CV 5 CV 5 CV 5 CV 2 CV 1/10 1/10 10 10 1 4/6 1 1.5 IEX GF With custom equilibration up to 5 columns of each type may be selected and up to 9 different solutions may be used. 0.8/8 0.8/8 8 8 0.8 3.2/4.8 0.8 1.2 AC Columns – Metal ion charging DS Customized equilibration is a procedure that equilibrates the column with an optional number of solutions. Only one column type can be included. The procedure is used, for example, to remove ethanol from new columns, using deionized water before equilibration with buffer. Customized equilibration is a Maintain method plan: • Select Maintain. • Select Column CIP (Rigorous) / Customized Equilibration • Make appropriate selections on each page. Note: The first column should be positioned in column position 1, the second column in position 2, etc. AC Columns – Affinity blank run AC AC Purpose Purpose To charge new or stripped HiTrap Chelating HP, or stripped HisTrap HP columns with metal ions (e.g. Ni2+). Several columns can be prepared simultaneously. It is recommended to run a blank run: • Prior to a first time usage of an affinity column, or • After metal ion charging. Manual preparations IEX columns – Ion exchange blank run Note: Put the outlet tubing F11 in a liquid container for collection of any metal waste, if required. Purpose IEX To provide ion-exchange columns with exchangable counter ions. A blank run should be performed before first-time usage and after long-term storage. Metal ion charging (1 ml / 5 ml) Step 1 Solution Inlet Volume Deionized A5 water 5 CV Flow [ml/min] RT CR 1/5 0.8/4 Step 2 Solution Inlet Metal ion A7 solution Volume Flow [ml/min] RT CR 1 CV 1/5 0.8/4 Step 3 Solution Inlet Deionized A5 water Volume Flow [ml/min] RT CR 5 CV 1/5 0.8/4 Affinity blank run (1 ml / 5 ml) Step 1 Buffer A A1 5 CV 1/5 0.8/4 Step 2 Buffer B A3 5 CV Step 2 Buffer B B2 5 CV 1/5 0.8/4 Step 3 Buffer A A1 10 CV Step 3 Buffer A A2 10 CV 1/5 0.8/4 1 4/6 1 0.8 3.2/4 0.8 Ion exchange blank run Step 1 Buffer A A2 5 CV – HiTrap Q/SP HP 1 – RESOURCE Q/S (1/6 ml) 4/6 – Mono Q/S 1 0.8 3.2/4.8 0.8 cue 18-1178-03 AA 2004-03 • p14 1 4/6 1 0.8 3.2/4 0.8 cue 18-1178-03 AA 2004-03 • p15 cue card cue card Automated multi-step protein purification Automated multi-step protein purification 4 Run preparations Run preparations, continued Creating a method plan for the protein purification Use the Method Editor wizard to create a new method plan or to change an existing method plan. • Select Purify. Advanced zone General Advanced Settings • Select procedures to be included in the run: – Include System Preparation see page 10 – Include Column Preparation see page 12 General Some options can be set in the Advanced zone page General Advanced Settings. • Select Purification Protocol and Include Tag Removal (see page 24) if required. Creating the method plan • Make appropriate selections on each page. • Click the Method Wizard icon in the Method Editor module. • On the last page, click Finish (or to enter the Advanced zone, click Next). • Select New or open an existing method plan to edit. • Finally, in the Save As page, type the name of the method plan and click OK. Note: Do not change default parameter values in a method plan unless the result is clearly understood understood. Changing one parameter value might require adjustments of other parameter values as well. Main Prepare Purify Maintain Purification protocol A: B: C: D: E: F: G: Include System Preparation Column Equilibration In the Advanced zone, all main parameter Advanced zone values can be viewed, and modified if required. Examples of parameters that may be modified: • Wash volumes • Flow rates • Gradient slopes • Watch parameters for peak detection AC–GF AC–DS AC–DS–IEX AC–DS–IEX–DS AC–DS–IEX–GF DS GF Getting help To get help about parameter setting, click Help in the wizard boxes. Setup Result file folders Select Store Results in a Main Folder and enter the name of a new or existing folder in Main Folder Name. Specify path to import file If an import file is going to be used for specification of sample information, the file path can be set in Import File Location. Tip If the import files are saved in the default folder MethodWizardImport, there is no need to specify the path. Disabling air sensor aided loading Air sensor aided loading of samples is enabled as default. To disable air sensor loading uncheck Enable AirSensor Controlled Sample Loading. Tag Cleavage Running Condition Prepare Column Last Page Fill Affinity ... ... Desalting Ion Exchange More information about Advanced zone Room Temp. Cold Room Prepare System See page 28. Additional parameters ... ... Tip Column selection Use the Note field in the Save As page to specify changes made to the default settings. Assisted... ... ... Gel filtration Method plan Save As Finish OK Method plan Save As Next > Advanced zone Finish OK View film in On-line Cue Card cue 18-1178-03 