Cardiovascular Research 39 (1998) 224–232 The atherosclerotic Yucatan animal model to study the arterial response after balloon angioplasty: the natural history of remodeling B.J.G.L. de Smet a ,b , J. van der Zande a ,b , Y.J.M. van der Helm a ,b , R.E. Kuntz c , C. Borst a , c, M.J. Post * a Department of Cardiology, Utrecht University Hospital, Utrecht, Netherlands Interuniversity Cardiology Institute of the Netherlands, Utrecht, Netherlands c Cardiovascular Division, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, MA, USA b Received 8 October 1997; accepted 20 January 1998 Abstract Objective: Remodeling in de novo atherosclerosis and in restenosis after balloon angioplasty constitutes a change in total arterial circumference which, together with plaque growth or neointimal formation, determines the lumen of the artery. To better understand the fundamental biology of neointimal formation, remodeling and their interaction, animal studies are needed. In this study, we described in detail the methodology used and the natural history of neointimal formation and remodeling after balloon angioplasty in atherosclerotic Yucatan micropigs. Methods and results: Atherosclerosis was induced in 60 peripheral arteries of sixteen Yucatan micropigs by a combination of denudation and atherogenic diet. Balloon angioplasty was performed in 38 arteries, with serial intravascular ultrasound (IVUS) and quantitative angiography before and after intervention and at 2, 4, 7, 14 or 42 days follow-up. Remodeling, expressed as late media-bounded area (MBA) loss, increased progressively over time. At 42 days, late MBA loss after balloon angioplasty was significantly different compared to late MBA loss in control arteries, 2.261.0 versus 20.361.1 mm 2 and p50.02. Late lumen loss increased over time and was highest at 42 days after balloon angioplasty (2.860.7 mm 2 ). The contribution of neointimal formation to late lumen loss decreased over time and the contribution of late MBA loss to late lumen increased over time and was highest at 42 days (78%). Medial necrosis was 48% at two days after balloon angioplasty and the repopulation of the media was almost completed at seven days. Conclusion: Remodeling following balloon angioplasty has an early onset and progresses with neointimal formation to cause restenosis over the standard 42-day time course for Yucatan micropigs. This correlates to six months renarrowing in humans. In this model, atherosclerosis and the natural history of restenosis, both with respect to neointimal formation and remodeling, resemble the human disease quite closely. 1998 Elsevier Science B.V. All rights reserved. Keywords: Experimental (discipline); Vasculature (object of study); Circulatory physiology (level); Remodeling; Angioplasty; Ultrasound; Arteries; Atherosclerosis (additional) 1. Introduction Coronary restenosis after balloon angioplasty, defined as arterial lumen renarrowing of the treatment site, has long been considered to be primarily the consequence of excessive neointimal formation. Recently, however, human [1,2] and animal studies [3–5] have shown that, in addition to neointimal formation, changes in the total arterial circumference (so-called geometric remodeling) also occur *Corresponding author: Tel.: 617 667 3790; Fax: 617 975 5201; E-mail: [email protected] 0008-6363 / 98 / $19.00 1998 Elsevier Science B.V. All rights reserved. PII: S0008-6363( 98 )00085-6 in response to arterial intervention. To better understand the fundamental biology of neointimal formation, geometric remodeling and their interaction, animal studies are needed. Preferably, such an animal model should have arteries with atherosclerotic plaques that bear maximum similarity to the human disease, and should illustrate the two paradigms of restenosis after balloon angioplasty, neointimal formation and remodeling. Small atherosclerotic animal models, especially in the mouse, should be developed for future transgenic and knockout approaches Time for primary review 27 days. B. J.G.L. de Smet et al. / Cardiovascular Research 39 (1998) 224 – 232 to address specific questions in the vascular biology of restenosis. In addition, a large animal model will remain necessary to study new recanalization devices that are usually designed according to human measures, and to evaluate local drug delivery strategies with specialized catheters. Sound and fastidious qualitative and