THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROEDURE PROCEDURE NO: MOD: PAGE: Procedure Type: General Equipment Procedure Title: Pipette (Gilson) 1. GEP 014 1st Issue 1 of 13 Risk Assessment: TASK PERFORMED Operation of Pipettes HAZARDS 1. RSI (Repetitive Strain Injury) 2. Exposure to biohazards/hazardous substances RISK ASSESSMENT 1. The risk of developing a muscle strain that leads to RSI is present whenever pipettes are used, however, risk of injury increases as the period of use increases. 2. The risk of exposure to biohazards and hazardous substances while using a pipette is present whenever these substances are being handled. Exposure may occur as a result of splashing or contact with vapours. RISK CONTROL 1. See Safety Precautions (point 4.1) in this GEP. 2. Ensure that biohazards and hazardous substances are only pipetted in appropriate environments such as in biohazard hoods or fume cupboards, refer to the MSDS. Always wear gloves (and safety glasses where necessary) when pipetting any substance. Draw and dispense volumes slowly and with care to avoid splashing. 2. Calibration: 2.1 Frequency All pipettes should be calibrated at six-month intervals, or after major cleaning. 2.2 Method - Calibration Set up WRITTEN BY NAME (signed) DATE Distributed To: GEP Master File / GEP Lab File CHECKED BY THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE Procedure Type: General Equipment Procedure Title: Pipette (Gilson) 2.2.1 PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 2 of 13 Pipettes must be calibrated using the forward pipetting technique (see step 4 - General Operation. 2.2.2 GILSON does not guarantee that its' pipettes conform to specifications when using other manufacturers tips. 2.2.3 Balances should be checked and if possible, calibrated before use in pipette calibration. 2.2.4 P2 and P10 GILSON pipettes must be sent away for calibration as they require a 6 decimal place balance. 2.2.5 A NATA calibrated thermometer must be used for all temperature measurements (see GLP 001 NATA Thermometer Calibration). 2.2.6 Double de-ionised or distilled water must be used in the calibration. 2.2.7 The weighing vessel should be filled to a volume of 10 - 50 times the test volume. The vessel should have a narrow neck (such as a conical flask) to minimise the effects of evaporation. Heat from a light source or radiated heat from the hands may increase the evaporation rate. Use a lid/cap for volumes below 20μL. Only a high ambient humidity will negate the need for a cover. The cover/lid should be made of a non-porous material and tweezers must be used to lift it on and off the flask. 2.2.8 For all models other than the P5000 and P10mL, racked tips must be used for the calibration. Do not touch the tip or white shaft of the pipette with the hands while fitting the tip. For the P5000 and P10mL models, take a tip (between the thumb and forefinger) and place it onto the pipette firmly with a slight twisting motion to ensure a positive and airtight seal. 2.2.9 It is vital that anyone attempting to calibrate pipettes can pipette accurately and repetitively. This can be tested by performing a series of 10 -15 weighings with a single pipette and noting the results. If repetitive results cannot be achieved, then someone else must perform the calibration. Even a pipette that is out of the correct calibration range should dispense repetitive volumes unless it is leaking or sticking. See step 6. Leak Test. 2.3 Method - Calibration 2.3.1 Allow the temperature of the pipettes, tips, thermometer, a beaker of water and an empty measuring vessel to eqilibrate for one hour. They should be placed as close as possible to the balance being used. THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE Procedure Type: General Equipment Procedure Title: Pipette (Gilson) 2.3.2 PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 3 of 13 The balance should be capable of weighing to at least one tenth of the maximum allowable error deviation. Therefore a 6-digit balance is required for volumes less than 20µL, a 5 digit balance for volumes less than 200µL and a 4-digit balance for volumes greater than 200µL. 2.3.3 An Estimation of Evaporation Loss is only required when calibrating the P20 model, as the evaporation loss for measurements above 20µL is insignificant. Evaporation loss measurements should be taken before and after the calibration measurements. Estimation of Evaporation Loss is determined as follows: a) Fill the weighing vessel to half its level with temperature stabilised distilled or double de-ionised water b) Cover the weighing vessel with the lid c) Tare the balance d) Take off the lid with tweezers e) Aspirate a sample f) Dispense the sample in a dummy vessel outside the balance g) Cover the weighing vessel h) Read and record the result (a negative value) i) Repeat the steps d - h three times j) Calculate the mean of these readings, e, usually between 20nL and 30nL, under normal conditions. Record all measurements and the mean (e) on the Pipette (Gilson) Calibration Record Sheet (Illustration 9.1). 