EZDILATE ENDOSCOPIC BALLOON DILATOR

EZDILATE ENDOSCOPIC BALLOON DILATOR
Quick Reference Guide
Inflation Device Preparation
Ensure inflation device plunger is in the unlocked position by
pushing the locking mechanism to the left.
14713
14670
14698
14669
1
Fill the inflator with 40 ml of water or contrast. Make sure the
OFF knob of the stopcock is in the three o’clock position.
Note: It is important that you do not overfill the inflation device
to ensure that an adequate vacuum is created to deflate the
balloon after dilation.
Remove air from the inflator.
1)Turn the inflator so that the pressure gauge is facing the
ceiling, and any air rises to the top of the inflator.
2)Push the plunger until a steady stream of contrast comes
out of the stopcock. Be sure to remove all air bubbles from
the syringe.
Remove Air from the Balloon
2
2
Connect the balloon catheter port to the stopcock by twisting
the two together.
1) Turn the inflator upside down so that the pressure gauge is
pointed towards the ground.
Pull the plunger back as far as possible
2) Thereafter, lock it by pushing the locking mechanism to the
center position.
14715
14705
14693
2
1) Turn the inflator so that the pressure gauge is facing the
ceiling, allowing any air to rise to the top of the inflator.
2) Turn the OFF knob of the stopcock to the 12 o’clock position.
You will hear the air being released.
3) With the air at the top of the inflator, unlock the plunger by
pushing the locking device to the left and push the plunger
up to remove any air until a steady stream of water or
contrast comes out of the stopcock.
Inflate the EZDilate Balloon
Return the OFF knob of the stopcock to the three o’clock
position and unlock the inflation device by pushing the locking
mechanism to the left.
14681
Push the plunger in until you meet resistance. Lock the plunger in
place by moving the locking mechanism to the center position.
14697
14705
14698
14669
3
Turn the plunger handle clockwise until you have reached the
atm pressure required to reach the desired balloon diameter.
Deflate the EZDilate Balloon
XXXXX
Unlock the inflation device by pushing the locking mechanism
to the left.
1) Grab the syringe and pull the plunger all the way back
(locking the mechanism to the left).
2) Then relock the device by pushing the locking mechanism to
the center position.
14708
14707
14700
14710
4
You may now let go of the device. You’ll see some air bubbles
coming. The generated vacuum supports rapid deflation time of
the balloon.
ORDERING INFORMATION
EZDilate Fixed Wire Balloon Dilator, Single-Use – Esophageal (1/Box)
For use in adult populations only.
Article Name
Article Number
Inflated Outer
Ø (mm)
Dilation Balloons
Inflated Outer Ø (Fr)
Inflation
Min. Working Channel
Pressure (atm)
Ø (mm)
Balloon Length (mm)
Working
Length (mm)
EZDilate 6-7-8 fixed wire
EGBD-400P-0880
6.0-7.0-8.0
18-21-24
2.0-5.5-9.5
2.8
90
1900
EZDilate 8.5-9.5-10.5 fixed wire
EGBD-400P-1080
8.5-9.5-10.5
25.5-28.5-31.5
2.0-5.5-9.5
2.8
90
1900
EZDilate 11-12-13 fixed wire
EGBD-400P-1380
11.0-12.0-13.0
33-36-39
2.0-3.5-5.5
2.8
80
1900
EZDilate 13.5-14.5-15.5 fixed wire
EGBD-400P-1580
13.5-14.5-15.5
40.5-43.5-46.5
2.5-4.0-6.0
2.8
80
1900
EZDilate 16-17-18 fixed wire
EGBD-400P-1880
16-17-18
48-51-54
2.5-4.0-5.5
2.8
80
1900
EZDilate 18-19-20 fixed wire
EGBD-400P-2080
18-19-20
54-57-60
2.5-3.5-5.0
2.8
80
1900
EZDilate Wire-Guided Balloon Dilator, Single-Use – Esophageal, Pyloric, Colonic, Biliary (1/Box)
For use in adult populations only.
Dilation Balloons
Guide Wire Ø
(Preloaded &
Compatible)
Inflated Outer Ø
Inflated Outer Ø
Inflation
Min. Working
Balloon
Working
(mm)
(Fr)
Pressure (atm)
Channel Ø (mm)
Length (mm)
Length (mm)
EGBD-410X-0855
6.0-7.0-8.0
18-21-24
2.0-5.5-9.5
2.8
65
2400
.035”
EZDilate 8.5-9.5-10.5 wire-guided
EGBD-410X-1055
8.5-9.5-10.5
25.5-28.5-31.5
2.0-5.5-9.5
2.8
65
2400
.035”
EZDilate 11-12-13 wire-guided
EGBD-410X-1355
11.0-12.0-13.0
33-36-39
2.0-3.5-5.5
2.8
55
2400
.035”
EZDilate 13.5-14.5-15.5 wire-guided
EGBD-410X-1555
13.5-14.5-15.5
40.5-43.5-46.5
2.5-4.0-6.0
2.8
55
2400
.035”
EZDilate 16-17-18 wire-guided
EGBD-410X-1855
16-17-18
48-51-54
2.5-4.0-5.5
2.8
55
2400
.035”
EZDilate 18-19-20 wire-guided
EGBD-410X-2055*
18-19-20
54-57-60
2.5-3.5-5.0
2.8
55
2400
.035”
Article Name
Article Number
EZDilate 6-7-8 wire-guided
* Esophageal, colonic and biliary application only.
Inflation Device For EZDilate Endoscopic Balloon Dilators, Single-Use – (1/Box)**
Article Name
MAJ-1740
Article Number
Syringe Volume (cc)
Max. Pressure (atm)
EGMAJ-1740
60
15
EZDilate wire-guided
14668
EZDilate inflation device
Caution:
This guide is only a summary of the instructions for use for EZDilate endoscopic balloon dilator. Be sure to follow the Olympus EZDilate endoscopic balloon dilator instruction manual that
was included in the device package.
Specifications, design, and accessories are subject to change without any notice or obligation on the part of the manufacturer.
Postbox 10 49 08, 20034 Hamburg, Germany
Wendenstrasse 14–18, 20097 Hamburg, Germany
Phone: +49 40 23773-0, Fax: +49 40 233765
www.olympus-europa.com
E0428595 · 1.000 · 10/15 · OEKG · ED
EZDilate fixed wire
14727
14727
**Optional accessory - EZDilate fixed and wire-guided balloon dilators are compatible with any 60 cc inflation device.