Use and Care of Light Microscopes Compound and dissecting light microscopes provide magnification ranges from the macroscopic to the limits afforded by light—from a few times actual size to 1000×. These microscopes must be treated with the care and respect that such precision instruments require. To get the most out of the microscopes, follow these precautions and instructions for use. Precautions Avoid getting fluids on the microscopes. Wipe any dishes and slides that have specimens in fluid with a towel before placing them on the microscopes. Never mount a slide on the compound microscope unless its specimen is covered by a coverslip. Sea water is especially corrosive, but other physiological fluids contain salts that can corrode microscope parts as well, and organic solvents can erode cements that hold the lenses in place; some can corrode lenses or their optical coatings. If spills do reach a microscope, wipe it dry as soon as possible; use only lens tissue or cotton-tipped applicators on the lenses themselves. Before putting a microscope away, blot up any water on it and then wipe it down with a second towel dampened with fresh water and dry it. Don’t touch the lenses. Occasional cleaning to remove oil or grease may be necessary; this should be done only with special lens tissue or cotton-tipped applicators, in some cases dampened with solvents. Because different microscopes have different cements holding their lenses, each may need to be treated differently. Please consult the instructor before attempting cleaning. Please do not loosen any set screws, such as the screw holding the head of the microscope in its mounting ring. (Please do not turn the head to show lab partners views under the scope; keep the microscope fully stationary and make them move to look in it.) Compound microscope Turning the illumination on and off should be done gradually by using the rheostat appropriately. Before turning on the rocker switch, make sure the rheostat is turned all the way down to 0; then turn on the rocker switch and dial the rheostat slowly up until some illumination can be seen. Similarly, in turning the illuminator off, turn the rheostat down to 0 before turning off the rocker switch. This gradual on-and-off is important for long life of the bulb. Do not run the illuminator at full throttle (6) except briefly when making observations at high magnification; this setting is higher than the voltage rating of the bulbs. Bright-field Köhler illumination. Optimum resolution is obtained only with appropriate setting of the lenses and apertures—specifically a condition called Köhler illumination in which the conjugate field planes and conjugate aperture plans are properly aligned. This illumination minimizes stray light that would degrade the image and balances resolution against contrast to afford the sharpest possible image, especially for photography. Köhler illumination can be set this way: . 1. Focus roughly on a specimen (starting with a low-magnification objective—i.e., the 4×) and then changing to at least the 10× objective. If the two eyepieces are not set so that both of your eyes are focused on the specimen, adjust them by turning the knurled knobs on them. (Most microscopes will have just one independently focusable eyepiece; focus first on a small, crisp detail in the specimen with just the one eye looking through the non-focusable eyepiece [e.g., tilt your head so that only your right eye sees the specimen through the right eyepiece]; then look through just the other eyepiece [e.g., tilt your head so that only your left eye sees the specimen through the left eyepiece] and adjust the knob on it until the same detail can be seen clearly with that eye. Make sure both eyes see the same field of view completely once the eyepieces are adjusted; adjust the interocular distance as necessary to accomplish this. Use of light microscopes 2. Check that the lamp filament is focused on the front aperture of the condenser. Close that aperture fully and see that the image of the filament is projected onto it. If a focusing knob is provided for the lamp, adjust to sharpen the image and center it on the aperture. 3. Pick the lens (10× – 100×) that you will be using and focus on the specimen with it. (Köhler illumination needs to be set independently for each objective lens, so you will need to run through the following steps for whichever lens you use for critical observation or photography before you start with that lens.) 4. Close down the field-stop diaphragm until you can see its edge as you look through the eyepieces. Center the image of the edge (the diaphragm aperture) in the field of view using the alignment screws on the condenser lens. 5. Focus the image of the edge of the field-stop diaphragm using the knob on the condenser lens and recenter it if necessary. 6. Expand the field-stop diaphragm until its aperture goes just beyond the edge of the field of view, no further. 