2.3.4 Set the volume of the pipettes as per the following table. MODEL SET VOLUME TO PERMITTED RANGE P2 (0.1 - 2µL) 0.5µL 0.483 - 0.517µL P10 (0.5 - 10µL) 1.0µL 0.983 - 1.017µL P20 (2.0 - 20µL) 2.0µL 1.93 - 2.07µL P100 (20 - 100µL) 20.0µL 19.8 - 20.2µL P200 (50 - 200µL) 50.0µL 49.7 - 50.3µL P1000 (200 - 1000µL)* 200µL 198 - 202µL P5000 (1000 - 5000µL) 1000µL 993 - 1007µL P10mL (1000 - 10000µL) 1000µL 979 -1021µL * The P1000 pipette must only be calibrated using C200 tips. THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 4 of 13 Procedure Type: General Equipment Procedure Title: Pipette (Gilson) 2.3.5 Use the Calibration Record Sheet (Illustration 9.1) to record all measurements. 2.3.6 Place a tip firmly onto the cone tip. 2.3.7 Measure the water temperature to the nearest 0.1oC and record it on the record sheet. 2.3.8 Fill the weighing vessel to 10-50 times the test volume. Place the weighing vessel onto the balance and press Tare. Pre-rinse the tip by drawing up the set volume of liquid and discarding it outside the weighing vessel. Pipette distilled water into the vessel at least ten times, recording the mass to the nearest 0.0001g each time. This should be done as quickly as practicable but without compromising the integrity of the liquid delivery, the precision of the operator technique or the time intervals. 2.3.9. If required, perform another 4 measurements for the Estimation of Evaporation Loss (step 2.3.3) Record measurements on the Calibration Record Sheet. 2.3.10 Measure and record the water temperature again and use the average of the two measurements as the calibration temperature. 2.3.11 Use the table below to determine the conversion factor (z) and record it on the Calibration Record Sheet. Temp oC Conversion Temp oC Factor(z) Conversion Temp oC Factor(z) Conversion Factor(z) 15 1.0020 20.5 1.0030 26 1.0043 15.5 1.0020 21 1.0031 26.5 1.0044 16 1.0021 21.5 1.0032 27 1.0045 16.5 1.0022 22 1.0033 27.5 1.0047 17 1.0023 22.5 1.0034 28 1.0048 17.5 1.0024 23 1.0035 28.5 1.0050 18 1.0025 23.5 1.0036 29 1.0051 18.5 1.0026 24 1.0038 29.5 1.0052 19 1.0027 24.5 1.0039 30 1.0054 19.5 1.0028 25 1.0040 20 1.0029 25.5 1.0041 2.3.12 Calculate the average mass of the balance readings (m). Record the result on the record sheet. THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE Procedure Type: General Equipment Procedure Title: Pipette (Gilson) PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 5 of 13 2.3.13 Calculate the average volume using the following formula: V = ( m + e ) X z, where; V = average volume m = average mass e = average evaporation loss* z = temperature conversion factor * used only in P20 model calibration. 2.3.14 Compare the average volume result with the values given in the Permitted Range column of the table in step 2.3.4. If the volume is within this range, then calibration is complete. Proceed to step 2.4. If the volume is outside this range, the pipette needs to be adjusted, as follows. 2.3.15 To adjust the pipette, dismantle as follows: ♦ Pull the tip ejector down and unscrew the connecting nut. ♦ Separate the handle from the bottom part of the pipette. ♦ Remove the piston from the tip holder then inspect the piston surface, oring and seal. (Never disassemble the upper part of the pipette). ♦ Clean the piston, o-ring and seal with detergent and water. Rinse with absolute alcohol and leave to dry. ♦ If required, replace the o-ring and seal. Always replace the o-ring and seal at the same time using genuine parts only. (These have a life span of approximately 12 months.) 2.3.16 Reassemble the pipette and repeat the calibration (steps 2.3.1 - 2.3.14). If, after the second calibration test the volume is still outside the Permitted Range, the pipette may need to be reseated. Do not attempt this step without seeking assistance from an experienced staff member. 2.3.17 Wind the volume adjuster right down to the bottom where it should read V2. If not, wind the volume adjuster up until it comes right out. Push the volume and piston back in until it rests against the lug but is not winding. Move the adjuster to read V1, then push down to engage the thread. Wind the volume adjuster to the bottom again, where it should read V2. THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 6 of 13 Procedure Type: General Equipment Procedure Title: Pipette (Gilson) 2.3.18 Repeat the calibration procedure (step 2.3.1 - 2.3.14). If the pipette is still out of range, then the calibration may be adjusted using the special calibration tool. Adjust as described in the following table: PIPETTE MODEL VOLUME INCREASE/DECREASE PER FULL TURN (μL) P20 1 P100 4 P200 10 P1000 40 P5000 200 P10000 400 2.3.19 Repeat the calibration procedure (step 2.3.1 - 2.3.14) and 2.3.18 if required. If the pipette is still out of range - send it to John Morris Scientific for professional diagnosis and calibration. 