Star Series Method Note M-1800 Fluoride in Toothpaste Key Words Introduction • Fluoride • Toothpaste • Fluoride Ion Selective Electrode (ISE) Fluoride is incorporated in oral health products in several chemical forms, including sodium fluoride, sodium monofluorophosphate, amine fluoride and stannous fluoride. This method provides rapid and uncomplicated determinations for fluoride in the form of sodium fluoride in toothpaste. This procedure eliminates interferences associated with the presence of color agents and solids. The direct measurement of fluoride ions with the fluoride ion selective electrode is a well-established technique that is frequently used as a standard method of analysis. Recommended Equipment Cat. No. 1. 4-Star Plus benchtop pH/ISE meter 1115000 2. Fluoride combination electrode 9609BNWP 3. Stirrer probe (or magnetic stir plate and bar) 096019 4. Swing arm electrode stand 090043 5. 500 mL volumetric flasks 6. 100 mL and 200 mL beakers 7. 50 mL graduated cylinders Required Solutions Cat. No. 1. 1 ppm fluoride standard with TISAB II 040906 2. 2 ppm fluoride standard with TISAB II 040907 3. 10 ppm fluoride standard with TISAB II 040908 4. Optimum Results™ A electrode filling solution 900061 5. TISAB II solution 940909 6. Deionized water Calibration Standard Preparation 1. Using a graduated cylinder, transfer 25 mL of the 1 ppm (mg/L) fluoride standard into a beaker. The 1 ppm (0.0001 % w/v) standard will be used as “0.002 % w/v” for the calibration. The “0.002 % w/v” value takes into account and corrects for the 1:20 dilution of the sample (25 grams dispersed in 500 mL of deionized water). 2. Using a graduated cylinder, transfer 25 mL of the 10 ppm (mg/L) fluoride standard into a beaker. The 10 ppm (0.0010 % w/v) standard will be used as “0.020 % w/v” for the calibration. The “0.020 % w/v” value takes into account and corrects for the 1:20 dilution of the sample (25 grams dispersed in 500 mL of deionized water). 1 of 4 Fluoride in Toothpaste Sample Preparation 1. Mix 25 grams of toothpaste in about 150 mL of deionized water. 2. Quantitatively transfer the mixture to a 500 mL volumetric flask. Dilute to the mark with deionized water and thoroughly mix the solution. 3. Using a graduated cylinder, transfer 25 mL of the prepared sample into a beaker. Using a graduated cylinder add 25 mL of the TISAB II solution. 4. Repeat steps 1 through 3 for additional samples. Results Five aliquots of diluted toothpaste were measured for fluoride. The data in the table below is representative of the results expected for fluoride in toothpaste. Toothpaste with Sodium Fluoride Fluoride (% w/v) Sample # 1: 0.111 Sample # 2: 0.110 Sample # 3: 0.111 Sample # 4: 0.110 Sample # 5: 0.110 Mean: 0.110 Standard Deviation: 0.001 %CV: 0.50 Electrode Storage For short-term storage between sample measurements, store the electrode in the 10 ppm (0.0010 % w/v) fluoride standard. The filling solution in the fluoride electrode should not be allowed to evaporate, which would cause crystallization. For long-term storage over one week, refer to the electrode use guide. Fluoride electrodes, whose response may become sluggish with time, can often be restored to working order by brushing the sensing element with fluoride toothpaste and a soft brush. Equipment Setup Electrode Setup – Fluoride Electrode 1. Remove the protective shipping cap that covers the electrode tip. 2. Attach the flip-spout cap to the Optimum Results A filling solution bottle and lift the spout to a vertical position. 3. Insert the spout into the fill hole in the outer sleeve of the electrode, add a small amount of filling solution to the chamber, tip the electrode upsidedown to wet the O-ring and return the electrode to its original, vertical position. 4. Holding the electrode by the body with one hand, use the thumb to push down on the electrode cap and allow a few drops of filling solution to drain wetting the inner cone. 5. Release the electrode cap and allow the outer sleeve to return to its original position. If the outer sleeve does not return to its original position immediately, check to see if the O-ring is moist and repeat steps 3 through 5 until the outer sleeve returns to its original position. 2 of 4 6. Add filling solution to the electrode until the amount of filling solution within the electrode body is level with the bottom edge of the electrode fill hole. Meter Setup – 4-Star Plus Benchtop pH/ISE Meter For initial meter setup, follow the steps in the quick start guide that is included with the meter. The quick start guide also contains a layout of the meter keypad for reference. The words in all capital letters, such as POWER, indicate a key on the meter and words in quotations, such as “UnIt”, indicate a display on the meter screen. 1. Connect the electrode to the meter. 2. Connect the stirrer to the meter. 