quantitative analysis of atherosclerosis and neointimal formation and remodeling after balloon angioplasty is a second requisite of a good animal model. We have discussed previously that serial intravascular ultrasound is the best way to assess remodeling after balloon angioplasty [6]. The current size of intravascular ultrasound catheters also necessitates the use of large animals. We developed an atherosclerotic model in the Yucatan micropig [7] that meets most of these conditions. We found that the morphology of the atherosclerotic plaque covers the whole range from hypercellular fibromuscular plaque to lipid lakes that are covered by a fibrous cap and complex lesions with calcification, cholesterol clefts and signs of continuous inflammation and prothrombotic tendency. After balloon angioplasty, 70% of restenosis is attributed to remodeling and 30% to neointimal formation, which fits nicely with the reported numbers for human coronary restenosis [1]. In contrast, neointimal formation is the sole cause of restenosis in stented arteries. The time course of remodeling after angioplasty has not been described yet. Results from the SURE trial with serial intravascular ultrasound suggested that there was a biphasic remodeling response of the coronary artery, with enlargement at one month and shrinkage at six months after balloon angioplasty [8]. Our aim was to study the possible differences in the time course of neointimal formation and remodeling in this animal model. In this manuscript, we describe in detail the methodology used and the natural history of neointimal formation and remodeling after balloon angioplasty in the atherosclerotic peripheral arteries of Yucatan micropigs. 225 US National Institutes of Health (NIH Publication No. 85-23, revised 1985) and was approved by the ethical committee on Animal Experiments of the Faculty of Medicine, Utrecht University. 2.1. Atherogenic diet and anesthesia In addition to essential nutrients, vitamins and salts, 1.5% cholesterol, 17.5% casein, 14% lard and 6% peanut oil formed the basic atherogenic components of the diet, which had a daily nutritional value equivalent to 2400 kcal (Hope Farms, Netherlands). In pilot experiments, it was shown that this regimen results in a sustained tenfold increase in the total cholesterol level to 14.9 (61.3) mmol / l, and an approximately 3.5-fold increase in the high-density lipoprotein (HDL) level to 2.2 (60.1) mmol / l. After discontinuation of the atherogenic diet, the levels of cholesterol and HDL cholesterol dropped to base levels (1.560.2 and 0.660.1 mmol / l, respectively). Water intake was not restricted. The diet during follow-up was a regular, non-atherogenic chow, with a nutritional value that was equal to that of the atherogenic diet. For denudation, intervention and termination, the animals were anesthetized with intravenous metomidate (4 mg / kg) and ventilated (Servo, EM 902) with a mixture of O 2 –N 2 O (1:2) and halothane, 1–2%. During each procedure, the animals were heparinized, starting with 100 IU / kg thromboliquine (Organon Technika), and then titrating to an activated partial thromboplastin time (APTT) above 80 s throughout the procedure. Every 15 min, 0.25 mg of atropine was given intravenously. One day before intervention, acetyl salicylic acid (125 mg) p.o. was given and administration was continued for two weeks after the intervention. During intervention and termination, a continuous infusion of nitroglycerin (20 mg / min) was given to prevent arterial spasm. 2.2. Procedures 2. Methods Sixteen Yucatan micropigs with an average weight of 24 kg at termination were used for this study. All animals were started on an atherogenic diet at the age of seven– eight months. Two weeks thereafter, they underwent Fogarty denudation of the internal iliac, external iliac and femoral arteries. The atherogenic diet was continued for an additional four months, after which, selected arteries were balloon dilated with a standard angioplasty balloon catheter, and the atherogenic diet was replaced by a regular diet. The immediate and long term results of balloon angioplasty were documented by quantitative angiography and by serial intravascular ultrasound. After 2, 4, 7, 14 or 42 days, the animals were killed and the arteries were harvested for histology. The investigation conforms with the Guide for the Care and Use of Laboratory Animals published by the For