GE Healthcare Kvick Lab™ cross flow system Operating Instructions Original Instructions Table of Contents Table of Contents 1 Introduction .............................................................................. 5 1.1 1.2 1.3 1.4 1.5 1.6 2 Safety instructions................................................................. 19 2.1 2.2 2.3 2.4 3 7 Operation overview .............................................................................................. 43 Starting the instrument ...................................................................................... 43 The pump control system ................................................................................. 43 Maintenance ........................................................................... 51 5.1 5.2 5.3 5.4 5.5 5.6 6 Site requirements .................................................................................................. 25 Unpacking ................................................................................................................. 25 Moving and carrying the system ................................................................... 26 Connections ............................................................................................................. 26 Installing a cassette holder .............................................................................. 28 Installing cassettes in the holder ................................................................... 28 Spare parts and accessories ........................................................................... 41 Operation ................................................................................ 43 4.1 4.2 4.3 5 Safety precautions ............................................................................................... 19 Labels .......................................................................................................................... 22 Emergency procedures ...................................................................................... 23 Recycling procedures .......................................................................................... 24 Installation .............................................................................. 25 3.1 3.2 3.3 3.4 3.5 3.6 3.7 4 Important user information ................................................................................. 6 Regulatory information ......................................................................................... 8 Instrument ................................................................................................................ 10 Flow path .................................................................................................................. 14 Kvick Lab cassette holder ................................................................................. 15 Cassettes ................................................................................................................... 16 General ....................................................................................................................... 51 User maintenance schedule ............................................................................ 51 Cleaning ..................................................................................................................... 51 Cleaning and storing cassettes ...................................................................... 52 Maintenance of the holder ............................................................................... 54 Replacement of fuses ......................................................................................... 55 Troubleshooting ..................................................................... 57 Reference information .......................................................... 63 7.1 7.2 7.3 7.4 Kvik Lab system specifications ....................................................................... 63 Chemical compatibility ....................................................................................... 66 Service and training ............................................................................................. 67 Ordering information ........................................................................................... 67 Kvick Lab Operating Instructions 28-9646-76 AA 3 Table of Contents Appendix A ................................................................................... 69 A.1 A.2 4 Drawings ............................................................................................................ 69 Using the cartridge adaptor kit ............................................................... 72 Kvick Lab Operating Instructions 28-9646-76 AA Introduction 1 1 Introduction Purpose of the Operating Instructions The Operating Instructions provide you with the instructions needed to handle Kvick Lab in a safe way. Prerequisites In order to operate Kvick Lab safely and according to the intended purpose the following prerequisites must be met: • You should be acquainted with the use of general laboratory equipment and with handling of biological materials. • You must read Chapter 2 Safety instructions‚ on page 19 of these Operating Instructions. • The system should be installed according to the instructions in Chapter 3 of these Operating Instructions. In this chapter This chapter contains important user information and a general description of Kvick Lab and its intended use. Kvick Lab Operating Instructions 28-9646-76 AA 5 1 1.1 Introduction Important user information 1.1 Important user information Read this before using the Kvick Lab All users must read the Safety Instructions in Chapter 2 of these Operating Instructions before installing, using or maintaining the system. Do not operate Kvick Lab in any other way than described in the user documentation. If you do, you may be exposed to hazards that can lead to personal injury and you may cause damage to the equipment. Intended use The Kvick Lab cross flow system is intended for laboratory scale filtration of biological solutions using membrane cassettes under the conditions stated in this user manual. If you use the system and cassettes in a manner not specified by GE Healthcare, you may impair the protection provided by the system and cassettes. When using any laboratory, pilot-scale, or production cross flow equipment, the potential exists for personal injury unless you follow established safety procedures. When using GE Healthcare products, you should follow OSHA, federal, state, and local safety regulations for equipment installation and operation. You should follow your company’s safety regulations. You should follow the specific safety instructions provided in this user manual and any original equipment manufacturer user manuals provided with your system. Only qualified personnel who are adequately trained and who understand the operating instructions should install, operate, maintain, and inspect the Kvick Lab system. Safety notices These Operating Instructions contain WARNINGS, CAUTIONS and NOTICES concerning the use of the product, with meanings as defined below. WARNING WARNING indicates a hazardous situation which, if not avoided, could result in death or serious injury. It is important not to proceed until all stated conditions are met and clearly understood. 6 Kvick Lab Operating Instructions 28-9646-76 AA Introduction 1 Important user information 1.1 CAUTION CAUTION indicates a hazardous situation which, if not avoided, could result in minor or moderate injury. It is important not to proceed until all stated conditions are met and clearly understood. NOTICE NOTICE indicates instructions that must be followed to avoid damage to the product or other equipment. Kvick Lab Operating Instructions 28-9646-76 AA 7 1 1.2 Introduction Regulatory information Notes and tips Note: A Note is used to indicate information that is important for trouble-free and optimal use of the product. Tip: A tip contains useful information that can improve or optimize your procedures. 1.2 Regulatory information This section lists the directives and standards that are fulfilled by Kvick Lab. Manufacturing information Requirement Content Name and address of manufacturer GE Healthcare Bio-Sciences AB, Björkgatan 30, SE 751 84 Uppsala Sweden Name and ID of notified body - Place and date of declaration Uppsala, Sweden, May 2009 Identity of person authorized to sign DoC See EC Declaration of Conformity CE Conformity Directive Title 2006/42/EC Machinery Directive (MD) 2006/95/EC Low Voltage Directive (LVD) 2004/108/EC ElectroMagnetic Compatibility (EMC) Directive International standards 8 Standard Description Notes EN 61010-1, IEC 61010-1, CAN/CSA-C22.2 no. 61010-1 Safety requirements for electrical equipment for measurement, control and laboratory use Kvick Lab Operating Instructions 28-9646-76 AA Introduction 1 Regulatory information 1.2 Standard Description Notes EN 61326-1 EMC emissions and immunity requirements for measurement, control and laboratory use Harmonized with 2004/108/EC EN-ISO 12100-1, 12100-2 Safety of machinery – Basic concepts, general principles for design Harmonized with 2006/42/EC EN-ISO 14121-1, 14121-2 Safety of machinery – Principles of risk assessment Harmonized with 2006/42/EC CE marking The CE marking and the corresponding Declaration of Conformity is valid for the instrument when it is: • used as a stand-alone unit, or • connected to other CE-marked instruments, or • connected to other products recommended or described in the user documentation, and • used in the same state as it was delivered from GE Healthcare, except for alterations described in the user documentation or explicitly authorized by GE Healthcare. Regulatory compliance of connected equipment Any equipment connected to Kvick Lab cross flow system should meet the safety requirements of EN 61010-1/IEC61010-1 or relevant harmonized standards. Within the European Union, connected equipment must be CE-marked. Kvick Lab Operating Instructions 28-9646-76 AA 9 1 1.3 Introduction Instrument 1.3 Instrument The main components of a Kvick Lab cross flow system include: • Stainless steel, jacketed reservoir • Rotary lobe feed pump • Kvick Lab cassette holder • Valves • In-line pressure gauges • Fittings and piping • Torque wrench and socket The following sections describe each component. 9 10 4 5 1 6 7 8 2 3 Figure 1-1. Kvick Lab cross flow system. Table 1-1. .Parts in Figure 1-1. 10 Part Description 1 Feed reservoir 2 Pump head 3 Drain valve 4 Kvick Lab cassette holder Kvick Lab Operating Instructions 28-9646-76 AA Introduction 1 Instrument 1.3 Part Description 5 Retentate pressure gauge 6 Retentate valve 7 Pump cabinet 8 Pump controls 9 Feed pressure gauge (not visible in picture) 10 Permeate valve Feed Reservoir The feed reservoir holds your process solution, feeds the solution to the feed pump, and collects the retentate stream. The feed reservoir's cover includes two inlet ports enabling you to vent the reservoir and introduce buffer during diafiltration. The feed reservoir's cooling-heating jacket enables you to control the temperature of the solution in the reservoir. The reservoir holds 2.5 liters of feed solution and is constructed of stainless steel. A clamp holds the reservoir cover in place. Six inlet and outlet ports support the functions of the feed reservoir. The feed port located on the bottom of the reservoir feeds the pump. The retentate port on the side of the tank accepts the retentate flow. The inlet and outlet ports for the cooling-heating jacket are located on the side of the reservoir and are supplied with standard pipe threads for connecting your cooling-heating lines. The small port on the reservoir cover enable you to attach a filter for sterile venting of the reservoir, and the large reservoir cover port enables you to introduce feed solution or diafiltration solution into the system. All the process line ports use sanitary connectors for connections. 