3. Press the POWER key on the meter to turn the meter on. 4. Note that the arrow icon on the left of the screen indicates the active line. If the top line is not active, press the LINE SELECT key to change the selected line to the top line. Press the UP or DOWN ARROW key to change the measurement mode of the top line to ISE. 5. Press the SETUP key. Press the UP or DOWN ARROW key until “ISE” is displayed on the top line. 6. Press the LINE SELECT key to select the middle line. Press the UP or DOWN ARROW key to select “rES”, which is for ISE resolution and the number of significant digits that will be displayed on the meter. Press the LINE SELECT key to select the bottom line. Press the UP or DOWN ARROW key to select “3” for “rES”. Press the LINE SELECT key to accept the setting. 7. Press the LINE SELECT key to select the middle line. Press the UP or DOWN ARROW key to select “nLIn”, which is an abbreviation for non-linear blank correction. Press the LINE SELECT key to select the bottom line. Press the UP or DOWN ARROW key to select “OFF” for “nLIn”. Press the LINE SELECT key to accept setting. 8. Press the LINE SELECT key to select the middle line. Press the UP or DOWN ARROW key to select “rAng”, which is an abbreviation for the measurement range type. Press the LINE SELECT key to select the bottom line. Press the UP or DOWN ARROW key to select “HIgH” for “rAng”. Press the LINE SELECT key to accept setting. 9. Press the LINE SELECT key to select the middle line. Press the UP or DOWN ARROW key to select “UnIt”. Press the LINE SELECT key to select the bottom line. Press the UP or DOWN ARROW key to select “PEr” for “UnIt”. Press the LINE SELECT key to accept setting. 10. Press the MEASURE key to return to the measurement mode. 11. If all steps were followed correctly, the meter display will show three digits in the top line and “ISE: %” to the right of the top line. The meter and electrode are now ready for calibration. 3 of 4 Calibration and Analysis 1. Allow all the standards and the samples to reach room temperature for precise measurements, since fluoride measurements are temperature sensitive. 2. Prepare the 0.002 % w/v (1 ppm) and 0.020 % w/v (10 ppm) fluoride standards as described in the Calibration Standard Preparation section and stir the standards at a moderate and uniform rate. 3. Place the electrode and stirrer into the beaker with the 0.002 % w/v standard so the electrode tip is fully immersed in the solution. The stirrer should be positioned slightly below the tip of the electrode. Press the STIR key on the meter to turn the stirrer on. Press the CALIBRATE key. 4. Wait for a stable reading, usually in 1 to 2 minutes. Enter the standard value of “0.002” using the DIGIT key and UP and DOWN ARROW keys. Press the DIGIT key until the first digit to be changed is flashing, press the UP and DOWN ARROW keys to change the value of the flashing digit and continue to change the digits until the meter displays the correct value of the standard. Once the value is set, press the DIGIT key until the decimal point is in the correct location. 5. Press the CALIBRATE key to accept the 0.002 % w/v standard and press the STIR key to turn the stirrer off. 6. Rinse the electrode and stirrer thoroughly with deionized water. Gently remove excess solution from the outer sleeve of the electrode by dabbing it with a clean paper tissue. Do not wipe or rub the sensing element of the electrode. 7. Place the electrode and stirrer into the beaker with the 0.020 % w/v standard so the electrode tip is fully immersed in the solution. The stirrer should be positioned slightly below the tip of the electrode. Press the STIR key on the meter to turn the stirrer on. 8. Wait for a stable reading, usually in 1 to 2 minutes. Enter the standard value of “0.020” using the DIGIT key and UP and DOWN ARROW keys. 9. Press the MEASURE key to accept the 0.020 % w/v standard and return to the measurement mode. Press the STIR key to turn the stirrer off. 10. Rinse the electrode and stirrer thoroughly with deionized water. Gently remove excess solution from the outer sleeve of the electrode by dabbing it with a clean paper tissue. Do not wipe or rub the sensing element of the electrode. 11. Prepare a sample as described in the Sample Preparation section. 12. Place the electrode and stirrer in a prepared sample with the electrode tip fully immersed in the solution. Press the MEASURE key on the meter and the stirrer will turn on. The ISE:% icon will flash as the measurement is being made. The ISE:% icon will become solid and the display value will freeze when a stable reading is achieved. The stirrer turns off automatically. 13. Repeat steps 10 through 12 for additional samples. When all samples have been measured, rinse the electrode with deionized water and store the electrode according to the Electrode Storage section. ©2008 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. Environmental Instruments Water Analysis Instruments North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031 Europe Denmark House, Angel Drove Ely, Cambridgeshire England, CB7 4ET Tel: 44-1353-666111 Fax: 44-1353-666001 Asia Pacific Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836 www.thermo.com/water 257483-001 Rev. A 07-08
© Copyright 2026 Paperzz