10/14/2013 Introduction to Electron Microscopy Alan Nicholls, PhD Associate Director RRC, Director Electron Microscopy Service Research Resources Center, UIC University of Illinois at Chicago RRC - Electron Microscopy Service Introduction Microscopy is one of the few methodologies applied to nearly every field of science and technology in use today. A microscope can be as simple as a hand held optical device or as complex as a multi-million dollar research multi-million dollar research tool. University of Illinois at Chicago RRC - Electron Microscopy Service 1 10/14/2013 Why do we need microscopy: • Resolution of the Human Eye ~ 0.1 mm • Apparent resolution of the Eye for a Magnified Object is: R~ 0.1 mm Magnification Magnification • 100 • 1,000 • 10,000 • 1, 000,000 Resolvable Distance 1 µm - (1,000 nm) 0.1 µm - (100 nm) 0.01 µm - (10 nm) 0.0001 µm – (0.1 nm) • What is the difference between Magnification and Resolution? Magnification: Apparent enlargement of an object Resolution: Capability of making distinguishable the individual parts of an objects University of Illinois at Chicago RRC - Electron Microscopy Service Imaging Resolution In 1870, Ernst Abbe derived mathematical expression for resolution of a microscope this was modified by Raleigh. R = d/2 = kλ µsinα k=0.5 Abbe k=0.61 Raleigh Center of one Airy disk is superimposed on the first minimum of another (26.5% dip ) Where λ is the wavelength of the illumination, µ the refractive index of the lens and α the illumination half angle. (In light microscopy µsinα is known as the numerical aperture (NA)) Ernst Abbe 1840 - 1905 University of Illinois at Chicago RRC - Electron Microscopy Service 2 10/14/2013 Imaging Resolution In light microscopy the NA of a light optical lens and therefore resolution can be increased by a) increasing the half angle (α) of illumination, b) increasing the refractive index (µ) of the lens by using Crown glass or oil immersion or c) decreasing the wavelength (λ) of illumination. In electron microscopy µ cannot exceed 1.0, α is very small, and the only thing that can be adjusted is decreasing the wavelength of illumination, λ, by increasing the acceleration voltage. University of Illinois at Chicago RRC - Electron Microscopy Service Why Electron Microscopy? So the main advantage of electron microscopes over light microscopes lies in the much shorter wavelength of the electron and therefore its much greater resolution. Light microscope : UV light, λ is ~400nm >> best resolution (oil immersion) ~ 150nm Electron microscope: 200kV electrons, λ is 0.0025nm >> resolution ~ 0.015nm 1MV electrons, λ is 0.0009nm >> resolution ~ 0.005nm Another advantage is that electromagnetic lenses have variable focal lengths unlike lenses used in light microscopes. and strength can be increased by using a pole piece to concentrate the magnetic field. University of Illinois at Chicago RRC - Electron Microscopy Service 3 10/14/2013 Why Electron Microscopy? BUT electrons have a short mean free path in air so column has to be under vacuum this limits the types of specimen we can look at – nothing WET! Electromagnetic lenses are not as good as optical lenses and suffer from aberrations which limits the resolution to ~10x the Abbe resolution (~0.15nm & 200kV). University of Illinois at Chicago RRC - Electron Microscopy Service Why Electron Microscopy? Most significant aberration is spherical aberration, Cs Light optical lenses Cs ~ µm Electron optical lenses Cs ~ mm The shorter the focal length, the smaller Cs and the smaller the amount of blur. Electrons away from the optical axis are focused to a different point to those on axis. This can now be corrected, at a price ($1,000,000), in an aberration corrected microscope. University of Illinois at Chicago RRC - Electron Microscopy Service 4 10/14/2013 Why Electron Microscopy? BUT electrons have a short mean free path in air so column has to be under vacuum this limits the types of specimen we can look at – nothing WET! Electromagnetic lenses are not as good as optical lenses and suffer from aberrations which limits the resolution to ~10x the Abbe resolution (~0.15nm). Non-conducting specimens may charge and may need coating with a conductive film (e.g. C, Au, Cr). RRC - Electron Microscopy Service University of Illinois at Chicago Why Electron Microscopy? Naked Eye 10-2 m Natural World 1 cm 10 mm Head of a pin 1-2 mm Ant ~ 5 mm 1,000,000 nanometers = 1 millimeter (mm) 0.1 mm 100 µm 10-5 m 0.01 mm 10 µm 10-6 m