detect and identify Revision of IEC 61005 Neutron ambient dose equivalent meters Alfred Klett Berthold Technologies, Bad Wildbad Germany Neutron Users Club Meeting NPL/Teddington, November 1, 2011 History of Standard IEC 61005 Radiation protection instrumentation – Neutron ambient dose equivalent (rate) meters prepared by IEC technical committee 45: Nuclear instrumentation subcommittee 45B: Radiation protection instrumentation working group B8 Active Electronic Dose Equivalent and Dose Equivalent Rate Meters and Monitors current project leader: Radoslav Radev / LLNL IEC 61005 1st edition IEC 61005 2nd edition IEC 61005 3rd edition (CD) Alfred Klett Nov 1, 2011 1990-09 2003-02 2011-11 2 IEC 61005 Revision Documents & Plan Document ID 45B/684/RR 45B/685/CD 45B/685/CC_MV Title Date Review Report (RR) 2011-02-18 Committee Draft (CD) 2011-02-18 Comments (CC) 2011-05-16 Schedule Committee Draft for Vote (CDV) Final Draft International Standard (FDIS) Publication of International Standard Alfred Klett Nov 1, 2011 2013-02 2014-02 2014-06 3 Scope assemblies for the measurement of ambient dose equivalent for neutrons up to 20 MeV, comprising: detection assembly (for example thermal neutron detector with neutron moderator/absorber) measuring assembly (display reading) not applicable to unconventional dose meters as spectroscopic neutron dose meters (more than 1 probe) meters based on bubble dosimeter technology meters based on silicon pillar neutron detectors electronic personal dose (rate) meters No requirements for pulsed radiation fields defined Alfred Klett Nov 1, 2011 4 Normative references Electromagnetic compatibility (EMC) IEC IEC IEC IEC IEC IEC IEC 61000-4-2:1995 61000-4-3:2002 61000-4-4:1995 61000-4-5:1995 61000-4-6:1996 61000-4-8:1993 61000-4-12:1995 ISO 8529 Reference neutron radiations ISO 8529-1:2001 Characteristics and methods of production. ISO 8529-2:2000 Calibration fundamentals ISO 8529-3:1998 Calibration and determination of response ISO 11929:2010 Determination of the characteristic limits Alfred Klett Nov 1, 2011 5 Tests Tests for influence quantities of Type F Type S Radiation sources 241Am-Be, 252Cf 241Am-Be, 252Cf moderated by D2O sphere ∅ 30cm accelerator target sources (ISO 8529) for thermal & epithermal reference fields, accelerator target, reactor beams, 241Am-Be or 252Cf sources with well defined moderator/filter arrangements may be used Alfred Klett Nov 1, 2011 6 Radiation sources for tests 241Am-Be, 252Cf bare 241Am-Be, 252Cf moderated by D O sphere ∅ 30cm 2 accelerator target sources (ISO 8529) for thermal & epithermal ref. fields, accelerator target, reactor beams, 241Am-Be or 252Cf sources with well defined moderator/filter arrangements Work place neutron fields simulated fields specified in ISO 12789 other work place environments whose fields are well defined by spectral calculations and/or measurements traceable to or recognized by a primary standards laboratory Alfred Klett Nov 1, 2011 7 General requirements: effective range For dose meters with an analogue display: omitted here for dose meters with a digital display, from an indication in the 2nd least significant digit up to the maximum indication on each range. (example: with max. indication 9 999,9, the effective range can extend from 1,0 to 9 999,9) for dose meters with digital & scientific display the mantissa shall have 3 digits at least (for instance 1,00 to 9,99). The manufacturer shall define the effective range of measurement (for instance 1,00 E–7 to 9,99 E–2 Sv h–1). The minimum effective range of measurement of dose equivalent rate: at least 4 orders of magnitude including 10 µSv h–1 dose equivalent: at least 4 orders of magnitude including 0,1 mSv. Alfred Klett Nov 1, 2011 8 General requirements Mechanical characteristics assembly labels shall include measured quantity effective range of measurement the type of radiation the dose meter is suitable for rated range of particle energy reference point & orientation (or in the manual) IP classification shall be stated by the manufacturer according to IEC 60529 Ease of decontamination Alfred Klett Nov 1, 2011 9 General requirements Interface requirements output connection for a remote readout is recommended recommended input and output connections input connection to the pre-amplifier input for electrical test