No. 45 – 2016 > > Pipetting with a Smile: How to Enjoy Pipetting (BN 45) JUNE 2016 PAGE 1 Faster Isolation of PBMCs Using Ficoll-Paque™ PLUS in the Eppendorf Centrifuge 5920 R >Handling of Difficult Liquids Made Easy >New Centrifuge 5920 R with 4 Liter Capacity >Eppendorf Sterilize-in-Place Bioprocess Systems Application Notes Faster Isolation of PBMCs Using Ficoll-Paque® PLUS in the Eppendorf Centrifuge 5920 R · Faster and More Reproducible Cell Viability Assays with the Eppendorf epMotion® 96 · etc. NICOLE SEELIGMÜLLER, EPPENDORF AG, HAMBURG, GERMANY Introduction Human blood consists of equal parts of blood plasma and blood cells. These include erythrocytes (red blood cells), leukocyctes (white blood cells), and thrombocytes (platelets). Leukocytes are further subdivided into different cell types. These include, for example, lymphocytes and monocytes, which (in co-operation with other cells) form the basis of the innate immune system and which, owing to their simple nucleus, are referred to as peripheral blood mononuclear cells (PBMCs). The term lymphocyte encompasses two major classes, B lymphocytes and T lymphocytes. B lymphocytes are responsible for antibody production, whereas T lymphocytes produce signal molecules which will finally lead to the removal of diseased or foreign cells. PBMCs play an important role in the areas of infection diagnostics as well as in clinical research [1]. PBMCs are isolated from buffy coats (whole blood concentrates without serum). Centrifugation using FicollPaque™ PLUS (density: 1.007 g/mL) allows separation of PBMCs from other blood components: erythrocytes, granulocytes, and dead cells which, due to their higher density, will pass through the Ficoll layer, whereas lymphocytes and monocytes will accumulate at the interphase between plasma and Ficoll gradient (Fig. 1), [2,3]. One prerequisite for a clean PBMC isolation with a maximum yield of living cells is the formation of a clear interphase. For this reason, the procedure must be carried out with the least vibration possible. Usually, mixing of the phases can only be avoided by centrifugation with the rotor brake deactivated [4], which constitutes an extremely time-consuming step within this application. This study shows that the new 4 liter benchtop Centrifuge 5920 R not only satisfies the high demands of this method, but that thanks to the option of selecting individual ramps, the user will enjoy considerable time savings. Materials and methods Ficoll-Paque PLUS density gradient centrifugation 1. Fill Eppendorf Conical Tubes 50 mL with 15 mL Ficoll-Paque PLUS (GE Healthcare®) each. 2. Using the slowest pipetting speed available, overlay the Ficoll with 15 mL of the blood/PBS mixture. 3. Centrifuge the samples for 30 min in the swing-bucket rotor at 400 x g, 20 °C, at ramp* 0/0 or ramp 3/3, respectively, while selecting the setting “at set rpm”. *Possible ramp settings Centrifuge 5920 R: deactivated acceleration or deceleration, respectively (0/0) to fastest acceleration/brake (9/9) Purification of lymphocytes 1. Carefully remove 2/3 of the upper layer (plasma). 2. Aspirate as much of the PBMC layer as possible in the smallest volume possible and transfer to a fresh tube. 3. Wash PBMCs several times in PBS and resuspend in suitable cell culture medium. Viability test and determination of yield 1. Dilute cells 1:1 in trypan blue and count. 2. Determine viability and yield. Detailed information about materials and methods is available in Application Note 372 [5]. Results and discussion In order to evaluate the quality of separation, it was determined whether a defined interphase with clearly delineated phase transitions was visible. Fig. 2 (A–D) shows the results obtained from the density gradient centrifugations carried out in the Centrifuge 5920 R using different swing-bucket rotors. For each rotor, the acceleration and deceleration rates 0/0 and 3/3 were tested. It is evident that optimal separation of blood components was achieved in all cases, and that vortex effects during deceleration of the rotor could be completely avoided (see negative example Fig. 2E). Plasma PBMC Ficoll Granulocytes Erythrocytes A B C D Fig. 2: Results obtained after density gradient centrifugation in the Centrifuge 5920 R Fig. 1: Density gradient obtained from buffy coat using Ficoll-Paque PLUS (schematic) A) Rotor S-4x1000 with high-capacity bucket (ramp 0/0) B) Rotor S-4x1000 with high-capacity bucket (ramp 3/3) C) Rotor S-4xUniversal-Large (ramp 0/0) D) Rotor S-4xUniversal-Large (ramp 3/3) E) Negative example: cloudy interphase, turbid plasma-/Ficoll phase (different centrifuge) Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com E 2 EDITORIAL · DEAR READER Imprint Editorial team Berrit Hoff (Editor-in-Chief), Axel Jahns, Jochen Müller-Ibeler, Tanja Musiol, Natascha Weiß Publisher Eppendorf AG, Barkhausenweg 1, 22339 Hamburg, Germany Telephone: (+49) 40-53801-0 Fax: (+49) 40-53801-556 E-mail: [email protected] Web: www.eppendorf.com We welcome all readers’ articles for this Dear Reader! publication. However, no responsibility is accepted for unsolicited manuscripts. Important note The new products described may be In 1962, Eppendorf launched a revolutionary invention: the modern microliter pipette. Since that time, the world of pipetting has continually evolved, but so has our product portfolio. > With pipetting solutions for every taste and demand, as well as a broad spectrum of support and training opportunities, we aim to make your daily pipetting routine as pleasant as possible, as you will discover on pages 4 – 5. Here we will also introduce to you our new internet platform www.eppendorf.com/pipetting, where you will find numerous tips, tricks, and stories all about pipetting. Perhaps you will even recognize yourself in one of our new cartoons and start your next pipetting series with a smile! The topic of liquid handling will recur several times throughout this issue: in the contribution on the handling of difficult liquids (p. 6), in the article featuring the high chemical resistance of our pipette controller Easypet® 3 (p. 10), as well as in the study on the influence of different liquid handling instruments on cell assay performance (Application Note p. 3 – 4). Remember part 1 of our Application Note series “The Tip of the Iceberg”? For the sequel, the authors have scrutinized the openings of different pipette tips using microscopy – with impressive results (Application Note p. 5 – 6). In this issue, we proudly present you the refrigerated Centrifuge 5920 R, a true miracle of performance and space. With similar dimensions as a 3 liter centrifuge, it provides a capacity of up to 4 x 1,000 mL or 52 x 50 mL conical tubes. The Centrifuge 5920 R is the perfect combination of user-friendliness and precision (p. 8 and Application Note p. 1 – 2). The family of Eppendorf Tubes® 5.0 mL continues to grow! The new screw cap tube 5.0 mL was developed for certain applications in the medium volume range, where a screw cap is beneficial to either experiment or sample handling. Learn more about the new “Eppi” on page 11. We wish you an exciting read! Your BioNews team launched at different times in various countries. Please contact your local Eppendorf organization or distributor for details. Technical specifications subject to change. Errors and omissions excepted. All rights reserved, including graphics and images. © Copyright Eppendorf AG, June 2016. Carbon neutrally printed in Germany. > CONTENTS 4 IN THE SPOTLIGHT STRAIGHT FROM THE LAB INNOVATION > 8 Pipetting – with a Smile 10 4 – 5 Proper Handling of Cells in the Lab (I) 7 State-of-the-Art DNA Measurement: Efficient, Fast, and Reproducible 9 Robust and Reliable: Easypet® 3 10 Holistic Approach to Ergonomics 12 Eppendorf Sterilize-in-Place (SIP) Bioprocess Systems 13 Handling of Difficult Liquids Made Easy 6 Centrifuge 5920 R: New Benchmark in Capacity and Performance 8 Eppendorf Tubes® 5.0 mL – Now Also Available with Screw Cap CLOSE-UP NEWS / TIPS SERVICE (BN 45) JUNE 2016 PAGE 1 Faster Isolation of PBMCs Using Ficoll-Paque™ PLUS in the Eppendorf Centrifuge 5920 R NICOLE SEELIGMÜLLER, EPPENDORF AG, HAMBURG, GERMANY Introduction Human blood consists of equal parts of blood plasma and blood cells. These include erythrocytes (red blood cells), leukocyctes (white blood cells), and thrombocytes (platelets). Leukocytes are further subdivided into different cell types. These include, for example, lymphocytes and monocytes, which (in co-operation with other cells) form the basis of the innate immune system and which, owing to their simple nucleus, are referred to as peripheral blood mononuclear cells (PBMCs). The term lymphocyte encompasses two major classes, B lymphocytes and T lymphocytes. B lymphocytes are responsible for antibody production, whereas T lymphocytes produce signal molecules which will finally lead to the removal of diseased or foreign cells. PBMCs play an important role in the areas of infection diagnostics as well as in clinical research [1]. PBMCs are isolated from buffy coats (whole blood concentrates without serum). Centrifugation using FicollPaque™ PLUS (density: 1.007 g/mL) allows separation of PBMCs from other blood components: erythrocytes, granulocytes, and dead cells which, due to their higher density, will pass through the Ficoll layer, whereas lymphocytes and monocytes will accumulate at the interphase between plasma and Ficoll gradient (Fig. 1), [2,3]. One prerequisite for a clean PBMC isolation with a maximum yield of living cells is the formation of a clear interphase. For this reason, the procedure must be carried out with the least vibration possible. Usually, mixing of the phases can only be avoided by centrifugation with the rotor brake deactivated [4], which constitutes an extremely time-consuming step within this application. This study shows that the new 4 liter benchtop Centrifuge 5920 R not only satisfies the high demands of this method, but that thanks to the option of selecting individual ramps, the user will enjoy considerable time savings. Materials and methods Ficoll-Paque PLUS density gradient centrifugation 1. Fill Eppendorf Conical Tubes 50 mL with 15 mL Ficoll-Paque PLUS (GE Healthcare®) each. 2. Using the slowest pipetting speed available, overlay the Ficoll with 15 mL of the blood/PBS mixture. 