Nanoworld 10-7 m ~10 nm diameter Pollen grain Red blood cells Infrared 10-4 m 10-8 m Zone plate x-ray “lens” Outer ring spacing ~35 nm 1,000 nanometers = 1 micrometer (µ µm) Visible Red blood cells with white cell ~ 2-5 µm Light optical microscope MicroElectroMechanical (MEMS) devices 10 -100 µm wide Scanning Electron Microscope (SEM) 0.1 µm 100 nm Ultraviolet Fly ash ~ 10-20 µm Microworld Microwave 10-3 m Man-Made World Self-assembled, Nature-inspired structure Many 10s of nm 0.01 µm 10 nm Nanotube electrode ATP synthase m Carbon buckyball 1 nanometer (nm) Soft x-ray 10-9 Atoms of silicon spacing ~tenths of nm 10-10 m 0.1 nm ~1 nm Carbon nanotube diameter ~1.3 nm diameter Quantum corral of 48 iron atoms on copper surface positioned one at a time with an STM tip Corral diameter 14 nm University of Illinois at Chicago Transmission Electron Microscopy (TEM) RRC - Electron Microscopy Service 5 10/14/2013 Electron Microscopy a brief history 1897 J J Thompson discovers electrons 1924 Louis de Broglie develops theory of wave nature of electrons 1926 Hans Busch discovers electric fields act as lenses for electrons 1932 Max Knoll & Ernst Ruska build the first TEM (x14.4) 1934 TEM with better resolution than LM (x12000) 1938 First Scanning TEM (von Ardenne) 1939 First Commercial TEM, 10nm resolution, 80keV - Siemens, Germany 1940 Commercial TEM with 2.4nm resolution - RCA, USA University of Illinois at Chicago RRC - Electron Microscopy Service Electron Microscopy a brief history 1942 First Scanning Electron Microscope 1944 First EELS experiments (Hillier) 1954 First lattice resolution image (1.2nm) ~1965 First analytical microscope using XWDS (EMMA) 1966 First Commercial Scanning Electron Microscope (Cambridge S1) ~1970 First commercial energy dispersive X-ray detector (EDAX) 1970 First images of single atoms (STEM, U. Chicago) 1973 First atomic resolution TEM image (0.2nm) ~1977 First commercial EELS spectrometer (Gatan) 1998 First TEM and STEM aberration corrected microscopes ~1999 First commercial Silicon Drift Detector (SDD) for X-rays 2004 First commercial sub-0.1nm resolution (FEI aberration corrected TEM) University of Illinois at Chicago RRC - Electron Microscopy Service 6 10/14/2013 Major Types of Electron Microscope Scanning Electron Microscope (SEM) - Beam is scanned across the surface of a bulk specimen and signal detected from each point using various detectors and displayed on a monitor. Transmission Electron Microscope (TEM) - Beam is transmitted through a sample 3mm in diameter and < 100nm thick. Image is viewed on a fluorescent screen and collected using a CCD camera. Scanning Transmission Electron Microscope (STEM) - Combination of TEM and SEM using Transmission specimen and detectors after the specimen. Beam is scanned across specimen. also known as Analytical Electron Microscope (AEM) - When combined with detectors for measuring the local composition of the specimen by X-ray analysis (XEDS) & Electron Energy Loss Analysis (EELS) University of Illinois at Chicago RRC - Electron Microscopy Service Comparison of types of microscope University of Illinois at Chicago RRC - Electron Microscopy Service 7 10/14/2013 Electron Sources Field Emission Tip Tungsten Wire LaB6 Crystal Thermal Emitters University of Illinois at Chicago Thermal electron source RRC - Electron Microscopy Service What can EM tell us? Electron Microscopes use a beam of highly energetic electrons to examine objects at high resolution. They can yield the following information: Morphology The shape and size distribution of the particles making up the object Composition The elements and compounds that the object is composed of and the relative amounts of them Crystallographic Information How the atoms are arranged in the object Topography The surface features of an object or "how it looks", its texture. University of Illinois at Chicago RRC - Electron Microscopy Service 8 10/14/2013 Beam Specimen Interaction The incident electron beam interacts with the specimen and produces a range of different imaging and analytical signals. TEM – transmitted and diffracted electrons. SEM – secondary, backscattered and absorbed electrons Analytical – X-rays (XEDS) and inelastically scattered electrons (EELS) University of Illinois at Chicago RRC - Electron Microscopy Service Scanning Electron Microscopy In an SEM: • The electron beam is focused to a small spot that is scanned across the specimen by beam deflectors • Electrons that are generated or back scattered from