denudation, intervention and termination, the arterial tree was accessed through a carotid cutdown and an arterial 8F sheath was inserted into the descending aorta under fluoroscopic guidance. An 8F guiding catheter was advanced to the aortic bifurcation. Through this, contrast (Telebrix, Laboratoire Guerbet, France) angiography was performed and the Fogarty balloon catheter, the intravascular ultrasound catheter and the balloon angioplasty catheter were advanced. After the denudation and the intervention, the carotid arteries were carefully sutured for future interventions. During denudation, 3–4 cm segments (measured by a radiopaque ruler) in the iliac and femoral arteries were denuded by triple withdrawal of a 4F Fogarty catheter that was manually inflated with a 50:50 (v / v) water–contrast mixture. Four months after the denudation, selected sites of 226 B. J.G.L. de Smet et al. / Cardiovascular Research 39 (1998) 224 – 232 angiographic arterial narrowing were balloon dilated. Before and after each intervention, angiography and intravascular ultrasound were performed under fluoroscopic guidance. For balloon dilation, a standard peripheral (length, 2–4 cm; diameter, 4–7 mm) or coronary (length, 2 cm; diameter, 2–4 mm) balloon catheter was advanced over a 0.03599 or a 0.01499 guidewire. The balloon was inflated three times for 1 min at a pressure of 10 atm. Of all of the arteries available, 21 arteries were used for another study that did not interfere with this study regarding drug therapy or repeated catheter handling. After a follow-up of 2, 4, 7, 14 or 42 days, the pigs were anesthetized and angiograms and intravascular ultrasound measurements were made. The animals were killed after angiography and intravascular ultrasound was performed. 2.3. Angiography and intravascular ultrasound Angiography was performed before and after each intervention and at follow-up. Contrast (Telebrix) was injected selectively into the artery under study through an 8F guiding catheter. The fluoroscopy was recorded at a cine rate of 12 images / s using a digital C-arm (Philips, Eindhoven, Netherlands). The image with the highest contrast was selected and stored on DAT tape for later analysis. The angiographic diameters of the arteries were measured using a semi-automated program [7]. In each artery, lumen diameters were measured at intervals of 0.5 cm, including the treated segment and a proximal and distal reference segment (RLD). Reference segments were chosen in an angiographically normal adjacent segment at a distance of at least 1 cm proximal and distal from the balloon-dilated site. To locate the treated segment repeatedly at different time points (pre- and post-procedure and at follow-up), its position was documented relative to an anatomic landmark with the support of a radiopaque ruler placed along the spine in a reproducible manner. The ruler was also used for calibration of the angiography [7]. All quantitative analyses were performed off-line. The minimum lumen diameter (MLD) of the lesion was determined. Angiographic acute gain was defined as the difference between post- and pre-procedure lumen diameters, and angiographic late lumen loss was defined as the difference between post-procedure and follow-up lumen diameters. Angiographic percentage of stenosis was calculated as (12MLD/ RLD)3100. Dilation ratio was defined as the diameter of the inflated balloon on fluoroscopy divided by the angiographic diameter of the reference segment. Intravascular ultrasound recordings were made before and after intervention and at follow-up, using a 30-MHz ultrasound transducer (Du-MED, Rotterdam, Netherlands), which rotated up to 16 times per second within a 4.1 French catheter. The axial resolution of the system was 0.1 mm. The images were displayed on a monitor and recorded on VHS video-tape (Fig. 1). Intravascular ultrasound Fig. 1. Representative serial IVUS images of an atherosclerotic lesion (A) before balloon angioplasty, (B) immediately after balloon angioplasty and (C) at 42 days follow-up. M5media-bounded area; the distance between indicators is 1 mm. B. J.G.L. de Smet et al. / Cardiovascular Research 39 (1998) 224 – 232 images were analyzed with a digital video analyzer, as described previously [9]. Fluoroscopy was performed during intravascular ultrasound (IVUS) so that the IVUS images were documented relative to an anatomic landmark, to match the IVUS images at different