1269 Journal of Cell Science 110, 1269-1278 (1997) Printed in Great Britain © The Company of Biologists Limited 1997 JCS1406 Tip localized Ca2+ pulses are coincident with peak pulsatile growth rates in pollen tubes of Lilium longiflorum Mark Messerli and Kenneth R. Robinson* Department of Biological Sciences, West Lafayette, IN 47907-1392, USA *Author for correspondence (e-mail: [email protected]) SUMMARY It is known that locally elevated Ca2+ at the growing tips of pollen tubes is necessary for pollen tube elongation. Here we show that this localized Ca2+ is also temporally regulated and is closely associated with pulsatile tip growth. Lilium longiflorum pollen tubes were injected with the photoprotein, aequorin, and the Ca2+-dependent light output was detected with a low noise photon-counting system. Ca2+ pulses with a mean period of 40 seconds were invariably associated with growth. The pulses were sporadic and of low amplitude for about the first 1.5 hours after germination. With subsequent growth, pulses increased in amplitude and the period between pulses became more regular. We have localized these Ca2+ pulses to the INTRODUCTION Pollen tube elongation is an essential process for the reproduction of flowering plants. After the pollen grain lands on the stigma, it germinates a tube that grows down the style to the ovary in order to transport the sperm from the grain to the ovum. Tube elongation is a result of directed vesicle fusion, which adds the necessary cell wall material and plasma membrane components (Lancelle and Hepler, 1992; Larson, 1965; Rosen et al., 1964). Lilium longiflorum pollen grains, typically about 90 µm in diameter, must grow tubes several centimeters long in order to reach the ovary. Pollen tube elongation is known to be a Ca2+-dependent process. The presence of Ca2+ in the growth medium is necessary for germination and growth (Brewbaker and Kwack, 1963). Tubes grown in 1% of normal Ca2+ elongate at half the usual rate and tend to burst, while even lower Ca2+ concentrations reduce growth further and induce more bursting (Weisenseel and Jaffe, 1976). Injection of tubes with the Ca2+ buffer, BAPTA, completely inhibits growth (Miller et al., 1992; Pierson et al., 1994). La3+, an inorganic Ca2+ channel blocker, has also been shown to stop pollen tube growth (Picton and Steer, 1985). It has long been thought that Ca2+ is spatially graded in growing pollen tubes, with higher Ca2+ at the tip (Jaffe et al., 1975). However, it was not until ratiometric fluorescent Ca2+ indicators, indo-1 and fura-2, were microinjected into pollen tubes that such cytosolic gradients were demonstrated unequiv- elongating end of the growing tube. The Ca2+ pulses are asymmetrical, rising more slowly than they fall. We estimate that the Ca2+ concentration at the peak of the pulses reaches nearly 10 µM. The addition of 100 µM La3+, a Ca2+ channel blocker, extinguished the pulses. An analysis of growth of elongating tubes establishes that extension is pulsatile, with a 42 second period between pulses. Calcium imaging, using the fluorescent indicator, Calcium Green dextran, shows that calcium pulses are coincident with peak growth rates. Key words: Ca2+ pulse, Aequorin, Calcium green, Tipgrowth, Pollen tube ocally (Miller et al., 1992; Rathore et al., 1991). Those experiments were done on Lilium longiflorum pollen tubes and it was shown clearly that there is a steep, tip-directed gradient of Ca2+ localized to the distal 30 µm or so of the growing tube. Furthermore, it was shown by both groups that the elevated Ca2+ at the tip was absolutely correlated with growth; if the gradient was disrupted by any means, growth ceased. Subsequent work has shown that the extent of the region of elevated Ca2+ is confined to 20 µm and that an influx of external Ca2+ is required (Pierson et al., 1994). The existence of the Ca2+ gradient is not unique to lily pollen; similar gradients have been found in Nicotiana sylvestris, Tradescantia virginiana and Agapanthus umbellatus pollen tubes (Malhó et al., 1994; Pierson et al., 1994). Growth in lily pollen is pulsatile (Pierson et al., 1996), mirroring the situation in fungal hyphae (López-Franco et al., 1994). Pierson et al. (1996) also detected fluctuations in Ca2+, but were unable to determine the periodicity due to limitations on the rate at which ratiometric images could be gathered. We too have detected Ca2+ fluctuations in lily pollen, using ratiometric fluorescent indicators, and have also failed to establish their period. In addition to the problem of the rate of gathering images, prolonged exposure to the ultraviolet excitation light used with ratiometric Ca2+ indicators damaged the tubes and interfered with growth. Consequently, we have elected to sacrifice spatial information (which is already well established), by using aequorin, in order to achieve better temporal resolution. We have also used the non-ratiometric fluorescent 1270 M. Messerli and K. R. Robinson indicator, Calcium Green, in order to establish the phase relationship between growth pulses and calcium pulses. The photoprotein, aequorin, emits light upon reacting with Ca2+. Once this reaction has occurred the aequorin molecule is irreversibly inactivated. The rate of light emission increases with the 2.2 power of the Ca2+ concentration (Shimomura et al., 1993), making aequorin an excellent indicator for spatially or temporally localized increases in Ca2+. It can be microinjected into cells and is non-disturbing to biological processes over periods of hours or even days (Miller et al., 1994). It is available in a variety of semi-synthetic forms with differing Ca2+ sensitivities (Shimomura et al., 1993). Its major drawback is the low level of light emission, which makes imaging difficult. However, total light emission from single cells can be recorded by fairly simple means, with excellent temporal resolution (Cork et al., 1987; Keating et al., 1994). We have employed this approach in order to detect Ca2+ fluctuations in growing