Transmission Electron Microscopy and Serial Sections for Light Microscopy from the Same Block in Routine Renal Pathology ERNESTO O. HOFFMANN, M.D., HERMAN B. GARRETT II, M.S., JOAN R. COOVER, MT, B.S.(ASCP), EMSA, AND TERESA R. FLORES, HT(ASCP) Routine serial sections for light microscopy (LM) from one sample and transmission electron microscopy (TEM) from selected areas of another sample are current procedures in renal pathology. At the present time, there are methods available to perform high-resolution light microscopy (HRLM) and TEM from the same block. However, this technology is not directly applicable or convenient for routine renal biopsies. Most methods presently available to obtain TEM studies from different areas of the same large HRLM block use reembedding methods, which considerably delay TEM studies, or mesa technics that fragment and waste the original block. Present methods to obtain serial sections from plastic blocks have been used in plant and animal morphology and need to be modified for renal pathology. In the present article, the authors present a technic that produces immediate TEM sections from different areas of the same HRLM block. This method does not produce fragmentation or waste of the original block, and TEM sections are obtained within two hours after large HRLM sections are screened. The latter sections are mounted serially on a regular microscopic slide on 8-15 separate wells painted with waterrepellent ink. The methods are simple and produce large HRLM and TEM sections in 24 hours, avoiding separate samples for LM and TEM studies. (Key words: High-resolution light microscopy; Renal pathology; Serial sections; Electron microscopy) Am J Clin Pathol 1983; 80: 441-444 RECENTLY, several authors have recommended the use of TEM histotechnology to improve the study of renal biopsies in the light microscope.1-2-78121314,27-29 This method produces HRLM sections not possible to produce with current wax methods.18,29 The advantages of HRLM are better preservation of tissues, seldom use of special stains, no overlay of structures, accurate morphometry analysis, better utilization of the resolution power of the LM, and potential use of the same block for immunohistology.9'0,25 In addition, the large plastic blocks prepared for HRLM can be sectioned serially and can be used directly for high-quality TEM studies, avoiding separate samples for LM and TEM studies. Different methods have been recommended for this latter purpose Received January 20, 1983; accepted for publication March 28, 1983. Address reprint requests to Dr. Hoffmann: Laboratory Service, Veterans Administration Medical Center, New Orleans, Louisiana 70146. Laboratory Service, Veterans Administration Medical Center and Department of Pathology, Louisiana State University Medical Center, New Orleans, Louisiana for tissues other than kidney. Reembedding of tissues from plastic sections (epoxy) or fragments of the original block prolongs (24-48 hours more) the processing time for TEM studies and/or destroys the original HRLM sections.4,5,6,17,22,26 The single or multiple mesa technics waste and destroy part of the original block. 2 ' 6 The total field method28 produces fragmentation and weakening of the block, interfering with thin sectioning or further sectioning for HRLM. Different methods also are reported in the literature for the serial mounting of plastic sections (other than kidney) on glass slides.21,24 The coating of the blocks with tacky glue does not work well with the epoxy blocks. Paraffin coating of the blocks is not applicable when the sections are mounted on the slides by heat, which is the most common procedure used for mounting plastic sections. The high temperatures melt the paraffin, interfering with the adherence of the sections to the slides. The individual transfer of the serial section into glass slide wells is not recommended for small blocks or when numerous serial sections are needed. But this latter method appears to be simple, fast, and applicable to large epoxy blocks of renal biopsies. In this article, we describe adaptations and modifications of some of the above technics 1519,20,23 for instant TEM thin sectioning of different parts of a large HRLM block. Serial sections from the same epoxy blocks also are obtained for routine renal pathology. Materials, Methods and Findings From 50 consecutive renal biopsies, the tips of both ends of the needle biopsy specimens or representative areas from open biopsy specimens are obtained for fluorescent microscopy. The central part of the needle biopsies or the main part of the open biopsies are fixed in paraformaldehyde fixative and postfixed on Os0 4 . 