E UROPACE Results: Kappa 900 pacemakers have been implanted in 104 patients, with 55 complete measurements from the l-month visit available for comparison. Data were excluded for one or more of the following reasons: lead dislodgmerit (3), missing EGM to verify measurements (X3), C M aborts due to fast rhythms (13), and patient not having a l-month visit prior to the trial cutoff date (18). All leads were placed in the right ventricular apex, and the following characteristics were represented: 50 bipolar including 45 long and 5 short ~14 mm) tip-ring spacing, 5 unipolar; 16 active and 39 passive; 13 low (520mV POL), 39 medium/high, and 3 unknown POL. On average, SD and C M voltage T were 0.482 V and 0.486 V, respectively. The average SD-CM T difference of these tests was -0.005V with a 95% bootstrap coniidence interval (CI) upper bound of 0.030 V The overall (2.sided) 95% CI for the average difference was (-0.091,0.036). No evoked response undersensing was observed during either 24.hour Halter recordings from 11 patients or device memory analysis between visits. Conclusion: The ventricular C M T test showed reliable T results compared to the manual SD T test when using a variety of ventricular leads. In addition, there was a 100% success rate of evoked response detection. I P 120 EFFECTIVENESS OF INTRACARDIAC A TOOL TO OPTIMIZE A-V DELAY IMPEDANCE AS A.P. Ravazzi, P. Diotallevi, M.P. Provera, L. Brandolisio’, C. Militelloz, R. Audoglio3. Div. of Cardiology Osp. S. Antonio E Biagio, Alessandria, Italy; ‘Biohvnik Seda, Trezzano SIN, Italy; 2Biohvnik, Erlangen, Germany; 3Sra, Pavia, Italy Introduction: Programming the optimal atriovenhicular delay (OAVD) in pacemaker patients allows the completion of the end-diastolic filling flow prior to ventricular contraction providing the longest diastolic filling time. Non-invasive, echocardiographic assessment of OAVD is feasible, but time and resource consuming. Aim of this preliminary study was to verify whether OAVD assessment is feasible using informations deducible from right ventricular intracardiac impedance (Z) variations measured by the pacemaker INOS” CLS (Biotronik, Germany). Methods: In 24 pts (16M; 67.5&11.5 yrs; 14 AVB, 10 SSS; NYHA I) implanted with an INOS’+ CLS, the OAVD was assessed by Ritter’s formula based on echo measurements. At different programmed AVDs, Z was measured by the pacemaker and downloaded via telemetry for post-processing. Results: The OAVD assessed by echo highly correlated to that calculated on the impedance when selecting the AV delay showing the highest value of the maximum of the squared 1” derivative (r=0.91, p<O.O05). Mean OAVD measured by echo and by Z were 162.0&25.7 ms and 160.4&26.4 ms respectively. No statistical differences were observed between these values Conclusion: Right ventricular intracardiac impedance seems to be a valid tool for optimal AV delay assessment. Such an algorithm, when implemented in a pacemaker would significantly improve patient hemodymmics and reduce follow-up time. I P 121 F. Zolezzi, Cardiology, PERSISTENT CLOSED LOOP STIMULATION AND HEART RATE VARIABILITY IN NEUROMEDIATED SYNCOPAL PATIENTS R. Negro, C. Orvieni, L. Brandolisio’, R. Audoglio’. Div. of Osp. Civile, Vigevano, Italy; * Biotronik-Seda, Treuano SIN, Italy Recent studies have shown that Closed Loop Stimulation (CLS), modulated by variations in myocardial contractility, is effective in preventing highly symptomatic, neurally mediated syncopal episodes whether of cardio-inhibitory vasovagal or of carotid sinus (CS) origin. Aim of OUT study was to assess if a normal pattern of heart rate vatiability (HRV) before and after implantation of an INOS’+CLS (Biotronik, D) pacemaker is maintained. The need of this investigation was dictated by the consideration that CLS algorithm requires a persistent (>85%) ventricular pacing to be effective, whereas these pts need to be paced in concomitance of syncopal spell only. It was assumed that CLS maintains an acceptable HRV if difference between data collected before and after device implantation are not statistically signiiicative.Eighteen consenting, syncopal pts (15 m, age 68.3&13.6 yrs, mean 3.4 syncopeiyear), with no evidence of atria1 or AV conduction diseases, were enrolled in the study. Syncope was reproducible in all pts, with positive response to HU’IT (Type 2A) in 6 pts, to CS massage (pause > 5s) in 10 pts and to both in 2 pts. A 24-h ECG Halter and RR variability analysis were performed before and after device implantation. Mean HR, percentage of delta of R-R>50ms (pNN50), standard deviation of averaged RR/5min (SDANN) and standard deviation (SDNN) were analyzed and compared. The results, respectively before and after