time points (pre- and post-procedure and at follow-up). In the IVUS images, the area circumscribed by the interface between the echodense intimal layer and the echolucent media was manually traced and taken as the media-bounded area (MBA). In addition, the lumen area (LA) was traced and the minimum lumen area was determined (MLA). Echographic acute gain was defined as the difference between the post- and pre-intervention lumen area, and echographic late lumen loss as the difference between post-intervention and follow-up lumen area. In addition to late lumen loss, late MBA loss, being a measure of remodeling after angioplasty, was introduced. Late MBA loss is defined as the difference between the post-intervention MBA and the follow-up MBA at either the site of the initial stenosis or the reference sites. Intimal area was taken as the difference between MBA and the lumen area. At follow-up, the intima is the sum of plaque and neointimal formation. Neointimal formation is therefore calculated as the difference between the intima at follow-up and the intima before intervention. Echographic percentage of stenosis is calculated as (12MLA / RLA)3100. The loss index was defined as the coefficient of the regression with echographic late lumen loss as the dependent variable and echographic acute gain as the independent variable. 227 buffer (pH 7.4) for 2 h, followed by overnight storage in sterile 15% sucrose–phosphate-buffered saline (PBS) solution at 48C. After fixation, the arteries were divided into segments of 3–4 mm and embedded in optimal cutting temperature compound (OCT) and frozen in liquid nitrogen. Serial cross-sections were either stained with hematoxylin and eosin or with elastin van Gieson stain. Three cross-sections per artery were studied to evaluate medial necrosis and the occurrence of thrombosis. In hematoxylin and eosin stained cross-sections, medial necrosis was quantitatively analyzed in the following fashion. Four fields per cross-section were chosen in clockwise order (at 3, 6, 9 and 12 h). In each field, the total number of cell nuclei in the media was counted. Cell counting was performed at 3200 magnification using a video microscope that projected the image onto a visual display unit. Nuclei were counted on the screen using a computer ¨ program, analySIS (Soft-Imaging Software, Munster, Germany), which allowed tagging of counted nuclei. The percentage medial necrosis was calculated as: (12number of cell nuclei in dilated artery / average number of cell nuclei of all control arteries)3100, assuming no medial necrosis in control arteries. Cell density was defined as the number of cell nuclei in the media per mm 2 . Thrombosis was evaluated at two days after balloon angioplasty in both hematoxylin and eosin stained and elastin van Gieson stained cross-sections. For this purpose, three cross-sections per artery, at 5 mm intervals, were evaluated. 2.5. Statistical analysis 2.4. Histology The arteries were removed ‘en block’ and fixed by immersion in 4% paraformaldehyde in 0.1 mol / l NaPO 4 All data in text and table are presented as mean6standard error of the mean. SPSS 6.1 was used for all statistical calculations. To identify differences between Fig. 2. Representative serial angiographic images of an iliac artery (A) before balloon angioplasty, (B) the balloon, (C) immediately after balloon angioplasty and (D) at 42 days follow-up. Bar indicates 1 cm. 228 B. J.G.L. de Smet et al. / Cardiovascular Research 39 (1998) 224 – 232 Fig. 3. Representative histological cross-sections of control iliac arteries (elastin van Gieson stain) (A), an eccentric lesion with calcifications (C) and a fibrocellular intima (F), (B) a concentric lesion with lipid accumulation (L) and a thin fibrous layer (F). I5internal elastic lamina, E5external elastic lamina. Bar indicates 1 mm. Note that the arteries are not perfusion fixed and are frozen sections, paraformaldehyde-fixed. B. J.G.L. de Smet et al. / Cardiovascular Research 39 (1998) 224 – 232 subgroups, we used one-way ANOVA and Duncan’s range test. Pearson correlation coefficients were calculated when indicated. 