Intralaboratory Quality Control of Hematology Comparison of Two Systems TERENCE R. J. LAPPIN, PH.D., CHRISTOPHER L. FARRINGTON, PH.D., MAURICE G. NELSON, M.D. AND JOHN D. MERRETT, PH.D. Lappin, Terence R. J., Farrington, Christopher L., Nelson, Maurice G., and Merrett, John D. Intralaboratory quality control in hematology. Comparison of two systems. Am J Clin Pathol 72:426-431,1979. Two systems for quality control have been compared, viz., the whole-blood control preparation method and the algorithm method using the geometric moving average X„ and a new estimator Y. The system involving wholeblood controls has the advantage of simplicity of operation, but the economic cost of commercial preparations is often high. The algorithm system has the advantage that results of all the test samples are used in the calculation; to some extent, this provides a buffer against random variation. The number of count-outs in a given channel is related to the precision of the channel, which in turn is a function of the number of determinations and calculations required for that result. An error of around 1% is introduced into the result each time a calculation is performed. A successful quality control scheme should contain elements of both control preparation and algorithm methods. (Key words: Algorithm methods; Control preparations; Geometric moving average; Quality control; Patient data.) TO ENSURE THE ACCURACY of hematologic test results, it is essential to employ quality control procedures to monitor laboratory equipment. Current methods may be broadly divided into two main categories. In the first approach, reference or control preparations of whole blood are used to adjust the instrument and monitor its performance, usually after every twentieth sample has been analyzed. Any consistent error or drift may be detected by the operator either by inspection of the results or by plotting the cumulative sum of differences between the reference value and successive observations. 5 In the second approach, the results of analysis of patient samples are statistically analyzed at various levels of mathematical sophistication, which range from calculation of the mean value of batches of results to complex algorithm methods. Dorsey 6 and Bull and Elashoff2 found that the data derived from erythrocytic indices of their patient populations showed no significant variation from Received May 1, 1978; received revised manuscript and accepted for publication July 18, 1978. Address reprint requests to Dr. Lappin: Department of Haematology, Royal Victoria Hospital, Belfast, BT 12 6BA, Northern Ireland. Department of Haematology, Royal Victoria Hospital, Belfast, and the Department of Medical Statistics, Queen's University, Belfast, Northern Ireland day to day; furthermore, these results were symmetrically distributed and were approximately gaussian in distribution. A recent advance in this statistical approach to quality control was the introduction of the algorithm X B by Bull, Elashoff, Heilbron and Couperus, 3 for monitoring automated analytical instruments using patient erythrocytic indices. The main purpose of the present investigation was to compare these two main approaches to quality control. It was hoped that such a comparison would provide information to enable operators of automated equipment to choose a scheme that would facilitate close tolerance monitoring of their analytical instruments. In addition, a further algorithm, Y, which is mathematically closely related to X B and designed to "amplify" any error caused by drift, was tested in parallel to X B . Methods In order to establish mean values for our hospital population, 4,535 consecutive routine patient samples were analyzed on the Coulter® Model S over a 20-day period. Each morning the Coulter counter was adjusted by sampling the commercial 4C control preparation three times, followed immediately by a secondary control, prepared by the method of Carville and Lee, 4 which was then assigned the observed values for the remainder of the day. Two different batches of secondary standard were used throughout the 20-day test period, and very little day-to-day variation was found. When the 4C control preparation was checked against manual methods a correction for trapped plasma was applied. Solely for the purpose of this part of the study, the patient samples were grouped together in batches of 19, according to whether the patients were inpatients, outpatients, or patients from a maternity hospital. 