50436_AN94r0.qxd 1/29/07 6:57 AM Page 1 Solutions for a nanoscale world.™ Practical Advice on the Determination of Cantilever Spring Constants By: Ben Ohler, Ph.D., Veeco Instruments Inc. INTRODUCTION Atomic force microscopy (AFM) is being used in a great variety of force measurement applications 1, including investigating the unfolding pathways of native membrane proteins 2, probing the structure of single polysaccharide molecules 3, and monitoring the response of living cells to biochemical stimuli 4. All of these techniques rely on the accurate determination of the cantilever spring constant in order to yield quantitative results. For although the cantilever deflection can be measured with great accuracy and sub-Angstrom sensitivity, converting these measurements to units of force via Hooke’s law, F = –k · x, requires that the spring constant, k, be determined for each cantilever. It has often been noted in the literature that spring constants can vary greatly from the values quoted by their manufacturers. In fact, these values are only provided as nominal indications of the cantilever properties and the manufacturers often specify the spring constant in a wide range that may span values up to four times smaller and four times larger than the nominal value (cf. ref. 5). This is because the techniques used to fabricate the probes can result in substantially different cantilever dimensions, especially thickness, from wafer to wafer and smaller variations within a single wafer 6,7. While it is sometimes possible to achieve tighter tolerances8, this generally is not practical for the economical production of probes for general imaging and force measurement applications. Fortunately, this issue was recognized early while some of the first AFM force measurement applications were being developed (e.g. ref. 9) and many techniques have since been proposed to characterize cantilever spring constants. These can generally be grouped into three categories: “Dimensional models” where fully theoretical analysis or semi-empirical formulas are used to calculate the cantilever spring constants based on their dimensions and material properties, “Static deflection measurements” where the spring constant is determined by loading the cantilever with a known static force, and “Dynamic deflection measurements” where the resonance behavior of the cantilever is related back to its spring constant. Many articles have appeared that have discussed and compared some of the techniques 1,10-16. Unfortunately, many of these consider only a small subset of the techniques, or the discussion of existing techniques is secondary to introduction of a new method, or else they are simply outdated and don’t include important recent work. However, among these the recent reviews by Sader 10, Cook et al. 11, Hutter 12, and Butt et al.1 are very good and are highly recommended reading. In this Application Note we hope to expand on these works by including: 1) a review of the most commonly used calibration methods, including their recent refinements, 2) advice on selecting an appropriate method, and, 3) practical advice on implementing the methods discussed. 50436_AN94r0.qxd 1/29/07 6:57 AM Page 2 DIMENSIONAL MODELS FOR PREDICTING CANTILEVER SPRING CONSTANTS There has been a great deal of work done on developing dimensional models to predict cantilever spring constants. Clifford and Seah recently published a very good review of many of these methods 17. However, there are a number of challenges that have limited the popularity of these techniques. In particular, the thickness of cantilevers is difficult to measure accurately. Even with electron microscopy, uncertainty is rarely better than 5% 18-20. These difficulties led at least one group to measure the thickness with AFM by detaching the cantilever and laying it on a flat substrate 21 and another to predict the cantilever’s dimensions based on the resonance frequencies of several of its bending modes 22. The elastic modulus is also a major source of uncertainty, especially for silicon nitride cantilevers where the exact stoichiometry, SixNy, is often unknown 10,23-25. The thin metal films (i.e. Cr, Au, Al) often applied to cantilevers to enhance their reflectivity can also affect the spring constant and must therefore be considered 17, 26. In spite of these issues, the basic dimensional models are still useful in situations where an estimate is adequate and in special cases where the thickness and modulus are known. The result for a simple cantilevered rectangular beam, derived from Euler-Bernoulli beam theory 27, is given by: k= Ewt 3 4 L3 (1) where E is the elastic modulus of the cantilever, w is its width, t is its thickness, and L is its length. For v-shaped cantilevers, the simplifying assumption is often made that the cantilever is equivalent to two parallel beams with the same dimensions. This “parallel beam approximation” (PBA) was first suggested by Albrecht et al.6 and later considered by Butt 28. Sader was the first to recognize the ambiguity in defining w and L for v-shaped levers and offered his own refinements 29, 30, resulting in his final form: (2) ⎡ ⎤−1 Ewt 3 4 w3 ⎢ k= cos θ 1 + 3 (3 cos θ − 2)⎥ ⎢ ⎥ b 2 L3 ⎣ ⎦ where b is the width at the base of the “V”, is half the angle between the two legs, w is the width of the legs measured parallel to the front edge of the substrate (not perpendicular to the leg edge), and L is the length of the cantilever measured straight out to the apex from the substrate (i.e. not parallel to the leg). (cf. Figures 1 and 2 of reference 30). Neumeister and Ducker arrived at more complex analytical results 31, though the modest improvement in accuracy may be offset by the additional complexity. These various analytical models are frequently compared against finite element analysis (FEA) models 23,29-31. The FEA models are themselves dimensional models and subject to many of the same limitations as the simpler analytical results. For rectangular levers, Cleveland et al. realized that the thickness could be eliminated from Equation (1) by combining it with an expression for the resonant frequency of a cantilevered beam 24: / t ⎛⎜ E ⎞⎟ ⎟ f0 ≈ ⎜ 2π L2 ⎜⎝ ρ ⎟⎟⎠ 1 2 (3) k≈ ( ) These methods are based on the simple premise that the spring constant can be measured by applying a known force to the cantilever and measuring its deflection. In practice, application of these methods is complicated by the difficulty in accurately applying a known force of appropriate size. Mass calibration standards, for instance, are not generally available under the milligram range 32, making them several orders of magnitude too large for use in calibrating most AFM cantilevers. One early attempt at static deflection calibration placed a mass calibration standard in a pendulum configuration, pressed the cantilever to its side, and compared the deflection to the calculated force required to displace the mass from its resting position 28. Another attached small (i.e. 10-50µm