3. Results Balloon angioplasty was performed in 38 arteries in 16 Yucatan micropigs. Twenty-two arteries were not balloon dilated and served as control arteries. Complete IVUS and angiographic imaging were performed in all arteries. An example of the serial IVUS images and serial angiography is shown in Figs. 1 and 2. The analyses in this study were based on 34 / 38 balloon-dilated arteries where a positive echographic acute gain was achieved. The negative gain in the remaining arteries resulted from spasm after balloon angioplasty, which potentially influences the relation between acute gain and late loss, and introduces a group of post-angioplasty arteries that are enlarged, but not necessarily remodeled. The four excluded arteries were comparable before balloon angioplasty to the arteries that were successfully dilated. There were no statistically significant differences between the excluded and included arteries in parameters such as angiographic and echographic reference diameter, angiographic and echographic percentage stenosis and dilation ratio. The average echographic stenosis was 2862%. The echographic acute gain was 5.760.8 mm 2 . Acute gain and the dilation ratio are considered to reflect the severity of injury imparted on the arterial wall. The mean pre-existing 229 plaque area was 1.860.2 mm 2 . Representative histologic cross-sections show the heterogeneous nature of the atherosclerotic plaques, varying from a concentric lesion with accumulations of lipids covered by a fibrocellular layer (Fig. 3A) to an eccentric lesion with calcifications (Fig. 3B). The echographic and angiographic percentage of stenosis, echographic and angiographic acute gain and the dilation ratio were not statistically different at the different time points (Tables 1 and 2). Lumen areas, MBAs and lumen diameters before balloon angioplasty are given in Table 1 for both the site of stenosis and the reference site. 3.1. Remodeling and neointimal formation after balloon angioplasty over time We used late MBA loss as a measure of remodeling after balloon angioplasty. Late MBA loss increased progressively over time (Fig. 4A). Even by seven days, we observed an increase in late MBA loss. At 42 days, late MBA loss after balloon angioplasty was significantly different compared to late MBA loss in control arteries, 2.261.0 versus 20.361.1 mm 2 and p50.02. Echographic and angiographic late lumen losses increased over time and were highest at 42 days after balloon angioplasty (2.860.7 mm 2 and 0.360.2 mm, respectively). The contribution of late MBA loss and neointimal formation to late lumen loss is shown in Fig. 4A. At 42 days, echographic late lumen loss correlated with late MBA loss (r50.86, p50.03) but not with neointimal formation (r5 20.17, p50.75). The relative contribution of late MBA Table 1 Angiographic parameters Number of days after balloon angioplasty Number of arteries MLD pre (mm) LD reference pre (mm) Stenosis angio (%) Acute gain angio (mm) Dilation ratio 2 4 7 14 42 Control 8 2.460.2 3.560.3 30.766.8 0.860.2 1.260.1 5 2.560.3 3.060.1 15.7610.9 1.060.4 1.460.4 7 2.460.2 2.960.2 18.764.0 0.960.3 1.460.3 8 2.760.2 3.460.1 21.965.9 0.860.2 1.260.1 6 2.360.3 3.260.2 29.067.7 0.860.2 1.260.0 22 2.660.1 3.260.1 15.162.4 Data are presented as the mean6SEM, pre5before balloon dilation, MLD5minimum lumen diameter and LD5lumen diameter. Table 2 Echographic parameters and loss index Number of days after balloon angioplasty Number of arteries MLA pre (mm 2 ) MBA pre (mm 2 ) LA reference pre (mm 2 ) MBA reference pre (mm 2 ) Stenosis IVUS (%) Acute gain IVUS (mm 2 ) Loss index 2 4 7 14 42 Control 8 7.660.5 9.460.6 12.562.2 13.762.1 29.969.6 5.161.8 20.160.06 5 6.160.3 7.560.8 11.062.6 11.462.6 33.9612.1 5.661.6 0.1460.08 7 6.060.9 7.461.2 8.562.0 8.761.9 26.767.9 7.162.1 0.2260.1 8 9.861.0 12.361.5 13.361.4 13.561.6 28.065.7 4.661.5 0.2660.22 6 7.061.2 8.961.4 11.262.0 12.062.0 34.768.1 6.361.5 0.4260.08 22 7.460.7 8.260.7 9.861.0 10.561.0 24.562.5 Data are presented as the mean6SEM, pre5before balloon dilation, MLA5minimum lumen area, LA5lumen area and MBA5media-bounded area. 230 B. J.G.L. de Smet et al. / Cardiovascular Research 39 (1998) 224 – 232 Fig. 5. Cell density after balloon angioplasty over time. Note that zero days are the control arteries. *5p,0.05 (ANOVA) and compared with 0, 7, 14 and 42 days. Fig. 4. Panel A: The contribution of remodeling (REM) and neointimal formation (NI) to late lumen loss (LL) over time after balloon angioplasty, as assessed by IVUS. Panel B: The contribution of remodeling (REM) and neointimal formation (NI) to late lumen loss (LL) in control arteries on different days of follow-up, as assessed by IVUS. loss to late lumen loss