LA-UR-02-2530 FRAM Isotopic Analysis of Uranium in Thick-Walled Containers Using High Energy Gamma Rays and Planar HPGe Detectors Thomas E. Sampson, Phillip A. Hypes, and Duc T. Vo Safeguards Science and Technology, Group NIS-5 Nonproliferation and International Security Division Los Alamos National Laboratory Los Alamos, New Mexico 87545 USA Abstract We describe the use of the Los Alamos FRAM isotopic analysis software to make the first reported measurements on thick-walled UF6 cylinders using small planar HPGe detectors of the type in common use at the IAEA. Heretofore, planar detector isotopic analysis measurements on uranium have used the 100-keV region and can be defeated by 10 mm of steel absorber. The analysis of planar detector measurements through 13-16 mm of steel shows that FRAM can successfully carry out these measurements and analysis in the 120-1024 keV energy range, a range previously thought to be the sole province of more efficient coaxial detectors. This paper describes the measurement conditions and results and also compares the results to other FRAM measurements with coaxial HPGe detectors. Introduction The technique of gamma-ray isotopic analysis of arbitrary samples is desirable for measuring the isotopic composition of uranium in UF6 cylinders because it does not require calibration with standards or knowledge of the cylinder wall thickness. The International Atomic Energy Agency (IAEA) uses the MGAU (Multi Group Analysis Uranium) uranium isotopic analysis software [1] with planar high purity germanium (HPGe) detectors to measure the isotopic composition of uranium. Measurements on UF6 cylinders with 13-16-mm thick steel walls are usually unsuccessful because of the strong absorption of the 89-100 keV gamma rays and x-rays that MGAU requires for the measurement. This paper describes the use of the Los Alamos FRAM isotopic analysis software [2-6] to make these measurements on UF6 cylinders. Uranium measurements with FRAM typically cover the energy range from 120-1001 keV and can easily be made through the walls of UF6 cylinders. While these measurements are usually performed with efficient coaxial HPGe detectors, this paper reports the first successful measurements using small planar HPGe detectors of the type in common use at the IAEA. Background Gamma ray isotopic analysis of arbitrary samples of uranium was first demonstrated at Los Alamos over 12 years ago [7]. The technique used the same FRAM analysis software that had been developed for plutonium isotopic analysis and measured uranium gamma rays in the energy range from 143 keV to 1001 keV. The Los Alamos uranium measurements using FRAM were typically performed using a modest-sized coaxial HPGe detector of about 50-mm diam by 50-mm thick with an efficiency of 25-30% in the usual definition. This type of detector has adequate high-energy efficiency and still provides good resolution (< 750 eV at 122 keV) at lower energies. MGAU measurements used by the IAEA are typically performed with planar HPGe detectors of the same type used for its plutonium sister code, MGA [8]. These detectors can be various sizes but a common size of 25-mm diam by 15-mm thick is widely used by the IAEA. Requested Task The task requested by the IAEA was to characterize the performance of FRAM for measuring the isotopic composition of uranium under three constraints 1. Must work through 13-16 mm of steel. 2. Must use planar HPGe detectors in common use at IAEA (25-mm diam by 15-mm thick) and possibly also 60-mm diam by 20-mm thick detectors. 3. Must work with multichannel analyzers with a maximum of 4096 channels of data storage. 1 LA-UR-02-2530 Previous FRAM Experience Coaxial Detector Los Alamos had previously demonstrated isotopic analysis of uranium through thick steel with coaxial detectors and 8192 data channels. Figure 1 below shows the geometry of these laboratory measurements. The samples were the EC-171/NBS-969 uranium reference materials certified for 235U isotopic abundance. The detector was a Canberra coaxial detector with dimensions 60 mm diam by 42 mm long with 28.4% relative efficiency, in the usual definition. EC-171/NBS-969 071 (Natural U) 28% rel eff Coax HPGe Fig. 1. Measurement geometry with detector, steel absorber, and sample simulating the measurement of a UF6 cylinder. 