- Physik Instrumente

Technical Note
KSch, 2014-01-27
P736T0001, valid for PD73Z2xxW
PD73Z2xxW PInano® Piezo Scanner Systems
Contents
System Description ............................................................................................................ 2
Possible System Components ......................................................................................................2
Available PInano® Piezo Scanner Systems ..................................................................................2
Accessory .....................................................................................................................................3
Symbols and Typographic Conventions .......................................................................... 3
Other Applicable Documents ............................................................................................ 4
Intended Use ....................................................................................................................... 4
Safety Precautions ............................................................................................................. 5
Product Description ........................................................................................................... 6
Product View.................................................................................................................................6
Scope of Delivery ..........................................................................................................................6
Installation........................................................................................................................... 7
General Notes on Installation ........................................................................................................7
Connecting the P­736.Zxx to a Protective Earth Conductor ..........................................................8
Mounting the P-736.Zxx to the M-687.Ux ......................................................................................9
Mounting a Sample Holder to the P-736.Zxx ...............................................................................11
Start-Up and Operation .................................................................................................... 12
General Notes on Start-Up and Operation ..................................................................................12
Starting Up and Operating the P-736.Zxx ...................................................................................13
Discharging the P-736.Zxx ............................................................................................... 14
Maintenance ...................................................................................................................... 14
Cleaning the P-736.Zxx .................................................................................................... 14
Customer Service ............................................................................................................. 14
Technical Data .................................................................................................................. 15
Dimensions ....................................................................................................................... 17
Maximum Ratings ............................................................................................................. 21
Ambient Conditions and Classifications ........................................................................ 21
Old Equipment Disposal .................................................................................................. 22
EC Declaration of Conformity.......................................................................................... 22
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 2 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
System Description
The PInano® piezo scanner systems are designed for inverse microscopy using microscopes by
Nikon and Olympus. Each system consists of a P-736.Zxx PInano® Z piezo scanner and an
E-709.xRG digital piezo controller.
Possible System Components
Standard versions of the P-736.Zxx PInano® Z piezo scanner:
Model
Name
P-736.ZCN
PInano® Z Piezo Scanner for Inverted Nikon Microscopes, 220 µm,
Well Plate Size Aperture, Capacitive Sensors
P-736.ZRN
PInano® Z Piezo Scanner for Inverted Nikon Microscopes, 220 µm,
Well Plate Size Aperture, Piezoresistive Sensors
P-736.ZCO
PInano® Z Piezo Scanner for Inverted Olympus Microscopes, 220 µm,
Well Plate Size Aperture, Capacitive Sensors
P-736.ZRO
PInano® Z Piezo Scanner for Inverted Olympus Microscopes, 220 µm,
Well Plate Size Aperture, Piezoresistive Sensors
Standard versions of the E-709.xRG digital piezo controller:
Model
Name
E-709.CRG
Digital Piezo Controller, 1 Channel, -30 to 130 V, Capacitive Sensor,
Bench-Top
E-709.PRG
Digital Piezo Controller, 1 Channel, -30 to 130 V, Piezoresistive
Sensor, Bench-Top
Available PInano® Piezo Scanner Systems
The following PInano® piezo scanner systems are available as combinations of Z piezo scanner and
digital piezo controller:
System
-
Z Piezo Scanner
Digital Piezo Controller
P-736.ZCN
P-736.ZRN
P-736.ZCO
P-736.ZRO
E-709.CRG
E-709.PRG
PD73Z2CNW
x
-
-
-
x
-
PD73Z2RNW
-
x
-
-
-
x
PD73Z2COW
-
-
x
-
x
-
PD73Z2ROW
-
-
-
x
-
x
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 3 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Accessory
The P-736.Zxx PInano® Z piezo scanners can be mounted to the M-687.UN and M-687.UO XY
microscope stages or any other suitable mounting surface, which is mounted to a corresponding
microscope.
The Z piezo scanners can be mounted, among others, to the following M-687.Ux XY microscope
stages:
P-736.Zxx model
M-687.UN
M-687.UO
P-736.ZCN
x
-
P-736.ZRN
x
-
P-736.ZCO
-
x
P-736.ZRO
-
x
To position slides or petri dishes PI offers a microscope universal holder.
Order Number
Description
M-687.AP1
Microscope Universal Holder for Slides and Petri Dishes for PI Stages
with 160 mm × 110 mm Free Aperture
Other inserts are available on request.
Symbols and Typographic Conventions
The following symbols and typographic conventions are used in this Technical Note:
CAUTION
Dangerous situation
If not avoided, the dangerous situation will result in minor injury.

