1) While the push rod to transmitted or reflected is to the working position “R”, fluorescent microscopy can be used. 2) Reduce the field diaphragm, If it is not centered, move it to the center using the special tool. Adjustment can be done by opening the field diaphragm. 3) The push rod to direct image to binocular eyepieces or camera port to the upper left of the trinocular head will send 100% of light to the binocular eyepieces or to the top port. It satisfies high-grape photography, see figure 7. LED Epifluorescence Microscope L3201LED User Guide Preventative Maintenance 1. Cleaning frame and stage Disconnect the plug from mains socket before cleaning. Clean the frame and stage with a soft cloth moistened with a mild detergent solution. Be sure the instrument is dry before using. 2. Cleaning optical parts Microscope eyepieces and objectives are coated. They should not be wiped while dry as dirt or dust may scratch the coating. It is best to remove parts from the frame prior to cleaning. Always blow loose dust away first. Use cotton swabs or lens tissue moistened with a lens cleaner or a small amount of alcohol, then wipe the surface clean with a good quality lens tissue. Solvents such as Xylene should NOT be used as cleaner. Cleaning 100X oil immersion lens 3. The immersion oil should be removed from the lens at the end of each workday using cotton swabs or lens tissue moistened with a lens cleaner or a small amount of alcohol. HAGAVISH 3,HOLON, ISRAEL TEL: 972-3-5595252, FAX. 972-3-5594529 11 [email protected] Congratulations and Thank You for your purchase of The Microscope This precision instrument has been designed to require a minimum of optical and mechanical maintenance. Its excellent design assures years of high quality, reliable service. We recommend you read this entire manual carefully before beginning to use the instrument. WARNING swing the attachment lens to optical center. 6. Condenser Alignment – to center the Abbe condenser into the optical light path, see figure 6: 1) Close the field iris diaphragm. 2) Observe the image using the 5X A B C objective and you will see an image similar to figure 6 A. The Figure 6 image may be out of focus and out of center. 3) To center the field iris diaphragm in the field of view, use the two centering screws at each side of the Abbe condenser mounting ring, see figure 6 B. PLEASE DISCONNECT PLUG FROM MAINS SOCKET BEFORE REMOVING THE BOTTOM PLATE, OPENING THE LAMP REPLACEMENT DOOR, OR REMOVING THE LAMP HOUSE 4) Once the image is in focus and centered in the field of view, open the field iris diaphragm until the image is almost as large as the field of view, see figure 6 C. 7. Fluorescent Illumination Alignment, see figure 7. WARNING 1 2 THE POWER CORD PROVIDED WITH THE EQUIPMENT HAS A GROUNDED PLUG. ALWAYS USE THE POWER CORD WITH A PROPERLY GROUNDED WALL OUTLET. DO NOT EXPOSE THE INSTRUMENT TO HIGH TEMPERATURES OR HUMIDITY. AVOID USING THE INSTRUMENT IN EXTREMELY DUSTY LOCATIONS OPERATING TEMPERATURE 5 DEGREE C. TO 35 DEGREE C. OPERATING HUMIDITY: 20% TO 80% @ 25 DEGREE C. CAUTION: NEVER IMMERSE THE INSTRUMENT IN WATER OR SOLVENT CAUTION: DO NOT PUT ANY FOREIGN OBJECTS IN THE FRAME OR INTO ANY MOVING MECHANICAL PARTS 1 5 4 3 1. Field Diaphragm Adjustment Knob 2. Camera Port 3. Push Rod To Transmitted or Reflected 4. Push Rod To Direct Image to Binocular Eyepieces or Camera Port Compensator 5. Field 10 Diaphragm Centering Screw 1) Diopter Adjustment Proper correction for individual vision is accomplished via the diopter adjustment located at the left eyepiece. Using the 10X objective, bring an Diopter Adjustment Ring image into focus with your right eye only. Once the image is well focused, observe with left eye, make fine adjustments with the diopter adjustment ring to correct for your vision. 4. 5. 2) Proper interpupillary distance, or the distance between eyepieces, is crucial to the comfort of the user. Adjusting the interpupillary distance is accomplished through a “folding” action of the optical head, at Figure 4, allowing for quick and easy adjustment. Mechanical Stage Control, see figure 5. The coaxial transverse / longitudinal knob allow for easy adjustment. Condenser Components – are located below the stage as follows, see figure 5. Abbe condenser; its mounting ring with set screw and two large centering screws at either side for adjusting condenser/iris assembly in the optical center of the light path; the substage adjustment knob for raising and lowering the condenser. Upon the collector is field diaphragm to hold the filters. Note: Using the 5X objective can be 9 1 2 3 Table of Contents Getting Started ................................................................................................... 3 Components ....................................................................................................... 4 Technical Specifications..................................................................................... 6 Setup Instructions............................................................................................... 7 Basic Operation.................................................................................................. 