Brit. J. Anaesih. (1967), 39, 659 THE HIGH OXYGEN MIXTURES DELIVERED BY THE AIR-MEX CONTROL OF THE BIRD MARK 7 VENTILATOR BY G. A. HARRISON Department of Anaesthetics, St. Vincent's Hospital, Sydney, Australia SUMMARY The reasons for the high oxygen mixtures delivered by the Bird Mark 7 ventilator were investigated by measuring the flow of gases through various parts of the ventilator. The major causes were the constant flow of 100 per cent oxygen from the nebulizer, the decrease and eventual cessation of flow through the venturi as the airway pressure rose, and the decrease in the proportion of entrained air at low flow settings. When the Bird Mark 7 automatic ventilator is driven by a source of compressed oxygen, an "airmix" control permits the administration of either 100 per cent oxygen or a mixture of oxygen and air. This oxygen-air mixture is obtained when the oxygen passes through an injector, entraining air through a filter (fig. 1). The mixture then passes through a spring-loaded one-way valve into the righthand chamber of the ventilator and on to the patient via a length of wide-bore tubing. Before reaching the patient the mixture also passes through a humidifier and through a nebulizer to FIG. 1 Simplified diagram of the Bird Mark 7 ventilator. The arrows indicate the flow of gases during the inspiratory phase. 1. Flow control 6. Oxygen supply for non8. Humidifier 2. Inspiratory pressure control return valve and 9. Nebulizer 3. Filter for entrained air nebulizer 10. Non-rebreathing valve 4. Venturi 7. Line for oxygen-air 11. Patient attachment 5. Spring-loaded valve mixture BRITISH JOURNAL OF ANAESTHESIA 660 which 100 per cent oxygen is added from the ventilator by a length of small-bore pressure tubing. This oxygen also exerts pressure on a nonrebreathing valve, keeping it closed during inflation. When the pressure in the righthand chamber reaches a figure set by the operator, a diaphragm snaps over to the left, the supply of oxygen is interrupted, and the machine cycles to expiration. Fairley and Britt (1964) investigated the mean inspired oxygen percentages when the air-mix control was used and found that they varied directly with the cycling pressure and inversely with the inspiratory flow rate. The oxygen concentrations ranged from 46 to 99 per cent but, at any fixed flow or pressure, there was considerable variation from one Mark 7 to another. Using a technique similar to Fairley and Britt, the author has confirmed their findings and has demonstrated that a decrease in compliance or an increase in airway resistance causes a further rise in the mean inspired oxygen percentage at any particular flow or pressure setting. This paper describes an investigation into the causes of the high inspired oxygen mixtures and the reason for the increase with changes in the compliance and airway resistance. lator and the nebulizer. This demonstrated the flow from the Bird ventilator of the oxygen-air mixture which had arisen from the injector system. (<f) In the wide-bore tubing between the nebulizer and the non-rebreathing valve. This demonstrated the total inspiratory flow (mixture from ventilator plus nebulizer oxygen). (e) Between the non-rebreathing valve and the endotracheal tube. This demonstrated the overall inspiratory and expiratory flow patterns. As the inspiratory flow pattern was expected to be the same as found in (d), this pneumotachometer acted as a check against any leaks or additional sources of oxygen around the non-rebreathing valve. The pneumotachometer head was connected to a Sanborn Model 270 differential transducer and the output recorded on a Sanborn two-channel recorder Model 321. On the other channel was recorded the airway pressure measured in the endotracheal tube by a Sanborn Model 267B pressure transducer. The ventilator was set to a pressure setting of 25 and recordings taken at flow settings of 6, 10, METHOD Three Bird Mark 7 ventilators in regular use in an intensive therapy ward were studied. The ventilator was connected by a size 10 Magill endotracheal rube and non-distensible plastic tubing to a rigid drum with a compliance of 23.2 ml/cm KLO (fig. 2). Two one-way valves ensured the mixing of the inflating gases which could be