Operating Instructions MAGNETIC FLOW/ COIL TEST PIECE MAGNETIC PARTICLE INSPECTION EQUIPMENT Product Description The Magnaflux magnetic flow/coil test piece contains artificial defects that simulate discontinuities when verifying MPI system performance. Part number 026T002 RRP58004, TP4 5 holes, 200 x 25 x 25 mm Operating Instructions Initially carry out the following procedure when the unit has just been calibrated, and fresh ink is in the tank: 1. 2. 3. 4. 5. 6. Ensure that the test piece is thoroughly degreased and free from corrosion and oxidation Clamp the test piece between the contacts on the head and tailstock of the bench unit for magnetic flow or position within the aperture of the coil Apply magnetic ink while the current is flowing Establish the current required to make the hole nearest the outer surface visible on the outer surface Further increase the current to establish indications from the other holes on the outer surface of the test piece Note the current required to establish each hole indication. If the magnetising system is functioning correctly each hole should be visible at each subsequent check with the test piece. Magnaflux (A Division of ITW Ltd), Faraday Road, South Dorcan Industrial Estate, Swindon, Wiltshire, SN3 5HE, UK Tel: +44 (0)1793 524566 Fax: +44 (0)1793 490459 Email: [email protected] www.eu.magnaflux.com Issue 3, 09/04/2014 Operating Instructions Test Pieces Part No: 026T002 page 1
© Copyright 2024 Paperzz