3 The specimen then is dehydrated and embedded in 0002-9173/83/1000/0441 $01.00 © American Society of Clinical Pathologists 441 ~-^v. CM 00 ON 1^ <(• Sf VO I CM 00 2 cm i <n z < u % V -1 a o 111 Z IIIM|IIII|IIII|IIII|IIII|IIII| 1 1 2 m m c —*- d * * ^ 3 Vol. 80 • No. 4 LM AND TEM IN RENAL PATHOLOGY 443 FIG. 1 (upper). Serial HRLM sections (b) and TEM studies (J) are obtained from the same large block (a). An extra-thick section (c) is sectioned from the original block (a); desired areas are selected and the extra-thick section trisected. Two selected fragments (d) are glued on the blank blocks to obtain secondary blocks (e). These are immediately ready for TEM thin sectioning (J) or for additional HRLM sections. FlG. 1A (lower, left). A HRLM photograph shows focal nephritis. B (lower, right). Two hours later, a TEM section from one of the secondary blocks confirmed the presence of subepithelial, subendothelial, and mesangial deposits. An area from the other secondary block (not shown here) presented normal glomeruli. (A, HRLM X950, B, TEM X8.500) < Spurr, medium viscosity.3 In most cases the molds used to make the epoxy blocks are inverted beem capsules. When the needle biopsies are too long (greater than 7 mm), these are bisected and trisected to fit into one block or each fragment is embedded in a separate block. As an alternative, and especially for the larger open biopsies, molding cups* (14 or 16 mm) also are used. However, the inverted beem capsules are preferred, because the blocks produced here can be sectioned directly (for HRLM and TEM) in a regular ultramicrotome. To obtain HRLM sections from the larger blocks (molding cup blocks) a special microtomef is used. Consecutive 1-3 /im sections for HRLM are obtained and mounted serially on glass slides (Fig. 1). Slides have 10-15 wells painted with a water-repellent marker (any nonwaxy, water-repellent marker is adequate) and contain a drop of distilled water per well. Printed slides with wells are commercially available.}: One to three sections may be mounted in each well. The slide, with the wells and sections, is placed on a 90-100°C hot plate until the sections are distended, the water has evaporated, and the sections are firmly adherent to the glass (10-15 minutes) (Fig. 1). The mounted sections are stained with toluidine blue and basic fuchsin." Immediately after obtaining 1-3-^m HRLM sections, the face of the block may be marked with adhesive tape for orientation purposes. This is optional, however, because the cutting surface (polished surface) may be recognized with a dissecting microscope. A parallel extra-thick section (50200 urn) then is obtained from the same block by hand with a razor blade or with a manual microtome with a glass knife.20 To prevent cracking of the extra-thick sections, the blocks are warmed up with an infrared light20 to between 60-90°C. The thickness of the extra-thick sections may vary (5-500 /tm), depending on the type of work to be done. For routine renal biopsies, 50-200 Mm is adequate. After studying the regular HRLM sections, glomeruli or other areas are selected for TEM studies. If more than one glomerulus (or tissue area) is needed for TEM studies and these are too far apart on the block face to be cut in a single thin section, the extra-thick section is placed in the top of a plastic petri dish and bisected or trisected with a razor blade. Blade, specimen, and cut* Sorvall, Biomedical Division, Newton, Connecticut. t Sorvall JB4, Biomedical Division, Newton, Connecticut. X Roboz Instruments Co., Inc., Washington, D.C. ting hand should be covered with a towel or petri dish top at the moment of sectioning the extra-thick section, because the fragments may pop out of the petri dish and be lost. Double-faced adhesive tape may be used on the bottom of the petri dish to prevent loss of the fragments of the extra-thick section. Each