Technical Notes # 014 Precise Mass Performance of the micrOTOF with either DualSprayer Internal Calibration or Highly Stable External Calibration The micrOTOF benchtop ESI-TOF system offers excellent mass accuracy over a wide dynamic range. A mass accuracy of 3 ppm is achieved with both internal and external calibration. This note describes the use of a dual sprayer setup on the Apollo source for the identification of natural products from green tea using exact mass determination. Introduction Because of their antioxidative properties flavanoids from plants have gained increasing interest in the last years. Different positive health effects are discussed, such as protection against cancer and coronary heart diseases [1], [2]. For tea, flavanoids are of great importance as they contribute to the colour and flavour of the tea brew. The content of flavanols (catechins) in fresh tea leaves can add up to 30% of the dry mass. Another major constituent in tea is caffeine with a quantity of about 4%. The following application shows the identification of catechins from green tea using exact mass determination with simultaneous consideration of the true isotopic pattern (TIP). purpose of internal calibration the Bruker Apollo source was equipped with two sprayers, a simple configuration option which is available without source modification. The setup is shown in figure 1. Experimental The sprayers are mounted opposite to each other. The position of each nebulizer relative to the mass spectrometer inlet may be adapted by spacers accor- The micrOTOF was coupled to an Agilent 1100 HPLC system. For the Plant extracts are normally very complex. The greatest inconvenience when analyzing flavanoids is the lack of commercially available standards that complicates the compound identification. Here we show HPLC separation in combination with precise measurement of both mass and isotopic pattern as useful tools in the identification of natural compounds. The micrOTOF offers the possibility of empirical formula determination using both accurate mass and statistical isotopic pattern analysis. A mass precision of 3 ppm can be achieved with internal calibration. The calibrant is introduced into the ion source at the same time as the sample via a second sprayer. In this way each spectrum can be internally recalibrated using automated software routines. Fig. 1: Dual Sprayer setup on the Bruker Apollo source. Left side: Sprayer with low flow spacer for introduction of calibration mix (2 µL/min). Right side: Sprayer with high flow spacer for connection of the HPLC flow at 200 µl/min. Enabling Life Science Tools Based On Mass Spectrometry™ Sample: Hot water was added to a bag of green tea and brewed for 5 min. After cooling, the tea was diluted 1:10 in H2O/acetonitrile (90/10). Results Fig. 2 shows the separation of catechins and caffeine from green tea. The corresponding structures and elemental formulas of the compounds are illustrated in figure 3. Seven major signals were detected. Two peaks were obtained for 291 m/z and 307 m/z. They belong to the two different epimers, catechin and epicatechin respectively gallocatechin and epigallocatechin. Fig. 2: Separation of caffeine and flavanols from a green tea brew. ding to the flow rate. A low-flow spacer is used for the nebulizer that delivers the calibration mixture at a flow rate of 2 μl/min, whereas the nebulizer for the HPLC eluent is fixed on a high-flow spacer. In order to adjust the nebulizer pressure for the two sprayers independently a second nitrogen pressure valve was used. For internal calibration a standard mixture of perfluoroalkylphosphazine (ES Tuning Mixture, Agilent) was used. The calibrant was infused via a syringe pump. isocratic post time (21 min) Flow: 0.2 ml/min Injection: 5 μl UV Detection: 280 nm MS parameters Mode: ESI positive Nebulizer: HPLC: 1.4 bar, syringe pump: 0.4 bar Dry Gas: 4 l/min at 220°C The first step in data processing is the extraction of the spectra. For all compounds, spectra were automatically detected and averaged over the whole peak width, a function made possible by the very stable mass accuracy over dynamic concentration range. Each average spectrum was internally recalibrated using the calibration masses m/z 118.0863, 322.0481, 622.0290 and 922.0098 m/z. Afterwards elemental formulas were calculated using the “GENERATE MOLECULAR FORMULA” option in DataAnalysis. By default, the elements C, H, O and N are considered. Other HPLC and MS parameters are given in the following: HPLC conditions Column: Prontosil 120-3C18 H 3.0 μm, 53 x 2.0 mm i.d. Eluents: A = 0.1 % acetic acid in water B = acetonitrile Gradient: 12% B, 4 min isocratic (4 min) to 25% B in 3 min (7 min) 25% B, 8 min isocratic (15 min) to 12% B in 1 min (16 min) 12% B, 5 min Fig. 3: Structures and formulas of catechins and caffeine. Moreover, an intensity-independent accuracy has the advantage that concentration of sample and calibrant do not need to be adjusted. Low quantities of the calibrant are sufficient and minimize the risk of signal suppression and interferences in the spectrum. The relative intensity of the calibration peaks is ≤ 5 % in our example. The results show that the dual sprayer setup on the micrOTOF works smoothly. For comparative interest, alternative ways of data processing