2.4 Calibration Control 2.4.1 Once successful calibration is complete a calibration control test must be performed. Set the pipette as indicated in the table below: MODEL SET VOLUME TO CONTROL LIMITS P20 20.0µL 19.9 - 20.1µL P100 100µL 99.5 - 100.5µL P200 200µL 199 - 201µL P1000 1000µL 995 - 1005µL P5000 5000µL 4979 - 5021µL P10mL 10000µL 9985 - 10042µL 2.4.2 Perform one pre-rinsing (aspirate a sample, and dispense it). 2.4.3 Using the same procedure described in 2.3, perform three consecutive measurements. 2.4.4 Calculate the mean volume dispensed for the three measurements, and compare these with the CONTROL LIMITS in the table above. If the mean value is out of the control limits, the pipette must be recalibrated. If the mean value is within these limits, the pipette has been calibrated successfully. THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE Procedure Type: General Equipment Procedure Title: Pipette (Gilson) 2.5 PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 7 of 13 Records Record all calibration measurements and adjustments on the Calibration Record Sheet (multiple sheets may be required - staple them together) and file in the appropriate section of the Equipment Records folder when complete. 3. Monitoring: 4. General Operation: 4.1 N/A SAFETY PRECAUTIONS - excessive pipette use can lead to RSI (Repetitive Strain Injury). Therefore the following guidelines should be noted: 4.1.1 Wrists should be held in a straight, neutral position - it may be necessary to incline the sample holders or solution flasks, rather than the wrists. 4.1.2 The workload should be shared between right and left hands. 4.1.3 Elbows should be held as close as possible to the users' side. 4.1.4 Reaching may be reduced by using shorter pipettes, shallower waste containers for used tips and shallower solution containers. 4.1.5 Items should be positioned on the bench and as close as possible to the worker to minimise twisting of the neck and torso. 4.1.6 The chair should provide adequate lower back and thigh support. 4.1.7 Regular breaks should be taken - even very short breaks help. 4.1.8 Work processes should be evaluated to spread pipetting throughout the day. 4.1.9 Pipetting should be rotated among several staff. 4.1.10 Pipetting tasks should be automated wherever possible. 4.2 The delivery volume is set using the device seated in the pipette handle. Ensure that the desired volume is in line with the indicator and that the digits are completely visible on the handle display. Do not set volumes outside the pipette's specified volume range. 4.3 The pipette is fitted with a tip ejector system on the top of the tip holder. Direct the tip towards a suitable waste receptacle and press on the ejector mechanism, disposing of the tip safely. 4.4 Forward Pipetting Technique THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE Procedure Type: General Equipment Procedure Title: Pipette (Gilson) PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 8 of 13 4.4.1 Depress the operating button to the first stop. 4.4.2 Lower the tip under the surface of the liquid and slowly release the operating button. Withdraw the tip from the liquid, touching it against the side of the container to remove excess liquid. 4.4.3 Deliver the liquid by gently depressing the operating button through the first stop all the way to the second stop. This action will empty the tip. 4.4.4 Release the operating button to the ready position. Replace tip if necessary and continue with pipetting. 4.5 Reverse Pipetting Technique 4.5.1 The reverse technique is suitable for pipetting liquids having high viscosity, a tendency to foam easily or for very small volumes. NB Performing a serial dilution is more accurate than a straight dilution, especially when very small volumes are involved (less than 10μL). For example, it would be more accurate to perform a serial dilution involving two dilutions of 1/10 than a straight dilution of 1/100. 4.5.2 Depress the operating button all the way to the second stop. 4.5.3 Lower the tips under the surface of the liquid and slowly release the operating button. This action will fill the tip. Withdraw the tip from the liquid touching it against the side of the container to remove excess liquid. 4.5.4 Deliver the preset volume by gently depressing the operating button to the first stop. Some liquid will remain in the tip and should not be included in the delivery. It should either be discarded or pipetted back into the container. 4.6 Filter tips should be used when pipetting corrosive substances to prevent damage to the internal parts of the pipette. After contact with corrosive liquids, the piston should be cleaned with alcohol and a soft tissue. 5. Cleaning: 5.1 Frequency 5.1.1 Before use - A routine external cleanliness check should be performed and the pipette cleaned if required. 5.1.2 Six Monthly - Disassemble and calibrate. THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE Procedure Type: General Equipment Procedure Title: Pipette (Gilson) 5.2 PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 9 of 13 Method 5.2.1 Before Use Check for dust and dirt on the outside surfaces of the pipette. Particularly the tip cone. Use absolute alcohol or detergent and water to clean exterior surfaces as required. If the inside of the cone tip is dirty, disassemble and clean, according to step 2.3.15, then perform a calibration control test as described in step 2.4. 