collected and sampled as desired in a Douglas bag en the expiratory side of the non-rebreathing valve. The flow of gases in various parts of the ventilator was measured by inserting a Fleisch pneumotachometer head in one of five positions in turn (fig. 2): (a) Connected to ± e lefthand side of the ventilator containing the air filter by a surrounding layer of thick plasticine which was free of leaks up to a pressure of 20 mm Hg. This measured the flow of air which was entrained into the Bird ventilator. (6) In the small-bore tubing leading to the nonrebreathing valve and nebulizer. (c) In the wide-bore tubing between the venti- FIG. 2 The arrangement of the Bird Mark 7 ventilator and the artificial lung (rigid drum). The flow of gases through different parts of the circuit was measured in turn by a pneumotachometer inserted in the positions shown. Airway pressure was measured at the ventilator end of the endotracheal tube. a. Pneumotachometer: entrained air filter b. Pneumotachometer: oxygen to valve and nebulizer c. Pneumotachometer: Bird side of nebulizer d. Pneumotachometer: patient side of nebulizer e. Pneumotachometer: between valve and "patient" f. Oxygen analyzer g. Douglas bag h. Endotracheal tube i. Rigid drum HIGH OXYGEN MIXTURES FROM BIRD MARK 7 VENTILATOR 20, 30 and 40. The flow setting was then set at 20 and recordings taken at pressure settings of 15, 25, 40 and at a setting slightly in excess of 40 (>40). Recordings were repeated three times for each of the three ventilators. The sensitivity was fixed at 15 and the rate at 16 b.p.m. at intermediate flows and pressures. The rate then varied with the duration of the inspiratory period which depended on the particular flow or pressure setting. Tracings of flow from positions (a) (entrained air), (c) (air and oxygen from injector) and (d) (air + oxygen from injector + nebulizer oxygen) were superimposed. It was possible to check that no unsuspected leaks or obstructions had occurred in the circuit during the change of site of the pneumotachometer by comparing the pressure curve and the flow pattern from the same site during different runs. The contribution of the entrained air to the total air-oxygen mixture was established by measurement of the areas under the curves from positions (a) and (c) by planimetry. Finally the experiments were repeated at a pressure setting of 20 and flow settings of 6, 10, 20, 30, and 40, (1) with the drum of compliance 23.2 ml/cm H2O and a size 10 Magill tube, (2) with the drum of compliance 23.2 ml/cm H,O and a size 3 Magill tube, and (3) with the drum of compliance 12.5 ml/cm H2O and a size 10 Magill tube. 661 ward the "patient" ceased, oxygen actually passed in the reverse direction from the injector through the filter to the atmosphere. When the flow was increased to a setting of 10 the total flow into the "patient" increased to 37 l./min but again there was a constant contribution of 8 l./min of oxygen from the nebulizer. The flow from the Bird eventually stopped and reverse flow occurred through the air filter. The period of infiltration of the "patient" by the nebulizer oxygen only was much shorter. At flow setting 20 the total flow had risen but the contribution of the nebulizer remained a RESULTS Compliance 23.2 ml/cm H2O with size 10 Magill tube. When the pressure was set at 25 the flow pattern at flow setting 6 showed that the total flow into the "patient" rose rapidly to 25 l./min but quickly fell to a constant flow of 8 l./min which was continued for a long inspiratory period. The flow from the Bird rose to a level below the total flow, the difference being due to the contribution from the nebulizer oxygen (fig. 3). In addition the flow from the Bird (oxygen-air mixture) quickly fell to zero and in Bird No. 1 there was actually a small flow towards the ventilator from the nebulizer. Once flow from the Bird had stopped, the constant flow from the nebulizer continued until the pressure in the system rose to that set for cycling. The entrainment of air followed the flow from the Bird but when flow from the Bird to- TIME 3 Flow (l./min) and pressure (mm Hg) produced by the Bird Mark 7 ventilator No. 1 at a fixed pressure setting of 25 and flow settings (F) of 6, 10, 20, 