AA 2004-03 • p16 cue 18-1178-03 AA 2004-03 • p17 Creating a Method Plan cue card cue card Automated multi-step protein purification Automated multi-step protein purification 5 Starting a run Tip Starting a run, continued Select run parameters Final preparations 3 Select method plan 8 Check the flow path • In UNICORN System Control, click Instant Run. Buffer and sample tubing inspection • If Fill sample inlets is included in the method plan, the inlets should be placed in affinity A buffer. During the run, the system will pause after initial buffer filling, and a message will appear requesting each sample inlet tubing to be moved to its sample. Buffer volumes required It is possible to print out a Summary check list to get a list of the buffer volumes and inlet positions for the run before preparing the system and columns. To do this: 1 Start with step 3 and continue to 7. 2 Print out the Summary check list and instead of starting the run, press Cancel. • Select the required Method plan from the list. 4 Select systems for the run • Select on which System(s) the method plan will be run and the Number of Samples on each system. 5 Specify samples • Check that the tubing is fully immersed in the containers. Fix the tubing, for example, by: – Fitting the inlet filter weights on the tubing, or – Running the tubing through the rubber cap included in the accessory kit. • If Fill sample inlets is not included, gently move the sample inlet tubing to each sample. • Immerse the waste and outlet tubing in appropriate containers according to the Summary check list. • Secure the tubing with the tubing holder. • Make sure: – there is enough buffer available – the correct inlet is placed in each buffer – the Superloop is attached, if to be used. Start the run Air bubbles inspection • Click Run to start the run on the selected systems. The systems will immediately enter Pause. • Inspect the inlet tubing for air bubbles. All tubing up to the pump heads should be inspected. Make sure that no air enters the tubing. Place the tubing close to the bottom of the liquid container but not too tight against the bottom. 11 Start the run RUN PAUSE 12 Final check S1 S2 S3 S4 • Peform a final check that tubing, columns and solutions are placed according to the Summary check list. Tip Prepare system and columns 1 Prepare the buffer, inlets, outlets and waste • Prepare buffers required for the run. • Immerse or connect all inlet tubing to the appropriate liquid containers as described on page 8 and make sure the system is prepared as described in System preparations page 10. • Immerse unused tubing in e.g. 20% ethanol to avoid air entering the tubing. • If Fill sample inlets is not included in the method plan to use, make sure the sample inlet tubing is filled with affinity buffer A, see System preparations page 10. 2 Connect columns and superloop • Enter identification names for the samples, either via the keyboard or using a bar code reader. • For each sample, enter: – isoelectric point, pI, – extinction coefficient for the protein, Ext Coeff, – molecular weight of the protein, MW. The data is automatically imported from an import file, if it has been prepared and placed in the specified folder. • Enter optional text, e.g. culture batch number. cue 18-1178-03 AA 2004-03 • p18 Buffers A1, A2, B1, B2 Buffers A3–A8 Samples S1–S4 9 Prepare the microplate • Press the Eject button on the system to eject the fraction collector sled, if required. 6 Edit result file location and names • Place an empty deep well microplate on the sled and check that the labelling H and A match the labelling on the system. • If required, edit the folder path and file names of the result files to be created. • Press the Eject button again to load the microplate. 7 Print Summary check list • Print out the Summary check list by clicking Print. • Connect selected columns as described on page 12 and prepare them according to the Column preparations description. • Connect the Superloop if it is to be used, see Affinity tag removal, page 24. Sensitive samples can be attached to the system at this point. 10 Prepare the samples • Prepare the samples and clarify them using centrifugation and/or filtration through a 0.45 µm filter. • Place the sample tubes in the tube or flask holder. View film in On-line Cue Card Initiating Runs 13 Continue the run • Click Continue on one system at a time or Continue all to continue the run on all systems. Alternatively, press the Continue button on each individual system. • If Tag Removal was selected, a wizard will give guidance on how to fill the Superloop manually. See Affinity tag removal page 24. Single step protocols (F and G) • For single step protocols, a wizard will give guidance on how to fill the loops with sample(s) manually. Tip For single step protocols, manually fill the loop to be used with the running buffer, before starting