Clinical Science (1997) 93, 3 17-324 (Printed in Great Britain) 317 Reproducibility of ultrasonographic measurements of different carotid and femoral artery segments in healthy subjects and in patients with increased intima-media thickness T. J. SMILDE, H. WOLLERSHEIM, H. VAN LANGEN* and A. F. H. STALENHOEF Departments of Medicine, Division of General Internal Medicine, University Hospital Nijmegen, PO Box 9 I0 I, 6500 HE3 Nijmegen, Nijmegen, The Netherlands, and *Clinical Vascular Laboratory, University Hospital Nijmegen, PO Box 9/01, 6500 H B Nijmegen, The Netherlands (Received 13 January/l9 June 1997; accepted 30 June 1997) 1. The reproducibility of measurements of the arterial wall thickness in both the carotid and femoral artery was investigated by means of high-resolution B-mode ultrasonography. For this purpose, subjects with normal and increased intima-media thickness were selected. Images were stored on a n optical disk and were analysed with a semi-automatic software program by two readers. Individuals were scanned twice by two independent observers. 2. Measurements were performed of the far and near wall of the common carotid artery and bulbous in 30 healthy subjects and 19 patients known to have a n increased intima-media thickness. Far-wall measurements were made of the internal carotid artery on both sides and common femoral artery on the right side only. 3. In healthy subjects the mean within-observer coefficient of variation was 1.8% and 3.0% for the far wall in the common carotid artery on the right side and left side, respectively. For the near wall the mean coefficient of variation of the common carotid artery was 2.8% on the right and 3.4% on the left side. The mean coefficient of variation was less than 4% for both far and near wall in the bulbous and far wall in the internal carotid artery. Even in patients with increased intima-media thickness the mean coefficient of variation of each segement was less than 4.5%. In the control subjects the betweenobserver coefficient of variation of the common carotid artery was 2.8% and 5.1% for the far wall on the right and left side, respectively, and 3.4% and 4.2% for the near wall on the right and left side. In healthy subjects a mean difference of 0.002mm within observers was found in the right far-wall common carotid artery, with limits of agreement of -0.048 to 0.052 mm. The coefficient of repeatability was 0.050mm. For patients with increased intimamedia thickness the mean difference in this segment was -0.006 mm (-0.094 to 0.082) with a coefficient of repeatability of 0.0i38mm. For the near wall in the common carotid artery and far and near wall in the bulbous and internal carotid artery the mean differences were larger, but were all below 0.1 mm. The differences and limits of agreements increased between observers. In patients the between-observer mean difference of the far wall of the common carotid artery was -0.055 mm (-0.255 to 0.145). For the common femoral artery of normal control subjects the within- and between-observer mean differences were 0.005 mm (-0.119 to 0.129) and 0.015 mm (-0.081 to O.lll), respectively. 4. I n conclusion, the reproducibility of intimamedia thickness measurements in the common carotid artery is reliable, even in patients with increased artery wall thickness. Also in other segments prone to atherosclerosis, such as the bulbous, internal carotid artery and common femoral artery, a good reproducibility was found. To obtain good reproducibility it is highly recommended to use the same ultrasonographer to scan patients in follow-up studies. INTRODUCTION Since the identification of lumen-intima and media-adventitia echoes [l], high-resolution B-mode ultrasonography has been used for noninvasive quantitative measurements of intima-media thickness (IMT) which is associated with the presence of atherosclerotic disease elsewhere [l-111. The assessment of IMT is an important tool in intervention trials [12-171. Ultrasonogrpahy has several advantages over contrast anteriography: it can repetitively be applied to asymptomatic subjects, it is relatively cheap and safe and it measures both wall thickness and lumen diameter. The quality of ultrasonographic assessment of atherosclerotic disease is highly dependent on the Key words: atherosclerosis,carotid artery, femoral artery, intima-mediathickness, reproducibility,ultmonography. Abbreviations: BUL,bulbous; CCA, common carotid artery; CFA, common femoral artery; CV,coefficient of variation; ICA, internal carotid artery; IMT intima-mediathickness; RC. coefficient of repeatability. Correspondence: Dr T. 1. Smilde. 