lily pollen tubes. Although spatially-resolved Ca2+ dynamics can be determined with aequorin, using ultralow-noise imaging equipment, a more accessible method was chosen. Scanning confocal fluorescence microscopy of injected Calcium Green dextran is sufficient to show relative changes in Ca2+ concentration. As the Calcium Green signal is not ratiometric, the fluorescence will depend not only on the Ca2+ concentration but also on the amount of indicator in the optical path. This factor can be controlled for by imaging a calcium-insensitive fluorescent indicator; we used fluorescein dextran. Both Calcium Green and fluorescein can be excited by the krypton/argon laser line at 488 nm, which avoids UV irradiation and permits long-term imaging of pollen tubes. MATERIALS AND METHODS Lilium longiflorum flowers were grown in a greenhouse and the pollen was separated from the anthers, dried at room temperature for 1 week and stored at −20°C. A small sample of pollen was hydrated in modified Dickinson’s medium (in mM: 0.16 H3BO3, 0.127 Ca(NO3)2, 1.0 KNO3, 1.0 succinic acid, 292.0 sucrose), pH 5.5, which was mixed with an equal volume of 2% low-temperature gelling agarose (FMC Co., Rockland, ME), in modified Dickinson’s medium. The agarose provided mechanical support during injection but also impeded the rapid change of medium near the tubes. The pollen/agarose mixture was then put into 35 mm ×10 mm Falcon culture dishes with No.1 coverglass bottoms, and placed at 4°C for 1 minute to gel the agarose thus forming a gelled layer about 1 mm deep. A few milliliters of modified Dickinson’s medium was added to immerse the gel and the dish was then placed in a 25°C incubator. Pollen began germinating in less than an hour under these conditions. Growth rate measurements and Ca2+ measurements were done at room temperature, 21-22°C. Aequorin A high pressure injection system (Kiehart, 1982) was used to inject pollen tubes with 1-2 pl of 10 mg/ml recombinant h-aequorin (provided by Dr O. Shimomura at the Marine Biological Laboratory, Woods Hole, MA). Briefly, mercury filled micropipettes were connected to a threaded syringe filled with a fluorocarbon oil, Fluorinert (Sigma Chemical Co., St Louis, MO), via Fluorinert-filled tubing. Positive pressure was applied to force the mercury to the micropipette tip. The micropipette was then dipped into a small volume of h-aequorin, which was drawn into the micropipette followed by a small volume of dimethylpolysiloxane, DMPS (Sigma). The DMPS oil cap in the micropipette prevented the aequorin from being exposed to Ca2+ before it could be injected into the tube. Tubes growing parallel to the coverslip plane and with lengths between 300 and 500 µm were selected for injection. Micropipettes were pulled from 1 mm OD Microcaps glass (Drummond Scientific, Broomall, PA) to approximately 1 µm OD and were silanized before use. For silanization, micropipettes were dried at 200°C for 1 hour, treated with N,N-dimethyltrimethylsilylamine (Fluka Chemical Co., Ronkonkoma, NY) for 1 hour at 200°C in an enclosed chamber and then cooled before use. Photomultiplier system Photon counting was performed with a system previously described (Cork et al., 1987; Keating et al., 1994). The equipment comprised an R464s photomultiplier tube (Hamamatsu, Bridgewater, NJ) and a Pacific Precision Instruments (Concord, CA) amplifier/discriminator. The photon-generated pulses were sent to a PC-LPM-16 data acquisition card (National Instruments, Austin, TX) installed in an IBMPC/XT, which recorded the number of pulses in each one second interval. The data were then converted to ASCII format and imported into Axum 5.0 (Mathsoft Inc., Cambridge, MA). Data were then averaged over 5 second intervals before graphing, except in Fig. 2 where the 1 second, unaveraged data were used. The average period of the Ca2+ pulses is given as the mean ± s.e.m. The coverslip-bottomed Petri dish was placed directly on the photomultiplier’s end-window photocathode, the photon-sensing component of the system, and a light-tight cover was placed over the dish. The position of the tube on the coverslip was marked to ensure that the whole tube and grain were centered on the photocathode. Likewise, when the tip of the tube was placed just off the photocathode, similar markings were made. Marks were made adjacent to the grain, opposite the side of germination (within 100 µm). When the tube had grown for a few hours, about 1.5 mm long, two marks were made approximately two-thirds down the tube, one on each side of the tube. During recording these marks were placed off the photocathode, so that the two marks along with the distal third of the tube were off the photocathode while the proximal two-thirds of the tube and the grain were maintained on the photocathode. In order to change the medium or reposition the Petri dish, the high voltage to the photomultiplier tube was turned off and the cover was removed. The dish was left in place on the photocathode while the medium was changed. After the treatment had taken place, the cover was replaced and the tube was powered up. This total procedure took approximately 2 minutes. Growth records Time-lapse bright field images of elongating pollen tubes were taken on an inverted microscope (Nikon Diaphot TMD) with a Newvicon camera and microchannel plate image intensifier (Video Scope International, Ltd, Sterling, VA). Collection and analysis of data were performed on a Sun Workstation using software from G. W. Hannaway and Associates (Boulder, CO). Each image was an average of 16 video frames taken over a period of 0.5 seconds. An image was collected every 5 seconds. Tubes could be imaged for about 3 minutes before the faster-growing tubes grew out of the field of view. Successive pollen images were then viewed on the monitor and a line representing the path of elongation was drawn through the leading edge of the tubes, using the ‘vector’ function of the image processing software. Individual images were then viewed with the