3. Centrifuge the samples for 30 min in the swing-bucket rotor at 400 x g, 20 °C, at ramp* 0/0 or ramp 3/3, respectively, while selecting the setting “at set rpm”. *Possible ramp settings Centrifuge 5920 R: deactivated acceleration or deceleration, respectively (0/0) to fastest acceleration/brake (9/9) Purification of lymphocytes 1. Carefully remove 2/3 of the upper layer (plasma). 2. Aspirate as much of the PBMC layer as possible in the smallest volume possible and transfer to a fresh tube. 3. Wash PBMCs several times in PBS and resuspend in suitable cell culture medium. Viability test and determination of yield 1. Dilute cells 1:1 in trypan blue and count. 2. Determine viability and yield. Detailed information about materials and methods is available in Application Note 372 [5]. Results and discussion In order to evaluate the quality of separation, it was determined whether a defined interphase with clearly delineated phase transitions was visible. Fig. 2 (A–D) shows the results obtained from the density gradient centrifugations carried out in the Centrifuge 5920 R using different swing-bucket rotors. For each rotor, the acceleration and deceleration rates 0/0 and 3/3 were tested. It is evident that optimal separation of blood components was achieved in all cases, and that vortex effects during deceleration of the rotor could be completely avoided (see negative example Fig. 2E). 11 ConvexAccess : Improved Handling, Increased Safety 7 Increase Your Pipetting Knowledge! 5 ™ Supporting You: Maintenance and Certification Services 10 Eppendorf Services 12 Eppendorf Prize Winners 2015/2016 Shigeki Watanabe & Adrian Liston 14 Trademarks Information 14 Win a Multichannel Pipette! 15 NICOLE SEELIGMÜLLER Faster Isolation of PBMCs Using Ficoll-Paque® PLUS in the Eppendorf Centrifuge 5920 R 1 – 2 ERIC GANCAREK, MAUD BRASSEUR, MURIEL ART, HANAË A. HENKE Faster and More Reproducible Cell Viability Assays with the Eppendorf epMotion® 96 3 – 4 ULRIKE GAST, LAURA KOCH The Tip of the Iceberg: How Pipette Tips Influence Results. Part 2: Perfect Geometry Makes a Difference 5 – 6 Plasma PBMC MAREIKE PANZ, JOHANNA PRIEB, INES HARTMANN Ficoll Granulocytes Erythrocytes A B C D Fig. 2: Results obtained after density gradient centrifugation in the Centrifuge 5920 R Fig. 1: Density gradient obtained from buffy coat using Ficoll-Paque PLUS (schematic) A) Rotor S-4x1000 with high-capacity bucket (ramp 0/0) B) Rotor S-4x1000 with high-capacity bucket (ramp 3/3) C) Rotor S-4xUniversal-Large (ramp 0/0) D) Rotor S-4xUniversal-Large (ramp 3/3) E) Negative example: cloudy interphase, turbid plasma-/Ficoll phase (different centrifuge) Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com E Improved Automated Single Cell Clone Analysis with the SynenTec Cellavista® System and Eppendorf Cell Culture Plates 3 7 – 8 > 4 IN THE SPOTLIGHT · PIPETTING – WITH A SMILE KAY KÖRNER, EPPENDORF AG Pipetting – with a Smile In the year 1962, Eppendorf launched a revolutionary invention to the market: the modern microliter pipette. Since that time, the world of pipetting has continually evolved. Previously unattainable demands on ergonomics and precision have become the new standard. Understandably, the expectations of modern science on the reproducibility of results have risen in parallel. As a variable which is difficult to control, the human factor can influence results significantly. This is where modern pipetting systems must exert their leverage – with a focus on easy handling. ppendorf invests much time and brainE power in the development of products which allow the user to enter the state of flow: lightweight pipettes, optimized tip fit, and high-contrast plate labeling. If you take a closer look or speak with one of our team members, you will discover a multitude of product features which can ease your life significantly. > Occasionally, potential problems may hide in the smallest details, invisible to the naked eye. For example, our laboratory had problems obtaining reproducible results when using pipette tips by other manufacturers. The Application Note on p. 5 – 6 in this issue “The Tip of the Iceberg: How Pipette Tips Influence Results. Part 2: Perfect Geometry Makes a Difference” provides more information on this topic. Really good pipetting is an acquired skill You too can pipette with a smile! Do you enjoy pipetting? The responses to this question range from “yes” with a smile to an eye-rolling “no”. As we know from experience, we enjoy activities when we are proficient, don’t suffer, and finally achieve success – and we enjoy a challenge. Psychologists call this state “flow” – the happy feeling of complete immersion in or concentration on a task. Some experience this state of flow during model construction or while playing soccer, while others find their happiness while pipetting: the purposeful striving to reach the 96th well and the trance-like state of repeating the same actions. However, blessing and burden are sometimes closely linked. A decisive factor which can make the difference between flow and frustration is also the quality of the instruments used. Getting into the flow A pipette that feels heavy, pipette tips that don’t fit. A plate on which the labeling is difficult to read. All these shortcomings not only tire the body, but also the spirit. A feeling of fun and ease is hard to come by. For these reasons, Really good pipetting, however, takes more than really good products. Mastering the right techniques and acquiring expertise are essential. This is why Eppendorf, beyond developing really good products, takes on the responsibility of providing the user with the knowledge required to reach the state of flow. For example, in addition to online webinars (see www.eppendorf.com/webinar), Eppendorf also offers pipetting training at different sites around the world. For example, do you recall off hand the exact density at which reverse pipetting > PIPETTING – WITH A SMILE · IN THE SPOTLIGHT 5 Tip Increase Your Pipetting Knowledge! Exact pipetting is a prerequisite for reliable results. Pipetting, at first glance, appears to be a simple action. In fact, the pipetting results can be influenced by numerous factors that are listed here from A to Z! Influencing factors on pipetting results becomes beneficial? This and other questions will be addressed, not only during our training sessions, but also on our new internet platform www.eppendorf.com/ pipetting. > On this new platform you will not only find a multitude of tips and tricks on pipetting. You can also test your pipetting expertise in a quiz challenge or discover if you recognize your daily lab routine in one of our cartoons which may help you survive the next long pipetting row with a smile! Automated pipetting In order to tame the human factor mentioned at the beginning, solutions for automated pipetting are available. The influence of the human hand as a source of uncertainty is eliminated to a large extent (in the case of an electronic pipette) or even entirely (in the case of a liquid handling workstation). A liquid handling workstation increases the reproducibility of pipetting results considerably as factors such as the pipetting angle, speed, and other similar parameters are largely kept constant. Especially since high accuracy is achieved, liquid handling workstations are capable of pipetting small volumes much more reliably than the human hand. This way, assays may be carried out using smaller volumes of r eagent – this saves money. A Adjustment settings B Body temperature and heat transfer C Calibration, last D Density of liquid E Environmental conditions F Foaming liquids G Geographical height and air pressure H Humidity of air I Immersion depth J Just wait: waiting time after aspiration K Knowledge of pipetting techniques While liquid handling workstations may hardly induce a state of flow, they will certainly help save a lot of time for “flow-worthy” pursuits. L Level of pipetting speed – work slowly M Maintenance status N Ninety degrees – hold pipette vertically O Operational vapor pressure P Prewetting Pipetting solutions for every taste Q Quality and design of pipette tip R Rhythm of pipetting S Size of air-cushion (as small as possible) T Temperature U Under stress: a tired hand does not pipette accurately V Viscosity W Wetting behavior of liquid X XXL temperature difference between sample and pipette Y Dispense against wall in same angle as the arms of a Y Z Zero patience With our today’s selection of pipetting solutions, we are sure to offer the right product for every taste – whether you are pipetting with a smile or not. Finding the right pipette back in 1962, however, was very easy. There was only one microliter pipette, and it was made by Eppendorf. Would you like to learn more? A wide selection of Eppendorf videos on pipetting can be found at www.eppendorf.com/ playlist-eppendorf-pipettes or by entering the QR code! The automated liquid handling systems of the epMotion® series by Eppendorf are available in four different formats and with various upgrading options. > 6 INNOVATION · HANDLING OF DIFFICULT LIQUIDS MADE EASY RUDOLF WALCZAK, EPPENDORF AG Handling of Difficult Liquids Made Easy Nearly every life science researcher works with liquids, and viscous or volatile liquids pose a significant challenge. Handling them correctly requires specialized knowledge as well as the proper tools. Eppendorf offers its customers a broad portfolio of products for liquid handling. The electronic dispensers Multipette® E3 and Multipette E3x (U.S./CAN: Repeater® E3/Repeater E3x) represent the latest innovations for working with difficult liquids and for handling long dispensing series. > Effortless handling of difficult liquids Water is the main component of living cells, and the majority of biological processes occur in an aqueous environment. As a consequence, biochemists, molecular biologists, and researchers in related disciplines conduct their experiments primarily in aqueous solutions. A number of options are available for properly handling difficult liquids. In addition to special pipetting techniques (e.g. reverse pipetting), specialized pipette tips may be employed (e.g. epT.I.P.S.® LoRetention or ep Dualfilter T.I.P.S.