the protocol. View film in On-line Cue Card cue 18-1178-03 AA 2004-03 • p19 RUN Note: If the run starts with a system wash, the system will enter pause again. Therefore, do not press Continue more than once. System Check and Start cue card cue card Automated multi-step protein purification Automated multi-step protein purification 6 Monitoring a run Monitoring a run, continued Monitoring a run is performed in System Control. Run data Right-click the view pane and select Properties to select run data to be displayed. View single or all systems The run data of all systems can be viewed simultaneously by clicking Expand >>. Return to single system view by clicking << Collapse. Example of a 4-step purification run Conductivity UV Empty loops, discarded Re-equilibration of IEX Wash 2 AC DS IEX Re-generation of columns GF Post-run procedures Panes Click Panes to select panes to display. wash salt fractionation Sample loading S1 S2 S3 S4 Zoom Use the mouse dragand-drop box to zoom in and out the curve. Right-click to reset zoom. Run status Each system’s status indicator shows the run status: Status Color Example of events Log book Right-click to select log book properties to be displayed. End .................. White Run or Manual Green Hold ................ Yellow ............... Sample loading Curves Right-click to select which curves to be displayed. Flow through F3–F6 Nonselected peaks F7–F10 Microplate Run finished When a system has entered End status the result can be evaluated, see Evaluating the results page 22. Tip If there was a problem See chapter Troubleshooting in the User Manual. Cleaning after a run After the run, if required: • Empty drip plate • Clean system • Clean columns Pause .............. Red .................. Pump-wash Error ................ Yellow, flashing .. Air detected (e.g. running out of buffer) cue 18-1178-03 AA 2004-03 • p20 View film in On-line Cue Card cue 18-1178-03 AA 2004-03 • p21 Monitoring a Run cue card cue card Automated multi-step protein purification Automated multi-step protein purification 7 Evaluating the results 1 Find and open the result files • Use the Recent Runs or the Find tab, in the Evaluation module to locate the result file. Note: Automatic pooling is not performed if the result file is opened via the Files tab. • Click to expand the list for the result file. A result file consists of: – Sample loading chromatogram (11 below) – Purification chromatograms for each sample • Double-click a sample chromatogram to open it. • If necessary, adjust the theoretical suggested pooling. If the extinction coefficient has been entered in the wizard, concentration and amount are automatically calculated. Otherwise: enter the extinction coefficient manually by marking a pool and then typing the value in the extinction coefficient field. The concentration is given in mg/ml, or M, and amount is given in mg, or mmole, depending on the coefficient used. The linearity of the UV monitor is limited. For more information, see Technical specification in the User Manual. Evaluating the results, continued Getting help 4 Print the Report On-line Manuals and Cue Cards • Select File : Report or click the Report icon. • Select Xpress1sample, Xpress2samples, Xpress3samples or Xpress4samples report format, depending on how many samples are included in the result file. • Open the Help menu and select Manuals to access the on-line instructions: – UNICORN User Manual – ÄKTAxpress User Manual – Cue Cards with films. • The report format XpressActive Chrom can also be selected if only the active zoomed window should be included in the report. • Click the Add to Pooling Protocol button to add the adjusted pools to the Pooling protocol. • Repeat the procedure for other chromatograms from the same, or other, result files. • Click the View Pooling Protocol button. 3 Print or save the Pooling protocol The Pooling protocol can be used as a help when making the physical pooling of the purified samples from the microplate. 2 Adjust pooling and add to Pooling protocol The chromatogram is displayed and UNICORN will automatically display a suggested theoretical pooling of the fractions. The pooled fractions are listed in a table below the chromatogram and the pooled peaks are numbered sequentially in the chromatogram. • To print the Pooling protocol: Click the Print button to print the protocol on the default printer. • Click Edit or Preview to inspect and modify the report format if needed. • Click the Print button. • Choose which pages and how many copies to print and click OK. • To save the Pooling protocol as a file: Click Export and save the protocol in one of the following formats: text (.txt), Excel (.xls), HTML (.htm). Context-specific Help • Click the Help button in the dialog box, or press F1. Example of Method Editor wizard help: Printed instructions Printed information for ÄKTAxpress: Only adjacent fractions will be pooled. The fraction numbers for each pool are listed in the table as a range in retention order, e.g. A6–A7 etc. Book PDF On-line User Manual • • • Safety Handbook • • – If the pooling suggestion is not performed, choose Operations : Pool. Installation Guide • • – UNICORN manuals • • • View film in On-line Cue Card Evaluating the Results cue 18-1178-03 AA 2004-03 • p22 cue 18-1178-03 AA 2004-03 • p23 For ordering numbers see page 40. cue card cue card Automated multi-step protein purification Automated multi-step protein purification System cleaning Affinity tag removal General Filling the Superloop Affinity tag removal can be performed in all multi-step protocols. The Superloop can be filled: – With assistance from a Purify method plan. – With assistance from a Prepare method plan, if Assisted Manual Loading of Superloop has been selected. – Manually. There is a higher risk of introducing air into the system with this procedure. Creating the method plan • In the Purify or Assisted Loading method plan, select Tag Cleavage to add an affinity tag removal step to the method plan. Assisted filling • Choose Incubation time and select Fill Columns with Cleavage buffer if the ordinary binding buffer should not be used. During Assisted Manual Loading of Superloop, dialog boxes will give guidance to filling the Superloop a the beginning of the run. The filling is performed via a syringe connected to the injection valve. • Make appropriate selections on each page. • On the last page, click Finish. • In the Save As page, type the name of the method plan and click OK. • During the run, follow the dialog boxes and press Continue between each step. 1 Fill syringe with buffer and insert syringe in valve. 2 Inject buffer. 3 Fill syringe with protease solution and insert syringe in valve. 4 Inject protease solution. Calculate protease amount • Calculate the volume and concentration of protease needed. Use, for example, the Excel file available on the Installation CD. Clean system Leaving the system for some days Purpose If the system is not going to be used for some days: Use a bacteriostatic solution, for example deionized water and 20% ethanol, as the two final wash solutions. To wash all used tubing, including the loops, within the system: up to 5 cleaning solutions can be used, one at a time. Creating the method plan Superloop off-line cleaning • Click the Method Wizard icon in the Method Editor module. See the Superloop instructions for guidelines on how to clean the Superloop off-line. • Select New or open an existing method plan to edit. Note: The Superloop can not be cleaned automatically. When starting a Loop Wash on position 5 make sure to disconnect the Superloop and replace it with a capillary loop. • Select Maintain. • Select Clean System. • Select the parts to be cleaned. • On the last page, click Finish (or to enter the Advanced zone, click Next). • In the Save As page, type the name of the method plan and click OK. Manual preparation • Prepare a sufficient volume of the wash solutions (max. 5). • Wash the outside of the inlet tubing with deionized water and/or ethanol. • Immerse the tubing in a container of cleaning solution. • If the column valve is to be cleaned: Remove the columns and reconnect the tubing to use as bypass column connections. • Check that loop 5 is a capillary loop if the loop valve is to be cleaned. Connect the Superloop Performing a Clean System run • Fit the Superloop in the right hand holder. • Run the Clean System run as any other run using Instant Run. See page 18. Top Note: Make sure the Superloop is not hindering the ejection of the microplate. • A message will prompt when to manually change cleaning solutions. Inlet A1 is used for cleaning outlets, loops and column valve. • Connect the Superloop to the Loop Valve: – Top to OUT 5 – Bottom to IN 5 2 3 Superloop 5 1 5 4 3 2 The protease can be injected directly into the Superloop before starting the run. IN 4 1 See page 32 for the flow chart of system cleaning. Manual filling off-line • Connect a syringe to the lower port of the Superloop and fill it. Avoid introducing air bubbles. 