Journal of Human Hypertension (2000) 14, 423–427 2000 Macmillan Publishers Ltd All rights reserved 0950-9240/00 $15.00 www.nature.com/jhh ORIGINAL ARTICLE Accuracy of the OMRON M4 automatic blood pressure measuring device JE Naschitz1, L Loewenstein1, R Lewis1, D Keren1, L Gaitini2, A Tamir3 and D Yeshurun1 Departments of 1Internal Medicine A, 2Anesthesiology and 3Epidemiology, Bnai Zion Medical Center and the Bruce Rappaport Faculty of Medicine, Technion Israel Institute of Technology, Haifa, Israel Despite widespread use of the automatic blood pressure (BP) measuring device OMRON M4, there is little formal validation on its accuracy. A study was conducted to assess the accuracy of the OMRON M4 compared with the true indirect BP measured by mercury sphygmomanometer (MS). A rapid method for the evaluation of automatic blood pressure measurement devices (READ) was applied for this study. The READ is based on numerous BP measurements at rest and during a standardised postural challenge in a small number of subjects who exhibit a wide range of BPs. Blood pressure measurements were done in supine position for 10 min followed by head-up tilt for 30 min and again supine for 10 min. The automatic device (AU) and the MS were connected to one arm-cuff by means of a T connector. A stethoscope with dual sets of ear-pieces was used for duplicate MS measurements (MS1 and MS2). The MS1, MS2 and AU measurements were taken simultaneously in a blinded manner. Three units of the automatic instrument were evaluated. An average of 111 measurements per unit were performed, every BP category being present in ⭓15 MS measurements. The differences between MS1 and MS2 measurements (⌬BP: MS1–MS2) were utilized to assess the consistency of true indirect BP and the differences between AU and MS measurements (⌬BP:AU-MS2) were utilized to assess the accuracy of the AU. The following characteristics of the OMRON M4 were assessed: (1) grade of accuracy, (2) aberration pattern, (3) consistency of the aberrationpattern, and (4) correlation between levels of BP and ⌬BP: AU-MS. For MS paired readings, 92–100% of systolic and 99–100% of diastolic readings fell within 5 mm Hg difference range, that is consistent with a British Hypertension Society grade A of both. For AU compared to MS2, 29–64% of systolic and 73–94% of diastolic readings fell within 5 mm Hg and 49–86% of systolic and 86–99% of diastolic readings fell within 10 mm Hg and the devices qualified C, D and C, respectively. All devices exhibited irregular and inconsistent aberration patterns, making the design of correction formulas impractical. In conclusion, the OMRON M4 device did not meet the requirements of the British Hypertension Society and, therefore, cannot be recommended for clinical use. Journal of Human Hypertension (2000) 14, 423–427 Keywords: validation; blood pressure measurement; tilt test Introduction The diagnosis and management of hypertension is traditionally based on casual blood pressure (BP) readings taken at the physician’s office. Although casual BP is easy to obtain, several studies have shown that patients with an elevated BP in the office may have lower BP if measured by the patients themselves at home.1,2 A greater number of readings can be obtained leading to a greater accuracy3 and the absence of the white-coat effect.4 Also, these measurements are free of observer bias.5 It would seem appropriate to encourage the widespread use of self-recorded BP as an important adjunct to the clinical care of the patient with hypertension.6,7 Particular attention must be paid to ensure the accuracy of the devices used.8 Ideally, rec- Correspondence: Dr Jochanan E Naschitz, Department of Internal Medicine A, Bnai Zion Medical Center, P.O. Box 4940, 31048 Haifa, Israel Received 19 October 1999; revised and accepted 3 February 2000 ommended devices should have been subjected to standard validation procedures. Validation of automatic BP measuring devices (AU) is a relatively new field of research. Most validation procedures assess the accuracy of the test device compared with the true indirect BP measured by mercury sphygmomanometer (MS). At present validation of automatic BP measurement devices is very cumbersome, time-consuming and expensive because a large number of subjects, a wide range of BPs, and many well trained observers are required for validation.9–11 We proposed a rapid method for the evaluation of automatic blood pressure measurement devices (READ).12 The READ is based on numerous BP measurements at rest and during a