®). Difficult liquids: not the exception However, liquids are as different as the people who work with them. As soon as their physical properties deviate significantly from water or simple salt solutions, liquids pose a special challenge. Multipette/Repeater E3 and E3x: innovative experts for difficult liquids and long dispensing series For example, viscous or especially dense solutions are difficult to pipette using air-cushion pipettes. Volatile liquids, on the other hand, will drip from the pipette tip due to their high vapor pressure. Solutions that contain detergents or proteins tend to foam, which bears the risk of sample contamination. Finally, even cold or warm liquids (relative to ambient temperature) may be classified as difficult. Due to the difference in temperature, the air-cushion above the liquid will expand or contract, which can lead to systematic pipetting errors. Furthermore, the formation of aerosols by warm liquids may contribute to corrosion of the pipette. Difficult liquids are more common than generally appreciated, and they play an important role in a number of different areas, for example in diagnostic laboratories as well as in the chemical and cosmetics industries. Common difficult liquids Property Examples High viscosity Glycerol, blood, mineral oil Volatile Ethanol, acetone, chloroform High density Sulfuric acid, cesium chloride solutions Foaming Tween™ 20, Triton™ X-100, protein solutions Cold Enzyme solutions, cooled buffers Warm Cell culture medium The safest solution, however, is using a positive-displacement system which dispenses the liquid directly, without the need for an air-cushion. For this purpose, Eppendorf has developed the Multipette (Repeater)/Combitips® system. All Multipettes/Repeaters enable precise and fast dispensing of a broad range of liquids. The Multipettes/Repeaters E3 and E3x (with their different dispensing modes) offer the additional benefits of an electronic instrument. They make a large number of liquid handling applications accessible and support increased throughput. This way, working with difficult liquids becomes really easy. More information Additional information is available in the product brochure or at www.eppendorf.com/ multipette-system. Multipette ® E3 and E3x (U.S./Can: Repeater ® E3 and E3x) • Ref. no. 290 > PROPER HANDLING OF CELLS IN THE LAB (I) · STRAIGHT FROM THE LAB 7 TANJA MUSIOL, EPPENDORF AG Proper Handling of Cells in the Lab (I) Cell culture work, which includes seeding, cultivation, and analysis, holds a Close-up ConvexAccess™: Improved Handling, Increased Safety Cell culture flasks are the preferred special place in the daily laboratory routine. Whenever cells are concerned, solution for routine applications involving experimental conditions are as different as the cells themselves. long-term cultivation or large-scale expansion of cells. Optimal protection of cells from contamination was one of the key requirements in developing the Eppendorf Cell Culture Flasks. To meet these demands, Eppendorf not only developed an exceptionally efficient > Sources of contamination Prevention of contamination No matter whether primary cells, stem cells, or cell lines are the focus of one’s work, the demands on cell handling can be quite diverse, depending on the specific cell type as well as the nature of the experiment. The multitude of possible contamination sources points out the limited usefulness of antibiotics added to the culture. While a large proportion of bacteria may be combatted in this manner, contamination of a different nature may be not. This fact harbors the additional threat of undetected low level contaminations with resistant strains which are difficult to detect. Nonetheless, the contamination persists within the culture and therefore poses a danger to the entire cell culture laboratory. In order to prevent carry-over contamination to other cultures, it is imperative that equipment be thoroughly cleaned, or exchanged, after each cell line. To summarize, sterility is still the overarching principle wherever cell culture is concerned. That being said, a few general facts need to be considered when working with cells, most importantly those concerning the prevention of contaminations. However, there are obviously more sources of contamination than spring to mind. Not only bacteria, but also mycoplasmas, yeasts, and cells from other cultures pose a potential contamination threat, where especially the latter often result from carry-over contamination. More information Basic tips on contamination prevention are available at www.eppendorf.com/cellbiology. In addition, we recommend the webinar “Preventing Contamination in Cell Culture Labs” (which may be found at www.eppendorf.com/webinar in the category “Recorded”). air filter technology. We also created the ConvexAccess-shape of the flask neck which is unique on the market. It greatly facilitates the access to the entire growth surface. Cell seeding and media change have now become more comfortable, reliable, and safe. Product features Eppendorf Cell Culture Flasks >> ConvexAccess geometry for easier and safe cell treatment >> Optimized gas exchange and protection against contamination by high efficiency air filter technology >> Anti-rolling cap which can be stably positioned on its side facilitating cultivation steps and minimizing the contamination risk >> In-line pressure testing for each flask guarantees leakage-free flasks for maximum workflow safety >> Three sizes: T–25, T–75, and T–175, with TC-treated or non-treated surfaces, filter caps or plug-seal caps More information at ccc.eppendorf.com/flasks.php Eppendorf Cell Culture Consumables • Ref. no. 270 > 8 INNOVATION · CENTRIFUGE 5920 R: NEW BENCHMARK IN CAPACITY AND PERFORMANCE PETER SCHREINER, EPPENDORF AG Centrifuge 5920 R: New Benchmark in Capacity and Performance The new refrigerated Centrifuge 5920 R by Eppendorf sets new standards for benchtop centrifuges. With its tremendous capacity of up to 4 x 1,000 mL or 52 x 50 mL conical tubes it is the ideal instrument to handle highthroughput applications. This absolute eye-catcher was awarded the Good Design Award 2015 of the Chicago Athenaeum Museum of Architecture and Design for its compact and ergonomic design. temperature, and time only have to be saved once to make them easily accessible via program key. This saves time! Reliable temperature management > As awarded by The Chicago Athenaeum: Museum of Architecture and Design. The state-of-the-art refrigeration system with modern temperature management ensures that the selected temperature is maintained as accurately as possible and that temperaturesensitive samples are safely centrifuged. The FastTemp pro® function allows programming of date and time for automatic pre-cooling. Versatile and innovative With its extensive accessories-program for rotors, adapters, and aerosol-tight caps and lids, the Centrifuge 5920 R is exceptionally versatile. The new universal buckets are especially innovative as they hold both tubes and plates (Fig. 2). Fig. 1. Centrifuge 5920 R: similar dimensions to those of a regular refrigerated 3 liter centrifuge, but with a capacity of 4 liters Extraordinary capacity Customers with high-throughput applications will be delighted: The new refrigerated Centrifuge 5920 R (Fig. 1) offers almost double the tube capacity of the popular Centrifuge 5810 R. The capacity has been increased from 28 x 50 mL to 52 x 50 mL for conical tubes and from 92 x 5 mL to 196 x 5 mL for blood collection tubes. The maximum volume has increased as well. The 5920 R holds four 1 liter Nalgene® bottles, setting a new standard for benchtop centrifuges. User-friendliness and precision: the perfect combination The Centrifuge 5920 R is equipped with an advanced operating system featuring a user-friendly interface for intuitive and comfortable handling. At the same time, the operating system was designed to enable accurate and reproducible results. Five program keys allow fast and easy access to routine programs. For frequently used applications, parameters such as speed, Fig. 2: Swing-bucket rotor S-4xUniversal-Large with innovative universal buckets This previously required separate rotor buckets, and sometimes even separate rotors. With this new product solution, one rotor will suffice. This saves money! More information at www.eppendorf.com/together Centrifuge 5920 R • Ref. no. 289 > (BN 45) JUNE 2016 PAGE 1 Faster Isolation of PBMCs Using Ficoll-Paque® PLUS in the Eppendorf Centrifuge 5920 R NICOLE SEELIGMÜLLER, EPPENDORF AG, HAMBURG, GERMANY Introduction Human blood consists of equal parts of blood plasma and blood cells. These include erythrocytes (red blood cells), leukocyctes (white blood cells), and thrombocytes (platelets). Leukocytes are further subdivided into different cell types. > These include, for example, lymphocytes and monocytes, which (in co-operation with other cells) form the basis of the innate immune system and which, owing to their simple nucleus, are referred to as peripheral blood mononuclear cells (PBMCs). The term lymphocyte encompasses two major classes, B lymphocytes and T lymphocytes. B lymphocytes are responsible for antibody production, whereas T lymphocytes produce signal molecules which will finally lead to the removal of diseased or foreign cells. PBMCs play an important role in the areas of infection diagnostics as well as in clinical research [1]. PBMCs are isolated from buffy coats (whole blood concentrates without serum). Centrifugation using FicollPaque PLUS (density: 1.007 g/mL) allows separation of PBMCs from other blood components: erythrocytes, granulocytes, and dead cells which, due to their higher density, will pass through the Ficoll layer, whereas lymphocytes and monocytes will accumulate at the interphase between plasma and Ficoll gradient (Fig. 1), [2,3]. One prerequisite for a clean PBMC isolation with a maximum yield of living cells is the formation of a clear interphase. For this reason, the procedure must be carried out with the least vibration possible. Usually, mixing of the phases can only be avoided by centrifugation with the rotor brake deactivated [4], which constitutes an extremely time-consuming step within this application. This study shows that the new 4 liter benchtop Centrifuge 5920 R* not only satisfies the high demands of this method, but that thanks to the option of selecting individual ramps, the user will enjoy considerable time savings. Materials and methods Ficoll-Paque PLUS density gradient centrifugation 1.Fill Eppendorf Conical Tubes 50 mL with 15 mL Ficoll-Paque PLUS (GE Healthcare®) each. 2.Using the slowest pipetting speed available, overlay the Ficoll with 15 mL of the blood/PBS mixture. 3.Centrifuge the samples for 30 min in the swing-bucket rotor at 400 x g, 20 °C, at ramp** 0/0 or ramp 3/3, respectively, while selecting the setting “at set rpm”. **Possible ramp settings Centrifuge 5920 R: deactivated acceleration or deceleration, respectively (0/0) to fastest acceleration/brake (9/9) Purification of lymphocytes 1.Carefully remove 2/3 of the upper layer (plasma). 