0 ml OUT Manual filling in-line See the User Manual. Bottom cue 18-1178-03 AA 2004-03 • p24 cue 18-1178-03 AA 2004-03 • p25 cue card cue card Automated multi-step protein purification Automated multi-step protein purification Column cleaning Column cleaning, continued Note: Make sure all inlet tubing is filled before start. See System preparations, page 10. AC Column CIP (Cleaning-In-Place) Metal Ion Stripping (chelating columns) Purpose Purpose To clean contaminated columns. For column cleaning procedures and column storage instructions, please refer to the instructions supplied with each column or to the Amersham Biosciences web page: www.chromatography.amershambiosciences.com To remove metal ions before regenerating the HiTrap chelating and HisTrap affinity columns. To recharge the column, follow the instructions in Column preparations page 12. Creating the method plan • Click the Method Wizard icon in the Method Editor module. • Select New or open an existing method plan to edit. Note: Always remove metal ions before or directly after storing the columns in ethanol. Creating the method plan • Select Maintain. • Select New or open an existing method plan to edit. • Select Maintain. • Select Strip Affinity Columns. • Select the type of column, number of columns, and running condition. • On the last page, click Finish (or to enter the Advanced zone, click Next). Rigorous can have up to 9 steps and the default cleaning parameters can be changed. • In the Save As page, type the name of the method plan and click OK. See the tables for specifications of the CIP procedures. Performing a Metal Ion Stripping run • On each page, select the parameter values. • On the last page, click Finish (or to enter the Advanced zone, click Next). • Run the stripping run as any other run using Instant Run. See page 18. • In the Save As page, type the name of the method plan and click OK. Performing a CIP run • Run the CIP run as any other run using Instant Run. See page 18. Step 1 Column1) Solution HisTrap HP, HiTrap Chelating (1 ml / 5 ml) 1M NaOH GSTrap HP/FF (1 ml / 5 ml) 6M GuaHCl or 1% Triton X-100 HiTrap Desalting 0.2M NaOH HiPrep 26/10 Desalting 0.2M NaOH HiTrap Q/SP HP 2M NaCl RESOURCE Q/S (1 ml / 6 ml) 2M NaCl Mono Q/S 2M NaCl HiLoad Superdex 75/200 pg 0.5M NaOH 1) • Click the Method Wizard icon in the Method Editor module. • Select cleaning type: – Column CIP (Regular) – Column CIP (Rigorous) Regular is a 2-step cleaning where the cleaning procedure is automatically adjusted for the specified column to clean. The solutions and inlets are specified on the Summary check list when starting the run. Inlet Flow [ml/min] Volume RT CR Step 2 Solution A7 A8 1 CV2) 2 CV 1/5 1/5 0.8/4 0.8/4 A4 A4 B2 B2 B2 A6 2 CV 2 CV 2 CV 3 CV 2 CV 1 CV 10 10 0.35 4/6 2 0.5 8 8 0.35 3.2/4.8 1.6 0.3 The columns must be filled with deionized water prior to CIP. 2) IEX GF Strip of HisTrap and HiTrap Chelating (1 ml / 5 ml) Procedure Strip Step 1 Solution Inlet His affinity A8 A-Buffer with 50mM EDTA Step 2 Deionized water A5 Volume 5 CV Flow [ml/min] RT CR 1/5 0.8/4 5–10 CV 1/5 Volume Flow [ml/min] RT CR Deionized water A5 GST A-buffer A1 10 CV 5 CV 1/5 1/5 0.8/4 0.8/4 Deionized water Deionized water IEX A-buffer IEX A-buffer IEX A-buffer Deionized water A5 A5 A2 A2 A2 A5 5 CV 5 CV 5 CV 5 CV 10 CV 2 CV 10 10 1 4/6 2 0.5 8 8 0.8 3.2/4.8 1.6 0.3 Step 2 Solution Inlet Flow [ml/min] Volume RT CR Inlet Pauses 1 hour. CIP Rigorous, recommended solutions and inlets Column HisTrap HP, HiTrap Chelating (1 ml / 5 ml) GSTrap HP/FF (1 ml / 5 ml) HiTrap Desalting Step 11) Solution 1M NaOH 6M GuaHCl 1mg pepsin/ml in 0.1M acetic acid, 0.5M NaCl HiPrep 26/10 Desalting 1mg pepsin/ml in 0.1M acetic acid, 0.5M NaCl HiTrap Q/SP HP 2M NaCl RESOURCE Q/S (1 ml / 6 ml) 1M NaCl Mono Q/S 1M NaCl HiLoad Superdex 75/200 pg 1M NaOH Inlet Flow [ml/min] Volume RT CR A7 1 CV2) A8 2 CV A7 or 1 CV3) A8 1/5 1/5 5 0.8/4 0.8/4 4 Deionized water A5 GST A-buffer A1 0.2M NaOH A4 10 CV 5 CV 2 CV 1/5 1/5 5 0.8/4 0.8/4 4 A7 or 1 CV3) A8 10 8 0.2M NaOH A4 2 CV 10 8 B2 B1 B1 A7 0.25 4/6 0.5 0.5 0.20 3.2/4.8 0.4 0.4 Deionized water Deionized water Deionized water Deionized water A5 A5 A5 A5 2 CV 2 CV 2 CV 4 CV 0.25 4/6 0.5 0.5 0.20 3.2/4.8 0.4 0.4 Deionized water A5 GST A-buffer A1 10 CV 5 CV 1/5 1/5 0.8/4 0.8/4 Deionized water Deionized water Deionized water Deionized water A5 A5 A5 A5 4 CV 2 CV 2 CV 2 CV 0.25 4/6 0.5 0.5 0.20 3.2/4.8 0.4 0.4 A5 A5 A5 A5 10 CV 4 CV 2 CV 2 CV 1/5 0.25 4/6 0.5 0.8/4 0.20 3.2/4.8 0.4 2 CV 2 CV 4/6 0.5 3.2/4.8 0.4 4 CV 5 CV 4 CV 4 CV Step 3 Strip HisTrap HP, HiTrap Chelating (1 ml / 5 ml) GSTrap HP/FF (1 ml / 5 ml) HiTrap Desalting HiPrep 26/10 Desalting HiTrap Q/SP HP RESOURCE Q/S (1 ml / 6 ml) Mono Q/S HiLoad Superdex 75/200 pg 0.8/4 30% 2-Propanol 1% Triton X-1004) Deionized water Deionized water 1M NaOH 1M NaOH 1M NaOH 30% 2-Propanol Step 4 A3 A7 A5 A5 A7 A7 A7 A3 10 CV 4 CV 5 CV 5 CV 4 CV 5 CV 4 CV 0.5 CV 0.5/2.5 1/5 7.5 15 0.25 4/6 0.5 0.5 0.4/2 0.8/4 6 12 0.20 3.2/4.8 0.4 0.4 Step 5 HisTrap HP, HiTrap Chelating (1 ml / 5 ml) HiTrap Q/SP HP RESOURCE Q/S (1 ml / 6 ml) Mono Q/S 2M NaCl 30% 2-Propanol 1M HCl 1M HCl Step 7 RESOURCE Q/S (1 ml / 6 ml) 1M NaCl Mono Q/S 1M NaCl Step 6 B2 A3 A1 A1 1 CV 2 CV 5 CV 4 CV 1/5 0.25 4/6 0.5 0.8/4 0.20 3.2/4.8 0.4 Deionized water Deionized water Deionized water Deionized water B1 B1 5 CV 4 CV 4/6 0.5 3.2/4.8 0.4 Step 8 Deionized water A5 Deionized water A5 Step 1 implies that the column is already filled with deionized water. If this is not the case, an additional step using water has to be added. 