5.2.2 Six Monthly Disassemble and clean, as described in step 2.3.15. Then Calibrate the pipette, as described in step 2. 5.3 Records Record all cleaning/calibration details on the Maintenance Record Sheet (Illustration 9.2). 6. Maintenance: 6.1 6.2 Frequency 6.1.1 Ongoing maintenance - Leak Test 6.1.2 Ongoing maintenance - Diagnostic Test 6.1.3 Periodic Maintenance - Six monthly Clean and Calibration. Method 6.2.1 Ongoing Maintenance - Leak Test To test the pipette for leaks (for pipettes P1000 to p10mL); ♦ Set the pipette to the maximum setting and draw some water into the tip, hold the pipette vertically and let it stabilise for 20 seconds. ♦ If a drop appears at the end of the tip, there is a leak. To test for leaks (for pipettes P2 to P200); ♦ Set the pipette to the maximum setting and draw some water into the tip, hold the pipette vertically and let it stabilise for 20 seconds. ♦ If a drop appears at the end of the tip, there is a leak. If no drop appears, re-immerse the tip in the test liquid. While immersed, the level of the THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE Procedure Type: General Equipment Procedure Title: Pipette (Gilson) PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 10 of 13 liquid in the tip should remain constant. If the level in the tip goes down, there is a leak. Leaks may be caused by the following: ♦ a scratched or damaged tip holder ♦ use of non-Gilson tips ♦ use of non-Gilson seals ♦ a damaged seal ♦ vapour pressure from organic solvents. When an organic solvent is used with an air-displacement pipette, leaks may occur. This is due to the difference in the solvent vapour pressure and the vapour pressure of the air cushion between the piston and the sample. To solve this problem is to use a positive-displacement pipette, as it has no air-cushion. Alternatively, saturating the air cushion of the pipette with solvent vapour by aspirating and distributing solvent repeatedly. The leak will stop when the pressure equilibrium is reached. 6.2.2 Ongoing Maintenance - Diagnostic Test To check if the pipette is performing normally: ♦ Go through the entire volume range using the push -button adjusting knob (or thumb wheel on the older pipettes). The knob should move smoothly. ♦ The min and max settings should correspond to the pipettes normal volume range. ♦ Check the alignment and the movement of the volumeter display. ♦ Set the volume at the maximum volume for that pipette and depress the push-button slowly. The movement should be smooth. 'Hitches' in the motion or variations in friction may be due to a scratch or corrosion on the piston, or to a bent operating rod. Listen for a spring noise, which would indicate incorrect positioning of the spring inside the pipette. ♦ Any problems may require an internal inspection or be sent away for a service. For dismantling and cleaning instructions, see step 6.1.3, Periodic Maintenance - Six Monthly Clean and Calibration THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE Procedure Type: General Equipment Procedure Title: Pipette (Gilson) 6.1.3 PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 11 of 13 Periodic Maintenance - Six Monthly Clean and Calibration ♦ All pipettes must be cleaned and calibrated every six months. Disassemble and clean pipettes, as described in step 2.3.15. Then calibrate the pipette, as described in step 2. 6.3 Records Record dates and details of all maintenance performed on the Maintenance Record Sheet (Illustration 9.2) and file completed sheets in the appropriate section of the Equipment Records folder. 7. Instruction Manual: 7.1 8. 9. 10. See specific ERN for availability. Repairs: 8.1 Check ERN for Service Company and warranty details. 8.2 Contact the University Workshop for advice and repairs. Illustrations: 9.1 Pipette (Gilson) Calibration Record Sheet 9.2 Pipette (Gilson) Maintenance Record Sheet Change History: 10.1 Issue Number: 1st Issue Date Issued: 10.2 Issue Number: Reason for Change: Date Issued: THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 12 of 13 Procedure Type: General Equipment Procedure Title: Pipette (Gilson) ILLUSTRATION 9.1 CALIBRATION RECORD SHEET PIPETTE (GILSON) ERN/MODEL: / INITIALS: DATE: Acceptable Weight Range: oC Average Water Temp Conversion Factor (z) Water temp 1 Water temp 2 EVAPORATION ESTIMATION MASSES (for P20 model only) BEFORE CALIBRATION MASS M1 M2 M3 M4 AFTER CALIBRATION MASS M5 M6 M7 M8 AVERAGE (e): READING 1 2 MASS AVERAGE MASS (m) AVERAGE VOLUME V=(m+e)X z 3 4 5 6 7 8 9 10 THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE PROCEDURE NO: GEP 014 MOD: 1st Issue PAGE: 13 of 13 Procedure Type: General Equipment Procedure Title: Pipette (Gilson) ILLUSTRATION 9.2 MAINTENANCE RECORD SHEET PIPETTE (GILSON) ERN: Date Maintenance Performed Signed
© Copyright 2026 Paperzz