diameter) tungsten spheres to cantilevers and compared the resulting deflection to the calculated mass of the spheres based on their measured diameters 33. However, by far the most popular technique among the static deflection measurements is use of a calibrated reference cantilever. Here the cantilever to be calibrated is used to measure a force curve on the end of a second cantilever that is calibrated. The slope of the contact portion of the force curve is compared to that measured on a hard surface (i.e. the deflection sensitivity in nm/V) and the spring constant calculated from: ⎛S ⎞ k = k ref ⎜⎜⎜ ref −1⎟⎟⎟ ⎟⎠ ⎜⎝ Shard where is the density of the cantilever, to obtain: 2π 3 w f 0 L ρ STATIC DEFLECTION MEASUREMENTS OF CANTILEVER SPRING CONSTANTS (5) 3 (4) E Note that this formula appears incorrectly in Tortonese and Kirk8 and is derived using a more exact expression for f0 in Clifford and Seah 17, which results in the 3 prefactor 2π being replaced by a constant α2= 59.3061. where kref is the spring constant of the reference cantilever, Sref is the deflection sensitivity measured on the reference cantilever, and Shard is the deflection sensitivity measured on a hard surface. One complication, however, arises in how the cantilever is positioned over the reference cantilever. The PAGE 2 PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS 50436_AN94r0.qxd 1/29/07 6:57 AM Page 3 cantilever must be positioned so that it contacts the reference cantilever as close to its end as possible because the reference cantilever appears progressively stiffer as loads are applied closer to its base. If the spring constant is measured with the tip offset from the end of the reference cantilever, it can be corrected by: ⎛ L ⎞⎟3 k = koff ⎜⎜ ⎜⎝ L − ∆L ⎟⎟⎠ (6) where koff is the spring constant measured offset from the end of the cantilever, L is the length of the reference cantilever, and ∆L is the distance that the tip is offset from the end of the reference cantilever 25. Note that this correction is exact for rectangular cantilevers and a good approximation (i.e. less than about 4% error 30) for v-shaped cantilevers provided that ∆L ≤ 0.10 L. Care should also be taken to position the tip near the midline of the cantilever to avoid errors due to torsional bending 34. Of course the key requirement for implementing the reference cantilever technique is that one has a calibrated reference cantilever. The earliest reported uses of the technique relied on reference cantilevers fashioned from glass fibers, which were themselves calibrated by optically measuring their deflection under the load of small calibration masses 35, 36. Later, handmade metal cantilevers were calibrated similarly 9, 37, 38. Others have used a regular AFM cantilever calibrated via a dimensional method 20. Fortunately, calibrated reference cantilevers are now commercially available (i.e. model CLFC-NOBO at Veeco Probes 5), which are manufactured with very well controlled dimensions and material properties 8 such that their spring constants are easily calculated from Equation (4). These probes have three cantilevers with spring constants ranging from 0.157 to 10.4N/m. This is important because the spring constant of the cantilever to be calibrated should generally be in the range 0.3kref < k < 3kref such that the deflection measured is not dominated by just one of the cantilevers 14. This allows the technique to be applied to a wide range of cantilever spring constants, which is an advantage over some other techniques. Also noteworthy is work being conducted by Cumpson and colleagues at the National Physical Laboratory in the U.K. 34,39-42 and Pratt and coworkers at the National Institute of Standards and Technology in the U.S. 32,43,44. Both groups are developing reference standards utilizing the static deflection approach with the ultimate goal of producing a device for cantilever calibration that is traceable to the Système International d'Unites. DYNAMIC DEFLECTION MEASUREMENTS OF CANTILEVER SPRING CONSTANTS This category includes three of the most widely used cantilever calibration methods: the added mass method 24, the thermal tune method 45, and the Sader method 26. The physical principles on which these methods are based differ greatly. However they all have in common that each require high speed measurement of the deflection signal in order to characterize the cantilever resonance behavior. Each also has distinct advantages and disadvantages, which will be discussed below. The added mass method, also known as the Cleveland method after one of its creators, is based on the following formula relating a cantilever’s fundamental resonance frequency, spring constant, and mass: f = 1 2π k M + m* Where m* is the “effective mass” of the cantilever, a quantity proportional to the actual mass of the cantilever, and M is an additional mass applied to the end of the cantilever 24. This additional mass consists of small (i.e. 3-10 micron diameter) tungsten microspheres 46 that are placed as close to the tip as possible. From Equation (7) we see that adding additional mass to a cantilever will reduce its resonance frequency. Of course the resonance frequency of the cantilever is easily measured by either performing a low amplitude TappingMode™ frequency sweep or else by looking at a power spectral density analysis of the cantilever’s thermally driven oscillations. Rearranging Equation (7) gives: M= k ( 2π f ) 2 − m* (8) Clearly if we were to add several masses and measure the new resonance frequency after each addition we would be able to plot 2 M vs. 1/(2πf) and find k from the slope of that line. However, one can also calculate the spring constant based on just one mass addition by comparing the resonance frequency of the original cantilever, f0, with that of the cantilever after addition of one mass, f1, by substituting the appropriate quantities into Equation (7) for the two cases and combining them to eliminate m*: For f=f0 and M=0: m* = k 2 (2π ) f 02 (9a) For f=f1 and M=M1: k = (2π ) f12 ( M 1 + m* ) 2 (9b) then combining to eliminate m*: k= (1 2 ( 2π ) M 1 f12 −1 f 02 ) (9c) (7) PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS PAGE 3 50436_AN94r0.qxd 1/29/07 6:57 AM Page 4 There are two major factors that limit the accuracy of this method. The first is that the position of the applied masses is critical. Masses applied closer to the base of the cantilever have less effect on the resonance frequency, effectively having a smaller mass. A correction for this effect is available, based on Equation (6), where: ⎛ L − ∆L ⎞⎟ = M meas ⎜⎜ ⎜⎝ L ⎟⎟⎠ 3 M eff (10) where Meff is the effective mass of the particle, Mmeas is the measured mass of the particle (discussed below), L is the length of the cantilever, and ∆L is the distance that the particle is offset from the tip of the cantilever. Note that this correction can be quite significant, for instance a 20µm offset on a 200µm long cantilever reduces the effective mass by almost 30%. The second significant source of