Datascope 3000 Multiparameter Monitor Specifications: Display

Datascope 3000 Multiparameter Monitor Specifications:
Display: Screen size: 10.5 cm x 14 cm. Channels: 3. Sweep speed: 25 mm/se. Display time: 4
seconds waveform. Message advisory: alphanumeric LCD for messages, alarms, and pulse level.
LED: 7 segments LEDs for SaO2 and NIBP measurements.
ECG Leads: I, II, III, aVR, aVL, aVF, V. Gain: 0.5, 1.0, 2.0, and 3.0 cm/mV. Frequency response:
0.5-25 Hz with automatic ESU filter on, 0.5-40 Hz screen and delayed output @ 25mm/sec. 0.5100 Hz at diagnostic output. CMRR: 100 dB min, at 50/60 Hz with 5K imbalance and 0 dc offset.
80 dB min, maximum output of 1 mVp-p over 60 second period with a 51K ohm lead imbalance
and ±300mVdc offset. Leakage current: Source: less than 10 µA from any patient connection to
ground. Sink: less than 10 µA from any patient connection to ground. Defibrillator protection: fully
protected against 500J discharge. Recovery time: 2 sec. Maximum.
Heart Rate Range: 30-250 – pediatric/adult, 30-250 bpm – neonatal. Accuracy: ±3 bpm or ±3%.
Source: ECG, pressure or pulse with auto selection, SaO2, and NIBP. Pacer rejection: rejects
spikes and displays pacer representation on screen.
Alarm Limits Alarms and limits are available for heart rate, systolic and diastolic pressure, for
IBP1 and IBP2, saturated O2, and NIBP.
Pressure Channels: two. Scales and ranges: 40 mm/cm scale: 20 to 300 mmHg range; 20
mm/cm scale: -10 to 150 mmHg; 10 mm/cm scale: -5 to 75 mmHg range; 5 mm/cm scale: -2.5 to
37.5 mmHg. Accuracy: 2 mmHg or 2%, whichever is greater. Frequency response: 15 Hz, +2/-0
Hz (+0/-3 db) for 1 cm of deflection. Isolation: 500Vac minimum from transducer circuit to chassis
ground. Auto zero: balance range of ±120 mmHg. Transducer requirements: 50 uV/V/cm Hg.
Calibration adjustment range: ±15%.
Temperature Range: 15-40°C. Accuracy: ±0.2°C. Number of channels: 2.
NIBP Technique: measures and displays systolic, mean and diastolic pressures using the
oscillometric technique. Cuff deflation: adaptive linear bleed. Adaptively adjusts bleed rate for
volume of cuff in use. Measurement cycle: from initiation of cuff inflation to display of readings
typically requires 30 seconds depending on BP, HR, and motion artifact.
O2 Saturation Range: 0-100%. Accuracy: 70-100 <±2%.
Recorder Recorder: dual trace; diagnostic and delayed ECG, pulse, pressure, CO2, SaO2, trend
graphics, and list trend. Annotates: time, HR, ECG lead and size, invasive and non-invasive
pressures, temperatures, CO2, O2 and respiration rate, SaO2, N2O, and agent. Speeds: 25
mm/sec for all traces except CO2. Choice of 1 mm/sec or 12.5mm/sec for CO2 print speed.
Mechanical Size: 10.5” H x 12” W x 13” L; 26.7 cm x 30.5 cm x 33 cm. Weight: 32.5 lbs fully
configured. Power requirements: 100/120V, 0.8 amp, 50/60 Hz; 220/240V, 0.4 amp, 50/60 Hz.
Battery Type: rechargeable, sealed lead-acid, 24V. Operating time: 30 minutes or longer, varies
with number of features.