Actions to take to avoid the situation.
NOTICE
Dangerous situation
If not avoided, the dangerous situation will result in damage to the equipment.

Actions to take to avoid the situation.
INFORMATION
Information for easier handling, tricks, tips, etc.
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 4 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Other Applicable Documents
The devices and software tools which are mentioned in this documentation are described in their own
manuals.
Product
Document
E-709 Digital Piezo Controller
PZ222E User Manual
E-709 Digital Piezo Controller for Capacitive Sensors
E700T0003 Technical Note
M-687.UN XY Microscope Stage for Inverted Nikon Microscopes
MP100E User Manual
M-687.UO XY Microscope Stage for Inverted Olympus Microscopes
MP107E User Manual
Intended Use
The P-736.Zxx PInano® Z piezo scanners and the E-709.xRG digital piezo controllers are laboratory
devices as defined by DIN EN 61010-1. They are intended to be used in interior spaces and in an
environment which is free of dirt, oil and lubricants.
Based on their design and realization, the Z piezo scanners are intended to position in Z direction.
They are designed to position a sample holder with suitable dimensions, as for example the
M-687.AP1 sample holder that can carry a slide or a petri dish.
The Z piezo scanners can carry a maximum load of 0.5 kg. They are intended to be used with
M-687.Ux XY microscope stages or with other suitable microscope stages.
The Z piezo scanners are to be mounted horizontally.
The intended use of the Z piezo scanners is only possible in combination with the E-709.xRG digital
piezo controllers. The E-709.xRG controllers provide the required operating voltages. To ensure
proper performance of the servo-control system, the electronics are able to read out and process the
signals from the capacitive or piezoresistive sensors.
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 5 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Safety Precautions
CAUTION
Dangerous voltage and residual charge on piezo actuators!
The P-736.Zxx is driven by piezo actuators. Temperature changes and compressive
stresses can induce charges in piezo actuators. After being disconnected from the
electronics, piezo actuators can also stay charged for several hours. Touching or
short-circuiting the contacts in the connector of the P-736.Zxx can lead to minor
injuries. In addition, the piezo actuators can be destroyed by an abrupt contraction.

Do not open the P-736.Zxx.

Discharge the piezo actuators before installation:
Connect the piezo actuator to the switched-off PI controller.

Do not pull out the connector from the electronics during operation.
For piezo actuators with sub-D connector:
Touching the contacts in the connector can lead to an electric shock (max.
130 V DC) and minor injuries.

Do not touch the contacts in the connector.

Secure the connector of the piezo actuator with screws against being pulled out
of the controller.
CAUTION
Risk of electric shock if the protective earth conductor is not connected!
If a protective earth conductor is not or not properly connected, dangerous touch
voltages can occur and there is a risk of electric shock. In the case of malfunction or
failure of the system, touching the P-736.Zxx can result in minor injuries.

Connect the P-736.Zxx to a protective earth conductor before start-up.

Do not remove the protective earth conductor during operation.

Use electrically conductive materials (e.g. screws and flat washers) for mounting
the protective earth conductor.

Make sure that the contact resistance is < 0.1 ohm at 25 A at all connection
points relevant for mounting the protective earth conductor.