line overlay and the pixel position of the line on the pollen tube’s leading edge of growth was recorded. Growth rates were calculated from the differences between the tube’s leading edge pixel position of successive images, along with a known 5 second time interval and known spatial parameters. Average growth rates and the average period of the growth pulses are given as the mean ± s.e.m. Fluorescence imaging Pollen tubes were injected 2 hours after germination with 1 pl of either Ca2+ pulses and pollen tube elongation 1271 2.5 mM Calcium Green dextran (Mr 10×103) or 2.5 mM fluorescein dextran (Mr 3×103), (Molecular Probes, Eugene, OR) in 100 mM KCl and imaged with an MRC 1024 laser scanning confocal imaging system (Bio-Rad, Hercules, CA) on an inverted Nikon microscope through a ×40, 1.3 NA Nikon objective. Both dyes were excited at 488 nm with the emission light viewed through a 522 nm filter with a bandpass of 25 nm. Laser intensity was set to 1% of maximum and the diaphragm was fully opened to give a brighter signal. The resulting confocal scanning section was approximately 10 µm thick. Under these conditions, photodamage and photobleaching were minimized, and pollen tubes could be studied for hours without apparent harm. Pollen tubes that were growing parallel to the bottom of the chamber were chosen for study. An average of 5 time-lapse recordings were taken for each tube. Each recording consisted of 30-45 images taken at 5 second intervals; each image took less than a second to collect. The limitation on the number of images collected in each recording was the growth of the pollen tube tip out of the field of view. Images were displayed with a 256 pixel intensity gray-scale and were exported and analyzed using MetaMorph imaging software (West Chester, PA). Growth rates were determined from the fluorescent images by drawing a line 20 pixels thick over the long axis of the image of the growing tube. Each set of 20 pixels was averaged and a single value was returned to the middle point on the line thus generating a profile of the fluorescence under the line. The coordinates of each point on the line as well as the averaged pixel intensity of that point was saved to a spreadsheet. The leading edge of growth was determined to be the leading edge pixel value with approximately half the peak intensity value of the line profile (see Fig. 8A, arrows). The pixel coordinates of this point on the line were recorded for each image in the recording. The vector difference between coordinates along with a known time interval and distance calibrated coordinate plane were used to calculate growth rates. The growth rate between each time point was averaged with its two nearest neighbors to reduce noise before graphing. Tip localized fluorescence was determined by drawing a circle/oval approximately 10 µm in diameter and placing it on top of the images of the tubes within a micron of the tip and front edges of the tubes. The average pixel intensity of the region beneath the circle/oval was recorded in this way for each of the images in a recording. The average pixel intensity under the region varied by 1-2 pixel values when moving the region a few pixels left/right or up/down. The average tip localized fluorescence for each image was averaged with its two nearest neighbors before determining the relative fluorescence. The relative fluorescence was calculated by dividing each of the averaged pixel values per time point by the lowest averaged pixel value. This allowed comparison of changing fluorescent intensities despite differing gain settings between recordings. RESULTS Calcium pulses We have monitored intracellular free Ca2+ in 12 elongating pollen tubes of Lilium longiflorum. Fig. 1 shows a complete recording of an aequorin-injected pollen tube along with three 15 minute segments of the recording. Fig. 1A shows a 6.25 hour recording of the Ca2+ in the pollen tube and grain beginning 30 minutes after germination. Pulses appeared superimposed on a baseline Ca2+ signal that changed only slowly with time. The sharp drop at about 180 minutes occurred when the growing tip of the tube was positioned off the photocathode for 10 minutes while the rest of the tube along with the grain remained on the photocathode (see below). The baseline of the recording decreased during this procedure but still remained above instrumental background, and the Ca2+ pulses ceased. The pulses returned when the whole tube was returned to the photocathode. The final pulse occurred when Fig. 1. (A-D) Recordings of aequorin luminescence from a growing lily pollen tube and grain. (A) Full length recording of a growing lily pollen tube and grain beginning about 30 minutes after germination. (B) A segment of the first 15 minutes of the recording in A with an expanded time scale. (C) A 15 minute segment of the recording in A starting about 3 hours after germination. (D) The final 15 minutes of the recording in A showing aequorin burnout and the resulting instrumental background signal. Aequorin luminescence (photons/s) A B 150 150 100 100 50 50 0 0 0 100 200 300 400 0 10 15 375 380 385 D C 150 150 100 100 50 50 0 150 5 155 160 165 0 370 Time (minutes) 1272 M. Messerli and K. R. Robinson Calcium pulse isolation As discussed above, it is known that the elevated calcium in elongating tubes is localized to the distal 20 µm of the tube. In order to confirm that the recorded Ca2+ pulses we have detected were at the elongating end of the growing tube, we repositioned the tube so that the distal third of the tube was located off the photocathode of the photomultiplier tube while the proximal two-thirds of the tube along with the grain was still on the photocathode. Fig. 3 shows recordings from two different tubes Aequorin luminescence (photons/s) A 110 100 90 80 70 60 50 40 30 20 -10 -5 0 5 10 Time (seconds) B 2.5 Relative [Ca2+ ] the membrane was permeablized with the detergent, Triton X100, allowing Ca2+ influx from the medium and causing complete inactivation, ‘burnout’, of