increased over time and was 79% at 42 days. The relative contribution of neointimal formation to late lumen loss decreased over time and was 21% at 42 days. The loss index increased from 20.1060.06 at two days follow-up to 0.4260.08 at 42 days follow-up. Two days after balloon angioplasty there was a tendency towards a negative late lumen loss. However, this was not significant compared to control arteries. For control arteries, the relative contributions of late MBA loss and neointimal formation are shown in Fig. 4B. Late lumen loss, late MBA loss or neointimal formation in the control arteries were not significantly different from zero at the different time points. 3.2. Thrombus formation and medial necrosis Two days after balloon dilation, thrombi were seen in six out of eight arteries. Only in one artery was a medium size thrombus present in all cross-sections, covering a large part of the luminal surface without compromising the lumen area. In five arteries, the thrombi were minor and present in only one of three cross-sections. In those five arteries, the thrombi were only present at medial tears, filling up the crevices in the tissue. The thrombi did not protrude into the lumen. Medial necrosis at two days was 48% and at four days, it was 45%. The cell densities at two and four days were significantly different (ANOVA, p,0.05) from the cell density at 7, 14 and 42 days and from the cell density in the control arteries (Fig. 5). The cell densities at 7, 14 and 42 days were not significantly different from each other and also were not significantly different from the cell density in control arteries. However, the cell density at 7 and 42 days was approximately 79% of the cell density in the control arteries (29346174 and 28896134, respectively, versus 36696188). At 14 days, the cell density was 92% of the cell density in the control arteries (33646168 versus 36696188). Medial necrosis was patchy and asymmetrically distributed over the circumference of the cross-section. 4. Discussion Remodeling after balloon angioplasty has been recognized as a major determinant of restenosis. After balloon angioplasty, shrinkage seems to be the predominant, although certainly not the only, mode of remodeling. Post-dilation remodeling, therefore, usually leads to a reduction in lumen size [1–5] that is additional to the lumen encroachment by neointimal formation. Previously, we reported that remodeling after balloon angioplasty importantly determines restenosis in this model [10], to an extent that is equivalent to that reported for human B. J.G.L. de Smet et al. / Cardiovascular Research 39 (1998) 224 – 232 coronary arteries [1]. Thus, this animal model seems to be highly suited for the study of remodeling after angioplasty. In this combined angiographic, serial ultrasound study in an atherosclerosis model in the Yucatan micropig, we studied the natural history of neointimal formation and remodeling after balloon angioplasty. Our principal finding is that remodeling following balloon angioplasty has an early onset and progresses with neointimal formation over a 42-day time course. The histology of the atherosclerotic plaque is heterogeneous and is very similar to that of advanced human atherosclerosis. However, spontaneous plaque rupture, an important source for thrombus accumulation and peripheral embolization [11], was not observed in these atherosclerotic animals. In this model, the relationship between acute gain and late loss after balloon angioplasty is almost identical to that seen in various human angioplasty trials, and we observed previously that this relationship is based on a proportional remodeling response [10]. The follow-up of 42 days was chosen because, by this time, the neointimal formation has reached a steady state [12,13]. Steele et al. [12] showed in a pig model that neointima thickness was maximal at 14 days after balloon angioplasty and did not change between 14 and 60 days after balloon angioplasty. A similar time course was observed by Groves et al. [13] in a study where neointimal area was already maximal at seven days after balloon angioplasty, without rupture of the internal elastic lamina, and was maximal at 21 days after balloon angioplasty, which had caused rupture of the internal elastic lamina. Furthermore, serial angiographic studies in humans have shown that the vast majority of restenosis occurs in the first three months after