11 1 2 4 5 3 6 7 Figure 1-2. Kvick Lab stainless steel, jacketed feed reservoir. Kvick Lab Operating Instructions 28-9646-76 AA 11 1 1.3 Introduction Instrument Table 1-2. Parts in Figure 1-2. Part Description 1 Vent port 2 Fill port 3 Cooling/heating water inlet 4 Clamp for reservoir cover 5 Cooling/heating water outlet 6 Retentate inlet 7 Feed outlet Rotary lobe feed pump The pump provides precise control over feed flow, provides low shear pumping, and is easy to operate. An original equipment user manual is provided with the feed pump. Read the feed pump user manual completely before operating the system. The 230 volt European model includes a high-pressure cut-off safety switch that is controlled by the feed line pressure gauge. If the feed pressure reaches 4 barg (60 psig), the pump shuts down. Kvick Lab cassette holder Your Kvick Lab system includes a Kvick Lab cassette holder and user manual. The user manual describes how to use and care for the holder and cassettes. Read the holder’s user manual before operating the system. The Kvick Lab holder simplifies operation with fewer parts and connections than other types of cross flow filtration cassette holders. The Kvick Lab holder can hold one to five Kvick Lab cassettes, each cassette having 0.11 2 2 2 2 2 2 m (1.2 ft ) of membrane surface area for a range of 0.11 m (1.2 ft ) to 0.55 m (6.25 ft ). The cassette holder includes a stand with height-adjustable feet to level and stabilize the cassette holder on the pump housing. You can adjust the height of the holder by loosening the lock nut on the holder feet, rotating the feet to adjust the height, then tightening the lock nut. Valves Four sanitary diaphragm valves mounted on the feed, permeate, retentate, and drain lines enable you to control the operating flows and pressures. The valve located at the feed pump outlet has one inlet port and two outlet ports. One port directs the feed to drain, and the other port directs the feed to the cassette holder. 12 Kvick Lab Operating Instructions 28-9646-76 AA Introduction 1 Instrument 1.3 The flow path is controlled by the operator by opening or closing the valve flow control handles. Of the three remaining valves, one controls the permeate flow, one controls the retentate flow, and one controls the flow of compressed air into the feed line when an integrity tester is attached to the system. Pressure gauges The system includes two, 0 to 4 barg (0 to 60 psig), glycerin-filled, in-line, pressure gauges to monitor feed and retentate pressure. The feed pressure gauge on European models includes a transmitter. If the feed pressure rises above 4 barg (60 psig), the signal from the transmitter triggers a safety control circuit that shuts the pump down. 1.3.1 Fittings and piping The system includes a number of stainless steel components, tubing, fittings, and sanitary clamps that come pre-assembled. In addition to the assembled components, the following parts are supplied with the system: Two, 1/2-inch NPT-to-3/8-inch hose barb fittings—screw these into the heating jacket ports so that you can attach heating/ cooling water tubing to the reservoir jacket. Two, 24-inch-long pieces of 3/8-inch ID tubing and three clamps—attach one piece of tubing to the permeate hose barb and secure with a clamp. Direct the other end of the tubing to either a drain, collection vessel, or the feed reservoir inlet, depending upon the application. Attach the other piece of tubing to the feed-line hose barb and secure with a clamp. Direct and attach the other end of the line to an integrity tester if needed. One, 1/2-inch sanitary cap—use this cap and a user-supplied sanitary clamp to blank off any 1/2-inch sanitary fitting, such as the feed line drain valve fitting. One, vortex diverter—the vortex diverter is a square piece of metal placed at the bottom of the feed reservoir to prevent the formation of a vortex. You can purchase additional piping, fittings, and clamps through your local GE Healthcare representative. Torque wrench and socket The system includes a torque wrench, a 9/16-inch socket for precise tightening of the cassette holder tie rod nuts, and a user manual. Read the torque wrench user manual before clamping the cassettes in the cassette holder. Kvick Lab Operating Instructions 28-9646-76 AA 13 1 1.4 Introduction Flow path 1.4 Flow path The standard flowpath for the Kvick Lab system enables you to complete many cross flow processes. 10 4 14 11 12 5 1 15 8 2 8 6 5 16 3 8 13 9 7 Figure 1-3. Kvick Lab cross flow system flow diagram. Table 1-3. Parts in Figure 1-3. 14 Part Description 1 Kvick Lab cassette holder 2 Retentate line 3 Feed line 4 Cooling /Heating connection for jacket 5 Pressure gauge 6 Permeate line 7 Collection vessel 8 Valve 9 Drain line Kvick Lab Operating Instructions 28-9646-76 AA Introduction 1 Kvick Lab cassette holder 1.5 Part Description 10 Vent port 11 Feed reservoir 12 Cooling/Heating jacket 13 Pump 14 Diafiltration (solution) feed line 15 Cooling/Heating water 16 Double valve 1.5 Kvick Lab cassette holder The Kvick Lab cassette holder is designed to hold 1 to 5 Kvick Lab cassettes. Together, the holder and the cassettes enable you to separate, concentrate, and diafiltrate biological solutions with precision. For example, you can complete ultrafiltration at laboratory scale volumes (starting volumes of less than 1 liter). Figure 1-4. Kvick Lab cassette holder. Table 1-4. Parts in Figure 1-4. Part Description 1 Flow distribution plate 2 Back plate Kvick Lab Operating Instructions 28-9646-76 AA 15 1 1.6 Introduction Cassettes Part Description 3 Spacers 4 Washer 5 Nut 6 Guide rod 7 Stand 8 Retentate port 9 Permeate port 10 Feed port 11 Tie rod 1.6 Cassettes The two sizes of Kvick Lab cassettes share design components, including inlet and outlet ports, labeling, and alignment notches (Figure 1-5). Figure 1-5. Cassette. Table 1-5. Parts in Figure 1-5. 16 Part Description 1 Feed port 2 Retentate port 3 Permeate port 4 Cassette information 5 Alignment notches 6 Incorporated gasket Kvick Lab Operating Instructions 28-9646-76 AA Safety instructions 2 Safety precautions 2.1 2 Safety instructions This chapter describes safety compliance, safety labels, general safety precautions, emergency procedures, power failure and recycling of Kvick Lab. 2.1 Safety precautions Introduction Before installing, operating of maintaining the system, you must be aware of the hazards described in the user documentation. Follow the instructions provided to avoid personal injury or damage to the equipment. The safety precautions in this section are grouped into the following categories: • General precautions • Flammable liquids • Personal protection • Installing and moving the instrument • Maintenance Kvick Lab Operating Instructions 28-9646-76 AA 19 2 2.1 Safety instructions Safety precautions General precautions WARNING If you overpressurize the system with compressed air during integrity testing, system components can rupture causing bodily harm. To avoid the rupture of a system component due to over pressurization with compressed air, use a pressure regulator to limit the pressure of your compressed air supply to the maximum operating pressure of your system. WARNING Wear the appropriate personal protection devices when testing and using the system. WARNING If the system leaks, it can release potentially hazardous process fluids or cleaning fluids causing bodily harm. To prevent leaks and the release of potentially hazardous process or cleaning fluids, limit feed pressure to 3 barg (45 psi), periodically check parts for wear, and ensure you assemble the components correctly. Notices NOTICE Failure to carefully plan the installation of the system can result in equipment damage or poor performance. To avoid equipment damage, carefully plan the installation of the system in reference to the effects of a power failure and utility connections. Read the installation instructions in the system’s Operating Instructions. NOTICE Hollow fiber cartridges are strong and safe, but mishandling can damage the membrane fibers or cartridge housing connectors. Follow guidelines in Operations section to prevent damage to the hollow fibers and cartridge housing connectors. NOTICE If you operate the pump dry (without any process fluid in it), you can damage the pump. To prevent damage to the pump, ensure that fluid is present in the reservoir and feed line whenever you run the pump. 20 Kvick Lab Operating Instructions 28-9646-76 AA Safety instructions 2 Safety precautions 2.1 Using flammable liquids WARNING An adequately dimensioned ventilation system shall be installed when flammable or noxious substances are used. Personal protection WARNING Wear the appropriate personal protection devices and clothing when operating and cleaning the system. Basic personal protection devices include protective clothing, safety glasses, and hearing protection. Use additional protective devices (fume hood, respirator, etc.) if the nature of your product or your company’s safety regulations require it. WARNING Wear hearing protection when operating the system at high speeds for extended periods of time. WARNING Wear protective glasses at all times when in vicinity of instrument. Installing and moving the instrument WARNING Use tubing clamps on all hose barb connections, particularly connections to the feed ports for leak-free connections at high pressure. Failure to do so may lead to unexpected leaks that my lead to bodily harm. WARNING The instrument must be connected to a grounded mains socket. WARNING Do not block the rear panel of the system. The mains power switch must always be easy to access. Kvick Lab Operating Instructions 28-9646-76 AA 21 2 2.2 Safety instructions Labels CAUTION Cassette holders are heavy. The weight of cassette holders requires that you carry them properly and support them well, especially when integrated into a cross-flow system. To avoid bodily injury or equipment damage from a falling cassette holder, ensure you have a secure, controlled grip and a sufficient number of people or the proper equipment to move it. Wear safety shoes. Ensure the holder and piping are properly supported and that piping supports meet local safety codes. System operation WARNING Ensure that the retentate valve is open before starting the feed pump. Failure to do so can cause the feed pressure to reach excessively high levels causing leaks, bodily harm, and damage to the cartridge. WARNING If you place a finger in the inlet or outlet port of the pump, your finger can reach the pump rotors. Hence, if the pump starts or is running while you place a finger into the pump inlet or outlet, bodily harm will result. To prevent bodily harm, never place fingers in the inlet or outlet ports of the pump. Maintenance WARNING Disconnect the power cable, the compressed air supply, and decompress the system before inspecting or maintaining the system. Failure to do so can result in bodily harm from the accidental starting of the pump or unexpected release of pressure. WARNING Decontaminate the equipment before decommissioning to ensure the removal of all hazardous residues. 2.2 Labels This section describes safety labels and labels concerning hazardous substances that are attached to the Kvick Lab. 22 Kvick Lab Operating Instructions 28-9646-76 AA Safety instructions 2 Emergency procedures 2.3 Labels on the instrument The pump label, located on the back of the pump housing, provides key identification and specification information (Figure 2-1). The CE mark for European models is located on the lower left-hand side of the back panel. The Kvick Lab label is shown in Fig 2-2. Figure 2-1. Typical pump label. XX-XXXX-XX Code No: Serial No Mfg Year: Kvick Lab XXXXXXXX XXXXXXX 2009 N12406 Made in Sweden GE Healthcare Bio-Sciences AB 751 84 Uppsala Sweden Figure 2-2. Example of Kvick Lab marking label. 