If the protective earth conductor has to be temporarily removed (e.g. for
modifications), reconnect the P-736.Zxx to the protective earth conductor before
starting it up again.
NOTICE
Unsuitable cables!
Unsuitable cables can damage the electronics.

Only use cables from PI for connecting the P-736.Zxx to the electronics.
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 6 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
INFORMATION
Extended cables can affect the performance of the P-736.Zxx.

Do not use cable extensions. If you need longer cables, contact our customer
service department (p. 14).
Product Description
Product View
Figure 1: P-736.ZCO as example
1
4 x M3 mounting holes for mounting a sample holder as for example the M-687.AP1
2
Movable platform
3
Cable exit
4
3 x counterbore holes for M4 screws for mounting the P-736.Zxx to a M-687.Uxx XY stage
5
Base body
Scope of Delivery
A PInano® piezo scanner system is delivered with the following components.
Order
Number
Component
P-736.Zxx
PInano® Z piezo scanner according to order
E-709.xRG
Digital piezo controller according to order
C-663.PS
Wide-range 24-V power supply, 42 W
3763
Line cord
000036360
USB cable (3 m, USB-A (m)/USB Mini-B (m)) for PC
connection
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 7 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Order
Number
Component
E709B0002
RS-232 adapter HD sub-D 26 (m) to sub-D 9 (m) for PC
connection via RS-232
C-815.34
Null-modem cable for PC connection via RS-232
E-709.04
Adapter cable for analog I/O (1m, HD sub-D 26 / 2 x
BNC)
000036450
M4 screw set for protective earth connection
Documentation, consisting of:
P736T0001
Technical note for PD73Z2xxW PInano® piezo scanner
systems (this document in printed form)
PZ222E
User manual for E-709 digital piezo controllers
E-709.CD
CD with PC software and documentation of the E-709
digital piezo controller
A000T0032
Technical note on updating the PI Software using the PI
Update Finder
A000T0028
Technical note on using the PI Update Finder
Only
PD73Z2CxW: Technical note for E-709.CRG, CHG, CR
E709T0003
Packaging materials
Installation
General Notes on Installation
NOTICE
Appropriate torque required!
Specifications are only achieved if a minimum torque is observed when mounting the
stage to the M-687.Ux or any other suitable microscope stage. A torque too high can
cause damage to the stage.

To ensure that specifications are achieved and to avoid damage to the stage
observe a torque range of 250 Ncm to 270 Ncm when you mount the stage to
the M-687.Ux or any other suitable microscope stage using the 3 M4 screws.
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 8 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
NOTICE
Unallowable forces on movable platform!
Executing a push force >50 N or a pull force >30 N on the movable platform can
damage the stage and render the mechanical design inoperable.

Ensure that only a maximum push force of 50 N or a maximum pull force of 30 N
is executed on the movable platform.
NOTICE
Lubricants, dirt, condensation!
Dirt, oil, lubricants and condensation will render the motor/drive inoperable.

Ensure that the piezo actuators of the P-736.Zxx do not come into contact with
lubricants.

Keep the P-736.Zxx free from dirt and condensation.
Connecting the P­736.Zxx to a Protective Earth Conductor
NOTICE
Ground loops with double connection of protective earth conductor!
In the case of P­736.Zxx models with sub-D connectors, ground loops can occur
when the P­736.Zxx is grounded via its protective earth connection or the mounting
holes as well as by the shield of the connection cable for the electronics.