the aequorin sample (Fig. 1A,D). The remaining signal subsequently is the instrumentgenerated background. Low intensity, irregularly spaced pulses were observed as soon as recording started (Fig. 1B), less than 10 minutes after injection. Pulses reached higher intensities and became more regular about 1.5 hours after germination. Fig. 1C shows a segment of the recording at 3 hours after germination. The baseline of this section is around 20 photons/second with averaged pulses being between 0.5 and 6.5 times higher. Pulses occur with 10-15 second breaks between them. It can be seen that not all of the Ca2+ pulses are of the same intensity; there is a second population of pulses that is two to three times higher than surrounding pulses. We could not determine any regular pattern for these higher pulses among the different tubes, but these occasional higher pulses did occur in a majority of the recordings, particularly in longer tubes. The pulse frequency changed only slightly over the course of a recording. The segment of recording in Fig. 1C displays pulses with a period of 47 seconds while the average period for the total record during regular pulsing was 44 seconds. A similar pattern of Ca2+ pulses was observed in all 12 tubes that were studied. The mean period of the pulses was 40±3 seconds with a range of 29 to 57 seconds. During regular period pulsing, the periods showed no consistent pattern of change. Tubes continued with higher pulses until they either burst and stopped growing or until the experiment was ended by either blocking Ca2+ influx with lanthanum or by permeabilizing the cell membrane and burning out the aequorin with Triton X-100. Toward the end of the recording, pulse intensity and frequency became more sporadic (Fig. 1D). As recording continues and the aequorin is being used up, we expected a slight decline in the baseline counts and the pulse intensity. However the sudden drop at 190 minutes (Fig. 1A) is too large to be accounted for by aequorin exhaustion and must represent a decline in Ca2+. In order to extract information about the shape of the pulses, we have averaged a total of 80 pulses from two different recordings, using the 1-second raw data. Pulses with peaks between 70 and 150 photons/second, during regular period pulsing, were chosen for the study. A graph of this averaging is shown in Fig. 2A. It is clear that the rising component (solid) of the pulses initially is higher than the falling component (dash), indicating an increased average free [Ca2+] during the rising component. Fig. 2B shows the calculated relative [Ca2+] of the two components. The mirror image of the falling component was drawn on top of the rising component to facilitate comparison. The asymmetry is evident in many of the individual pulses as well, especially when viewed on an expanded time scale (data not shown). 2.0 rising component 1.5 falling component 1.0 -10 -5 0 5 10 Time (seconds) Fig. 2. (A) An averaging of a total of 80 Ca2+ pulses from 2 different tubes. Pulses were selected whose peaks were between 70 and 150 photons/second. The peaks of the pulses were aligned and the data were averaged second by second for 15 seconds on either side of the peak. The solid line represents the rising component while the dashed line represents the falling component. (B) The relative [Ca2+] levels of the rising (solid) and falling (dashed) components. The falling component is shown as the mirror image of itself in order to facilitate comparison. To calculate the relative [Ca2+] from the curve shown in A, the instrumental background along with the signal from the grain and the proximal 2/3 of the tube (5 photons/second) were subtracted from the average value of each time point. The ratio of these points with the adjusted baseline (15 photons/second) and the 2.2 power dependence of light on Ca2+ were used to calculate the resulting [Ca2+] values. before and after the tip of each tube was moved off the photocathode. There are breaks in the recordings of about 2 minutes when the photomultiplier tube was turned off in order to reposition the pollen tube. Both Fig. 3A and Fig. 3B show a loss of the Ca2+ pulses as well as an overall drop below the baseline of the tip recording. The instrumental background signal was still lower than the signal where Ca2+ was only being measured from the grain and proximal two-thirds of the tube. Source of calcium pulses We wanted to identify the source of Ca2+ that produced the pulses. Our first approach was to chelate the extracellular Ca2+ by adding 1 mM BAPTA to the medium of growing tubes. We found that the tubes burst before adequate measurements could be made. As an alternative, we added LaCl3 to a final concentration of 100 µM. Fig. 4 shows two such recordings. The La3+ caused these two tubes to swell at the tip before growth stopped. The arrows on the figures indicate the beginning of the recording after the La3+ was added to the medium. The Ca2+ pulses and pollen tube elongation 1273 A 250 Aequorin luminescence (photons/s) 200 150 100 50 0 135 145 155 165 B 100 90 80 70 60 50 40 30 20 10 0 60 70 80 90 Time (minutes) Aequorin luminescence (photons/s) pulses stopped within 10 minutes of adding La3+. Most of this delay can be accounted for by the time required for La3+ to diffuse through the 1 mm of agarose. In both recordings, there is a rise and fall of pulses just before pulses are extinguished. 100 90 80 70 60 50 40 30 20 10 0 272 Fig. 3. (A,B) Aequorin luminescence recordings from two different tubes. The first and last segment of the recording is the signal from the entire tube and grain while the middle segment of the recording (separated by breaks) is the signal from the grain and the proximal 2/3 of the tube. This may be the Ca2+ signal that causes the tubes to swell at their tips before growth stops. The latter portion of these recordings shows the signal after aequorin burnout. These recordings show that active aequorin is still present in the tube A 282 292 302 312 322 332 B 100 90 80 70 60 50 40 30 20 10 0 85 95 105 115 Time (minutes) 125 135 145 Fig. 4. (A,B) Recordings of aequorin luminescence in two pollen tubes before and after the addition of 100 µM La3+. Arrows indicate the beginning of the recording after addition of La3+. In both records, Triton X100 was added near the end of the recording in order to burn up the remaining aequorin. 