2.3 Emergency procedures This section describes how to do an emergency shut down of the Kvick Lab system. The section also describes the result in the event of power failure. Emergency procedures In an emergency situation, do as follows to stop the run: Press the Stop button on pump control panel. Kvick Lab Operating Instructions 28-9646-76 AA 23 2 2.4 Safety instructions Recycling procedures Power failure In the event of a power failure: 1 Unplug the system from the mains power receptacle until the power is restored. 2 Drain the system reservoir to recover as much process fluid as possible. 2.4 Recycling procedures The equipment shall be decontaminated before decommissioning and all local regulations shall be followed with regard to scrapping of the equipment. Disposal, general instructions When taking Kvick Lab out of service, the different materials must be separated and recycled according to national and local environmental regulations. Recycling of hazardous substances Kvick Lab contains hazardous substances. Detailed information is available from your local GE Healthcare representative. Disposal of electrical components Waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment. 24 Kvick Lab Operating Instructions 28-9646-76 AA Installation 3 Site requirements 3.1 3 Installation 3.1 Site requirements To set up your system you must choose a location, move the system to the location, and make several utility connections. Choosing a location Select a location that includes: • appropriate drains or collection vessels to collect waste and processed fluids • electrical power supply • instrument-quality compressed air if needed for customer-supplied integrity test equipment • a supply of cooling/heating water and a system to pump it through the cooling/ heating jacket if needed • sufficient room to perform maintenance on the system. Table 3-1. Electrical and environmental requirements. Parameter Requirement Voltage* 104-126 VAC, 200-230 VAC, 50/60 Hz Current 2.2 to 4.5 A Placement On sturdy laboratory bench Ambient temperature 25ºC maximum *Not suitable to Ground Fault Circuit Interrupt circuitry with 30mA breakers. Use unprotected mains to avoid nuisance tripping of breakers. 3.2 Unpacking Your Kvick Lab system arrives assembled and packaged in a wooden crate. The wooden crate includes these components: • Kvick Lab system including a Kvick Lab cassette holder and rotary lobe pump • Torque wrench and socket Kvick Lab Operating Instructions 28-9646-76 AA 25 3 3.3 Installation Moving and carrying the system • Torque wrench OEM user guide • Rotary lobe pump OEM user guide To unpack the system, place the crate on the ground. Remove the five bolts located along the bottom edge of the crate. With an assistant, slide the cover (top and sides) up off the base. Remove the foam surrounding the system. After unpacking do the following: • Check the equipment for any apparent damage before starting installation. • Document any damage carefully and contact your GE Healthcare representative. 3.3 Moving and carrying the system CAUTION The system weighs about 80 kg (175 lbs). At least four people or a lifting device are required when lifting the equipment. Failure to follow this instruction may lead to strain or injury. After unpacking, you must lift and move the system to the operating location. The system weighs about 80 kilograms (175 pounds). Therefore you should: • get help from at least 3 assistants to ensure safe, controlled lifting, • use a sturdy cart to transport the system, and • wear gloves to protect your hands. When lifting the system, support it from the pump head and the pump housing, not from the reservoir, cassette holder, or piping. Transport and lift the system onto a sturdy laboratory bench or other work table. Ensure the system is stable and level. 3.4 Connections Process lines Connect a drain line to the drain port of the feed/drain valve using a 1.2-inch sanitary connector fitting, gasket, and clamp. (Attaching a flexible drain line to the port will enable you to move the drain line between various collection vessels or a drain.) 26 Kvick Lab Operating Instructions 28-9646-76 AA Installation 3 Connections 3.4 Electrical connections Connect the Kvick Lab system power cord to an electrical outlet of the proper voltage. You can determine the voltage requirements of your system by reading the label located on the back of the pump housing. The electrical supply circuit should be fuseWARNING The instrument must be connected to a grounded mains socket. protected, but the system will not work if the supply circuit is protected with a ground fault circuit interrupt. (A ground fault circuit interrupt is already included in the pump’s inverter.) Compressed air connections If you will be using an integrity tester to document the proper performance of your cassettes, cassette holder, and system piping, ensure you have a supply of instrumentquality compressed air. The air supply should include a pressure regulator to limit incoming air pressure to 3 bar g (45 psi g). Connect the instrument quality air to your integrity tester. Connect the air output from your integrity tester to the 3/8-inch hose barb located in the system’s feed line. Make the connection using 3/8-inch ID compressed air tubing and hose clamps. Tip: Expert users normally check their cross flow systems for leaks by pressurizing the system with compressed air and observing any loss in air pressure over time. Expert users also check the integrity of cassettes before using them by pressurizing the system and measuring the diffusion of air through the cassettes. If the diffusion rate meets known standards, then the user is assured that the cassettes do not contain pinholes and are securely mounted in the cassette holder. GE Healthcare recommends that you use or purchase an integrity tester and test your system and cassettes before use following the procedure described Cooling and heating water If you plan to use the cooling and heating jacket to control product temperature in the feed reservoir, you will need a system to supply or circulate cooling and heating fluid through the cooling jacket. The inlet and outlet ports for the cooling/heating jacket are located on the side of the reservoir. The inlet port is closest to the bottom of the reservoir, and the outlet port is located closer to the top of the reservoir. Attach the two 1/2-inch NPT-to-3/8-inch hose barb fittings supplied with the system to the inlet and outlet ports. Attach 3/8-inch ID heating/cooling lines to the hose barbs and secure with hose clamps. Set up your heating/cooling system to circulate fluid through the reservoir heating/ cooling jacket. See Figure 1-2 for location of connections. Kvick Lab Operating Instructions 28-9646-76 AA 27 3 3.5 Installation Installing a cassette holder 3.5 Installing a cassette holder Gather the following materials to install your GE Healthcare cassette holder into your cross-flow system: • A cross-flow system including a feed reservoir, feed pump, a minimum of two glycerinfilled pressure gauges, and three flow control valves • System piping and fittings to connect the feed pump, permeate collection vessel, and feed reservoir to the holder. The system piping must end in 1/2-inch sanitary clamps to connect to the Kvick Lab holder. • Three, fractional sanitary clamps and 1/2-inch gaskets Making the connections Place the Kvick Lab holder on the pump as shown in Figure 1-1. Stabilize and level the holder using the adjustable feet. Keep piping runs short to minimize holdup volume. Connect the holder’s feed, retentate, and permeate ports to your system’s piping using the three gaskets and sanitary clamp. To install a cassette in your holder, place the cassette and a gasket in the holder and clamp them in place using the tie rod spacers and nuts. You must tighten the nuts to torque of 20.3 Nm (180in-lbs) to ensure the proper clamping force is applied to seal the cassette in place. Final checks Before using the system, check the cassette holder to ensure it sits squarely on the pump housing and is not producing strain on the system piping. Finally, check the tightness of all sanitary clamps. 3.6 Installing cassettes in the holder To install a cassette in your holder, place the cassette and a gasket in the holder and clamp them in place using the tie rod spacers and nuts. You must tighten the nuts to a torque of 20.3 Nm to ensure the proper clamping force is applied to seal the cassette in place. Determining how many cassettes you can install The Kvick Lab holder can hold multiple cassettes depending upon the cassette type and the membrane area your application requires (Table 3-2). Table 3-2. Cassette installation guide for the Kvick Lab cassette holder. 28 Cassette type Cassette size No. of cassettes (install) Nut torque Kvick Lab packet 100 cm² (0.11 ft²) 1 to 10 13.5 Nm (120 in-lb) Kvick Lab Operating Instructions 28-9646-76 AA Installation 3 Installing cassettes in the holder 3.6 Cassette type Cassette size No. of cassettes (install) Nut torque Kvick Lab 0.11 m² (1.2 ft²) 1 to 5 20.3 Nm (180 in-lb) Installing the cassettes Follow these steps to install cassettes in Kvick Lab cassette holders: 1 Remove the tie rod nuts, washers, and spacers. Slide the back plate away from the flow distribution manifold (Figure 1-4). 2 To install one cassette: 3 a) Clean and wet a silicone gasket by rinsing it with DI water or water-forinjection (WFI). b) Place the gasket against the flow distribution manifold, aligning the holes in the gasket with the holes in the manifold. The gasket will stick to the manifold. c) Place the cassette into the holder with the gasket- side of the cassette facing the backing plate, and slide the cassette against the gasket. To install two or more cassettes: a) Complete the step above for installing one cassette. b) Place the second cassette into the holder between the first cassette and the backing plate. Ensure the gasket-side of the second cassette faces the ungasketed side of the first cassette. c) Place subsequent cassettes in the holder in a similar fashion. 4 Slide the backing plate forward to hold the cassettes in the hardware, ensuring the holes in the manifold, gasket, and cassettes line up. 5 Add spacers to the tie rods until about 0.75-inches (16mm) of thread is exposed on each tie rod. Reinstall the washers and nuts. Finger tighten the nuts evenly. 6 Tighten each nut alternately 1/4 turn at a time with the supplied torque wrench and socket until each nut is tightened to 20.3 Nm (180 in-lbs). If a seal is not achieved, there may be a cassette, gasket, or surface finish irregularity. Discontinue torque tightening and inspect the sealing surfaces. To learn how to use the torque wrench, consult the torque wrench user’s guide. Tip: The nut torque values are based on the tie rod threads being dry. You should not oil the tie rod threads. Oiled tie rod threads will result in too much clamping pressure being applied to the cassette. Tip: The nut torque values in the above procedure are based on the holder and cassette being at 18 to 23°C (64 to 74 °F). Lower or higher temperatures Kvick Lab Operating Instructions 28-9646-76 AA 29 3 3.6 Installation Installing cassettes in the holder experienced during processing and cleaning can change the clamping force applied to the cassette (due to expansion and contraction of the holder and cassette). Increases in temperature, in particular, can damage cassettes. Therefore, you should periodically check the nut torque. If the nut torque is too high, excessive clamping force can damage the cassette. If the nut torque is too low, insufficient clamping force can cause leaks or give false integrity test results. Before using a new Kvick Lab cassette, you must flush it with DI water or WFI to remove the storage solution. Many applications require that you complete several of the other preparatory stages listed below to ensure the cassette and the holder perform properly and consistently: Rinse Storage Solution from the Cassette—before using a new cassette—or a used one that you cleaned and stored—you must flush the storage solution from the cassette. Determine the Water Flux—you determine the water flux by