AJCP / Original Article Performance of Various Laboratory Assays in the Measurement of Dabigatran in Patients Receiving Therapeutic Doses A Prospective Study Based on Peak and Trough Plasma Levels Robert Gosselin, CLS,1 Emily Hawes, PharmD,2 Stephan Moll, MD,3 and Dorothy Adcock, MD4 From the 1Department of Medical Pathology and Laboratory Medicine, UC Davis Medical Center, Sacramento, CA; 2Department of Pharmacy, University of North Carolina Hospitals and Clinics, Chapel Hill; 3Department of Medicine, University of North Carolina School of Medicine, Chapel Hill; and 4Esoterix, Englewood, CO. Key Words: Dabigatran; Direct thrombin inhibitor; Mass spectrophotometry; Ecarin clotting time; Ecarin chromogenic assay; Dilute thrombin time DOI: 10.1309/AJCPRNUMI4PVSJ7Q ABSTRACT Objectives: To study dabigatran etexilate, a new oral anticoagulant that functions as a direct thrombin inhibitor. Methods: This study evaluates four methods, one of which is performed in three different laboratories, and compares results against dabigatran levels measured by BoehringerIngelheim (Ingelheim, Germany) using mass spectrometry. Results: Although routine monitoring is not required, measurement of plasma concentrations may be necessary in certain clinical situations. Routine coagulation assays such as the prothrombin time, activated partial thromboplastin time, and thrombin time do not reliably determine levels of dabigatran anticoagulation. Alternative assays, when calibrated with a dabigatran standard, such as the modified dilute thrombin time, ecarin clotting time, and ecarin chromogenic assay, may be appropriate, although a comparison of these methods using samples from patients taking dabigatran has not been performed. Conclusions: Although results using all methods in this study demonstrate adequate correlation, measured dabigatran levels varied in a statistically significant manner, even when the same method was used by different laboratories. The clinical significance of this variation in dabigatran concentrations is uncertain. 262 262 Am J Clin Pathol 2014;141:262-267 DOI: 10.1309/AJCPRNUMI4PVSJ7Q Dabigatran extilate (Pradaxa; Boehringer-Ingelheim, Ingelheim, Germany) is a novel oral anticoagulant that belongs to the direct thrombin inhibitor class of anticoagulants. Dabigatran’s mechanism of anticoagulant action is through inhibition of free, fibrin-bound, and clot-bound thrombin. It is approved by the US Food and Drug Administration for stroke reduction in individuals with atrial fibrillation not caused by a heart valve problem. Unlike warfarin, dabigatran has predictable pharmacodynamic and pharmacokinetic profiles, has fewer drug-drug interactions compared with warfarin, is not affected by diet, requires no bridging therapy, and is distinct from other direct thrombin inhibitors (eg, bivalirudin) because it does not require routine monitoring. Patients receiving dabigatran therapy may, however, require assessment of plasma drug concentration due to unforeseen circumstances, such as suspicion of overdose, the occurrence of bleeding, or in situations of increasing renal insufficiency, emergency surgery, or trauma. Published studies have illuminated the limitations of routine coagulation tests, such as prothrombin time (PT), activated partial thromboplastin time (APTT), and thrombin time (TT), to determine the relative concentration of plasma dabigatran.1-3 Alternative assays, when calibrated with a dabigatran standard, such as the dilute TT (dTT), ecarin clotting time (ECT), ecarin chromogenic assay (ECA), or the prothrombinase-induced clotting time, have been suggested methods to quantify the amount of dabigatran in plasma.1-5 It is not known, however, whether these methods provide comparable results. In this study, we assessed four different means for measuring dabigatran levels compared with a standard method in patients administered dabigatran for thromboprophylaxis. © American Society for Clinical Pathology AJCP / Original Article Materials and Methods The study was approved by the University of North Carolina Institutional Review Board for human subject experimentation. Detailed study methods