318 T.J. Smilde et al. instrumentation and the observer. Since ultrasonographic scanning cannot be automated, the observer is an important source of measurement variability. Most studies involving reproducibility and variability report on far-wall measurements of the common carotid artery only (CCA) and are mostly performed in healthy subjects with normal IMT [2, 18, 191. Although the femoral artery is a preferred site for atherosclerosis, only few studies report on femoral artery IMT [20-221, while reproducibility figures are lacking. The purpose of the present study was to invesigate the within- and between-observer variability and reproducibility of measurements of the mean IMT at different sites: the far and near wall of the CCA, bulbous (BUL) and far wall of the internal carotid artery (ICA) on both sides and far wall measurements of the right common femoral artery (CFA) in subjects with normal and increased IMT. METHODS This cross-sectional study was performed in 30 healthy subjects (50% male), without cardiovascular risk factors except for smoking (n = 5 ) with an IMT < 1.1mm, and in 18 male smokers known to have intima-media thickening of 21.1 mm in at least one segment. The latter were selected from a study of various risk factors and vessel wall changes. All subjects were investigated twice with an interval between 5 and 90 days. The first time observer 1 conducted the scanning, the second visit scannings were made by two observers, blind for each other. In four control subjects scannings were performed by observer 1 on five consecutive days. The study was approved by the Ethics Committee of our hospital. Informed consent was obtained from each individual. Ultrasound scanning protocol The ultrasound examinations were performed using a Biosound Phase 2 real-time scanner equipped with a 10 MHz transducer. Two black and white monitors displayed the B-mode ultrasound images with spectrum analysis of the Doppler signals. Images were grabbed by a computer, stored on a hard disk and analysed with a semi-automatic software program (Eurequa; TSA company, Meudon, France) [18]. The scannings were performed with the subject in a comfortable supine position, the head rotated approximately 45 degrees away from the side being scanned. The scanning sites involved were: the far and near wall of the distal 1.0 cm of the straight part of the CCA, the far and near wall of the carotid bifurcation, beginning at the tip of the flow divider and extending 0.8-1.0 cm proximal, and the far wall of the proximal 1.0cm of the ICA. Measurements were performed on the right and left side. Three angles of interrogation were used: anterolateral, lateral and posterolateral. In each individual the most optimal scanning position (i.e. the head position and scanning angle which images the clearest and thickest projection of interfaces) was noted on a worksheet. The sonographer had the responsibility of differentiating the ICA from external carotid artery. This was accomplished by using several criteria. Often the ICA is the furthest from the skin surface when the V-shape of the flow divider is seen. Other criteria are the larger luminal diameter of the ICA and the dilatation in the ICA on the lateral or posterolateral view. Apart from these characteristics Doppler analysis was used to avoid misinterpretation. Finally, scanning of the far wall of the right CFA was performed 1 cm proximal to the descent of the deep femoral artery. A fixed angle of insonation, anteroposterior, was used. The scan converter enabled freezing of the images during scanning. Callipers were placed on the anatomic references and on the edges of the far and near wall. Subsequently, the digitized frozen images with the clearest and thickest projection were stored on disk. The worksheet with data on head position and scanning angle was used for the second scanning and passed on to the second observer. Ultrasound analysis and report The images stored on disk were read