30 and 40. The compliance of the rigid drum was 23.2 ml/cm H,O and the airway represented by a size 10 Magill tube. The flow patterns at each flow setting consist of three superimposed tracings from above down. Total flow into "patient" (nebulizer oxygen + airoxygen mixture) from position (d) (fig. 2). Flow from ventilator (air-oxygen mixture only) from position (c) (fig. 2). Flow through air filter (entrained air) from position (a) (fig. 2). The area of vertical hatching represents the contribution of the nebulizer oxygen, and the horizontal hatching the contribution of atmospheric air. FIG. 662 FIG. 4 Flow and pressure produced by the Bird Mark 7 ventilator No. 2 with a fixed flow setting of 20 and pressure settings (P) of 15, 25, 40 and a pressure setting slightly greater than 40 (P>40). The tracings are represented in the same manner as figure 3. steady 8 l./min. The flow from the nebulizer and Bird stopped as the cycling pressure was reached. At flow setting 30, as the total flow increased the nebulizer contribution remained the same but the contribution from the Bird, including the entrained air, had only just begun to fall as the cycling pressure was reached. At flow setting 40 the situation was similar but the inspiratory period had become very shon and the cycling pressure fell. Apparently the balance of the magnets within the machine was upset by the high flow of gases with resultant early cycling. With the flow setting kept constant at 20 and the pressure setting raised in stages to 15, 25, 40, > 4 0 , a picture similar to the above was seen (fig. 4). As the pressure was raised the contribution of the nebulizer remained constant but flow from the Bird with the entrained air fell to zero and the "patient" was inflated with nebulizer oxygen only and oxygen flowed out the air filter. Increasing the pressure only increased the duration of this state as the pressure in the system rose slowly due to the constant oxygen flow until cycling occurred. BRITISH JOURNAL OF ANAESTHESIA 6 10 30 30 40 Flow setting FIG. 5 The percentage of atmospheric air in the air-oxygen mixture delivered by three Bird Mark 7 ventilators at flow settings of 6, 10, 20, 30 and 40. The volumes were measured from the areas under the flow-time tracings from positions (c) and (a) (fig. 2). These findings on the contribution of the nebulizer system were confirmed by the flow through pneumotachometer in position (b) (fig. 2). Entrained air. In all of the ventilators the percentage of air in the air-oxygen mixture leaving the Bird before reaching the nebulizer rose as the flow rate was increased (fig. 5) but there was considerable variation in the percentage amongst the three machines. Bird 3 in particular had a low percentage of entrained air with a higher mean inspired oxygen concentration than Birds 1 and 2. Effect of a decreased compliance (compliance 12.5 ml/cm H2O with size 10 Magill tube). At flow settings 6 and 10 flow from the ventilator stopped as the pressure rose and the "patient" was then inflated with 100 per cent oxygen from the nebulizer while the oxygen flowing through the injector passed out through the air filter (fig. 6). The duration of this state was shorter than HIGH OXYGEN MIXTURES FROM BIRD MARK 7 VENTILATOR 663 8 l./min. The pressure and flow patterns otherwise were similar to the patterns produced by a size 10 Magill tube (fig. 6). Entrained air. In all three ventilators there was a fall of 1-3 per cent in the contribution of the entrained air to the air-oxygen mixture at any particular flow setting when the compliance was reduced or the resistance increased. This probably resulted from the decreased flow through the injector. TMElMoordU 6 Comparative flow and pressure tracings from the Bird Mark 7 No. 3 for a compliance of 23.2 ml/cm HjO and size 10 Magill tube (C23 RIO), a compliance of 23.2 ml/cm H,O and a size 3 Magill tube (C23 R3), and a compliance of 12.5 ml/cm H,O and a size 10 Magill tube (C12 RIO). The pressure setting was fixed at 20 and flow setting varied. The results at flow settings 6 and 30 only are shown. FIG. Validity of some of the observations in patients. To confirm in patients the occurrence of the flow patterns seen in the experiments with