17 mm Steel We made 15 replicate measurements on each standard in the set. The relative standard deviation of a single measurement obtained from the distribution of the replicates is plotted in Fig. 2. Fig. 2. Single measurement precision of the 235U fraction calculated from 15 replicates. The five standards were measured in 8192 channels with a 28% relative efficiency coaxial HPGe detector and analyzed with the FRAM software. Large Diameter Semi-Planar Detector We repeated these measurements using a detector that was 70 mm diam by 30 mm thick to see if the larger detector diameter would improve the peak-to-background ratio at 258 keV. We obtained the same results as for the coaxial detector. This is thought to occur because much of the continuum underneath the 258-keV peak arises from bremsstrahlung, not Compton scattering. Coaxial Detector Measurements on 30B UF6 Cylinders Los Alamos has made measurements, analyzed with FRAM, on type 30B UF6 cylinders at the ABB Atom fuel fabrication facility in Vasteras, Sweden [9]. A type 30B UF6 cylinder has 13-mm thick steel walls. These measurements were made with a 26% relative efficiency coaxial detector in 8192 channels. The measurements on 30B cylinders showed a measurement precision or repeatability of 6-8 % (1 RSD) for 20-30 min 2 LA-UR-02-2530 measurements covering enrichments from natural to 4.7%. This is consistent with the results in Fig. 2 and even better than Fig. 2 for natural uranium. Considerations for Uranium Measurements with FRAM [9-10] Coincidence Summing One of the principal considerations is the sample-detector distance sensitivity of the measurement that arises from coincidence summing. Vo and Sampson [5] have completed a solution for this problem that is implemented in version 4 of FRAM. The planar detector data to be presented were analyzed with a version of FRAM that incorporated these corrections. Relative Efficiency Curve Another potential difficulty for uranium isotopic analysis with FRAM occurs because of the large “gap” in the relative efficiency curve between the 235U peaks at low energies and the 234mPa (238U daughter) peaks at high energy. The coaxial detector data discussed above have all been successfully analyzed with the standard empirical relative efficiency curve that has worked well in all FRAM versions to date. However, for weak uranium spectra we feel that more robust relative efficiency formalism is appropriate. We have developed an improved, physics-based relative efficiency curve [10] that greatly aids in spanning this gap and preserving the correct analysis for the ratio of 235U/238U. This formalism has been used in the analysis of the planar HPGe data discussed later. The 258-keV Peak from the 234mPa Daughter of 238U This peak plays an important role in FRAM for defining the relative efficiency curve for uraniumbearing materials. Data points from 235U at energies of 143-, 163-, 186-, and 205 keV usually define the lowenergy portion of the relative efficiency curve. The high-energy portion of the curve is usually defined by a series of gamma rays from 234mPa beginning at about 700 keV with only the single point at 258 keV to make the normalization between the high- and low-energy portions of the curve. The importance of the 258-keV peak in establishing the relative efficiency curve is illustrated in Fig. 3. Relative Efficiency Fig. 3. The relative efficiency curve for a depleted uranium sample (0.31% 235U) measured with a coaxial HPGe detector. There was no steel “shielding” present for this measurement. This relative efficiency curve uses the improved physics-based relative efficiency formalism now incorporated in FRAM. Energy (keV) Also complicating the measurement is the very weak intensity of the 258-keV peak that also rides on an intense background continuum. This peak is strongly attenuated by the steel walls of a UF6 cylinder and also becomes vanishingly small and cannot be used for highly enriched uranium samples. Figures. 