measuring water flow through the cassette’s membrane under controlled process conditions. By measuring the water flow before and after cleaning, you can determine cleaning effectiveness and monitor the cassette’s performance. Test the Integrity of the Cassette—in this two-part process, you check the system and cassette for leaks or damage to the membrane. Sanitize the Cassette—when you require sanitary processing conditions, you can sanitize the cassette with sanitizing agents. Condition the System with Buffer— conditioning exposes your crossflow system’s wetted parts to an appropriate buffer before you introduce your product to the system. The conditioning minimizes unwanted chemical reactions between your product and the wetted parts. The buffer can also bring the system to the proper operating temperature before processing begins. Tip: 30 Expert users normally establish pump curves for the system configuration they plan to use. By establishing a pump curve, they can quickly and accurately set the pump speed to achieve the recommended flow rate during processing. To determine the flow rate without a pump curve, you would have to disconnect the retentate line and measure the flow rate using a graduated cylinder and stop watch. If you need help in developing a pump curve, contact GE Healthcare technical support team. Kvick Lab Operating Instructions 28-9646-76 AA Installation 3 Installing cassettes in the holder 3.6 Stage 1-Rinse storage solution from the cassette Opening the cassette bag safely WARNING The cassette and cassette bag contains an aqueous solution containing 0.1 to 0.2 N NaOH and 20 to 22% glycerin by weight. When opening the cassette bag, follow the standard safety procedures for handling aqueous NaOH, including the use of safety glasses, safety gloves, and protective lab coat. To avoid unexpected spilling, hold the cassette package upright over a sink and trim the top of the cassette bag with scissors. Drain and dispose of the excess storage solution in accordance with environmental regulations. To avoid inadvertent contact with the storage solution after the cassette is removed from the bag, rinse the outside surface of the cassette with high purity water. If you allow an ultrafiltration cassette to dry out, the membrane will be damaged. Therefore, do not store the cassette without rewetting it with an approved storage solution. At a minimum, before using a new or used-and-stored cassette, rinse the storage solution from the cassette following these steps: 1 Open the cassette bag and remove the cassette following the safety instructions above. 2 Install the cassette in the cassette holder. (See your cassette holder’s user guide for specific installation instructions.) 3 Configure the system so that the retentate and permeate lines discharge to waste (Figure 3-1). Close the feed, retentate, and permeate valves. 4 Fill the reservoir with an appropriate volume of clean water (6 liters per 0.11 m² [1.2 ft²] cassette). 5 Open the feed and retentate valves and close the permeate valve. Pump about 10 percent of the water through the retentate line to waste, maintaining a feed pressure of 0.35 barg (5 psig). NOTICE Using tap water to flush or rinse cassettes can plug the pores of membrane with the relatively large particulates (rust, dirt, minerals, bacteria, etc.) found in tap water. To prevent plugging the pores of the cassette, always use 0.2 µm filtered water or water-for-injection when rinsing or flushing the cassette, or when making up cleaning solutions or adding dilution water to process fluids. Kvick Lab Operating Instructions 28-9646-76 AA 31 3 3.6 Installation Installing cassettes in the holder 4 1 3 2 5 6 7 8 10 9 Figure 3-1. Typical setup for rinsing cassettes. Table 3-3. Parts in Figure 3-1. Part Description Part Description 1 Feed reservoir 6 Kvick Lab holder 2 Waste 7 Permeate 3 Valve 8 Pump 4 Pressure gauge 9 Drain line 5 Retentate 10 Feed 6 Open the permeate valve and close the retentate valve and pump the remaining water through the permeate line to waste, maintaining a feed pressure of 0.35 barg (5 psig). 7 If needed for your application, sample and analyze the last volume of drain water pumped from the system for traces of storage or cleaning solution. Repeat steps 4 through 7 until you achieve the needed level of storage solution removal. 8 If needed, circulate buffer to prepare the cassette and system for processing. Stage 2—Measuring water flux Measuring water flux involves measuring the flow of water through the cassette’s membrane under controlled conditions. You convert the measurement into units that represent water flux. The water flux value you obtain provides an indicator of the performance of the cassette. GE Healthcare recommends that you measure water flux when the cassette is new and before each use. By tracking the water flux measurements, you can; (1) 32 Kvick Lab Operating Instructions 28-9646-76 AA Installation 3 Installing cassettes in the holder 3.6 determine the effectiveness of cleaning cycles, and (2) determine when cassette reaches the end of its service life. Measuring water flux manually requires about 30 minutes and this equipment: graduated cylinder, stopwatch, thermometer, calculator, pen and paper. If you have a flow meter installed on the permeate line, you can use the flow meter for measuring flux. Summary of water flux measurement procedure The following steps summarize how to measure water flux: 1 Measure the water flow from the permeate port at two transmembrane pressures. 2 Calculate water flux in l/m²/hr (LMH) for the two water flow measurements. 3 Plot the two flux values. Then interpolate Kvick Lab cassette water flux at 0.7 barg (10 psig) transmembrane pressure. 4 Normalize the water flux value to 20°C and 1 psig to arrive at flux in LMH/psig @ 20°C. 5 Record the flux value as a benchmark for future comparison. Compare the flux with previously obtained values. The following text describes how to complete each step. Measure flow from the permeate port Follow these steps to measure the water flow rate from the permeate port: 1 Rinse the cassette following the instructions in Stage 1—Rinse the Kvick Lab cassette. 2 Setup your crossflow system to circulate the retentate and permeate back to the feed reservoir (Figure 3-2). 3 Fill the feed reservoir with a volume of deionized water or WFI. Fully open the feed, retentate, and permeate valves. 4 Start the feed pump and adjust it and the retentate valve to obtain a transmembrane pressure of 0.3 barg (5 psig). Using a graduated cylinder and stop watch, measure and record the permeate flow rate in ml/min. Record the temperature of the water in the feed reservoir. 5 Adjust the feed pump and retentate valve to produce a transmembrane pressure of 1 barg (15 psig). Using a graduated cylinder and stop watch, measure and record the permeate flow rate in ml/min. Record the temperature of the water in the feed reservoir. 6 Drain the system. Kvick Lab Operating Instructions 28-9646-76 AA 33 3 3.6 Installation Installing cassettes in the holder 3 1 2 4 6 12 5 9 7 10 8 11 Figure 3-2. Typical configuration for measuring water flux. Table 3-4. Parts in Figure 3-2. Part Description 1 Feed reservoir 2 Valve 3 Pressure gauge 4 Retentate 5 Kvick Lab holder 6 Permeate 7 Pump 8 Drain line 9 Feed 10 Optional flow meter 11 Graduated cylinder 12 Pressure gauge Calculate flux in l/m²/hr Convert the permeate flow rate (l/min.) measurements to flux using this equation: 34 Kvick Lab Operating Instructions 28-9646-76 AA Installation 3 Installing cassettes in the holder 3.6 Flux in LMH (l/m²/hr) = permeate flow in ml/min ÷ cassette surface area in square meters × 60 Plot flux and determine flux at 0.7 barg Plot the calculated flux values against the transmembrane pressure (Figure 3-3). Determine the water flux in LMH at 0.7 barg (10psig). In Figure 3-3, the flux at 0.7 barg (10 psig) is 115 LMH. 200 F l ux ( L M H ) 150 100 50 5 10 15 20 25 Transmembrane Pressure (PSIG) Figure 3-3. Plot of water flux versus transmembrane pressure. Normalize water flux to 20°C and 1 psig Follow these steps to normalize the water flux value to 20°C and standardize to 1 psig to arrive at flux in LMH/psig at 20°C: Multiply the calculated flux value by the appropriate viscosity correction factor in Table For example, if the flux at 0.7 barg (10 psig) is 115 LMH, and the temperature of the water during permeate flow measurement was 18°C, then the flux normalized to 20°C is: 115 LMH x 1.053 = 121 LMH/10 psig To standardize the flux value to 1 psig, divide the normalized water flux value by 10 psig. For example, 121 LMH ÷ 10 = 12.1 LMH/psig Record the flux value Record the normalized flux value in your laboratory notebook or production records as a benchmark for future comparison. If the cassette is used, compare the flux with previously obtained values to determine the performance of the cassette over time. Many users replace cassettes when, after cleaning, the normalized water flux falls below 60 to 80 percent of the normalized flux value when the cassette was new. If after cleaning, the flux is low, you can try alternative, aggressive cleaning processes. Contact GE Healthcare for more information. Kvick Lab Operating Instructions 28-9646-76 AA 35 3 3.6 Installation Installing cassettes in the holder Table 3-5. Viscosity correction factors. Temperature in ºC when permeate flow was measured Viscosity correction factor Temperature in ºC when permeate flow was measured Viscosity correction factor 4 1.567 25 0.890 5 1.519 26 0.871 6 1.472 27 0.851 7 1.428 28 0.833 8 1.386 29 0.815 9 1.346 30 0.798 10 1.307 31 0.781 11 1.271 32 0.765 12 1.235 33 0.749 13 1.202 34 0.734 14 1.169 35 0.719 15 1.139 36 0.705 16 1.109 37 0.692 17 1.081 38 0.678 18 1.053 39 0.665 19 1.027 40 0.653 20 1.000 41 0.641 21 0.978 42 0.629 22 0.955 43 0.618 23 0.933 44 0.607 24 0.911 45 0.597 Stage 3—Test the integrity of the cassette Before using your cross-flow system and cassette to process product, test the crossflow system to ensure it does not leak. Then test the integrity of the cassette. 36 Kvick Lab Operating Instructions 28-9646-76 AA Installation 3 Installing cassettes in the holder 3.6 To test your cross-flow system for leaks, pressurize it with water and check the tubing or piping connections for leaks. To test the integrity of the cassette, pressurize the system with compressed air and measure the diffusional flow of air through the cassette membrane. A high flow rate (above the cassette’s specification) indicates there might be a hole or tear in the membrane. Follow the instructions below to test the system for leaks, and to test the integrity of the cassette. Test your system for leaks To test your cross-flow system for leaks, fill the system with clean water. Open the feed valve. Open the retentate valve. Close the permeate valve. Start the feed pump on slow speed and throttle the retentate valve until the feed pressure reads 3 barg (45 psig). Check all connections for leaks. Test the integrity of the cassette To measure diffusional air flow through the membrane via the permeate port, you will need to add a compressed air line, pressure regulator, and cut-off valve to your crossflow system. You also need the following equipment: a 500-ml beaker, 50-ml graduated cylinder, stopwatch, and a piece of flexible tubing (Figure 3-4). Test the integrity of the system and cassette manually by following the procedure described below. 2 1 3 4 5 6 7 8 10 11 9 12 13 14 15 16 Figure 3-4. Flow diagram for testing a crossflow system and cassette before use. Kvick Lab Operating Instructions 28-9646-76 AA 37 3 3.6 Installation Installing cassettes in the holder Table 3-6. Parts in Figure 3-5. Part Description Part Description 1 Flexible tubing connected from permeate line and inserted into mouth of inverted graduated cylinder 2 Inverted graduated cylinder filled with water 3 Beaker filled with water 4 Feed reservoir 5 Waste 6 Valve 7 Pressure gauge 8 Retentate 9 Kvick Lab cassette 10 Permeate 11 Pump 12 Feed 13 Drain line 14 Air shutoff valve 15 Pressure regulator 16 Instrument quality compressed air Check the integrity of the cassette When you establish that the system is leak-free, test the integrity of the Kvick Lab cassette by following these steps. 