Analysis of whole lipid extracts using on-line high resolution LC-MS Catharina Crone1, Eric Genin2 and Helmut Muenster1 1Thermo Fisher Scientific, Bremen, Germany, 2Thermo Fisher Scientific, Les Ulis, France Purpose: Analysis of a complex lipid extract with high resolution and high mass accuracy including fast polarity switching and fragmentation. FIGURE 2. Base peak chromatogram of the lipid extract in positive ionization mode. FIGURE 3. Base peak chromatogram of the lipid extract in negative ionization mode. FIGURE 5. Positive HCD MS/MS spectrum of the PEs at RT 5.3 minutes FIGURE 6. Negative HCD MS/MS spectrum of the PEs at RT 5.3 minutes. Inset shows the fatty acid region with assignment of the fatty acids chain length and degree of saturation. Methods: On-line LC-MS using a stand-alone OrbitrapTM. 9.83 100 10.05 All on-line LC-MS Experiments were performed on an ExactiveTM mass spectrometer (Thermo Fisher Scientific, Bremen, Germany) equipped with an AccelaTM HPLC system (Thermo Fisher Scientific, San Jose, CA) using a 5 um 150x2.1 mm Hypersil Silica Column (Thermo Fisher Scientific, USA) at a flow rate of 300 ul/min. Solvents: A – CHCl3 / MeOH / 80:20, 5 mM NH4OAc, B – CHCl3 / MeOH / H2O 60:34:6, 5 mM NH4OAc. Gradient: 0-2 min 20% B, 2–10 min 20% B – 100% B,10–18 min 100% B,18–19 min 100% B – 20% B, 19–27 min 20% B. The mass spectrometer was operated with standard electrospray ionization performing alternating full scan and CID fragment ion scans carried out in the HCD collision cell (see Figure 1). Resolution settings of 100,000 were applied in both polarity modes. For the fast polarity switching experiments a resolution setting of 25,000 was used to achieve a full cycle (one positive full scan and one negative full scan) in less than one second. FIGURE 1. Schematic layout of the instrument C-Trap 40 2.34 SM 6.25 PE 1.91 20 9.44 C22H48N 3.26 LPC 11.87 11.48 6.97 5.30 0 1 2 3 4 5 7 8 9 Time (min) 10 11 12 13 14 10.11 PI 40 4.03 PE 5.36 7.67 11.97 0 15 0 1 2 3 4 5 6 7 8 9 Time (min) 10 11 12 13 14 15 Therefore, resolution settings of up to 100,000 were used in both polarity modes to ensure the best possible mass spectrometric separation for these complex extracts. Mass accuracies in the low to sub ppm range were obtained without averaging data and were applied to identify and confirm the different lipids in each scan. Due to the complexity of the extracts with many isobaric compounds amongst the different lipid classes it is essential to obtain fragmentation information in addition to the full scan accurate mass data. This allows confirmation of the identity of the different lipid classes and furthermore to determine the fatty acid pattern within these lipid classes. The high mass accuracy fragmentation data inherent to HCD offers the ability to generate pseudo parent ion mass chromatograms which are ideally suited to confirm the specific lipid classes. In combination with high resolution (key for separation of interferences) the applied narrow mass tolerance window (as a result of stable accurate masses) ensures correct identification of the lipid classes even for very complex samples. Examples for the benefits of these instrumental features are shown using complex natural lipid extracts. Split Lens Mass Flatapole Lens RF Lens Skimmer Tubelens 792.5536 100 Heated Capillary 80 API Ion Source Results As can be seen from the total ion chromatograms generated in both ionization mode polarities (figures 2 and 3), the chromatographic conditions employed allow good separation of the different lipid classes but not complete separation of the different lipid species within one lipid class. Therefore, resolution settings of up to 100,000 were used in both polarity modes to ensure the best possible mass spectrometric separation for these complex extracts. Mass accuracies in the low to sub ppm range were obtained without averaging data and were applied to identify and confirm the different lipids in each scan. Due to the complexity of the extracts with many isobaric compounds amongst the different lipid classes it is essential to obtain fragmentation information in addition to the full scan accurate mass data. This allows confirmation of the identity of the different lipid classes and furthermore to determine the fatty acid pattern within these lipid classes. Relative Abundance Orbitrap Mass Analyzer O 651.5340 C 43 H 71O 4 -1.03 ppm 792.5530 C 45H79O8 NP -0.99 ppm 18:0 0 350 400 450 500 550 m/z 600 650 700 750 40 16:0 303.2335 C20H31O2 1.86 ppm 22:4 283.2646 C 18 