A Normal Incidence Scanning Reflectometer of High Precision Ulrich Gerhardt and Gary W. Rubloff A near-normal incidence (6°) reflectometer system is described that records continuously and directly the reflectance R(co) as a range of photon energies is scanned. The system has an absolute error of ±2 X 10-2 and a relative error of ±t2 X b5. It incorporates a quartz light pipe rotating at 70 Hz which captures light from the incident and the reflected beam, respectively, during about 20% of its period of rotation in either case. A gating circuit separates the output signal of the photomultiplier into two channels, corresponding to the incident and the reflected beam, respectively. The signal corresponding to the incident beam is kept constant by a servo system which regulates the gain of the photomultiplier. The reflectance is thus proportional to the signal of the second channel, which is recorded as a function of photon energy. Portions of the reflectance spectrum of Ge are given as examples. No trace of a fine structure in the reflectance of Ge below 2 eV is found. 1. Introduction The complex dielectric function (co) of a solid is closely connected with its electronic structure. Thus, the accurate eperimental determination of this quantity is of great importance. One method to determine e consists of measuring the reflectance at or close to normal incidence over a sufficiently large range of photon energies and then performing a Kramers-Kronig inversion.I Reflectance measurements require a separation of the incident and the reflected beam. This requirement gives rise to two major difficulties. First, the noise spectrum of the light source, such as a high pressure discharge, and of the detector quite frequently contains large low frequency components. Such a noise spectrum is particularly harmful when measuring the reflectance by a point-by-point method, in which one normalizes the incident beam to 100%, changes the optical path of the light, and then measures the reflected beam. The change of the optical path takes typically 10 see or longer, thus making the method susceptible to instabilities with a time constant of this order of magnitude. Second, Io, the output of the detector, is normally a strong function of photon energy, so that intensity variations of several orders of magnitude often occur within the spectral range of interest. It is possible to record Io () and then R(cw) Io(W) and divide the two quantities afterwards in order to get the reflectance. However, this method performs poorly if Io (co) depends The authors are with The James Franck Institute, The University of Chicago, Chicago, Illinois 60637. Received 15 August 1968. strongly on c, because it requires a very high reproducibility of the wavelength setting. The point-by-point method is capable of very accurate results, provided the low frequency components of the noise spectrum are small enough.2 The method consisting of first recording o (co) and then R(w) Io(W) requires both a good long-term stability and a weak dependence of o on in order to be accurate. An ideal method should be capable of bypassing both difficulties, i.e., it should alternate between Io and RIO at high frequency in order to avoid the low frequency components of the noise spectrum and it should be capable of operating under strong variations of lo(U). Furthermore, a high signal-to-noise ratio requires that a large fraction of the photons available are used in the measurements, i.e., the on-time should be a large fraction of the period of the system. The purpose of this paper is to describe a reflectometer which approaches the ideal system mentioned above. The paper describes the optical and electronic components of the system and its characteristics in operation. It also gives some portions of the reflectance spectrum of germanium samples recorded with this system. 11 The Scanning Reflectometer Design of the System Our procedure for continuously recording R () at near normal (.6) incidence is shown schematically in Fig. 1. The beam emerging from the monochromator is focused on the surface of the sample. A Suprasil fused quartz light pipe 4.8 mm in diameter rotates at about 70 Hz, sampling incident (Io) and reflected (RIo) beams alternately. Figure 2 shows the top view of the light pipe, the top part of which is bent slightly off-axis so that it does not interfere with the incident beam February 1969 / Vol. 8, No. 2 / APPLIED OPTICS 305 Sample is attached to the light pipe and which sweeps across the poles of two induction coils, Fig. 1. The gating signals are adjusted to sample the output signal of the multiplier during the flat part of the pulses. The relative duration of the gating signals with respect to the period of the light pipe is about 11%; it is given by the separation of the two little spikes on top of the flat parts of the main pulses in the lower half of Fig. 3. The lo voltage signal is applied to a high input impedance differential preamplifier with variable bias voltage as a reference. The output of the differential amplifier determines the power applied to a servomotor. A multiturn potentiometer which is mechanically coupled to the servomotor controls the high voltage, and therefore the gain, of the photomultiplier. Hence the anode current of the multiplier corresponding to the incident beam is kept constant even as the monochromator output intensity changes due to the scanning of h and to fluctuations in the light source. Light