38 1 Ensure the compressed air valve is closed and the pressure regulator is set to 0 barg (0 psig). Close the feed and drain valves, and open the retentate and permeate valves. 2 Remove water from the cassette feed-retentate path by opening the compressed air valve, and slowly adjusting the pressure regulator to 0.3 barg (5 psig). Flow air through the system until water stops draining from the retentate line. Close the retentate valve to allow the air pressure to drive the water from the permeate port and permeate line. 3 Slowly adjust the pressure regulator to 1 barg (15 psig). 4 Fill a 50-ml graduated cylinder with water and place it inverted in a 500-ml beaker half filled with water. Connect flexible tubing to the permeate outlet (Figure 3-4). 5 When the air bubble rate is stable, note the time and the standing air volume in the graduated cylinder. When 5 to 10 ml of air has collected in the graduate cylinder, record the time and air volume again. 6 Using the time and air volume data, calculate the air diffusion rate in ml/min/ft² @ 1 barg (15 psig). 7 Compare your results with the specifications below. If your measured flow rate is less than the specification, the Kvick Lab cassette is acceptable for use. The Kvick Lab Operating Instructions 28-9646-76 AA Installation 3 Installing cassettes in the holder 3.6 specifications are: 10 ml/min/Kvick Lab cassette at 1 barg for 10K and 10KS cassettes. 8 When the Kvick Lab cassette passes integrity testing, you can sanitize it (Stage 4), condition it with buffer (Stage 5), or process product as required. 9 If the Kvick Lab cassette fails the integrity test, contact your GE Healthcare representative for help. Stage 4- Sanitizing the Kvick Lab cassette Clean and rinse the Kvick Lab cassette, then follow these steps to sanitize the cassette: 1 Configure the system so that the retentate and permeate lines discharge into the feed reservoir (Figure 3-5). Close the drain valve. 2 Fill the feed reservoir with 1.5 to 2 liters of 0.1 N sodium hydroxide for each cassette in your system. The temperature of the solution should be 35 to 45°C. 3 Open the feed, retentate, and permeate valves. Start the pump. Adjust the retentate valve and pump speed to obtain a retentate pressure of 0.7 barg (10 psig) and the following retentate flow rate: 875 ml/min for each 0.11 m² (1.2 ft²) Kvick Lab cassette installed. 4 Circulate the sanitizing solution for minimum of 60 minutes at a temperature of 35 to 45°C. Stop the pump. If you are going to store the Kvick Lab cassette, rinse and add storage solution as described on page 52. Kvick Lab Operating Instructions 28-9646-76 AA 39 3 3.6 Installation Installing cassettes in the holder 5 If you are sanitizing the Kvick Lab cassette before use, drain the sanitizing solution from the system and then rinse the Kvick Lab cassette following the rinsing procedure on page 31. 1 3 2 4 5 7 6 8 9 Figure 3-5. Typical setup for sanitizing a Kvick Lab cassette and system. Table 3-7. Parts in Figure 3-5. Part Description Part Description 1 Feed reservoir 2 Valve 3 Pressure gauge 4 Retentate 5 Kvick Lab cassette 6 Pump 7 Permeate 8 Feed 9 Drain line - - Stage 5—Conditioning the system with buffer Before processing your sample, precondition the cross-flow system with a buffer similar in pH and ionic strength to that of your sample. Conditioning the system removes trapped air and minimizes unwanted chemical reactions between your sample and the wetted parts of the cross-flow system. You can also use buffer conditioning to stabilize the temperature of the cross-flow system. Follow these steps to condition the system with buffer: 40 Kvick Lab Operating Instructions 28-9646-76 AA Installation 3 Installing cassettes in the holder 3.6 1 Setup your cross-flow system to circulate the retentate and permeate back to the feed reservoir (Figure 3-6). 2 Prepare the buffer solution, and if necessary, condition it to the proper temperature. The recommended volume of buffer solution is 1 liter per cassette. 3 Put the buffer in the feed reservoir. Open the feed, retentate, and permeate valves. Start the feed pump and increase the feed rate until solution flows from the retentate and permeate lines. 4 Close the retentate valve to produce about 0.3 to 1 barg (5 to 15 psig) of retentate pressure. Run until no bubbles appear in the permeate stream. 5 Open the retentate valve and close the permeate valve. Increase the retentate flow rate to the recommended operating cross-flow rate: 600 to 800 ml/min per 0.11m² (1.2-ft²) Kvick Lab cassette. Run until no bubbles appear in the retentate stream. 6 Open the permeate valve. Adjust the retentate valve until the permeate flow is about 10 percent of the feed flow. Adjust the feed pump to maintain a feed pressure of 1.6 to 2.8 barg (25 to 40 psig) for ultrafiltration cassettes. 7 Circulate the buffer for four minutes to condition the system for pH and ionic stability. If conditioning for temperature control, continue circulating until the temperature of the system stabilizes. 8 Remove the buffer from the feed reservoir. Keep buffer in other parts of the system to prevent air from entering the system. 3 1 2 4 7 6 5 8 Figure 3-6. Typical configuration for conditioning the system with buffer. Kvick Lab Operating Instructions 28-9646-76 AA 41 3 3.7 Installation Spare parts and accessories Table 3-8. Parts in Figure 3-6. Part Description Part Description 1 Feed reservoir 2 Valve 3 Pressure gauge 4 Retentate 5 Kvick Lab cassette 6 Pump 7 Permeate 8 Feed 3.7 Spare parts and accessories For correct up to date information on spare parts and accessories, visit: www.gelifesciences.com/filtration 42 Kvick Lab Operating Instructions 28-9646-76 AA Operation 4 Operation overview 4.1 4 Operation 4.1 Operation overview The actual operation of the system will vary considerably depending upon your process and product objectives. If you need assistance with application information, contact your GE Healthcare representative. Equipment configuration and process conditions for cross-flow processes vary widely depending upon the application, system design, process objectives, and product objectives. However, many cross-flow processes share three main stages: • Startup • Steady state operation • Shut down and product recovery 4.2 Starting the instrument NOTICE If you operate the pump dry (without any process fluid in it), you can damage the pump. To prevent damage to the pump, ensure that fluid is present in the reservoir and feed line whenever you run the pump. Before operating the pump, you should read the pump manufacturer’s user manual that was supplied with the Kvick Lab system. 4.3 The pump control system To operate the pump, press the UP ARROW or DOWN ARROW until the desired pump RPM is displayed on the control panel display. Then press the FWD button to start the Kvick Lab Operating Instructions 28-9646-76 AA 43 4 4.3 Operation The pump control system pump. You can control other pump operations by using the buttons on the control panel. Figure 4-1. Pump control panel buttons and display. Table 4-1. Description of Figure 4-1. 44 Key Description FWD Starts the pump running in the forward direction, pumping feed solution to the cassette holder. If the pump is running in reverse when the FWD button is pressed, the pump will slow to a stop and then speed up in the forward direction. REV Starts the pump running in the reverse direction. If the pump is running forward when the REV button is pressed, the pump will slow to a stop and then speed up in the reverse direction. STOP Slows the pump to a stop UP ARROW If the pump is running, pressing this button will increase the speed of the pump to that shown on the display. If the pump is stopped, pressing this button increases and displays the RPM speed that the pump will obtain when you start it. DOWN ARROW If the pump is running, pressing this button will decrease the speed of the pump to that shown on the display. If the pump is stopped, pressing this button lowers and displays the speed that the pump will obtain when you start it. ENTER Pressing this button at anytime will store the RPM shown on the display as the default RPM, which will be displayed whenever the system is powered up. SHIFT Used by authorized personnel for programming. Kvick Lab Operating Instructions 28-9646-76 AA Operation 4 The pump control system 4.3 Key Description PROG This enables those personnel with proper authorization to enter the programming menus and program the operation of the pump. STATUS light The STATUS light on the control panel glows red when the pump is powered up and not rotating. The STATUS light glows green when the pump is running. Process flow in a cross-flow system A typical system for many cross-flow applications includes two or three pressure gauges (for precise process control and safety), feed tank, permeate collection vessel, and a drain line (Figure 1-3). Recommended operating conditions When using your Kvick Lab cassette, fully open the permeate valve and adjust your pump speed and retentate valve to achieve the recommended cross-flow rate. The recommended operating conditions for many applications are: Table 4-2. Recommended operating conditions. Characteristic Parameter Transmembrane pressure 0.5 to 2.4 barg (25 to 35 psig) Cross-flow rate 600 to 800 ml/min per 0.11 m³ (1.2 ft³) cassette pH range, long-term (operating or storage) 2 to 13 pH range, short-term (cleaning) 1 to 14 Maximum operating temperature 50 °C Maximum inlet pressure 4 barg (60 psig) If your feed is particularly viscous or particulated or has other unusual physical characteristics, contact the technical support team at GE Healthcare for assistance. Stage 1—Starting the cross flow process Follow these steps to start a typical cross-flow process: Kvick Lab Operating Instructions 28-9646-76 AA 45 4 4.3 Operation The pump control system 1 Drain residual water from the system and equilibrate the system with water or buffer. 2 Close the permeate valve; open the feed and retentate valves; and close the drain valve (if equipped). 3 Direct the permeate line to an appropriate container or drain; for example, a permeate container, waste container, or process drain. 4 Fill the feed tank with product. Start the feed pump and slowly circulate the product through the system for about three to four minutes. Recirculation helps purge trapped air from the flow channels, maximizing membrane performance. 