balloon angioplasty [14,15], a time period that corresponds with the 42 days in pigs. However, we cannot exclude the possibility that late lumen loss and late MBA loss will further increase after 42 days. 4.1. Relationship between neointimal formation and remodeling Although in the present study both neointimal formation and remodeling progress over time and contribute to late lumen loss, only remodeling was positively correlated with late lumen loss. The extent of neointimal formation did not correlate with remodeling [10]. This suggests that the pathogeneses of both are dissimilar and that changes in the size and probably the composition of the intima do not affect total arterial size. Lafont et al. [5] found adventitial changes to correlate with remodeling and, indeed, by its high collagen content, the adventitia is most likely involved in the remodeling process. In a previous study in this animal model [16], we showed that expression of mRNA of collagen types I and III was markedly increased seven and fourteen days after balloon angioplasty and that this expression was highest in the adventitia. This is in agreement with the results of Strauss et al. [17,18], who 231 showed an increase of total collagen synthesis one week after balloon angioplasty in a rabbit model. The repopulation of the media is almost finished at seven days. The finishing of the repopulation of the media seems to coincide with both the start of the remodeling process and the increase in collagen synthesis, which is maximal at seven days [16]. This suggests that a repopulated media is needed for the start of the remodeling process, in which the increase in collagen synthesis is likely to play an important role. 4.2. Drawbacks of the Yucatan atherosclerosis model The drawbacks of this Yucatan atherosclerosis model are shared with other large animal atherosclerosis models. The model is expensive since it requires long term housing and care of large animals that are being fed an expensive, custom-made atherogenic diet. The incubation period of four months precludes quick screening of a new device or a new drug. Furthermore, the time pattern of the development of restenosis seems to be faster in pigs than in humans. In this model, the relative contributions at 42 days of late MBA loss and neointimal formation to late lumen loss at 42 days were 79 and 21%, respectively. This is similar to the results in human coronary arteries at six months [1,2]. Also, the neointima formation appears to be more rapid in the pig than in the human. Neointima formation after balloon dilation stabilizes between two and four weeks after balloon dilation in the pig, but in humans, it increases for up to six months after balloon dilation [8]. Another possible drawback is that the average age of these pigs is approximately one year, which does not reflect the average age in the clinical situation. The micropigs are expensive to purchase, but over four months, they end up being cheaper than domestic Yorkshire pigs because of the smaller amount of feed required. Given these limitations, the atherosclerotic Yucatan model is extremely useful for fundamental research on vascular wall biology after balloon angioplasty. It can also be effectively used in the final, preclinical, stages of deviceor drug testing. 4.3. Limitations of the study This study was performed in peripheral arteries in an animal model of advanced atherosclerosis and it remains to be determined if results obtained in peripheral arteries in animal atherosclerosis models can be extrapolated to atherosclerotic human coronary arteries. Separate serial intravascular ultrasound studies by Mintz et al. [1] and by Di Mario et al. [2], however, showed similar results with respect to the relative contribution of remodeling to late lumen loss [3–5]. Although the arteries contained considerable pre-existing plaques, the echographic stenoses were moderate, with an average of 2862% and a lumen reduction of 4.060.8 232 B. J.G.L. de Smet et al. / Cardiovascular Research 39 (1998) 224 – 232 mm 2 . The average acute gain was 5.760.8 mm 2 and, thus, 30% of the acute gain was achieved through over dilation of the arteries. However, the dilation ratio we employed in this study was 1.2, which is similar to the dilation ratios in coronary arteries of humans. Moreover, we found that the observed loss index of 0.4260.08 at 42 days follow-up was also similar to values reported for clinical studies [19]. 