standardised postural challenge in a small number of subjects who exhibit a wide range of BPs. It could serve as a pre-validation test, disclosing inadequacies in automatic BP measuring devices which may be amenable to correction by the manufacturer and thus prevent them needlessly going through cumbersome validation protocols. Second, the READ could be useful as a post-validation test to Accuracy of the OMRON M4 automatic BP measuring device JE Naschitz et al 424 examine whether or not accurate devices preserve their accuracy in field. Third, the READ could be helpful for testing automatic devices which are to be used in scientific studies. The OMRON M4 is an automatic BP measurement device particularly designed for self-monitoring of the BP. This device is relatively inexpensive. Its precision is ±3 mm Hg according to the manufacturer, yet there is little formal validation in the literature concerning its accuracy. The aim of the present study was to assess the accuracy of the OMRON M4 according to the grading system of the British Hypertension Society.10 having been utilized before. The cuff automatic inflation time is approximately 25 sec and the deflection time is 30 to 80 sec, depending on the measure of the systolic BP. Each device was calibrated before the study. READ protocol Informed consent was obtained from all patients and controls under an investigational protocol approved by the institutional review board for clinical studies. For evaluation of each of the three OMRON M4 devices, the HUTT was performed in four subjects, each subject representing a different category of the sitting BPs: optimal-normal BP, mild hypertension, moderate hypertension, and severe hypertension. The differences between MS1 and MS2 measurements (⌬BP: MS1–MS2) were determined to evaluate the consistency of the ‘true indirect BP’. The differences between AU and MS2 measurements (⌬BP: AU-MS2) were calculated to assess the parameters of accuracy of the automatic device. Patients READ parameters Adult patients of both sexes with a mean age of 51 years (range 20 to 75 years) and mean body mass index 23 (range 21 to 24.8) were assigned to the study. All patients exhibited sinus rhythm during the test. According to their sitting BPs the patients were classified to the following categories:13 optimal BP and normal BP, mild hypertension, moderate hypertension, and severe hypertension. The following parameters of each of the OMRON M4 units were determined: (1) grade of accuracy according to the British Hypertension Society grading system,10 (2) the pattern of aberration, (3) consistency of the aberration-pattern, and (4) correlation between ⌬BP: AU-MS and levels of BP. Patients and methods The head-up tilt test (HUTT) The tilt test comprises phases of supine rest and postural challenge.14 The patient lay supine on the tilt table, secured to the table at chest, hips and knees using adhesive girdles, the cuff of the BP recording device attached to the left arm which was supported at heart level at all times during the study. The automatic device (AU) and the MS were connected to one arm-cuff by means of a T connector. Stethoscopes with dual sets of ear-pieces permitted duplicate MS measurements (MS1 and MS2). The MS1, MS2 and AU measurements were taken simultaneously in a blinded manner. The MS measurements were done with a standard Baumanometer (Standby Model 0661–0250) by two physicians certified in the BP measurement technique according to the American Heart Association recommendations.15 The BP was measured initially in the supine position for 10 min. This was followed by slowly (during 30 sec) tilting the patient to 70°C head-up position. The tilt phase lasted 30 min and was followed by another supine phase of 10 min duration. The BP measurements were performed at 2-min intervals. On the average, 28 BP measurements were obtained in each patient. During the course of the tilt test an electrocardiogram was recorded continuously. Instruments tested Three OMRON M4 devices were evaluated, two of them having been in clinical use (approximately 150 and 200 measurements, each) and one device not Journal of Human Hypertension The grade of accuracy: The test devices were evaluated by calculating the percentage of readings that pertain within the ⌬BP: AU-MS2 ⭐5 mm Hg, ⭐10 mm Hg and ⭐15 mm Hg. The data for each test device were referred to the grading chart of the British Hypertension Society.10 According to the latter, a test device is acceptable for clinical practice when at least 65% of the readings fall