7 Preventative Maintenance ................................................................................ 11 Figure 4 4 5 6 7 Figure 5 1.Substage Adjustment Knob 2.Slide Holder 3. Condenser Set Screw 4. Condenser Centering Screw 5.Iris Diaphragm Control Lever 6.Longitudinal Adjustment Knob Knob 7.Transverse Adjustment 2 3) Reflection Illumination Press the transmitted / reflection illumination Getting Started This reference guide is written base on whole series, it covers some capable optional attachments and functions, you may read the similar operation what you need. toggle switch on the side of “ Ⅱ ”, which means the reflection illumination is used. 2. Focusing Controls, see figure 3 1) Focusing adjustment is accomplished by using the large coarse adjustment knobs located comfortably on each side of the frame. Fine adjustment is accomplished using the smaller knobs located on the same focus shaft. This coaxial arrangement allows for easy, precise adjustment without drift or discomfort. 2) Focus Control Turning either of the coarse focus control knobs will raise or lower the stage. The smallest graduation on the fine adjust knob index scale is 2μm of vertical. 3) Focus Tension Adjustment The tension of the coarse focus is adjustable and preset at the factory for ease of use. If you wish to adjust the coarse focus tension, first locate the tension adjustment ring. Turning the ring toward the front of the microscope increases the tension, and toward the rear of the microscope loosens it. Tension is too high if you experience physical discomfort. 4) Pre-focusing or Focus Stop Control. Use of this feature will insure that the short working distance objectives don’t contact the stage or slide glass when using the microscope. Its use also simplifies focusing. After focusing on the slide glass with the coarse adjustment, rotation of the lever toward the rear of the microscope will set an upper limit on the coarse adjustment movement. After changing slide glass or objectives, focusing is easily accomplished by rotating the coarse adjustment knob to reach the pre-focused position, then making fine adjustments with the fine adjustment knob. Focusing movement with the fine adjustment isn’t affected by using the pre-focusing lever. 3. 3 Diopter and Interpupillary Adjustments 8 Adapter: Input: 110V~240V 50/60Hz Components Output: 12V 1 2 Setup Instructions 1. Remove all parts from their packing materials and retain the packaging in the event you need to transport the product. 2. Connect the power cord to power outlet. 14 3 Basic Operation 1. Illumination controls, see figure 3 1) The external power switch and transmitted / reflection illumination toggle switch are located on the base. The brightness control is located on the chassis. The electrical system is fuse protected and the fuse holder located on the power Inlet. 2) Transmitted Illumination Press the transmitted / reflection illumination toggle switch on the side of “ Ⅰ ” , which means the 7 6 5 4 12 5 11 6 10 1 2 3 9 Figure. 3 1. Transmitted / Illumination Toggle Switch 2. Power Switch 3. 7 Reflection Brightness circuits is got through. If the light Control 4. Up Stop Knob 5. Focus does not appear to be ON, check Tensional Adjustment 6. Coarse the brightness control to see if it’s Focus Control Knob 7. Fine Focus on a sufficiently low setting. Then Control Knob adjust the brightness control until image can be observed comfortably. Note: Using the light at brightest setting reduces life span of bulb. 7 4 13 8 Figure. 1 1. Field Diaphragm Adjustment Knob 2. Push Rod To Direct Image to Eyepieces or Camera Port 3. Baffle 4. Condenser Centering Screw 5. Longitudinal Adjustment Knob 6. Transverse Adjustment Knob 7. Field Iris Diaphragm 8. External Power Switch 9. Transmitted / Reflection Illumination Toggle Switch 10. Fine Focus Control Knob 11.Coarse Focus Control Knob 12. Focus Tensional Adjustment Knob 13. Substage Adjustment Knob 16 . Push Rod To Transmitted or Reflected 4 Technical Specifications 1 2 Trinocular: Inclined 30°. Objectives: Plan achromatic objectives 3 13 12 4 5 Magnification N.A. W.D. (mm) 4X 0.10 19.75 10X 0.25 5.0 40X 0.65(Spring) 0.66 Nosepiece: 100X 1.25(Spring, Oil) 0.36 Quadruple/Quintuple nosepiece ball-bearing reversed nosepiece with positiv Eyepieces: 10X Plan wide field, focal length 25mm, Field Φ20mm. Stage: 210mm X 140mm, range of motion 75mm X 50mm. Integrated mechanical stage with vernier, low-position coaxial control knobs. 6 Focus Mechanism: Adjustable tension control to prohibit drift and adjustable up-stop to protect slides, slide glass and objective lenses; Coaxial coarse and fine adjustments with dial markings at 2μm increments. 7 Condenser: 10 N.A. 1.25 Abbe condenser with iris diaphragm, rack and pinion focusing. 9 Transmitted Illumination: 3W High brightness white LED, brightness adjustable. 8 Reflected fluorescence system: 3W blue LED (460~470nm), with blue fluorescent filter system. Brightness cannot adjustable. Figure. 2 1. Eyepiece 2. Trinocular 3. LED lamp box 4. Stage 5. Filter Seat 6. Up Stop 7. Brightness Control Knob(Transmitted Illumination) 8. Attachment Lens 9. Condenser Iris Diaphragm 10. Condenser Set Screw 11. Light guard 12. Objective 13. Nosepiece 5 6
© Copyright 2026 Paperzz