error is in measurement of the tungsten masses. This is usually done by measuring the diameter of the particles with an optical microscope, calculating their volume based on the formula for a sphere, 3 V = (1/6)π D , and using the bulk density of tungsten (19300 kg/m3) to convert to mass. However, the particles are not perfectly spherical so it is preferable to measure their diameter in two axes and take the 1/2 geometric average, Davg=(D1D2) . Better yet would be to examine the particles by SEM and calculate the volume as an ellipsoid, though that obviously increases the effort required. Despite these potential limitations on accuracy, this method has gained a reputation as a sort of “gold standard” for cantilever calibration. Considering that measurements of the position of the sphere and its diameter could easily each have 5% error and each enters cubically in the calculation, it is unclear if this reputation is deserved. Certainly it is a very time intensive procedure with a high risk of damage to the cantilevers. Overall we believe that other methods are better choices for routine use. The second technique to be discussed here is the so-called Sader method. Note that Sader has been quite prolific in his study of cantilever calibration 25,26,29,30 and dynamics47,48, but here we specifically consider his technique for rectangular cantilevers that requires only their plan view (i.e. top down) dimensions, resonance frequency, quality factor (Q), and the density and viscosity of the fluid in which these are measured (typically air) 26. This method is quite convenient, but as formulated here it is only applicable to rectangular cantilevers and the frequency spectrum must generally be measured in air such that the response is not highly damped (i.e. Q>>1). More specifically, the theory requires that L>>w>>t, though in practice it is found that ratios of at least L/w>3 are acceptable 49. Sader’s result then is: k = 7.5246ρ f w2 LQ f 02Γi ( Re) (11) where: Re = 2πρ f f 0 w2 4η f where f is the density of the fluid in which the measurement is taken (typically air), f is the viscosity of that fluid, Q is the quality factor of the cantilever oscillation, and Γi is the imaginary component of the hydrodynamic function, which is a function of the Reynolds number, Re. While the other variables are familiar from the previous discussion, these new ones require explanation. Note that the prefactors here differ by 2 factors of (2π) and 2π, respectively, because Sader used the radial frequency in his original formulas. The resonance frequency and Q of the cantilever are measured by performing a power spectral density analysis of the cantilever’s thermally driven oscillations. The resonance peak is then fit with the simple harmonic oscillator (SHO) model: A = Awhite + A0 f 04 (f 2 −f 2 0 ) 2 ⎛f f ⎞ + ⎜⎜ 0 ⎟⎟⎟ ⎜⎝ Q ⎟⎠ 2 (12) where Awhite is a white noise fit baseline, A0 is the zero frequency amplitude, and f0 and Q are again the resonance frequency and quality factor. These four parameters are fit to the data using a least-squares method. We have found that this method gives more consistent results than attempting to use values taken from a TappingMode tuning curve. The hydrodynamic function is a more complicated calculation. Sader gives an analytical expression that is an approximate correction to the exact solution for a cylindrical beam 47. That solution is a complex function using modified Bessel functions of the third kind. The function is shown graphically in Sader et al. 26. However, Sader also makes available on his website 50 a Java applet to perform the calculation and downloadable Mathematica code. A posting also appeared on the Veeco SPM Digest that provided tabulated results versus Reynolds number and simple fit equations for these data 51. Note that the Reynolds number depends on the air viscosity and density. While the viscosity will be -5 -1 -1 1.84·10 kg m s at 20°C for all ambient pressures, the density is directly proportional to atmospheric pressure and can vary significantly from its sea level value of -3 1.18 kg m at higher elevations 11. Although mathematically complex, the Sader method is very convenient experimentally. The resonant frequency and Q can be measured very accurately and do not depend on any calibration of the AFM. PAGE 4 PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS 50436_AN94r0.qxd 1/29/07 6:57 AM Page 5 The cantilever dimensions can be measured optically and the air density and viscosity are essentially constants for a given laboratory location. Clearly the most severe limitation is that the method is only applicable to rectangular cantilevers. Even nominally rectangular cantilevers where the cantilevers taper to a long point (e.g. OTESPA probes 5) have been reported to give inaccurate results 15. However, minor deviations from rectangular shape like clipped corners on the cantilever ends (e.g. Veeco TESP, FESP, ESP probes) do not seem to introduce significant errors 15. Though this method is not applicable to v-shaped cantilevers, Sader does mention a way to calibrate v-shaped cantilevers as long as there is a rectangular cantilever on the same probe. By calibrating the rectangular cantilever using the method here, the elastic modulus and thickness of the cantilever can be deduced from Equation (1) and then the spring constant of v-shaped cantilevers on the same probe calculated with Equation (2) using those same values 26. However, this method is not generally applicable because it is rare for a probe to have both rectangular and v-shaped cantilevers. The known exceptions are Veeco Microlevers, which are silicon nitride probes that have five v-shaped cantilevers and one rectangular cantilever (e.g. model MLCT-AUNM 5). The last calibration method to be discussed here is the thermal tune method. This is probably the most popular and widely available method. It comes as a standard feature on all Veeco systems that use the new NanoScope® V controller. It is also available on the MultiMode® PicoForce™ on any NanoScope controller. The method is based on modeling the cantilever as a simple harmonic oscillator. Making use of the equipartition theorem, Hutter and Bechhoefer related the thermal (i.e. Brownian) motion of the cantilever’s fundamental oscillation mode to its thermal energy, kBT, via the formula: k T k = B2 (13) zc where kB is the Boltzmann constant -23 (1.38·10 J/K), T is the temperature, 2 and 〈z c 〉 is the mean square displacement of the cantilever 45. This quantity is found by performing a power spectral density analysis of the cantilever oscillations (i.e. of the vertical deflection signal) and integrating the area under the peak of the fundamental mode. Analysis in the frequency domain also has the effect of eliminating external noise sources from the measurement since they will likely be either broadband noise (e.g. white noise) that can be baseline subtracted or else they will occur at discrete frequencies separated from the cantilever’s resonance and can simply