0002-9173/79/0900/0426 $00.80 © American Society of Clinical Pathologists 426 QUALITY CONTROL IN HEMATOLOGY Vol. 72 • No. 3 427 Table 1. Hematologic Results from Hospital Patients (Mean Values) General Hospital Outpatients General Hospital Inpatients Maternity Hospital Patients All Hospital Patients Hemoglobin, g/dl Erythrocyte count, x 106//xl Hematocrit, % Mean corpuscular volume (MCV), fl Mean corpuscular hemoglobin concentration (MCHC), g/dl Mean corpuscular hemoglobin (MCH), Pg 13.78 4.83 41.85 86.65 13.15 4.56 40.36 88.46 12.26 4.11 35.88 87.24 13.09 4.52 39.44 87.28 32.94 28.54 32.58 28.82 34.18 29.82 33.20 28.99 No. of patient samples 1,129 2,362 854 4,345 Every twentieth sample tested throughout the day was either the commercial control or the secondary control. The rather formidable mathematical expressions for XB and Y are detailed in the Appendix. Korpman and Bull9 have outlined the program steps necessary for X B . In order to provide data to compare the control preparation method with the algorithm methods, a further 1,900 consecutive patient samples were analyzed by the Coulter S. These patient samples were tested in the ordinary manner in batches of 18-20, but were not sorted according to source. For this part of the study only the 4C control preparation was used, because it was considered more stable than our own secondary control. Therefore, after each batch of patient samples, the 4C control preparation was analyzed, and the results were considered "in control" when they varied by less than 3% from the manufacturer's asigned mean value. Conversely, they were considered "out of control" when the variation exceeded 3%. When the XB or Y value for a batch of patient results exceeded a calculated 3% limit, this also was adjudged to be a loss of control. Loss of control as defined by either of these criteria is hereafter referred to as a "count-out." The control preparation and patient sample data were analyzed and the numbers of agreements and disagreements recorded. Results As can be seen from Table 1, the mean corpuscular volume (MCV) appeared to be little influenced by the sample source. However, the mean values for hemoglobin (Hb), erythrocyte count, and hematocrit (Hct) were lower in the maternity hospital group than in either of the other two groups. For the mean corpuscular hemoglobin concentration (MCHC) and mean corpuscular hemoglobin (MCH), the converse was found, with both these values approximately one unit higher for the maternity hospital group than for the rest of the general hospital patients. This last finding is in agreement with that of Holly. 8 Quality Control Trial Based on 1,900 Patient Samples Commercial Control Preparation. Table 2 shows the number of times the instrument was adjudged out of control when using the commercial control preparation. For this part of the study, 105 batches of samples were analyzed, but a temporary printing error in the Coulter S reduced the number of MCH batches avail- Table 2. The State of Control of the Coulter Model S as Adjudged by Three Different Methods of Quality Control during Analysis of 1,900 Consecutive Patient Samples Number of Times 'In Control' Within Mean ± 3% Total Number of Batches Channel* Control Preparation Method 105 105 100 105 105 105 MCV MCHC MCH Hb RBC Hct 102 80 80 90 95 83 ' MCV >= mean corpuscular volume; MCHC = mean corpuscular hemoglobin con- Number of Times 'Out of Control" Outside Mean ± 3% Algorithm Method x„ Y 94 98 85 99 99 91 — — — — — — Control Preparation Method 3 25 20 15 10 22 centration; MCH = mean corpuscular hemoglobin; Hb count; Hct = hematocrit. Algorithm Method X„ Y 11 7 15 6 6 9 — — — — — — hemoglobin; RBC erythrocyte 428 LAPPIN£7"AL. able for the subsequent statistical analysis to 100. The three calculated parameters (Hct, MCHC and MCH) showed a higher number of "count-outs" (outside the mean ± 3% limits) than the three directly measured values. Of the six channels, the MCV channel was clearly the most stable. Algorithm Data. The XB and Y formulas were applied to the individual patient results in batches of 1820 samples. The figures thus obtained were plotted on linear graph paper on which the grand mean and the preset limits (±3%) are indicated for the parameter under consideration. Table 2 shows the number of times that the analytical instrument was adjudged out of control by the algorithms XB and Y. When the instrument was in control, as adjudged by XB, it was also in control as adjudged by Y, with only one exception, in the total of 310 batches. On the other hand, the instrument was out of control on 13 occasions as adjudged by XB when deemed in control by Y. This discrepancy between the two algorithms XB and Y can be explained when the basic differences between the methods are considered. The algorithm XB as described by Bull and co-workers3 is essentially a moving average using the mean of previous batches to weight the present batch. However, the algorithm Y was used in a discrete batch mode using the grand mean instead of the mean of previous batches to weight each batch of patient results. The effect of this difference is that XB can indicate a loss of control due to an accumulation of several batches, whereas Y starts fresh on each batch and will only indicate whether an individual batch is out of control. A second difference between the two algorithms is that part of the equation for Y is raised to the power of 3, which leads to an ever-increasing Y value for a given deviation from the