have been described elsewhere, and the data presented here are a further analysis of data from that study.3 In brief, a total of 35 participants, taking 150 mg dabigatran twice daily at steady state, provided written consent and were enrolled in the study. Blood was collected by standard phlebotomy techniques into 3.2% sodium citrate tubes for trough (just before dose) and peak (2.5 hours after dose) dabigatran levels. Once collected, the blood was processed within 2 hours, centrifuged to obtain platelet-poor plasma, aliquoted into cryovials, and stored at –70°C until analysis. Each stored sample was labeled with a study ID number with designation of “trough” or “peak” level. Frozen samples were distributed to testing laboratories via express delivery in Styrofoam containers with ample dry ice to ensure maintenance of a frozen state. Frozen samples were sent to a central laboratory for predicate dabigatran measurements using mass spectrometry (courtesy of Boehringer-Ingelheim, Ingelheim, Germany [designated BI-MS]) and to five expert laboratories based in the United States. Reagents, calibration materials, instrumentation used, and trough/peak dabigatran results were provided by each participating site. Of the five participating laboratories reporting dabigatran levels, three used the ECA method,6 one applied a commercial dTT method,1 and one employed liquid chromatography/tandem mass spectrometry (LC-MS/MS). One site that reported the ECA also provided results using the classic ECT method.7 Laboratory sites A, B, and C used a commercially prepared ECA (ECA-T Kit; Diagnostica Stago, Asnieres, France), and laboratory D used commercially prepared dTT (HEMOCLOT; Hyphen-BioMed, NeuvilleSur-Oise, France). All ECA and dTT methods used the same dabigatran calibrator source and control material (Dabigatran Calibrator Plasma and Dabigatran Control Plasma; HyphenBioMed) but not necessarily the same lot of materials. Laboratory E used an internally validated LC/MS-MS measurement for dabigatran levels. Laboratory A also performed a classic noncalibrated ECT, and for this reason, results are reported as clotting times and do not quantify dabigatran concentration. All values were reported in nanograms/milliliters of drug, except for ECT, which was reported in seconds. Laboratory A also calibrated the ECT using aforementioned dabigatran calibrators to determine whether the ECT quantification of dabigatran would be equivalent to other methods. Regression analysis was used to determine the correlation between the predicate method (BI-MS) and other methods, with an R2 greater than 0.95 indicating acceptable agreement. Biases between the reporting laboratory and predicate method were assessed using the Bland-Altman calculation (bias = laboratory – [laboratory + BI-MS]/2). The paired t test P value of less than .05 was used to determine whether statistically significant differences exist between methods. Comparison evaluation between laboratory and predicate dabigatran measurement methods was determined for (1) all measured dabigatran (trough and peak) values, (2) expected dabigatran trough levels based on the RE-LY trial (95th percentile range, ~40-240 ng/mL),8 and (3) changes in drug concentrations between trough and peak measures. Results Thirty-five pairs (trough and peak levels) of samples were drawn for each enrolled participant. The laboratory testing matrix included three ECAs on two different analyzers, one dTT assay, one ECT, and one LC/MS-MS method ❚Table 1❚. ❚Table 1❚ Testing Matrix Used for Determining Dabigatran Levels Laboratory SiteMethod Reagent Calibrator Instrument A Chromogenic ECA-T Kit (Diagnostica Stago, Dabigatran Calibrator Plasma STA Compact (Diagnostica Stago) ecarin assay Asnieres, France) (Hyphen-BioMed, Neuville- Sur-Oise, France) A Ecarin clotting time Ecarin Reagent None STart 4 Hemostasis Analyzer (Diagnostica Stago) (Diagnostica Stago) B Chromogenic ECA-T Kit (Diagnostica Stago) Dabigatran Calibrator Plasma BCS XP System (Siemens Healthcare ecarin assay (Hyphen-BioMed) Diagnostics, Marburg, Germany) C Chromogenic ECA-T Kit (Diagnostica Stago) Dabigatran Calibrator Plasma STA-R