fragment contains the different glomeruli (or tissue areas) selected for TEM studies. When the sections are thin (5-20 ^irn) and for orientation purposes, these may be stained with toluidine blue. This is optional, however, because with appropriate lighting the glomeruli can be seen in the unstained osmicated sections. On the other hand, toluidine blue staining does not interfere with TEM studies. When 2-3 or more glomeruli are close together, all of them are processed for TEM in a single block. Blank (dummy) blocks are prepared in advance with the same resin (Spurr). Either side of the blank block is used, depending on the size of the extra-thick section (or fragment of it) selected for TEM studies. When the bottom of the block (beem capsule block) is to be used, caution should be taken to produce a flat bottom. This may be accomplished by polymerizing the block upside down in a capped beem capsule, which is filled through the cut off tip. Blank blocks are also commercially available.§ The extra-thick sections or the parts selected from it now can be glued to the blank block with cyanoacrylate glue** or epoxy glueff (Fig. 1). A drop of the glue is placed on a toothpick and thinly smeared on the flat surface of the blank block. As soon as the glue becomes tacky (30-60 seconds), the selected part of the extra-thick section (marked faced up) is placed on the glued part on the block. Cyanoacrylate glue solidifies in one hour. The glued block is placed in an incubator at 70°C for the first 10-15 minutes and left at room temperature for the remaining 40-45 minutes. The epoxy glue polymerizes in the incubator in 15 minutes. The piece of adhesive tape now is removed from the face of the secondary block. The areas selected for TEM studies now are glued firmly to the blank block and are ready for trimming (if necessary) and thin sectioning for TEM studies (Fig. 2). These secondary blocks also may be used for additional 1-3-jum HRLM sections as needed. The unused fragments of the extra-thick section may be saved in an empty beem capsule. It is very important to leave very § Ladd, Burlington, Vermont. ** Hysol, E. F. Fullam, Inc., Schenectady, New York. t t Devcon epoxy, Devcon Co., Danvers, Massachusetts. HOFFMANN ET AL. 444 little glue on the surface of the blank block and to wait until the glue is tacky before placing the extra-thick section. Too much glue or fluid glue (especially cyanoacrylates) will penetrate deep into the extra-thick section and destroy the tissues. This is especially true with thinner sections. Polymerization of the glues for less than the suggested times will produce detachment of the extra-thick section. 4. 5. 6. 7. Comments The present method takes advantage of the high quality of plastic (epoxy) embedding for i_M8'12-13,14,27'29 and at the same time offers serial sections and immediate TEM studies of selected areas of renal biopsies. All glomeruli studied by light microscopy may be studied by TEM, even those mounted in the HRLM slides.22 The methods are applicable to other areas of pathology where serial sections and TEM studies are needed from different parts of the same HRLM block. HRLM serial sections are obtained in 24 hours by individual transfer of the sections into the glass slides. A technic to obtain ribbons from epoxy blocks has been published recently21 and is apparently easily adaptable for renal biopsies. TEM studies may be ready immediately (1-2 hours later) from parallel extra-thick sections (Fig. 2). No waste or fragmentation of the block occurs. The original and secondary blocks remain intact with parts of the biopsy ready for additional HRLM or TEM studies as needed. An additional advantage is the increase in the number of glomeruli routinely studied by TEM from each biopsy. Technicians become acquainted with sectioning large plastic blocks for HRLM and TEM and secondary blocks may be thin-sectioned without trimming. Our average number of glomeruli examined by TEM before this method was employed was two glomeruli per biopsy. With the present method, it is 3-4 and a maximum of 12 glomeruli for TEM studies in a single large thin section. Potentially parallel sections from the same epoxy blocks also may be used for immunohistology.91025 This would avoid totally the current fragmentation of the renal biopsy for light, fluorescent, and electron microscopy. Acknowledgment. The authors are in debt to Karen M. Dunn for her excellent secretarial work. References 1. Agodoa LC, Striker GE, Chi E: Glycol methacrylate embedding of renal biopsy specimens for light microscopy. Am J Clin Pathol 1975; 64:655-660 2. Collan Y, KJockars M, Heino M: Revision of light-microscopic kidney biopsy diagnosis in glomerular disease. Nephron 1978; 20:24-31 3. Carson FL: Histologic techniques in electron microscopy. Mono- 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28. 