the specimen surface, are collected to form the image • The specimen can be bulky! • The image is built up one pixel at a time, as the beam is scanned across the specimen University of Illinois at Chicago E0 0.1eV-30kV RRC - Electron Microscopy Service 9 10/14/2013 SEM Imaging Signals Secondary Backscattered Low energy < 50eV High energy (50% > 0.5Eo) Atom is ionized by secondary generation. Relaxation can lead to generation of a characteristic X-ray Number backscattered depends on atomic number of specimen (Z contrast) University of Illinois at Chicago RRC - Electron Microscopy Service SEM Interaction Volume University of Illinois at Chicago RRC - Electron Microscopy Service 10 10/14/2013 Scanning Electron Microscopy Hitachi S-3000N Variable Pressure SEM 0.1-30kV SEM with X-ray. Able to operate in high vacuum and low vacuum (1-270Pa), where gas ions generated by the beam neutralize build up of negative charge on non-conducting specimens. Backscattered Electron Viscoelastic Sponge (20kV, 30Pa, uncoated, backscattered) University of Illinois at Chicago RRC - Electron Microscopy Service Scanning Electron Microscopy Secondary Electron Virtually no specimen prep needed if specimen is dry and small enough to fit in column (6” diameter or less and less than 1” high) Silicon Nanotubes (10kV, uncoated) Typical XEDS spectrum showing Si, C and O (10kV) > University of Illinois at Chicago RRC - Electron Microscopy Service 11 10/14/2013 Scanning Electron Microscopy Backscattered Electron Mouse skin Split magnification showing red blood cells Wet life science specimens have to be fixed (Glutaraldehyde & and OsO4) and dehydrated without the cells collapsing (water replaced by hexamethyldisilazane (HMDS)) and may need a conductive coating (C or Au/Pd) to stop charging. University of Illinois at Chicago RRC - Electron Microscopy Service Transmission Electron Microscopy The TEM has the same basic components as a light microscope • Radiation source (100-300 KeV electrons) • Condenser lens • Specimen • Objective lens • Projector lens • Viewing system Microscope column is evacuated. Lenses are electromagnetic. The specimen must be dry or dehydrated and VERY thin (<100 nm), to permit the electrons travel through the sample. The entire field of view is illuminated during imaging. The resulting image is a 2D representation of a 3D object. University of Illinois at Chicago RRC - Electron Microscopy Service 12 10/14/2013 Transmission Electron Microscopy Apertures University of Illinois at Chicago RRC - Electron Microscopy Service TEM Image Contrast Contrast in the TEM Nearly all electrons are transmitted through the specimen in TEM. Contrast arises due to scattering of the beam. A TEM image is made up of non-scattered electrons (which strike the screen) and scattered electrons which do not and therefore appear as a dark area on the screen Concentrated in smaller angles Contribute more to analytical signals Contribute most to image contrast (dark areas) Contribute most to image (bright areas) University of Illinois at Chicago RRC - Electron Microscopy Service 13 10/14/2013 TEM Image Contrast Contrast in the TEM Nearly all electrons are transmitted through the specimen in TEM. Contrast arises due to scattering of the beam. Mass-thickness contrast (small objective aperture to select direct beam) Dominant contrast mechanism for amorphous specimens. Diffraction contrast (small objective aperture selecting single spot) Dominant contrast mechanism in crystalline specimens. Phase contrast (large objective aperture to select multiple spots) Contrast is due to the phase of the electron wave at the exit plane on the specimen. University of Illinois at Chicago Mass-thickness contrast Atrial Muscle (80kV) Contrast is enhanced by staining with heavy metals (U, Pb) which are taken up non-uniformly by the specimen. Dominant in amorphous specimens. University of Illinois at Chicago RRC - Electron Microscopy Service Diffraction contrast Stacking faults in Al Bragg scattering from crystalline planes in specimen. Dominant in crystalline specimens RRC - Electron Microscopy Service 14 10/14/2013 Phase contrast Dislocation in Silicon “Atomic” resolution images can be obtained, close to focus, on oriented crystalline specimens due to elastic scattering of the beam by the atomic nuclei. Contrast due to phase of electron wave at exit plane of specimen. Diffraction Pattern Tin oxide nano-crystals Changing projector lens strength a diffraction pattern can be displayed