5. Conclusion From this serial intravascular ultrasound study in an atherosclerotic Yucatan micropig model, we conclude that remodeling following balloon angioplasty has an early onset and progresses with neointimal formation to cause restenosis over the standard 42-day time course for Yucatan micropigs. This correlates to six months renarrowing in humans. In this model, atherosclerosis and the natural history of restenosis, both with respect to neointimal formation and remodeling, resemble that of the human disease quite closely. Acknowledgements This study was supported by the Netherlands Heart Foundation (grant NHS 92.365) and in part by the Interuniversity Cardiology Institute of the Netherlands. [5] [6] [7] [8] [9] [10] [11] [12] [13] [14] [15] References [1] Mintz GS, Popma JJ, Pichard AD, et al. Arterial remodeling after coronary angioplasty: a serial intravascular ultrasound study. Circulation 1996;94:35–43. [2] Di Mario C, Gil R, Camenzind E, et al. Quantitative assessment with intracoronary ultrasound of the mechanisms of restenosis after percutaneous transluminal coronary angioplasty and directional atherectomy. Am J Cardiol 1995;75:772–777. [3] Post MJ, Borst C, Kuntz RE. The relative importance of arterial remodeling compared with intimal hyperplasia in lumen renarrowing after balloon angioplasty: A study in the normal rabbit and the hypercholesterolemic Yucatan micropig. Circulation 1994;89:2816– 2821. [4] Kakuta T, Currier JW, Haudenschild CC, Ryan TJ, Faxon DP. [16] [17] [18] [19] Differences in compensatory vessel enlargement, not intimal formation, account for restenosis after angioplasty in the hypercholesterolemic rabbit model. Circulation 1994;89:2809–2815. Lafont A, Guzman LA, Whitlow PL, et al. Restenosis after experimental angioplasty: Intimal, medial, and adventitial changes associated with constrictive remodeling. Circ Res 1995;76:996– 1002. Post MJ, Pasterkamp G, Borst C. Arterial remodeling in atherosclerosis and restenosis: A vague concept of a clear phenomenon. Atherosclerosis 1995;118:S115–S123. De Smet BJGL, Kuntz RE, van der Helm YJ, et al. Relationship between plaque mass and neointimal hyperplasia after stent placement in Yucatan micropigs. Radiology 1997;203:484–488. Kimura T, Kaburagi S, Tamura T, et al. Remodeling of human coronary arteries undergoing angioplasty or atherectomy. Circulation 1997;96:475–483. Wenguang L, Gussenhoven EJ, Zhong Y, et al. Validation of quantitative analysis of intravascular ultrasound images. Int J Card Imaging 1991;6:247–253. Post MJ, De Smet BJGL, van der Helm YJ, Borst C, Kuntz RE. Arterial remodeling after balloon angioplasty or stenting in an atherosclerotic experimental model. Circulation 1997;96:996–1003. Falk E. Why do plaques rupture?. Circulation 1992;86(III):30–42. Steele PM, Chesebro JH, Stanson AW, et al. Balloon angioplasty. Natural history of the pathophysiological response to injury in a pig model. Circ Res 1985;57:105–112. Groves PH, Banning AP, Penny WJ, et al. Kinetics of smooth muscle cell proliferation and intimal thickening in a pig carotid model of balloon injury. Atherosclerosis 1995;117:83–96. Nobuyoshi M, Kimura T, Nosaka H, et al. Restenosis after successful percutaneous transluminal coronary angioplasty: serial angiographic follow-up of 229 patients. J Am Coll Cardiol 1988;12:616– 623. Serruys PW, Luijten HE, Beatt KJ, et al. Incidence of restenosis after successful coronary angioplasty: a time-related phenomenon. A quantitative angiographic study in 342 consecutive patients at 1, 2, 3, and 4 months. Circulation 1988;77:361–371. de Smet BJ, van der Helm YJ, Horowitz A, Simons M, Post MJ. Matrix remodeling in balloon dilated and stented arteries in atherosclerotic yucatan micropigs: procollagen alpha1(I), alpha1(III) and MMP1 expression over time. Circulation 1996;94:1–403. Strauss BH, Chisholm RJ, Keeley FW, et al. Extracellular matrix remodeling after balloon angioplasty injury in a rabbit model of restenosis. Circ Res 1994;75:650–658. Strauss BH, Robinson R, Batchelor WB, et al. In vivo collagen turnover following experimental balloon angioplasty injury and the role of matrix metalloproteinases. Circ Res 1996;79:541–550. Kuntz RE, Gibson CM, Nobuyoshi M, Baim DS. Generalized model of restenosis after conventional balloon angioplasty, stenting and directional atherectomy. J Am Coll Cardiol 1993;21:15–25.
© Copyright 2026 Paperzz