by two independent readers. Each segment was analysed separately. The reader selected the best measurable portion of the image. Three measurements were made in a preselected segment with a length of 0.5 cm. The measurements were performed automatically from significant changes in density on a section perpendicular to the vessel wall from the lumen towards subadventitial structures. The procedure was repeated over 0.5 cm adjacent to the first portion and the mean thickness over 1 cm was calculated. When it was not possible to measure the IMT over the whole length of the selected segments, for example in the bifurcation or when a plaque was present, a smaller sample size was taken with a minimum of 0.2cm. In neither case did the readers have access to the IMT data of previous examinations. Measurements of the 12 individual segments were noted. Missing data were scored. Statistical analysis Coefficients of variation (CV) were calculated as the proportion of the SD of the mean. For CVs describing the between-observer variability, SD was computed over the first measurement of observer 1 and the measurement of observer 2. The CV was estimated for each individual segment and of combinations of far and near wall per segment on the right and left side and of the total mean combined Reproducibility of arterial wall thickness score of the 10 measurements in the carotid artery (far and near wall of the CCA and BUL and far wall of the ICA on left and right side). In addition, mean differences between the first and second scanning of observer 1 and the first scanning of observer 1 and observer 2 were determined. By using the method of limits of agreement as described by Bland and A t - 319 man [23], data are plotted in Figures 1 and 2, showing differences against the mean for each subject. In addition, the coefficient of repeatability (RC) was calculated for all segments as the SD of the estimated difference between two measurements, assuming the mean difference to be zero. The 95% confidence interval of the expected difference is cal- 0.15 g .- 0,1 b a 8 2 0.05 g z o L -0.05 B d -0,15 -O'l -0.2 4 I 032 0 44 0.6 0,s 1 1,2 1,6 1,4 Mean IMTof the far wall of the CCA in mm Fig. 1. Difference between the first and second scanning of observer I of the IMT of the far wall of the CCA plotted against their mean. In this Figure the data of all subjects are included. The mean difference of 0.002 mm with the limits of agreement (-0.048 to 0.052 mm) for healthy subjects is also indicated. 0.2 , b Mean IMT of the far wall of the CCA in mm Fig. 2. Difference between observer I and observer two of the IMT of the far wall of the CCA plotted against their mean. In this Figure the data of all subjects are included. The mean difference of -0.004 mm with limits of agreements (-0.082-0.074) for healthy subjects is also indicated. T. J. Smilde et al. 320 results are listed of the measurements of the IMT of the far wall of the right CCA, the BUL and CFA in the 30 healthy control subjects. In Table 2 the CV within and between observers in healthy control subjects and patients with increased IMT is given. Although the CV remained low, the number of missing data increased moving distally from CCA to ICA (Table 3). Both observers visualized the near and far wall of the CCA in all sessions. Of the far and near walls in the bulb obtained in the first session, only 3% could not be found the second time by the same observer. This increased to 18% when compared with observer 2, meaning that 6% of the far and 12% of the near walls obtained by observer 1 could not be visualized by observer 2 and vice versa. Table 4 gives the absolute differences in mm (SD) within and between observers, as measured in the different segments of the carotid artery and the CFA. In the control subjects the within-observer RC ranged from 0.050mm for the far wall of the right CCA to 0.070 mm for the far wall in the right CCA (Table 5). The reproducibility of the combined culated as 1.96 RC (definition adopted by the British Standards Institution) [23], meaning that repeated measurements are expected to differ by more than the confidence interval with a probability of only 5%. In the four subjects measured on five consecutive days we calculated the SD of the serial measurements for that subject against