Evolution (Diagnostica Stago) ecarin assay (Hyphen-BioMed) D Dilute thrombin time HEMOCLOT (Hyphen-BioMed) Dabigatran Calibrator Plasma STA Compact (Diagnostica Stago) (Hyphen-BioMed) E Liquid chromatography, In-house developed mass spectrometry © American Society for Clinical Pathology 263 Am J Clin Pathol 2014;141:262-267 DOI: 10.1309/AJCPRNUMI4PVSJ7Q 263 263 Gosselin et al / Performance of Assays in Measuring Dabigatran For the chromogenic ecarin assays, Pearson correlation was 0.948, 0.984, and 0.979 for laboratories A, B, and C, respectively ❚Figure 1A❚, but all sites varied significantly (P < .001) from the predicate BI-MS method. Bland-Altman bias plots demonstrate that most of these result differences are directional, indicating that overall, the ECA method underestimates drug level compared with the BI-MS method ❚Figure 1B❚. There was no statistical improvement when comparing the range of expected trough dabigatran levels ❚Table 2❚. When comparing the change in drug concentration between trough and peak drug concentration, only laboratory B demonstrated no significant differences (P = .10) (when compared with changes between drug concentrations as measured by BI-MS). When comparing among laboratories that report the ECA, there were no significant differences between values reported between laboratories A and B (P = .10). 500 450 400 350 300 250 200 150 100 50 0 B Lab A Lab B Lab C Lab D Lab E Lab A Lab B Lab C 60 40 ECA Dabigatran Difference (ng/mL) Dabigatran (ng/mL) A There was borderline correlation (R2 = 0.937) and statistically significant differences (P < .001) between the commercial dTT kit (laboratory D) and BI-MS–measured dabigatran levels (Figure 1A). The bias plots indicate that values determined using the dTT method are primarily lower than BI-MS measurements, with increasing bias as drug concentration increases ❚Figure 1C❚. There was no observed improvement in correlation or drug concentration differences between this dTT and predicate BI-MS when evaluating those samples in the expected trough range of results (~40-240 ng/mL) or the observed changes between trough and peak drug levels (Table 2). Results obtained with the dTT method demonstrated no significant differences (P = .54) between this method and the chromogenic ecarin method for only one of the three laboratories that performed the chromogenic method. 20 0 –20 0 50 100 150 200 250 300 350 400 450 500 550 –40 –60 –80 –100 –120 0 50 100 150 200 250 300 350 400 450 500 Mean (Lab + BI-MS)/2 (ng/mL) BI-MS Dabigatran (ng/mL) 100 75 50 25 0 –25 0 –50 –75 –100 –125 –150 –175 D 50 100 150 200 250 300 350 400 450 500 Mean (Lab D + BI-MS)/2 (ng/mL) Lab E LC/MS-MS Dabigatran Difference (ng/mL) Lab D dTT Dabigatran Difference (ng/mL) C 100 90 80 70 60 50 40 30 20 10 0 –10 0 50 100 150 200 250 300 350 400 450 500 BI-MS Dabigatran (ng/mL) ❚Figure 1❚ A, Regression analysis between laboratories A, B, and C performing the ecarin chromogenic assay (ECA); laboratory D performing dilute thrombin times (dTT); and laboratory E performing liquid chromatography/tandem mass spectrometry (LC-MS/MS) for determining dabigatran levels and predicate measurements using Boehringer-Ingelheim (Ingelheim, Germany) mass spectrometry (BI-MS). Laboratory A = 0.85x + 11.8; R2 = 0.948. Laboratory B = 0.92x – 2.58; R2 = 0.984. Laboratory C = 0.78x – 4.16; R2 = 0.979. Laboratory D = 0.81x – 5.56; R2 = 0.937. Laboratory E = 1.06x + 8.28; R2 = 0.977. B-D, An overlay of Bland-Altman–style plots demonstrating the bias between (B) each respective laboratory site’s ECA-determined dabigatran concentration and predicate mass spectrometry quantitation of dabigatran concentration, (C) dTT-determined dabigatran concentration and predicate mass spectrometry quantitation of dabigatran concentration, and (D) LC/MS-MS–determined dabigatran concentration and predicate mass spectrometry quantitation of dabigatran concentration. 