If a ground loop occurs, contact our customer service department.
INFORMATION

Observe the applicable standards for mounting the protective earth conductor.
The P­736.Zxx features a separate protective earth connection.
Figure 2: Example of a protective earth connection
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 9 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Tools and accessories



Suitable protective earth conductor: Cross-sectional area of the cable
≥0.75 mm2
Suitable screwdriver
M4 CE screws for connecting the protective earth conductor
Connecting the protective earth conductor to the separate protective earth connection
Figure 3: Recommended mounting of the protective earth conductor (profile view)
1
Case of the P-736.Zxx
2
Flat washer
3
Safety washer
4
Screw
5
Cable lug
6
Protective earth conductor
1. If necessary, fasten a suitable cable lug to the protective earth conductor.
2. Fasten the cable lug of the protective earth conductor to the M4 screw on the protective earth
connection of the P­736.Zxx as shown in the profile view.
3. Tighten the M4 screw with at least three rotations and a torque of 1.2 Nm to 1.5 Nm.
Mounting the P-736.Zxx to the M-687.Ux
Prerequisite
 You have read and understood the General Notes on Installation (p. 7).
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 10 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Tools and accessories


3 socket head screws per P-736.Zxx model (included with the P-736.Zxx model)
P-736.Zxx model
Suitable screws
P-736.ZCN
(part of PD73Z2CNW system)
M4x10
P-736.ZRN
(part of PD73Z2RNW system)
M4x10
P-736.ZCO
(part of PD73Z2COW system)
M4x14
P-736.ZRO
(part of PD73Z2ROW system)
M4x14
Suitable screwdriver
Mounting the P-736.ZxN to the M-687.UN
INFORMATION
The upper moving platform of the M­687.UN features an adapter plate which has to
be removed before mounting the P-736.ZRN or P-736.ZCN PInano® Z piezo
scanners on the M­687.UN.
Figure 4: Fixing screws of the adapter plate of the M-687.UN
1. Loosen the three M4 fixing screws of the adapter plate.
2. Remove the three M4 fixing screws of the adapter plate.
3. Remove the adapter plate.
4. Mount the P-736.ZxN on the M­687.UN using three M4x10 screws observing a
torque range of 250 to 270 Ncm.
For mounting, use the threaded holes where the adapter plate was affixed.
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 11 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Mounting the P-736.ZxO to the M-687.UO
Figure 5: M-687.UO: Threaded holes in the upper platform (black arrows = 3 x M4)
1. Choose the mounting position so that the three M4 holes in the upper platform
(Y axis; see arrows in the figure) can be used for affixing the P-736.ZxO.
2. Mount the P-736.ZxO with three screws M4x14 to the threaded holes observing
a torque range of 250 to 270 Ncm; see "Dimensions" in the MP107E user
manual for M-687.UO.
Mounting a Sample Holder to the P-736.Zxx
Prerequisite
 You have read and understood the “General Notes on Installation” (p. 7).
Tools and accessories



Sample holder of suitable dimensions as, for example, M-687.AP1
Suitable screwdriver
4 x M3x4 screws per P-736.Zxx model (included with the M-687.AP1 model)
Mounting a sample holder to the P-736.Zxx
 Only mount the sample holder to the P-736.Zxx using the four mounting holes for M3
screws intended for this purpose (see Figure 6).
 Optional, when an M-687.AP1 sample holder is used: The sample holder can be snapped
in place and held in place by the retaining spring of the sample holder
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 12 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Figure 6: Dimensions of a P-736.Zxx; a sample holder, as for example the M-687.AP1, is to be fastened to the
P-736.Zxx using the mounting holes marked by the grey arrows
Start-Up and Operation
General Notes on Start-Up and Operation
NOTICE
Destruction of the piezo actuator by electric flashovers!
The use of the P-736.Zxx in environments that increase the electrical conductivity
can lead to the destruction of the piezo actuator by electric flashovers. Electric
flashovers can be caused by moisture, high humidity, liquids and conductive
materials such as metal dust. In addition, electric flashovers can also occur in certain
air pressure ranges due to the increased conductivity of the air.

Avoid operating the P-736.Zxx in environments that can increase the electric
conductivity.