their mean and then calculated the CV. Between-reader reproducibility was also expressed as CV and mean difference. RESULTS The mean age (SD) of the 30 control subjects was 45 (12) years. Fifteen subjects were male; two of them smoked while three were past smokers. Mean total serum cholesterol was 5.6 (1.3) mmol/l. All participants with known thickened intima-media were male smokers. Their mean age was 55 (8) years and their mean total serum cholesterol concentration was 6.2 (1.5) mmol/l. In Table 1 the Table 1. Detailed overview of measurement of the mean IMT of the far wall of the right CCA, BUL and CFA in 30 healthy subjects. -, Only measured by observer I. Abbreviations: Obs I, observer I; Obs 2, observer 2. CCA Far wall Obs I Subject no. I 2 3 4 5 6 7 8 9 10 II I2 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 BUL Far wall Obs 2 Obs 2 Obs I Age (Years) First time Second time First time First time Second time 32 32 58 33 28 32 38 69 44 66 38 31 26 58 38 33 36 64 42 45 68 38 49 49 50 45 50 41 63 44 0.43 0.6 I 0.44 0.55 0.83 0.61 0.54 0.66 0.87 - 0.5 I 0.8 I 0.72 0.96 0.70 0.53 0.98 0.95 0.80 0.6 I 0.53 0.65 0.90 I.oo 0.64 0.98 0.84 0.67 0.6 I I.02 0.62 0.77 0.77 I.05 0.83 0.58 0.83 0.70 0.63 0.66 0.73 0.59 0.87 0.85 0.8I 0.73 1.01 0.69 I .03 0.80 0.64 0.6 I I.oo 0.6 I 0.77 0.78 I.05 0.84 0.57 0.84 0.70 0.58 0.67 0.7 I 0.60 0.86 0.85 0.80 0.75 0.59 0.60 0.69 0.84 0.99 0.71 I.05 0.84 0.63 0.6 I 0.99 0.63 0.74 0.76 I.05 0.84 0.59 0.8 I 0.70 0.6 I 0.64 0.69 0.62 0.96 0.8 I 0.84 0.75 CFA Far wall First time 0.73 0.95 0.68 0.97 0.97 Obs I First time Second time 0.41 0.43 I.04 0.54 0.44 0.36 0.95 I.06 0.45 0.42 I.04 0.53 0.43 0.38 0.90 I.09 0.80 0.89 0.71 0.86 0.85 0.84 0.98 - 0.77 0.89 0.93 0.76 0.93 0.89 0.93 0.85 0.86 I.02 0.90 0.76 0.85 0.87 0.85 0.49 0.85 0.76 I .oo 0.99 0.78 0.86 0.89 0.8I 0.44 0.83 0.74 I.03 0.90 0.80 0.81 0.92 I .08 0.99 0.80 0.78 0.64 0.75 0.78 0.77 0.69 0.76 0.77 0.79 0.58 0.76 I.08 0.96 0.82 I.04 0.97 0.82 0.89 0.97 0.82 0.86 0.83 0.84 0.71 0.77 0.9 I 0.69 0.84 0.93 0.73 0.76 0.99 0.77 0.64 I.06 0.88 0.7 I 0.83 0.7 I 0.80 0.74 0.83 Obs 2 0.87 0.75 0.87 0.83 0.94 0.89 0.75 0.93 0.88 First time 0.55 0.45 0.40 0.88 1.12 0.81 0.88 0.72 0.85 0.82 0.80 0.9 I 0.88 0.92 0.88 0.85 0.78 0.44 Reproducibilityof arterial wall thickness 32 I Table 2. CVs (SD) of different segments in the carotid and femoral artery. CVs are given between and within observers for subjects with normal IMT and for subjects with increased IMT. Subjects with normal IMT CCA far wall right CCA far wall left CCA near wall right CCA near wall left BUL far wall right BUL far wall left BUL near wall right BUL near wall left ICA far wall right ICA far wall left CFA right side CCA far and near wall on right and left side Combination of 10 carotid measurements (left and right) Within observer Between observer Within observer Betweenobserver 1.8 (I .8) 3.0 (4.0) 2.8 (3.7) 3.4 (3.8) 2.1 (1.8) 3.2 (4.4) 1.5 (1.3) 3.1 (4.7) 3.8 (5.1) 3.9 (4.4) 2.3 (2.1) 1.5 (1.6) 2.8 (2.4) 5.1 (6.2) 3.4 (3.7) 4.2 (4.4) 5.1 (5.7) 3.4 (2.6) 5.0 (3.3) 4.8 (6.0) 5.1 (4.2) 6.4 (4.9) 2.5 (2.5) 2.0 (I.9) 2.3 (2.4) 2.7 (4.6) 2.8 (6.0) 2.8 (3.6) 4.1 (5.2) 3.1 (1.7) 2.7 (3.7) 2.0 (I .3) 1.7 (1.8) 3.4 (2.4) 2.9 (I.9) 2.2 (3.7) 3.5 (5.0) 4.7 (5.9) 8.2 (9.9) 5.9 (8.1) 7.5 (9.2) 2.5 (0.5) 6.2 (4.3) 2.4 (1.1) 3.2 (2.6) 6.4 (I 0.5) 4.5 (10.5) 4.4 (4. I) 2.4 (2.8) 3.2 (3.5) I.9 (2.0) 4.8 (5.0) Table 3. Percentage of missing data in scannings made by observer I (n = 96) and observer 2 (n = 45) Missingdata (%) Observer I, first time Observer I, second time 0 0 31 60 42 0 0 31 58 43 62 38 54 Observer 2, first time ~ CCA far and near wall right CCA far and near wall left BUL far wall right BUL far wall left BUL near wall right BUL near wall left ICA far wall right ICA far wall left 64 38 53 Subjects with increased IMT 0 I 43 56 42 65 38 56 measurements of the far and near wall on both sides of the CCA were better than the individual measurements, except for the between-observer variability in patients with increased wall thickness (Table 2). The CV of the total carotid artery, combining all measurements, was also low (betweenobserver CV, 3.2 and 4.8% for healthy subjects and patients respectively, see Table 2). The CV of the measurement of the far wall of the CFA was <5% although the