be ignored. Later, it was realized that two important corrections are necessary. These corrections are sometimes lumped together in the literature, which makes it somewhat confusing to compare different methods. The first correction takes into account that cantilevers do not behave as ideal springs and therefore the energy of their oscillatory modes differs from that of a simple harmonic oscillator. To correct for this, Butt and Jaschke derived a formula similar to Equation (13) by using beam theory to explicitly consider the actual bending modes of the cantilever: k= 12k BT αi4 zi2 k BT z12 More significantly, Butt and Jaschke also recognized that the cantilever displacement as measured by the optical lever detection scheme (i.e. using a reflected laser spot on a photodetector) is different from the actual displacement of the cantilever because it is proportional to angular changes in the cantilever position, not its absolute deflection, and these angular changes depend on the bending mode of the cantilever. They derived another formula, accounting for both this effect and the effect of considering only a single bending mode: (14a) where for i =1 (fundamental mode): k = 0.971 of these theoretical calculations were experimentally verified by independently calibrating the same cantilever with both the fundamental resonance and the resonance of the second bending mode, which resulted in good agreement between the two results 53. For v-shaped cantilevers this correction is more difficult to calculate because there is no analytical expression for the shape of the cantilever bending modes. Using finite element analysis, Stark et al. examined one particular v-shaped cantilever, the Veeco Microlever cantilever “E”, which is 140µm long and has a nominal spring constant of 0.1N/m. They found that the appropriate correction was very nearly the same, 0.965 instead of 0.971 for the rectangular cantilever 54. While this is the result of a numerical simulation on a specific v-shaped cantilever, it is probably safe to assume that there will also not be large variations in this correction for other common v-shaped cantilevers. (14b) where αi is a constant equal to 1.8751 for i = 1, the fundamental 2 mode of a rectangular lever, and 〈 zi 〉 is the mean squared displacement of a single bending mode 52. The results k= 16k BT 3αi2 zi*2 ⎛ sin α sinh α i i ⎜⎜ ⎜⎜⎝ sin α + sinh α i ⎞⎟2 ⎟⎟ (15a) ⎟ i⎠ where for i =1: k = 0.817 k BT z1*2 (15b) *2 where the asterisk in 〈 z1 〉 indicates that it is the “virtual” cantilever displacement (i.e. the displacement PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS PAGE 5 50436_AN94r0.qxd 1/29/07 6:57 AM Page 6 as measured by the optical lever scheme). Note that this second correction frequently appears as a correction to the deflection sensitivity, such as in Walters et al. 55 where the deflection sensitivity was adjusted upwards by a factor of 1.09. This factor comes from comparison of Equations (14b) and (15b), which we can then rewrite as: k= 0.971 k BT χ 2 z1*2 (16) which, when = 1.09, is equivalent to Equation (15b). Again note that this value is calculated based on the assumption of a rectangular cantilever. As with the first correction, it is difficult to calculate the proper correction for a v-shaped cantilever. Though again Stark et al. have addressed the question numerically for the same Veeco Microlever cantilever “E” and found that = 1.12, which when combined with the first correction (0.965), results in an overall prefactor for Equation (15b) of 0.764 instead of 0.817 for rectangular cantilevers 54. It is unknown how much may vary for other v-shaped cantilevers, though we don’t expect it to vary greatly. There have been reports noting that the original value of = 1.09 is based on the assumption of an infinitely small laser spot and that more accurate values can be calculated by taking into account the finite spot size, the cantilever size, and the spot position on the cantilever. The value of = 1.09 is the asymptotic limit of this calculation as the spot size goes to zero 56, 57. However we find that this calculation only results in significantly different values of when the cantilever length is significantly shorter than normal, the spot size is significantly larger than normal, or the laser is not properly aligned on the cantilever. For Veeco MultiMode and Dimension™ systems the laser spot size is typically <20µm, so when aligned at the end of a cantilever at least 100µm in length the value of only varies between 1.08 and 1.09, which is insignificant compared to other uncertainties in the measurements. We do acknowledge that this correction will become more important if so-called “small” cantilevers are used (cf. ref. 55) where the cantilever length is similar to the laser spot size] or if using an instrument with a large spot size. In these cases an online calculator for can be found at Schaffer’s website 58. It should also be noted that this calculation is only applicable to rectangular cantilevers. Since v-shaped cantilevers tend to have relatively small areas near the apex on which the laser spot can be positioned, this tends to dictate that the spot must be small and very near the end of the cantilever. Therefore we predict that these considerations would not be significant when using these cantilevers. Overall then the thermal tune method is an attractive method for spring constant calibration because of its simplicity and its general applicability to both rectangular and v-shaped cantilevers. The only real limitation to the technique is that it is best applied to relatively soft cantilevers where the thermal noise is well above the noise floor of the deflection measurement 61. EXPECTED UNCERTAINTY IN SPRING CONSTANT CALIBRATION It is very difficult to define the absolute error for any of these calibration methods. Often, estimates of error in the literature are based on comparisons of one method to another. However, we know that the overall error is due A NOTE ON THE EFFECT OF CANTILEVER TILT AFM cantilevers are typically mounted at a slight angle from horizontal in order to increase the distance between the sample and probe substrate. It has been noted that this can lead to errors in force measurements if proper corrections are not made 59, 60. Specifically, if the intrinsic spring constant is used then force 2 measurements are overestimated by a factor of cos α, where α is the tilt angle off horizontal 60, typically about 12°. As it turns out, however, the calibrated spring constant is often 2 underestimated by a factor of cos α due to the tilt and is therefore not the intrinsic spring constant. This occurs for both the reference cantilever and thermal tune methods. So for these two methods the errors cancel each other out and no further correction is required. The dimensional, added mass, and Sader methods, however, result in the correct intrinsic spring constants. So spring constants calibrated with those methods should be modified to obtain the “effective” spring