2.Aspirate as much of the PBMC layer as possible in the smallest volume possible and transfer to a fresh tube. 3.Wash PBMCs several times in PBS and resuspend in suitable cell culture medium. Viability test and determination of yield 1.Dilute cells 1:1 in trypan blue and count. 2. Determine viability and yield. Detailed information about materials and methods is available in Application Note 372 [5]. Results and discussion In order to evaluate the quality of separation, it was determined whether a defined interphase with clearly delineated phase transitions was visible. Fig. 2 (A–D) shows the results obtained from density gradient centrifugation carried out in the Centrifuge 5920 R using d ifferent swing-bucket rotors. For each rotor, the acceleration and deceleration rates 0/0 and 3/3 were tested. It is evident that optimal separation of blood components was achieved in all cases, and that vortex effects during deceleration of the rotor could be completely avoided (see negative example Fig. 2E). Plasma PBMC Ficoll Granulocytes Erythrocytes A B C D Fig. 2: Results obtained after density gradient centrifugation in the Centrifuge 5920 R Fig. 1: Density gradient obtained from buffy coat using Ficoll-Paque PLUS (schematic) A) Rotor S-4x1000 with high-capacity bucket (ramp 0/0) B) Rotor S-4x1000 with high-capacity bucket (ramp 3/3) C) Rotor S-4xUniversal-Large (ramp 0/0) D) Rotor S-4xUniversal-Large (ramp 3/3) E) Negative example: cloudy interphase, turbid plasma-/Ficoll phase (different centrifuge) Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com E > PAGE 2 (BN 45) JUNE 2016 Faster Isolation of PBMCs Using Ficoll-Paque® PLUS in the Eppendorf Centrifuge 5920 R Conclusion In order to obtain a high yield of viable PBMCs, Eppendorf recommends the use of the new Centrifuge 5920 R for density gradient centrifugation using Ficoll-Paque PLUS. According to available test results, consistent high quality of yield can be expected with all available swing-bucket rotors for 50 mL conical tubes. With its large capacity and the option of significantly reducing the time required for rotor deceleration, the Centrifuge 5920 R is also ideally suited for laboratories processing high sample quantities. *This centrifuge is an in vitro diagnostic accessory and therefore itself is an in vitro diagnostic device according to Directive 98/79/EC of the European Parliament and the Council dated October 27, 1998. > In addition to the visual inspection of the centrifuged samples, yield and viability of the PBMCs obtained from samples centrifuged in the rotor S-4xUniversal-Large were analyzed as an example to confirm the quality of the PBMC isolation using Ficoll-Paque PLUS. The data provided by GE Healthcare, which routinely achieved a viability of 95 % (+/−5 %) during internal testing, served as a reference [6]. The studies performed at Eppendorf showed an average viability of 94 %, which is in line with expected values. According to the literature, the expected yield of mononuclear cells falls between 0.8 and 3.2 x 106 cells/mL of blood [7]. A result of 3.0 x 106 cells/mL (ramp 0/0) or 2.16 x 106 cells/mL of buffy coat (ramp 3/3), respectively, places the total yield of viable cells within a very good range. These results demonstrate that the Centrifuge 5920 R fully meets the demands of low vibration performance, independent of the swing-bucket rotor used, and independent of whether an acceleration/deceleration rate of 0/0 or 3/3 was selected. Furthermore, a ramp of 3/3 allows considerable time savings of up to 19.7 min (36 %) compared to the centrifugation parameters recommended in the literature (deactivated brake) (Fig. 3) [8]. [2] Bøyum A. Isolation of mononuclear cells and granulocytes from human blood. Scan. J. Clin. Lab. Invest. 1968, 21 (Suppl. 97): 77-89. [3] Graham J. Biological Centrifugation: The Basics from Background to Bench. Oxford: BIOS Scientific Publishers Limited; 2001. [4] Heine H, Uschkureit T. Software controlled acceleration and braking ramp for optimized isolation of mononuclear cells. Eppendorf Application Note 074; www.eppendorf.com/applications [5] Seeligmüller N. Faster isolation of PBMCs using Ficoll-Paque® PLUS in the Eppendorf Centrifuge 5920 R. Eppendorf Application Note 372; www.eppendorf.com/applications [7] HIV/AIDS Network Coordination: CrossNetwork PBMC Processing Standard Operating Procedure. www.hanc.info 34.6 S-4x750 (0/0) [8] Luttman W, Bratke K, Küpper M, Myrtek D. Der Experimentator: Immunologie. 4. Auflage. Heidelberg: Springer Verlag; 2014. 54.3 S-4x1000 (3/3) [1] Theml H, Diem H, Haferlach H. Taschenatlas der Hämatologie: Morphologische und klinische Diagnostik für die Praxis. 5. Auflage. Stuttgart: Georg Thieme Verlag; 2002. [6] Ficoll-Paque® PLUS Instructions. www.gelifesciences.com Total run times swing-bucket rotors Centrifuge 5920 R (centrifugation time 30 min, ramps 0/0 and 3/3) S-4x750 (3/3) Literature 34.7 S-4x1000 (0/0) 54.1 S-4xU-Large (3/3) 33.4 S-4xU-Large (0/0) 47.36 0 10 20 30 40 50 60 Minutes Readers’ service Fig. 3: Total centrifugation times for density gradient centrifugation in the new Eppendorf Centrifuge 5920 R while using different swing-bucket rotors and ramp settings, 0/0 or 3/3, respectively Centrifuge 5920 R • Ref. no. 289 Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com > (BN 45) JUNE 2016 PAGE 3 Faster and More Reproducible Cell Viability Assays with the Eppendorf epMotion ® 96 ERIC GANCAREK, MAUD BRASSEUR, MURIEL ART, EPPENDORF APPLICATION TECHNOLOGIES S.A., NAMUR, BELGIUM HANAË A. HENKE, EPPENDORF AG, HAMBURG, GERMANY Abstract Cell-based assays are an established standard tool in research. The most common formats for these assays are 96- and 384-well plates, requiring multichannel pipetting to ensure quick, efficient, and reproducible handling. In this Application Note, a cell viability assay was used to determine the speed, reliability, and reproducibility of a manual 12-channel pipette, an electronic 12-channel pipette, and the semi-automated 96-channel electronic pipette epMotion 96. Compared to a manual 12-channel pipette it was shown that the epMotion 96 and the electronic 12-channel pipette delivered more reproducible results, with the epMotion 96 being significantly faster. Introduction Cell-based assays are an important method in life science and pharmaceutical laboratories. They allow limiting the use of animal models and providing more representative information generated in a biological context like a normal physiological situation [1]. A large panel of cell responses can be tackled by this method such as cell proliferation, activation of specific signaling pathways, cytotoxic impact of a molecule, or cell viability. This could be solved by automated systems, but these can be an investment smaller labs cannot afford. The epMotion 96, a semi-automated pipette equipped with 96 channels, is an alternative solution for scientists interested in increasing their throughput. This study clearly illustrates the benefits of using the epMotion 96 for cell-based assays. Method: cell viability assay Cells seeded into 96-well Eppendorf Cell Imaging Plates are grown over night and then treated with eleven increasing concentrations of Staurosporine (STS) between 1 nM and 20 µM to induce cell death. The DMSO concentration does not exceed 1 %. The plates are incubated for 48 h in presence of the cytotoxic agent. After incubation, 20 µL of freshly prepared CellTiter-Blue® Cell Viability reagent (Promega®) are added to each well of the 96-well plates. The CellTiter-Blue reagent contains highly purified resazurin. Viable cells retain the ability to reduce resazurin into highly fluorescent resorufin. Nonviable cells rapidly lose metabolic capacity, and thus do not generate a fluorescent signal. Fluorescence is read in each well at two wavelengths (excitation at 535 nm; emission at 595 nm). epMotion® 96 Results and discussion Impact of liquid handling instrument on cell-based assay performances In the cell-based assay chosen for this study, volumes dispensed were between 10 µL and 90 µL. Moreover, solutions handled vary from a cell suspension to a cytotoxic agent dissolved in DMSO. The epMotion 96 was compared to a 12-channel electronic pipette and a 12-channel manual pipette for each of the following experiments. Dose-dependent viability curves The epMotion 96, a 12-channel electronic pipette, and a 12-channel manual pipette were used to seed HepG2 cells into 96-well plates (three plates in parallel) and to add the assay reagent to all wells. The HepG2 cells viability was evaluated after treatment by a STS dose-response curve. An electronic 12-channel pipette was used to add eight replicates of this curve to the nine microplates. A widely distributed technique is to remove the data generated from curves displayed in rows A and H in all plates, so six replicates were used for calculation and are shown in Fig. 1. Profiles generated were comparable for each instrument used for cell seeding and reagent adding. Nevertheless, with a manual pipetting system, the assay reproducibility was affected as indicated by higher standard deviations. Manual pipette Electronic pipette 45,000 45,000 40,000 40,000 40,000 35,000 35,000 35,000 30,000 25,000 20,000 15,000 10,000 5,000 0 -10 Fluorescence (RFU) 45,000 Fluorescence (RFU) Fluorescence (RFU) > Taken together, the epMotion 96 is the ideal tool for cell-based assays in plates, including cell seeding, media change, and compound and assay reagent addition. Compatibility of the liquid handling tool with the cell-based assay has to be demonstrated and assay reproducibility is an essential factor which has to be evaluated [2]. 