264 264 Am J Clin Pathol 2014;141:262-267 DOI: 10.1309/AJCPRNUMI4PVSJ7Q © American Society for Clinical Pathology AJCP / Original Article ❚Table 2❚ Statistical Analysis for Each Laboratory Testing Sitea Laboratory Characteristic A BC D E All samples (n = 70) Pearson correlation 0.948 0.984 0.979 0.937 0.977 P value (paired t test) <.001 <.001 <.001 <.001 <.001 Samples 40-240 ng/mL (n = 57) Pearson correlation 0.855 0.962 0.956 0.849 0.937 P value (paired t test) <.001 <.001 <.001 <.001 <.001 Difference in drug concentration between trough and peak levels Pearson correlation 0.933 0.958 0.966 0.836 0.937 P value (paired t test) <.001 .10 <.001 .03 .57 Testing laboratories A, B, and C used the ecarin chromogenic assay, laboratory D used dilute thrombin times, and laboratory E used liquid chromatography/tandem mass spectrometry for measuring dabigatran levels. Each site was compared with dabigatran levels obtained from the parent drug company using Boehringer-Ingelheim (Ingelheim, Germany) mass spectrometry (BI-MS), with a P value of less than .05 indicating significant differences between methods. A subset analysis of samples between 40 and 240 ng/ mL dabigatran by BI-MS would reflect the 90th percentile of trough levels in patients receiving a 150-mg twice-daily regimen. The difference between trough and peak drug concentration represents the absolute change in drug value between these time frames. There was acceptable correlation (R2 = 0.977) for laboratory E using LC/MS-MS compared with the predicate BI-MS values (Figure 1A). Bias plots indicate that the LC/MS-MS method of laboratory E overestimates drug concentration (P < .001) compared with BI-MS ❚Figure 1D❚. There was no statistical improvement between methods when evaluating drug levels from 40 to 240 ng/mL, but no statistically significant difference between methods (P = .57) was observed when assessing the change between trough and peak drug concentrations (Table 2). There was acceptable correlation (R2 = 0.970) between BI-MS and the ECT, but a relatively flat slope ❚Figure 2❚ was evident. As expected, there were significant differences between BI-MS, which measures drug concentration, and ECT, which merely reports time in seconds. These differences did not improve when evaluating expected trough levels or differences between trough and peak samples. There was good correlation (R2 = 0.962) but significant differences (P < .001) between dabigatran levels obtained from calibrated ECT and BI-MS measurements. The ECT-dabigatran results correlated with other laboratory methods, with R2 values ranging from 0.874 (laboratory A) to 0.934 (laboratory E). There were no significant differences between ECT-dabigatran levels and those levels reported by laboratory C using ECA (P = .43) and laboratory D using LC-MS/MS (P = .23). measure. While the use of a dTT for measuring dabigatran has been advocated by some, these data suggest that the chromogenic ecarin method is also acceptable when both assay methods are calibrated using a dabigatran standard.1,3,9,10 Significant differences in dabigatran concentrations were demonstrated between some laboratories in this study, despite the fact that they used the same methods and calibration source. It is unclear why there were significant differences between laboratory sites that used ECA methods, even though the same calibrator source was used. The only two significant Dabigatran (ng/mL) or ECT (s) a 350 325 300 275 250 225 200 175 150 125 100 75 50 25 0 –25 0 –50 ECT-Dabigatran ECT 50 100 150 200 250 300 350 400 450 500 BI-MS Dabigatran (ng/mL) Discussion This multisite study used different methods of determining dabigatran concentration in the plasma of patients treated with 150 mg dabigatran twice daily. All methods demonstrate a linear relationship with drug concentrations determined by BI-MS, which in this study is considered the predicate © American Society for Clinical Pathology ❚Figure 2❚ Regression analysis of a single site performing both classic ecarin clotting time (ECT); y = 0.311x + 28.0; R2 = 0.970) and dabigatran levels obtained from calibrated ECT; y = 0.920x – 25.48; R2 = 0.962) compared with dabigatran levels measured by Boehringer-Ingelheim (Ingelheim, Germany) mass spectrometry (BI-MS). 