1) 2) 3) 4) cue 18-1178-03 AA 2004-03 • p26 DS CIP Regular, recommended solutions cue 18-1178-03 AA 2004-03 • p27 Pause 1 hour. Pause over night at RT. Alternatively 3–4 CV 70% ethanol. cue card cue card Automated multi-step protein purification Automated multi-step protein purification Advanced zone Advanced zone General In the Advanced zone, all main parameter Advanced zone values can be viewed, and modified if required. Examples of parameters that may be modified: • Wash volumes • Flow rates • Gradient slopes • Watch parameters for peak detection Note: Do not change default parameter values in a method plan unless the result is clearly understood. Changing one parameter value understood might require adjustments of other parameter values as well. Tip Use the Note field in the Save As page to specify changes made to the default settings. Advanced zone, continued Some considerations in Advanced Zone Flow rates and pressure limits No warning will be issued if the flow rate or pressure limit are set higher than the values recommended for the columns used. Note: Changing the flow rate will change the slope of an eluting peak. Peak detection and collection The default values for detecting peak start and peak end are set to match typical peaks eluting from the supported columns. Different values are used for normal and high sample loading. The slope and levels values though might need to be adjusted if other parameters are changed. Example of peak collection in loops: UV max UV max' The slope of a peak is changed if: 2 • The flow rate is changed, or 2' • The gradient slope is changed. 2:nd wash of AC chelating column A second wash of affinity chelating columns can be used to wash out loosely bound proteins. The second wash is an additional step before the elution of the target protein. For a powerful wash, the imizadole concentration %B should be set to above 0%B. If no imidazole is used (%B=0), the volume used in this step should be decreased to minimize the run time. Note: Do not raise the imidazole concentration too much as it might cause the target protein to come off as well. Note: The second wash is by default not included for GSTrap columns. Peak volume in AC and IEX elution steps Peak volumes for peaks eluted from affinity and ion exchange columns should be as small as possible. It will increase the possibility to: • Load the entire peak volume onto any desalting or gel filtration column in the next step since their maximum loading volumes are limited. • Optimize the recovery in the ion exchange step by making sure that the entire peak volume is collected when eluted in the ion exchange gradient. Only one loop is available for each peak, so peak volumes larger than 7.5 ml will cause sample loss. To minimize the peak volume: • Choose a suitable column type. • Choose a suitable column size. • Adjust the gradient slope. A steeper gradient will narrow the peaks. Protease injection volume This applies to protocols that include tag cleavage. If necessary, change Watch Slope Greater than. 1' 3' 1 3 The default level values might need adjustment if: • Small peaks are expected, or • The peaks are broadened due to e.g. a more flat gradient. If necessary, change Watch Level Greater than and Watch Level Less than. 1 Greater_Than AND Slope_Greater_Than 2 Peak_Max with Factor 0.6 3 Less_Than_or_Valley OR Stable_Plateau Note: If the level values are decreased for very large or broad peaks, the peak volume collected might be too big for the loop available or exceed the loading volume of the next column. Watch instructions for peak collection Greater_Than The signal exceeds a certain value. Peak fractionation parameters The parameters are based on slope or level. Example of peak fractionation in microplate: Less_Than The signal falls below a certain value. UV max Slope_Greater_Than The slope of the the signal curve exceeds a specified value. Slope_Less_Than The slope of the the signal curve falls below a specified value. It is always used in combination with Peak_Max. 3 Less_Than_Or_Valley The signal falls below a specified value or a valley is detected. A valley is detected only after a Peak_Max has been detected, and the valley is defined by a local minimum followed by an increase to 102% of the local minimum plus the Delta_Peak value. Peak_Max The signal falls to a specified fraction of the most recent peak maximum minus the Delta_Peak value. Stable_Baseline The signal is stable within the limits of the Delta_Base value for a specified time interval. Stable_Plateau The signal is stable within the limits of the Delta_Plat value for a specified time interval. The default value for the protease injection is 0.7 CV. If using a larger value, the target protein might be lost since it will pass through the column before peak collection is activated. cue 18-1178-03 AA 2004-03 • p28 Intermediate peak collection parameters cue 18-1178-03 AA 2004-03 • p29 1 2 