models, Fleisch pneumotachometers were inserted on the ventilator side and patient side of the nebulizer in patients being ventilated by Mark 7 Bird ventilators after cardiac surgery. Similar patterns were seen to those occurring in the experiments with models (fig. 7). DISCUSSION It appears that there are several reasons for the high oxygen mixtures delivered by the Bird Mark 7 when the air-mix control is used. As the oxygen-air mixture from the Bird and the oxygen flow from the nebulizer into the patient's lungs the pressure in the righthand with the higher compliance because the pressure chamber of the ventilator rises and flow through in the system rose more rapidly and the cycling the venturi falls. As the flow of oxygen into the pressure was reached sooner. However, as the patient from the nebulizer quickly becomes conpeak flow and duration of the flow of the oxygen- stant the concentration of oxygen flowing into the air mixture were markedly decreased the nebulizer patient rises as inflation proceeds. The lower the oxygen represented a higher percentage of the flow setting, and hence the flow through the total inspired mixture. At higher flow settings, the injector, the greater the percentage contribution flow from the Bird did not stop until the cycling of the nebulizer oxygen to the inhaled mixture. pressure was reached but the nebulizer oxygen There is some variation in the flow through the flow at 8 l./min represented a large part of the injector at any given flow setting in different reduced total flow and smaller tidal volume when Mark 7s which will contribute to the difference compared with the same flow setting with the from one Bird to another. higher compliance. When a graph of airway pressure and flow from the Bird through the venturi is drawn for Effect of an increased airway resistance (com- various flow settings (fig. 8) it is evident that the pliance 23.2 ml/cm H2O with size 3 Magill lower the flow setting the lower the pressure at which the flow from the venturi ceases as the tube). There was decrease in the total flow and tidal one-way valve in front of it shuts. If the cycling volume at all flow settings due to a decrease in the pressure is above that at which the injector flow oxygen-air flow from the ventilator associated with stops then inflation of the lungs can only continue a slightly more rapid rise in pressure in the system from the nebulizer oxygen. Similarly at a fixed but the flow of oxygen in the nebulizer remained pressure setting the flow from the Bird falls to P?0F6 PATIENT SIDE OF NEBULISER i 1 i . i.... 1 : , , ..;.... . . • : ..;... .. . . L i .i. ' ! i ' i 1 "" ..; ...1 ! . : . . ! . ! ' !.. . . ; j : BIRD SIDE OF NEBULISER I • • • • i • -I1 • • • .... S- i ... ! ! ! ii , I1'"L71 ...,. ..., ,.. . :. . !. ... 1 : : ; • : . , ! : • ' .. j ... .. . ; . • 1 1 • '••• ; • t. : . ........ i : . I ! . . . . 1 P2OF6 1j . . P2OF3O PATIENT SIDE OF NEBULISER P2OF3O FIG. 7 Pneumotachometer tracings from a patient who was being ventilated by a Bird Mark 7 through a size 10 oral Oxford tube after open heart surgery. Tracings A and C were taken on the patient's side of the nebulizer and tracings B and D on the Bird side. The pressure setting was fixed at 20 (P20). At low flow settings (F6) flow from the Bird stopped early in inspiration, leaving only th; oxygen from the nebulizer to inflate th: lungs. Flows at F6 and F30 were measured at different attenuations. BIRD SIDE OF NEBULISER TIME (seconds) HIGH OXYGEN MIXTURES FROM BIRD MARK 7 VENTILATOR 665 70 -i P40 60 - P>40 Flow (l/min) 50- 4020 Ftow (L/min) 30 Pressure (mm Hg) FIG. 9 The relationship between the flow of the air-oxygen mixture from the Bird Mark 7 No. 2 ventilator and the airway pressure at a fixed flow setting of 20. The pressure settings and the cycling pressure they produced are represented by the series of vertical lines (PI5, 25, etc.). At this flow setting, flow from the Bird stopped at a pressure setting of 25. At higher pressure settings, the cycling pressure was reached because of the flow of the nebulizer oxygen. 