265 Am J Clin Pathol 2014;141:262-267 DOI: 10.1309/AJCPRNUMI4PVSJ7Q 265 265 Gosselin et al / Performance of Assays in Measuring Dabigatran differences between assays at the ECA testing sites were (1) use of a 0-ng/mL calibrator, not included in the commercial dabigatran calibrator kit, and (2) instrumentation. Although the 0-ng/mL calibrator may enhance the sensitivity to lower drug concentrations (<30 ng/mL), this would not affect results greater than 30 ng/mL. These data suggest that subtle differences in processing, reading, calibration, accuracy, and result extrapolation between instrumentation, possibly even within the same instrument manufacturer, may account for differences seen in ECA measurements. This study is distinct from other published findings in several ways. Douxfils and colleagues6 used pooled normal plasma spiked with different concentrations of dabigatran, but the calibration of dTT and ECA curves is unclear. Another study measuring dabigatran levels used local laboratory-prepared calibrators to compare levels obtained using either dTT or ECT.11 In the study reported by Samama and colleagues,12 the observed dabigatran concentration in patients receiving dabigatran after major orthopedic surgery was similar between dTT and ECA when using the same instrumentation and calibrator at a single testing center. Current recommendations from the Scientific and Standardization Committee (SSC) of the International Society for Thrombosis and Haemostasis indicate that either the APTT or a dabigatran-calibrated dTT can be used to estimate drug level.10 This SSC report also notes that ECT is sensitive and linear to 300 ng/mL of dabigatran, but it takes no position on chromogenic ecarin methods to determine dabigatran levels. We also demonstrated that the classic ECT, when calibrated with dabigatran calibrators, correlates well with other dabigatran measurement methods and therefore may also be considered a laboratory-developed (and equivalent) method for determining dabigatran levels. To our knowledge, this is the first multisite comparison of methods for measuring dabigatran concentration in patients receiving the drug for thromboprophylaxis. Recently, a European-based proficiency assessment survey program (ECAT; Leiden, the Netherlands) incorporated plasma samples containing novel anticoagulants, including dabigatran, as part of the program. For the 2012 ECAT proficiency dabigatran sample survey, 26 participating laboratories determined dabigatran levels under the category “anti-IIa assay” (which is a chromogenic prothrombin inactivation assay similar to the anti-Xa used to measure heparin), one laboratory used the ECT, and 36 used the dTT. No participating laboratories used mass spectrometry. The coefficients of variation associated with anti-IIa and dTT methods were 8.3% to 14.3% and 11.9% to 18.9%, respectively, with the acceptable result range of 235 to 410 ng/mL and 85 to 200 ng/mL for samples containing approximately 300 ng/mL and 100 ng/mL, respectively.13 This proficiency survey data would suggest that variability for dabigatran measurements is not uncommonly observed in coagulation testing. 266 266 Am J Clin Pathol 2014;141:262-267 DOI: 10.1309/AJCPRNUMI4PVSJ7Q There are some limitations to our findings. First, the chromogenic ecarin method was used by most laboratories. This is in contrast to ECAT findings that suggest that dTT and anti-IIa methods are more abundant. However, the ECAT proficiency assessment program does not offer ECA as a method option, and perhaps those laboratories using ECA are recording under the chromogenic method (anti-IIa) in lieu of a clotbased method (dTT). In our study, only a single site used the dTT method for reporting dabigatran levels, and none used a chromogenic prothrombin inactivation assay. The calibrators used by each ECA testing site are optimized for dTT methods and not ECA per the manufacturer’s instructions. However, the calibrator values reported for these calibrators are assessed using BI-MS and thus could be used in different methods, assuming that the lyophilization processing of this material does not interfere with the testing method. Indeed, our data would suggest that these dTT-optimized commercial calibrators are suitable for ECA methods and measuring