mean. The effect is to provide the operator with a more striking indication of loss of control than is possible with XB. Comparison of Control Preparation and Algorithm Data. The algorithm methods were applied only to those channels, viz., MCV, MCHC and MCH, which are known to be relatively stable within the patient population. The algorithm data were then compared with the appropriate results obtained from the control preparations (Table 2). For the MCHC and MCH channels, loss of control was indicated more frequently by the control preparation method than by either of the algorithm methods. In contrast, the MCV was shown to be out of control more often by the algorithm methods than by the commercial control method. A.J.C.P. • September 1979 Percentage Error in Calculations made by the Coulter S To investigate the magnitude of the error of the calculations performed by the Coulter S, we analyzed the data from this trial by performing independently the calculations for the Hct, MCHC and MCH. It was clear that each time a calculation is performed by the Coulter S an error of the order of 1% was introduced into the result. Therefore, the MCHC determination might be expected to be less precise than the MCH determination. This trend was confirmed by the results obtained in an interlaboratory trial involving five separate Coulter S counters. Our calculations based on the data of Pinkerton and associates10 gave coefficients of variation of 0.91% for MCV, 0.94% for erythrocyte count, 1.20% for Hb, 1.19% for MCH, 1.22% for Hct, and 1.43% for MCHC, indicating that the precision of a particular parameter is in general related to the number of calculations involved. Relative Costs of the Two Systems The relative costs of the systems are dependent upon whether an on-line computing facility is available. Where such facilities exist for data processing, the extra cost of running a quality control scheme using the algorithm approach would be negligible. On the other hand, where only a desk top calculator is available, it could take one technician VA-2 hours to perform the necessary computations for 300 specimens on the three channels under consideration. For a quality control system based on a control preparation the overall cost is dependent upon several factors. First is the cost of the preparation, whether commercial or prepared locally in the laboratory; second, the cost of analyzing the control samples in terms of technician time and reagent consumption; third, the cost of any necessary computations, e.g., a cusum. This last factor may again be negligible when a laboratory computer is used, but would require 15-20 min for 15 control results (300 specimens, with 1 in 20 a control) for four channels. Discussion Both methods of quality control have inherent advantages and disadvantages. The algorithm methods generally show fewer "count-outs" than the wholeblood control method, presumably because the large reservoir of patient results exerted a smoothing effect on the data. As may be seen from Table 1, both MCHC and MCH showed a higher mean value for maternity hospital patients than for either the general Vol. 72 • No. 3 QUALITY CONTROL IN HEMATOLOGY hospital outpatients or inpatients. It would be expected, therefore, that these two channels would be susceptible to source weighting. While about half the MCH count-outs were due to source weighting, only one of the MCHC count-outs was due to this cause. There is no apparent reason for this discrepancy. Since the number of count-outs involved is relatively small, the problem is easily circumvented by examining any suspect batches for specimen source. In our experience, sample source weighting has not been found to be a serious problem. A further point worthy of note is that the assay values supplied by the manufacturers for any batch of commercial control are not necessarily correct. This in turn could lead to incorrect setting of the Coulter S. When an algorithm method is used to monitor the performance of the instrument, any error in setting would be reflected in a consistently high or low XB or Y. During three years' experience of using the algorithm Y routinely in the laboratory, one such error was detected in the assigned values of one batch of commercial control. In this case the Y values for the MCV, MCHC, and MCH of each batch of 20 patient samples were consistently more than 3% above the expected values. The commercial control preparation was reassayed and new values assigned. When the Coulter S was reset using these corrected values, subsequent calculation of the Y algorithm for patient sample batches indicated a return to a satisfactory state of control. There are two main disadvantages of the whole-blood control method of quality control. The first is the high cost of commercial fixed whole-blood preparations, but this may be partially overcome by the preparation of a secondary control suspension, albeit with less stability. The second disadvantage of using a whole-blood