constant: k keff = cos2 α Also, note that if the tip is long compared to the cantilever length then this can also result in a non-trivial error 60. However, most tips only protrude <20µm from the cantilever for Si probes and <4µm for SiN probes, so this correction is not often significant. PAGE 6 PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS 50436_AN94r0.qxd 1/29/07 6:57 AM Page 7 to both measurement errors and intrinsic errors and approximations in the methods themselves. Lacking a true “gold standard,” perhaps the best we can do is to try to examine the relative uncertainty of each method separately. This will only tell us about the absolute error of the calibration to the extent that we trust the model on which the method is based. However, it is useful and instructive to answer the question of how much uncertainty there is in a single calibration, i.e. how much scatter would we expect in the spring constant if we calibrated the same lever many times with the same method? To examine this issue, Monte Carlo calculations were performed by preparing 10,000 data sets for each method with estimated average errors that were normally distributed for each measurement type. The spring constants for each set were calculated and the standard deviation of those taken as representative of uncertainty resulting for each method due to measurement errors. The estimated average errors for each measurement type are listed in Table 1. The results of the analysis are shown in Table 2, which lists the estimated uncertainty for each method due to measurement errors along with the particular measurements that dominate the overall uncertainty. The estimates for the PBA and thermal tune methods are based on typical values for the long, thin (k~0.06N/m) cantilever on a silicon nitride DNP probe. The estimates for the other methods are based on typical values for a silicon FESP (k~2N/m) probe. This choice is especially significant to the error estimate for the PBA formula, which suffers from the large uncertainty in the silicon nitride elastic modulus. The other surprisingly large error is that for the added mass method. Here even modest 5% uncertainty in the particle diameter contributes almost 15% uncertainty to the Table 1: Estimated measurement errors Average error Measurement Note Cantilever length 1% Optical (e.g. 2µm out of 200µm) Cantilever width 4% Optical (e.g. 1µm out of 25µm) Cantilever thickness 5% SEM (e.g. 150nm out of 3µm) Length offsets (∆L) 10% Optical (e.g. 1µm out of 10µm) Particle diameter 5-10% Elastic modulus 5% (Si) 20% (SixNy) Estimate 5% Estimate Density (Si or W) Resonance frequency Optical (e.g. 1µm out of 10-20µm) 0.1% Cantilever tune, reference [26] Quality factor 1% SHO fit, reference [26] Deflection sensitivity 3% Estimate Reference lever k 3.3% Density of air 5% Viscosity of air 2.5% Temp. at cantilever 2 <z >, voltage PSD Reference [8] Typical atmospheric pressure changes Changes with temperature, 20-40°C 3% Uncertain due to laser heating 3% Estimate Table 2: Overall uncertainty in spring constants Method Uncertainty Main source of error Simple beam, Eqn. (1) ~16% Cantilever thickness PBA, Eqn. (2) ~26% Elastic modulus of SiN Freq. scaling, Eqn. (4) ~9% Si density Reference cantilever, Eqn. (5) and (6) ~9% Deflection sensitivity Added mass, Eqn. (9c) and (10) 15-30% Particle diameter Sader, Eqn. (11) ~4% Cantilever width Thermal tune, Eqn. (15a) ~8% Deflection sensitivity spring constant. In order to reduce the error it would be important to use larger particles where the relative uncertainty in diameter is lower. For any of the methods, it is important to realize that the main source of error listed in Table 1 is the largest contributor to uncertainty by a large margin. Any efforts to improve accuracy should be directed toward reducing uncertainty in these measurements. Although we believe that these are reasonable uncertainty estimates, we do emphasize that they do not consider any weaknesses of the methods themselves, but rather only uncertainty in measurements. For instance, the thermal tune method has a fair amount on uncertainty in its various corrections. The Sader method assumes a perfect rectangular cantilever, which is often only an approximation of reality. PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS PAGE 7 50436_AN94r0.qxd 1/29/07 6:57 AM Page 8 The added mass method is actually based on simple beam theory and is only extended to v-shaped levers by the basic parallel beam approximation. While difficult to quantify the uncertainty that they contribute, it is best to understand these various limitations. RECOMMENDATIONS ON SELECTING A SPRING CONSTANT CALIBRATION METHOD Any of the experimental methods are well accepted and appear frequently in the literature. Increasingly, equipment availability is becoming less of a limiting factor. In the past the dynamic deflection measurement methods frequently required external hardware capable of higher sampling rates than the standard AFM controller. However, especially with the introduction of Veeco’s new NanoScope V controller, researchers can now perform all of these measurements with no additional hardware. Even analysis has become much simpler, now that the thermal tune method is directly supported by the NanoScope V and PicoForce and the Sader method calculations can be performed using Sader’s online calculator 50. Selecting a spring constant calibration method should then be based on the limitations of the methods, their accuracy, and convenience. For relatively stiff, (i.e. k > 1N/m), rectangular cantilevers the Sader method has a lot of advantages. The uncertainty of the method is very good and the measurements are convenient to make. The reference cantilever method can also be a good choice, though it is somewhat less convenient and generally has more uncertainty. For relatively soft cantilevers, (i.e. k < 1N/m), the thermal tune method is convenient and its uncertainty is relatively low. Moreover, many of these soft cantilevers are v-shaped, so the Sader method cannot be used. The reference cantilever method can be a reasonable alternative. such that the nominal spring constant of the cantilever to be calibrated is in the range of 0.3k ref < k < 3k ref In general, we see few reasons to recommend the dimensional methods and the added mass method. There is simply too much uncertainty in the required measurements and the added mass method is experimentally very time consuming. 2. Mount the reference cantilever on a sample disc using double-sided tape. Align the reference cantilever on the AFM sample stage such that its long axis is aligned with the cantilever to be calibrated, but facing the opposite direction, as shown below in Figure 1. Although we haven’t discussed it here, calibration of the lateral spring constant can also be done. This is important for quantifying frictional forces, for instance. While an in-depth discussion of the methods is beyond the scope of this Application Note, we will note that the method by Ogletree et al.62 and refined by Varenberg et al.63 is well accepted. The more recent contribution by Green et al.64, which is very similar to the Sader method discussed here, also appears convenient. IMPLEMENTING THE CALIBRATION METHODS ON VEECO AFM SYSTEMS REFERENCE CANTILEVER METHOD Implementing the reference cantilever method is straightforward on all Veeco AFM systems. We recommend the use of Veeco reference cantilevers (i.e. model CLFC-NOBO at Veeco Probes 5). 