29. A.J.C.P. • October 1983 graph published by the American Society of Medical Technologists, 1979, pp 12, 13-15 Cox US, Seely JC: A simplified resectioning technic. Stain Technol 1974; 49:329-333 Dalley BK, Seliger WG: A new technique for the rapid screening and selection of large pieces of tissue for ultrastructural evaluation. Stain Technol 1980; 55:129 Davidowitz J, Pachter BR, Breinin GM: Acrylic slides for the clearing and reembedding of thick ultramicrotome sections. Stain Technol 1976; 51:247 Ditscherlein G, Schneider W: Comparison of the results of semithin sections with those of conventional paraffin histology and immunohistology in renal biopsy specimens. Int Urol Med Nephrol 1976; 8:227-235 Eastham WE, Essex WB: Use of tissues embedded in epoxy resin for routine histological examination of renal biopsies. J Clin Pathol 1969; 22:99-106 Enestrom S: Immunofluorescent staining of open embedded kidney sections through simultaneous use of two different fluorochromes-conjugated antisera. Stain Technol 1982; 57:31-38 Erlandsen SL, Parsons J A, Rodning BC: Technical parameters of immunostaining of osmicated tissues in epoxy sections. J Histochem Cytochem 1979; 27:1286-1289 Flores TR, Hoffmann EO: In search for a Paragon substitute. Louisiana Society for Electron Microscopy Newsletter 1981; 7: 8-11 Germain JP: Epoxy resin embedding and light microscopy, its advantages and disadvantages. Science Tools 1974; 21:30-32 Heaton JM, Turner DR, Cameron JS: Localization of glomerular "deposits" in Henoch-Schoenlein nephritis. Histopathology 1977; 1:93-104 Hoffmann EO, Flores TR: High resolution light microscopy in renal pathology. Am J Clin Pathol 1981; 76:636-643 Johnson PC: A rapidly setting glue for sectioning and remounting epoxy embedded tissue. Stain Technol 1976; 51:275 Jones G, Gallant P, Butler WH: Improved techniques in light and electron microscopy. J Pathol 1977; 121:141-148 King DG, Kammlade N, Murphy J: A simple device to help reembed thick plastic sections. Stain Technol 1982; 57:307-310 Lindner M, Richards P: Long edged glass knives (Ralph knives), their use and the prospects for histology. Science Tools 1978; 25:61-67 Nadelhaft I, Rose LM: Serial sectioning of epon embedded sections for autoradiography. Stain Technol 1974; 49:178-181 Ribi WA: A golgi-electron microscope method for insect tissue. Stain Technol 1976; 51:13-16 Richards PR: A technique for obtaining ribbons of epoxy and other plastic sections for light microscopy histology. J Microscopy 1980; 122:213-216 Schabtach E, Parketing TA: A method for sequential high resolution light and electron microscopy of selected areas of the same material. J Cell Biol 1974; 61:261-264 Seidl W: A simple technique for serially mounting one micron sections of resin embedded (epon) material. Stain Technol 1976;51:311-313 Shoukrey NM: A device for serially mounting 0.5-2.0 micrometer sections of 1-2 *im epon embedded tissue blocks for light microscopy. Stain Technol 1980; 55:81-85 Sisson SP, Vermier RL: Methods for immunoelectron microscopy. J Histochem Cytochem 1980; 28:441-452 West RW: Superficial warming of epoxy blocks for cutting of 2 5 150 jim sections to be resectioned in the 40-90 jim range. Stain Technol 1972;47:201-204 Xipell JM, Gladwin RC: Routine renal preparation of thin epoxy resin-embedded sections of renal biopsies for light microscopy. Am J Clin Pathol 1972; 58:469-474 Yang GCH, Morrison AB: Total field electron microscopy examination of thin tissue slices via surface-embedding and multiple-mesa technics. Am J Clin Pathol 1980; 74:237-238 Zamboni L: Preparation of percutaneous kidney biopsies for high resolution light microscopy. Hum Pathol 1972; 3:128-132
© Copyright 2026 Paperzz