5 Open the permeate valve. Slowly increase the pump speed until the optimal crossflow rate is achieved. Using the retentate valve, adjust the retentate back-pressure until the system reaches the optimum transmembrane pressure. Tip: If you do not know the relationship between your pump speed setting and the feed flow rate, it may be difficult to consistently achieve the process conditions you desire. Therefore, to ensure consistent performance and ease of use, prepare a pump curve on your feed pump before processing so that you know the relationship between pump speed and flow rate. 6 Monitor the product level in the feed tank to ensure the pump does not run dry. As the level drops, you can add additional product to the feed tank. 7 If needed, transfer diafiltration solution into the feed tank at a controlled rate to maintain a working volume in the feed tank. (The required number of diafiltration volume exchanges varies greatly with respect to the particular characteristics of a product.) Typically, in crossflow processes, six volume exchanges of buffer are commonly processed through the feed tank. 8 Continue processing following the steps below in the section entitled, Stage 2— Operating the system in a steady state. Stage 2—Operating the system in a steady state During steady state operation, you monitor the process and record the process information that you need for product development, process development, process validation, or quality control. Key operational data includes these: 46 • Transmembrane pressure • Feed pressure • Retentate pressure • Product temperature • Feed volume (initial and final) Kvick Lab Operating Instructions 28-9646-76 AA Operation 4 The pump control system 4.3 • Diafiltrate volume • Permeate flow rate • Permeate volume • Retentate cross-flow rate • Various product quality analyses Stage 3—Shutting the system down and recovering product Before shutting your cross-flow filtration system down, you can complete two recovery methods to maximize the recovery of product. One recovery method helps remove the buildup of product on the surface of the membrane inside the cassette. The other method helps you recover the most product volume from the system piping and cassette. Capturing product from the membrane surface The first product recovery method involves reducing the cross-flow rate at the end of your processing step. For example, when you finish processing your product, reduce the cross-flow rate to 1/10 of the recommended processing cross-flow rate. Close the permeate valve or reduce the feed pressure to 0.3 barg (5 psig). Retentate pressure should be 0 barg (0 psig). Circulate the remaining product for 15 minutes. This procedure will help recover product buildup from the surface of the membrane. Maximizing the volume of product recovered The second product recovery method maximizes the volume of product you can recover from the system piping and cassette. The method involves slightly over concentrating the product, then returning a small volume of permeate back into the system to flush concentrated product from the feed-retentate loop. This method is described below in the following steps: 1 As the process nears completion, decrease the pump speed to minimize flow rate, vortexing in the feed tank, and the possibility for product foaming. 2 When you reach the slightly over concentrated volume, collect about 100 to 500 ml of permeate depending upon the installed membrane area and nature of the product) in a container. (Typical volumes of permeate are about 100 to 500 ml for lab/pilot systems or considerable more for pilot/production systems.) Stop the pump and close the permeate valve. 3 Attach a line to the drain valve and connect the other end aseptically to your collection vessel (Figure 4-2). Tip: If maximum recovery is required, you can remove product from the system by applying a low-pressure nitrogen flow to the retentate side of the system. Contact your GE Healthcare’ representative for additional details. Kvick Lab Operating Instructions 28-9646-76 AA 47 4 4.3 Operation The pump control system NOTICE Many types of pumps will be damaged if you run them dry. Therefore, monitor the liquid level in the tank and shut off the pump when the tank is empty. 4 Open the drain valve, and with the retentate valve open, start-up the pump at a slow speed. Apply back-pressure by slightly closing the retentate valve if needed to assist in removal of product from the drain line. Stop the pump as the tank fully drains. 5 Close the drain valve. Add a portion of saved permeate to the feed tank and circulate for two to three minutes. Open the drain valve and circulate (and apply back-pressure by slightly closing the retentate valve if needed) to drain the product from the feed-retentate loop. Stop the pump. Repeat step 5 as needed to recover the maximum volume of product. 2 1 9 3 8 10 4 5 6 12 11 7 13 Figure 4-2. Typical flow path for shutting down and recovering product. Table 4-3. Description of flow path in Figure 4-2. 1 48 Retentate Kvick Lab Operating Instructions 28-9646-76 AA Operation 4 The pump control system 4.3 2 After recovering a product from the drain line, add a portion of the collected permeate, circulate, and drain into your product collection vessel. 3 Feed reservoir 4 Feed 5 Pump 6 Drain line 7 Collection 8 Valve 9 Pressure gauge 10 Kvick Lab holder 11 Collection 12 Permeate 13 As you near the end of processing collect 200 to 500 ml of permeate for flushing the system. Kvick Lab Operating Instructions 28-9646-76 AA 49 4 4.3 50 Operation The pump control system Kvick Lab Operating Instructions 28-9646-76 AA Maintenance 5 General 5.1 5 Maintenance 5.1 General The system should be flushed after each use and before storage. Normally, when you flush the cassette and holder after use, you simultaneously flush and clean the Kvick Lab system, and no further flushing is needed. 5.2 User maintenance schedule Inspection Periodically—based on usage—you should make these inspections: • Inspect and perform maintenance on the Kvick Lab cassette holder as described on 54. • Inspect and perform maintenance on the rotary lobe pump as described in the rotary lobe pump user manual. • Inspect piping connections for tightness and worn gaskets after each run. 5.3 Cleaning If the last operation you perform with the system is to flush the cassette with a storage solution, you must then flush the system and holder with deionized (DI) water or waterfor-injection (WFI) to remove the storage solution. Use an old cassette in the holder to flush the system with water. You can also cut the center out of two gaskets and install the gaskets in the holder for flushing (Figure 5-1). After use, you should wipe down the exterior components of the Kvick Lab system with DI water or WFI to remove dirt and residual process/cleaning solutions. Wipe the holder dry with a clean, lint-free cloth. Figure 5-1. Gasket with center cut out for flushing a holder with water. Kvick Lab Operating Instructions 28-9646-76 AA 51 5 5.4 Maintenance Cleaning and storing cassettes Storing the system To store the system for more than a few days, clean it as described in the maintenance section. Drain all rinse water from the system. Remove the cassettes from Kvick Lab holder. Blank off open lines or ports in the system and reservoir. Store the cassette in accordance with the instructions in the cassette user manual. 5.4 Cleaning and storing cassettes Introduction You can reuse cassettes if you clean and store them properly. The cleaning and storage stages include these: Cleaning the Cassette Rinsing the cassettes with water prior to cleaning ensures the most effective use of the cleaning solution and prevents solubility changes that can prevent effective cleaning. Flushing with the cleaning solution dissolves and physically removes contaminates from the cassette. Flushing with water removes the cleaning solution from the cassette. Measure the Water Flux You measure the water flux and compare it to the water flux when the cassette was new. You can then determine cleaning effectiveness. Store the Cassette and Holder Storing the cassette and holder involves adding a storage solution and preparing the system for short-term or long-term storage. The goal is to keep the membrane wet and to prevent microbial growth during storage. Stage 1—Cleaning a cassette after use You can reuse a cassette if you clean and store it properly. To clean a used cassette, flush it with buffer, WFI, or DI water as described in Stage 1—Rinse storage solution from the cassette on page 31. Then clean the cassette following these steps: 52 1 Configure the system so that the retentate and permeate lines discharge to waste (Figure 3-1). Close the feed, retentate, and permeate valves. 2 Prepare the appropriate volume of a cleaning solution (1.5 to 2 liters per 0.11 m² (1.2 ft ²) Kvick Lab cassette. Table 5-1 describes recommended cleaning solutions. 3 Fill the feed reservoir with the cleaning solution. Kvick Lab Operating Instructions 28-9646-76 AA Maintenance 5 Cleaning and storing cassettes 5.4 4 Open the feed and retentate valves and pump about 10 percent of the cleaning solution through the retentate line to waste. Open the permeate valve and close the retentate valve and pump about 10 percent of the cleaning solution through permeate line to waste. 5 Stop the pump and configure the system so that the retentate and permeate lines discharge into the feed reservoir. Fully open the feed and retentate valves. Crack open the permeate valve. 6 Start the pump. Adjust the retentate valve and pump speed to obtain 1 to 1.5 times your processing cross-flow rate. 7 Circulate the cleaning solution for 30 to 60 minutes. 8 Drain the cleaning solution from the system using the drain line. 9 Rinse the cassette and system with clean water as described in Stage 1—Rinse storage solution from the cassette on page 31. Table 5-1. Cleaning agents and conditions. Cleaning agent Cleaning conditions 0.5 M NaOH 30 minutes at room temperature 1 M NaOH 2 hours at 50°C 0.5 M NaOH with 300 ppm NaOCl 30 minutes at room temperature 0.5 M NaOH then 0.5 M H2SO4 30 minutes at room temperature for each solution Double cleaning with 0.5 M NaOH with 300 ppm NaOCl 30 minutes at room temperature for each cycle Stage 2—Measuring water flux To determine the effectiveness of cleaning and the end of the service life of the cassette, you should measure the water flux after cleaning. Compare the most recent water flux measurement with previous flux measurements. To determine the cassette’s water flux, follow the procedure described in this user guide. However, instead of measuring the flux at the two recommended pressures, measure the flux flow at the target pressure for your type of cassette—0.7 barg (10 psig) for ultrafiltration cassettes—and determine the cassette’s water flux normalized to a water temperature of 20°C. Interpreting water flux values Compare the post-cleaning flux value with the as new flux values. If the post-cleaning flux values are less than 60 to 80 percent of the new flux values, you may have to repeat or change the cleaning procedure. If, after all cleaning attempts, the post-cleaning flux Kvick Lab Operating Instructions 28-9646-76 AA 53 5 5.5 Maintenance Maintenance of the holder values are still less than 60 to 80 percent of the new flux values, you may want to replace the cassette. Stage 3—Storing the cassette and system Short-term storage After completing stages 1 and 2, flush the cassette with a storage agent as described below to keep the membrane wet and minimize biological growth: • For storage less than 3 days, use 0.2 µm filtered water • For storage less than 6 months, use 0.1N NaOH • For storage longer than 6 months, use a storage solution of 0.1N NaOH and store the cassette at 4°C. For long-term storage (more than ten days), remove the cassette from the holder and place it into an air-tight plastic tub or heavy duty, zip-lock-type plastic bag. Add about 50 to 100 ml of the storage agent to the plastic bag and seal it. Place the sealed bag in a protected location at a temperature of 4°C. For short-term storage (five to ten days), keep the cassette in the holder and circulate the storage solution for 10 to 15 minutes, close the system valves, disconnect the electrical power to the feed pump, and ensure the feed reservoir is properly sealed. On manually-clamped holders, loosen the tie rod nuts alternately turn at a time until you can just turn the nuts by hand. Tighten the nuts to a torque of 10 to 15 Nm (88 to 133 in-lbs). On hydraulic systems, keep the cassettes and gasket in the holder and pressurize the hydraulic system to 35 to 69 barg (500 to 1000 psig) to prevent fluid from leaking from the cassette holder. 5.5 Maintenance of the holder WARNING Inspecting the cassette holder without shutting down the cross-flow system to which it is attached (disconnecting the power cable, disconnecting the compressed air supply, and decompressing the system) can result in bodily injury from the accidental starting of the pump or unexpected release of pressure.To prevent bodily injury, disconnect the power cable, disconnect the compressed air line, and depressurize the system before performing inspections and maintenance on the system. 