1. Calibrate the reference cantilever as directed according to its documentation. Briefly, the resonance frequency of the cantilever is measured using the TappingMode cantilever tune and Equation (4) is then used to calculate the spring constant based on the cantilever size and material properties, which are specified in the documentation. Recall that Equation (4) appears incorrectly in Tortonese and Kirk 8. Be sure to select a reference cantilever Figure 1: Initial alignment of reference cantilever 3. Align the cantilever to be calibrated over the substrate of the reference cantilever, as shown below in Figure 2, and engage normally in contact mode. Measure the deflection sensitivity in ramp mode. It is good practice to keep the vertical deflection signal near 0V and measure the deflection sensitivity over about 100nm of deflection. Repeat the deflection sensitivity measurement a few times and average the values. Figure 2: Deflection sensitivity on hard surface 4. Withdraw and realign the cantilever close to the end of the reference cantilever as shown below in Figure 3. Engage normally in contact mode and PAGE 8 PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS 50436_AN94r0.qxd 1/29/07 6:57 AM Page 9 repeat the deflection sensitivity measurement. Take care that the vertical deflection signal is near the same value as for the last measurement and measure the deflection sensitivity over the same range. Again, take several measurements and average the results. Figure 3: Measurement on reference cantilever 5. Measure the offset of the tip from the end of the reference cantilever. This is most conveniently done by using the top down optics of the AFM. You can use the length of the reference cantilever as a standard to calibrate the optical view. 6. Use equations (5) and (6) to calculate the spring constant. SADER METHOD The Sader method requires a power spectral density analysis of the cantilever’s thermal noise. This is easily measured on a Veeco NanoScope V system by using the included thermal tune interface, as follows: 3. From the workspace bar, enter the thermal tune view. If it doesn’t show up there, you can access it through the Acquire>Thermal Tune menu. If you get an error that the deflection sensitivity is not calibrated, you should simply go to Tools>Calibrate>Detector… and enter a reasonable value (e.g. 60 nm/V). The Sader method does not require that the deflection signal be calibrated, so there is no reason to do a real deflection sensitivity calibration. 4. In the thermal tune view, select the appropriate range for your cantilever, either 1-100 kHz or 5-2000 kHz. Click the “Get Data” button and wait for the acquisition to complete. 5. You should see a peak on the graph near the nominal resonance frequency for the cantilever. Sometimes you will see more than one peak. These might include resonances of higher modes of the cantilever. 6. Zoom in on the peak by holding down the Ctrl key and dragging a box over the peak. If you need to zoom back out, click on the magnifying glass icon in the bottom left corner of the graph. The peak should look something like the one in Figure 4 below. 7. The resonance frequency and quality factor must be determined by fitting the peak with the SHO equation (Equation 12). Drag two markers in onto the graph by clicking between the edge of the graph box and the vertical axis of the graph and dragging. Position them on each side of the peak, as shown in Figure 4. Select the "Simple Harmonic Oscillator (Fluid)" fit option and click "Fit Data." A red line showing the fit function will appear over the peak. The resonance frequency and Q will appear among the parameters to the right. 8. If you are far from sea level, calculate the density of air using the ideal gas equation: ρair = PM RT where P is the pressure, M is the average molar mass of air -1 (0.02897kg mol ), R is the molar -1 -1 gas constant (8.314 J mol K ), and T is the temperature. At sea -3 level, the density is 1.18kg m . -5 -1 -1 The viscosity is 1.86·10 kg m s . 9. Calculate k using the online calculator at Sader’s website 50, or using the other methods for calculating the hydrodynamic function discussed earlier. Note: Be sure to use a recent version of the NanoScope software. You can tell if you have a new enough version by comparing your thermal tune view to the one in Figure 4. The fluid fit (circled) should be labeled "Simple Harmonic Oscillator (Fluid)." If it is labeled simply "Fluid," please contact Veeco customer support for an upgrade. 1. Measure the length and width of the cantilever you wish to calibrate. You could use the top down optics of the AFM after calibrating them with a standard calibration grating. 2. Setup the AFM with the same cantilever, zero the vertical deflection, and configure the software for contact mode. You do not need to engage. Figure 4: Using the thermal tune view to collect data for the Sader method. PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS PAGE 9 50436_AN94r0.qxd 1/29/07 6:57 AM Page 10 THERMAL TUNE METHOD As previously mentioned, the thermal tune method is fully supported on Veeco NanoScope V systems. The procedure is as follows: 1. Engage in contact mode on a hard surface. Enter ramp mode and calibrate the deflection sensitivity. Again, it is good practice to keep the vertical deflection signal near 0V and measure the deflection sensitivity over about 100nm of deflection. It is very important for the deflection sensitivity to be accurate. 2. Withdraw the tip by clicking the withdraw button 2-3 times. 3. From the workspace bar, enter the thermal tune view. If it doesn’t show up there, you can access it through the Acquire>Thermal Tune menu. 4. In the thermal tune view, select the appropriate range for your cantilever, either 1-100kHz or 5-2000kHz. You can also enter the temperature if it is significantly different from the default values. You will not generally need to change the “PSD bin width” and “Median filter width” parameters. 5. The “Deflection sensitivity correction” parameter is applied the same way as the correction discussed earlier. However, it must also include the thermal tune correction of 0.971 or 0.965 from Equation (14b). Therefore we recommend the following values: rectangular cantilevers: 1.106 v-shaped cantilevers: 1.144 These values result in the appropriate prefactors as discussed for Equation (15b). wait for the data collection to finish. The PSD will appear in the graph. 