APPLICATION NOTE Thermal cyclers Thermal cycler sample amplification uniformity: a comparison of several models Introduction Thermal cycler well-to-well amplification uniformity is a critical factor that can affect analysis of PCR results. Without uniform reaction conditions across the block, users cannot make robust conclusions when comparing data obtained from wells at two different locations. This study compares the reaction yield of several thermal cyclers by measuring post-reaction absolute fluorescence. Well-to-well amplification uniformity is then compared by calculating the coefficient of variance (CV) of fluorescence readings obtained by each thermal cycler. Materials and methods The instruments tested in this study are shown in Table 1. The same equipment, methods and reagents were use for each instrument. Table 1. Instruments tested for amplification uniformity. Manufacturer Model name Cat. No. Thermo Fisher Scientific Applied Biosystems™ ProFlex™ 96-Well PCR System 4484075 Thermo Fisher Scientific Applied Biosystems SimpliAmp Thermal Cycler A24811 Bioer GeneMax™ Thermal Cycler BYQ6067 Bio-Rad C1000 Touch Thermal Cycler with 96-Well Fast Reaction Module 185-1196 Bio-Rad T100™ Thermal Cycler 186-1096 Eppendorf ™ Mastercycler Nexus Gradient 6331 000.017 Eppendorf Mastercycler™ Nexus GX2 6336 000.015 Eppendorf Mastercycler Pro S 6325 000.510 SensoQuest Labcycler Gradient 011-101, 012-103 Takara Dice Touch TP350 ™ ™ ™ ™ ™ Sample preparation A single bulk reaction was prepared using Applied Biosystems™ AmpliTaq Gold™ 360 Master Mix (Cat. No. 4398886) according to the standard protocol. Invitrogen™ Lambda DNA Standard (component C) (Cat.No. P7589) was used as the template at a final concentration of 0.01 ng/µL with forward primer (5´-GATGAGTTCGTGTCCGTACAACT-3´) and reverse primer (5´-ACGGCTGCACGGAGTTCAGTATG-3´) at 0.2 µM. The bulk reaction was divided between nonoptical 8-tube strips and non-optical 96-well plates (not created by the instrument manufacturers) at 10 µL or 100 µL testing volume. The following thermal profile was used for all instruments in this study: a primary stage at 94°C for 10 minutes, a secondary stage of 25 cycles at 94°C for 15 seconds and 70°C for 90 seconds, and a final stage at 72°C for 7 minutes. Data acquisition and analysis After completion of thermal cycling, Applied Biosystems™ Power SYBR™ Green PCR Master Mix (Cat. No. 4368577) was added to each reaction and incubated for 10 minutes. The reactions were then transferred to an Applied Biosystems™ MicroAmp™ Optical 96-Well Reaction Plate (Cat. No. N8010560) for fluorometric analysis with the Applied Biosystems™ ViiA™ 7 Real-Time PCR System (Cat. No. 4453534). Amplification uniformity was calculated by %CV as stated in the following formula: %CV = Standard deviation of SYBR Green signal over 3 runs Average SYBR Green signal over 3 runs x 100% Results Figure 1 shows the average fluorescence intensity obtained from each thermal cycler over two reaction volumes and two plastic consumable types. The individual values are specified below the chart, and the group average of 7,253,062 fluorescence units is indicated as a red line. Figure 2 shows the %CV for each plastic consumable and reaction volume. The %CV group average of 10.3% is shown as a red line. Average fluorescence intensity 14,000,000 Fluorescence intensity 12,000,000 10,000,000 8,000,000 7,253,062 6,000,000 4,000,000 2,000,000 0 ProFlex 96-Well SimpliAmp PCR System Thermal Cycler GeneMax Thermal Cycler C1000 Touch Thermal Cycler T100 Thermal Cycler Mastercycler Nexus GX2 Mastercycler Nexus Gradient Mastercycler Pro S Labcycler Gradient Dice Touch 10 μL, 96-well plate 8,051,637 7,910,946 9,239,160 7,946,722 8,504,935 ND 8,819,051 8,920,168 5,550,320 8,055,912 10 μL, 8-tube strips 7,121,495 7,367,679 8,041,636 7,175,663 7,869,934 7,815,814 7,170,831 8,212,778 6,656,877 7,514,640 100 μL, 96-well plate 9,581,911 10,207,855 7,977,774 8,614,839 5,016,273 ND 3,116,432 6,761,303 2,931,623 3,624,982 100 μL, 8-tube strips 10,470,426 11,303,278 9,375,210 9,690,691 6,355,783 3,785,679 3,273,543 8,073,197 3,265,825 4,243,518 Instrument Figure 1. Average fluorescence intensity obtained from thermal cyclers. ND = not determined. Uniformity measured by %CV 25.0 20.0 %CV 15.0 10.3 10.0 5.0 0.0 ProFlex 96-Well SimpliAmp PCR System Thermal Cycler GeneMax Thermal Cycler C1000 Touch Thermal Cycler T100 Thermal Cycler Mastercycler Nexus GX2 Mastercycler Nexus Gradient Mastercycler Pro S Labcycler Gradient Dice Touch 22.0 10 μL, 96-well plate 5.8 5.2 5.1 10.6 14.9 ND 7.0 6.7 11.1 10 μL, 8-tube strips 7.3 9.0 11.0 5.8 12.2 4.9 6.3 6.8 6.3 9.7 100 μL, 96-well plate 7.3 7.3 12.7 7.5 15.0 ND 8.6 14.7 6.7 21.6 100 μL, 8-tube strips 9.6 8.0 15.6 6.0 20.0 13.1 10.5 12.6 7.2 20.8 Instrument Figure 2. Average %CV of thermal cyclers. ND = not determined. Discussion The ability of a thermal cycler to uniformly amplify each well of the reaction plate is a pivotal and often overlooked requirement for the user when designing experiments. In this study, we have shown that PCR amplification uniformity varies between thermal cyclers. This was demonstrated as a side-by-side comparison while utilizing the same reaction chemistry. For best results, we recommend using a thermal cycler with both high reaction yield and good wellto-well uniformity (as indicated by low %CV). The combined results for each volume and plastic consumable suggest that the Applied Biosystems™ brand of thermal cyclers meet these criteria. Find out more at thermofisher.com/thermalcyclers For Research Use Only. Not for use in diagnostic procedures. © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. AmpliTaq Gold is a trademark of Roche Molecular Systems, Inc. C1000 Touch and T100 are trademarks of Bio-Rad Laboratories, Inc. Mastercycler is a trademark of Eppendorf AG. Dice is a trademark of Takara Bio Inc. GeneMax is a trademark of Bioer. CO210556 0715
© Copyright 2026 Paperzz