between-observer reproducibility was clearly less (CV 4.4) than the within-observer reproducibility (CV 2.2) in subjects with increased IMT. Fig. 1 shows the limits of agreement of the IMT of the CCA measured by observer 1 in all subjects. In this Figure the differences between the first and second session are plotted against their mean. Fig. 2 shows the difference between measurements made Table 4. Absolute differences in mm (SD)within and between-observers, as measured in the different segments of the carotid artery and the far wall of the CFA in subjects with and without intima-media thickening Subjects with normal IMT CCA far wall right CCA far wall left CCA near wall right CCA near wall left BUL far wall right BUL far wall left BUL near wall right BUL near wall left ICA far wall right ICA far wall left CFA right side CCA far and near wall on right and left side Combination of 10 carotid measurements(left and right) Subjects with increased IMT Within observer Between observer Within observer Between observer 0.017 (0.017) 0.026 (0.035) 0.031 (0.042) 0.026 (0.035) 0.025 (0.020) 0.041 (0.044) 0.023 (0.017) 0.040 (0.058) 0.032 (0.037) 0.030 (0.030) 0.03 I (0.03 I) 0.016 (0.018) 0.027 (0.022) 0.045 (0.055) 0.041 (0.042) 0.045 (0.055) 0.064 (0.072) 0.042 (0.034) 0.070 (0.042) 0.049 (0.070) 0.055 (0.046) 0.054 (0.043) 0.036 (0.036) 0.026 (0.025) 0.033 (0.029) 0.058 (0.I 15) 0.059 (0.152) 0.045 (0.045) 0.079 (0.099) 0.058 (0.030) 0.055 (0.081) 0.036 (0.025) 0.030 (0.050) 0.045 (0.030) 0.048 (0.043) 0.021 (0.019) 0.063 (0.101) 0.08 I (0.105) 0.140 (0.190) 0.100(0.163) 0.100 (0.140) 0.050 (0.008) 0. I26 (0.096) 0.040 (0.016) 0.062 (0.083) 0. I33 (0.298) 0.089 (0.218) 0.077 (0.076) 0.022 (0.026) 0.03 I (0.029) 0.036 (0.060) 0.072 (0.071) T. J.Smilde et al. 322 Table 5. Mean differences in mm with (limits of agreements) within and between observers measured in the different segments of the carotid and CFA in 30 healthy subjects and 18 subjects with increased IMT Subjects with increased IMT Subjects with normal IMT Within observer CCA far wall right CCA far wall left CCA near wall right CCA near wall left BUL far wall right BUL far wall left BUL near wall right BUL near wall left ICA far wall right ICA far wall left CFA right side CCA far and near wall on right and left side Combination of 10 carotid measurements (left and right) 0.002 (-0.048 to 0.052) -0.0012 (0.102 to 0.078) -0.013 (-0.I I3 to 0.087) -0,009 (-0.133 toO.115) 0.002 (-0.072 to 0.068) -0.01 I (-0.133 to 0.1 I I) 0.009 (-0.043 to 0.061) -0.036 (-0. I58 to 0.086) -0.014 (-0.1 14 to 0.086) 0.0 I2 (-0.082 to 0. I 18) 0.005 (-0.119to0.129) -0.006 (-0.052 to 0.040) 0.002 (-0.066 to 0.070) by observer 1 and 2 plotted against their mean. Table 5 shows the difference between measurements with their limits of agreement. For those control subjects measured on five consecutive days the CV did not change over time; it ranged from 2.1% in the CCA to 4.2% in the BUL. All observations were read by two different readers. The inter-reader variability was less than 2% in all segments, with a mean difference of lower than 0.015 mm. Variability did not differ between images of normal control subjects and patients with increased IMT. DISCUSSION With the ultrasound protocol used in this study it was possible to visualize and measure IMT in different segments of the carotid artery, including in patients with known increased IMT. The anatomical location of a biological structure is always defined by a leading edge of an echo, and the thickness of a structure as the distance between the leading edges of two different echoes. It has therefore been argued that, in spite of the similarities of near- and far-wall images, IMT can only be measured accurately in the far-wall position, because only the farwall IMT is defined by leading edges [24, 251. Because of the echogenicity of the adventitia in the Between observer Within observer Between observer -0.004 -0.006 (-0.094 0.030 (-0.222 - 0.044 (-0.358 -0.006 (-0.128 -0.017 (-0.273 t00.I 16) -0.055 (-0.255 0.022 (-0.242 -0.028 (-0.502 -0.05 (-0.422 to 0.239) 0.001 (-0.55 I to 0.553) (-0.082 to 0.074) - 0.003 (-0.151 t00.145) - 0.007 (-0.117to0.103) -0.1 15 (-0.259 to 0.029) 0.033 (-0.151 to 0.217) -0.010 (-0.I20t00.100) 0.012 (-0. I56 to 0.180) -0.032 (-0.202 to 0.138) -0.018 (-0.163 t00.146) 0.036 (-0.086 to 0.158) 0.0 I5 (-0.081 toO.l I I) -0.006 (-0.06 to 0.048) -0.014 (-0.094 to 0.066) to 0.082) to 0.282) to 0.270) - 0.00 I (-0.141 t00.139) -0.038 (-0.222 to 0.146) -0.021 (-0.101 to0.059) -0.023 (-0.131 t00.106) -0.005 (-0.121 to0.III) -0.002 (-0.146 to 0.142) 0.009 (-0.131 t00.149) -0.003 (0.09 I to 0.085) to 0.145) to 0.286) to 0.446) to 0.322) 0.024 ' (-0.062 to 0.1 10) 0.022 (-0.316 to 0.360) -0.004 (-0.098 to 0.090) -0.051 (-0.233 to0.131) 0.095 (-0.533 to 0.723) -0.048 (-0.5I2t00.416) -0.036 (-0.228 to 0.156) -0.019 (-0.249 to0.21 I) near wall, the reflections from the intima-media may be blurred. However, by scanning in such a way that the jugular vein is placed adjacent to the carotid artery and with the development of new software it was possible to measure near walls reliably, which may provide additional information. In this study we measured both far and near wall with the aid of a semi-automatic analysing program in a standardized way. Using the clearest and thickest projection of the IMT leads to bias towards thicker values. This scanning procedure was performed in order to see whether measurements of increased IMT are reproducible. The mean differences between and within sonographers were small, although from the limits of agreement we concluded that, particularly in patients with increased IMT, the between-observer variation is larger in an individual patient. How far apart measurements may be without causing difficulties is a matter of judgement and should be defined in advance. The limits of agreement are only estimates of the values which apply to the whole population and one should realize that a different sample would give different limits. The reproducibility of the combined IMT of the far and near wall of the CCA was better than the individual measurements. This is probably due to averaging more measurements. The CV was somewhat larger on the left than on the right side. This might be due to the different anatomical situation or Reproducibility of arterial wall thickness by the right-handiness of the observers, which has also been mentioned by others [19]. The CV and mean differences did not change between consecutive days, indicating that the scanning method is reliable and repeatable. The between-reader variation was small. In this protocol the sonographer determined the best image, which was stored on disk. Consequently, the readers’ influence was less prominent since he was confronted with a preselected image. The storage of images on optical disks instead of taping them on video has the advantage of full resolution preservation. Measurement error of IMT tended to increase with increasing levels of IMT, as reported earlier [19]. This study shows that also in segments other than the CCA, IMT can be measured with great reliability, even in patients with known increased IMT. This is important, since intervention trials will most often be performed in patients in which an increased IMT is anticipated. Moving from the CCA to the ICA, the amount of missing data increased mostly because the site of the bifurcation was located high up in the neck. Other studies encountered the same problem [5]. It is important to realize that reliable, repeatable measurements in the BUL and ICA are not possible in all patients. In calculating numbers of patients required for clinical trials with IMT one should correct for these expected missing data. Another important finding in this study was that the same segments could be visualized twice by the same observer in almost all cases. Since the within-observer variability is smaller than the between-observer variability and two observers are not always able to visualize the same segments, it is highly recommended that patients should be scanned during follow-up by the same observer. In several studies the reproducibility of the IMT has only been determined in the CCA in healthy subjects [2, 18, 191. The CV in the study of Salonen et al. [2] was approximately 5% and the mean difference in IMT between observers was 0.087mm. In the multi-centre Cardiovascular Health Study the between-sonographers difference was 0.20 [7]. In both studies the differences between observers was based on two and one measurements, respectively, in