96-113757_1 Hydrim M2_G4_QRG_EN.indd

HYDRIM M2
®
GENERAL USAGE
1
Operator’s manual should be consulted before using this washer disinfector.
3
Load instruments
TOUCH SCREEN MENU OPTIONS
Select cycle and Press
Network Connectivity –
Red*: No internet connection
Yellow*: Unit has IP, but no internet
Green: Unit is fully connected
Rinse and Hold Cycle*
Prevents soil from drying
on instruments.
Incomplete loads that
will not be washed within an hour.
*Please contact your network administrator
USB Storage Device –
Access last 5 cycle data and
last 5 cycle faults. Indicator also
confirms proper USB connection.
*This is not a cleaning cycle
Regular Cycle
No Disinfection
Moderately soiled
loose instruments.
2
Close the door
Water Softener/Detergent Level/Door
Status – Indicator will notify user if there
is sufficient salt in water softener system,
if there is sufficient detergent to run a
cycle and if the door is properly closed.
Heavy Duty Cycle
With Disinfection
Lock/Unlock Door – Indicator
will notify user if the door is locked.
Heavily soiled instruments
and cassettes.
User/Technician Menu –
Access unit information and menus to change
unit settings.
Custom Cycle
With Disinfection
Achieve between A0 = 3,500 and A0 = 6,000
Cycle must be validated by technician before use.
ROUTINE PROCEDURES
Check your Operator’s Manual for more details
Replacing the HIP cleaning solution
™
1
2
3
4
5
6
71
2
3
MSDS is available on SciCan’s website (www.scican.com)
41 52 63 741 52 63 74
1
5
2
6
3
7
4
5
6
7
8.
9.
1
2
3
4
5
6
7
10.
1.
2.
3.
4.
5.
6.
7.
BACK
Push button
to return to
previous menu.
ENTER
Push button
to view/select
menu options.
Heavily soiled instruments
and cassettes.
11.
12.
Open the cleaning solution door (picture 1).
Pull out the empty solution container and rotate backwards into the upright position (with the connector at the top).
Disconnect the cleaning solution connector by unscrewing counter-clockwise (picture 2).
Remove and dispose of/recycle the empty container as appropriate and in accordance with local guidelines.
If appropriate, remove any dried chemical by immersing the connector in a cup of water.
Place the new container of solution on the solution door (picture 3).
Remove the tear-off portion of the outer carton. (picture 4).
Cleaning the Chamber
13.
14.
Pull out the internal chemical bag nozzle and
secure it in the cutout (picture 5a). Remove
the cap.
Connect the cleaning solution connector to the
new solution container by screwing clockwise
until it stops rotating (picture 5b). Caution –
avoid cross threading the connector as damage
may occur and the joint may leak.
Insert the new solution container by rotating
forwards into the correct location in the solution
door (picture 6).
Close the cleaning solution door.
Press OK on the touch screen. The machine
display will indicate “Priming. Please wait”.
When the cycle selection menu appears, priming
is complete and the machine is ready for use.
Proceed with the next cycle.
Refilling the Water Softener salt reservoir
Every 25 cycles, you will be reminded to run a cleaning cycle.
Pour 1.0L of vinegar or citric acid into
the chamber and follow these steps:
When the error message “salt level low” or the water softener/detergent icon
indicated by a red X appears, the salt reservoir is in need of filling.
To fill the salt reservoir proceed as follows:
1. Unscrew the salt container lid.
salt container
Cleaning
2. F
ill the salt container to the top
(maximum 1 kg / 2.2 lbs.)
3. Screw the salt container lid on tightly.
4. Run a complete cycle, with no instruments.
Important notes:
Failure to clean the chamber
can promote corrosion.
If the salt container is being filled for the first time, it MUST
be filled with at least 1 litre of water before adding the salt.
nly tablet salt should be used in the softening unit. Fine
O
grained salt such as table salt should NOT be used as it
could clog the unit.
Common Cycle Faults
Error
What it Means
CF2 Chamber Filling Water is not filling the
chamber.
What to Do
• Check that the water hoses
are not kinked
• Check that water shut-off
valves are opened
• Call for Service
CF15 Water
Overflow switch triggered. • Restart the program
Reservoir Overflow
• Call for Service
CF20 Condenser
Unit is not able to cool
• Check that the water hoses
down.
are not kinked
System
• Call for Service
CF21 Dosing System Detergent is not
• Tighten detergent cap
dispensing
• Check for leaks
• Call for Service
Spare Parts and Accessories
Description
SciCan Part Number
HIP™ Cleaning Solution
(2 x 3.8L)
Water Softener Salt
Water Test Kits
HEPA Filters
(Reminded to change every 600 cycles)
IMPORTANT NOTE:
Terminal sterilization
of instruments MUST
follow processing in
the HYDRIM M2 G4.
CS-HIPL
01-112594S
01-108305S
01-111780S
Facebook: SciCan
Twitter: @SciCan
Youtube: SciCanTV
HYDRIM is a registered trademark and HIP and Your Infection Control Specialist are trademarks of SciCan Ltd.
www.scican.com
96-113757 Rev. 1.0
Refer to Operator’s Manual for a more comprehensive list.
Call for service if suggested recommendations do not resolve
the error message.
he water softener salt is also available from SciCan.
T
Please contact your local dealer.