Pipe Fig. 1. Schematic diagram of the system. Incident Beam Io B Reflected Beam RIO Fig. 2. Top view showing the geometry of the sample and bent light pipe. 3 while measuring the reflected beam. The light pipe catches each beam fully for about 20% of its period, producing wide, flat pulses, shown schematically in Fig. 1. The actual signal is reproduced in Fig. 3. An induction motor rotates the light pipe assembly. The coupling between motor and light pipe consists of two pulleys and a flexible rubber belt. The rubber belt decouples the high frequency vibrations of the motor from the light pipe. The light pipe assembly, which is mounted in ball bearings, is balanced dynamically to minimize vibrations. The gating circuit shown in Fig. 4 separates the 1 and RIO channels and measures the peak values of the signals. The field effect transistors act as switches to apply the voltage across the 10-kQ anode resistor alternately to the load capacitors as long as the gating signals persist. The gating signals are the output voltages of two monostable multivibrator circuits. They are triggered by pulses produced by a small magnet which 306 APPLIED OPTICS / Vol. 8, No. 2 / February 1969 Fig. 3. Photomultiplier output measured across 10 kg of the gating circuit (see Fig. 4). The larger peaks correspond to the Io channel and the smaller peaks to the Rlo channel. The upper trace is with the gating circuit turned off and the lower trace is with the gating circuit operating. From Photomultiplier 20/± A RI, 10 sv 5V 0 7l-7Jl 1L -57V trp J~ - From Monostoble Multivibrators Fig. 4. Gating circuit. 0.7 Ge, 3000 K w , 0.6 IU 0.5 2 3 4 PHOTON ENERGY 5 (eV) Fig. 5. The reflectance of P type Ge (0.074 02cm) from 1.6 eV to 5.2 eV. Vibration coupling to the monochromator causes a wavelength modulation, which produces a bump in the reflectance near the strong xenon line at 4.85 eV. The energy resolution given in the bottom part is the half-width of atomic mercury lines, as recorded with our system. The RIo voltage signal is amplified by a Keithley 150B Microvolt-Ammeter and then applied to the Y channel of an X-Y recorder. The Y channel is calibrated for absolute values of the reflectance by applying the Io signal to the RIO channel and adjusting the gain to full scale of the recorder. The X channel of the recorder monitors the wavelength setting of the monochromator. A variable offset voltage is available at the input of the Keithley amplifier to compensate for most of the RIo signal. We are thus able to record fine structure in R (w) by using a much higher gain of the amplifier or the recorder. Optimization of the System The servo system keeps the lo signal constant to within This is achieved by using a large dc gain of the feedback loop. A damping force proportional to the velocity of the motor reduces the tendency of the system to oscillate and reduces the response time. An aluminum flywheel mounted on the shaft of the servomotor moves in a strong magnetic field to provide the damping mechanism. The values of the anode resistance and the load capacitance in Fig. 4 are chosen to fulfill two requirements. First, the RC time constant is set equal to the gating time, resulting in a fast response while still averaging over variations of the intensity during the gating time, which are about 5%. Second, the anode resistance is kept small compared with the load resistances of the differential preamplifier in the I channel and of the amplifier in the RIO channel. In this way the capacitors discharge negligibly when not connected to the anode resistance; hence the signal levels read are very close to the peak values and independent of differences in the load resistances of the two channels. Dynamical balancing of the light pipe assembly is essential to the success of this method, since vibrations cause microphonics of the photomultiplier and wavelength modulation of the monochromator output. i 2 X 10-'. Furthermore, high precision, low friction ball bearings in the mounting of the light pipe are needed to minimize vibrations generated in the bearings themselves. The 70-Hz rotation speed of the light pipe is well above the frequency response of the servo system, which is about 10 Hz. Thus, low frequency signals from intensity fluctuations and wavelength scanning are compensated by the feedback circuit, whereas higher frequency signals are attenuated by the filter circuits of the system. The absolute error of the reflectance measured with our system is (AR/R)bsoluto = ±i2 X 10-2. This is the deviation of the measured reflectance of quartz from the theoretical reflectance, calculated using the tabulated refractive index. This value is the maximum error within the spectral range of the system, i.e., from 1.5 eV to 5.5 eV. The absolute error depends smoothly on photon energy. Therefore, the ability of the system to detect fine structure in the reflectance superimposed on a smoothly varying background is given by the performance of the servo system, i.e., the relative error of the system is (R/R)oiative = 2 X 10-5. 