MSP80D
www.muesentechnik.de
Monocrystalline silicon smart differential pressure transmitter
MSP80D
High stability, low static pressure
MSP80D smart differential pressure
transmitter uses monocrystalline silicon
sensor chip and world's original design
of monocrystalline silicon suspension that
made of advanced MEMS technology. So it
achieves the world's leading overpressure
performance and also ensures excellent
signal stability. Built-in manometry capsule
and signal processing module that are
imported from Germany achieve the
perfect combination of static pressure and
temperature compensation, which provides
high measurement accuracy and stability
under a wide range of static pressure and
temperature.
®
MSP80D smart differential pressure (flow)
transmitter can measure differential
pressure accurately and convert it into DC
4 ~ 20mA output signal. This transmitter
can be operated through three buttons
locally or through universal communicator
or configuration software remotely. Without
affecting the output signal of DC 4 ~ 20mA,
it can display and configurate at the same
time.
Detailed Information
www.muesentechnik.de
Differential pressure transmitter
Application
the flow of gas, vapor and liquid
measurement with device (volume or
mass flow)
Measurement of level, volume and mass
Differential pressure monitoring
Through a number of international
certifications, widely used
Powerful instruments for process pressure
High stability
Low static pressure
Long term stability
Worldwide support and service
Benefits
High accuracy
Differential pressure (flow) transmitter
can measure high accurately in the 0.5 ~
500kPa measuring range.
The accuracy of standard calibration
range: ±0.05%
Positive and negative zero shift, without
adjusting linearity
Excellent static pressure speciality within
a small range
overpressure of 1kPa nominal range
chip:
1.5MPa (1500 times)
overpressure of 4kPa nominal range
chip:
2.5MPa (625 times)
Without central diaphragm, simplify the
structure, improve the accuracy and the
stability
Excellent suitability of environment
The intelligent static pressure and
temperature compensation protect
transmitter from the effects of
temperature, static pressure and
overpressure, so that the consolidated
measurement error will be kept to the
minimum.
Flexible range of compression
M1,M2 Range ratio: 20:1
M3 Range ratio: 40:1
M4,M5,M6 Range ratio: 100:1
Excellent operability & convenient to use
▪ Five-digit LCD with backlight
▪ View of units (Pa, kPa, MPa, bar, mbar,
%, psi, mmH2o)
▪ Quickly adjusted through built-in three
buttons
▪ A variety of corrosion-resistant materials
▪ Comprehensive self-diagnosis function
1
MSP80D
www.muesentechnik.de
Specifications
Functional specifications
Performance specifications
▪ Testing fluid : liquid, gas, vapor
▪ Measuring range :
▪ Accuracy :
range/
scope
M1
M2
M3
M4
M5
M6
kPa
sensor cell
inH2o
mbar
M1,M2,M3, M4,M5,M6
mmH2o
range
0.5~10
2~40
5~100
50~1000
scope
-10~10
-40~40
-100~100
-1000~1000
range
1~20
4~80
10~200
100~2000
scope
-20~20
-80~80
-200~200
-2000~2000
range
1~40
4~160
10~400
100~4000
scope
-40~40
-160~160
-400~400
-4000~4000
range
1~100
4~400
10~1000
100~10000
scope
-100~100
-400~400
-1000~1000
-10000~10000
range
2~200
8~800
20~2000
200~20000
scope
-200~200
-800~800
-2000~2000
-20000~20000
range
5~500
20~2000
50~5000
500~50000
scope
-500~500
-2000~2000
-5000~5000
-50000~50000
Note:
wetted material
working pressure
M1
M2
M3
M4
M5
M6
01, 02, 03, 04, 05
01, 02, 03, 04, 05
01, 02, 03, 04, 05
01, 02, 03, 04, 05
01, 02, 03, 04, 05
01, 02, 03, 04, 05
25MPa
25MPa
25MPa
25MPa
25MPa
25MPa
code of sensor cell & wetted material refers to the code table
DC4~20mA+HART
▪ Output signal :
▪ Allowable load resistance : 0~600Ω (DC 24V)
Note: when communicates with the
handheld communicator, you need a
standard (250Ω) load resistance
▪ Power supply :
general situation
DC10.5~45V
intrinsically safe explosion-proof DC10.5~26V
▪ Communication line
condition :
Length of line: max 2km
(When 0.75~1.25mm2 control instrument cable
exceeds 1 km, you should use twisted pair
cable)
load resistance :
250~600Ω
(DC 24V, include cable resistance)
load capacity :
under 0.55mF
load inductance :
under 3.3mH
Spacing above power line : above 15cm
(Please avoid parallel wiring)
upper limit 20.8mA
▪ Saturation current :
lower limit 3.8mA
upper limit 22.8mA
▪ Alarm current :
lower limit 3.6mA
(Mode can be set)
The zero & full span point can be operated
▪ Adjustment function :
through three buttons from the top of
the housing or through configuration software
remotely.