1274 M. Messerli and K. R. Robinson Fig. 5. (A-D) Growth records of growing pollen tubes at two different times after germination: (A,B) 30 minutes after germination, (C,D) 3 hours after germination. Pixel coordinates of the leading edge of growing pollen tubes were recorded from successive bright field images. The difference between successive, calibrated coordinates at 5 second time intervals was used to calculate growth rates. Final values were averaged with the two nearest neighbors. Lily Pollen Tube Growth Rates (µm/s) A B 0.35 0.35 0.32 0.32 0.29 0.29 0.26 0.26 0.23 0.23 0.20 0.20 0.17 0.17 0.14 0.14 0.11 0.11 0.08 0.08 0 50 100 150 0 200 50 100 150 200 D C 0.35 0.35 0.32 0.32 0.29 0.29 0.26 0.26 0.23 0.23 0.20 0.20 0.17 0.17 0.14 0.14 0.11 0.11 0.08 0.08 0 50 100 and grain after the pulses and growth have stopped. As La3+ is a more potent aequorin activator than Ca2+ (Blinks et al., 1982) the recordings show that significant amounts of La3+ were not entering the cell. If La3+ were entering the tube, the pulses just before the tube stopped growing would be expected to be higher than preceding pulses. Growth rates It has been shown that pollen tube growth is pulsatile (Pierson et al., 1996). In order to investigate the temporal relationship between pulsatile tip growth and the calcium pulses reported here, we have measured growth rates with sufficient resolution to detect growth pulses. Fig. 5 shows growth records from tubes at 0.5 and 3 hours after germination. Although these growth records are from different tubes than the ones used for Ca2+ measurements, they can be compared to the Ca2+ pulse records in Fig. 1B and C which were also taken at 0.5 and 3 hours after germination, respectively. Fig. 5A and B show two growth records about 0.5 hours after germination. Growth rates fluctuated between 0.2 and 0.1 µm/second for the two tubes, with an average growth rate of 0.17±0.01 µm/second (n=6). No clearly defined period can be obtained for these early recordings. However, the growth records taken at 3 hours after germination clearly show periodic growth (Fig. 5C,D). Growth rates fluctuated between 0.34 and 0.08 µm/second. The mean period of growth pulses for these tubes was 42±2 seconds (n=8) with a range of 38 to 55 seconds. The average growth rates for these longer tubes was 0.18±0.01 µm/second. Fluorescence imaging We have determined the phase relationship between the tip localized Ca2+ pulses and pulsatile growth by imaging Calcium 150 0 50 100 150 200 Time (seconds) Green dextran in elongating pollen tubes between 2.5 and 4 hours after germination. Growth rates and calcium pulses were measured from the same fluorescent images. Data collected from two recordings of a single tube are shown in Fig. 6A. The temporal resolution for these data is 5 seconds and the relative Calcium Green fluorescence data is plotted on the 5 second interval starting at 0 seconds. The growth rate data, however, were calculated as the distance a tube grew between images, so these data have been plotted at the time point half-way between successive images, starting at 2.5 seconds and incremented every 5 seconds. The growth rate and Calcium Green relative fluorescence data both appear oscillatory with similar periods and closely related phasing. The phase relationship was studied by comparing the time difference between the minimum and maximum extreme points of growth pulses and tip localized Calcium Green fluorescence. We found that the Calcium Green fluorescence extremes lagged the growth rate extremes by 3.8±0.9 seconds (136 extremes from 4 tubes). This is below our temporal resolution of 5 seconds, so the Ca2+ pulses, within our limits of resolution, are coincident with growth pulses. In order to determine that the Calcium Green fluorescence changes were not simply due to the changing shape of the pollen tube tip, we imaged fluorescein dextran as a control. Fig. 6B shows data from two recordings of a single pollen tube. Growth rates were still oscillatory but the fluorescein signal showed only small differences in intensity, compared to the Calcium Green signal. The small oscillations in the fluorescein intensity were out of phase with the growth pulses. We compared the growth maximum and minimum extremes with the nearest fluorescein minimum and maximum extremes, respectively, and found that they differ only by 3.9±0.9 seconds (74 extremes for Ca2+ pulses and pollen tube elongation 1275 Growth Rate Calcium Green Fluorescence A 0.32 1.24 Growth Rates (µm/s) 0.29 1.20 0.26 1.16 0.23 1.12 0.20 1.08 0.17 1.04 0.14 1.00 0.11 0.08 50 0 100 150 0 100 50 150 Relative Calcium Green Fluorescence 0.35 0.96 Time (seconds) B Growth Rate Fluorescein Fluorescence 0.35 0.32 0.29 Growth Rates (µm/s) 1.20 0.26 1.16 0.23 1.12 0.20 1.08 0.17 1.04 0.14 1.00 0.11 0.08 0 50 100 150 200 0 50 Time (seconds) 3 tubes). Not all of the growth extremes were accompanied by a change in fluorescein fluorescence. Those extremes were excluded from the preceding analysis. Growth extremes are a half-cycle out of phase with fluorescein oscillations. We have noticed periodic distortions in the growing tips. When tubes grew out rapidly, the tip was narrower and when the growth slowed, the tube thickened. These periodic distortions leading to a changing level of fluorescein in the optical path may explain the weak fluorescein oscillations. These same factors will apply to the Calcium Green data; the shape-induced oscillations are 100 150 0.96 Relative Fluorescein Fluorescence 1.24 Fig. 6. Simultaneous growth and tip fluorescence of a Calcium Green dextran injected pollen tube (A) and a fluorescein dextran injected