Only operate the P-736.Zxx within the permissible ambient conditions and
classifications (p. 21).
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 13 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
NOTICE
Destruction of the piezo actuator by continuously high voltage!
The constant application of high voltage to piezo actuators can lead to leakage
currents and flashovers that destroy the ceramic.
If the P-736.Zxx is not used, but the controller is to remain switched on to ensure
temperature stability:

Set the piezo voltage to 0 V on the controller.
NOTICE
Uncontrolled oscillation!
Oscillations can cause irreparable damage to the P-736.Zxx. Oscillations are
indicated by a humming and can result from the following causes:

The load and/or dynamics of operation differ too much from the calibration
settings.

The P-736.Zxx is operated near its resonant frequency.

If you notice oscillations, stop the P-736.Zxx immediately.
Starting Up and Operating the P-736.Zxx
INFORMATION
Systems are calibrated at the factory to achieve optimum performance. Replacing
the system components will cause a loss in performance.

Note the assignment of the stage axes to the controller channels which is given
by the calibration label of the piezo servo controller.

If the piezo servo controller or the stage has to be replaced, recalibrate the axis
displacement (see controller manual) or contact our customer service
department (p. 14).
Prerequisite
 You have read and understood the following sections:

Safety Precautions (p. 5)

General Notes on Start-Up and Operation (p. 12)
Starting up and operating the P-736.Zxx
 Follow the instructions in the manual of the used piezo controller for start-up and operation of
the P-736.Zxx.
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 14 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Discharging the P-736.Zxx
The P-736.Zxx must be discharged before demounting. Demounting is necessary e.g. before cleaning
or transporting the P-736.Zxx as well as for modifications
Discharging an P-736.Zxx that is connected to the controller
 Set the piezo voltage to 0 V on the controller.
Discharging an P-736.Zxx that is not connected to the controller
 Connect the Z piezo scanner to the switched-off PI controller for 10 seconds.
Maintenance
NOTICE
Misalignment from manipulating Z piezo scanner!
The P-736.Zxx is maintenance-free and precisely aligned.

Do not try to manipulate or to open the P-736.Zxx.
Cleaning the P-736.Zxx
Prerequisites
 You have discharged the piezo actuators of the P-736.Zxx.
 You have disconnected the P-736.Zxx from the controller.
Cleaning the P-736.Zxx
 Clean the surface of the P-736.Zxx with a towel that is only lightly dampened with a mild
cleanser or disinfectant, with alcohol or with isopropanol.
 Do not do any ultrasonic cleaning.
Customer Service
For inquiries and orders, contact your PI sales engineer or send us an e-mail ([email protected]).
If you have questions concerning your system, have the following information ready:




Product codes and serial numbers of all products in the system
Firmware version of the controller (if present)
Version of the driver or the software (if present)
Operating system on the PC (if present)
The latest versions of the relevant user manuals for your system are available for download on our
website (http://www.pi.ws).
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 15 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Technical Data
PD73Z2ROW /
PD73Z2RNW
PD73Z2COW /
PD73Z2CNW
Lower cost system
with P-736.ZRO
stage for Olympus
microscopes /
P-736.ZRN stage for
Nikon microscopes
Higher precision
system with
P-736.ZCO stage
for Olympus
microscopes /
P-736.ZCN stage
for Nikon
microscopes
Z
Z
Integrated sensor
Piezoresistive
Capacitive
Travel
(closed-loop)
220
220
µm
Resolution
(closed-loop)
1
1
nm
Resonant frequency
(loaded, at 100 g)
250
250
Hz
Load capacity
500
500
g
PICMA® P-885
PICMA® P-885
Operating temperature range
15 to 40
15 to 40
Material
Aluminum
Aluminum
Mass
850
850
Active axes
Unit
Tolerance
Motion and positioning
typ.
Mechanical properties
max.
Drive properties
Ceramic type
Miscellaneous
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
°C
g
±5 %
Page 16 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
PD73Z2ROW /
PD73Z2RNW
PD73Z2COW /
PD73Z2CNW
Unit
Tolerance
Cable length
1.5
1.5
m
±10 mm
Piezo controller
E-709 controller
included in delivery
Interface / communication
USB, RS-232, SPI
I/O connector
HD Sub-D 26-pin
1× analog input 0 to10 V
1× sensor monitor 0 to10 V
1× digital input (LVTTL, programmable)
1× analog output
5× digital outputs (LVTTL, 3× predefined, 2× programmable)
Command set
PI General Command Set (GCS)
User software
PIMikroMove
Software drivers
LabVIEW drivers, shared libraries for Windows and Linux. Is
supported by MATLAB, MetaMorph, µManager, Andor iQ
Supported functionality
Wave generator, data recorder, auto zero, trigger I/O,
Controller dimensions
160 mm x 96 mm x 33 mm
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 17 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Dimensions
Decimal places are separated in commas.
Dimensions of P-736.ZCN
Figure 7: P-736.ZCN, Note 1: Cable length 1.7 meter, termination is a sub-D7W2
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 18 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Dimensions of P-736.ZCO
Figure 8: P-736.ZCO, Note 1: Cable length 1.7 meter, termination is a sub-D7W2
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 19 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Dimensions of P-736.ZRN
Figure 9: P-736.ZRN, Note 1: Cable length 1.7 meter, termination is 9-pin sub-D
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 20 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Dimensions of P-736.ZRO
Figure 10: P-736.ZRO, Note 1: Cable length 1.7 meter, termination is a 9-pin sub-D
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 21 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Maximum Ratings
The P-736.Zxx is designed for the following maximum ratings:
Maximum operating
voltage
Maximum operating
frequency (unloaded)
Maximum power
consumption
-30 to 130 V
100 Hz
20 W
Ambient Conditions and Classifications
The following ambient conditions and classifications must be observed for the P-736.Zxx:
Area of application
For indoor use only
Maximum altitude
2000 m
Air pressure
1100 hPa to 0.1 hPa
(corresponds to roughly 825 torr to
0.075 torr)
Relative humidity
Highest relative humidity 80% for
temperatures up to 31°C
Decreasing linearly to 50% relative
humidity at 40°C
Operating temperature*
15°C to 40°C
Storage temperature
0°C to 70°C
Transport temperature
-25°C to 85°C
Overvoltage category
II
Protection class
I
Degree of pollution
1
Degree of protection
IP20
according to IEC 60529
*Specifications assured from 17°C to 23°C, performance may be reduced outside this range.
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws
Page 22 / 22
2014-01-27
P736T0001, valid for PD73Z2xxW
Old Equipment Disposal
In accordance with the applicable EU law, electrical and electronic equipment may not be disposed of
with unsorted municipal wastes in the member states of the EU.
When disposing of your old equipment, observe the international, national and local rules and
regulations.
To meet the manufacturer’s product responsibility with regard to this product, Physik Instrumente
(PI) GmbH & Co. KG ensures environmentally correct disposal of old PI equipment that was first put
into circulation after 13 August 2005, free of charge.
If you have old PI equipment, you can send it postage-free to the following address:
Physik Instrumente (PI) GmbH & Co. KG
Auf der Römerstr. 1
D-76228 Karlsruhe, Germany
EC Declaration of Conformity
For the PInano® piezo scanner systems, an EC Declaration of Conformity has been issued in
accordance with the following European directives:
2006/95/EC, Low Voltage Directive
2004/108/EC, EMC Directive
2011/65/EU, RoHS Directive
The applied standards certifying the conformity are listed below.
Electromagnetic Emission: EN 61000-6-3:2007, EN 55011:2009
Electromagnetic Immunity: EN 61000-6-1:2007
Safety (Low Voltage Directive): EN 61010-1:2010
Physik Instrumente (PI ) GmbH & Co. KG_Auf der Roemerstrasse 1_76228 Karlsruhe/Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019 E-mail [email protected], www.pi.ws