30- 20- 10 20 30 Pressure (mm Hg) FIG. 8 The relationship between the flow of the air-oxygen mixture from the Bird Mark 7 No. 1 and the airway pressure at flow settings of 6, 10, 20, 30 and 40. When the cycling pressure was set at 25 mm Hg (represented by the vertical line on the right), flow from the ventilator stopped at lower pressures at flow settings 6 and 10. The cycling pressure was reached because of the flow of oxygen from the nebulizer. zero as the pressure rises (fig. 9). Unless the cycling pressure is equal to or less than the pressure at which flow stops (P25 and P15), the nebulizer oxygen continues until the cycling pressure is reached (P40 and P>40). There is some difference in the pressure at which flow stops in different Birds. The lower the flow of oxygen through the injector the lower the percentage of air entrained into the air-oxygen mixture. The percentage at any flow varies in different Birds as suggested by Fairley and Britt (1964). When there is an increase in ±e airway resistance, the flow of the air-oxygen mixture is decreased for any given pressure reached in the airway (fig. 10). As a result, the nebulizer oxygen contributes a higher proportion of the reduced tidal volume. At any given flow setting the flow through the venturi stops at the same pressure as occurs with a lower airway resistance. At low flow settings a period of inflation by the nebulizer oxygen alone is likely to occur. When there is a decrease in compliance of the lungs, the results are similar (fig. 10), but because of the rapid rate of rise of the airway pressure the duration of the changes is shorter and the tidal volume is considerably reduced. Besides the possibility of oxygen toxicity in the respiratory tract, the ventilator-nebulizer characteristics of the Mark 7 can probably lead to an inspired mixture low in water vapour content. This appears likely to occur at low flow and high pressure settings when the common practice is adopted of humidifying the air-oxygen mixture but leaving the nebulizer empty (as in fig. 1). Attempts to increase the tidal volume in patients BRITISH JOURNAL OF ANAESTHESIA 666 C23R3 3 W C12RK) 15 PRESSURE (mm Hg) FIG. 10 Relationship between the flow of the air-oxygen mixture from the Bird Mark 7 ventilator and the airway pressure with (a) a compliance of 23.2 ml/cm H , 0 and a size 10 Magill tube (C23 RIO), (b) a compliance of 23.2 ml/cm H,O and a size 3 Magill tube (C23 R3), (c) a compliance of 12.5 ml/cm H,O and a size 10 Magill tube (C12 RIO). with a high airway resistance or diminished compliance by decreasing the flow setting or increasing the pressure setting will increase the contribution of the nebulizer oxygen until the patient may be inflated with completely dry and undiluted oxygen. ont £t£ drudie'es par la mesure du flux gazeux dans diffirentes parties du ventilateur. Les principales causes sont les suivantes: le flux constant de 100% d'oxygene a partir du ne'buliseur, la diminution et eventuellement rarre't du flux a travers l'injecteur a mesure que la pression montait dans les voies aeriennes et enfin la diminution de la proportion d'air entrain^ pour les flux faibles. ACKNOWLEDGEMENTS This work was supported by a grant from the Research Committee of St. Vincent's Hospital. My thanks are due to Mrs. M. McCutcheon for the preparation of the illustrations. REFERENCE Fairley, H. B., and Britt, B. A. (1964). The adequacy of the air-mix control in ventilators operated from an oxygen sourc:. Canad. med. Ass. J., 90, 1394. LES MELANGES A FORT POURCENTAGE D'OXYGENE FOURNIS PAR LE OONTROLE AIR-MELANGE DES VENTILATEURS DU TYPE BIRD MARK 7 SOMMAIRE Les raisons pour lesquelles un ventilateur du type Bird Mark 7 foumit des melanges a forte teneur en oxygene DIE HOHEN SAUERSTOFF-GEMISCHE, DIE AUS DER LUFTGEMISCHKONTROLLE VON BIRD MARK 7—BEATMERN FREIGEGEBEN WERDEN ZUSAMMENFASSUNG Die Griinde fur die hohen Sauerstoff-Gemische, die aus dtm Bird Mark 7—Beatmungsgerat freigegeben werden, wurden durch Messung der Gasstrdmungen in den verschiedenen Teilen des GerStes untersucht. Die Hauptursachen waren die standige Stromung von lOOprozentigem Sauerstoff aus dem Vernebler, die Verminderung und das schliefihche Aufhfiren des Stromes durch den Injektor, wahrend der Druck in den Luftwegen anstieg, und die Abnahme des Anteils an aufgeladener Luft bei niedrigen Stromungseinstellungen.
© Copyright 2025 Paperzz