tube (B). Two recordings for each dye are plotted and have their own time scales. The growth rate and relative fluorescence scales, plotted on the yaxis, are identical for both dyes. (A) Tip localized Calcium Green fluorescence oscillates with large intensity changes and is in phase with pulsatile tip growth. (B) Tip fluorescein fluorescence oscillates with weak intensity changes and is out of phase with pulsatile tip growth. out of phase with the Ca2+ oscillations. Correcting the Calcium Green results for the shape-induced changes in fluorescence would increase the amplitude of the Ca2+ oscillations. Alternatively, the fluorescein fluorescence oscillations may represent weak pH oscillations as fluorescein’s emission decreases with decreasing pH. Fig. 7 shows the images that were used to generate the final growth pulse/calcium pulse shown in Fig. 6A (arrow). The middle image (#2) was used to generate the value at the peak of the Calcium Green fluorescence curve while the images 1276 M. Messerli and K. R. Robinson for the fluorescein injected tube. The #2 profile for fluorescein is the profile for the first of the two images used to calculate the peak growth rate for the pulse labeled in Fig. 6B (arrow). Profiles #1 and #3 are the profiles of the images taken 20 seconds before and after the #2 profile. A reference line was drawn at a pixel intensity of 200. The peak fluorescence was located a few microns further from the tip for fluorescein than for Calcium Green and the fluorescein intensity changes only slightly for the different images both among different tubes and within the same tube (n=3). Uninjected tubes were not detected by the imaging system even at maximum gain, indicating that autofluorescence was insignificant. 1 Calcium Green Fluorescence DISCUSSION 2 3 0 10 µm Fig. 7. Time-lapse images of a Calcium Green dextran-injected pollen tube. Images are 10 seconds apart and were used to generate the indicated pulse shown in Fig. 6A (arrow). Each of these images was filtered with a low-pass filter and the pixel intensities below 20 were masked to reduce background noise. The tip localized fluorescence intensity rises between the first image and the middle image, and then falls during the last two images. Growth between the first two images and the last two images is lower than the growth between the second and third and between the third and fourth images. above were taken 10 and 20 seconds before and the images below were taken 10 and 20 seconds afterward. The profiles of the labeled images in Fig. 7 are shown in Fig. 8A. These profiles not only show the growth of the tube’s leading edge but give an idea of the relative fluorescence down the tube. A reference line was drawn at a pixel intensity of 175 to facilitate comparison with the tip localized fluorescence. Profile #2, the peak of the calcium pulse, clearly shows a more intense fluorescence in the tip than images taken 20 seconds before (#1) and 20 seconds afterward (#3). Fig. 8B shows similar profiles The data presented here establish that the cytosolic Ca2+ in growing Lilium longiflorum pollen tubes is pulsatile, with large pulses of increased Ca2+ superimposed on a steady background. During the early stages of growth, the pulses are sporadic, but as the tubes elongate, the pulses become quite regular. We find that pollen tube growth is similarly pulsatile, with the growth pulses being irregular initially but becoming large and regular in longer tubes. Both growth pulses and Ca2+ pulses continue until growth ceases, typically several hours after germination; thus pulsing is the dominant mode of growth. If our in vitro results can be extrapolated to the in vivo situation, then pulsatile growth accompanied by Ca2+ pulses would be the mode of growth during nearly all of the tube’s trek from the stigma to the ovary. Our data further establish that the changes in Ca2+ are in phase with the changes in growth rate; that is, the time of fastest growth corresponds to the time of highest Ca2+ and slowest growth occurs when Ca2+ is lowest. The method that we have used for detecting Ca2+, photon counting of aequorin luminescence by means of a lens-less system, allows us little spatial resolution of the dynamic changes in Ca2+ that occur during growth. Nevertheless, we are confident that the Ca2+ pulses that we detect occur at the growing tip of the pollen tube. The ratiometric fluorescence images of calcium distribution indicate that the only regions of elevated Ca2+ in growing lily pollen tubes are at the tip (Miller et al., 1992; Pierson et al., 1996; Rathore et al., 1991). The limited spatial resolution that we were able to attain, by positioning the distal 1/3 of the tube off the photocathode, showed that the Ca2+ signal from the remainder of the tube and the grain was unchanging and small. A major advantage of the aequorin method is the 2.2 power dependence of light output on Ca2+ concentration. This non-linearity greatly amplifies increases in Ca2+, which in turn allows for better temporal resolution. The analysis of the shape of the pulses shows that they are asymmetric about the peak, with the rising component being slower than the falling component. Initially, Ca2+ rises in a nearly linear manner, until it becomes about 1.5 times the basal level. It then increases in a non-linear manner, perhaps exponentially, suggesting a regenerative process. The falling phase is also non-linear, with an identical time course to the nonlinear part of the rise in Ca2+. The periodicity and amplitudes of the Ca2+ pulses reflect the variations in the rate of pollen tube elongation. In shorter tubes, the Ca2+ pulses are of low amplitude and sporadically distrib- Ca2+ pulses and pollen tube elongation 1277 Pixel Intensity Fig. 8. Fluorescence profiles A B Calcium Green Fluorescein 250 250 1 from images taken from a 1 200 200 Calcium Green dextraninjected tube (A) and a 150 150 fluorescein dextran-injected 100 100 tube (B). The profiles are from the images used to generate the 