Technical specifications
Ref.
XLS
392
Internal dimensions - without work surface
Width
Depth
Height*
Width
Depth
Height
39 1/2’’
31 1/2’’
51 3/4’’ min
59 7/8’’ maxi
38’’
26 1/4’’
41’’
*According to filtration column type
Total power consumption
(Including electricity for the lights)
Structure
Anti-corrosion metallic alloy protected by a thermo-hardened
anti-acid polymer coating
Metallic parts
Polypropylene
Filtration module
Amperage absorbed
38 - 88 Watts
3.6 amp.
Modular filtration column
1/4’’ thick acrylic
Side and front panels
Modular filtration column
2
Number of fans (IP44)
2
Processed air flow
270 cfm
79 fpm to 118
Air velocity at the openings in
working position
fpm
Voltage / Frequency
90 - 264V / 50 - 60 Hz
Number of columns
Type 1P
Class 2 - (NFX 15-211 : 2009 standard)
Class 1 - (NFX 15-211 : 2009 standard)
Type 1C
Type 2C
Liquid chemicals
handlings
Liquid chemicals
handlings
Type 1P 1C
Type 1P 2C
Liquid chemicals
and powders
handlings
Liquid chemicals
and powders
handlings
Type 1C 1P
Type 2C 1P
Liquid chemicals
and powders
handlings in clean
room
Liquid chemicals
and powders
handlings in clean
room
Powders handlings
Molecular filter
Equipments
H14 HEPA filter
Types of filters available
Standard Equipment
Options
GF4 AS For organic vapours
Air flow meter
Permanent air face velocity monitoring
system
Molecode S*
Automatic alarm to detect filter saturation by solvents according to
NF X 15 211 : 2009 standard (*compulsory for class 1 configuration)
Flow monitor
Permanent ventilation control device
according to EN 14175-2:2003 standard
Work surface #1
Tempered glass work surface with seeled frame
GF4 F
For formaldehyde vapours
Work surface # 2
In phenolic resin with built-in retention tray
GF4 K
For ammonia vapours
Internal lighting
Compact tubular fluorescent lighting 18 Watts - 500 Lux - IP 67
Benchcap
Fixed work bench made of anti-corrosion metallic alloy
Sampling port
For filter saturation detection (N/A if
Molecode S installed)
Adjustable slinding shelf for Mobicap and Benchcap
Programmable timer
Timer - 60 hours setting by default according to NFX 15 211: 2009 standard
Energy ports
For the passage of cables
Chemical Listing
List of approved chemicals
Shelf
Transparent back
panel
Access door on
rear panel
Particular
Pre-filter
GF4 BE + For organic vapours + acid vapours
GF4 HP
Compliance to standards
Acrylic
For maintenance interventions
40 % filtration efficiency for particles > 0.3 µm
H14 HEPA Filter for powders
AFNOR NF X 15-211:2009 - France
ANSI/ASHRAE 110-1995 - USA
BS 7258 - England
CSA Z316.5 - Canada
ERLAB INC. - 388 Newburyport Turnpike - Rowley, MA 01969 - USA
Phone: 1 (800) 964-4434 - Fax: 1 (978) 948-3354 - e-mail : [email protected]
P Please consider the environment before printing this document.
CE Marking
NF P.92.507 (Acrylic part)
EN 1822 : 1998 (H14 HEPA filter )
w w w. c ap t a i r. c o m
Technical data are not contractual. Erlab® reserves the right to modify the present document without prior notice.
Specifications
Dimensions (inches)
External dimensions
Mobile ductless filtration fume hood for personal protection
during handlings of liquid, gaseous and solid chemicals.
Molecular filtration by adsorption on activated carbon filter
and particulate filtration on H14 HEPA filter.
XLS 392
37 3/8’’
39 1/2’’
25 3/4’’
59 7/8’’ max
22 1/4’’
10 3/4’’
3’’
26 1/4’’
35 1/2’’
31 1/2’’
* Total height according to filtration column type (in)
Type 1P
Type 1C
Type 1P 1C
Type 1C 1P
51 3/4’’
51 3/4’’
57 1/2’’
57 1/2’’
Type 2C
Type 1P 2C
Type 2C 1P
57 1/2’’
59 7/8’’
59 7/8’’
ERLAB INC. - 388 Newburyport Turnpike - Rowley, MA 01969 - USA
Phone: 1 (800) 964-4434 - Fax: 1 (978) 948-3354 - e-mail : [email protected]
P Please consider the environment before printing this document.
w w w. c ap t a i r. c o m
Technical data are not contractual. Erlab® reserves the right to modify the present document without prior notice.
51 3/4’’ min
41’’
XLS 392
Mobile ductless filtration fume hood for personal protection
during handlings of liquid, gaseous and solid chemicals.
Molecular filtration by adsorption on activated carbon filter
and particulate filtration on H14 HEPA filter.
38’’
Work surface #1
24 3/4’’
35 3/8
Work surface #2
38’’
26 5/8’’
35 3/8’’
1 1/4’’
32 3/4 ‘’
ERLAB INC. - 388 Newburyport Turnpike - Rowley, MA 01969 - USA
Phone: 1 (800) 964-4434 - Fax: 1 (978) 948-3354 - e-mail : [email protected]
P Please consider the environment before printing this document.
w w w. c ap t a i r. c o m
Technical data are not contractual. Erlab® reserves the right to modify the present document without prior notice.
2 3/8’’