dabigatran concentrations in plasma. In conclusion, we find good correlation between existing laboratory methods (ECT, ECA, and dTT) and predicate BI-MS measurements for dabigatran, but significant differences in measured dabigatran concentrations exist between laboratories and methods. It is unclear whether these differences have any clinical significance or impact, since this drug is not routinely monitored For those clinical situations that may require virtual absence of drug, employing a more sensitive measure for assessing dabigatran (eg, TT testing) should be considered. Address reprint requests to Dr Adcock: 8490 Upland Dr, Suite 100, Englewood, CO 80112-7116. Acknowledgments: We thank Cheryl Jeanneret, University of North Carolina at Chapel Hill, McAllister Heart Institute, for patient enrollment; Herbert Whinna, University of North Carolina at Chapel Hill, Department of Pathology and Laboratory Medicine; Mike Taylor, Esoterix Coagulation, and Anne Winkler, Emory University, Department of Pathology, for dabigatran results provided from their respective institution; Joann van Ryn, Boehringer-Ingelheim, for providing predicate dabigatran measurements; and Abigail Cook, Loyola University Health System, for assistance with study design. References 1. van Ryn J, Stangier J, Haertter S, et al. Dabigatran etexilate—a novel, reversible, oral direct thrombin inhibitor: interpretation of coagulation assays and reversal of anticoagulant activity. Thromb Haemost. 2010;103:11161127. 2. Dager WE, Gosselin RC, Kitchen S, et al. Dabigatran effects on the international normalized ratio, activated partial thromboplastin time, thrombin time, and fibrinogen: a multicenter, in vitro study. Ann Pharmacother. 2012;46:16271636. © American Society for Clinical Pathology AJCP / Original Article 3. Hawes EM, Deal AM, Funk-Adcock D, et al. Performance of coagulation tests in patients on therapeutic doses of dabigatran: a cross-sectional pharmacodynamic study based on peak and trough plasma levels. J Thromb Haemost. 2013;11:1493-1502. 4. Harenberg J, Giese C, Marx S, et al. Determination of dabigatran in human plasma samples. Semin Thromb Hemost. 2012;38:16-22. 5. Stangier J, Fuering M. Using the HEMOCLOT direct thrombin inhibitor assay to determine the plasma concentrations of dabigatran. Blood Coagul Fibrinolysis. 2012;23:138-143. 6. Douxfils J, Mullier F, Robert S, et al. Impact of dabigatran on a large panel of routine or specific coagulation assays. Thromb Haemost. 2012;107:985-997. 7. Nowak G. The ecarin clotting time, a universal method to quantify direct thrombin inhibitors. Pathophysiol Haemost Thromb. 2003;33:173-183. 8. Connolly SJ, Ezekowitz MD, Yusuf S, et al; and the RE-LY Steering Committee and Investigators. Dabigatran versus warfarin in patients with atrial fibrillation. N Engl J Med. 2009;361:1139-1151. 9. Siegel DM, Crowther MA. Acute management of bleeding in patients on novel oral anticoagulants. Eur Heart J. 2013;34:489-498. © American Society for Clinical Pathology 10. Baglin T, Hillarp A, Tripodi A, et al. Measuring oral direct inhibitors (ODIs) of thrombin and factor Xa: a recommendation from the Subcommittee on Control of Anticoagulation of the Scientific and Standardisation Committee of the International Society on Thrombosis and Haemostasis [published online January 24, 2013]. J Thromb Haemost. doi:10.1111/jth.12149. 11. Freyburger G, Macouillard G, Labrouche S, et al. Coagulation parameters in patients receiving dabigatran etexilate or rivaroxaban: two observational studies in patients undergoing hip or total knee replacement. Thromb Res. 2011;127:457465. 12. Samama MM, Guinet C, Le Flem L, et al. Measurement of dabigatran and rivaroxaban in primary prevention of venous thromboembolism in 106 patients, who have undergone major orthopedic surgery: an observational study. J Thromb Thrombolysis. 2013;35:140-146. 13. ECAT Foundation. External Quality Control for Assays and Tests With a Focus on Thrombosis and Haemostasis. Leiden, the Netherlands: ECAT Foundation; 2012. Survey report 2012-2. 267 Am J Clin Pathol 2014;141:262-267 DOI: 10.1309/AJCPRNUMI4PVSJ7Q 267 267
© Copyright 2026 Paperzz