purposes output connection from the amplifier output connection from the discriminator manufacturer shall deliver the evaluation algorithm of the indicated value Alfred Klett Nov 1, 2011 10 Radiation detection requirements indication of the assembly shall be in units of ambient dose equivalent (rate), for example in millisieverts (per hour) variation of relative response due to the nonconstancy of the dose rate response shall not exceed –17 % to +25 % over the whole of the effective range of measurement for neutron reference radiations chosen The statistical fluctuations of the indication measured as coefficient of variation shall fulfill the requirements given in Table 3. Alfred Klett Nov 1, 2011 11 it is permissible to employ an equivalent electrical test in order to determine the intrinsic error at the ambient dose equivalent rates that cannot be provided by the sources of radiation. Alfred Klett Nov 1, 2011 12 Variation of the response due to neutron radiation energy and angle of incidence Combination of requirements for angular and energy response was intended (like for instance IEC 60841-1:2009) But it was unclear if any existing instrument could meet these requirements they are now independent in the CD Alfred Klett Nov 1, 2011 13 Variation of response due to neutron energy “it would be desirable for the variation of the response with the neutron energy over the whole energy range from thermal to the maximum energy specified by the manufacturer not to exceed 50%. However, at the time of publication, such requirement is not practically achievable” (6.4.1) appropriate response calculations, the results of these calculations shall be made available for the entire energy range with data at least at two energy points per decade of radiation energy. Alfred Klett Nov 1, 2011 14 Neutron Survey Instruments International Atomic Energy Agency IAEA, Technical Reports Series No. 403, Compendium of Neutron Spectra and Detector Responses for Radiation Protection Purposes, Supplement to Technical Report Series No. 318, Vienna, 2001, Table 3.II. p. 27 Alfred Klett Nov 1, 2011 15 Energy Dependent Dose Response Rem-Counters Calculated from the data of IAEA TRS 418 Alfred Klett Nov 1, 2011 16 Variation of response due to neutron energy Energy Range Requirements thermal - 50 keV manufacturer shall specify variation of ambient dose equivalent with neutron energy. The recommended variation should not be more than 20%-800% 50 keV - 10 MeV within 50%-150% of response for 0° incident radiation and reference energy 10 MeV – max. Energy manufacturer shall specify variation of ambient dose equivalent with neutron energy. The recommended variation should not be more than 20%-100% Alfred Klett Nov 1, 2011 17 Method of test of energy response at least at two neutron energies below 50 keV or with a moderated source at least at three neutron energies in the energy range between 50 keV and 10 MeV at least with 1 broad source (e.g. 252Cf or 241Am-Be) at least at one neutron energy exceeding 10 MeV only tests at energies within the manufacturer’s specified energy range are required In addition, it is recommended to state response to standardized simulated work place neutron fields Alfred Klett Nov 1, 2011 18 Dose & dose rate alarm requirements when the dose equivalent (rate) meter is subjected to a dose equivalent (rate) of 0,8 of the dose equivalent (rate) alarm set point for 10 min, the alarm shall not be activated Similarly, at a dose equivalent (rate) of 1,2 times the alarm level set, the alarm shall be activated the alarm shall actuate within 5 s or within a time such that the product of this time and the dose equivalent rate of the alarm point is less than 10 µSv Alfred Klett Nov 1, 2011 19 Additivity of indicated value requirements The indicated value shall be additive with respect to simultaneous irradiation with different types of radiation (for example, photon and neutron) and with different neutron energies and with different angles of radiation incidence Alfred Klett Nov 1, 2011 20 Test additivity of indicated value Perform subsequently two irradiations under the two different irradiation conditions K and L (different energies, different angles of incidence or even different types of radiations) with the conventional quantity values HK and HL Determine the indicated