4 1 5 2 4 Start Slope OR Start Level UV max × Peak Max Factor 0.5 3 5 Stop Level OR Stop Slope cue card cue card Automated multi-step protein purification Automated multi-step protein purification Flow charts Flow charts, continued • Valve positions after reset (END), no flow • Column equilibration • Sample loading Waste Frac • Wash out unbound sample • Elution or Waste F3 A H Outlet Valve F3 F4 F11 Frac F4 F11 F10 A H Outlet Valve F5 F5 F6 F9 F10 F7 F8 F6 F9 Fraction Collector F7 F8 Fraction Collector OutV OutV 0.2 MPa Loop L1 Restr L3 L2 L4 0.2 MPa Loop L5 C3in C5in C3 C4 C2 Bypass L4 BP C5in L2 UV Bypass L4 L5 C5 C3 C4 C1 C2 C4 C3 L5 L2 L4 L3 L2 L1 Loop Valve C5out C2out C4out C3out Manual inject Waste C1 Waste Column Valve L4 L5 L3 UV C1out ColV BP L1 BP Bypass BP LoopV2 C5 Bypass L1 C4out Manual inject C1 L5 L2 Loop Valve C3out C2 Cond UV L3 UV C2out L4 L1 BP C1in L3 C5out C1out Cond C2in L1 L5 BP BP C4in Cond C1 L3 L2 C3in L1 BP C1in C5 L1 Restr Cond C2in C4in C2 C4 C3 LoopV2 C5 Waste ColV Waste Column Valve Injection Valve Injection Valve MixInj MixInj P A8 Inlet Valve P S4 A7 Press A6 A5 Mix2 Mixer S2 A4 Mix1 A8 S3 S1 IN IN NC NO A2 NC A1 Samples NO B1 IN B2 SW2 IN SW1 SW2 SW1 NC Samples NO B1 A2 SW2 AirS Pump Buffers Pump Air sensor AirS Pump Buffers Pump cue 18-1178-03 AA 2004-03 • p30 S1 A3 SW1 SW2 SW1 S2 A4 Mix1 B2 S3 A5 Mix2 Mixer NC A1 S4 A6 A3 NO Inlet Valve A7 Press cue 18-1178-03 AA 2004-03 • p31 Air sensor cue card cue card Automated multi-step protein purification Automated multi-step protein purification Flow charts, continued Flow charts, continued • Intermediate fractionation • System wash • Pump wash Waste Waste Frac F3 F3 A H Outlet Valve Frac F4 F4 F11 F11 F5 F5 F10 F10 F6 F6 F9 F9 F7 F7 F8 F8 Fraction Collector Fraction Collector OutV OutV Loop 0.2 MPa 0.2 MPa Loop L1 Restr L3 L2 L4 L5 Cond C2in C4in C3 C4 C2 L1 Cond C1 L4 BP Bypass C5in C5 C3 C4 C2 L4 L5 L3 L2 L1 C1 Waste ColV L2 L1 Loop Valve C4out C3out C2 Manual inject C4 C3 LoopV2 C5 Waste ColV Waste Column Valve L3 C5out C1out LoopV2 C5 L4 L5 L3 UV C2out Manual inject C4 C3 L2 L4 BP Bypass BP BP L1 L5 Loop Valve C3out Bypass UV C4out C2 L5 L1 L3 C5out C1out C1 L4 Cond C1 UV C2out L3 L2 BP C1in L2 UV BP BP L1 L5 Cond C2in C4in Bypass BP C1in C5 L1 Restr C3in C3in C5in A H Outlet Valve Waste Column Valve Injection Valve Injection Valve MixInj MixInj P P A8 A8 Inlet Valve A6 A5 Mix2 Mixer S3 S2 A4 Mix1 S1 IN Mix1 NC A2 A1 Samples IN Buffers IN SW2 Air sensor Samples NO A2 SW2 AirS Pump Buffers Pump cue 18-1178-03 AA 2004-03 • p33 SW1 SW2 NC Pump cue 18-1178-03 AA 2004-03 • p32 B2 SW1 B1 AirS Pump S1 A3 NO B1 S2 NC SW1 SW2 S3 A5 Mix2 Mixer NO B2 SW1 S4 A6 NC A1 IN Press A4 A3 NO Inlet Valve A7 S4 A7 Press Air sensor cue card cue card Automated multi-step protein purification Automated multi-step protein purification Flow charts, continued • Sample loading from loops • Loop wash Waste Frac F3 A H Outlet Valve F4 F11 F5 F10 F6 F9 F7 F8 Fraction Collector OutV 0.2 MPa Loop L1 Restr L3 L2 L4 L5 C3in C5in Cond C2in C4in C5 C3 C4 C2 L1 Bypass BP C1in Cond C1 L5 L2 UV L4 BP Bypass BP BP L1 L4 L5 L3 L3 L2 L1 UV Loop Valve C5out C1out C2out C1 C4out C3out C2 Manual inject C4 C3 LoopV2 C5 Waste ColV Waste Column Valve 7 Injection Valve MixInj P A8 Inlet Valve S4 A7 Press A6 A5 Mix2 Mixer S3 S2 A4 Mix1 S1 A3 NO NC A1 IN B2 IN SW1 SW2 SW1 NC Samples NO B1 A2 SW2 AirS Pump Buffers Air sensor Pump cue 18-1178-03 AA 2004-03 • p34 cue 18-1178-03 AA 2004-03 • p35 cue card cue card Automated multi-step protein purification Automated multi-step protein purification Reference information Overview of columns and solutions Protocol A • Flow overview, simplified B C D E F G Column position 1 AC(1) AC(1) AC(1) AC(1) AC(1) – – 2 AC(2) AC(2) AC(2) AC(2) AC(2) – – 3 AC(3) AC(3) AC(3) AC(3) DS – – 4 AC(4) AC(4) IEX IEX IEX – – 5 GF DS DS DS GF DS GF outlet valve A1 AC binding AC binding AC binding AC binding AC binding – – B1 AC wash AC wash AC wash AC wash AC wash – – A2 – – IEX binding1) IEX binding IEX binding1) – – B2 – – IEX elution IEX elution IEX elution – – A3 AC elution AC elution AC elution AC elution AC elution – – A4 GF buffer DS buffer – DS buffer GF buffer A5 – – – – – A6 A7 0.5 M NaOH2) 0.5 M NaOH2) 0.5 M NaOH2) 0.5 M NaOH2) 0.5 M NaOH2) flow through, sample 1–4 not used peaks, sample 1–4 Cond. Cell loops 1-5 loop valve UV cell DS buffer GF buffer – – – – Cleavage Cleavage Cleavage Cleavage Cleavage – buffer3) buffer3) buffer3) buffer3) buffer3) – – – – – – – – – S1 Sample 1 Sample 1 Sample 1 Sample 1 Sample 1 S2 Sample 2 Sample 2 Sample 2 Sample 2 Sample 2 Sample 2 Sample 2 S3 Sample 3 Sample 3 Sample 3 Sample 3 – Sample 3 Sample 3 S4 Sample 4 Sample 4 – – – Sample 4 Sample 4 A8 0.2 MPa waste, metal ion waste Buffer inlet 96-well microplate column valve alternative : superloop columns 1-5 Sample inlet 1) 2) 3) Sample 1 Sample 1 mixer