111. Experimental Examples: Germanium The room temperature reflectance of P type Ge ( = 0.074 Q cm) is given in Figs. 5-7. The surface of this sample is prepared by grinding with increasingly finer abrasives down to mesh 1000 and subsequently etching in a mixture of nitric, hydrofluoric, and acetic acid (ratio 3:1: 1). The etching process is stopped by rinsing in methanol and quickly removing the thin methanol layer on the surface by a warm stream of air. Figure 5 shows the actual recorder trace of the reflectance of Ge from 1.6 eV to 5.2 eV. The zero of the reflectance scale is suppressed. The bump at 4.85 eV is caused by a wavelength modulation of the monochromator output. The source intensity changes rapidly in this region because of the strong xenon line at 4.85 eV, superimposed on the continuum of the high pressure xenon lamp used in our measurements. In Fig. 6, the sensitivity has been increased fourfold compared with O. 0 Z 0 i 1: a: 0.45 2.0 Fig. 6. 2.5 PHOTON ENERGY 3.0 ( eV) The reflectance of Ge from 1.6 eV to 3.2 eV in expanded scale (same sample as Fig. 5). February 1969 / Vol. 8, No. 2 / APPLIED OPTICS 307 W , 0.512 Z W0.510 ( w 0.508 2.2 2.!1 PHOTON ENERGY (eV) 2.3 Fig. 7. The reflectance of Ge from 2.04 eV to 2.32 eV in drastically expanded scale (same sample as Fig. 5). The vertical arrows indicate the relative error of the measurement. Fig. 5. The figure is again an actual recorder trace and spans the photon energy from 1.6 eV to 3.2 eV. A trace taken with a fortyfold increased sensitivity is shown in Fig. 7 for the region around the 2-eV doublet. The energy resolution given in Figs. 5-7 is the halfwidth of atomic mercury lines, as recorded with our monochromator. The relative error of the reflectance is indicated on Fig. 7 only; it is given by the width of the line enclosed by the arrows. IV. Discussion The scanning reflectometer described in this paper approaches the features of the ideal system discussed in Sec. I. The relatively high operating frequency allows us to take advantage of the high intensity of high pressure discharge lamps. The large low frequency fluctuations of these light sources do not appreciably affect the operation of our system, except when ultimate stability is required. We mention in this context that Figs. 5 and 6 were taken using a high pressure xenon lamp, whereas Fig. 7 was taken using a tungsten-iodine lamp. Repeating the measurement given in Fig. 7 with the high pressure xenon lamp increases the noise approximately fivefold. The combination of the small relative error with the possibility to scan the spectrum makes our system ideally suited to detect fine structure in reflectance. We used this feature to investigate the reflectance of Ge between 1.6 eV and 2 eV. Potter 4 previously reported fine structure in this region which is about AiR = 0.01 above the background reflectance and assigned it to the L3 ' -> L1 transition. The recorder trace given in Fig. 6 shows no indication of such a fine structure. Similar measurements using differently doped samples show no fine structure either. The accuracy of our method enables us to detect fine structure 500 times smaller than the one reported by Potter. The absolute error of the reflectance of our system is dominated by the asymmetries of the optical system, i.e., the beam is convergent for the Io position and divergent for the RIO position of the light pipe, as shown in Fig. 2. Furthermore, the cross section of the beam on the front surface of the light pipe might be slightly different in size for the I and the RIO position and the light might not hit the same spot on the light pipe. The asymmetries of the electronic circuit contribute very little to the absolute error of the reflectance. The on-time of our system is about 22% of its period. One could, in principle, approach an on-time equal to the period if a square wave motion of the light pipe were used. However, the vibrations generated by such a motion might be difficult to control. The on-time could also be increased by an oscillatory motion of the light pipe, with its extreme positions identical to the Io and RIo position. The task of minimizing vibrations of the multiplier and the monochromator would still be more difficult than in the case of a rotary motion. The spectral range of the scanning reflectometer described here can be extended into the vacuum uv by coating the front surface of the light pipe with sodium salicylate. We have in fact already used the system in the vacuum uv.* H. Fritzsche for his help and The authors thankI advice and for his interest in the subject. This work was supported by the U.S. Air Force Office of Scientific Research under a contract. References 1. T. S. Robinson, Proc. Phys. Soc. London B65, 901 (1952); H. R. Philipp and E. A. Taft, Phys. Rev. 113, 1002 (1959). 2. See, e.g., H. E. Bennett and W. F. Koehler, J. Opt. Soc. Amer. 50, 1 (1960). 3. The same geometry of the light pipe has been used previously for reflectance measurements in the vacuum uv. For this application, the front surface was coated with a sodium salicylate layer. The light pipe was turned by hand. Results obtained with this set up were reported by U. Gerhardt and E. Mohler, Phys. Stat. Sol. 18, K45 (1966). 4. R. F. Potter, J. Phys. Soc. Japan 21, 107 (1966). The optical constants given in this paper were obtained from polarimetry measurements, using large angles of incidence. This fact might make polarimetry measurements more sensitive to surface layers than measurements of the reflectance at normal incidence. * We use the synchrotron radiation produced by the electron storage ring located at the Physical Science Laboratory of the University of Wisconsin in Stoughton, Wisconsin, for our measurements in the vacuum uv. March 1969 features Interferometry vs Grating Spectroscopy 308 APPLIED OPTICS / Vol. 8, No. 2 / February 1969
© Copyright 2025 Paperzz