4–5 display the relation of the 258-keV peak to the remainder of the uranium gamma ray spectrum. As an example Table I displays (Net Peak/Background) data where Net Peak is the height of the 258-keV peak above background. The attenuation at 258 keV caused by the steel is fairly consistent with that calculated from attenuation coefficients. Table I. Net Peak/Background at 258 keV for Natural Uranium, Nominal 25% Relative Efficiency Coaxial HPGe detector @ 0.125 keV/ch Conditions Net Peak/Background 200 g U3O8, NBS 071, bare 1330 g U3O8, 16-mm steel 1330 g U3O8, bare 30B UF6 cyl, 13-mm steel 0.67 0.20 0.63 0.13 3 LA-UR-02-2530 Fig. 4. A gamma ray spectrum from a 200-g sample of natural uranium (235U = 0.71%) as U3O8 with the 258keV peak designated. Fig. 5. The 258-keV gamma ray for several different 235U enrichments. The plots are offset and the Y axis is logarithmic. The 258-keV peak cannot be used in the analysis of highly enriched uranium. Measurements with a Planar HPGe Detector Samples We chose a set of uranium samples from NIS-5’s extensive inventory of well-characterized uranium oxide samples [11, 12] with 235U enrichments from 0.3% to 10%. Their characteristics are given in Table II. Table II. Characteristics of U3O8 Uranium Reference Materials Sample U3O8 Mass (g) EC-171-031 A1-408-1 EC-171-194 EC-171-446 A1-324-1 200 1339 200 200 1170 wt% 234 U wt% 235 U wt% 236 U wt% 238 U 0.0020 0.0049 0.0171 0.0359 0.0512 0.3166 0.7135 1.9420 4.4623 10.0863 0.0146 0.0010 0.0003 0.0068 0.0903 99.6668 99.2805 98.0406 95.4950 89.7722 The EC171 set of standards is also available under the issue NBL-CRM-969. 4 LA-UR-02-2530 Detector The detector characteristics are given in Table III. Table III. Planar HPGe Detector Characteristics Manufacturer Model Serial No. Bias Voltage Diameter Thickness Resolution 122 keV, 3 µs, 1 kHz 122 keV, 1µs, 50 kHz ORTEC SH-SGD-25520-S 39-N21194D 2000 V Negative 25.4 mm 16.2 mm 509 eV 549 eV Data Acquisition Data acquisition and storage were provided by an ORTEC DSPEC Plus digital spectroscopy system. The system was operated with a rise time = 4.0 µs, cusp = 0.6 µs, and flat top = 0.8. The gross counting rates ranged from ~500 counts/s for the depleted U sample with 16-mm steel to ~1100 counts/s for the 10% enriched sample with 13-mm steel. We made 24 measurements on each sample at counting times (true time) of 60 min, 30 min, and 15 min. Each sample-count time sequence was carried out with absorbers of 16 mm and 13 mm of steel. The measurement geometry was identical to that shown in Fig. 1 except for the use of a planar detector. The measured results are the average of the 24 individual measurements and the % RSD is the relative standard deviation from the distribution of the 24 individual measurements. Tables IV-VI shows the results for the measurements at the three measurement times with the two absorbers. Table IV. Twenty-Four 60 Minute Measurements 16 mm Steel Absorber Sample EC-171-031 A1-408-1 EC-171-194 EC-171-446 A1-324-1 13 mm Steel Absorber Accepted Measured wt% 235U wt% 235U Meas./Accept. % RSD 0.3166 0.7135 1.9420 4.4623 10.086 0.2960 0.7009 1.9516 4.5219 10.271 0.9349 0.9823 1.0049 1.0134 1.0183 Average 0.9908 15.6 8.8 9.7 8.5 9.0 Table V. Twenty-Four 30 Minute Measurements 16 mm Steel Absorber Sample EC-171-031 A1-408-1 EC-171-194 EC-171-446 A1-324-1 Accepted wt% 235U 0.3166 0.7135 1.9420 4.4623 10.086 1.1317 1.0153 0.9792 1.0388 1.0325 Average 1.0395 48.2 15.0 13.0 10.6 12.1 5 Meas./Accept. % RSD 0.2955 0.7411 1.9533 4.5284 10.101 0.9333 1.0387 1.0058 1.0148 1.0015 13.8 9.5 6.6 6.6 7.7 Average 0.9988 13 mm Steel Absorber Measured wt% 235U Meas./Accept. % RSD 0.3583 0.7244 1.9016 4.6354 10.414 Measured wt% 235U Measured wt% 235U Meas./Accept. % RSD 0.3049 0.6938 1.9344 4.5710 10.246 0.9630 0.9724 0.9961 1.0244 1.0159 18.1 12.2 8.0 9.3 5.7 Average 0.9944 LA-UR-02-2530 Table VI. Twenty-Four 15 Minute Measurements 16 mm Steel Absorber Sample EC-171-031 A1-408-1 EC-171-194 EC-171-446 A1-324-1 Accepted wt% 235U 0.3166 0.7135 1.9420 4.4623 10.086 13 mm Steel Absorber Measured wt% 235U Meas./Accept. % RSD 0.3131 0.7231 1.9495 4.5950 10.093 0.9889 1.0135 1.0039 1.0297 1.0007 Average 1.0073 37.4 16.9 18.6 20.9 14.9 Measured wt% 235U Meas./Accept. % RSD 0.3556 0.7100 1.9968 4.3954 10.453 1.1232 0.9951 1.0282 0.9850 1.0364 27.9 14.1 15.7 11.9 8.0 Average 1.0336 Discussion These measurements on uranium using a planar HPGe detector operating at 0.25 keV/channel (4096 channels spanning 0–1024 keV) were very difficult for FRAM to analyze. The spectra were weak because of the small detector. Even more important was the low number of channels/keV resulting in very narrow peaks at low channel numbers for the high-resolution planar detector. The full width at half