control is that a large number of count-outs occur on some channels when limits of ±3% around the mean are used. It may be calculated from the data of Pinkerton and colleagues10 that 2 SD represent a variation of the order or ±2% or ±3% on all the channels under consideration. A small degree of instrumental drift, say 1.5%, which would not in itself justify resetting, could by summation with the inherent statistical variation give an increased number of count-outs. If it is accepted that the Coulter S results should lie within ±3% limits, then the essential problem of any quality control scheme is to detect when these limits have been reached. The control preparation method is probably most useful for the directly measured values. The algorithm methods were directly applied to the MCV, MCH, and MCHC. However, so long as one of the primary determinations, e.g., hemoglobin, is 429 checked against a reference method daily, the primary determinations will remain in control while the algorithms show the indices to be in control.1 The main advantage of the whole-blood control method lies in its simplicity. When the error on any channel is greater than about 5%, it will nearly always be obvious to the operator from the results of sampling such a control. Provided that a stable whole blood suspension is available, and it is sampled regularly throughout the day, useful statistics about the overall performance of the instrument may be derived. The main advantage of the algorithm method is that the results for all the samples tested are used for the calculations. In this sense the large amount of patient sample data available is acting as a buffer against random variation. In addition to use in daily quality control, the algorithms have a further advantage of being able to detect any error in the values assigned to a commercial control preparation that is used either for setting or quality control. The inherent problem with both the algorithm and control sample methods of quality control is the number of false count-outs. When algorithms only are used, the problem may be overcome by either rerunning a patient sample previously assayed when the instrument was known to be in control1 or by using a whole-blood control sample. When a whole-blood control sample is used for quality control, probably the simplest method to check a count-out is to reanalyze the control preparation. In our experience running the two methods of quality control in parallel prevents any disruption of the laboratory routine by eliminating the need to rerun either a patient sample or a control specimen when a count-out occurs. Thus, the algorithms provide a cross-check in the event of the control showing a count-out, and vice versa. With this approach one method can successfully complement the other to provide a very comprehensive scheme of quality control for any automated instrument. During this trial it became increasingly clear that correct setting of the Coulter S is an essential prerequisite for the production of accurate results. A very extensive calibration procedure is detailed in Appendix 3 of the Coulter S Operator's Manual; however, this protocol may prove to be too cumbersome for routine daily use. Many laboratories may prefer to use a control preparation for setting the instrument. Since no complete reference material is yet available, the best substitute for a commercial control preparation may well be a preparation rigorously analyzed within the laboratory by the methods suggested by Gilmer and associates.7 When a control preparation is used for 430 A.J.C.P. • September 1979 LAPPIN£7/AL. setting the Coulter S, it should be sampled a minimum of three times to allow for the inherent random variation. Thus, the mean of each parameter can be calculated and the instrument set accordingly. By use of a control in this manner the entire operation of the Coulter S is dependent on the assigned values of the control. Assuming that the grand mean values of the algorithm methods were determined over a long period, covering at least several batches of control preparation, an independent intralaboratory check for the instrument is thus available. The algorithms also serve to ensure not only the within-day accuracy but also accuracy from day to day and month to month. On the basis of the results of this trial and our experience over the past three years with the algorithm methods, we feel that an efficient quality control scheme for the Coulter S should contain elements of both the whole-blood control method and an algorithm method. Where a dedicated computer is available, the calculation of the algorithms in an on-line mode can be readily accomplished. Acknowledgment. Mr. A. Lamont, Senior Chief Technician, Department of Haematology, assisted throughout the course of this work. References 1. Bull BS: A statistical approach to quality control, Quality Control in Haematology. Symposium of the International-Committee for Standardization in Haematology. Edited by SM