30,000 25,000 20,000 15,000 10,000 -9 -8 -7 -6 -5 log10 (Staurosporine) M -4 25,000 20,000 15,000 10,000 5,000 5,000 0 -10 30,000 -9 -8 -7 -6 -5 log10 (Staurosporine) M -4 0 -10 Fig. 1: Dose-dependent STS toxic effect on HepG2 cells (n=six replicates) Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com -9 -8 -7 -6 -5 log10 (Staurosporine) M -4 > PAGE 4 (BN 45) JUNE 2016 Faster and More Reproducible Cell Viability Assays with the Eppendorf epMotion ® 96 Intra-plate CV (n = 96) Global intra-plate CV (n = 96 x 3) Global inter-plate CV (n = 96 x 3) Plate 1 Plate 2 Plate 3 epMotion® 96 9.4 7.9 8.8 8.7 9.8 Electronic pipette, 12 channels 7.2 8.4 10.1 8.6 10.4 Manual pipette, 12 channels 13.1 15.9 16.8 15.3 17.9 Table 1: Intra- and inter-plate CVs calculated for CellTiter-Blue cell viability assay for the respective instruments used (three replicates) Intra- and inter-plates reproducibility > Time needed to fill one assay plate with different liquid handling tools 20 18 epMotion® 96 Electronic pipette, 12 channels 16 Time in minutes The epMotion 96 was used for all three successive assay steps (cell seeding, cell treatment by a cytotoxic agent, and assay reagent addition). A unique STS concentration (1,000 nM) was applied on HepG2 cells to induce cell death. The assay reproducibility was assessed by calculating intra-plate and inter-plate coefficients of variation (CV) in between the three replicates presented in Table 1. Manual pipette, 12 channels 14 12 10 > The global intra-plate CVs and the interplate CVs generated by the epMotion 96 and the electronic pipette were comparable and significantly lower than CVs obtained with a manual instrument. These instruments allow almost complete elimination of human error as one of the major sources of variability. Besides the assay reproducibility improvement, the use of an electronic instrument also allows to reduce the hands-on time as indicated in Fig. 2. Fig. 2: Hands-on time needed for one assay plate for each CellTiter-Blue Cell Viability assay step depending on the instrument used The epMotion 96 appears as the fastest option since a complete cell-based assay requires less than four minutes, while with the electronic pipette one needs approx. eight minutes and with the manual pipette almost 20 minutes. medium to high throughput. This instrument allows saving hands-on time and combines the ease of use of pipettes with a high precision at a price more affordable than fully automated systems. Conclusion Literature In this Application Note, we show the capability to use the epMotion 96 for processing a cell-based assay. With all assay performance parameters studied, it was demonstrated that the assay reproducibility was significantly better when using the epMotion 96 or the electronic pipette. When a large number of tests have to be handled in parallel, the epMotion 96 can represent a good choice for scientists interested in [1] Zang R, Li D, Tang I-C, Wang J, Yang S-T. Cell-based Assays in High-Throughput Screening for Drug Discovery. International journal of Biotechnology for Industries 2012; 1:31-51. 8 6 4 2 0 Cell seeding Cytotoxic agent addition Assay reagent addition [2] An W.F, Tolliday N. Cell-based Assays for High-Throughput Screening. Mol Biotechnol 2010; 45:180-186. Readers’ service epMotion® 96 • Ref. no. 275 Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com All assay steps (BN 45) JUNE 2016 PAGE 5 The Tip of the Iceberg: How Pipette Tips Influence Results. Part 2: Perfect Geometry Makes a Difference ULRIKE GAST, EPPENDORF AG, HAMBURG, GERMANY LAURA KOCH, EPPENDORF POLYMERE, HAMBURG, GERMANY Abstract We performed a study including standard tips from different manufacturers in order to investigate the tip-related influences on the pipetting result. The study showed a dramatic impact of the tip on the pipetting accuracy. Small volume pipetting accuracy is significantly influenced by the quality of the tip orifice. Any impairment of geometry and shape can lead to retention of liquid. Introduction Tip-related influencing factors on pipetting results can be compared to an iceberg. Some influencing factors are easy to observe, like the necessity for high tip attachment forces in order to achieve a secure and tight fit of non-recommended tips on the pipette cone. Other factors stay rather unknown, like the geometry of the tip orifice. round geometry lead to deflection of liquid drops to the outside of the thinner wall. Flashes cause liquid retention. Such production flaws are generated within tip production by tools with low quality or long maintenance cycles and/or a non-optimized injection molding process. Since the production obviously has a high impact on the geometry and shape of the tip, influencing the pipetting result, it makes sense to take a closer look at production. The next part of this series will focus on the production of pipette tips and, in this context, the use of additives. (BN 44) JANUARY 2016 PAGE 1 The Tip of the Iceberg: How Pipette Tips Influence Results. Part 1: Tip Fit Is Not All Users Should Look for MURIEL ART, VINCENT DUFEY, IOAN GLIGOR, EPPENDORF APPLICATION TECHNOLOGIES S.A., NAMUR, BELGIUM ULRIKE GAST, RONJA KUBASCH, EPPENDORF AG, HAMBURG, GERMANY Part 1 of this series of Application Notes described pipette and tip as a system which may not perform within permissible error tolerances with all tips [1]. Material and methods The fact that a tip fits onto a pipette cone does not say anything about the performance of the pipetting system comprising the components “Pipette and Tip”. We performed a study including standard tips from 15 different manufacturers in order to investigate the tip-related influence on the pipetting result. The study results showed a dramatic influence of the tip on the pipetting accuracy. order to achieve efficient tip fit. Other problems often remain unrecognized like decreased pipette accuracy when using other tips than recommended by the pipette supplier. The ISO 8655-2:2002 standard [1] defines pipette and tip as a system, which requires extra calibration for the use of other manufacturers’ tips. But why does this standard put so much focus on a product that is to be discarded after usage? The standard ISO 8655 regards pipette and tip to be a system. Our study results emphasize the validity of this statement and the need of calibration / pipette adjustment if other tips than recommended by pipette manufacturer are to be used. This series of articles answers this question. It shows the influence of tips on the pipetting result explaining the main tip-related impact factors. Introduction Eppendorf Xplorer ® plus 50 –1,000 µL and 0.5 –10 µL were used. Racked 10 µL and 1,000 µL standard tips of Eppendorf and 14 other manufacturers have been tested. Exceptions: Manufacturer H did not offer racked 10 µL standard tips, manufacturers K and N offered only 1,250 µL tips for 1,000 µL pipettes. Within the scientific community, a rising number of published experiments cannot be reproduced by other groups. In general, plastics are not taken seriously leading to problems with analysis results caused by e.g. leachables or incorrect pipetting volumes. This may lead to results not being reproducible if performed by other groups using other consumables. Some problems with pipette tips are obvious, e.g. the need to push tips with force onto the pipette cone in Material and methods General material Calibration by gravimetric method The performance of the system “Pipette and Tip” was determined by calibration according to [1]. Environmental conditions were set according to requirements [1]. Calibration was performed using analytical balance Model XP26PC (MettlerToledo®) at 100 % nominal volume and 10 % nominal volume. Two series of 10 pipettings were performed. Systematic error and random error were determined for each series of 10 measurements and compared to specifications [1] and [2]. For further information please refer to [3]. Results and discussion While being perfectly within error limits with Eppendorf tips, we found the system “Pipette and Tip” to be out of specifications when using other manufacturers’ tips. As shown in Fig. 1 and 2, the allowed systematic error was exceeded with 4 manufacturers’ tips at a volume of 1,000 µL and with 5 manufacturers’ tips at 1 µL test volume. These tips with 1,000 µL test volume exceeded not only the manufacturer specifications but also the wider limits for systematic error as stated by the ISO 8655:2002 standard [1]. The random error was noticeably increased but stayed within allowed tolerances. When comparing the calibration results with the outcome of dimensional measurements, it becomes clear that with 1,000 µL the biggest impact factor is the air-cushion size: Those tips that produced error limits beyond the pipette 1 µL 0.5 0.45 0.4 1.5 1 0.5 0.35 0.3 0.25 0.2 0.15 0.1 0.05 0 Results and discussion 10 µL 2 Random error [%] The orifice microscopy of 10 µL tips from Eppendorf and three other suppliers (manufacturers E, F, and H) was performed using a microscope (Leica®) with 25-fold magnification and a DFC 280 digital camera (Leica). Random error [%] > This is in concert with ISO 8655 [2] which declares that pipette and tip build a system which needs to be calibrated if alternative tips are used. In part 1, the air-cushion size was detected to be the main influencing factor on bigger volumes like 1,000 µL. In contrast, with 10 µL tips other f actors come into play. Here, the quality of the tip’s orifice was found to be of high importance. For further investigation, the orifices of tips from different manufacturers were scrutinized and compared to calibration results [1,3]. Abstract -4 -2 0 2 4 6 8 10 0 -1.5 Systematic error [%] -1 -0.5 0 0.5 1 1.5 Systematic error [%] Fig. 1: Calibration results using 10 µL tips of different manufacturers. The red shaded area shows the span of the maximum permissible errors stated for the system “Pipette and Tip” by pipette supplier. All data points within the red shaded area are within the specifications. The zone where liquid leaves the tip during dispensing is very important for the accuracy of results. At this part of the tip, the drop cut-off occurs. Any imperfection of geometry or shape, e.g. by production errors, can lead to liquid retention. This especially plays a role with small volumes. A poor drop cut-off may not only impair the pipetting result but can make it impossible to dispense small volumes below 1 µL onto a solid surface. Fig. 1 (see next page) shows examples of such production flaws. Tips from these manufacturers lead to the system performing outside permissible systematic error tolerances [1,3]. This underlines the direct influence of tip orifice geometry on pipetting of small volumes. A good tip has a perfectly round geometry and uniform wall thickness. The orifice does not show flashes or loose material residues. Walls with varying thickness and not perfectly Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · e-mail: [email protected] · www.eppendorf.com Missed part 1? A rising number of published experiments cannot be reproduced by other groups. In general, only little attention is paid to the purchase of plastic laboratory consumables, like pipette tips and microtest tubes. The fact that a pipette tip fits onto a pipette cone does not say anything about the pipetting accuracy of the pipetting system comprising the components “Pipette and Tip”. We have studied this issue thoroughly in part 1 of our series “The Tip of the Iceberg: How Pipette Tips Influence Results”. If you missed part 1, then check out BioNews No. 44 in our archive at www.eppendorf.com/bionews. Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com > PAGE 6 (BN 45) JUNE 2016 The Tip of the Iceberg: How Pipette Tips Influence Results. Part 2: Perfect Geometry Makes a Difference Lying flash Flash Eppendorf epT.I.P.S.® 10 µL Manufacturer E The orifice has a good geometry and the function is not negatively influenced by production errors. Problem 1: Lying flashes caused by non-harmonized ejection molding process. Cavity has not been fully filled with liquid PP. Result: Risk of deflection of water drop because of varying diameter of frontal area Problem 2: Flashes at exterior wall Result: Risk of keeping liquid residues Molding flash > Flash Molding flash Manufacturer F Manufacturer H Problem 1: Flashes at interior and exterior wall caused by long maintenance cycles of tool. Problem 1: Molding flashes caused by a non-tightening tool or problems with injection molding process where too much liquid PP has been injected. Result: Risk of liquid residues and risk of PP particles falling into sample. Additionally, particles inside the tip displace water leading to wrong liquid volume. Such an error pattern makes it impossible to dispense e.g. 0.8 µL sample onto a solid surface. Result: Liquid retention Problem 2: Noticeable core shift. Not all walls have the same thickness. This error pattern can be caused by e.g. a poorly manufactured tool. Result: Risk of deflection of liquid beside instead of into the target vessel. Fig. 1: Microscopic pictures of 10 µL tip orifices of different manufacturers. Manufacturer E, F, and H failed the calibration by impaired systematic error. The examples have been chosen to explain production errors. PP = polypropylene. Literature [1] The Tip of the Iceberg: How Pipette Tips Influence Results. Part 1: Tip Fit Is Not All Users Should Look for. BioNews 44, January 2016. www.eppendorf.com/bionews [2] DIN EN ISO 8655:2002. Piston-operated volumetric apparatus. Beuth-Verlag, Berlin, Germany [3] Application Note 354: The Tip of the Iceberg: How Pipette Tips Influence Results. www.eppendorf.com/applications Readers’ service Eppendorf Liquid Handling Consumables • Ref. no. 288 Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com > (BN 45) JUNE 2016 PAGE 7 Improved Automated Single Cell Clone Analysis with the SynenTec Cellavista® System and Eppendorf Cell Culture Plates MAREIKE PANZ, JOHANNA PRIEB, SYNENTEC GMBH, ELMSHORN, GERMANY INES HARTMANN, EPPENDORF AG, HAMBURG, GERMANY Abstract Single cell cloning is one critical step in stable cell line development. To ensure the monoclonal character of the population, a reliable microscopic identification of single cells is essential. Automated cell analyzers like the SynenTec Cellavista System (Fig. 1) allow a high throughput and a reliable tracking and documentation from single cell status to colony formation. the method it is critical to verify that the resulting colony derives from a single cell. Manual microscopic observation of single cell clones and colony formation can be time-consuming, unreliable, and difficult to document. Automated cell analyzers like the SynenTec Cellavista System can monitor cell growth of colonies from single cell status by repeated measurements of the same microplate over a span of typically two to three weeks. The system is able to follow the growth of the clones and can completely document this process. Here we analyze the influence of the consumable in single cell cloning analysis. A uniform well illumination without interfering shadows at the well edges is important to detect cell colonies growing at the well periphery. The choice of the consumable can affect the analysis as well: Evaporation and meniscus effects can lead to optical interferences with the result that single cell clones may stay undetected or be falsely analyzed. In Eppendorf Cell Culture Plates (Fig. 2), the liquid meniscus is minimized resulting in an excellent well illumination. Introduction Stable transfected cell lines arisen from one single cell clone allow continuous gene expression by a genetically homogeneous population. Independent from Fig. 2: The Eppendorf 96-Well Cell Culture Plate is equipped with an outer moat that can be filled with liquid to insulate specifically the edge wells to minimize evaporation. The inter-well space and the outer moat of the Eppendorf Cell Culture Plates were filled with sterile, distilled water to minimize evaporation of medium over time. Plates were kept in the incubator until cell seeding. Competitor plates were placed in the incubator at the same time prior to seeding. Adherent CHO-GH3 cells were prepared and seeded in a concentration of 0.5 cells per well in 200 μL medium. Day 0 Images (n=4) were taken on day 0 and day 1 followed by measurements every 2 – 3 days until day 10 –14 post-seeding. In order to prevent cell loss no change of medium was performed. Meniscus effect Plates were seeded with different volumes of medium from 1–6 μL/mm2. Volumes were adapted to volume/surface area to cover variances in growth areas of plates. In a second experiment cells were seeded using 3.31 μL/mm2 and stained with trypan blue. All plates were analyzed using the Cellavista System. Edge effect Single cell seeding procedure Fig. 1: The automated cell analyzer SynenTec Cellavista System Edge effect Results and discussion Materials and methods > Analysis Day 1 Evaporation in 96-well plates is a critical factor. With the Eppendorf Cell Culture Plates evaporation can be reduced to a minimum [1]. Fig. 3 shows that filling the moat and the inter-well space of the Eppendorf plate leads to clear well illumination throughout the entire duration of incubation. Colony growth in the edge wells of the plate can be detected clearly throughout the whole incubation time. Day 6 Day 14 Fig. 3: Well brightness and illumination uniformity. Outer well 6A of the Eppendorf 96-Well Cell Culture Plate was measured at indicated time points using the Cellavista System. Filling of the outer moat minimizes evaporation and enables optimal illumination of the well also after 14 days of incubation. The complete colony is clearly visible even in the well edge. (Upper panel: 4x objective, brightfield; lower panel: image detail) Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com > PAGE 8 (BN 45) JUNE 2016 Improved Automated Single Cell Clone Analysis with the SynenTec Cellavista® System and Eppendorf Cell Culture Plates Competitor Eppendorf 1 µL/mm2, BF 1 2 µL/mm2, BF 1 3 µL/mm2, BF 1 4 µL/mm2, BF 2 5 µL/mm2, BF 3 6 µL/mm2, BF 3 Fig. 4: Comparison of meniscus effect in Eppendorf Cell Culture Plates (96-Well, TC-treated) and competitor plates. The lower the volume the more critical is the dark shadow formation. Volumes were adapted to volume/surface area (Cellavista, 2x objective, brightfield adjustment: BF1 = 8 ms, BF2 =12 ms, BF3 = 20 ms). Competitor Eppendorf > No shadow visible Shadow visible Fig. 5: Comparison of meniscus effect with trypan blue stained cells. No shadow at the edge is visible in the Eppendorf Cell Culture Plate. Volumes were adapted to volume/surface area (Cellavista, 2x objective, brightfield). Meniscus effect Conclusion Literature A common problem when imaging 96well formats is the meniscus formation of the liquid. The curvature of the liquid causes a refraction of the light with the result of uneven illumination of the sample which may lead to a distorted image as if looking through a porthole. To analyze the meniscus effect, different liquid filling heights were measured in the tested plates. Single cell cloning using SynenTec’s automated Cellavista analyzing system together with the Eppendorf Cell Culture Plates is best suited for successful single cell clone analysis. The Cellavista can monitor cell growth of colonies from single cell status by repeated measurement of the same microplate over a defined span. By this, documentation of colony development is made easy. Eppendorf Cell Culture Plates have an optimized microscopic performance: The reduced meniscus enables a uniform illumination without disturbing shadows at the edges. Due to the possibility to reduce evaporation in the plate to a minimum, Eppendorf Cell Culture Plates enable an improved microscopic performance also in long-term incubation experiments. [1] Wagener J, Plennevaux C. Eppendorf 96Well Cell Culture Plate – A simple method of minimizing the edge effect in cell-based assays. Eppendorf Application Note 326; www.eppendorf.com/applications Fig. 4 shows that the Eppendorf plate allows an excellent illumination of the whole well area also at reduced filling heights. In comparison to that the competitor plate already shows disturbing shadows and uneven illumination when using higher filling levels. In Fig. 5 trypan blue stained cells were analyzed in Eppendorf plates and comparable plates from a competitor using the Cellavista Imaging System. An interfering shadow can be observed in the competitor plate at the edge of the well whereas in the Eppendorf plate the shadow is reduced to a minimum. Readers’ service Eppendorf Cell Culture Consumables • Ref. no. 270 Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany · E-mail: [email protected] · www.eppendorf.com > STATE-OF-THE-ART DNA MEASUREMENT: EFFICIENT, FAST, AND REPRODUCIBLE · STRAIGHT FROM THE LAB 9 SOPHIE MANUELLO, EPPENDORF AG State-of-the-Art DNA Measurement: Efficient, Fast, and Reproducible Molecular biologists consider quantification of dsDNA an essential routine technique. Measurement data obtained are of great importance for subsequent experiments. In a worst case scenario, faulty underlying data may lead to the failure of downstream applications. The Eppendorf BioSpectrometer® fluorescence combines UV/Vis and fluorescence measurements in a single device to reliably determine both concentration and purity of your DNA samples. > Why not quantify using UV/Vis spectroscopy? What does fluorescence have to do with it? Is fluorescence the only method required for quantification? Increasing discussion within scientific forums points to the fact that UV/Vis