within the 5 mm Hg, 85% within 10 mm Hg, and 95% within 15 mm Hg of the MS. In a similar way, the differences between MS1 and MS2 measurements permitted to assess the consistency of the ‘true indirect BP’ referred to the grading chart of the British Hypertension Society. Patterns of aberration: The aberration pattern of automatic measurements was evaluated by observation of the time-curves of systolic ⌬BP (AU-MS2) and diastolic ⌬BP (AU-MS2) (Figure 1). We proposed that the aberration is regular and ‘positive’ when ⬎90% of ⌬BPs (AU-M2) in a subject are positive values. The aberration is regular and ‘negative’ when ⬎90% of ⌬BPs (AU-MS2) in a subject are negative values. The 90% cut-off was proposed taking in consideration that inconsistency of a competent examiner should not exceed 10% of measurements.10,16 A ‘regular biphasic’ aberration was diagnosed when positive and negative aberrations succeed in a predictable order. An ‘irregular’ aberration was diagnosed when serial AU-MS2 measurements were irregularly positive and negative. Consistency of aberration: The consistency of the aberration pattern is assessed by observing similarity or dissimilarity of aberration patterns on several tests performed with a device. An instrument’s aberration pattern is ‘consistent’ when the aber- Accuracy of the OMRON M4 automatic BP measuring device JE Naschitz et al 425 Figure 3 Correlation between diastolic sphygmomanometric BP and ⌬BP: AU-MS2. ration is either positive or negative in each patient examined. Correlations between levels of BP and ⌬BP: AUMS2: The differences ⌬BP:AU-MS2, systolic and diastolic, were correlated with the values of MS2, systolic and diastolic, respectively. For statistical analysis the t-test and Bland-Altman plots were utilized as appropriate. Results Figure 1 Blood pressure differences on serial measurements during the tilt test recorded in a patient with mild systolic and diastolic hypertension. (a) systolic BP differences. (b) diastolic BP differences. The paired sphygmomanometric measurements MS1 and MS2 were similar, attesting the consistency of ‘true indirect BP’. The paired automatic versus manual measurements AU and MS2 showed discrepancies that often exceeded 10 mm Hg, thus demonstrating inaccuracy of automatic device. Figure 2 Correlation between systolic sphygmomanometric BP and ⌬BP: AU-MS2. An average of 27.6 simultaneous AU and MS measurements (range 24 to 36) were performed per patient, and an average of 111 measurements (range 106 to 116) were performed per device tested. For each device, every BP category of the JNC VI13 was met in at least 15 manual BP measurements. Differences between paired BP measurements are shown in Table 1, demonstrating overlap of paired MS measurements (⌬BP: MS1–MS2) and considerable discrepancy of AU and MS measurements (⌬BP: AU-MS2). Consistency of the true indirect BP was tested according to the standards of the British Hypertension Society (Table 2). For paired MS readings, 92– 100% of systolic and 99–100% of diastolic readings fell within 5 mm Hg difference range, that is consistent with a British Hypertension Society grade A of both.10 The grade of accuracy of the OMRON M4 was evaluated based on the same principle (Table 2). For AU compared to MS2, 29–64% of systolic and 73–94% of diastolic readings fell within 5 mm Hg and 49–86% of systolic and 86–99% of diastolic readings fell within 10 mm Hg. Based on both systolic and diastolic ⌬BP: AU-MS2, the automatic devices classified C, D and C, respectively, according to the British Hypertension Society grading chart. Patterns of aberration and their consistency are shown in Table 3. Formulas for correction of aberrant BP values could not be proposed for the OMRON M4 device since there was no apparent rule commanding the instrument’s deviation from true indirect BP. Weak correlation was found between ⌬BP: AU-MS2 and levels of the sphygmomanometric BP. Systolic BP differences ⌬BP: AU-MS2 correlated positively to values of sphygmomanoJournal of Human Hypertension Accuracy of the OMRON M4 automatic BP measuring device JE Naschitz et al 426 Table 1 Differences between paired BP measurements: ⌬BP: MS1-MS2 and ⌬BP: AU-MS2 ⌬BP: MS1-MS2 (mm Hg) mean (s.d.) OMRON M4 Systolic a Diastolic −1.10 (2.31) −0.28 (1.52) 0.02 (1.96) Device 1 Device 2 Device 3 ⌬BP: AU-MS2 (mm Hg) mean (s.d.) Systolic −1.12 (2.01) −0.58 (1.52) 0.19 (1.58) Diastolic −5.46 (7.01) −14.5 (12.28)a −2.44 (6.41) −3.71 (8.27)a −0.65 (5.33) 0.28 (3.03) Not satisfying requests of accuracy of the American National