maximum for the 185 keV 235U peak was less than 3 channels. With the weak, narrow peaks it was not possible to perform a shape calibration on each spectrum. We used a fixed set of default peak shapes for the analysis of these data. This is standard procedure for analyzing weak spectra with FRAM. We also used a beta release of v4 of FRAM to analyze the spectra. FRAM version 4 has several features that enhanced the analysis of these data. It incorporates a coincidence summing correction to reduce the sample-detector distance sensitivity as well as using an improved physics-based relative efficiency function in the data analysis [5,9,10]. We found that both of these features were necessary for the analysis. The measurement results were nearly unbiased with quite large counting statistics uncertainties, especially for the lowest enrichments at the shortest counting times. Of the 720 measurements tabulated in Tables IV-VI, only one measurement was discarded for an obviously bad analysis, a tribute to the robustness of the FRAM software. The discarded measurement was for 16mm steel and a 15 minute count time on the depleted 235 U = 0.31 wt%--the worst case studied. Figures 6–8 display the 130 to 200 keV region of a 15 min measurement through 16 mm of steel on samples with 0.31, 1.94, and 10.1% 235U respectively. The depleted U (0.31% 235U) sample is on the edge of being able to be analyzed successfully as one of the 24 run sequence did not analyze correctly. Note the background continuum levels above the 185-keV peak. For low enriched samples we expect these to be fairly constant with respect to enrichment as this level is generally proportional to the 238U content, arising mainly from bremsstrahlung from 238U daughters. These levels are seen to be approximately the same for the 0.31% and 1.94% samples. The level for the 10.1% sample is about twice as high as for the lower-enriched samples, arising, we believe, from the fact that the 10.1% sample contains over five times as much uranium as the lower-enriched samples. This observation leads us to believe that measurements on actual UF6 cylinders containing depleted uranium will be more difficult than for the mockup measurements reported here because of the UF6 cylinder’s much greater uranium content. We also observe that the measurement repeatability for the depleted uranium sample, 16-mm steel, and 15-min count time was over 35% (1 RSD). The 2σ error bars for 15-min measurements on the depleted and natural uranium samples essentially touch but do not overlap. This is close enough to suggest that the ability to perform discrimination between depleted and natural uranium using planar HPGe detectors should be tested in the field on actual UF6 cylinders. 6 LA-UR-02-2530 1500 235 U = 0.31% Fig. 6. Region of the spectrum from sample EC-171-031 with 235U = 0.31%. Count time was 15 min. Steel thickness was 16 mm. This case was the most difficult to analyze as the 143- and 163-keV peaks are essentially not visible. Counts 1000 500 0 130 140 150 160 170 keV 180 190 200 10000 235 U = 1.94 % 8000 Fig. 7. Region of the spectrum from sample EC-171-194 with 235U = 1.94%. Count time was 15 min. Steel thickness was 16 mm. This spectrum is easily analyzed, as the 143-and 163-keV peaks are easily visible. Counts 6000 4000 2000 0 130 140 150 160 170 keV 180 190 7 200 LA-UR-02-2530 60000 235 U = 10.1 % Fig. 8. Region of the spectrum from sample A1-324-1 with 235U = 10.1%. Count time was 15 min. Steel thickness was 16 mm. This spectrum is easily analyzed, as the 143-and 163-keV peaks are easily visible. 50000 Counts 40000 30000 The photopeaks are more intense relative to the 1.94% sample because of the higher enrichment and physically larger sample. 20000 10000 0 130 140 150 160 170 keV 180 190 200 Summary We have documented the performance of FRAM on extensive sets of data from various uranium items including the first successful measurements reported using a planar HPGe detector in the 120–1024 keV range. These measurements were undertaken to show the feasibility of IAEA use of existing planar detectors and 4096 channel MCAs for measurements on UF6 cylinders with wall thickness between 13 and 16 mm. We have demonstrated that FRAM can successfully analyze planar HPGe detector (25 mm diam x 16 mm thick) data taken through as much as 16 mm steel in 4096 channels at a gain of 0.25 keV/channel. The measurements were demonstrated for 235U enrichments ranging from 0.3% to 10% and counting times as low as 15 min. Measurement repeatability at the 1 RSD level for 235U ranged from a low of about 6.5% (one hour count time, 13 mm steel absorber, and 10% enrichment) to a high of approximately 35% (15 minute count time, 16 mm steel absorber, and 0.3% enrichment). The latter measurement likely bounds the range of conditions under which planar detector measurements can be successful. In comparison we have shown coaxial detector (approx. 