7. Find the fundamental resonance peak. There may be additional peaks (e.g. resonances of higher bending modes) at higher frequencies. The nominal resonance frequency provided by the manufacturer should help you find the correct peak. 8. Drag markers in onto the graph by clicking between the edge of the graph box and the vertical axis of the graph and dragging. Bracket the resonance peak such that the markers are position on each side and far enough away from the peak to include some of the baseline. 9. Select the “Air” or “Fluid” button. This selects either a Lorentzian or SHO model fit, respectively. The Lorentzian fit generally fits the data well in either air or fluid. 10. Click the “Fit data button.” A fit curve will appear on top of the PSD data. Verify that the fit appears valid. Readjust the marker positions as needed if the fit is not good. 11. Click the “Calc spring constant” button. A box will pop up with the calculated spring constant. The thermal tune view varies slightly for PicoForce and Dimension Hybrid systems on NanoScope IIIa and IV controllers. The frequency range is limited to 30kHz. However, this will still allow you to calibrate most soft cantilevers. Note that on Dimension Hybrid systems you will also be asked to manually toggle a switch on the thermal tune adaptor. ADDED MASS METHOD As we noted before, this method is not generally recommended because of its considerable uncertainties and the time consuming nature of attaching the particles. However, the general procedure is as follows: 1. Measure the resonance frequency of the cantilever by using the TappingMode cantilever tune window. Use very small drive amplitudes. You only want to be able to accurately find and measure the frequency of the resonance. 2. Attach a tungsten sphere near the end of the cantilever. It is usually best to attach it to the tip side of the cantilever so that you can still align the laser on the backside. Refer to Veeco Support Note 226, “Attaching particles to AFM cantilevers”, for helpful hints on this procedure. Note that capillary forces may be sufficient to hold the particle in place and therefore allow you to avoid using an adhesive. Measure the diameter of the particle as accurately as possible. Also measure the distance between the particle’s center and the end of the cantilever. 3. Re-measure the resonance frequency of the cantilever. 4. Using Equations (9c) and (10), calculate the spring constant. 5. Alternatively, remove the mass, attach another, and re-measure the resonance frequency. Repeat, collecting a small data set of M vs. f values. Perform a linear regression analysis on this data set, extracting k as the slope according to Equation (8). The averaging obtained may improve the overall uncertainty. 6. Click the “Get data” button and PAGE 10 PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS 50436_AN94r0.qxd 1/29/07 6:57 AM Page 11 REFERENCES 1. H.-J. Butt, B. Cappella, M. Kappl, "Force measurements with the atomic force microscope : Technique, interpretation and applications", Surf. Sci. Rep. 59 (2005) 1-152. 2. H. Janovjak, J. Struckmeier, M. Hubain, A. Kedrov, M. Kessler, D.J. Muller, "Probing the Energy Landscape of the Membrane Protein Bacteriorhodopsin", Structure 12 (2004) 871-879. 3. M. Kawakami, K. Byrne, B. Khatri, T.C.B. Mcleish, S.E. Radford, D.A. Smith, "Viscoelastic Properties of Single Polysaccharide Molecules Determined by Analysis of Thermally Driven Oscillations of an Atomic Force Microscope Cantilever", Langmuir 20 (2004) 9299-9303. 4. A. Trache, J.P. Trzeciakowski, L. Gardiner, Z. Sun, M. Muthuchamy, M. Guo, S.Y. Yuan, G.A. Meininger, "Histamine Effects on Endothelial Cell Fibronectin Interaction Studied by Atomic Force Microscopy", Biophys. J. 89 (2005) 2888-2898. 5. Veeco Probes website, http://www.veecoprobes.com 6. T.R. Albrecht, S. Akamine, T.E. Carver, C.F. Quate, "Microfabrication of cantilever styli for the atomic force microscope", J. Vac. Sci. Technol. A8 (1990) 3386-3396. 7. M. Tortonese, "Cantilevers and tips for atomic force microscopy", IEEE Engineering in Medicine and Biology Magazine 16 (1997) 28-33. 8. M. Tortonese, M. Kirk, "Characterization of application specific probes for SPMs", Proc. SPIE 3009 (1997) 53-60. 9. E.L. Florin, V.T. Moy, H.E. Gaub, "Adhesion forces between individual ligand-receptor pairs", Science 264 (1994) 415-417. 10. J.E. Sader, Calibration of atomic force microscope cantilevers, in: A. Hubbard (Eds.), Encyclopedia of Surface and Colloid Science, Marcel Dekker, Inc., 2002, pp. 846-856 11. S.M. Cook, T.E. Schaffer, K.M. Chynoweth, M. Wigton, R.W. Simmonds, K.M. Lang, "Practical implementation of dynamic methods for measuring atomic force microscope cantilever spring constants", Nanotechnology 17 (2006) 2135-2145. 14. C.T. Gibson, G.S. Watson, S. Myhra, "Scanning force microscopy-calibrative procedures for `best practice'", Scanning 19 (1997) 564-581. 28. H.-J. Butt, P. Siedle, K. Seifert, K. Fendler, E. Bamber, K. Goldie, A. Engel, "Scan speed limit in atomic force microscopy", J. Microsc. 169 (1993) 75-84. 15. C.T. Gibson, D.A. Smith, C.J. Roberts, "Calibration of silicon atomic force microscope cantilevers", Nanotechnology 16 (2005) 234-238. 29. J.E. Sader, L. White, "Theoretical analysis of the static deflection of plates for atomic force microscope applications", J. Appl. Phys. 74 (1993) 1-9. 16. E.L. Florin, M. Rief, H. Lehmann, M. Ludwig, C. Dornmair, V.T. Moy, H.E. Gaub, "Sensing specific molecular interactions with the atomic force microscope", Biosensors and Bioelectronics 10 (1995) 895-901. 30. J.E. Sader, "Parallel beam approximation for V-shaped atomic force microscope cantilevers", Rev. Sci. Instrum. 66 (1995) 4583-4587. 17. C.A. Clifford, M.P. Seah, "The determination of atomic force microscope cantilever spring constants via dimensional methods for nanomechanical analysis", Nanotechnology 16 (2005) 1666-1680. 18. U.D. Schwarz, P. Koster, R. Wiesendanger, "Quantitative analysis of lateral force microscopy experiments", Rev. Sci. Instrum. 67 (1996) 2560-2567. 19. J.D. Holbery, V.L. Eden, M. Sarikaya, R.M. Fisher, "Experimental determination of scanning probe microscope cantilever spring constants utilizing a nanoindentation apparatus", Rev. Sci. Instrum. 71 (2000) 3769-3776. 20. C.T. Gibson, G.S. Watson, S. Myhra, "Determination of the spring constants of probes for force microscopy/spectroscopy", Nanotechnology 7 (1996) 259-262. 21. N. Maeda, T.J. Senden, "A method for the calibration of force microscopy cantilevers via hydrodynamic drag", Langmuir 16 (2000) 9282-9286. 22. A.W. McFarland, M.A. Poggi, L.A. Bottomley, J.S. Colton, "Characterization of microcantilevers solely by frequency response acquisition", J. Micromech. Microeng. 15 (2005) 785-791. 23. J.L. Hazel, V.V. Tsukruk, "Spring constants of composite ceramic/gold cantilevers for scanning probe microscopy", Thin Solid Films 339 (1999) 249-257. 24. J.P. Cleveland, S. Manne, D. Bocek, P.K. Hansma, "A nondestructive method for determining the spring constant of cantilevers for scanning force microscopy", Rev. Sci. Instrum. 64 (1993) 403-405. 