Standard.11 Table 2 Percentage of BP differences ⭐5 ⭐10 and ⭐15 mm Hg ⌬BP: MS1–MS2 (mm Hg) Device 1 2 3 a ⌬BP: AU-MS2 (mm Hg) BHSa ⭐5 ⭐10 ⭐15 Systolic 92% Diastolic 99% Systolic 100% Diastolic 99% Systolic 98% Diastolic 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% A A A A A A BHSa ⭐5 ⭐10 ⭐15 57% 75% 29% 73% 64% 94% 77% 86% 49% 95% 86% 99% 92% 99% 58% 99% 97% 100% C C D B C A British Hypertension Society grading system. Table 3 Patterns of aberration of the OMRON M4 Device Aberration pattern Systolic ⌬BP: AU-MS2 Consistency of aberration pattern Diastolic ⌬BP: ⌬BP ⌬BP: AU-MS2 1 Negative, negative, irregular, negative Irregular, irregular, positive, negative inconsistent 2 negative, negative, negative, negative Negative, positive, positive, negative inconsistent 3 Negative, negative, irregular, negative negative, irregular, irregular, irregular inconsistent metric systolic BP (r = 0.4256). Diastolic BP differences ⌬BP: AU-MS2 correlated negatively to values of sphygmomanometric diastolic BP (r = −0.34198) (Figures 2 and 3). Discussion Because of the demand for long-term antihypertensive treatment, the involvement of the patient in BP control is desirable. Such involvement is feasible if user-friendly and precise automatic BP measurement devices are available. Trained patients are able to provide reliable self-measured BP with automatic devices. Such measurements could be adequate for clinical and for research purposes as well.6,7 However, it should be emphasised that the accepted standard for non-invasive BP measurements is the MS. An automatic device may be a convenient surrogate for self-measurement of BP for those patients who have difficulty in achieving skill and accuracy in self-measuring BP with a MS. When using an autoJournal of Human Hypertension matic instrument, preference should be given to devices which have been authorised by a regular validation procedure. Also, the automatic devices need to be periodically re-calibrated. For research purposes, in addition to re-calibration the validity of each automatic device should be periodically reassessed. The accuracy of various models of OMRON devices has been evaluated, their precision ranging from excellent to bad.17–22 We could not find in the literature data on validation of the OMRON model M4. In the present study, the OMRON M4 device invariably underestimated the diastolic BP and unpredictably over- or underestimated the systolic BP. Each of the three devices evaluated by us qualified within category C or D of the British Hypertension Society grading system.10 A similar category of error can be understood if the automatic device is grossly inaccurate. Alternatively, the possibility that the MS—the gold standard with which automatic devices are compared23—was inaccurate has been excluded since differences between manual measurements by blinded observers were insignificant. The automatic BP measurements were inaccurate in all ranges of BP, whether normal, high-normal as well as mild, moderate and severe hypertension. Beyond establishing the degree of accuracy of an automatic BP measuring device, identification of the pattern of aberration of an instrument could be the basis for the ‘diagnosis’ of a faulty device as well as for calculating a formula for correction of the measured BP. Such formula must be tailored for each inadequate unit. The OMRON M4 units exhibited irregular aberration patterns, therefore making proposition of correction formulas impossible. For our investigation we have used a recently introduced test for the rapid evaluation of automatic Accuracy of the OMRON M4 automatic BP measuring device JE Naschitz et al BP measurement devices, the READ.12,16 Conventional procedures for validation of automatic BP measurement devices are very cumbersome, timeconsuming and expensive because a large number of subjects are tested across a wide range of BPs, and employment of many well trained observers.11 There is a need to simplify the validation of automatic BP measurement devices by enlisting fewer subjects, with a more restricted range of BPs and carrying out numerous BP measurements in each subject. For the READ, we utilized the setting of the head-up tilt test. The major purpose of tilting in the present study was to increase the range in BP values for each subject, thereby permitting to restrict the number of patients enlisted. The preliminary power analysis for READ takes advantage of the established principles utilized for validation methods.10 The total number of BP measurements for both the