EZDILATE ENDOSCOPIC BALLOON DILATOR Quick Reference Guide Inflation Device Preparation Ensure inflation device plunger is in the unlocked position by pushing the locking mechanism to the left. 14713 14670 14698 14669 1 Fill the inflator with 40 ml of water or contrast. Make sure the OFF knob of the stopcock is in the three o’clock position. Note: It is important that you do not overfill the inflation device to ensure that an adequate vacuum is created to deflate the balloon after dilation. Remove air from the inflator. 1)Turn the inflator so that the pressure gauge is facing the ceiling, and any air rises to the top of the inflator. 2)Push the plunger until a steady stream of contrast comes out of the stopcock. Be sure to remove all air bubbles from the syringe. Remove Air from the Balloon 2 2 Connect the balloon catheter port to the stopcock by twisting the two together. 1) Turn the inflator upside down so that the pressure gauge is pointed towards the ground. Pull the plunger back as far as possible 2) Thereafter, lock it by pushing the locking mechanism to the center position. 14715 14705 14693 2 1) Turn the inflator so that the pressure gauge is facing the ceiling, allowing any air to rise to the top of the inflator. 2) Turn the OFF knob of the stopcock to the 12 o’clock position. You will hear the air being released. 3) With the air at the top of the inflator, unlock the plunger by pushing the locking device to the left and push the plunger up to remove any air until a steady stream of water or contrast comes out of the stopcock. Inflate the EZDilate Balloon Return the OFF knob of the stopcock to the three o’clock position and unlock the inflation device by pushing the locking mechanism to the left. 14681 Push the plunger in until you meet resistance. Lock the plunger in place by moving the locking mechanism to the center position. 14697 14705 14698 14669 3 Turn the plunger handle clockwise until you have reached the atm pressure required to reach the desired balloon diameter. Deflate the EZDilate Balloon XXXXX Unlock the inflation device by pushing the locking mechanism to the left. 1) Grab the syringe and pull the plunger all the way back (locking the mechanism to the left). 2) Then relock the device by pushing the locking mechanism to the center position. 14708 14707 14700 14710 4 You may now let go of the device. You’ll see some air bubbles coming. The generated vacuum supports rapid deflation time of the balloon. ORDERING INFORMATION EZDilate Fixed Wire Balloon Dilator, Single-Use – Esophageal (1/Box) For use in adult populations only. Article Name Article Number Inflated Outer Ø (mm) Dilation Balloons Inflated Outer Ø (Fr) Inflation Min. Working Channel Pressure (atm) Ø (mm) Balloon Length (mm) Working Length (mm) EZDilate 6-7-8 fixed wire EGBD-400P-0880 6.0-7.0-8.0 18-21-24 2.0-5.5-9.5 2.8 90 1900 EZDilate 8.5-9.5-10.5 fixed wire EGBD-400P-1080 8.5-9.5-10.5 25.5-28.5-31.5 2.0-5.5-9.5 2.8 90 1900 EZDilate 11-12-13 fixed wire EGBD-400P-1380 11.0-12.0-13.0 33-36-39 2.0-3.5-5.5 2.8 80 1900 EZDilate 13.5-14.5-15.5 fixed wire EGBD-400P-1580 13.5-14.5-15.5 40.5-43.5-46.5 2.5-4.0-6.0 2.8 80 1900 EZDilate 16-17-18 fixed wire EGBD-400P-1880 16-17-18 48-51-54 2.5-4.0-5.5 2.8 80 1900 EZDilate 18-19-20 fixed wire EGBD-400P-2080 18-19-20 54-57-60 2.5-3.5-5.0 2.8 80 1900 EZDilate Wire-Guided Balloon Dilator, Single-Use – Esophageal, Pyloric, Colonic, Biliary (1/Box) For use in adult populations only. Dilation Balloons Guide Wire Ø (Preloaded & Compatible) Inflated Outer Ø Inflated Outer Ø Inflation Min. Working Balloon Working (mm) (Fr) Pressure (atm) Channel Ø (mm) Length (mm) Length (mm) EGBD-410X-0855 6.0-7.0-8.0 18-21-24 2.0-5.5-9.5 2.8 65 2400 .035” EZDilate 8.5-9.5-10.5 wire-guided EGBD-410X-1055 8.5-9.5-10.5 25.5-28.5-31.5 2.0-5.5-9.5 2.8 65 2400 .035” EZDilate 11-12-13 wire-guided EGBD-410X-1355 11.0-12.0-13.0 33-36-39 2.0-3.5-5.5 2.8 55 2400 .035” EZDilate 13.5-14.5-15.5 wire-guided EGBD-410X-1555 13.5-14.5-15.5 40.5-43.5-46.5 2.5-4.0-6.0 2.8 55 2400 .035” EZDilate 16-17-18 wire-guided EGBD-410X-1855 16-17-18 48-51-54 2.5-4.0-5.5 2.8 55 2400 .035” EZDilate 18-19-20 wire-guided EGBD-410X-2055* 18-19-20 54-57-60 2.5-3.5-5.0 2.8 55 2400 .035” Article Name Article Number EZDilate 6-7-8 wire-guided * Esophageal, colonic and biliary application only. Inflation Device For EZDilate Endoscopic Balloon Dilators, Single-Use – (1/Box)** Article Name MAJ-1740 Article Number Syringe Volume (cc) Max. Pressure (atm) EGMAJ-1740 60 15 EZDilate wire-guided 14668 EZDilate inflation device Caution: This guide is only a summary of the instructions for use for EZDilate endoscopic balloon dilator. Be sure to follow the Olympus EZDilate endoscopic balloon dilator instruction manual that was included in the device package. Specifications, design, and accessories are subject to change without any notice or obligation on the part of the manufacturer. Postbox 10 49 08, 20034 Hamburg, Germany Wendenstrasse 14–18, 20097 Hamburg, Germany Phone: +49 40 23773-0, Fax: +49 40 233765 www.olympus-europa.com E0428595 · 1.000 · 10/15 · OEKG · ED EZDilate fixed wire 14727 14727 **Optional accessory - EZDilate fixed and wire-guided balloon dilators are compatible with any 60 cc inflation device.
© Copyright 2026 Paperzz