were investigated: Spectra of peak 27 were externally recalibrated using the calibration equation obtained from peak no. 1. Secondly, another HPLCMS run was performed under identical conditions. The spectra of the second analysis were externally recalibrated Fig. 4: Average spectrum of peak 3: Coelution of caffeine and catechin. Signal intensity of catechin is about 1% relative to caffeine. Other Parameters were: • mass tolerance 3 ppm • electron configuration even • apply nitrogen rule. The results are summarized in table 1. The elemental formulas of all expected compounds could be determined with a mass error of less than 3 ppm. Unique results were obtained for the peaks 1-5. Only peak 6 and 7 gave more than one formula, the top hit is listed in table 1. It corresponds to the result with the best SigmaFitTM (lowest Sigma value), meaning the best match for measured and theoretical isotope pattern (for details see [3]). The excellent accuracy over a wide dynamic range is demonstrated with the spectrum of the co-eluting peaks number 3 and 4, namely caffeine and catechin. The averaged mass spectrum of peak 3 is shown in figure 4. Caffeine is by far the dominant signal. Catechin has an intensity of about 4000 cts, which is approximately 1 % relative to caffeine. Its mass peak is only visible if the m/z range of interest is greatly enlarged. The results for caffeine and catechin calculated with “GENERATE MOLECULAR FORMULA” are shown in figure 5. For both compounds a unique formula with very good mass precision and fit of the isotope pattern is obtained, showing that co-elution is easily handled even if there is a wide difference in terms of signal intensity. Table 1: Flavanols and caffeine identified by accurate mass data obtained from the micrOTOF ESI-TOF MS with internal calibration. Peak no. RT [min] Measured m/z Calc. Formula Calc. [M+H]+ Sigma value Error [ppm] Substance 1 2.2 307.08114 C15H15O7 307.08123 0.0070 0.298 Gallocatechin 2 3.1 307.08109 C15H15O7 307.08123 0.0032 0.437 Epigallocatechin 3 4.0 195.08725 C8H11N4O2 195.08765 0.0046 2.056 Caffeine 4 4.0 291.08675 C15H15O6 291.08631 0.0152 -1.505 Catechin 5 6.6 291.08592 C15H15O6 291.08631 0.0053 1.359 Epicatechin 6 7.2 459.09321 C22H19O11 459.09219 0.0007 -2.233 Epigallocatechin gallate 7 11.0 443.09842 C22H19O10 443.09727 0.0057 -2.590 Epicatechin gallate Table 2: Comparison of mass errors calculated from 3 different experiments. A: First run; all spectra internally recalibrated. B: First run; all spectra externally recalibrated using the calibration coefficients from peak no. 1. C: Second run; all spectra externally recalibrated with the calibration equation from peak 1 of the first run. Peak no. RT [min] Calc. Formula A Error [ppm] B Error [ppm] C Error [ppm] Substance 1 2.2 C15H15O7 0.298 0.298 1.396 Gallocatechin 2 3.1 C15H15O7 0.437 0.513 1.900 Epigallocatechin 3 4.0 C8H11N4O2 2.056 1.217 1.147 Caffeine 4 4.0 C15H15O6 -1.505 -1.894 0.235 Catechin 5 6.6 C15H15O6 1.359 2.228 0.972 Epicatechin 6 7.2 C22H19O11 -2.233 -2.244 -3.088 Epigallocatechin gallate 7 11.0 C22H19O10 -2.590 -0.405 -0.819 Epicatechin gallate 1.497 1.257 1.365 Average absolute Error [ppm] 0.878 0.868 0.916 Standard Deviation [ppm] For caffeine and catechin, which are present in the same LC peak, unique formulas with a mass error around 2 ppm were calculated simultaneously and confirmed by SigmaFitTM even though catechin is detected at a 100fold lower intensity. Although the dual sprayer setup works simply and smoothly on the micrOTOF it should be remembered that similar mass accuracy can be achieved with external calibration, because of the extreme inherent stability of micrOTOF accuracy over time. References [1] M. Wang et al., Cancer Res. 52 (1992) 1162-1170 [2] M.G.L. Hertog et al., Lancet 342 (1993) 1007-1011 [3] Bruker Daltonics Application Note ET-07 / LCMS-42 Authors: Andrea Kiehne, Oliver Räther Fig. 5: Result from “Generate Molecular Formula” for caffeine and catechin. Mass accuracy with internal calibration is better than 3 ppm. with the value obtained from peak no. 1 of the first run (20-30 min earlier). The mass errors from the three different experiments are listed in table 2. The mass error is ≤ 3 ppm for all three calibration methods. No significant difference is observed between internal and external calibration, even for consecutive LC runs. Here, the long-term stability of the micrOTOF is demonstrated resulting in excellent mass accuracy over time. Bruker Daltonik GmbH Conclusion The Bruker Apollo source can be easily converted from a single to a dual sprayer source and vice versa. Because of the wide dynamic range of the micrOTOF ESI TOF system it is not necessary to adjust sample and calibrant concentrations. The calibrant can be introduced at a low concentration in order to avoid signal suppression and interferences. © 2005 Bruker Daltonics For more information, please visit our worldwide web sites on the internet, call your local Bruker sales representative, or contact us at one of our application centers. www.bdal.de www.bdal.com Bruker Daltonik GmbH Bremen · Germany Phone +49 (421) 22 05-0 Fax +49 (421) 22 05-104 E-Mail [email protected] Bruker Daltonics Inc. Billerica, MA · USA Phone +1 (978) 663-36 60 Fax +1 (978) 667-59 93 E-Mail [email protected]
© Copyright 2026 Paperzz