SGC 120 Tissue Culture Chamber Benefits of the - Spectro-Lab

SGC 120
Tissue Culture Chamber
The Fitotron® SGC 120 Tissue Culture
Chamber offers a future proof solution
with outstanding control performance to
cope with the changing demands of tissue
culture research.
Supplied with 5 lit, stainless steel shelves
with individually controllable lighting
intensity, the range of capabilities of the
SGC 120 is further enhanced by the
option to add an ultrasonic humidifier. This
gives a humidity range of 35%RH to 80%
RH.
The SGC 120 has been designed with the
user in mind and as such has a host of
features that reflect this, such as the Colour
Touch Panel for programming, digital safety
thermostat, and eye level controls.
Benefits of the SGC 120
• Supplied ready to use- no assembly required
• Shelving area of 3.4m2
• Fully adjustable lamp trays
• Interchangeable lamps including UV lamps
• Shelf lighting individually controllable
• Dimmable lighting intensity from 100% to 10%
• 3.5” Colour Touch Panel for programming
• Minimal footprint size, using less laboratory space
• Optional low energy ultrasonic humiditity generator available up to 80% relative humidity
• User friendly, easy to use eye level controls
SGC 120 Specifications
Chamber size
Internal External Volume 1310mm (w) x 675mm (d) x 1410mm (h)
1440mm (w) x 810mm (d) x 1975mm (h)
1200 litres nominal
Temperature range -2°C to +40°C lights off
+5°C to +40°C lights on
Temperature stability +/-0.3°C
Humidity (optional) 35% - 80% RH – temperature dependent
Low energy ultrasonic humidity generator & capacitive humidity sensor
Shelving & Lighting Configuration
Five shelves:
2 x 36W fluorescent tubes per shelf
Max intensity approx 100 mmol m-2 s-1@ 150mm below tube @ 25°C
Max height between shelves 200mm
Total shelf area 3.4m2
Safety
User settable hi/low temperature alarms with remote contact
Audible & visual alarm indicators
Construction
Interior:
Corrosion resistant stainless steel
Exterior:
Painted, protected steel with a textured finish
Mobility:
Lockable castors enable the chamber to be moved
Service Requirements 230V, 50Hz, 13 amp max.
Water Requirements
De-mineralised water (humidity option only)
Optional CO2 Control Available
Please note: All figures subject to change
Weiss Gallenkamp Ltd
Units 37/38 The Technology Centre
Epinal Way, Loughborough,
Leicestershire LE11 3GE
Tel 01509 631590
Fax 01509 211133
www.weiss-gallenkamp.com
[email protected]