H35O 2 1.34 ppm 331.2648 C22H35O 2 1.83 ppm 255.2332 C16 H31 O 2 0.93 ppm 332.2682 317.0679 303.2335 0 260 270 280 290 300 m/z 310 320 330 40 0 C40 H75 O10 N P 0.60 C40 H79 O10 N P 90 786.5289 -0.22 C42 H77 O10 N P 788.5442 -0.58 C42 H79 O10 N P 790.5599 -0.56 C42 H81 O10 N P 70 792.5758 -0.21 C42 H83 O10 N P 65 794.5920 0.40 C42 H85 O10 N P 60 806.4986 1.03 C44 H73 O10 N P 808.5142 0.94 C44 H75 O10 N P 810.5296 0.63 812.5447 0.05 C44 H79 O10 N P 35 814.5597 -0.86 C44 H81 O10 N P 30 818.5924 0.91 C44 H85 O10 N P 820.6062 -1.33 C44 H87 O10 N P 832.5124 -1.25 C46 H75 O10 N P 834.5293 0.30 C46 H77 O10 N P 836.5455 0.92 C46 H79 O10 N P 838.5608 0.55 C46 H81 O10 N P 842.5929 1.48 C46 H85 O10 N P 85 80 75 60 776.5591 40 748.5280 20 0 752.5594 702.5433 718.5383 728.5592744.5544 706.5406 690 700 710 720 730 740 750 768.5543 780.5906 796.5852 764.5229 814.5358 760 770 m/z 780 790 800 810 820 830 700.5277 702.5432 704.5600 716.5222 716.5586 718.5382 724.5262 728.5588 730.5744 732.5540 740.5220 742.5389 744.5535 746.5693 748.5275 750.5429 752.5579 754.5751 756.5898 758.6060 764.5223 766.5385 768.5533 772.5850 774.5426 774.6008 776.5586 778.5750 840 780.5898 790.5375 792.5534 796.5848 Composition Delta(ppm) C39 H75 O7 N1 P1 C39 H77 O7 N1 P1 C39 H79 O7 N1 P1 C39 H75 O8 N1 P1 C40 H79 O7 N1 P1 C39 H77 O8 N1 P1 C41 H75 O7 N1 P1 C41 H79 O7 N1 P1 C41 H81 O7 N1 P1 C40 H79 O8 N1 P1 C41 H75 O8 N1 P1 C41 H77 O8 N1 P1 C41 H79 O8 N1 P1 C41 H81 O8 N1 P1 C43 H75 O7 N1 P1 C43 H77 O7 N1 P1 C43 H79 O7 N1 P1 C43 H81 O7 N1 P1 C43 H83 O7 N1 P1 C43 H85 O7 N1 P1 C43 H75 O8 N1 P1 C43 H77 O8 N1 P1 C43 H79 O8 N1 P1 C43 H83 O8 N1 P1 C45 H77 O7 N1 P1 C43 H85 O8 N1 P1 C45 H79 O7 N1 P1 C45 H81 O7 N1 P1 C45 H83 O7 N1 P1 C45 H77 O8 N1 P1 C45 H79 O8 N1 P1 C45 H83 O8 N1 P1 0.20 -0.01 1.43 -0.37 -0.38 0.02 -1.91 -0.12 -0.16 0.24 -0.60 1.09 -0.42 -0.21 -0.05 -0.49 -1.25 0.74 -0.51 0.18 -0.19 0.50 -0.65 -0.16 -0.79 0.13 -0.35 0.56 -0.41 -0.75 -0.47 -0.30 835.5328 50 789.5488 45 281.2489 800 Composition 1.18 331.2648 200 250 300 40 774.5459 766.5410 792.5467 621.4468 255.2332 253.2175 25 431.5618 20 700.5304 350 400 450 500 550 m/z 600 650 700 750 800 850 Detailed structural analysis of the phosphatidylethanolamines (PE) Within the different lipid classes it is difficult to separate by LC the individual species. The lack of chromatographic separation is compensated for by high resolution in the mass spectrometer with the full scan and HCD fragmentation scans. We used a resolution setting of 100,000 for the alternating full scan / HCD scan experiment and performed separate runs in each polarity mode. As can be seen from the mass spectrum of the PE section of the lipid extract, a number of different PEs co-elute (figure 4). Using the accurate masses for the determination of the elemental composition and subsequently for the identification of the different PE species, we did find a number of PEs with one ether function in the molecule (figure 4, blue masses). In any case, as can be seen from the list of PEs in Figure 4, all compounds could be identified with a mass error of less than 2 ppm. Similar results are obtained from the full scan data taken in negative ionization mode (showing [M-H]-, table not shown) and ensures that all detected species within the lipid class. All HCD spectra show the same good mass accuracy as the full scan spectra (figure 5 and 6). Mass deviations of less than 2 ppm are routinely obtained using external calibration. The fragmentation pathway can be easily confirmed with the use of the accurate masses of the fragments allowing the determination of the elemental composition of the fragments in both polarity modes. 786.5301 15 FIGURE 7. Example of alternating positive and negative scan during fast polarity switching at the retention time of PE species. OSPE/SOPE 5 760.5144 760 774.5306785.4658 770 780 DOPE 790 positive scan # 348 40 DOPE 60 1.26 ppm 742.5399 C 41 H 77 O 8 N P negative scan # 349 DOPE: R’ = R’’ = C17H33 OSPE: R’ = C17H35, R’’ = C18H33 SOPE: R’ = C18H33, R’’ = C17H33 20 742 744 746 m/z 748 830 840 850 860 Fatty acid area 747.4981 C43 H72 O8 P 1.36 ppm 419.2576 C21 H40 O6 P 1.79 ppm 80 0 40 0 820 16:0 -C3H5O2N (serine) 750 In order to obtain all the accurate mass information of lipids ionizing in the positive as well as in the negative ion mode, the instrument was operated in the fast polarity switching mode. Two consecutive scans show the [M+H]+ and [M-H]- ion respectively for the different fatty acid containing phosphatidylethanolamines (figure 7). One full cycle (one