spectroscopy overestimates dsDNA concentrations and that especially small amounts are not quantified correctly. The reason is that all nucleic acids absorb light at 260 nm; therefore no general distinctions can be made between dsDNA, ssDNA, and RNA. Certain fluorophores can only fluoresce if bound to a specific molecule. Fluorescence kits take advantage of the binding specificity to dsDNA, ssDNA, or RNA and thus create a direct correlation between signal and concentration. Owing to the high binding affinity to the target molecule, quantification via fluorescence is also considerably more sensitive. In addition to the exact concentration of dsDNA, certain downstream applications require a statement on sample purity. Fluorescence measurements do not provide this information. Sample purity may be determined by UV/Vis spectroscopy using the ratios A260/A280 and A260/ A230 as well as the shape of the spectrum. The combination makes the difference! UV/Vis spectroscopy and fluorescence are methods which complement each other. The Eppendorf BioSpectrometer fluorescence is the first instrument on the market to combine both methods of measurement in one small, compact device. In combination with the microvolume cuvette Eppendorf µCuvette® G 1.0, the field of application for dsDNA measurement is now completed for a range from 1 pg/µL to 1,500 ng/µL. Several videos on the topic of “Detection” are available at www.eppendorf.com/ playlist-detection or by QR-Code. Eppendorf BioSpectrometer ® • Ref. no. 242 > 10 STRAIGHT FROM THE LAB · ROBUST AND RELIABLE: EASYPET® 3 HANAË A. HENKE, EPPENDORF AG Robust and Reliable: Easypet® 3 Tip Supporting You Maintenance and Certification Services Preventive Maintenance is a schedule of planned service measures aimed at the When conducting sterile work in cell culture, microbiology, or diagnos- prevention of unexpected instrument tics, the use of cleaning and disinfection agents cannot be avoided. For downtimes and failures through routine this reason, we have tested our electronic pipette controller Easypet 3 for problems. resistance to 67 of the most common laboratory chemicals. Certification Services include calibration, maintenance and early detection of verification, and Installation & Operational Qualification (IQ/OQ) to ensure that your results are consistent and reliable. Furthermore we offer support to ensure For clear distinction, materials were classified as either “resistant”, showing “limited resistance” or “increased risk/ wear”. > compliance with your laboratory guidelines by providing the necessary certification documents. > The test results In our test, all parts of the Easypet 3 which could come into direct contact with one of the test chemicals proved to be “resistant” to almost all chemicals tested. These test results offer you the safety you need when working with chemicals; for example when you use your Easypet 3 in a cell culture or molecular biology laboratory, as well as in the chemical or biotechnology industries. Test report for downloading In addition to commercial cleaning agents and disinfectants, this included saline solutions, buffers, acids and bases as well as organic solvents. The detailed report “Chemical Resistance Easypet® 3” may be downloaded as a PDF file at www.eppendorf.com/easypet under “Downloads/manuals”. When performed on a regular basis, our Performance Plans* are designed to keep your Liquid Handling, Sample Handling, and Cell Handling instruments in best condition over years and to meet manufacturer accuracy and precision specifications: >> ESSENTIAL CHECK >> ADVANCED MAINTENANCE >> PREMIUM SERVICE >> Installation Qualification (IQ) >> Operational Qualification (OQ) For more information and service requests go to www.eppendorf.com/epServices. Rigorous testing All individual parts of the Easypet 3 were tested separately. This approach provided certainty that all parts displayed comparable resistance to the chemicals tested. Individual parts are made from very different materials such as silicone (pipette adapter), polypropylene (aspiration and dispensing buttons, housing, wall holder, and shelf stand), or cyclic olefin copolymer (battery status display). Eppendorf Performance Plans Easypet ® 3 • Ref. no. 250 *Performance Plans are available in selected countries and services may vary. EPPENDORF TUBES® 5.0 ML – NOW ALSO AVAILABLE WITH SCREW CAP · INNOVATION 11 NILS GERKE, EPPENDORF AG Eppendorf Tubes® 5.0 mL – Now Also Available with Screw Cap With the development of the first Eppendorf Tubes over 50 years ago, Eppendorf ignited the evolution of the reaction tubes. Since that time, Eppendorf Tubes have been used worldwide on a daily basis for routine and special procedures. In order to further optimize applications in the life sciences, the Eppendorf Tubes 5.0 mL were developed, thus closing the gap between existing tube size variants. When working in the medium volume range, with this new tube at hand, it is no longer necessary to handle large tube formats which can often be impractical, inconvenient, and prone to contamination. With system Naturally, all these advantages can only be enjoyed if seamless integration into the existing laboratory environment, with all its established laboratory protocols, is possible. This is achieved by the carefully thought-out Eppendorf 5.0 mL system. The comprehensive solution with matching accessories allow all common laboratory workflows to be carried out in an easy, practical, and cost-efficient manner. > Eppendorf offers custom system components for efficient and easy pipetting, gentle mixing, fast centrifugation, incubation, automated liquid handling as well as secure storage. Furthermore, thanks to its conical shape, the Eppendorf Tube 5.0 mL is compatible with the accessories for the common 15 mL tubes, i.e. many adapters and racks can be utilized. Eppendorf Tubes 5.0 mL: with snap cap or screw cap for individual application demands in the medium volume range With screw cap The introduction of the screw cap tube represents the next major innovation within the Eppendorf Tubes 5.0 mL system. In addition to utmost seal security, the grooved, multi-surface sides of the newly designed screw cap ensure a slip-free grip. These tubes complete the application spectrum for sample processing in the medium volume range. They are especially suitable for applications where a screw cap can be beneficial to either experiment or handling: >> Long-term storage and archiving of samples and solutions at temperatures as low as −86 °C >> Secure sample transport >> Incubation at high temperatures up to 100 °C In addition they are well suited for sample preparation requiring centrifugation – the new Eppendorf Tubes 5.0 mL with screw cap feature the same centrifugation stability of up to 25,000 x g as the 5.0 mL snap cap tubes. Now the user can choose the ideal tube for each individual application and thus further optimize laboratory protocols. The 5.0 mL system is perfectly adapted to the Eppendorf Tubes 5.0 mL More information at www.eppendorf.com/5mL Eppendorf Tubes® 5.0 mL • Ref. no. 264 > 12 STRAIGHT FROM THE LAB · HOLISTIC APPROACH TO ERGONOMICS TANJA MUSIOL AND JAN-HENDRIK BEBERMEIER, EPPENDORF AG Holistic Approach to Ergonomics Tip Eppendorf Services Our online Service & Support portal www.eppendorf.com/epServices is not only THE gate to Eppendorf’s comprehen- Since the early 1970s, further to considering the technical specifications of its products, Eppendorf has also incorporated their ergonomic properties within their respective work environments. In 2003, the holistically de- sive service offerings. It also features a great deal of additional useful information. Application Support Our highly trained Scientific Support Staff signed PhysioCare Concept® for Liquid Handling products was launched is available to assist you with any questions and brought with it a fresh perspective on daily laboratory processes. concerning your Eppendorf products. Seminars and Training Find the latest information on seminars and webinars conducted by experienced specialists. Our courses have a practical orientation and will help you master your daily lab routine challenges. Technical Support > For maintenance, certification and repair service, please contact your local service organization here. Maintenance and Certification Services Learn more about the Eppendorf Performance Plans. The information is sorted by product groups and structured in a clear and concise manner. For conveniently Since that time, the PhysioCare Concept has seen continuous development as well as expansion to other product groups. Today it links the three spheres of action in the lab: the user, the laboratory, and the laboratory workflow, with the goal of harmonizing the workflow in the laboratory with the health and well-being of the staff. Ergonomics encompasses more aspects than meets the eye. For example, with manual Eppendorf pipettes, probably the most well-known products with PhysioCare Concept features, the focus is on the pipetting force: The smaller the force needed for pipetting, the less strain will be imposed on the muscles, tendons, and joints. In Eppendorf laboratory instruments, on the other hand, features like easy-to-read displays, simple operation, and intuitive menu navigation as well as low noise levels contribute to the overall ergonomics. One further crucial component of ergonomics is laboratory organization: How do I prevent repetitive stressful movements, unnecessary walking, or frequent interruptions of my work? How do I optimize the organization of my work space? Eppendorf, with its decades of experience, will offer useful ideas that are easily put into practice. Conclusion The PhysioCare Concept offers holistic solutions for optimizing the workflow in the laboratory. Ergonomic product features are complemented by recommendations on how to equip and organize ergonomic workflows. The result: an improved work environment, paired with higher efficiency. More information at www.eppendorf.com/physiocare getting in touch with the Eppendorf Service Team, you’ll find easy to use contact forms. Quality and Regulatory Affairs Your source for certificates, material safety data sheets, and information on purity grades of Eppendorf consumables. Knowledge Base The Knowledge Base is a true treasure box! Download all kinds of literature, such as operating manuals, application notes, protocols, posters, user guides, white papers, just to name a few. Register or update Eppendorf