injection valve pressure sensor The IEX binding buffer is also used for desalting. If NaOH wash between samples is chosen. If Tag cleavage and Fill columns with cleavage buffer are chosen. waste inlet valve pump gradient valves air sensor cue 18-1178-03 AA 2004-03 • p36 cue 18-1178-03 AA 2004-03 • p37 buffers A1, B1 buffers A2, B2 buffers A3-8 Samples 1-4 cue card Automated multi-step protein purification cue 18-1178-03 AA 2004-03 • p38 cue card Automated multi-step protein purification cue 18-1178-03 AA 2004-03 • p39 cue card Automated multi-step protein purification Ordering information Product Pack Size Code Number Product ÄKTAxpress module, system extension 1 18-6645-01 Ion exchange chromatography RESOURCE Q 1 x 1 ml ÄKTAxpress, set of 4 systems 18-6645-04 RESOURCE Q 1 x 6 ml 17-1179-01 RESOURCE S 1 x 1 ml 17-1178-01 17-1180-01 Systems 1 Pack Size Code Number 17-1177-01 ÄKTAxpress, set of 4 systems with computer 1 18-6645-05 RESOURCE S 1 x 6 ml Flask holder 1 18-1177-79 HiTrap Q HP 5 x 1 ml 17-1153-01 HiTrap SP HP 5 x 1 ml 17-1151-01 Supported columns Mono Q 5/50 GL 1 x 1 ml 17-5166-01 HisTrap HP 5 x 1 ml 17-5247-01 Mono S 5/50 GL 1 x 1 ml 17-5168-01 HisTrap HP 100 x 1 ml 17-5247-05 Gel filtration HiLoad 16/60 Superdex 75 prep grade 1 x 120 ml 17-1068-01 HiLoad 16/60 Superdex 200 prep grade 1 x 120 ml 17-1069-01 Affinity chromatography HisTrap HP 5 x 5 ml 17-5248-02 HisTrap HP 100 x 5 ml 17-5248-05 HiTrap Chelating HP 5 x 1 ml 17-0408-01 HiTrap Chelating HP 5 x 5 ml 17-0409-03 HiTrap Chelating HP 100 x 5 ml 17-0409-05 Superloop GSTrap HP 5 x 1 ml 17-5281-01 Superloop GSTrap HP 100 x 1 ml 17-5281-05 Documents GSTrap HP 5 x 5 ml 17-5282-02 UNICORN manual set 1 11-0003-68 GSTrap HP 100 x 5 ml 17-5282-05 ÄKTAxpress User Manual 1 18-1178-00 18-1178-02 1 18-1113-82 GSTrap FF 5 x 1 ml 17-5130-01 ÄKTAxpress Safety Handbook 1 GSTrap FF 100 x 1 ml 17-5130-05 ÄKTAxpress Installation Guide 1 18-1178-01 GSTrap FF 5 x 5 ml 17-5131-02 HiTrap Column Guide 1 18-1129-81 GSTrap FF 100 x 5 ml 17-5131-05 Affinity Chromatography Handbook 1 18-1022-29 Gel Filtration Handbook 1 18-1022-18 Desalting HiPrep 26/10 Desalting 1 x 53 ml 17-5087-01 Ion Exchange Chromatogr. Handbook 1 18-1114-21 HiPrep 26/10 Desalting 18-1132-29 4 x 53 ml 17-5087-02 Protein Purification Handbook 1 HiTrap Desalting 5 x 5 ml 17-1408-01 Recombinant Protein Handbook 1 18-1142-75 HiTrap Desalting 100 x 5 ml 11-0003-29 GST Gene Fusion System Handbook 1 18-1157-58 100-packs are special packs delivered on customer order. Includes connector package, domed nuts and instructions. Code numbers for accessories and user replaceable spare parts can be found in the ÄKTAxpress User Manual. Asia Pacific Tel: +852 2811 8693 Fax: +852 2811 5251 Australasia Tel: +61 2 9899 0999 Fax: +61 2 9899 7511 Austria Tel: 01 57 606 16 19 Fax: 01 57 606 16 27 Belgium Tel: 0800 73 888 Fax: 03 272 1637 Canada Tel: 1 800 463 5800 Fax: 1 800 567 1008 Central, East, South East Europe Tel: +43 1 982 3826 Fax: +43 1 985 8327 Denmark Tel: 45 16 2400 Fax: 45 16 2424 Finland & Baltics Tel: +358 (0)9 512 3940 Fax: +358 (0)9 512 39439 France Tel: 0169 35 67 00 Fax: 0169 41 9677 Germany Tel: 0761 4903 490 Fax: 0761 4903 405 Italy Tel: 02 27322 1 Fax: 02 27302 212 Japan Tel: 81 3 5331 9336 Fax: 81 3 5331 9370 Latin America Tel: +55 11 3933 7300 Fax: +55 11 3933 7306 Middle East and Africa Tel: +30 2 10 96 00 687 Fax: +30 2 10 96 00 693 Netherlands Tel: 0165 580 410 Fax: 0165 580 401 Norway Tel: 2318 5800 Fax: 2318 6800 Portugal Tel: 21 417 7035 Fax: 21 417 3184 Russian & other C.I.S. & N.I.S. Tel: +7 (095) 232 0250,956 1137 Fax: +7 (095) 230 6377 South East Asia Tel: 60 3 8024 2080 Fax: 60 3 8024 2090 Spain Tel: 93 594 49 50 Fax: 93 594 49 55 Sweden Tel: 018 612 19 00 Fax: 018 612 19 10 Switzerland Tel: 0848 8028 12 Fax: 0848 8028 13 UK Tel: 0800 616 928 Fax: 0800 616 927 USA Tel: +1 800 526 3593 Fax: +1 877 295 8102 UNICORN, ÄKTAxpress, HiTrap, HisTrap, GSTrap, HiLoad, HiPrep, RESOURCE, Superdex, Mono Q, Mono S, Superloop, PreScission and Drop Design are are trademarks of Amersham Biosciences Limited. Amersham and Amersham Biosciences are trademarks of Amersham plc. Amersham Biosciences AB Björkgatan 30, SE-751 84 Uppsala, Sweden. Amersham Biosciences UK Limited Little Chalfont, Buckinghamshire, England HP7 9NA. Amersham Biosciences Corporation 800 Centennial Avenue, PO Box 1327, Piscataway, NJ 08855 USA. Amersham Biosciences Europe GmbH Munzinger Strasse 9, D79111 Freiburg, Germany. Amersham Biosciences KK Sanken Building, 3-25-1, Hyakunincho, Shinjuku-ku, Tokyo 169-0073, Japan. All goods and services are sold subject to the terms and conditions of sale of the company within the Amersham Biosciences group that supplies them. A copy of these terms and conditions is available on request. © Amersham Biosciences AB 2004 – All rights reserved. cue 18-1178-03 AA 2004-03 • p40
© Copyright 2026 Paperzz