54 Kvick Lab Operating Instructions 28-9646-76 AA Maintenance 5 Replacement of fuses 5.6 Inspection Periodically—based on usage—you should make these inspections: • Inspect the tie rods and nuts for excessive wear and replace as needed. • Inspect the mating surfaces of the flow manifold and backing plate for scratches. • Inspect the sanitary connection gaskets for damage or wear and replace the gaskets as needed. • Check the rigidity of the stand. If the stand wobbles, tighten the screws that hold the legs and cross braces together. If the mating surfaces of the flow manifold and backing plate become excessively scratched, contact your GE Healthcare representative for assistance. Maintenance After using the holder, you should wash down the exterior components with clean water to remove dirt and residual process/cleaning solutions. Blow the wash water out of the flow manifold channels with compressed air. Wipe the holder dry with a clean, lint free cloth. Tip: The nut torque values are based on the tie rod threads being dry. You should not oil the tie rod threads. Oiled tie rod threads will result in too much clamping pressure being applied to the cassette. Storing the holder To store the holder, clean it as noted above and place several layers of a clean cloth between the flow manifold and the backing plate. Install the spacers, washers, and nuts, and lightly hand tighten to hold the cloth and backing plate in place. Place the holder in a dry, protected area such as a cabinet to prevent contamination or accidental dropping. 5.6 Replacement of fuses The 115- and 230-volt versions of the pump use one fuse located in the power switch to protect the pump from damage. Follow these steps to replace the fuse: 1 Disconnect the pump and cross flow system from power sources. Unplug the power cord from the pump’s electrical supply outlet and the power switch module on the back of the pump housing. 2 Insert a small screwdriver into the slot on the top of the power switch module and gently pry the power switch module cover outward from cabinet. 3 Insert a small screwdriver into the slot at the top of the red fuse holder and gently pry the fuse holder from the power switch module. As you pull the fuse holder from Kvick Lab Operating Instructions 28-9646-76 AA 55 5 5.6 Maintenance Replacement of fuses the power switch module, one fuse and one dummy fuse will pop out from the sides of the fuse holder. 4 Replace the old fuse with a new fuse—250 volt, 10 amp, fast blow, ABC10A type, 31.75 mm × 6.35 mm (1.2 in × 0.25 in). 5 Orient the fuse holder so that the proper voltage designation faces up. Place and hold the new fuse and the dummy fuse in the fuse holder. Insert and press the fuse holder into the power switch module until it clicks into place. 6 Reposition the power switch cover over the fuse holder and press the cover against the fuse holder until it clicks into place. Ensure the correct voltage designation (115 VAC or 230 VAC) shows through the window in the power switch module. The voltage that shows in the window should match the voltage on the pump label on the back of the pump cabinet. WARNING Make sure that the voltage displayed on the power module is the same as the voltage available from the power outlet. The wrong voltage setting may lead to electrical and fire hazards. System disposal When the service life of the unit is reached, flush the system to remove hazardous process fluids. Dispose of the unit in accordance with the government and environmental regulations that pertain to your location. Replacing the power cord If the power cord becomes worn or damaged, always replace it with an identical power cord. 56 Kvick Lab Operating Instructions 28-9646-76 AA Troubleshooting 6 6 Troubleshooting Consult Table 6-1 if the Kvick Lab system does not work properly. Table 6-1. Troubleshooting chart. Symptom Possible cause Suggested remedy Leaks at the tubing nipple at permeate connection of cartridge. Inside dimension of tubing is too large. Ensure tubing i.d. is sized properly for the connection. Split tubing at barb Check tubing for wear or splitting. Replace tubing if necessary. Tubing clamp incorrectly installed. Ensure the clamp is installed and tight. Process solution leaks from sanitary connections. Clamp is loose. Check and tighten clamp. Gasket worn, damaged, improperly mounted or wrong size. Inspect gasket for wear, damage and proper sizing. Replace if necessary. Feed pressure increases from starting values slowly at first, then suddenly very rapidly, exceeding process limit. Blockage or partial blockage of flow path. Flush through to clean blockage. If necessary replace tubing and connectors. Please refer to order information for tubing and connectors. Feed inlet at cartridge is blocked. Switch feed and retentate lines to reverse flow and flush with deionized water (DI) or water for injection (WFI). Fouled cartridge. For feed flow rate versus pressure drop and cleaning of cartridge refer to Operating Handbook - Hollow fiber cartridges for membrane separations. Retentate valve closed or restricted. Open valve. Speed of pump (RPM) is too high. Reduce pump speed. No or low retentate pressure. Kvick Lab Operating Instructions 28-9646-76 AA 57 6 Troubleshooting Symptom Possible cause Suggested remedy Feed pressure is too low. Pump head is not turning. Check for error messages on pump display, check voltage setting is correct on power supply and power is delivered to the motor. Air leaking from cartridge or housing during cartridge integrity testing (Cartridge integrity testing employs air on the retentate side of a fully wetted cartridge). Cartridge not wetted properly. Cartridge not been wetted properly. Please read the chapter New cartridge rinsing procedure in the Operating Handbook - Hollow fiber cartridges for membrane separations. For pre-treatment with alcohol please read the document Hollow fiber cartridge preparation instructions. Membrane is not integral. Please contact your local GE Healthcare technical support representative. Faulty diaphragm valve. Get valve serviced by GE Healthcare service engineer or replaced. Gasket worn, damaged, improperly mounted or wrong size. Inspect gasket for wear, damage and proper sizing. Replace if necessary. Leakage from process connectors. Check and tighten connectors. Water leaking during system integrity testing (System integrity testing employs water on the retentate side). 58 Kvick Lab Operating Instructions 28-9646-76 AA Troubleshooting 6 Symptom Possible cause Suggested remedy Permeate flow is not adequate (microfiltration membrane). Air trapped in the pore structure of the membrane. Air-locked membrane, in which case, water layers trap a pocket of air between inner and outer surfaces of the membrane; effectively reducing the available surface area. This may be observed following autoclaving. Remedy: recirculate cold water (5-15C) on the retentate side with the permeate side closed, very gradually increase retentate back-pressure until a stable 20 psig is reached and hold this condition for 10 minutes or until expected permeate flux is reached Note that this does not affect feed pressure or TMP; due to the permeate side being closed, but only raises the absolute pressure on both sides of the membrane; causing the trapped air to diffuse rapidly into the water. Kvick Lab Operating Instructions 28-9646-76 AA 59 6 Troubleshooting Symptom Possible cause Suggested remedy Permeate flow is not adequate (ultrafiltration membrane). Air trapped in the pore structure of the membrane, or glycerol humectant not properly removed. Cartridge not wetted properly. Please read the chapter New cartridge rinsing procedure in the Operating Handbook Hollow fiber cartridges for membrane separations. For pre-treatment with alcohol please read the document Hollow fiber cartridge preparation instructions. Auxiliary equipment such as manometers and pumps not working properly. Check the function of all auxiliary equipment. Repair/ replace if necessary. Bent tubing. Check that the flow path is not restricted. Liquid viscosity too high. Check the viscosity of the liquid. Viscosity is a function of temperature. (Lower temperature gives higher viscosity.) Let low-temperature liquid reach operating temperature before starting the run. Microbial growth in buffers. The buffer normally becomes opalescent due to microbial growth. Check buffers, especially those with phosphate, for microbial growth. Replace with fresh buffer if necessary. Dimension of tubing is incorrect. Ensure tubing i.d. is sized properly for the connection. Fouled cartridge. Clean the cartridge according to instructions. Split tubing at barb Check tubing for wear or splitting. Replace tubing if necessary. Tubing clamp incorrectly installed. Ensure the clamp is installed and tight. Permeate pressure is dropping. 60 Kvick Lab Operating Instructions 28-9646-76 AA Troubleshooting 6 Symptom Possible cause Suggested remedy Unexpected passage of target protein to the permeate side. Membrane pore size is too large for target molecule. Select the largest pore size that retains the target molecule (in general choose a membrane 35× smaller than the target protein). For example, either 30,000 or 50,000 NMWC ultrafiltration membranes are recommended for mAb concentration and diafiltration. Unexpected retention of target molecule. Membrane pore size is too small for target molecule. Select the smallest pore size that still allows efficient passage of the target molecule (in general 10 x larger opening than the target protein) For most mammalian cell culture clarification 0.2 to 0.65 micron pore sizes are used. Kvick Lab Operating Instructions 28-9646-76 AA 61 6 62 Troubleshooting Kvick Lab Operating Instructions 28-9646-76 AA Reference information 7 Kvik Lab system specifications 7.1 7 Reference information 7.1 Kvik Lab system specifications Size and dimensions Parameter Specification Dimensions, approximate (W×L×H) 38×61×53 cm (15×24×21 inches) Weight, approximate 68 kg (150 lbs) Materials of construction Parameter Specification Wetted parts, excluding cassette 316L stainless steel, EPDM gaskets and diaphragms Operating parameters Parameter Specification Recommended operating temperature 5 to 50°C Maximum temperature, steaming/ autoclaving (no cassettes) 121°C (250°F), not including pump or pressure gauges in autoclave. Maximum temperature, operating (pump rotor limited) 60°C (140°F) Recommended maximum inlet pressure 3 barg (45 psig) Operating pH 1 to 14 Reservoir volume 2500 ml 2 175 ml Minimum working volume with five 0.11 m 2 (1.2 ft ) cassettes 2 250 ml System holdup volume 20 ml Minimum working volume with one 0.11 m 2 (1.2 ft ) cassette Kvick Lab Operating Instructions 28-9646-76 AA 63 7 7.1 Reference information Kvik Lab system specifications Pump Parameter Specification Inlet and outlet ports (250 pump) 3/4-inch sanitary fittings Rotors, material of construction Stainless steel Maximum temperature, operating (pump rotor limited) 50°C (122°F), steamable rotors, not autoclavable Maximum temperature, steaming/ autoclaving (pump stopped) 121°C (250°F) Input voltage, 250 pump, 115 VAC version 104-126 VAC, 60/50 Hz Max. input current 8.9 A Input voltage, 250 pump, 230 VAC version 200-230 VAC, 50/60 Hz Max. input current 3.8 A Pump fuse rating 250 volt, 10 amp, fast-blow, ABC10A type, 31.75 mm × 6.35 mm (1.2 in × 0.25 in) Flow rate 0 to 12 liters (0 to 3 gallons) at 1.4 barg (20 psig) of back pressure High pressure cut-off system Stops pump at feed pressure of 4 barg (60 psig) See the pump manufacturer’s user manual for additional details. 