instead of an image University of Illinois at Chicago RRC - Electron Microscopy Service Transmission Electron Microscopy JEOL JEM-3010 300kV Materials Science TEM with Gatan CCD camera for imaging and Noran X-ray detector for microanalysis. Atomic resolution image of polycrystalline Barium Titanate (300KeV) University of Illinois at Chicago RRC - Electron Microscopy Service 15 10/14/2013 TEM Specimen Preparation Preparation can take from minutes (particles) to days (cross sectional) University of Illinois at Chicago RRC - Electron Microscopy Service TEM Specimen Preparation Specimens are nearly always circular and 3mm in diameter. Perpendicular to this disk, the specimen must be thin enough for electrons to pass through (t<100nm). Self supporting, electrochemically thinned to perforation, Al-4% Zn University of Illinois at Chicago Spore supported on a carbon film on a locator grid RRC - Electron Microscopy Service 16 10/14/2013 TEM Specimen Holder Single tilt specimen holder Tip is removable and is supported while the specimen is loaded University of Illinois at Chicago RRC - Electron Microscopy Service Transmission Electron Microscopy Er coated Si nanowire showing faults in the Si nanowire and crystallinity in the Er coating. dispersed onto a carbon film (300KeV) Dark Field 1 Dark Field 2 Bright Field DF1 DF2 University of Illinois at Chicago RRC - Electron Microscopy Service 17 10/14/2013 Transmission Electron Microscopy SiGe quantum well srtructure at atomic resolution - self supporting specimen 110 diffraction pattern from area DF1 DF2 University of Illinois at Chicago Low mag of similar area RRC - Electron Microscopy Service Analytical Electron Microscopy JEOL JEM-ARM200CF 200kV Cold Field Emission Probe Aberration Corrected Materials Science TEM/STEM with Gatan Enfina Electron Energy Loss Spectrometer and Oxford SDD X-ray detector for microanalysis. University of Illinois at Chicago RRC - Electron Microscopy Service 18 10/14/2013 Analytical Electron Microscopy A TEM becomes an Analytical Electron Microscope (AEM) when spectroscopic information can be collected as well as images. Typically X-Ray (XEDS) and Energy Loss (EELS). AEMs will often be capable of operating in Scanning Transmission (STEM) mode. In a field emission STEM the probe size can be smaller than interplanar spacing. Some electrons suffer a near nuclear collision and are elastically scattered through large angles. When collected by a high angle annular detector, the image not only contains atomic resolution, but the intensity of the atom columns is proportional to Z. Using the annular dark field detector allows simultaneous acquisition of the EELS spectrum. University of Illinois at Chicago RRC - Electron Microscopy Service Analytical Electron Microscopy Incident Probe Annular Detector Z-contrast Image BF-Image Spectrometer CCD-EELS Detector University of Illinois at Chicago RRC - Electron Microscopy Service 19 10/14/2013 Analytical Electron Microscopy Inelastic Scattering (fast electron – electron) Elastic Scattering (fast electron – nucleus) XEDS HAADF STEM Mass/ Thickness EELS Diffraction University of Illinois at Chicago Spectroscopy RRC - Electron Microscopy Service Analytical Electron Microscopy Atomic resolution HADF STEM image of a bismuth, strontium, copper, calcium oxide specimen containing a fault. The heaviest atomic number element (Bi) is the brightest with darker Sr atoms on each side of the Bi. In between there are three Cu/Ca columns. In the faulted region there are two more Cu/Ca columns but the structure recovers within two unit cells by loosing a Cu/Ca column. University of Illinois at Chicago RRC - Electron Microscopy Service 20 10/14/2013 AEM – X-ray analysis Sum Sumspectra spectrafrom fromrectangular circular area area X-Ray Spectrum Imaging Collecting full spectrum from every point in scanned image Data can be interrogated after collection University of Illinois at Chicago RRC - Electron Microscopy Service AEM - EELS Z contrast image showing atomic core structure of mixed dislocation. EELS from core and bulk showing increase in first peak Z contrast image of partial dislocation. EELS spectra from various positions on partial dislocation compared to bulk showing differences in electronic state University of Illinois at Chicago RRC - Electron Microscopy Service 21 10/14/2013 Aberration Corrected STEM Uncorrected, Cs=0.5mm JEM-2010F 0.13nm probe size, Si <110>, Dec 1998 Corrected, Cs=-0.5µm JEM-ARM200CF 0.08nm probe size, Si <110>, Oct 2011 There are similar analytical improvements University of Illinois at Chicago RRC - Electron Microscopy Service Aberration Corrected STEM Pt on Alumina catalyst Catalytic performance dependent on Pt particle size. Now able to image individual platinum atoms What effect do these have on catalytic properties? Visible also on carbon film, not as attracted to substrate as clusters? Are they present on an active catalyst after use? Higher resolution answers some questions but results in many more! University of Illinois at Chicago RRC - Electron Microscopy Service 22 10/14/2013 Aberration Corrected STEM Atomic-resolution Z-contrast image of epitaxial LaCoO3 films grown in LaAlO3 showing a superstructure. Z-contrast images of SrTiO3 film on GaAs for two different growth conditions. The SrTiO3 is shown in the [001] orientation while the GaAs is seen along the [110] in both samples. a): SrTiO3 film deposited on GaAs after depositing a Ti pre-layer; b): same growth conditions forSrTiO3 except that no Ti pre-layer was deposited. University of Illinois at Chicago RRC - Electron Microscopy Service Aberration Corrected STEM Z-contrast image of Ca3Co4O9 [110] Simulated HADF ABF of Ca3Co4O9 [110] Simulated ABF Atomic Resolution EELS Imaging Ca L (green), Co L (blue), O K (red) R.F. Klie, Q. Qiao, T. Paulauskas, Q. Ramasse, M.P. Oxley, J.C. Idrobo, Phys. Rev. B, 85(5), 054106 (2012) University of Illinois at Chicago RRC - Electron Microscopy Service 23 10/14/2013 Aberration Corrected STEM Atomic Resolution XEDS Imaging University of Illinois at Chicago RRC - Electron Microscopy Service The Need for 3-D Analysis A two-headed rhino – a projection artifact A TEM produces a 2-D image of a 3-D object Hayes (1980) – “When we see this image we laugh, but when we see equivalent (but more misleading) images in the TEM, we publish!” University of Illinois at Chicago RRC - Electron Microscopy Service 24 10/14/2013 The Need for 3-D Analysis By looking only in projection we can be fooled! >> Electron Tomography RRC - Electron Microscopy Service University of Illinois at Chicago Tomographic Reconstruction Representation of sampling in Fourier space Data points (b) Weighted backprojection – to account for under sampling of high frequencies ORIG High frequencies undersampled WBPJ BPJ (c) Iterative backprojection (SIRT) – iteratively constrain the reconstruction to be give re-projected images that are identical to the original projections. The Principle of (a) Backprojection 0 2 4 6 8 10 P. A. Midgley and M. Weyland, Ultramicroscopy 96, 413-431 (2003) University of Illinois at Chicago RRC - Electron Microscopy Service 25 10/14/2013 Tomographic Reconstruction Catalyst tilt series Pt on SiO2 One axis tilt through ±60o Images acquired every 1.5o University of Illinois at Chicago RRC - Electron Microscopy Service In Situ – Heating/ Cooling Stages Gatan 652 double tilt heating stage Gatan 636 double tilt LN2 cooling stage University of Illinois at Chicago RRC - Electron Microscopy Service 26 10/14/2013 In Situ – SPM inside the TEM SPM Probe Specimen SPM Probe Nanofactory STM-TEM stage Specimen SPM probe being bought into contact with nanowire, eventually voltage applied which melts nanowire University of Illinois at Chicago RRC - Electron Microscopy Service Conclusions – Electron Microscopy • Electron microscopy is a powerful tool for the study of materials. • Transmission EM is a full field viewing technique, like conventional light microscopy • Scanning EM and STEM are point scanning techniques • Specimen preparation techniques are often complex and timeconsuming especially for cross sectional materials microscopy. • The higher the resolution, the poorer the sampling. A 1970 estimate was that all TEMs in the world from 1940 to 1970 had only looked at 0.3mm3 of material. • SEM images are of the surface of the specimen. • TEM & STEM images are 2D images of 3D objects – generally all TEM information is averaged through the thickness of the specimen. • Ionizing radiation can damage the specimen. Contamination from hydrocarbon build up can occur – specimens should be as clean as possible • Is what you see representative of the whole specimen? University of Illinois at Chicago RRC - Electron Microscopy Service 27 10/14/2013 Electron Microscopy Service @ UIC JEM-1220 AXIS-165 XPS JEM-3010 JEM-ARM200CF VT-SPM University of Illinois at Chicago HB601UX Ramascope 2000 JSM-6320F S-3000N RRC - Electron Microscopy Service 28
© Copyright 2026 Paperzz