Lewis, JF Coster. London, New York, San Francisco, Academic Press, 1975, pp 111-121 2. Bull BS, Elashoff RM: The use of patient-derived hematology data in quality control. Preprint, Proceedings of the San Diego Biomedical Symposium 13:1-5, 1974 3. Bull BS, Elashoff RM, Heilbron DC, et al: Study of various estimators for the derivation of quality control procedures from patient erythrocyte indices. Am J Clin Pathol 61:473— 481, 1974 4. Carville JM, Lee D: Red cell suspension as a working standard. J Clin Pathol 22:738, 1969 5. Cavill I, Jacobs A: Quality Control in Haematology. Broadsheet 75, Association of Clinical Pathologists, 1973 6. Dorsey D: Quality control in hematology. Am J Clin Pathol 40:457-464, 1963 7. Gilmer PR, Williams LJ, Koepke JA, et al: Calibration methods for automated hematology instruments. Am J Clin Pathol 68: (Suppl)185-190, 1977 8. Holly R: The iron and iron-binding capacity of serum and the erythrocyte protoporphyrin in pregnancy. Obstet Gynecol 2: 119-126, 1953 9. Korpman RA, Bull BS: The implementation of a robust estimator of the mean for quality control on a programmable calculator or a laboratory computer. Am J Clin Pathol 65:252253, 1976 10. Pinkerton PH, Spence I, Ogilvie JC, et al: An assessment of the Coulter counter model S. J Clin Pathol 23:68-76, 1970 APPENDIX Mathematical Treatment of Results Consider X B , the "robust" geometric moving average described by Bull and Elashoff2: XB,i = XB,,_, + sgn ( X j ! — X B > 1 _•i ;) 2 V | X J t - XB.,-,1 where XB,,., is the average after i - 1 batches and Xji are individual patient values, and N is the number of patient samples considered. To calculate XB.i, a starting value of X B , i.e., X B 0 must be used. The rather complex mathematical expression for XB,i is readily calculated using a desk-top calculator with a program of about 80 steps. 9 It is possible to simplify equation 1 as follows: L e t d = Xji — X B ,i_i i.e., d represents the difference between the individual value and the mean of the preceding batch for the parameter under consideration. Equation 1 can then be simplified to: XB.i = XB,i-i + sgn 2 2-*VRTJ -JN2 (2) We have examined the effect of raising the function ( I d V j d j ) 2 to a higher degree, i.e., ( £ dVTd]) 3 . Instead of using the mean of the preceding batch as a smoothing function for X B , as suggested by Bull, 1 we chose to use the mean value for all patient samples. Thus, for the new esti- (Xji — XB>i_!) . ii A V |Xji — XB>1_i I (1) N B,1-1 mator, which we have designated Y, Y = m i N2 (3) where d x „ - XD; XD is the known mean value of the hospital patient population and N is the number of results considered. The value of Y has no real meaning in the physical sense since the higher power in the "cumulative" function has units of measurement different from X B . The practical differences between these two algorithms is best illustrated using the case where the MCV is set one unit high. The graph for X B rises by one unit and remains at that level, whereas the value of Y rises with each succeeding sample result (Fig. 1). The X B value may be used to recalibrate the instrument, whereas that for Y cannot. However, in practice, it is very difficult to reset the potentiometers on the Coulter S by a margin in order of 1%, and in practice the instrument would be reset only when the deviation reached 3%. 3 While Y does not allow a decision to be made with respect to the amount of recalibration required, simply by Vol. 72 • No. 3 QUALITY CONTROL IN HEMATOLOGY 431 20 14 3 > > o CD IX 15 12 O Y 10 • 01 • X B o 10 • 2 (0 > 00 • \20 X, 8 • IX I o o 4 ' N u m b e r of s a m p l e s w i t h r e s u l t s one u n i t above the mean FIG. 1. The relationship between the number of samples with results one unit above the mean and the XB and Y values. inspection of the plot of the results, it is possible to plot a graph of Y against percentage deviation (Fig. 2) for 20 sample results, i.e., a Y20 value. As a linear relationship exists between Y and the number of patient specimens tested, it is considered valid to calculate a Y20 when fewer than 20 specimens are analyzed, although for practical purposes an arbitrary minimum of 18 has been set. One striking feature of Figure 2 is the rapidly increasing Y value with increasing deviation of the results from the mean, resulting in an "amplification" of the error, leading to easier recognition of a loss of control by the laboratory staff. 0 1 2 3 4 5 6 D e v i a t i o n from the mean value FIG. 2. The relationship between the deviation from the mean value and the Y and XB values. The use of the patient grand mean for all Y computations means that Y is not a moving average, yet it still retains the "robust" element of X B . In practice it means that significant changes (>3%) in the average of a batch of 20 patient samples will not affect succeeding batches. This means that a batch of samples with predominantly abnormal results will not influence subsequent batches.
© Copyright 2026 Paperzz