Thermal Analysis & Rheology THERMAL SOLUTIONS Thermal Stability Determination of Bonded Silicas for use in Packing Columns by Thermogravimetric Analysis PROBLEM Silica materials used as chromatography support material gave inconsistent performance when packed in columns. This inconsistency was believed to be associated with poor bonding between the silica support and the active phase such that leaching occurred. Hence, a quick test which gave an indication of the stability of the bonding between the silica and active phase was desired. SOLUTION Thermogravimetric analysis (TGA), which is a thermal analysis technique used to measure weight changes in a material Figure 1. with temperature, can provide information about the relative thermal stabilities of these silica packing materials which in turn provides an indication of the bond strength between the silica and active phase. Figure 1, for example, shows the comparative TGA results for a good and bad silica material. The good material (silica A), as expected, exhibited a smaller total weight loss over the temperature range evaluated. In both materials, the initial weight loss (Figures 2 and 3) represented the loss of surface moisture, while the second weight loss corresponded to loss of the active phase. Total analysis time was about an hour, but with faster heating rates or the use of an autosampler, the productivity of this TGA testing could easily be improved to levels acceptable for QC analysis. COMPARISON OF SILICA MATERIALS 102 100 Weight (%) 98 96 SILICA A 94 92 SILICA B 90 88 0 100 200 300 400 500 Temperature (oC) 600 700 800 TS-7 Figure 2. WEIGHT LOSS - SILICA B 102 100 LOSS (1) 96 4.825 % (0.7996 mg) 0.10 LOSS (2) = 5.772 % (0.9566 mg) 0.08 0.06 370.97oC 94 Residue: 89.35% (14.81 mg) 92 90 88 0.00 N2 PURGE - 100 mL/MIN 0 200 400 Temperature (oC) Figure 3. -0.02 800 600 WEIGHT LOSS - SILICA A 0.05 100 Weight (%) LOSS 0.02 (2) 10oC/MIN HEAT RATE 101 LOSS (1) = 2.159 % (0.4223 mg) LOSS (2) = 1.169 % (0.9667 mg) LOSS (1) 99 0.04 0.03 Residue: 96.67 % (18.91 mg) 98 0.02 0.01 97 96 0.04 LOSS (2) HEAT RATE = 10oC/MIN N2 PURGE - 100 mL/MIN 0 200 400 Temperature (oC) 600 [- - - - -] Deriv. Weight (%/oC) Weight (%) 98 LOSS (1) = [- - - - -] Deriv. Weight (%/oC) 0.12 0.00 -0.01 800 Acknowledgment: TA Instruments would like to thank Phase Separation Ltd. for their cooperation in writing this application note. For more information or to place an order, contact: TA Instruments, Inc., 109 Lukens Drive, New Castle, DE 19720, Telephone: (302) 427-4000, Fax: (302) 427-4001 TA Instruments S.A.R.L., Paris, France, Telephone: 33-01-30489460, Fax: 33-01-30489451 TA Instruments N.V./S.A., Gent, Belgium, Telephone: 32-9-220-79-89, Fax: 32-9-220-83-21 TA Instruments GmbH, Alzenau, Germany, Telephone: 49-6023-30044, Fax: 49-6023-30823 TA Instruments, Ltd., Leatherhead, England, Telephone: 44-1-372-360363, Fax: 44-1-372-360135 TA Instruments Japan K.K., Tokyo, Japan, Telephone: 813-5434-2771, Fax: 813-5434-2770 Internet: http://www.tainst.com TS-7A Thermal Analysis & Rheology A SUBSIDIARY OF WATERS CORPORATION
© Copyright 2026 Paperzz