▪ Environmental temperature : -40 ~ +85℃
(With filling fluorine oil : -10~+60℃ )
-40~+90℃
▪ Storage temperature :
DIN40040GPC
▪ Weatherability :
▪ EMC applicable standard : EN1326-1: 2006
▪ TD>15:1
=±(0.0015×TD+0.05)%
*TD- range ratio
▪ square-root output precision:
output
precision
same as reference precision
≥ 50%
50%~drop point reference precision×50
square-root output (%)
▪ Effect of environmental
temperature :
The total amount of effect/28℃ (50°F)
sensor cell
effect
M1,M2,M3 ±[0.08%range+0.015%upper limit of range]
M4,M5,M6 ±[0.08%range+0.03%upper limit of range]
▪ Effect of static pressure :
▪ Static pressure :
sensor cell
reference precision
=±0.05%
▪ TD1:1...TD15 :1
sensor cell
M1,M2,M3
M4,M5,M6
effect
±[0.05%range]/10MPa
±[0.05%range]/10MPa
▪ Effect of overpressure : ±0.05%upper limit of range/10MPa
▪ Stability :
sensor cell
M1,M2,M3
M4,M5,M6
effect
±0.05%upper limit of range/year
±0.05%upper limit of range/year
▪ Effect of Power supply : ±0.005%/1V
▪ Effect of Mounting
position :
The changes of mounting position in the
direction parallel to the diaphragm will not
cause zero - drift effects. If the changes
between the mounting position and the
diaphragm is more than 90°, which can be
corrected through zeroing corrected within
the range of 0.4KPa.
90ms
▪ Response time :
The time constant can be adjusted from 0 to
▪ Damping :
99.9 seconds.
can be adjusted from 0 to 160 uA
▪ Filter constant :
▪ Self-stabilization
coefficient :
can be adjusted from 0 to 2%
600
Load resistance
R (Ω)
Communicated
scope HART
250
10.5
16.4
24.7
45
Supply voltage DC (V)
Diagram of supply voltage and external load
2
MSP80D
www.muesentechnik.de
Structure material
▪ wetted material:
code of
material
01
02
03
04
05
Electrical connection diagram
refers to the code table
detecting part
isolated diaphragm
installed plywood
SUS316L
SUS316
SUS316
Hastelloy C-276
SUS316
Tantalum
SUS316
SUS316L Gold plating
Hastelloy C
Hastelloy C-276
▪ Wetted sealing ring : fluororubber, chemigum
▪ Housing of transmitter: low copper aluminum die casting +
polyurethane coating
▪ Cover of transmitter : aluminum casting + polyurethane coating
IP67
▪ Ingress protection :
SUS304
▪ Nameplate:
silicon oil & fluorocarbon oil (optional)
▪ Filling oil :
according to outline
▪ Size :
about 2.2~2.7kg (noumenon)
▪ Weight :
G1/2,1/2-14NPT,M20×1.5
▪ Cable intake :
M3 bolt
▪ External terminal :
▪ Process connection
port :
Rc1/4 or 1/4 -18NPT
U-type mounted on 50mm (2▪ Installation :
inch) pipe, or on the wall
(depend on the model specification)
Dimensional drawing of outline
Shell
Internal grounding wire
Power supply
signal
(4...20)mA
Testing signal
External grounding
wire
3
MSP80D
www.muesentechnik.de
Model and specification code table
Type
Specifications code
MSP80D
· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ···
Output signal H · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Range
( sensor cell)
M1
M2
M3
M4
M5
M6
Wetted material
··················································
··················································
··················································
··················································
··················································
··················································
01
02
03
04
05
99
Filling oil
················································
················································
················································
················································
················································
················································
01 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
02 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
99 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
01 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
02 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
99 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Process connection
Explanation
Monocrystalline silicon differential pressure transmitter
4~20mA, HART digital communication protocol
range:0.5~10kPa
range:1~20kPa
range:1~40kPa
range:1~100kPa
range:2~200kPa
range:5~500kPa
isolation diaphragm : SUS316L
isolation diaphragm : hastelloy C-276
isolation diaphragm : Tantalum