values GK and GL for the two irradiations. Then perform a third simultaneous irradiation under the two irradiation conditions K and L with the conventional quantity value HK+L = HK + HL and determine the indicated value GK+L Alfred Klett Nov 1, 2011 21 Test additivity of indicated value shall be determined for any value of HK and HL and any simultaneous combination of radiation fields SK and SL As simultaneous irradiations are very difficult to perform, the use of calculations as a replacement for the simultaneous irradiations is permitted and recommended for this test ∆gmix shall be lower than 0.1 Alfred Klett Nov 1, 2011 22 Variation of angular response The response values shall be measured for polar angles of incidence of α = 0°, ±45°, ±90° for azimuth angles β of 0°, 90°, 180°, 270°, 360° while keeping α = 0° measurements shall be performed at one monoenergetic or one broad neutron spectrum. Test distance should be at least 3 times the sum of the largest linear dimension of source & detector The scatter contribution to the indicated value of the device shall not exceed 20 % (ISO 8529-2). Alfred Klett Nov 1, 2011 23 Overload characteristics The dose equivalent meter shall read off-scale on the high side or shall indicate overload when exposed to doses greater than the maximum of its measuring range When subjected to dose rates high enough to cause wrong dose indication, there shall be indication that the equipment is not able to provide correct dose indication Alfred Klett Nov 1, 2011 24 Overload chracteristics test ?!? CHECK Subject the dosemeter to a dose in excess of 10 times the maximum dose that can be indicated, at most 50 Sv, at least 1 Sv. Dose rate during exposure shall be less than the maximum dose rate capability as specified by the manufacturer. The equipment shall not be reset or switched off for at least 30 min after the equipment has been subjected to the test dose. Subject the dose equivalent meter to a dose rate 10 % in excess of that specified as the dose rate limit by the manufacturer for a period of 100 s. In Alfred Klett Nov 1, 2011 25 case, no error of the dose Response time 30 s for the increases or decreases of the ambient dose equivalent rate less than 0,1 mSv⋅h–1 10 s for the increases or decreases of the ambient dose equivalent rate between 0,1 mSv⋅h1and 1 mSv⋅h–1; 4 s for the increases or decreases of the ambient dose equivalent rate greater than 1 mSv⋅h–1. The manufacturer shall state the response time. Alfred Klett Nov 1, 2011 26 Response time for the increases or decreases of the ambient dose equivalent rate 30 s for less than 0,1 mSv⋅h–1 10 s between 0,1 mSv⋅h1and 1 mSv⋅h–1; 4 s for greater than 1 mSv⋅h–1. The manufacturer shall state the response time Definition Alfred Klett Nov 1, 2011 27 Response time tests The test may be carried out either with a suitable neutron source or by the injection of a suitable electrical signal The initial and final ambient dose equivalent rates shall differ by a factor of at least 10 and measurements shall be carried out for both an increase and a decrease in the ambient dose equivalent rate by this factor Alfred Klett Nov 1, 2011 28 Variation of the dose response due to dose rate dependence - Requirements The variation of the relative dose response due to dose rate dependence shall not exceed the range from –13 % to +18 % In addition, the variation of the rel. response due to low dose rates down to natural environmental radiation shall be stated by the manufacturer Alfred Klett Nov 1, 2011 29 Test of variation of the dose response due to dose rate dependence Determine dose responses at 80 % of each order of magnitude of the effective dose range when meter is exposed to a source Perform the test for each dose at one dose rate of each order of magnitude of the rated range of use for dose rate Since at lower dose values, the exposure times are too short, whilst at high dose values, the exposure times are too long for the lower rates, these tests shall exclude any exposure involving measurement times less than 10 s or in excess of 10 h The variation of the relative dose response due to dose rate dependence shall not exceed the range from –13 % to +18 % If integration is done digitally a test of ≈100 s per dose rate is sufficient. Alfred Klett Nov 1, 2011 30 Dose equivalent rate alarm