contrast to six measurement sites in our study. Furthermore, more observers were taken into account, while in our study two experienced observers performed all measurements. The use of an optical disk instead of video tapes makes the readers’ influence less and this may decrease the variability. The latter may explain the better results in our study compared with those of Riley et al. [26]. They report on the reproducibility of combined carotid artery wall thickness. The mean absolute difference within and between observers was 0.06 and 0.12mm, compared with 0.04 and 0.07mm in our study in patients with increased IMT. Another explanation for this difference may be the fact that the former authors included the near wall of the 323 ICA. The reproducibility data of the Rotterdam study are very similar to our data: a mean difference of 0.013mm on the right and 0.05 on the left side between sonographers at different visits in the CCA ~91. The variability of measurements is determined by the sonographer, reader, instrumentation and by differences between subjects. Especially in patients with increased IMT of plaques, measurements become more imprecise, because of tortuous arteries, eccentric plaque and irregularities. Since these factors cannot be influenced, it is important to reduce the effect of other factors determining variability. This means regular check-ups of the instrumentation, a repeated training programme of the sonographers and readers with quality control assessment and a follow-up schedule that allows scanning by the same sonographer. High-resolution B-mode ultrasonographic measurement of carotid arterial IMT is a suitable pseudo end-point in clinical trials [13-17, 27, 281, although the relation between IMT reduction and clinical events as coronary heart disease and stroke need to be established. The non-invasive imaging of the arterial wall can be performed repeatedly in symptomatic and asymptomatic patients, carried negligible risk and quantifies early atherosclerosis and atherosclerotic changes due to risk factor modification. Atherosclerotic changes in the carotid artery are not equally distributed. Increased IMT occurs more often in the carotid BUL than in the CCA. As shown by this study measurements at other sites than the CCA can be done with good reproducibility even in patients with thickened IMT. In conclusion, measurements of the IMT in different segments of the carotid artery and of the femoral artery were highly reproducible. Although the measurement error tended to increase with increased mean IMT, the reproducibility remained good. The reproducibility at the site of the BUL and ICA were good whenever obtained, although it was not always possible to visualize these segments. With the possibility of direct storing images on optical disk, the resolution is preserved. The betweenreader variability is very low when using a (semi) automatic reading system. In follow-up studies it is highly recommended that the same ultrasonographer be used to scan patients. REFERENCES I. Pignoli P, Tremoli E, Poli A, Oreste P, Paoletti R lntimal plus medial thickness of the arterial wall: a direct measurement with ultrasound imaging. Circulation 1986; 7 4 1399-406. 2. Salonen R Haapanen A Solonen IT. Measurement of intima-media thickness of common carotid arteries with high-resolution B-mode ultrasonography: interand intra-observer Variability. Ultrasound Med Biol 1991; 17: 225-30. 3. Penson J, Stavenow L. Wikstrand J, lsraelsson B, FormgrenJ, Berglund G. Noninvasive quantification of atherosclerotic lesions. Reproducibilityof ultrasonographic measurement of arterial wall thickness and plaque size. Arteriosclerosis Thromb 1992; I2 26 1-6. 324 T. J. Smilde et al. 4. Okin PM, Roman MJ, Devereux RB, Kligfield P. Association of carotid atherosclerosis with electrocardiographicmyocardial ischemia and left ventricular hypertrophy. Hypertension 1996; 28: 3-7. 5. Salonen JT, Salonen R Ultrasonographically assessed carotid morphology and the risk of coronary heart disease. Arteriosclerosis Thromb I99 I; I I : 1245-9. 6. Heiss G, Sharrett AR, Barnes R, Chambless LE, Szklo M, Alzola C. Carotid atherosclerosis measured by B-mode ultrasound in populations; associations with cardiovascular risk factors in the ARK study. Am J Epidemiol 1991; 134 250-6. 