50 50 arrow-labeled pulses in Fig. 0 0 6A,B. The #2 profiles represent 0 10 20 30 40 50 0 10 20 30 40 50 60 the peak of the growth pulses 250 250 while the #1 and #3 profiles 2 2 represent images taken 20 200 200 seconds before and 20 seconds 150 150 after the image used to generate profile #2. Arrows 100 100 indicate the leading edge of 50 50 growth. (A) A reference line 0 0 has been drawn at pixel 0 10 20 30 40 50 0 10 20 30 40 50 60 intensity 175. The peak intensities start at the front of 250 250 3 the tube and are changing, with 3 200 200 the middle profile having the highest intensity. (B) The peak 150 150 intensities are located a few 100 100 microns behind the tip and do not change significantly. The 50 50 peak fluorescence intensity is 0 0 more similar in magnitude to 0 10 20 30 40 50 0 10 20 30 40 50 60 the fluorescence intensity Imaged Length (µm) further back in the tube than in the Calcium Green profiles. Profiles were generated by drawing a line 20 pixels wide on top and parallel to the long axes of the images of the tubes. The 20 pixels were averaged to return a single value to each point on the line. The data for each tube were saved to a spreadsheet and graphed. uted; growth pulses are also small and irregular in shorter tubes. The increase in the size and regularity of the Ca2+ pulses in longer tubes is accompanied by a similar change in the growth pulses. The mean period of the Ca2+ pulses, 40 seconds with a range of 29 to 57 seconds, is not significantly different (P=0.5) from the period of the growth pulses, 42 seconds with a range of 38 to 55 seconds. A rough estimate of the Ca2+ concentration change due to the pulses can be made by assuming that the baseline level of light emission from a 1 mm long tube is about 2.5 photons/second (2/3 tube and grain signal minus instrumental background, Fig. 3A,B), with a basal Ca2+ level of about 200 nM (Miller et al., 1992; Pierson et al., 1996; Rathore et al., 1991). The Ca2+ pulses are localized to the distal 20 µm of the tube (Miller et al., 1992; Pierson et al., 1996) and the peaks of the unaveraged pulses are commonly 150 photons/second. This 60-fold increase in light occurs in about 1% of the tube volume; i.e. in the distal 20 µm that we model as a hemisphere connected to the 1 mm long cylindrical tube. Thus, the peak light emission per unit volume from the tip is about 6,000 times as high as that from the rest of the tube. As the light emission increases with the 2.2 power of [Ca2+], [Ca2+] at the peak of the pulses is about 50 times the basal level concentration found in the tube, or 10 µM. A similar calculation indicates that [Ca2+] at the tip is about 3 µM between pulses. While these estimates of absolute [Ca2+] levels are crude, it should be noted that the values obtained are fairly insensitive to changes in the starting assumptions due to the 2.2 power dependence. The other method that we have used to detect Ca2+ changes, scanning confocal imaging of Calcium Green fluorescence, allowed us simultaneously to measure Ca2+ and growth rates. While the temporal resolution of this method was not as good as the aequorin measurements, it was sufficient to establish the phase relationship between growth and Ca2+. Parallel experiments done with fluorescein make it clear that the changes revealed by Calcium Green are due to changes in Ca2+, and not to the amount of dye in the optical path. We regard the aequorin-detected pulses as a better representation of the temporal aspects of Ca2+ changes than the Calcium Green measurements. The aequorin measurements had a temporal resolution of 1 second, while the fluorescence measurements had a temporal resolution of 5 seconds. For this reason the Ca2+ dynamics are better described as pulses rather than oscillations. The spatial averaging and the reduced temporal resolution of the Calcium Green images made the Ca2+ appear as oscillations rather than pulses. Our results are connected in an interesting way with earlier measurements of ionic current in lily pollen. Weisenseel et al. (1975) found a steady ionic current that entered the pollen grain, predicting the site of germination. Current then entered the growing tip of the elongating tube. In some older tubes this steady current continued to enter the growing tip of the tube, but a new phenomenon appeared, which they describe as ‘an endless train of spontaneous monophasic pulses [riding] on top of the steady current’. The current pulses entered only the growing tip and continued until growth stopped. The pulses were sym- 1278 M. Messerli and K. R. Robinson metrical and had periods of about 46 seconds. We are confident that these current pulses are the electrical correlate of the calcium and growth pulses that we report. The only substantial difference is that they did not see current pulses in every case while we detected Ca2+ pulses in every pollen tube studied. Interestingly, all three quantities, growth, tip Ca2+ concentration and ionic current, have steady components on which the pulsatile components are superimposed. The Ca2+ concentration is always higher in the tip than in more proximal regions, even between the Ca2+ pulses; growth continues between growth pulses, and there is substantial current entry between current pulses. Another correlate of our Ca2+ pulses may have been detected by Jaffe et al. (1975). They exposed growing lily pollen tubes to 45Ca, then did low-temperature autoradiography on washed, frozen tubes. They always detected higher levels of 45Ca at the tip, but in some cases the tips were extremely hot, resulting in autoradiographs with high grain density at the pollen tube tips. It may be that the exposure to 45Ca was terminated during, or just following, a Ca2+ pulse in these cases. There are two other examples of oscillating changes in Ca2+ associated with definite events in higher plants. The self-incompatibility response of Papaver rhoeas pollen tubes appears to be mediated by cytosolic free Ca2+ (Franklin-Tong et al., 1993) and a slow-moving