12. J.L. Hutter, "Calibration of AFM Cantilevers", www.physics.uwo.ca/ ~hutter/calibration/afmcal.html, last updated: July 17, 2005. 25. J.E. Sader, I. Larson, P. Mulvaney, L.R. White, "Method for the calibration of atomic force microscope cantilevers", Rev. Sci. Instrum. 66 (1995) 3789-3798. 13. N.A. Burnham, X. Chen, C.S. Hodges, G.A. Matei, E.J. Thoreson, C.J. Roberts, M.C. Davies, S.J.B. Tendler, "Comparison of calibration methods for atomic-force microscopy cantilevers", Nanotechnology 14 (2003) 1-6. 26. J.E. Sader, J.W.M. Chon, P. Mulvaney, "Calibration of rectangular atomic force microscope cantilevers", Rev. Sci. Instrum. 70 (1999) 3967-3969. 27. W.C. Young, R. Budynas, Roark's Formulas for Stress and Strain, McGraw-Hill, 2001. 31. J.M. Neumeister, W.A. Ducker, "Lateral, normal, and longitudinal spring constants of atomic force microscopy cantilevers", Rev. Sci. Instrum. 65 (1994) 2527-2531. 32. J.R. Pratt, D.T. Smith, D.B. Newell, J.A. Kramar, E. Whitenton, " Progress toward Système International d'Unités traceable force metrology for nanomechanics", J. Mater. Res. 19 (2004) 366-379. 33. T.J. Senden, W.A. Ducker, "Experimental Determination of Spring Constants in Atomic Force Microscopy", Langmuir 10 (1994) 1003-1004. 34. P.J. Cumpson, C.A. Clifford, J. Hedley, "Quantitative analytical atomic force microscopy: a cantilever reference device for easy and accurate AFM spring-constant calibration", Meas. Sci. Technol. 15 (2004) 1337-1346. 35. Y.Q. Li, N.J. Tao, J. Pan, A.A. Garcia, S.M. Lindsay, "Direct Measurement of Interaction Forces between Colloidal Particles Using the Scanning Force Microscope", Langmuir 9 (1993) 637-641. 36. Y.I. Rabinovich, R.-H. Yoon, "Use of Atomic Force Microscope for the Measurements of Hydrophobic Forces between Silanated Silica Plate and Glass Sphere", Langmuir 10 (1994) 1903-1909. 37. A. Torii, M. Sasaki, K. Hane, S. Okuma, "A method for determining the spring constant of cantilevers for atomic force microscopy", Meas. Sci. Technol. 7 (1996) 179-184. 38. S.K. Jericho, M.H. Jericho, "Device for the determination of spring constants of atomic force microscope cantilevers and micromachined springs", Rev. Sci. Instrum. 73 (2002) 2483-2485. 39. P.J. Cumpson, J. Hedley, "Accurate analytical measurements in the atomic force microscope: a microfabricated spring constant standard potentially traceable to the SI", Nanotechnology 14 (2003) 1279-1288. 40. P.J. Cumpson, J. Hedley, C.A. Clifford, X. Chen, S. Allen, "Microelectromechanical system device for calibration of atomic force microscope cantilever spring constants between 0.01 and 4 N/m", J. Vac. Sci. Tecnol. A22 (2004) 1444-1449. PRACTICAL ADVICE ON THE DETERMINATION OF CANTILEVER SPRING CONSTANTS PAGE 11 50436_AN94r0.qxd 1/29/07 6:57 AM Page 12 41. P.J. Cumpson, J. Hedley, P. Zhdan, "Accurate force measurement in the atomic force microscope: a microfabricated array of reference springs for easy cantilever calibration", Nanotechnology 14 (2003) 918-924. 49. J.W.M. Chon, P. Mulvaney, J.E. Sader, "Experimental validation of theoretical models for the frequency response of atomic force microscope cantilever beams immersed in fluids", J. Appl. Phys. 87 (2000) 3978-3988. 42. P.J. Cumpson, P. Zhdan, J. Hedley, "Calibration of AFM cantilever stiffness: a microfabricated array of reflective springs", Ultramicroscopy 100 (2004) 241-251. 50. J.E. Sader, "Atomic Force Microscope Cantilevers (Calibration method of Sader)", <http://www.ampc.ms.unimelb. edu.au/afm/index.html>, last updated: August 11, 2005. 43. J.R. Pratt, J.A. Kramar, D.B. Newell, D.T. Smith, "Review of SI traceable force metrology for instrumented indentation and atomic force microscopy", Meas. Sci. Technol. 16 (2005) 2129-2137. 51. M. Berkahn, "Hydrodynamic function calculated", <http://spmlist.di.com/ pipermail/spm/>, last updated: February 25, 2004. 44. R.S. Gates, J.R. Pratt, "Prototype cantilevers for SI-traceable nanonewton force calibration", Measurement Science and Technology 17 (2006) 2852. 45. J.L. Hutter, J. Bechhoefer, "Calibration of atomic-force microscope tips", Rev. Sci. Instrum. 64 (1993) 1868-1873. 46. BioForce Nanosciences website, www.bioforcenano.com 47. J.E. Sader, "Frequency response of cantilever beams immersed in viscous fluids with applications to the atomic force microscope", J. Appl. Phys. 84 (1998) 64-76. 48. J.E. Sader, J. Pacifico, C.P. Green, P. Mulvaney, "General scaling law for stiffness measurement of small bodies with applications to the atomic force microscope", J. Appl. Phys. 97 (2005) 124903. 52. H.-J. Butt, M. Jaschke, "Calculation of thermal noise in atomic force microscopy", Nanotechnology 6 (1995) 1-7. 53. R. Levy, M. Maaloum, "Measuring the spring constant of atomic force microscope cantilevers: thermal fluctuations and other methods", Nanotechnology 13 (2002) 33-37. 54. R.W. Stark, T. Drobek, W.M. Heckl, "Thermomechanical noise of a free v-shaped cantilever for atomic-force microscopy", Ultramicroscopy 86 (2001) 207-215. 55. D.A. Walters, J.P. Cleveland, N.H. Thomson, P.K. Hansma, M.A. Wendman, G. Gurley, V. Elings, "Short cantilevers for atomic force microscopy", Rev. Sci. Instrum. 67 (1996) 3583-3590. 56. R. Proksch, T.E. Schaffer, J.P. Cleveland, R.C. Callahan, M.B. Viani, "Finite optical spot size and position corrections in thermal spring constant calibration", Nanotechnology 15 (2004) 1344-1350. 57. T.E. Schaffer, "Calculation of thermal noise in an atomic force microscope with a finite optical spot size", Nanotechnology 16 (2005) 664-670. 58. T.E. Schaffer, "Online Calculator for the detection sensitivity of the optical beam deflection method", <http://bioforce. centech.de//>, last updated: 2006. 59. L.-O. Heim, M. Kappl, H.-J. Butt, "Tilt of Atomic Force Microscope Cantilevers: Effect on Spring Constant and Adhesion Measurements", Langmuir 20 (2004) 2760-2764. 60. J.L. Hutter, "Comment on Tilt of Atomic Force Microscope Cantilevers: Effect on Spring Constant and Adhesion Measurements", Langmuir 21 (2005) 2630-2632. 61. C. Su, L. Huang, K. Kjoller, "Direct measurement of tapping force with a cantilever deflection force sensor", Ultramicroscopy 100 (2004) 233-239. 62. D.F. Ogletree, R.W. Carpick, M. Salmeron, "Calibration of frictional forces in atomic force microscopy", Rev. Sci. Instrum. 67 (1996) 3298-3306. 63. M. Varenberg, I. Etsion, G. Halperin, "An improved wedge calibration method for lateral force in atomic force microscopy", Rev. Sci. Instrum. 74 (2003) 3362-3367. 64. C.P. Green, H. Lioe, J.P. Cleveland, R. Proksch, P. Mulvaney, J.E. Sader, "Normal and torsional spring constants of atomic force microscope cantilevers", Rev. Sci. Instrum. 75 (2004) 1988-1996. Veeco Instruments Inc. For more information visit www.veeco.com or call 800-873-9750 AN94, Rev A0 Solutions for a nanoscale world. ™ © 2007 Veeco Instruments Inc. All rights reserved. Dimension, PicoForce, and TappingMode are trademarks of Veeco Instruments Inc. NanoScope is a registered trademark of Veeco Instruments Inc.
© Copyright 2026 Paperzz