requirements when the dose meter is subjected to 0,8 of the dose equivalent rate alarm set point for 10 min the alarm shall not be activated for more than 10% of the period of test Similarly, at a dose equivalent rate of 1,2 times the alarm level set, the alarm shall be activated for 90% of the test period. alarm shall actuate within 5 s or within a time such that the product of this time and the dose equivalent rate of the alarm point is less than 10 µSv Alfred Klett Nov 1, 2011 31 Dose equivalent alarm requirements when the dose equivalent meter is subjected to a dose equivalent of 0,8 times the dose equivalent alarm set, no alarm shall be given when the dose equivalent meter is subject to a dose equivalent rate of 1,2 times the dose equivalent alarm set point, the alarm shall be actuated. Alfred Klett Nov 1, 2011 32 Response to photon radiation Requirements Photon response shall be quoted in terms of units of the photon ambient dose equivalent rate Photons may not only cause the assembly to give an indication, but it may also modify the response to neutron radiation. Therefore, there are two separate requirements Indication to 137Cs photons at 10 mSv⋅h–1 shall not exceed indication of 0,1 mSv⋅h–1 neutron high-energy photon radiation (6 MeV) response shall, by agreement between the manufacturer and the purchaser, be checked at higher energies Alfred Klett Nov 1, 2011 33 Test of response to photon radiation assembly shall be exposed to a 137Cs source in a field having an ambient dose equivalent rate of 10 mSv⋅h–1 indicated value shall not exceed 0,1 mSv⋅h–1 Radiation sources used shall conform with the ISO 4037 series indication produced by a 137Cs and higher energy (> 1.5 MeV) photon dose rate of 10 mSv⋅h–1 is less than the indicated value due to a neutron ambient dose equivalent rate of 0,1 mSv⋅h–1 Alfred Klett Nov 1, 2011 34 Additivity of indicated value The indicated value shall be additive with respect to simultaneous irradiation with different types of radiation (f. e. photon and neutron) different neutron energies and angles of incidence If dosemeter uses only one signal (measured with one detector) for evaluation, then this is fulfilled more than one signal (measured either with several detectors or with one detector using for example pulse height analysis), then this is not automatically fulfilled Alfred Klett Nov 1, 2011 35 Additivity of indicated value In that case, it shall be assured that the relative change in indication,mixg∆, caused by the mix of radiation shall not exceed ± 0,1 NOTE If the algorithm used to evaluate the indicated value, (see 5.5) is either a linear combination of the signals or a linear optimization of them, then this requirement is fulfilled or no tests are required Alfred Klett Nov 1, 2011 36 Other Requirements Environmental Temperature Temperature shock Humidity Mechanical Requirements Mechanical shock Drop Test Vibration Test Electromagnetic Requirements Documentation Alfred Klett Nov 1, 2011 37 References ICRU The International Commission on Radiation Units and Measurements, Fundamental Quantities and Units for Ionizing Radiation, ICRU Report 60, ICRU Publications, Bethesda, MD, 1998 ICRU The International Commission on Radiation Units and Measurements, Determination of Operational Dose Equivalent Quantities for Neutrons, ICRU Report 66, Journal of the ICRU, Vol. 1 No. 3, 2001 IAEA Compendium of Neutron Spectra and Detector Responses for Radiation Protection Purposes, Technical Reports Series No. 318, International Atomic Energy Agency, Vienna, 1990 IAEA Compendium of Neutron Spectra and Detector Responses for Radiation Protection Purposes, Supplement to Technical Reports Series No. 318, Technical Reports Series No. 403, International Atomic Energy Agency, Vienna, 2001 Alfred Klett Nov 1, 2011 38 References R.J. Tanner, C. Molinos, N.J. Roberts, D.T. Bartlett, L.G. Hager, L.N. Jones, G.C. Taylor, D.J. Thomas, Practical Implications of Neutron Survey Instrument Performance Report HPA-RPD-016 , Nov. 2006, Health Protection Agency, Chilton, Didcot, United Kingdom ICRP The International Commission on Radiological Protection, Conversion Coefficients for use in Radiological Protection against External Radiation Publication 74, Annals of the ICRP, Vol. 26, No. 3-4, 1996 Alfred Klett Nov 1, 2011 39
© Copyright 2026 Paperzz