implantation, were: HR 63.W12.7 vs. 69.6&10.2 bpm; pNN50 20.5&21.6 vs. 21.3&12.4%; SDANN 138.4&58.9 2003 vs. 99.2&27.2 ms; SDNN 172.4&63.8 vs. 128.3&35.8 ms. Furthermore, no occurrences of syncopal spell were observed in these pts during a mean FU of 10.4 months. Data collected before and after implantation are comparable and differences between groups are not statistically signiiicative. Therefore, this study confirms that Closed Loop Stimulation is highly specific and effective in the treatment of mummediated syncope and preserves a physiologic HRV, despite a persistent ventricular pacing. I P 122 RIGHT VENTRICULAR INLET SEPTAL PACING WITH STEROID ELUTING TINED J-LEAD Y. Shimoyama. National Tokyo Medical Center Introduction: Although right ventricular pacing leads are traditionally placed in the apex, right ventricular inlet septal (near His-bundle) pacing is recommended in order to obtain more physiological pacing (in order to obtain minimal change of the intravenhicular conduction). However, this septal pacing often appeared high thresholds when screw-in leads were used. The very good results using steroid eluting J shaped tined leads were obtained to pace right ventricular inlet Septulll Patients and Methods: Fifteen patients were paced with steroid eluting J-shaped tined leads (CapSure Z Novas: model 5554, Medrtonic, Minneapolis, MN, USA) in the right ventricular inlet septum during DDD pacemaker implantation. Five patients were diagnosed sick sinus syndrome and ten were atria1 fibrillation with bradycardia Ventricular sensing, pacing thresholds (0.5 ms) and lead impedance at 0.5ms-2.5V were measured with ERA 300B (Biotronik, Berlin, Germany) at implant and followed up. Results: At implant, sensing amplitude were 5.6&3.7 mV, pacing thresholds were l.O&O.S V and lead impedance were 766&150 ohms. Three postoperative months (POM), mean sensing amplitude were 5.3&2.9 mV, pacing thresholds (0.41~) were 1.3&1.2 V and lead impedance (0.4ms-2.5V) were 800&146 ohms. At 6 POM, those were 6.3&3.8 mV, 1.7&1.4 V, and 820&125 ohms. At 12 POM, those were 6.3&2.9 mV, 1.6&1.2 V, and 817&144 ohms. Conclusion: Right ventricular inlet septal pacing with J-shaped steroid eluting tined leads revealed very good results. I P 123 TRICUSPID REGURGITATION PERMANENT PACEMAKER M. Yahalom, M. Gellerman, Galilee Hospital, Nahariya, IN PATIENTS WITH N. Roguin. Department of Cardiology, Western Faculty of Medicine, Technion, Haifa, Israel Introduction: The prevalence of Tricuspid Regurgitation (TR) in patients (pts) Pacemaker (PP) is still controversial. Objectives: The aims of the study were to investigate tricuspid regurgitation (TR) in patients (pts) with Permanent Pacemaker (PP) and it’s severity and association with Mitral and Aortic valve disease and with decreased LV function. Materials and Methods: Eighty one consecutive pts, who attended the pacemaker clinic during the second half of the year 2001 [46 males and 34 females, mean age 74.5 year (range 21.94)] were studied. All 81 pts were examined by two-dimentional and Doppler Echocardiography (2 months to 10 years following PP implantation). Results: The amount of TR was judged as normal or trivial in 28(35%), mild in 35(43%), moderate in 14(17%), and severe in 4(5%). Out of 22 pts with Mitral valve disease, 14(64%) had moderate/severe TR. In 59 pts with “normal flow” Mitral valve only 4(7%) had moderate/severe TR @<O.OOl). In 7(58%) out of 12 pts with severely decreased LV function there was moderate/severe TR. In 69 patients with EF>35%, moderate/severe TR was seen in 11(16%), @<0.004). Aortic Valve (Av) d’wax was found in 11 pts: 4128 (14%) with trivial TR, 4135 (11%) with mild TR, 3114 (21%) with moderate TR. None of the severe TR pts had AV disease (p=ns). Conclusions: The severity of TR in pts with PP is signiiicantly correlated with Mitral valve disease and with decreased LV function, not with Aortic valve disease and perhaps not with the implantation per-se. with Permanent I P 124 ELECTRICAL AND HANDLING CHARACTERISTICS OF A N E W DOWNSIZED PASSIVE FIXATION PACING LEAD A. Schuchert, E. Stammwitz, B. Now& I. Rows, D. Schmitz, N. Doll. For The Is&x S European Evaluation Group; UKe Hamburg Germany, Kreiskrankenhaus Leer Germany, Ccb Frankjirt Germany, University of Tartu Estonia, St. Johannes Dortmund Germany,Heruenhwn Leipzig Germany) Purpose: This study evaluated the performance of the new St. Jude Medical IsoFlex S lead family. These silicone leads feature a reduced lead body diameter, which is coated with Fast-Passm to improve handling. The atria1 “.I” version of the lead has a bipolar tip-ring spacing of 10 mm, the straight version 12 mm. Europace Supplements, Vol. 4, December 2003 B95
© Copyright 2026 Paperzz