wave can be triggered in these pollen tubes by localized elevation of inositol(1,4,5)trisphosphate (FranklinTong et al., 1996). Upon reaching the growing tip, the wave of elevated calcium inhibits growth, perhaps by disrupting the Ca2+ gradient there. Calcium spiking is also induced in legume root hairs in response to nodulation factors released by Rhizobium bacteria, and the calcium response seems to be necessary for inducing the nodulation response (Ehrhardt et al., 1996). The situation in growing lily pollen is quite different. No external signal is required for growth or for the Ca2+ pulses. Furthermore, the pulses do not spread in a wave down the pollen tube, but remain localized at the tip. Entry of external calcium is necessary for both growth and Ca2+ pulses. At this point, we do not know if the Ca2+ pulses also involve the release of Ca2+ from internal stores. The La3+ data indicate that at least some component of the Ca2+ signal involves Ca2+ influx. In some ways, the more analogous situation is that of neurotransmitter release. In both pollen and neurons, there is a massive, calcium-regulated exocytosis of vesicles that must be restricted spatially. In the case of neurons, action potentials open voltage-gated calcium channels allowing Ca2+ influx into the presynaptic cell. The interaction of Ca2+ with neuronal proteins leads to the secretion of neurotransmitter-containing vesicles. The proteins involved in this process are beginning to be understood, and it is clear that yeast and non-neuronal secretory cells have homologous proteins (Calakos and Scheller, 1996). It may be that similar proteins are involved in transducing Ca2+ regulated growth in pollen. We thank Theresa Kirk of Purdue University, Horticulture Department, for supplying the pollen for this work. We also thank J. Paul Robinson and Huw S. Kruger Gray of the Purdue Cytometry Laboratory for their help with the scanning confocal imaging. This research was funded by the US Department of Agriculture (95-37304-2222). REFERENCES Blinks, J. R., Wier, W. G., Hess, P. and Prendergast, F. G. (1982). Measurement of Ca2+ concentrations in living cells. Prog. Biophys. Mol. Biol. 40, 1-114. Brewbaker, J. L. and Kwack, B. H. (1963). The essential role of calcium ion in pollen germination and pollen tube growth. Am. J. Bot. 50, 859-865. Calakos, N. and Scheller, R. H. (1996). Synaptic vesicle biogenesis, docking, and fusion: a molecular description. Physiol. Rev. 76, 1-29. Cork, R. J., Cicirelli, M. F. and Robinson, K. R. (1987). A rise in cytosolic calcium is not necessary for maturation of Xenopus laevis oocytes. Dev. Biol. 121, 41-47. Ehrhardt, D. W., Wais, R. and Long, S. R. (1996). Calcium spiking in plant root hairs responding to Rhizobium nodulation signals. Cell 85, 673-681. Franklin-Tong, V. E., Ride, J. P., Read, N. D., Trewavas, A. J. and Franklin, F. C. (1993). The self-incompatability response in Papaver rhoeas is mediated by cytosolic free calcium. Plant J. 4, 163-177. Franklin-Tong, V. E., Drøbak, B. K., Allan, A. C., Watkins, P. A. and Trewavas, A. J. (1996). Growth of pollen tubes of Papaver rhoeas is regulated by a slow-moving calcium wave propagated by inositol 1,4,5trisphosphate. Plant Cell 8, 1305-1321. Jaffe, L. A., Weisenseel, M. H. and Jaffe, L. F. (1975). Calcium accumulations within the growing tips of pollen tubes. J. Cell Biol. 67, 488492. Keating, T. J., Cork, R. J. and Robinson, K. R. (1994). Intracellular free calcium oscillations in normal and cleavage-blocked embryos and artificially activated eggs of Xenopus laevis. J. Cell Sci. 107, 2229-2237. Kiehart, D. P. (1982). Microinjection of echinoderm eggs; apparatus and procedures. Meth. Cell Biol. 25, 13-33. Lancelle, S. A. and Hepler, P. K. (1992). Ultrastructure of freeze-substituted pollen tubes of Lilium longiflorum. Protoplasma 167, 215-230. Larson, D. A. (1965). Fine-structural changes in the cytoplasm of germinating pollen. Am. J. Bot. 52, 139-154. López-Franco, R., Bartnicki-Garcia, S. and Bracker, C. E. (1994). Pulsed growth of fungal hyphal tips. Proc. Nat. Acad. Sci. USA 91, 12228-12232. Malhó, R., Read, N. D., Pais, M. S. and Trewavas, A. J. (1994). Role of cytosolic free calcium in the reorientation of pollen tube growth. Plant J. 5, 331-341. Miller, D. D., Callaham, D. A., Gross, D. J. and Hepler, P. K. (1992). Free Ca2+ gradient in growing pollen tubes of Lilium. J. Cell Sci. 101, 7-12. Miller, A. L., Karplus, E. and Jaffe, L. F. (1994). Imaging [Ca2+]i with aequorin using a photon imaging detector. Meth. Cell Biol. 40, 305-338. Picton, J. M. and Steer, M. W. (1985). The effects of ruthenium red, lanthanum, fluorescein isothiocyanate and trifluoperazine on vesicle transport, vesicle fusion, and tip extension in pollen tubes. Planta 163, 20-26. Pierson, E. S., Miller, D. D., Callaham, D. A., Shipley, N. M., Rivers, B. A. and Hepler, P. K. (1994). Pollen tube growth is coupled to the extracellular calcium ion flux and the intracellular calcium gradient: effect of BAPTA-type buffers and hypertonic medium. Plant Cell 6, 1815-1828. Pierson, E. S., Miller, D. D., Callaham, D. A., van Aken, J., Hackett, G. and Hepler, P. K. (1996). Tip-localized calcium entry fluctuates during pollen tube growth. Dev. Biol. 174, 160-173. Rathore, K. S., Cork, R. J. and Robinson, K. R. (1991). A cytoplasmic gradient of Ca2+ is correlated with the growth of lily pollen tubes. Dev. Biol. 148, 612-619. Rosen, W. G., Gawlik, S. R., Dashek, W. V. and Siegesmund, K. A. (1964). Fine structure and cytochemistry of Lilium pollen tubes. Am. J. Bot. 51, 6171. Shimomura, O., Musicki, B., Kishi, Y. and Inouye, S. (1993). Light-emitting properties of recombinant semi-synthetic aequorins and recombinant flourescein-conjugated aequorin for measuring cellular calcium. Cell Calcium 14, 373-378. Weisenseel, M. H. and Jaffe, L. F. (1976). The major growth current through lily pollen tubes enters as K+ and leaves as H+. Planta 133, 1-7. (Received 24 January 1997 – Accepted 11 March 1997)
© Copyright 2026 Paperzz