positive and one negative scan) was acquired in less than one second at 25,000 resolution. Mass accuracy was well below 3 ppm for all ions. 834.5299 C46 H77 O10 NP 0.96 ppm 18:0 283.2647 C18 H35 O2 1.48 ppm 437.2680 C21 H42 O7 P 1.38 ppm 480.3100 327.2336 C23 H47 O7 NP 22 H31 O2 18:1 C1.89 0.97 ppm ppm 417.2421 20:4 375.2064 557.0342 P 1.31 ppm 744.5555 C 41 H 79 O 8 N P 80 810 m/z 255.2332 C16 H31 O2 0.93 ppm 20 OSPE/SOPE 100 800 FIGURE 9. Negative HCD MS/MS spectrum of the lipid extract recorded during the elution of the PSs. 20 0 806.4984 As an example for a different lipid class the negative full scan spectrum (figure 8) and the HCD spectrum (figure 9) are shown for the phosphatityl-serines (PS). Similar to the other lipid classes, the full scan data for the PSs are used to determine the accurate molecular weights and subsequently their elemental compositions (figure8). In the HCD scan significant fragments are seen for the loss of the serine group and for the fatty acid residues. This allows the determination of the lipid group and the distribution of the fatty acids in the PS molecules (figure 9). 40 -0.13 ppm 744.5536 C 41 H 79 O 8 N P 60 791.5548 C44 H77 O10 N P 60 -0.15 ppm 746.5693 C 41 H 81 O 8 N P 80 810.5292 836.5346 816.5769 838.5600 839.5630 817.5802 832.5139 856.5104 846.6247 790.5519 10 100 100 834.5295 55 790.5404 20 300 60 20 60 60 40 80 NH2 18:1 18:0 Delta (ppm) 764.5452 788.5454 100 0 Lens Curved Lens System P OH 14.16 FIGURE 4. Positive full scan mass spectrum of the PE section taken at RT 5.3 min and list of identified PEs obtained from the positive full scan experiment. Octapole O 20 8.42 14.16 6 60 20 O 22:6 80 2.90 80 m/z 760.5143 95 22:6 20:4 Relative Abundance Methods 60 100 100 Relative Abundance The analysis of lipids is challenging to mass spectrometry due to its complexity and variety of lipid classes. Electrospray ionization both in positive and in negative ionization mode has been shown to be a useful method for structural studies of lipids and especially for phospholipids because of their zwitterionic structure. Information can be obtained regarding the molecular weight, the acid moieties and the residue attached to the phosphoric acid. Some lipid classes like the phosphatidylinositols (PI) are best analyzed in negative ionization mode where most other lipid classes are best ionized in positive mode of operation. Nevertheless, both ionization modes deliver complementary information and therefore, to save time, it is beneficial to run the mass spectrometer in the alternated mode of operation. The non hybrid benchtop orbitrap mass spectrometer is capable of providing a full cycle (one positive and one negative high resolution full scan) in less than one second while maintaining high mass accuracy. In addition the instrument can perform CID experiments in the HCD collision cell delivering accurate fragment mass information which is of high value for structural analysis. Relative Abundance Introduction Relative Abundance 80 PS 1.92 80 385.2735 C 25 H 37O 3 -0.67 ppm Relative Abundance PC SM:7G 327.2335 Relative Abundance RT:0.00 - 15.72 Relative Abundance Results: Accurate mass data of different lipid species within the different lipid classes ionizing in different polarity modes could be obtained in one single LC-MS run. HCD Collision Cell 327.2335 C22H31O 2 1.84 ppm 100 RT:0.00 - 15.25 SM: 7G 2.32 100 FIGURE 8. Negative full scan spectrum of the lipid extract recorded during the elution of the PSs together with an accurate mass list of all detected PS species. Relative Abundance Overview 250 300 621.4589 C44 H62 P 701.5142 -0.91 ppm C46 H69 O5 -1.22 ppm 22:6 350 400 450 500 550 m/z 792.5774 C42 H83 O10 NP 1.72 ppm 836.5359 629.4539 600 650 700 750 800 850 Conclusions The ExactiveTM is an ideal instrument for the analysis of complex lipid extracts. The ability to perform full scans and MS/MS scans in both polarity modes at very high resolution ensuring accurate masse measurements independent of the polarity allows fast, precise and unambiguous identification of the lipid classes as well as the individual lipid species. Operation of the instrument in the alternated mode of operation saves time without sacrificing analytical performance. Acknowledgements We would like to thank Anne ATHIAS, IFR 100 Santé-STIC/Plateau Technique Lipidomique, Dijon, France and Arnaud BERNARD, LBMN - INSERM U866, Dijon, France for supplying the samples. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. IMSC09_W391_CCrone
© Copyright 2026 Paperzz