7. OLeary DH, Polak IF, Wolfson SKI, et al. Use of sonography to evaluate carotid atherosclerosis in the elderly. The Cardiovascular Health Study. CHS Colloborative Research Group. Stroke I99 I; 2 2 I 155-63. 8. Wendelhag I,Wiklund 0, Wikstrand J. Arterial wall thickness in familial measurement of intima-media thickness in the common carotid artery. Arteriosclerosis Thromb 1992; I 2 70-7. 9. Bond MG, Barnes RW, Riley WA, et al. High-resolution B-mode ultrasound scanning methods in the atherosclerosis risk in communities study (ARIC). Am Soc Neuroimag 1991: 68: 73. 10. Poli A, Tremoli E, Colombo 4 Sirtori M, Pignoli P. Paoletti R Uitrasonographic measurement of the common carotid artery wall thickness in hypercholesterolemicpatients. A new method for the quantitation and follow-up of preclinical atherosclerosis in living human subjects. Atherosclerosis 1988; 70: 253-61. I I. Bonithon Kopp C, Scarabin PY, Taquet A, Touboul PJ, Malmejac A, Guize L. Risk factors for early carotid atherosclerosis in middle-aged French women. Arteriosclerosis Thromb I99 I ; I I: 966-72. 12. Furberg CD, Borhani NO, Byington RP, Gibbons ME, Sowers JR Calcium antagonist and atherosclerosis. Multicenter Isradipine/DiureticAtherosclerosis Study. Am J Hypertens 1993; 6 245-95. 13. Hodis HN, Mack WJ, LaBree L, et al. Reduction in carotid arterial wall thickness using lovastatin and dietary therapy. A randomized, controlled clinical trial. Ann Intern Med 1996; I 2 4 548-56. 14. Kroon AA, Asten WNJC, Stalenhoef AFH. Effects of apheresis of low-density lipoprotein on peripheral vascular disease in hypercholesterolemicpatients with coronary artery disease. Ann Intern Med 1996; 125: 945-54. 15. Blankenhorn DH, Selzer RH, Crawford DW, et al. Beneficial effects of colestipol-niacin therapy on the common carotid artery. Two- and four-year reduction of intima-mediathickness measured by ultrasound. Circulation 1993; 88: 20-8. 16. Crouse JR, Byington RP, Bond MG. Pravastatin, lipids, and atherosclerosis in the carotid arteries (PLAC-11). Am J Cardiol 1995; 75: 455-9. 17. de Groot E, JukemaJW, van Boven AJ, et al. Effect of pravastatin on progression and regression of coronary atherosclerosis and vessel wall changes in carotid and femoral arteries: a report from the Regression Growth Evaluation Statin Study. Am J Cardiol 1995; 7 6 4OC-6C. 18. Touboul PJ. Prati P. Scarabin PY, Adrai V, Thibout E, Ducimetiere P. Use of monitoringsoftware to improve the measurement of carotid wall thickness by B-mode imaging. J Hypertens 1992; 10 (Suppl.): 537-41. 19. Bots ML, Mulder PG, Hofman A, van Es GA, Grobbee DE. Reproducibilityof carotid vessel wall thickness measurements. The Rotterdam Study. J Clin Epidemiol 1994; 47: 921-30. 20. Gariepy J, Simon 4 Massonneau M, Linhart A, Segond P, Levenson J, Groupe PCVMETRk Echographic assessment of carotid and femoral structure in men with essential hypertension. Am J Hypertens 1996; 9: 126-36. 21. Beach KW, Isaac CA. Phillips DJ, StrandnessJrE. An ultrasonic measurement of superficial femoral artery wall thickness. Ultrasound Med Biol 1989; 15: 723-8. 22. JoensuuT, Salonen R Winblad I, Korpela H, Slaonen JT. Determinants of femoral and carotid artery atherosclerosis. J Intern Med 1994; 236 79-84. 23. Bland JM. Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; k 307-10. 24. Wendelhag I, Gustavsson T, Suurkula M. Berglund G, Wikdstrand J. Ultrasound measurement of wall thickness in the carotid artery: fundamental principles and description of a computerized analysing system. Clin Physiol 1991; I I: 565-77. 25. Wong M, EdelsteinJ,Wollman J, Bond MG. Ultrasonic-pathologicalcomparison of the human arterial wall. Verification of intima-media thickness. Arteriosclerosis Thromb 1993; I 3 482-6. 26. Riley WA, Barnes RW, Applegate WB, et al. Reproducibilityof noninvasive ultrasonic measurement of carotid atherosclerosis. The asymptomatic carotid artery plaque study. Stroke 1992; 23: 1062-8. 27. Blankenhorn DH, Hodis HN. George Lyman Duff Memorial Lecture. Arterial imaging and atherosclerosis reversal. Arteriosclerosis Thromb 1994; 1 4 177-92. 28. Furberg CD, Byington RP, Borhani NA. Multicenter isradipine diuretic atherosclerosis study (MIDAS). Design features. The Midas Research Group. Am J Med 1989; 8 6 37-9.
© Copyright 2026 Paperzz