64 Kvick Lab Operating Instructions 28-9646-76 AA Reference information 7 Kvik Lab system specifications 7.1 Kvick Lab cassette holder Fittings Parameter Specification Feed, retentate, and permeate port fittings 1/2 inch sanitary Wetted parts Parameter Specification Flow distribution manifold Polished 316L stainless steel, Ra <0.8 µm Back plate Polished 316L stainless steel Non-wetted parts Parameter Specification Tie rods, washers, stand Stainless steel Tie rod nuts Bronze Operating parameters Parameter Specification Recommended operating flow rate 7 to 8 l/min/m of installed membrane 2 area (650 to 750 ml/min/ft of installed membrane area) with a retentate pressure of 0.7 barg (10 psig) Recommended operating temperature 5°C to 50°C Maximum operating temperature 121°C (250°F) Recommended operating pressure 0.34 to 3 barg (5 to 45 psig) Operating pH 1 to 14 Hardware holdup volume (feed/ retentate) less than 2 ml Maximum membrane area 0.55 m (6.0 ft ) 2 2 2 Type and number of cassettes: Parameter Specification Kvick Lab 0.11 m (1.2 ft ), maximum of 5 cassettes installed Kvick Lab Operating Instructions 28-9646-76 AA 2 2 65 7 7.2 Reference information Chemical compatibility Parameter Specification Cassette holdup volume 20 ml 7.2 Chemical compatibility You can damage a Kvick Lab cassette holder by exposing it to chemicals incompatible with stainless steel. The gaskets used to connect piping to the sanitary fittings of the holder are made of EPDM, and are resistant to many types of chemicals (Table 7-1). Consult the GE Healthcare Cassettes: Instructions for Use manual for guidelines on the chemical compatibility of cassettes. Table 7-1. Chemical resistance of Kvick Lab and cassette holder wetted parts. Chemical SS 316L resistance EPDM resistance Acetic acid 1.7 M Resistant Resistant EtOH 20% Resistant Resistant EtOH 40% Resistant Resistant Ethylene glycol Resistant Resistant Formaldihyde Resistant Resistant Formic acid 10% Resistant Resistant Glycerol 100% Resistant Resistant Hydrochloric acid 0.1 M Not recommended Resistant Isopropyl alcohol 30% Resistant Resistant Nitric acid 0.1 M Resistant Resistant Phosphoric acid 25% Limited Resistance Resistant 2 Resistant 3 Sodium Chloride 0.5 M Resistant Sodium hydroxide 2 M Resistant Resistant Trifluoroacetic acid 0.1% Resistant Resistant Resistant Resistant Resistant Resistant Triton TM X-100 100% Urea 8 M 66 2 Sodium chloride can cause corrosion on stainless steel at pH less than 5. Do not use sodium chloride in storage solutions. 3 Maximum recommended exposure is 4 hours. Kvick Lab Operating Instructions 28-9646-76 AA Reference information 7 Service and training 7.3 7.3 Service and training Contact GE Healthcare or visit www.gelifesciences.com for information on service and training. 7.4 Ordering information For spare parts and other ordering information visit www.gelifesciences.com. Kvick Lab Operating Instructions 28-9646-76 AA 67 7 7.4 68 Reference information Ordering information Kvick Lab Operating Instructions 28-9646-76 AA Appendix A Drawings A.1 Appendix A A.1 Drawings A.1.1 Perspective 2 1 3 4 5 6 7 8 9 Figure A-1. Exploded view of Kvick Lab. Kvick Lab Operating Instructions 28-9646-76 AA 69 Appendix A A.1 Drawings Table A-1. Parts in Figure A-1. 70 Part Description 1 Manual feed port and sight glass 2-inch TC (optional) 2 Vent or diafiltration port 3 Permeate line, 3/8-inch stainless steel hose barb 4 Cooling water outlet (Use is optional) use supplied propylene, 1/2-inch male NPT to 3/8-inch hose barb fittings to connect heating and cooling water inlet and outlet. 5 Pressure gauge, 3/4-inch TC on both ends 6 Compressed air inlet, 3/8-inch stainless steel hose barb 7 Drain line side of diverter valve 8 Reservoir outlet and retentate inlet, 3/4-inch TC 9 Pump inlet and outlet, 3/4-inch TC Kvick Lab Operating Instructions 28-9646-76 AA Appendix A Drawings A.1 A.1.2 Front view 1 2 3 4 Figure A-2. Exploded front view of Kvick Lab. (some versions of this system include fully jacketed tanks) Table A-2. Part in Figure A-2. Part Description 1 Heating/cooling water outlet 2 Heating cooling water inlet 3 Feed line 4 Compressed air inlet Kvick Lab Operating Instructions 28-9646-76 AA 71 Appendix A A.2 Using the cartridge adaptor kit Note: All undesignated sanitary connections are 1/2-inch. A.2 Using the cartridge adaptor kit Installing the hollow fiber cartridge adapter kit on a Kvick Lab system This instruction describes how to install the GE Healthcare cartridge adaptor kit that enables you to use Xampler™, size 3, 4, 5 and 6 hollow fiber cartridges with your Kvick Lab system. Your adapter kit contains the following parts: 72 • 8-inch x 8-inch base plate with 2-piece support rod • Cartridge mounting bracket (top) with manifold • Cartridge mounting bracket (bottom) with mounting blocks • Two, 0 to 4 barg (0 to 60 psig) pressure gauges • Two pressure gauge mounting adapters with TC ends • Drain valve • Three elbows with TC ends • Short-outlet tee with TC ends • Feed manifold with TC ends • Two, 1 1/2-inch TC clamps • Twelve, 1/2-inch gaskets • Twelve, 1/2-inch clamps • Three, 3/8-inch hose barb fittings with TC ends • Three, 3-foot sections of 3/8-inch ID tubing • Three hose clamps • Two, 1 1/2-inch TC to 1/2-inch TC adaptors, polysulfone • Two, 1 1/2-inch TC gaskets, silicone • Two, 1 1/2-inch TC quick disconnect clamps, nylon Kvick Lab Operating Instructions 28-9646-76 AA Appendix A Using the cartridge adaptor kit A.2 Preparing to install the adapter kit 1 Gather an adjustable wrench and a 7/16-inch box wrench. 2 Unpack the adapter kit and lay out components on the laboratory bench next to your Kvick Lab system. 3 Depressurize your Kvick Lab system and disconnect any lines from the drain side of the diverter valve. 4 Learn where to install the adapter kit components by reviewing the flow chart (Figure A-3). Assembling the stand 1 Screw the lower half of the support rod into the stand and tighten just snug using the adjustable wrench and the support rod bosses. 2 Screw the top half of the support rod into the bottom half of the support rod and tighten just snug with the adjustable wrench. 3 Install the lower cartridge mounting bracket on the support rod using the mounting blocks and the 7/16-inch wrench (Figure A-4). 4 Install the upper cartridge mounting bracket by slipping it over the support rod and tightening the hand lock. Install the pressure gauge and piping (Figure A-5). 1 3 2 5 4 6 10 7 11 8 9 12 16 13 17 14 15 18 19 20 Figure A-3. Process flow diagram of a Kvick Lab system with the cartridge adapter kit and hollow fiber cartridge installed. Kvick Lab Operating Instructions 28-9646-76 AA 73 Appendix A A.2 Using the cartridge adaptor kit Part Description Part Description 1 Kvick Lab cassette holder 11 Reservoir 2 Pressure gauge 12 Cooling and heating jacket 3 Feed 13 Capped permeate port 4 Retentate line tubing 14 Compressed air for integrity testing 5 Permeate collection 15 Pump 6 Valve 16 Retentate inlet 7 Hollow fiber cartridge 17 Cooling/heating water outlet 8 Permeate collection 18 Feed 9 Cooling /heating water inlet 19 Diverter valve 10 Vent or diafiltration inlet 20 Drain and recovery Assembling the piping Without fully-tightening the sanitary clamps, assemble the feed-line components (piping, elbows, adapters, pressure gauge, valve, gaskets, and clamps). Use Fig 1-1, Fig 1-2, and Fig 1-3 as guides. Once assembled, connect the feed-line assembly to the feed piping on the lower mounting bracket and the system’s diverter valve. Tighten the sanitary clamps. Cut and attach tubing to the retentate hose barb using a tubing clamp. Attach a hose barb adaptor and clamp to the other end of the tubing. Attach the tubing to the feed reservoir port using a TC gasket and clamp. Note: Adapters for Size 5 and 6 Cartridges—To install a size 5 or 6 cartridge, attach the two 1 1/2-inch TC to 1/2-inch TC adapters to the cartridge using the 1 1/2-inch gaskets and clamps. Then follow the procedure in Step 4, Install a hollow fiber cartridge. Note: Minimizing the working volume—To minimize the working volume of the system, cut a 3 1/2-inch piece of tubing and attach it to the bottom of the reservoir dip tube. 74 Kvick Lab Operating Instructions 28-9646-76 AA Appendix A Using the cartridge adaptor kit A.2 1 2 3 6 4 7 5 8 Figure A-4. Lower (left) and upper (right) cartridge mounting brackets. Part Description Part Description 1 Feed manifold 5 Feed elbow 2 Mounting bracket 6 Hose barb adapter on tubing clamp 3 Rubber mounting block 7 Pressure gauge adapter 4 Support rod 8 Upper cartridge mounting bracket and Retentate manifold Figure A-5. Feed line assembly with blue arrow pointing at diverter valve. with the adapter installed, one outlet feeds the cassette holder and one outlet feeds the cartridge. Kvick Lab Operating Instructions 28-9646-76 AA 75 Appendix A A.2 Using the cartridge adaptor kit Install a hollow fiber cartridge Cartridge with sanitary connectors 1 Gather the cartridge and its two gaskets, and two sanitary clamps of the appropriate size. 2 Raise upper cartridge-mounting bracket so that distance between upper and lower manifolds exceeds cartridge length. 3 Place a gasket on the sanitary flange of the feed piping. Place the end of the cartridge on the gasket and sanitary flange and secure with a sanitary clamp (Figure A-3). 4 Place a gasket on upper end of the cartridge. 5 Loosen the upper cartridge mounting bracket and slide it down to mate with gasket and upper end of cartridge. 6 Tighten upper manifold hand knob to secure it in place. 7 Secure the end of the cartridge to the upper manifold with a sanitary clamp. Cartridge with tubing nipples (QAK-2 kit required) 1 Gather the Xampler size 4 cartridge, two 1/2-inch silicone gaskets and fractional sanitary clamps, two cartridge-to- manifold adapters, two long tube standoffs, two hose barb O-rings, and two short tube standoffs. 2 Raise upper manifold so that distance between upper and lower manifolds exceeds cartridge length. 3 Clamp the tubing nipple adapters to the upper and lower manifolds using the silicone gaskets and sanitary clamps. 4 Push a long tube standoff onto each cartridge tubing nipple. 5 Push an O-ring onto each cartridge tubing nipple and seat it against the standoff. 6 Push a short tube standoff onto each cartridge tubing nipple and seat it against the O-ring. 7 Wet the O-rings with water, and with the standoffs and O-rings in place, use a twisting motion and firmly slide the lower end of the cartridge into the lower manifold adapter. 8 Slide the upper manifold assembly down to mate with the upper end of the cartridge and press firmly. 9 Tighten upper manifold hand knob to secure it in place. Which cartridges can you install? The Kvick Lab cartridge adapter kit can hold the following types of cartridges depending upon the cassette type and the membrane area your application requires: 76 Kvick Lab Operating Instructions 28-9646-76 AA Appendix A Using the cartridge adaptor kit A.2 • Xampler size 3M and 3×2M • Xampler size 4, 4M, and 4×2M • Pilot size 5 and 6 Operating parameters Parameter Value Recommended operating temperature 5°C to 50°C (41°F to 122°F) Maximum steaming/autoclaving temperature (no cartridge) 121°C (250°F) Maximum operating temperature (pump rotor limited) 60°C (140°F) Recommended maximum inlet pressure 3 barg (45 psig) Operating pH 1 to 14 Reservoir volume 2500 ml System holdup volume 20 ml System working volume with size 3M cartridge 200 ml System working volume with size 6 cartridge 500 ml Kvick Lab Operating Instructions 28-9646-76 AA 77 Appendix A A.2 78 Using the cartridge adaptor kit Kvick Lab Operating Instructions 28-9646-76 AA For local office contact information, visit www.gelifesciences.com/contact GE, imagination at work and GE monogram are trademarks of General Electric Company. GE Healthcare Bio-Sciences AB Björkgatan 30 751 84 Uppsala Sweden All third party trademarks are the property of their respective owners. www.gelifesciences.com/filtration Kvick Lab and Kvick are trademarks of GE Healthcare companies. © 2010 General Electric Company—All rights reserved. First published Feb. 2010. All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact your local GE Healthcare representative for the most current information. GE Healthcare UK Ltd Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK GE Healthcare Bio-Sciences Corp 800 Centennial Avenue, P.O. Box 1327, Piscataway, NJ 08855-1327, USA GE Healthcare Europe GmbH Munzinger Strasse 5, D-79111 Freiburg, Germany GE Healthcare Japan Corporation Sanken Bldg. 3-25-1, Hyakunincho, Shinjuku-ku, Tokyo 169-0073, Japan imagination at work 28-9646-76 AA 02/2010
© Copyright 2026 Paperzz