25–30% efficiency) data taken in 8192 channels on some of the same samples. Coaxial detector measurements show significantly improved repeatability when compared with planar measurements. With coaxial detectors, measurements through 16-mm steel in 10–20 min show repeatability at the 1 RSD level in the range of 20% (0.3% enriched) to 8% (2–4% enriched). The coaxial detector repeatability is about a factor of two better than that from the planar detector. We conclude that although coaxial detectors are preferred, one can successfully make UF6 cylinder measurements with 25-mm diam by 16-mm thick planar HPGe detectors in count times as low as 15 min with enrichments as low as 0.3% 235U. We suggest that the extremes of this range, depleted uranium with 15-min count times, need further field testing on actual UF6 cylinders. 8 LA-UR-02-2530 References [1] R. Gunnink, W. D. Ruhter, P. Miller, J. Goerten, M. Swinhoe, H. Wagner, V. Verplancke, M. Bickel, and S. Abousahl, “MGAU: A New Analysis Code for Measuring U-235 Enrichments in Arbitrary Samples,” Proceedings of the IAEA Symposium on International Safeguards, Vienna, Austria, (1994). [2] Thomas E. Sampson, George W. Nelson, Thomas A. Kelley, “FRAM: A Versatile Code for Analyzing the Isotopic Composition of Plutonium from Gamma-Ray Pulse Height Spectra,” Los Alamos National Laboratory report LA-11720-MS (December 1989). [3] Thomas A. Kelley, Thomas E. Sampson, Dorothea DeLapp, “PC/FRAM: Algorithms for the GammaRay Spectrometry Measurement of Plutonium Isotopic Composition,” Proceedings of the Fifth International Conference, Facility Operations-Safeguards Interface, Jackson Hole, Wyoming, September 24-29, 1995. [4] T. E. Sampson, T. A. Kelley, T. L. Cremers, T. R. Konkel, and R. J. Friar, “PC/FRAM: New Capabilities for the Gamma-Ray Spectrometry Measurement of Plutonium Isotopic Composition,” Proceedings of the Fifth International Conference, Facility Operations-Safeguards Interface, Jackson Hole, Wyoming, September 24-29, 1995. [5] Duc T. Vo, Thomas E. Sampson, “Uranium Isotopic Analysis with the FRAM Isotopic Analysis Code,” Los Alamos National Laboratory report LA-13580-MS (May 1999). [6] Thomas E. Sampson, “Gamma-Ray Isotopic Analysis Development at Los Alamos,” Los Alamos National Laboratory report LA-13667-MS (November 1999). [7] Thomas E. Sampson, George W. Nelson, and Thomas A. Kelley, “FRAM: A New, Versatile Gamma-Ray Spectrometry Code for Measuring the Isotopic Composition of Plutonium,” Nucl. Mater. Manage. XIX, 402 (1990). [8] R. Gunnink, “MGA: A Gamma-Ray Spectrum Analysis Code for Determining Plutonium Isotopic Abundances,” Lawrence Livermore National Laboratory report UCRL-LR-103220 (1990). [9] T. E. Sampson, G. P. D. Verrecchia, M. T. Swinhoe, P. Schwalbach, J. Gustafsson, A. M. Anderson, J. Myatt, B. Metcalfe, “Test and Evaluation of the FRAM Isotopic Analysis Code for EURATOM Applications,” Nucl. Mater. Manage. XXVIII, Proceedings CD-ROM (1999). [10] T. E. Sampson, T. A. Kelley, and D. T. Vo, “Improvements in the PC/FRAM Isotopic Analysis Software,” Proceedings of the Sixth International Conference, Facility Operations-Safeguards Interface, Jackson Hole, Wyoming, September 20-24, 1999, p. 206. [11] J. L. Parker, M. Brooks, “Accurate, Wide-Range Uranium Enrichment Measurements by Gamma-Ray Spectroscopy, I. Characterization of Enrichment Standards, II. Demonstration of Enrichment Measurement Capability,” Los Alamos National Laboratory report LA-11227-MS (September 1988). [12] Certified Nuclear Reference Material, Certificate of Analysis, EC Certified Nuclear Reference Material 171, 235U isotope Abundance Certified Reference Material (U3O8) for Gamma-Spectrometry, Central Bureau for Nuclear Measurements, CBNM, Geel, Belgium (June 1985). 9
© Copyright 2026 Paperzz