7. Adjust the condenser diaphragm: while looking at a specimen, open the diaphragm fully, then close it down slowly until the brightness of the image just starts to dim. (If an eyepiece telescope or Bertrand lens is available, view the condenser aperture with it, and adjust the aperture so that it blocks a quarter of the width of the full width of the aperture of the lens [i.e., when about 3 /4 of the maximum diameter of the aperture remains illuminated]. In the absence of such visual aids, closing the condenser just to that point at which the image first begins to dim accomplishes that quarter-width closure.) With specimens of high contrast, the aperture can be opened further to achieve better resolution; closing it any further, however, reduces resolution. 8. Adjust illumination: adjust brightness with the rheostat on the illuminator as you view specimens, not with the condenser aperture. (It is also possible, in the absence of lamp-brightness control, to use neutral-density filters to reach desired brightness.) . This is Köhler illumination. It is the optimum for observing detail as you study a specimen and the optimum, especially, for photomicrography. As you change objective lenses, you will need to adjust the apertures and condenser lens for each, starting with step 4 above, to achieve Köhler illumination with it. Contrast can potentially be enhanced on transparent specimens by closing down the condenser aperture, but realize that by doing so you sacrifice resolution—you are leaving Köhler illumination. Contrast can also be enhanced (again sacrificing resolution, but not as severely) by making the illumination oblique—e.g., placing a narrow strip of paper across the substage illuminator. (It is best to use phase contrast or differential interference contrast to view transparent specimens—i.e., special illumination and lenses.) Use the dissecting microscope to find and mount specimens on slides for the compound microscope. Specimens on slides must be covered with a coverslip (never use the microscope to examine a specimen without a cover slip.) Live specimens also usually need to be compressed with the cover slip but not squashed by it. To achieve the correct compression, place small wax or clay feet on the corners of the coverslip by picking each corner into a piece of wax or clay and getting just a very small fleck on one side at the corners. Then lower the coverslip onto the specimen by resting first one edge on the slide and then lowering the other gradually with a dissecting needle, thus minimizing trapping of air bubbles on the specimen. Compression can be adjusted by pressing gently on the wax/clay feet of the coverslip and by wicking out fluid from under the coverslip with bibulous paper or filter paper (or other absorbent paper). 2 Use of light microscopes Substage controls on the Olympus microscope: 2: Focusing knob for objective lens 3: Field-stop aperture 3a: Condenser aperture 4: Focusing knob for condenser lens 5: Centering knobs for condenser lens Setting the condenser aperture: 3 Views of field-stop aperture during setup: 3: out of focus and uncentered 4: in focus, still uncentered 5: in focus and nearly centered 6: expanded but not quite centered 7: centered and expanded just beyond field of view Use of light microscopes Dissecting microscope The eyepieces (oculars) adjust in focus independently and should be adjusted when you first start work on the microscope. The spacing between the eyepieces (interpupilary spacing) is also adjustable and should be set so that both eyes see the specimen simultaneously. It is best to focus on a specimen with the large focus knob of the microscope while looking first through the right eyepiece (only) with your right eye; then look through the left eyepiece (only) with your left eye and turn the knurled ring on it until the same object you saw with your right eye in the other eyepiece is in focus. Magnification is changed by rotating the knurled knob on the right side of the head. To obtain parfocal zoom in magnification (i.e., so that the image stays in focus at all magnifications), first focus on a specimen at high magnification, then zoom to low magnification and adjust focus with the eyepieces only. Illumination. For reflected illumination, place the fiber-optic heads relatively high above the specimen and facing downward. Their position should not be so close to the specimen that they heat it up unnecessarily, and their angle should be adjusted to avoid producing glare on your eye. Adjust the substage mirror to give a high-contrast background (dark or light). For transmitted illumination, aim one of the fiber-optic heads horizontally through the rear hole on the stage and adjust the mirror to give a fully illuminated field of view. The binocular head should be firmly clamped in its support ring. (Please do not loosen the set screw on this ring and turn the head to show lab partners views under the scope.) Putting microscopes away Remove all specimens from the microscope. Return the magnification to the lowest lens/setting available. Ensure that no sea water or other fluids are on the microscope. Wipe off sea water with a towel dampened in fresh water and then dry the spot. Turn off the illumination. Replace the plastic dust cover on the microscope. 4
© Copyright 2026 Paperzz