software. Check out our FAQ database. Visit the Media Center to watch the latest videos from Eppendorf, or download our mobile apps. > EPPENDORF STERILIZE-IN-PLACE (SIP)BIOPROCESS SYSTEMS · STRAIGHT FROM THE LAB 13 SOPHEAP SUN, EPPENDORF, INC., ENFIELD, USA ULRIKE BECKEN, EPPENDORF AG, BIOPROCESS CENTER EUROPE, JUELICH, GERMANY Eppendorf Sterilize-in-Place (SIP) Bioprocess Systems Small footprint Critical to the success of industrial bioproduction, the scale-up of bioprocesses is currently receiving much attention. Eppendorf offers comprehensive fermentor and bioreactor portfolios for pilot through production applications. Lab space is precious. That is why Eppendorf SIP systems were constructed with a small footprint. At just 116 cm wide x 86 cm deep (45.5 x 34.0 in), the compact BioFlo 510 can fit on a lab bench. Also the other SIP systems are designed to make efficient use of space. Bioprocessing is indispensable for manifold industrial applications, like the production of recombinant therapeutic proteins in mammalian cells, the conversion of biomass into biofuels by bacteria or fungi, vaccine production, the fermentative production of building blocks for bioplastics, and various others. > Flexible design Often process requirements change over time. Eppendorf SIP systems provide much needed flexibility. A large number of ports in the vessel head plate and sidewall allow adjustments. The systems facilitate bioprocessing in batch, fedbatch, and continuous mode. The optimization of strains and cell lines, culture conditions, and process parameters is usually carried out in small volumes. But once a process is established, process development scientists will scale up for production. With its comprehensive portfolio of sterilize-inplace (SIP) fermentors and bioreactors, Eppendorf bridges the gap between process development in the milliliter scale and pilot/production manufacturing. The customer can choose from various impeller options optimized, for example, for fermentation, the cultivation of shearsensitive cells, or cell culture on microcarriers. Many of the options can also be field-upgraded if the customer needs to make a configuration change. Fig. 1: The Eppendorf New Brunswick BioFlo Pro A variety of vessels altogether covers a volume range of 5.5 L to 2,400 L (Table 1). Results gained by the Eppendorf bio process application team as well as data gathered by our customers demonstrate the excellent scale-up capabilities of these systems. Our SIP systems portfolio comprises the New Brunswick™ BioFlo® 510, 610, and Pro fermentors for microbial applications and the New Brunswick CelliGen® 510 and Pro bioreactors for cell culture (Fig. 1, Table 1). BioFlo 510 BioFlo 610 BioFlo Pro CelliGen 510 CelliGen Pro 5.5 – 32 L 13 – 100 L 32 – 2,400 L 5.5 – 32 L 18.8 – 520 L Bacteria/yeasts/fungi • • • Plant cells/algae • • • • • Mammalian cells • • Stem cells • • • • Working volume Insect cells • Table 1: Eppendorf SIP bioprocess systems • • Reliable function Industrial processes demand reliable fermentor and bioreactor systems. Eppendorf strives to fulfill this expectation by delivering robust devices to minimize maintenance and downtime and by applying rigorous quality control. Eppendorf service begins with short delivery times and comprises training and ongoing support to enable the user to make the most of their equipment. Brochure “We Know Bioprocessing” • Ref. no. 274 > 14 NEWS · EPPENDORF PRIZE WINNERS 2015/2016 SHIGEKI WATANABE & ADRIAN LISTON CAROLYN TAUBERT AND BERRIT HOFF, EPPENDORF AG Eppendorf Prize Winners 2015/2016 Shigeki Watanabe & Adrian Liston > Congratulations to the new winners of the Eppendorf research prizes! The Japanese scientist Shigeki Watanabe, Ph.D., (Johns Hopkins University, Baltimore, USA) won the 2015 Eppendorf & Science Prize for Neurobiology endowed with 25,000 US$ for his work on vesicle endocytosis. Shigeki Watanabe has developed two novel techniques in electron microscopy that allow visualization of protein and membrane dynamics at synapses at a millisecond temporal resolution. Using these techniques, he has discovered an ultrafast mechanism that removes synaptic vesicle components rapidly from the plasma membrane following exocytosis and further demonstrated that synaptic vesicles are regenerated from endosomes. Shigeki Watanabe has started his own lab where his research will focus on understanding the mechanisms underlying the rapid reorganization of membrane, proteins, and organelles that mediate synaptic plasticity. More information at www.eppendorf.com/prize The 2016 Eppendorf Award for Young European Investigators went to Prof. Adrian Liston, Group leader at VIB Translational Immunology Lab, University of Leuven, Belgium, for his seminal work in elucidating key mechanisms by which the immune system avoids attacking its own organism while remaining effective against pathogens. His experiments have paved the way for understanding key steps in controlling regulatory T-cells that are critical for balancing between autoimmunity and immunosuppression. His work opens up the way for new therapeutic approaches towards diseases resulting from a dysregulated immune homeostasis. In the Translational Immunology Lab, Adrian Liston and his team work on both the discovery (Treg biology, diabetes) and applied (genetics, human disease) aspects of translational immunology. More information at www.eppendorf.com/award Both prize winners will visit Eppendorf in Hamburg during 2016. Check out the next BioNews issue for more info! Trademark Information Amazon® is a registered trademark of Amazon Tech, Inc., USA. CellTiter-Blue ® and Promega® are registered trademarks of Promega Corporation, USA. Cellavista® is a registered trademark of SynenTec GmbH, Germany. Ficoll-Paque ® is a registered trademark of GE Healthcare Bio-Sciences AB, Sweden. GE Healthcare ® is a registered trademark of General Electric Company, USA. Leica® is a registered trademark of Leica Microsystems IR GmbH, Germany. Nalgene ® is a registered trademark of Thermo Fisher Scientific, USA. Triton™ and Tween™ are trademarks of Thermo Fisher Scientific Inc., USA. Eppendorf ®, the Eppendorf logo, Combitips®, Easypet ®, ep Dualfilter T.I.P.S.®, epMotion®, Eppendorf µCuvette®, Eppendorf BioSpectrometer ®, Eppendorf Reference ®, Eppendorf Tubes®, epPoints ®, the epServices ® logo, epT.I.P.S.®, FastTemp pro ®, Multipette ®, PhysioCare Concept ®, and Repeater ® are registered trademarks of Eppendorf AG, Germany. ConvexAccess™, Eppendorf Quality™, and New Brunswick™ are trademarks of Eppendorf AG, Germany. BioFlo ® and CelliGen® are registered trademarks of Eppendorf, Inc., USA. > PRIZE COMPETITION · SERVICE 15 Win a Multichannel Pipette! The solution of the prize competition of BioNews No. 43 was “Cell Handling Quality Products“. Damjan Franjevic (University of Zagreb, Croatia) won the first prize. Good luck in our new competition! How to find out the solution: Simply arrange all letters in the light gray boxes of the crossword in the correct order. Send us the solution until October 31, 2016. 1 2 3 4 5 6 10 14 22 30 > 1st Prize: 1 Eppendorf Reference® 2 Pipette (multichannel, adjustable-volume) of your choice 19 24 26 27 2nd to 5th Prize: 29 31 32 35 36 39 18 23 25 28 9 12 17 21 8 15 16 20 All correct answers will be considered for a prize. Winners will be notified in writing. Cash payment of the prize is not possible. No recourse to legal action. The judges’ decision is final. Eppendorf employees and their families may not participate. The winner of the first prize will be published in BioNews No. 47. 7 11 13 You can either send us an e-mail to [email protected], or participate online at www.eppendorf.com/bn-service. 33 1 Amazon® Voucher worth 50.00 EUR 34 37 38 40 41 42 43 44 6th to 15th Prize: 45 46 47 48 51 52 53 49 400 bonus epPoints® each 50 (epPoints registration required) ACROSS DOWN 1 Opportunity, possibility 6 President of the United States 2009 –2016 10 Website accepting changes, contributions, corrections 12 Short form of Samuel 13 Chinese politician 14 Devoted to technical subjects, computers, etc. 16 What do rubidium and rhythm and blues have in common 17 Software optimizer 20 System of signals/symbols for communication 22 Transcription factor (abbrev.) 24 17th letter of the Greek alphabet 25 Used in molecular biology (when magnetic) 27 Group of people working together 29 Alcoholic drink 30 German capital 32 Amyotrophic lateral sclerosis (abbrev.) 35 Immunoglobulin (abbrev.) 36 Electronic pipette controller 39 Gemstone 41 ISO country code for Turkey 42 Area of land for recreation 45 French personal pronoun 46 Postscript (abbrev.) 47 Source of trouble or worry 50 ISO country code for Spain 51 Saint (abbrev.) 52 Standard deviation (abbrev.) 53 A royal hall in London is named after him 1 2 3 4 5 7 8 9 11 15 18 19 20 21 23 25 26 27 28 Curriculum vitae (abbrev.) Prize, medal Tantalus' daughter Creatine kinase (abbrev.) Developed the general theory of relativity Chemical symbol for arsenic Finely ground green tea Before noon (abbrev.) Chemical symbol for strontium Dots per inch (abbrev.) Italian for three One's domicile, house Positive-displacement tips State in the Northwest region of the United States (abbrev.) Frequently asked questions (abbrev.) Severe snowstorm French writer in the 19th century Flat piece used for carrying things Chemical symbol for argon 31 Egomaniac, ego-tripper (syn.) 33 Treasure made of vinyl, reaches 33 1/3 rpm 34 Type of gundog 37 Hard metal 38 Confidence, faith 40 Computer program running on mobile devices (pl.) 43 ISO country code for Romania 44 1,000 base pairs 48 Home of CSULA, CSUN, USC, UCLA, etc. (abbrev.) 49 1,000 times 44 down Solution hint for prize competition of BioNews No. 45: E T A I Send us the solution until October 31, 2016, via e-mail to [email protected], or participate online at www.eppendorf.com/bn-service. > ABN 4513020 want new technologies? antibodies apoptosis biomarkers cancer cytometry data diseases DNA epigenetics genomics immunotherapies > medicine microbiomics microfluidics microscopy neuroscience proteomics watch our webinars Learn about the latest breakthroughs, new technologies, and ground-breaking research in a variety of fields. Our expert speakers explain their quality research to you and answer questions submitted by live viewers. sequencing toxicology transcriptomics VIEW NOW! webinar.sciencemag.org Brought to you by the Science/AAAS Custom Publishing Office @SciMagWebinars >
© Copyright 2026 Paperzz