READ and for validation studies is comparable, and covers the spectrum of low, medium and high BP in similar proportions for both methods.10,12,24 Since discrepancies between automatic devices and the mercury standard are influenced by age,25 subjects covering a spectrum of ages from 20 to 75 years were examined in this study. In conclusion, our investigation showed that the OMRON M4 automatic BP measuring device for selfmeasurement of the BP is inaccurate and cannot be recommended for clinical use. 9 10 11 12 13 14 15 16 17 18 References 1 Mancia G, Bertinieri G, Grassi G et al. Effects of blood pressure measurement by the doctor on patient’s blood pressure and heart rate. Lancet 1983; 2: 695–698. 2 Hoegholm A, Kristensen KS, Madsen NH, Svendsen TL. White coat hypertension diagnosed by 24-hour monitoring: examination of 159 newly diagnosed hypertensive patients. Am J Hypertens 1992; 5: 64 –70. 3 Pickering T, for the American Society of Hypertension Ad Hoc Panel. Recommendations for the use of home (self) and ambulatory blood pressure monitoring. Am J Hypertens 1995; 9: 1–11. 4 Fagard R, Staessen J, Thijs L. Ambulatory blood pressure during antihypertensive therapy guided by conventional pressure. Blood Press Monit 1996; 1: 279– 281. 5 Stergiou GS, Voutsa AV, Achimastos AD, Mountokalakis TD. Home self-monitoring of blood pressure: is fully automated oscillometric technique as good as conventional stethoscopic technique? Am J Hypertens 1997; 10: 428– 433. 6 Thijs L et al. Reference values for self-recorded blood pressure. A meta-analysis of summary data. Arch Intern Med 1998; 158: 481– 488. 7 Vidt DG. Self-recorded blood pressures. Arch Intern Med 1998; 158: 425. 8 Jones D, Engelke MK, Brown ST, Swanson M. A com- 19 20 21 22 23 24 25 parison of two noninvasive methods for blood pressure measurement in the triage area. J Emerg Nurs 1996; 22: 111–115. O’Brien E, Atkins N, Staessen J. Factors influencing validation of ambulatory blood pressure measuring devices. J Hypertens 1995; 13: 1235–1240. O’Brien E et al. British Hypertension Protocol: evaluation of automatic and semi-automatic blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990; 8: 607–619. White WB et al. National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers. Hypertension 1993; 21: 504 –509. Naschitz JE et al. Rapid estimation of automatic blood pressure measuring devices (READ). J Hum Hypertens (in press). The sixth report of the Joint National Committee on Prevention, Detection, and Treatment of High Blood Pressure. NIH Publication, 1997, p 11. Naschitz JE et al. Comparison of the 10-minute supine 30-minute-tilt test with 24-hour ambulatory blood pressure monitoring for diagnosis of diastolic and systemic hypertension. Am J Cardiol 1995; 76: 366–369. Curb JD et al. Training and certification of blood pressure observers. Hypertension 1983; 5: 610–614. Naschitz JE et al. In-field validation of automatic blood pressure measuring devices. J Hum Hypertens 2000; 14: 37–42. Watson S et al. Accuracy of a new wrist cuff oscillometric blood pressure device: comparisons with intraarterial and mercury manometer measurements. Am J Hypertens 1998; 12: 1469–1474. Mengden T et al. Reliability of self-measured blood pressure values by hypertensive patients. Am J Hypertens 1998; 12: 1413–1417. Brown MA, Robinson A, Buddle ML. Accuracy of automated blood pressure recorders in pregnancy. Aust N Z J Obstet Gynecol 1998; 38: 262–265. Stergiou GS. Voutsa AV, Achimastos AD, Moutokalakis TD. Home self-monitoring of blood pressure: is fully automated oscillometric technique as good as conventional stethoscopic technique? Am J Hypertens 1997; 10: 428– 433. Bald M, Westhues R, Bonzel KE. Blood pressure monitoring at the wrist: is it reliable in children and adolescents? Z Kardiol 1996; 85 (Suppl 3P): 106–108. Walma EP et al. Accuracy of an oscillometric automatic blood pressure device: the OMRON HEM 403C. J Hum Hypertens 1995; 9: 169–174. Association for the Advancement of Medical Instrumentation. American National Standard for Electronic or Automatic Shygmomanometers. Arlington, Va: Association for Advancement of Medical Instrumentation, 1987. Polk BF, Rosner B, Feudo R, Vandenburgh M. An evaluation of the Vita-Stat automatic blood pressure measurement device. Hypertension 1980; 2: 221–227. Miller ST, Elam JT, Graney MJ, Applegate B. Discrepancies in recording systolic blood pressure of elderly persons by ambulatory blood pressure monitor. Am J Hypertens 1992; 5: 16–21. 427 Journal of Human Hypertension
© Copyright 2026 Paperzz