isolation diaphragm : SUS316L gold planting
isolation diaphragm : hastelloy C -276
other requirements
silicon oil
fluorocarbon oil
other requirements
female thread , 1/4” - 18NPT
female thread , 1/2” - 14NPT
other requirements
01 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
02 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
99 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
fluororubber
chemigum
other requirements
Electrical connection
01 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
99 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
M20x1.5
other requirements
Display header
D · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ··
N · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ··
Sealing ring
2-inch pipe v bracket
H ······························
V ······························
O ······························
Additional code of type selection
*Reference type:MSP80D-H-M1-01-01-01-01-01-D-H
overpressure:2MPa
overpressure:3.5MPa
overpressure:3.5MPa
overpressure:7MPa
overpressure:7MPa
overpressure:7MPa
with LCD
nothing
flat bracket
right angle bracket
nothing
Type selection specification
plywood : SUS304
plywood : SUS316
plywood : SUS316
plywood : SUS316
plywood : hastelloy C
4
MSP80D
www.muesentechnik.de
Additional specifications (explosion- proof)
Projeckts
Explanation
NEPSI flame-proof and dust explosion-proof approval :
Applicable standard : GB3836.1-2010 GB3836.2-2010
d Ⅱ CT6 ; DIP A21T T6 A
T6 : maximum permissible surface temperature 85℃ Environmental termperature : -40~60℃
Chinese standard Electrical interface : 1/2NPT female thread , G1/2 female thread , M20 female thread
NEPSI
NEPSI intrinsic safety and dust explosion-proof approval :
Applicable standard : GB3836.1-2010 GB3836.4-2010
ia Ⅱ CT6 ; DIP A20T T6 A
T6 : maximum permissible surface temperature 85℃
Environmental termperature : -40~60℃
FM flame-proof approval
Applicable standard : FM3600 , FM3615 , FM3810 , ANSI/NEMA250
Flame-proof : Ⅰ rating , 1 district , B , C , D group
Flame-proof burning : Ⅱ / Ⅲ rating , 1district , E , F , G group
Hazardous locations : Indoor and outdoor ( NEMA4X )
T6 : environmental termperature : -40~60℃
Electrical interface : 1/2NPT female thread
FM intrinsic safety approval
Certification of Applicable standard : FM3600 , FM3610 , FM3611 , FM3810 , ANSI/NEMA250
factory
Intrinsic safety : Ⅰ rating , 1 district , A , B , C , D group ;
Ⅱ rating,1district , E , F , G group and Ⅲ rating , 1district hazardous locations
federation
Non combustibla : Ⅰ rating , 2 district , A , B , C , D group
Ⅱ rating , 1 district , E , F , G group and Ⅲ rating , 1 district hazardous locations
( FM )
seal : NEMA 4X
Termperature rating : T4
Environmental termperature : -40~60℃
Electrical interface : 1/2NPT female thread
FM flame retardant approval
Ⅰ rating , 2district , A , B , C , D group , Termperature rating T4 , Type 4X
Ⅱ rating , 2 district , F , G group , Termperature rating T4 , Type 4X
Ⅲ rating , 2 district , Termperature rating T4 , Type 4X
European
Community
(EC)
(ATEX)
Code
ND1
NX1
FD1
FX1
FX2
CENELEC (KEMA) flame-proof approval :
Applicable standard : EN60079-0,EN60079-1
Identification : KEMA 02ATEX2148
Ⅱ 2G Ex d Ⅱ C T4 , T5 , T6
Environmental termperature : T5 : -40~80℃ ;T4andT6 : -40~75℃
The highest wetted temperature : T4:120 ℃ T5:100 ℃ T6:85 ℃
Electrical interface : 1/2NPT female thread,M20 female thread
AD1
CENELEC (KEMA) intrinsic safety approval :
Applicable standard : EN50014,EN50020,EN50284 Identification : KEMA 02ATEX1030X
Ⅱ 1G Eex ia Ⅱ C T4;
Environmental termperature : -40~60℃ ;
Electrical interface : 1/2NPT female thread , M20 female thread
AX1
Including KF21,KS2 and N (non-sparking) approval :
Applicable standard : EN 60079-0 : 2009 , E,60079-15 : 2005
Ⅱ 3G Ex nl Ⅱ C T4 ;
Environmental termperature : -30~60℃ ;
Electrical interface: 1/2NPT female thread , PG13.5 female thread , M20 female thread
AX2
CENELEC ATEX Type n approval :
Applicable standard : EN 60079-0 : 2009 , E,60079-15 : 2005
Ⅱ 3G Ex nl Ⅱ C T4 ;
Environmental termperature : -30~60℃
Muesen Technik GmbH
Produktionsadresse:
Wilhelm-Gutbrod-Straße 19
60437 Frankfurt am Main, Germany
Telefon:
Telefax:
Email:
Internet:
+49 (0)69 9050 5554
+49 (0)69 5098 3597
[email protected]
www.muesentechnik.de
AX3
5